Topic 8 Mass Spectrometry

download Topic 8 Mass Spectrometry

of 63

Transcript of Topic 8 Mass Spectrometry

  • 7/29/2019 Topic 8 Mass Spectrometry

    1/63

    SKA6014

    ADVANCED ANALYTICAL CHEMISTRY

    TOPIC 9Mass Spectrometry 1

    Azlan Kamari, PhD

    Department of ChemistryFaculty of Science and Mathematics

    Universiti Pendidikan Sultan Idris

  • 7/29/2019 Topic 8 Mass Spectrometry

    2/63

    Mass Spectrometry

    Mass Spectrometry (a.k.a. MS or mass spec) a method

    of separating and analyzingionsby their mass-to-chargeratio

    MS does not involve a specific region of the

    electromagnetic spectrum (because it is not directly

    interested in the energies of emitted photons, electronicor vibrational transitions, nuclear spin transitions, etc)

    Ion

    abundance

    Up to m/z = 100000!m/z

    Ion

    IonIon

  • 7/29/2019 Topic 8 Mass Spectrometry

    3/63

    General Notes on Atomic and Molecular Mass

    Helpful units and conversions:

    1 amu = 1 Da = 1/12 the mass of a neutral 12C atom.

    1 kDa = 1000 amu

    Atomic weights of other elements are defined bycomparison.

    Mass-to-charge ratio (m/z): the ratio of the mass of an

    ion (m) to its charge (z)

    Molecular ion: an ion consisting of essentially the whole

    molecule

  • 7/29/2019 Topic 8 Mass Spectrometry

    4/63

    Mass Spectrometers

    A block diagram of a generic mass spectrometer:

    IonizationSource

    MassAnalyzer

    Detector

  • 7/29/2019 Topic 8 Mass Spectrometry

    5/63

    Ionization Sources

    Electron Ionization (EI)

    Chemical Ionization (CI/APCI) Photo-ionization (APPI)

    Electrospray (ESI)

    Matrix-assisted Laser Desorption (MALDI)

    Field Desorption (FD) Plasma Desorption (PD)

    Fast atom bombardment (FAB)

    High-temperature Plasma (ICP)

    Desorption

    Gas Phase

    Ionization

    SourceMass

    AnalyzerDetector

  • 7/29/2019 Topic 8 Mass Spectrometry

    6/63

    EI: Electron Ionization/Electron Impact

    The electron ionization

    (EI) source is designedto produce gaseous

    ions for analysis.

    EI, which was one of theearliest sources in wide

    use for MS, usually

    operates on vapors

    (such as those elutingfrom a GC)

    Heated Incandescent

    Tungsten/Rhenium Filament

    Accel!

    E-

    Vaporized

    Molecules

    70 eV

    IonsTo

    Mass

    Analyzer

  • 7/29/2019 Topic 8 Mass Spectrometry

    7/63

    EI: Electron Ionization/Electron Impact

    How EI works:

    Electrons are emitted froma filament made of

    tungsten, rhenium, etc

    They are accelerated by a

    potential of 70 V

    The electrons andmolecules cross (usually at

    a right angle) and collide

    The ions are primarly

    singly-charged, positive

    ions, that are extracted by asmall potential (5V) through

    a slit

  • 7/29/2019 Topic 8 Mass Spectrometry

    8/63

    EI: Electron Ionization/Electron Impact

    When electrons hit the molecules undergo

    rovibrational excitation (the mass of electrons is toosmall to really move the molecules)

    About one in a million molecules undergo the reaction:

    M+ e- M

    + + 2e-

  • 7/29/2019 Topic 8 Mass Spectrometry

    9/63

    EI: Electron Ionization/Electron Impact

    Advantages:

    Results in complex mass spectra with fragment ions,

    useful for structural identification

    Disadvantages: Can produce too much fragmentation, leading to no

    molecular ions! (makes structural identification

    difficult!)

  • 7/29/2019 Topic 8 Mass Spectrometry

    10/63

    CI: Chemical Ionization

    Chemical ionization (CI) is a form of gas-phase

    chemistry that is softer (less energetic) than EI

    Ionization via proton transfer reactions

    A gas (ex. methane, isobutane, ammonia) is introduced

    into the source at ~1 torr.

    Example: CH4 reagent gas

    CH4EI

    CH4+

    CH4+ + CH4

    CH5+ + CH3

    AH + CH5+ AH2

    + + CH4

    Strong acid

  • 7/29/2019 Topic 8 Mass Spectrometry

    11/63

  • 7/29/2019 Topic 8 Mass Spectrometry

    12/63

    APCI: Atmospheric-Pressure Chemical

    Ionization

    A form of API (atmospheric pressure ionization) arange of ionization techniques that operate at higher

    pressures, outside the vaccuum MS regions.

    APCI a form of chemical ionization using the liquideffluent in a spray chamber as the reagent

  • 7/29/2019 Topic 8 Mass Spectrometry

    13/63

    APCI: Atmospheric-Pressure Chemical

    Ionization The APCI process:

    The sample is in a flowing stream of a carrier liquid (or gas)and is nebulized at moderate temperatures.

    This stream is flowed past an ionizer which ionizes the carrier

    gas/liquid.

    63Ni beta-emitters

    Corona (electric) discharge needle at several kV

    The ionized stream (which can be an LC solvent) acts as the

    primary reactant ions, forming secondary ions with the

    analytes.

    The ions are formed at AP in this process, and are sent into the

    vaccuum

    In the vaccuum, a free-jet expansion occurs to form a Mach

    disk and strong adiabatic cooling occurs.

    Cooling promotes the stability of analyte ions (soft

    ionization)

  • 7/29/2019 Topic 8 Mass Spectrometry

    14/63

    APCI: Chemical Ionization

    APCI (diagram from Agilent)

    Diagram from Agilent Technologies

    760 torr

    10-6 torr

  • 7/29/2019 Topic 8 Mass Spectrometry

    15/63

    APCI: Chemical Ionization

    An APCI mass spectrum:

    Diagram from Agilent Technologies

  • 7/29/2019 Topic 8 Mass Spectrometry

    16/63

    Electrospray Ionization (ESI)

    The ESI process:

    Electrospray ionization (ESI) is accomplished by flowing a

    solution through an electrically-conductive capillary held at high

    voltage (several keV DC).

    The capillary faces a grid/plate held at 0 VDC.

    The solution flows out of the capillary and feels the voltage

    charges build up on nebulized droplets, which then begin to

    evaporate

    Coulombic explosions occur when the repulsion of the charges

    overcomes the surface tension of the solution (holding the drop

    together) known as the Rayleigh limit.

    Depending on whose theory you believe

    the analyte ion is eventually the only ion left

    orthe analyte ion is evaporated from a small enough droplet

  • 7/29/2019 Topic 8 Mass Spectrometry

    17/63

    Electrospray Ionization (ESI)

    A picture of two ideas for the electrospray process:

    Diagram from John B. Fenn (Nobel Lecture), 2002Picture from http://www.newobjective.com/electrospray/electrospray.html

    Noteions which are

    surface-active will be

    preferentially ionized

    this can lead to ion

    suppression!

    http://www.newobjective.com/electrospray/electrospray.htmlhttp://www.newobjective.com/electrospray/electrospray.html
  • 7/29/2019 Topic 8 Mass Spectrometry

    18/63

    Electrospray Ionization (ESI)

    An ESI source:

    Diagram from Agilent Technologies

  • 7/29/2019 Topic 8 Mass Spectrometry

    19/63

    Typical ESI Spectra

    An ESI mass spectrum:

    Diagram from Agilent Technologies

  • 7/29/2019 Topic 8 Mass Spectrometry

    20/63

    Typical ESI Spectra

    An ESI Mass Spectrum of a 14.4 kDa enzyme:

    Diagram from http://www.nd.edu/~masspec/ions.html

  • 7/29/2019 Topic 8 Mass Spectrometry

    21/63

    ESI and APCI

    ESI and APCI complementary techniques:

    Figure from Agilent Instruments

  • 7/29/2019 Topic 8 Mass Spectrometry

    22/63

    ESI and APCI

    ESI and APCIcomplementary techniques:

    ESI APCI

    Very soft ionization

    can ionize thermally

    labile samples

    Some sample volatility

    needed (nebulizer)

    Ions formed in solution Ions formed in gas

    phase

    Singly- and multiply-charged ions [M+H]+

    Singly-charged ions,[M+H]+ and [M-H]-

  • 7/29/2019 Topic 8 Mass Spectrometry

    23/63

    Atmospheric Phase Photo-ionization

    APPI can ionize things that ESI and APCI cant:

  • 7/29/2019 Topic 8 Mass Spectrometry

    24/63

    Atmospheric Phase Photo-ionization

    APPI can ionize things that ESI and APCI cant:

  • 7/29/2019 Topic 8 Mass Spectrometry

    25/63

    Comparison of Ionization Methods

    How to choose an ionization technique:

    Figure from Agilent Instruments

  • 7/29/2019 Topic 8 Mass Spectrometry

    26/63

    MALDI: Matrix-Assisted Laser

    Desorption/Ionization

    A method for desorbing

    a sample with a laser,

    while preventing thermal

    degradation

    A sample is mixed with a

    radiation-absorbing

    matrix used to help it

    ionize

    MALDI is mostly used for

    large biomolecules and

    polymers.

  • 7/29/2019 Topic 8 Mass Spectrometry

    27/63

    MALDI: Matrix Effects The role of the matrix

    Must absorb strongly at the laser wavelength

    The analyte should preferably not absorb at this wavelength

    Common matrices include nicotinic acid and many other

    organic acids

  • 7/29/2019 Topic 8 Mass Spectrometry

    28/63

    MALDI at Atmospheric Pressure

    Advantages: fast, easy and sensitive

    Disadvantages: no LC, matrix still needed

    S. Moyer and R. Cotter, Atmospheric Pressure MALDI, Anal. Chem., 74, 468A-476A (2002)

  • 7/29/2019 Topic 8 Mass Spectrometry

    29/63

    FAB: Fast Atom Bombardment

    A soft ionization technique

    Often used for polar, higher-mwt, thermally labile molecules

    (masses up to 10 kDa) that are thermally labile.

    Samples are atomized by bombardment with ~keV range Ar or

    Xe atoms.

    The atom beam is produced via an electron exchange process

    from an ion gun.

    Xee-

    Xe+ + 2e-

    Advantages:

    Rapid sample heating reduced fragmentation

    A glycerol solution matrix is often used to make it easier to

    vaporize ions

    Xe+accel

    Xe+ (high KE)

    Xe+ (high KE) + Xe Xe (high KE) + Xe+

  • 7/29/2019 Topic 8 Mass Spectrometry

    30/63

    SIMS: Secondary Ion MS

    Focused Ion Beam3He+, 16O+, 40Ar+

    Beam energy 5 to 20 keV

    Beam diameter 0.3 to 5 mm

    Beam Hits Target

    A small % of the target material is sputtered off and enters the

    gas phase as ions (usually positive)

    Advantages:

    Imaging of ions (characteristic masses) on a surface or in

    biological specimens

    Surface analysis using beam penetration depth/angle

    Can be used for bothatomicandmolecularanalysis

    Sensitive to low levels, picogram, femtogram and lower

  • 7/29/2019 Topic 8 Mass Spectrometry

    31/63

    Desorption Electrospray: DESI and DART

    Desorption-

    electrospray

    ionization(DESI)

    A new technique

    for desorbingions using

    supersonic jets

    of solvents

    (charged like inelectrospray)

  • 7/29/2019 Topic 8 Mass Spectrometry

    32/63

    Inductively Coupled Plasma (ICP)

    The inductively-coupled

    plasma serves as anatomization and

    ionization source (two-

    in-one!) for elemental

    studies.

    Photo by Steve Kvech, http://www.cee.vt.edu/program_areas/environmental/teach/smprimer/icpms/icpms.htm#Argon%20Plasma/Sample%20Ionization

    See optical electroniclecture for more details

    Solution flow rates up to:

    50-100 mL/min

  • 7/29/2019 Topic 8 Mass Spectrometry

    33/63

    Mass Analyzers - Outline

    Sector Mass Analyzers (Magnetic and Electrostatic)

    Quadrupole Analyzers

    Ion Traps

    Ion Cyclotron Resonance

    Time-of-Flight

    and many more.

    Ionization

    Source

    Mass

    AnalyzerDetector

  • 7/29/2019 Topic 8 Mass Spectrometry

    34/63

    Properties of Mass Analyzers

    Resolution (R):

    R = m/m

    m = mass difference of two adjacent resolved peaks

    (typically

    m = mass of first peak or average

    Example: R= 500 (low resolution)

    resolves m/z=50 and 50.1, and m/z=500 and 501

    Example: R= 150000 (high resolution)

    resolves m/z=50 and 50.0003, and m/z=500 and

    500.0033

  • 7/29/2019 Topic 8 Mass Spectrometry

    35/63

    Sector Mass Analyzers

    Basic Features

    A sector: a geometrical construction

    that has two arcs inside of one

    another.

    (Technically, a pie slice!)

    Types:

    Magnetic

    Electrostatic Combination (e.g. double-focusing)

  • 7/29/2019 Topic 8 Mass Spectrometry

    36/63

    Magnetic Sector Mass Analyzers

    Ion kinetic energy:

    V

    erB

    z

    m

    2

    22

    2

    2

    1

    mvzeVT

    BzeVFm

    rmvFc

    2

    mc FF

    Forces:

    Only ions with equal

    forces will pass:

    Therefore:

    Where:

    Tis kinetic energyzis charge on ion

    e is electron charge (1.60 x 10-19 C)

    B is magnetic field (T)

    v is velocity (m/s)

    Vis the accelerating voltage

    m is the mass

  • 7/29/2019 Topic 8 Mass Spectrometry

    37/63

    Electrostatic Sector Mass Analyzers

    2

    v

    reV

    z

    m

    Therefore:

    Ion kinetic energy:

    221 mvzeVT

    eVFm

    rmvFc

    2

    Mc FF

    Forces:

    Only ions with equal

    forces will pass:V

    can be varied to bring ions ofdifferent KE (and different m/z

    ratio to the exit)

  • 7/29/2019 Topic 8 Mass Spectrometry

    38/63

    Double-Focusing Sector Mass Analyzers

    If a batch of ions of equal

    m/z but with differentkinetic energies enters a

    magnetic sector

    instrument, this will result

    in a spread-out beam

    Soution: minimize

    directional and energy

    differences between ions

    of the same m/z.

    Example of a double-

    focusing MS: the Nier-

    Johnson geometry

  • 7/29/2019 Topic 8 Mass Spectrometry

    39/63

    Double-Focusing Sector Mass Analyzers

    Another design, the Mattauch-Herzog geometry

    This geometry is analogous to CCD-based opticalelectronic spectroscopy systems, while Nier-Johnson

    instruments are similar in nature to traditional scanning

    monochromator spectrometers.

  • 7/29/2019 Topic 8 Mass Spectrometry

    40/63

    Time-Of-Flight (TOF) Mass Analyzers

    The principle of Time-of-flight mass analysis:

    A batch of ions is introduced into a chamber by anpulse of accelerating current.

    This chamber has no fields, and is a drift tube

    Since the ions have the same kinetic energy, their

    velocities vary inversely with their mass during theirdrift.

    Notes:

    Typical flight times are 1-30 us Lighter ions arrive at the detector first

    2

    21 mvT

  • 7/29/2019 Topic 8 Mass Spectrometry

    41/63

    Time-Of-Flight (TOF) Mass Analyzers

    Delayed extraction anything you can do to

    tighten the KE spread will help a TOFinstrument

    m/zis mass-to-charge ratio of the ion

    Eis the extraction pulse potential (V)

    s is the length of flight tube over which E is applied

    dis the length of field free drift zone

    tis the measured time-of-flight of the ion

    zeEsmvT 221

    2

    2

    v

    eEs

    z

    m

    2

    2

    d

    t

    eEsz

    m

    Ti Of Fli ht (TOF) M A l

  • 7/29/2019 Topic 8 Mass Spectrometry

    42/63

    Time-Of-Flight (TOF) Mass Analyzers The reflectrona method of compensating for different ion KEs

    Figure from http://www.abrf.org/ABRFNews/1997/June1997/jun97lennon.html

  • 7/29/2019 Topic 8 Mass Spectrometry

    43/63

    Time-Of-Flight (TOF) Mass Analyzers

    The reflectrona method of compensating for different ion KEs

    Figure from http://www.abrf.org/ABRFNews/1997/June1997/jun97lennon.html

  • 7/29/2019 Topic 8 Mass Spectrometry

    44/63

    Quadrupole Mass Analyzers

    The quadrupole (named for its electrical structure) is one of the

    simplest and most effective mass spectrometers.

    Quadrupole Mass Analyzers

  • 7/29/2019 Topic 8 Mass Spectrometry

    45/63

    Quadrupole Mass Analyzers

    How a quadrupole works:

    Most important points:

    It is easier for an applied AC field to deflecta light ion than a heavier ion

    Conversely, it is easier for an AC field to

    stabilize a light ion

    Using this knowledge a combined AC/DC

    potential is applied to the rods. Via the DC,the ion is attracted to one set of rods and

    repelled by the other

    The DC serves to stabilize heavy ions in one

    direction (high pass filter). The AC serves to

    stabilize light ions in the other direction (lowpass filter).

    The ion must pass through the quadrupole to

    make it to the detector

  • 7/29/2019 Topic 8 Mass Spectrometry

    46/63

    Quadrupole Mass Analyzers

    Another view and the concept

    of the mass scan

    Images from http://www.jic.bbsrc.ac.uk/SERVICES/metabolomics/lcms/single1.htm

    Light ion:

    (ex. m/z = 100)

    Dragged by AC

    Heavy ion:

    (ex. m/z = 500)

    Dragged by DC

    Just right:

    Dragged by both,

    But equally balanced

    I T M A l

  • 7/29/2019 Topic 8 Mass Spectrometry

    47/63

    Ion Trap Mass Analyzers

    Ion trap: a device for

    trapping ions and confiningthem for extended periods

    using EM fields

    Used as mass analyzers because

    they can trap ions and eject themto a detector based on their mass.

    Theory is based on Mattieus work

    on 2nd order linear differential

    equations (in the 1860s), and onWolfgang Pauls Nobel Prize

    winning implementations

  • 7/29/2019 Topic 8 Mass Spectrometry

    48/63

    Ion Trap Mass Analyzers

    The stability region of an

    ion trap based ondifferential equations

    22

    0

    8

    mr

    eU

    az

    22

    0

    4

    mr

    eVqz

    )cos(0 tVU

    Most ITMS systems dont

    use DC (U), i.e. only qzis

    controlled

  • 7/29/2019 Topic 8 Mass Spectrometry

    49/63

    Ion Trap Mass Analyzers

    Layout of an ion trap mass analyzer:

    Diagram courtesy of M. Olsen, GlaxoSmithKline

    +

    Main RF

    Ring

    Endcap

    Lenses

    Octopole

    Optimized Asymptote Angle

    End Cap

    Shutter

    Focus

    Electron Multiplier

    Conversion Dinode

    Low Amplitude Dipole Field

    (1/3 frequency of main RF)

    ++

    ++

    ++

  • 7/29/2019 Topic 8 Mass Spectrometry

    50/63

    Ion Trap Mass Analyzers

    The BrukerEsquire ESI

    ITMS - a typical

    ion-trap LC-MS

    system:

    I C l t R

  • 7/29/2019 Topic 8 Mass Spectrometry

    51/63

    Ion Cyclotron Resonance

    FT-ICR:a FT-based mass spectral method that offers

    higher S/N, better sensitivity andhigh resolution

    Also contains a form of ion trap, but one in which ion

    cyclotron resonance occurs.

    When an ion travels through a strong magnetic field, itstarts circulating in a plane perpendicular to the field

    with an angular frequency c:

    m

    zeB

    r

    vc

    Ion Cyclotron Resonance

  • 7/29/2019 Topic 8 Mass Spectrometry

    52/63

    Ion Cyclotron Resonance

    How ICR works:

    The ions are circulated in a field

    An RF field is applied to match the cyclotron frequency of the ionsthis field brings them into phase coherence(forming ion packets)!

    The image current is produced as these little packets of ions get

    near the plates. The frequency of the image current is characteristic

    of the ion packets m/z ratio.

    http://www-methods.ch.cam.ac.uk/meth/ms/theory/fticr.html

    Ion Cyclotron Resonance and Magnetic Field

  • 7/29/2019 Topic 8 Mass Spectrometry

    53/63

    Ion Cyclotron Resonance and Magnetic Field

    Parallels between NMR/EPR and ICR:

    B

    B= q B

    m

    =

    B

    Picture courtesy Prof. Alan Marshall, FSU/NHMFL

  • 7/29/2019 Topic 8 Mass Spectrometry

    54/63

    The OrbitrapTM:A Hybrid Trap Between IT

    and ICR

    The Orbitrap is a recently developed electrostatic iontrap with FT/MS read-out of image current, coupled

    with MS/MS

    Advantages

    Ease of use

    Resolving power (superior to TOF)

    Precision and accuracy

    Versatility, dynamic range

    A lower-resolution, more economical ICR

    LTQ O bit h ti

  • 7/29/2019 Topic 8 Mass Spectrometry

    55/63

    LTQ Orbitrap schematic

    API Ion source Linear Ion Trap C-Trap

    Orbitrap

    Finnigan LTQ Linear Ion Trap

    Differential pumping

    Differential pumping

    Image/animation from Thermo Electron Inc. See A. Makarov et al.,Anal. Chem.2006,78, 2113-2120.

    LTQ Orbitrap Operation Principle

  • 7/29/2019 Topic 8 Mass Spectrometry

    56/63

    LTQ Orbitrap Operation Principle

    1. Ions are stored in the Linear Trap

    2. . are axially ejected

    3. . and trapped in the C-trap4. . they are squeezed into a small cloud and injected into the Orbitrap

    5. . where they are electrostatically trapped, while rotating around the central electrode

    and performing axial oscillation

    The oscillating ions induce an image current into the two

    outer halves of the orbitrap, which can be detected using

    a differential amplifier

    Ions of only one mass generate a sine

    wave signal

    Image/animation from Thermo Electron Inc. See A. Makarov et al.,Anal. Chem.2006,78, 2113-2120.

  • 7/29/2019 Topic 8 Mass Spectrometry

    57/63

    The axial oscillation frequency follows the formula

    Where = oscillation frequencyk = instrumental constant

    m/z = mass-to-charge ratio

    zm

    k

    /

    Frequencies and Masses

    Many ions in the Orbitrap generate a complex

    signal whose frequencies are determined using a

    Fourier Transformation

    Image/animation from Thermo Electron Inc. See A. Makarov et al.,Anal. Chem.2006,78, 2113-2120.

  • 7/29/2019 Topic 8 Mass Spectrometry

    58/63

    Multiple-Stage MS: MS-MS, and MSn

    Also known as Tandem MS or MSn

    Mass

    Analyzer

    Mass

    Analyzer

    Multiple quadrupoles are very common (e.g. triple-quad or

    QQQ systems, EB for double-focusing, Q-TOF for quad time-

    of-flight)

    Why tandem MS? Because of the possibility of doing CIDcollisionally induced dissociation. Ions are allowed to collide

    with a background gas (He) for several millliseconds, prior to

    analysis. Allows for MSn experiments in an ion trap.

    Comparison of Mass Analyzers

  • 7/29/2019 Topic 8 Mass Spectrometry

    59/63

    Comparison of Mass Analyzers

    A brief overview of the properties of common mass

    analyzers

    Analyzer Cost Scan speed Resolution

    Double-

    focusing

    High Slow High

    Quadrupole Low Medium Low-medium

    Trap Low Medium Medium

    TOF Medium Medium Medium-high

    ICR High Fast High

  • 7/29/2019 Topic 8 Mass Spectrometry

    60/63

    Detectors for Mass Spectrometry

    Electron multipliers: like a

    photomultiplier tube. Ions strike asurface, cause electron emission. Each

    successive impact releases more

    electrons.

    Faraday Cups: Ions striking a cup cause

    charge to flow across a load. The

    potential across the load is monitored.

    Ionization

    Source

    Mass

    AnalyzerDetector

    Figure from D. W. Koppenaal, et al.;Anal. Chem., 77; 2005, 418A-427A.

    Detectors: Electron Multipliers

  • 7/29/2019 Topic 8 Mass Spectrometry

    61/63

    Detectors: Electron Multipliers

    Electron multiplier (EM): most common design in current

    use

    High gain (107), low noise, good dynamic range (104-106) Several designs:

    Figure from D. W. Koppenaal, et al.;Anal. Chem., 77; 2005, 418A-427A.

    D t t Oth

  • 7/29/2019 Topic 8 Mass Spectrometry

    62/63

    Detectors: Others

    Super-conducting tunner junction high mass

    range, used with MALDI Can detect fmol of 150 kDa proteins

    Can measure both energy and arrival time (2D MS

    plots of m/z vs. kinetic energy)

    Focal-plane array detectors/CCD

    Like in electronic spectroscopy, much more

    challenging to design for ion detection

    Would combine well with mini-traps or other smallMS systems

    Figure from D. W. Koppenaal, et al.;Anal. Chem., 77; 2005, 418A-427A.

  • 7/29/2019 Topic 8 Mass Spectrometry

    63/63

    MS-Chromatography Interfaces

    GC-MS: gas eluent from a column is piped directly to the

    MS source

    LC-MS: the ionization methods themselves serve as

    interfaces techniques like ESI, APCI and APPI work on

    liquid phase samples. The methods are generally

    tolerant to RP LC solvents and some NP solvents.

    Some buffers can quench ionization of analytes though:

    Bad: Phosphate leaves a solid upon evaporation.

    Also ionizes preferentially. Bad: any other non-volatile additives are also bad

    Good: TFA, ammonium acetate, formic acid

    Good: lower concentrations,