Tomotherapy Patient QA OmniPro-I’mRT...

52
Tomotherapy Patient QA using OmniPro-I’mRT MatriXX - the Marshfield experience Sugata Tripathi PhD Marshfield Clinic/Saint Joseph’s Hospital Marshfield, WI

Transcript of Tomotherapy Patient QA OmniPro-I’mRT...

Page 1: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the

Tomotherapy Patient QA using

OmniPro-I’mRT MatriXX- the Marshfield experience

Sugata Tripathi PhDMarshfield Clinic/Saint Joseph’s HospitalMarshfield, WI

Page 2: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the

(Treatment) Delivery Quality Assurance

• Goal is to verify that the measured dose for a fraction matches the planned dose/fluence map.

• Performed for every Tomotherapy treatment plan.

• Delivery QA Setup & Delivery QA Analysis done in Tomo PS.

• Dose plane computed and extracted to and compared with MatriXX measurement within the OmniPro I’mRT.

Page 3: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 4: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the

OmniPro-I’mRT & MatriXX

• MatriXX – The measurement device: 1020 vented ion chamber array detector in 32 x 32 grid.

• OmniPro-I’mRT – Software that controls the MatriXX and provides analysis tools and storage of measured data.

Page 5: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the

OmniPro-I’mRT & MatriXX

• MatriXX – 1020 vented ion chamber array detector in 32 x 32 grid, active area 24.4 x 24.4 cm2. • 7.6 cm center to center distance between chambers• 4.5 mm diameter 5 mm depth: 0.08 cm3 chamber volume

• Water equivalent depth from MatriXX surface to effective point of measurement ~ 3.3 mm. MatriXX mass is 10.5 Kg.

• Multicube Phantom: Individual Plastic Water blocks of 5-cm thickness interlock to form phantoms of different sizes.

Page 6: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the

OmniPro-I’mRT & MatriXX

• Measured current is proportional to dose rate.

• 1020 non-multiplexed ADC channels digitize measured current.

• Measured data are transmitted to a PC via standard Ethernet interface. OmniPro-I’mRT interface.

Page 7: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the

DQA: Delivery QA Setup

• Delivery QA Setup requires an approved patient plan and DICOM compatible phantom image volume.

Page 8: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 9: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 10: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 11: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 12: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 13: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 14: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 15: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 16: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 17: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 18: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 19: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 20: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 21: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 22: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 23: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 24: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 25: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 26: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 27: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 28: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 29: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 30: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 31: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 32: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 33: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 34: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the

Intensity map measurements using I’mRT MatriXX

• Mij and Bij: Obtained from the MatriXX measurement

• NDW(60Co) * Kuni ij : Measured and calculated at production site.

• Kuser : The kuser factor is created on site.• Koff i,j : The Koff i,j is user defined, replaces system

calibration output. Created on site, as needed.• kpT: Measured and calculated by MatriXX system.

The pressure factor is corrected with the offset obtained when comparing the P value as measured by the system and by a reference meter.

Page 35: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the

More on Kuser

• Specific to the Selected LINAC (dose rate)

• Energy type

• Energy value

•The system calculates kuser using the entered Dose(refernce)

value and the average of the values of four middle MatriXX

chambers. The new factor is displayed in the kuser field.

Page 36: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 37: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 38: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 39: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 40: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 41: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 42: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 43: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 44: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 45: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 46: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 47: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 48: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the

Criteria for QA Pass/Fail: Gamma Method

If comparing two dose distributions, the simplest methods, e.g. comparing isodoses, or difference, may not be appropriate.

Page 49: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the

Criteria for QA Pass/Fail: Gamma Method

Misjudgment when checking the agreement of two dose distributions may occur in the following cases:

• The difference between two dose-distributions can be large in high-gradient regions, even if the isodoses are relatively close to each other. • The isodose distance between two dose distributions can be large in regions with a flat dose distributions, though the difference in dose may be quite small.

Page 50: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the

Criteria for QA Pass/Fail: Gamma Method

The gamma method combines both methods mentioned above.

The user can define dose difference-and isodose distance pass/fail criteria.

If both parameters (dose and isodose distance) are outside their pass/fail criteria, the agreement "fails" according to the gamma method.

If only one parameter is outside the defined pass/fail criteria but the other well inside, the result of the comparison can still pass the calculation.

Page 51: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the
Page 52: Tomotherapy Patient QA OmniPro-I’mRT MatriXXchapter.aapm.org/nccaapm/z_meetings/2009-04-24/2009-Spring_talk… · 24/04/2009  · Tomotherapy Patient QA using OmniPro-I’mRT MatriXX-the

Criteria for QA Pass/Fail: Gamma Method

General Pass/Fail Criteria used for all IMRT QA in Marshfield:

3%, 3 mm DTA, 95% pass rate.