Tips & Tricks GPC/SEC: How to Treat Your RI Detector & Tricks GPC/SEC: How to Treat Your RI Detector...

4
Tips & Tricks GPC/SEC: How to Treat Your RI Detector Daniela Held, PSS Polymer Standards Service GmbH, Mainz, Germany Gel permeation chromatography/size-exclusion chromatography (GPC/SEC) is the method of choice for determining the molar mass distribution of synthetic and natural polymers. Every GPC/SEC system requires at least one concentration detector in the setup to allow for the detection of the fractions eluting from the column. Typical detectors used are refractive index detectors (RIs), ultraviolet (UV), diode array detectors (DAD), photodiode array detectors (PDA), or evaporative light scattering detectors (ELSDs). Each of these detectors has its advantages and disadvantages. This instalment of Tips & Tricks compares the RI to other concentration detectors and provides practical tips when using it. How Does an RI Detector Work? In general, RI detectors respond to a change of the refractive index, n . Most detectors are differential refractometers. They measure the deflection of a light beam from the difference in the refractive index between pure solvent and solvent with sample. This difference is referred to as ∆n and is expressed in refractive index units (RIU). Figure 1 shows how that can be achieved by using a detector flow cell divided diagonally into two parts. The cell therefore comprises a sample-containing side filled with the column effluent and a non-sample-containing reference side. If both sides contain the same mobile phase the detector is in optical balance. When sample elutes from the column, the refractive index in the sample-containing side will change and the light beam will be deflected. RI detectors are therefore very universal detectors. Compared to UV detectors they do not require chromophores in samples. They can detect all types of samples including The refractive index (RI) detector is the most common detector in gel permeation chromatography/size-exclusion chromatography (GPC/SEC). The advantage of this universal detector is that it detects everything; the disadvantage is that it detects everything. This instalment of “Tips & Tricks” offers some advice when working with RI detectors. Photo Credit: Benbou/Shutterstock.com 24 2 24 4 Instrumental Innovations 2 I I I n n ns s st t t r r ru u um m me e en n nt t ta a al l l I I I n n nn n no o ov v va a at t t i i i o o on n ns s s Tips and Tricks 24 T T Ti i i p p p s s s a a an n nd d d T T T r r r i i i c c ck k ks s s News 21 21 2 N N Ne e ew w ws s s Mallet 28 M M Ma a al l l l l l e e et t t Q&A: Steinhaus 34 Q Q Q & & &A A A: : : S S S t t t e e ei i i n n nh h ha a au u us s s Training and Events 43 T T T r r r a a ai i i n n ni i i n n ng g g a a an n nd d d E E E v v ve e en n nt t ts s s Moreau and Loftus 37 37 3 M M Mo o or r r e e ea a au u u a a an n nd d d L L Lo o of f ft t t u u us s s HTC Event Preview 41 41 H H HT T TC C C E E Ev v ve e en n nt t t P P Pr r re e ev v vi i i e e ew w w Staff 45 S S S t t ta a af f ff f f

Transcript of Tips & Tricks GPC/SEC: How to Treat Your RI Detector & Tricks GPC/SEC: How to Treat Your RI Detector...

Page 1: Tips & Tricks GPC/SEC: How to Treat Your RI Detector & Tricks GPC/SEC: How to Treat Your RI Detector Daniela Held, PSS Polymer Standards Service GmbH, Mainz, Germany Gel permeation

Tips & Tricks GPC/SEC: How to Treat Your RI DetectorDaniela Held, PSS Polymer Standards Service GmbH, Mainz, Germany

Gel permeation chromatography/size-exclusion

chromatography (GPC/SEC) is the method

of choice for determining the molar mass

distribution of synthetic and natural polymers.

Every GPC/SEC system requires at least one

concentration detector in the setup to allow for

the detection of the fractions eluting from the

column. Typical detectors used are refractive

index detectors (RIs), ultraviolet (UV), diode array

detectors (DAD), photodiode array detectors

(PDA), or evaporative light scattering detectors

(ELSDs). Each of these detectors has its

advantages and disadvantages. This instalment

of Tips & Tricks compares the RI to other

concentration detectors and provides practical

tips when using it.

How Does an RI Detector Work?

In general, RI detectors respond to a change

of the refractive index, n. Most detectors are

differential refractometers. They measure the

defl ection of a light beam from the difference in

the refractive index between pure solvent and

solvent with sample. This difference is referred

to as ∆n and is expressed in refractive index

units (RIU).

Figure 1 shows how that can be achieved

by using a detector fl ow cell divided diagonally

into two parts. The cell therefore comprises a

sample-containing side fi lled with the column

effl uent and a non-sample-containing reference

side. If both sides contain the same mobile

phase the detector is in optical balance. When

sample elutes from the column, the refractive

index in the sample-containing side will change

and the light beam will be defl ected.

RI detectors are therefore very universal

detectors. Compared to UV detectors they

do not require chromophores in samples.

They can detect all types of samples including

The refractive index (RI) detector is the most common detector in gel permeation chromatography/size-exclusion chromatography (GPC/SEC). The advantage of this universal detector is that it detects everything; the disadvantage is that it detects everything. This instalment of “Tips & Tricks” offers some advice when working with RI detectors.

Ph

oto

Cre

dit

: B

en

bo

u/S

hu

tte

rsto

ck.c

om

242244

Instrumental Innovations2 IIInnnssstttrrruuummmeeennntttaaalll IIInnnnnnooovvvaaatttiiiooonnnsss Tips and Tricks24 TTTiiippppsss aaannnddd TTTrrriiiccckkksssNews21212 NNNeeewwwsss Mallet28 MMMaaalllllleeettt Q&A: Steinhaus34 QQQQ&&&AAA::: SSSttteeeiiinnnhhhaaauuusss

Training and Events43 TTTrrraaaiiinnniiinnngggg aaannnddd EEEvvveeennntttsssMoreau and Loftus37373 MMMooorrreeeaaauuu aaannnddd LLLoooffftttuuusss HTC Event Preview4141 HHHTTTCCC EEEvvveeennnttt PPPrrreeevvviiieeewww Staff45 SSStttaaaffffff

Page 2: Tips & Tricks GPC/SEC: How to Treat Your RI Detector & Tricks GPC/SEC: How to Treat Your RI Detector Daniela Held, PSS Polymer Standards Service GmbH, Mainz, Germany Gel permeation

The Column www.chromatographyonline.com

polysaccharides, starches, and polyesters.

Compared to ELSDs, which do not require

chromophores either, they are much more

linear in response and evaporating the solvent

is not required. Therefore, RIs can be used with

all types of solvents with or without salt. In

addition, RIs are able to detect oligomers and

contaminants as residual monomer, which could

be evaporated when using ELSDs.

Disadvantages of RI Detectors

At a fi xed wavelength of incident light the

changes in refractive index are generally linear

in relation to the changes in medium density.

However, the density of a medium is also

affected by composition, temperature, and

pressure. Therefore, the refractive index will

change with solvent composition, traces of

contamination, degassing level, temperature

fl uctuations, and more. One consequence of

this is that an RI detector can only be applied

in isocratic mode—solvent gradients are not

possible. Another consequence is that the RI

will respond to any change in the experimental

conditions with a change in its signal and a

variation of its baseline. It sometimes appears

as if RIs are unstable or generate unstable

baselines, when in fact, RIs are simply displaying

the instabilities of the environment and the rest

of the system.

It should be also kept in mind that RIs

are in general not as sensitive as UV

detectors or ELSDs are. They need relatively

high concentrations compared to other

detectors.

How to Handle an RI Detector

The cell design and the principle of the RI

detector have some consequences for practical

work. In a multidetector setup, which is very

typical in GPC/SEC, the RI often has to be the

last detector in the daisy-chain (1,2). The main

reasons for this are that the cell can withstand

only low backpressures, the cell and the tubing

in the detector often have large dimensions

leading to band broadening, and the signal

itself is pressure sensitive.

The vast majority of an RI needs fl ushing or

purging of the reference cell. To achieve

a stable baseline, the solvent in the reference

cell and the mobile phase must be the same

quality. The reference cell must be fl ushed or

purged regularly, and at the very least when

the mobile phase in the reservoir is changed.

The best time to purge or fl ush the RI is

directly prior to the analysis as the last action

after the system and the columns have been

thoroughly fl ushed and had time to stabilize. If

Figure 1: RI detector cell diagonally split into the reference and the sample part. The RI detects difference in the refractive index between pure solvent and solvent with sample.

Figure 2: RI chromatogram of a mixture of four PMMA reference materials with different molar masses. The system peaks at the end of the chromatogram indicating the end of size exclusion mode are typical. Drift, noise, wander, and S/N are compared to the requirements of ISO13885-1, when applicable.

Tips & Tricks

252255

Instrumental Innovations2 IIInnnssstttrrruuummmeeennntttaaalll IIInnnnnnooovvvaaatttiiiooonnnsss Tips and Tricks24 TTTiiippppsss aaannnddd TTTrrriiiccckkksssNews21212 NNNeeewwwsss Mallet28 MMMaaalllllleeettt Q&A: Steinhaus34 QQQQ&&&AAA::: SSSttteeeiiinnnhhhaaauuusss

Training and Events43 TTTrrraaaiiinnniiinnngggg aaannnddd EEEvvveeennntttsssMoreau and Loftus37373 MMMooorrreeeaaauuu aaannnddd LLLoooffftttuuusss HTC Event Preview4141 HHHTTTCCC EEEvvveeennnttt PPPrrreeevvviiieeewww Staff45 SSStttaaaffffff

Page 3: Tips & Tricks GPC/SEC: How to Treat Your RI Detector & Tricks GPC/SEC: How to Treat Your RI Detector Daniela Held, PSS Polymer Standards Service GmbH, Mainz, Germany Gel permeation

The Column www.chromatographyonline.com

purge or fl ush is performed while the system (or

the columns) are still equilibrating, the solvent

in the reference cell will be of a different quality

to the sample cell. No optical balance can be

achieved before the run and low signal quality

may be a problem during the run.

Troubleshooting for RI Detectors

Figure 2 shows a common RI trace obtained

for a mixture of four different polymethyl

methacrylate (PMMA) reference materials

with different molar masses. The system

peaks at the end of the chromatogram are

very typical. These peaks can be only negative

(as in the example), but a combination of

negative and positive peaks is also common.

The system peaks indicate the end of size

exclusion mode.

It is good practice in GPC/SEC to prepare

the samples in solvent taken from the mobile

phase reservoir to reduce the system peaks to

a minimum. The larger the difference between

the solvent used for sample preparation and the

mobile phase, the larger the system peaks will

be. System peaks are identifi ed by measuring

a blank sample within the sequence and

overlaying this with the samples. It is important

that the blank is treated the same way as the

samples themselves including all fi ltration steps.

Problems that can occur with RI detectors

are mainly the same as for other detectors in

GPC/SEC. Typical are baseline drift (the steady

movement of the baseline, either up or down

the scale), baseline wander (variation of the

detector with frequencies between 6 and

60 cycles/h), and detector noise (variation of the

detector signal with frequencies above 1 cycle/s)

(3). The inset in Figure 2 shows the values

for baseline drift, signal noise, signal wander,

and signal-to-noise ratio (S/N), if applicable

compared to the requirements of ISO13885-1.

S/N Ratio: When discussing the S/N, it is

important to understand the RI detector

response to a sample. The RI signal intensity

depends on the concentration and the

refractive index increment, dn/dc.

For a suffi cient S/N two parts need to be

considered:

1) The larger the concentration, the larger

the signal area. In Figure 2 all PMMAs have

the same dn/dc, so the same concentration

should result in the same area. This is true for

this example because the highest molar mass

(eluting around 6 mL) has only half of the

concentration compared to the three other

lower molar mass samples resulting in only half

of the peak area (4).

2) The larger the dn/dc, the larger the signal

area. For example, if a polystyrene (PS) mixture

with the same molar masses and concentrations

is measured, the PS signals will be higher than

the PMMA signals because of the higher dn/

dc of PS compared to PMMA (conditions: THF,

35 °C). If the dn/dc is 0 (isorefractive samples),

such as for polydimethylsiloxane (PDMS) in THF,

no signals or only small signals resulting from

end groups will be obtained when using an RI.

Here another solvent, such as toluene, needs

to be used to obtain representative signals with

good S/N. As the dn/dc is negative for PDMS

in toluene, negative signals will be obtained.

It is possible to either inverse the signal at the

RI directly or in the corresponding software to

evaluate them properly.

Baseline Drift: As mentioned earlier, RIs

show instabilities of the environment and

the rest of the system. There are three

major sources for baseline drift: low mobile

phase quality, temperature differences

between the columns and the RI cell, or a

not yet equilibrated system. Using freshly

prepared, high-quality solvent, thoroughly

Figure 3: RI detector wander example. A faulty degasser was responsible for the fl uctuations. As RIs also respond to the density of a medium the degassing level is important. Similar signals have been reported for malfunctioning ovens (heating cycle) or laboratories with bad air conditioning.

Tips & Tricks

262266

Instrumental Innovations2 IIInnnssstttrrruuummmeeennntttaaalll IIInnnnnnooovvvaaatttiiiooonnnsss Tips and Tricks24 TTTiiippppsss aaannnddd TTTrrriiiccckkksssNews21212 NNNeeewwwsss Mallet28 MMMaaalllllleeettt Q&A: Steinhaus34 QQQQ&&&AAA::: SSSttteeeiiinnnhhhaaauuusss

Training and Events43 TTTrrraaaiiinnniiinnngggg aaannnddd EEEvvveeennntttsssMoreau and Loftus37373 MMMooorrreeeaaauuu aaannnddd LLLoooffftttuuusss HTC Event Preview4141 HHHTTTCCC EEEvvveeennnttt PPPrrreeevvviiieeewww Staff45 SSStttaaaffffff

Page 4: Tips & Tricks GPC/SEC: How to Treat Your RI Detector & Tricks GPC/SEC: How to Treat Your RI Detector Daniela Held, PSS Polymer Standards Service GmbH, Mainz, Germany Gel permeation

The Column www.chromatographyonline.com

GPC/SEC light scattering setups they are

often used to estimate the refractive index

increment, dn/dc.

• The disadvantages of RIs are that they

not very sensitive, they are affected by

environmental conditions, and they have a

reference cell.

• RI detectors require high solvent quality for

stable baselines.

• Purging or fl ushing of the RI reference cell

should be done as the last action prior to

the analysis and only for fully equilibrated

systems.

References

1. D. Held and P. Kilz, The Column 8(18), 9–12 (2012).

2. D. Held and W. Radke, The Column 13(11), 9–14

(2017).

3. D. Held, The Column 8(6), 13–18 (2012).

4. D. Held, The Column 10(10), 12–15 (2014).

5. D. Held, The Column 9(2), 2–5 (2013).

Daniela Held studied polymer chemistry

in Mainz, Germany, and works in the PSS

software and instrument department.

She is also responsible for education and

customer training.

The air conditioning installed in the laboratory

could also evoke detector wander.

Other Problems: A very interesting

troubleshooting example is shown in Figure 4.

Here the signal of the internal standards (like

that of the other samples in the following

injections) starts negative and then becomes

directly positive with a steep increase in

between. The reason for this behaviour was

located in a defective purge valve. The reference

part of the cell was therefore not completely

shut during analysis and the sample was fl owing

through both cell parts, resulting in one positive

and one negative signal. The same behaviour

would be observed if a sample was run in

activated RI purge or fl ush mode. A replacement

of the purge valve took care of this problem.

The signal split does not have to be as

pronounced as in the example. Sometimes the

process starts with a small negative dip before

the signal starts. However, once the valve is

fully damaged the signal will be as shown in

the example.

Summary

• RI detectors are universal detectors with a

linear response suitable for many different

types of macromolecules. They do not

require chromophores because they respond

to a change in refractive index.

• RIs are typically used as concentration

detectors in GPC/SEC. When applied in

After removing the degasser completely from

the setup, the signal quality was stable again.

Similar traces are also observed when the

degasser is not working properly (because

of low solvent quality or malfunctioning of

the vacuum pump) or if the column oven

temperature is not stable enough (heating

cycles). In both cases working without a

degasser or column oven yields better results

than working with low quality components.

thermostatted columns and RI detector cells,

and providing enough equilibration time for

the system should take care of this issue.

Baseline Wander: Worse than drift—which

can be easily corrected (especially when

using the recommended two-step evaluation

procedure [5])—is detector wander. Figure 3

shows a nice example of detector wander.

The reason for the signal fl uctuations was a

degasser not suited for this GPC/SEC solvent.

E-mail: [email protected]: www.pss-polymer.com

Figure 4: RI trace obtained for an RI with a defective purge valve. The observation of a negative signal with a steep increase to a positive signal is typical. Replacement of the purge valve was required to solve the problem.

Tips & Tricks

272277

Instrumental Innovations2 IIInnnssstttrrruuummmeeennntttaaalll IIInnnnnnooovvvaaatttiiiooonnnsss Tips and Tricks24 TTTiiippppsss aaannnddd TTTrrriiiccckkksssNews21212 NNNeeewwwsss Mallet28 MMMaaalllllleeettt Q&A: Steinhaus34 QQQQ&&&AAA::: SSSttteeeiiinnnhhhaaauuusss

Training and Events43 TTTrrraaaiiinnniiinnngggg aaannnddd EEEvvveeennntttsssMoreau and Loftus37373 MMMooorrreeeaaauuu aaannnddd LLLoooffftttuuusss HTC Event Preview4141 HHHTTTCCC EEEvvveeennnttt PPPrrreeevvviiieeewww Staff45 SSStttaaaffffff