THz Wave 3D Imaging and Tomography

download THz Wave 3D Imaging and Tomography

of 22

Transcript of THz Wave 3D Imaging and Tomography

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    1/22

    Chapter 6

    THz Wave 3D Imaging and Tomography

    THz waves are transparent to most of the dry dielectric materials. This property

    makes THz wave a promising candidate for nondestructive evaluation of the inter-nal structures of targets. THz wave time-of-flight imaging method is one of the

    techniques to extract the information about the layered structures of a target. If there

    is no layer structure within the target, or if the interesting features are not located on

    those layer structures, one needs to use tomographic imaging techniques to extract

    those interesting information [1].

    T-Ray Computerized Tomography

    Initially, Computerized tomography (CT) technique was developed for X-ray imag-

    ing, and currently is used in medical diagnostics and other nondestructive evaluation

    applications. Figure6.1illustrates the concept of CT. Collimated X-ray beam trans-

    mits through the target and the transmitted intensity is recorded. The transmission

    of the X-ray beam through different paths defined by lateral position and angles is

    obtained by rotating and laterally shifting the target. For each beam path, the inten-

    sity of transmitted beam is determined by the integral absorption along the optical

    path. If the absorption distribution of the target is defined as f(x,y,z), then the total

    absorption along the path with rotation angle , height z and lateral displacementlis described as

    p(,l,z) =

    L(,l,z)

    f(x,y,z)dl (f(x,y,z), (1a)

    where L(,l,z) denotes the beam path through the target. The transmission Tand

    absorptionp of the beam path are defined as

    T= exp

    L(,l,z)

    f(x,y,z)ds

    exp[ p(,l,z)]. (1b)

    127X.-C. Zhang, J. Xu,Introduction to THz Wave Photonics,

    DOI 10.1007/978-1-4419-0978-7_6, C Springer Science+Business Media, LLC 2010

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    2/22

    128 6 THz Wave 3D Imaging and Tomography

    Fig. 6.1 Concept of

    computerized tomography

    (CT). Intensity of carrier

    wave beam is recorded after

    transmission through the

    target with certain lateral

    shifts and rotation angles.

    Inverse Radon transform

    of the recorded signal gives

    3D information of the target

    Equation (1) is known as the Radon transform [2]. The Fourier transform ofEquation (1) gives

    P(,,z) = p(,l,z)exp( il)dl= exp[ i(x y)]f(x,y,z)dxdy. (2)

    Here v is lateral spatial frequency, which indicates the spatial resolution of the

    CT imaging process. = vsin()and= vcos()are spatial frequency componentsalong the xand ydirection, respectively. Equation (2) indicates that, Fourier trans-

    form of the Radon transform is equivalent to 2D Fourier transform of the absorptiondistribution.

    CT image directly measures transmission of the carrier wave corresponding to

    each beam path. To get the 3D structure of the target, Equations (1) and (2) are used

    to calculate the spatial distribution of the absorption, f(x,y,z), in the target

    f(x,y,z) = P(,v,z) exp2 i(xcos +y sin ) dd=

    P(,l,z)exp( il) exp

    2 i(xcos +y sin )dldd.

    (3)

    This equation is called the reverse Radon transform. Reverse Radon transform of

    the beam path absorptions at height zgives cross section image of the target at the

    heightz. The above process is repeated for all different heights to obtain the cross

    section images at the different heights. These different cross section images provide

    the 3D structure of the target.

    Unlike X-ray imaging, which provides only the intensity information, the THz

    wave imaging provides both the intensity and the phase information. The THz CT

    enables one to obtain the absorption and the refractive index information of the

    target. An experimental setup of THz wave CT system is presented in Fig6.2.To

    ensure the spatial resolution, THz beam is focused onto the target using either an

    off-axis parabolic mirror or a lens. The focal spot size determines the spatial reso-

    lution of the CT image, while its depth of focus should be at least as long as lateral

    dimensions of the target. Therefore, the focusing conditions of THz beam needs to

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    3/22

    T-Ray Computerized Tomography 129

    Fig. 6.2 Experimental setup of THz wave CT. THz wave is focused with an off-axis parabolic mir-

    ror onto the target. Temporal waveform of THz pulses transmitted through the target is recorded.

    The target is driven by a rotation stage and a linear translation stage. THz waveform at each rotation

    and lateral transition position is recorded

    be well considered so that it provides sufficient depth of focus along with the high

    spatial resolution. In order to obtain a tomographic image, the target is set on a rota-

    tion stage which is able to rotate about the zaxis and the target moves along land

    z directions by a 2D translation stage. Thus, the target is scanned in three dimen-

    sions corresponding to variables of z, l and . Another off-axis parabolic mirror

    collects the transmitted THz waves and a THz wave detector records its temporal

    waveform.

    Refractive index of the materials can be approximately treated as 1 for almostall common materials when performing X-ray CT. As a result, X-ray beam can

    be treated as a simple straight line without any beam bending in the CT process.

    However, different materials may have different refractive index in THz band. For

    example, refractive index of plastic is1.5, of glass is2, and of semiconductormaterials could be as high as 34. As a result, refractive behavior of THz wave may

    not be negligible in CT process unless specific conditions are satisfied. Further, the

    wavelength of THz wave is usually comparable to spatial features of the target, such

    as textures, or microstructures, thus diffraction and scattering of the beams may also

    affect propagation of THz wave through the target. To reconstruct the THz wavetomographic image using Equation (3), the materials refractive index should have

    slow variation within a THz wavelength, otherwise, ray-tracing techniques need to

    be used in order to correctly retrieve 3D image of the target. In pulsed THz wave

    tomography, the entire THz waveform is recorded. Different physical quantities can

    be used to represent different features of the target. For example, the amplitude of

    transmission provides absorption coefficient, the phase delay represents refractive

    index of the target and the spectroscopy provides the composition information of

    the target.

    Polystyrene foam material has very low absorption coefficient (

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    4/22

    130 6 THz Wave 3D Imaging and Tomography

    Fig. 6.3 Photo of a target

    used in THz wave CT

    imaging process. Cubic

    polystyrene foam is carved

    with THZ through its

    height

    wave CT process. Three letters T, H and Z were carved in the target. Since

    there is no change in z direction, the cross section at z= z0 reflects the full 3Dinformation of the target. In order to obtain the image, the sample was driven along

    the x-axis by a linear stage. For each xposition, the sample was rotated within a

    range of= 0180 with an interval of 1.8. Imaging system recorded the THzwaveform for each x and . Figure 6.4 shows 3 THz waveforms recorded in the

    experiment. Those waveforms were obtained for the lateral displacement x= 0,and rotation angles of 0, 60 and 120, respectively. Both amplitude and positionof peak in THz waveform vary for different beam paths. Figure 6.5ashows time

    delay of THz pulses in the xspace. In the frequency domain, time delay of THz

    pulses corresponds to the phase delay. Figure6.5bgives the phase distribution at 1

    THz (the central frequency of THz pulses) in the x space. Using time delay to

    replace absorption in Equation (3), the 3D image of the target can be reconstructed

    according to its refractive index distribution. Figure6.6shows the retrieved cross

    section image of the target of Fig.6.3.The profile of target and the carved letters are

    reconstructed due to refractive index difference between air and polystyrene foam.Similarly, the phase distribution in the frequency domain can be used to retrieve 3D

    image of the target and the result is very similar to Fig. 6.6. Further, one can also

    use variation of THz amplitude to retrieve the absorption coefficient (or extinction

    coefficient) distribution of the target. The image based on amplitude attenuation of

    THz wave is less sharp due to scattering of the waves by the target.

    When the sample is scanned in the vertical direction along with the lateral scan-

    ning and rotation, the cross section of the target can be obtained for each height

    that provides a true 3D image of the target. To save the scanning time, single pulse

    detection method discussed in Chapter 3 can be used to record THz waveforms.

    Figure6.7presents cross sections images of a ping-pong ball recorded by THz wave

    CT at different heights. When all these cross sectional images are combined the

    THz wave CT retrieves cavity structure of the ping-pong ball, which is presented

    in Fig 6.8. The cavity structure of the ping-pong ball can be clearly seen in the

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    5/22

    T-Ray Computerized Tomography 131

    Fig. 6.4 Temporal waveforms of THz pulses transmitted through the target at 0, 60 and 120,respectively. THz pulse propagation through the target following different path does not only lead

    to different absorption, but also gives different phase delay. The phase delay reflects refractiveindex distribution of the target

    Fig. 6.5 The distribution of physical qualities in a THz wave CT measurement presented in x

    space. (a) Distribution of THz peak timing, and (b). distribution of phase delay at 1 THz

    THz wave CT; however, the thickness of the retrieved shell is much thicker than

    the real one. This is due to the fact that the physical model of THz CT is based on

    gradual change of refractive index; however, the real target has a sharp change at

    boundaries of the shell. Further, the shell of the ping-pong ball has relatively high

    refractive index and the Fresnel losses at its surface are not negligible. The Fresnel

    losses also affect the image quality. Figure6.9shows an optical photo of a piece

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    6/22

    132 6 THz Wave 3D Imaging and Tomography

    Fig. 6.6 3D image of the

    target reconstructed via

    inverse Radon transform of

    the peak timing distribution

    in thexspace

    of dry and defat turkey bone and the corresponding THz CT image [3]. The optical

    photo only shows profile of the bone, while THz CT image shows the profile as wellas the central cavity structure. However, there are rich microstructures in the bone,

    which are not retrieved in THz CT image. The inability of reconstruction to resolve

    microstructure is because the CT algorithm requires that the target has low scattering

    and absorption, which are not satisfied for the turkey bone. The heavy scatter-

    ing by those complicated microstructures reduces image quality of the retrieved

    image.

    Spatial resolution is the most important feature for an imaging system. The def-

    inition of spatial resolution may vary from one imaging system to anther system.

    If the imaging system is diffractive limited, the Rayleigh criteria is the most com-monly used definition. For THz CT imaging, however, structure or composition of

    the target may become a dominant factor affecting the spatial resolution. As a result,

    a more practical definition of spatial resolution is commonly used in study of THz

    wave CT. It is defined as the smallest distance between two spots that can be dis-

    tinguished in the retrieved image. Figure6.10shows photos of a cylindrical sample

    made using polystyrene foam. The sample contains eight holes with a diameter of

    2.5 mm each and two slots with 2.5 mm width. The smallest distance between two

    holes lying on the line AA is 0.5 mm. Figure16.11agives the reconstructed crosssection image of the sample using THz CT. The image indicates that THz CT pro-

    vides a spatial resolution of at least 0.5 mm, because even the closest holes are

    distinguishable in the image. Ray approximation used in the reconstruction does

    not consider the diffractive behavior of THz wave limited by the imaging system

    aperture. In reality, the measured phase change is an involution between the sample

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    7/22

    T-Ray Diffraction Tomography 133

    Fig. 6.7 THz wave CT images of a ping-pong ball

    caused phase delay and the diffraction effect. Using Wiener filter, the diffraction

    induced phase change can be removed from the THz CT image to obtain the pure

    sample information. Figure6.11bshows the enhancement in the sharpness of the

    structure of the target obtained by applying Weiner filter on the image in Fig.6.11a.

    T-Ray Diffraction Tomography

    Diffraction phenomenon plays a strong role in THz wave imaging due to the rela-

    tively long wavelength of THz waves. Interaction between THz wave and the target

    can be described by the following Maxwell equation[4]:

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    8/22

    134 6 THz Wave 3D Imaging and Tomography

    Fig. 6.8 3D image of the

    ping-pong ball cavity

    Fig. 6.9 Photo of a turkey bone (left) and its THz wave CT image (right)

    2E+ (r)(r)c2

    2

    t2E+ [ln (r)] ( E) + [E ln (r)] = 0, (4)

    where r denotes a position in space, c is the speed of light, (r) and (r) are

    the complex permittivity and permeability of the material, respectively. If the

    variation of both and is small within a wavelength, the above equation can

    be simplified to

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    9/22

    T-Ray Diffraction Tomography 135

    Fig. 6.10 Side view and top view of a target used in THz wave CT. The target is made of

    polystyrene foam

    Fig. 6.11 (a) THz wave CT cross section image of the target shown in Fig. 6.10(b) the cross

    section image with Wiener filter applied

    2E+ n(,r)2

    c2

    2

    t2E= 0 (5)

    .

    Here n = is the complex refractive index of the target. If polarization effectsof the EM wave are negligible, the vector equation can be further simplified to the

    Helmholtz scalar equation:

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    10/22

    136 6 THz Wave 3D Imaging and Tomography

    2u(r) + k20n(,r)2u(r) = 0, (6)

    wherek0= 2/is the wave number of EM wave in vacuum and u(r)is the complexamplitude function of the electric field. In an imaging process,u(r)can be described

    as the sum of two components: u(r)= u0(r) + us(r), where the former is electricfield of EM wave which does not penetrate through the target or is a solution to the

    equation

    2 + k20

    u0(r) = 0, (7)

    while the later is the electric field of EM wave after transmission through the target.

    If the target modulates only phase of the EM wave, the electric field can be

    written as

    u(r) = exp [(r)] = exp[0(r) + s(r)], (8)

    where is the complex phase of the electric field. Assuming that the scattering

    induced phase shift changes slowly, Equation (7) can be further simplified using the

    first order Rytov approximation to

    s(r) = 1u0(r)

    G(r r)V(r)u0(r)dr, (9a)

    whereG(rr)is a Green function [5]and is described as

    G(r r) = exp

    ik0r r

    4r r (9b)

    andV(r) is defined as

    V(r) = k20[n2(,r) 1]. (9c)

    Compared to optical waves, diffraction of THz pulses shows its unique prop-

    erties. The THz pulse contains broadband spectrum and its interference pattern is

    generated by sum of all frequency components. Figure6.12 illustrates an experi-

    mental setup of Youngs double slit interference measurement for THz wave. In this

    experiment, the sample was a 0.17 mm thick aluminum foil with double slits of

    width 1 mm each and the distance between those two slits being 6 mm. The dis-

    tance from the double slits plane to the THz wave detector was 48 mm. Since THz

    pulse has a short pulse width, its interference pattern also varies in time. The num-

    ber of peaks and valleys in the interference pattern increases with time delay. This

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    11/22

    T-Ray Diffraction Tomography 137

    Fig. 6.12 Experimental setup of THz wave Youngs double slit measurement. The double slits are

    carved on 0.17 mm thick aluminum foil with width of 1 mm each. The distance between those two

    slits isd= 6 mm the distance between the double slit and the detection crystal is D = 48 mm

    phenomenon indicates that, after THz pulse pass through the double slits, lower fre-

    quency components propagate faster than the higher frequency components. Figure

    6.13ashows the evolution of THz wave interference pattern along the direction per-

    pendicular to the double slits. Figure 6.13b shows position of peak and valley in

    Fig. 6.13a, which presents the wavefront propagation of THz wave starting from

    each slit.

    a b

    Fig. 6.13 The temporal evolution of THz field along xaxis. (a) experimental result, and (b) the

    normalized electric field (according to peak of THz waveform)

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    12/22

    138 6 THz Wave 3D Imaging and Tomography

    Fig. 6.14 The concept of

    diffraction tomography (DT).

    The carrier wave in DT

    measurement is a plane wave.

    Wavefront of the carrier wave

    is recorded after propagation

    through the target, and which

    is used to retrieve 3D image

    of the target

    In THz diffraction tomography, a planar THz wave is used to illuminate a target

    and the targets structural information is extracted by measuring and processing the

    diffracted THz distribution by the target. Figure 6.14shows the concept of diffrac-

    tion tomography using a monochromatic wave. The light source in a diffraction

    tomography is a plane wave. The THz wave is detected in a detection plane per-

    pendicular to its propagation direction after transmission through the target. The

    Fourier transform of the forward scattered electric field of the THz wave on the

    detection plane is proportional to 3D Fourier transform of the object function in adisplaced hemisphere in the frequency domain. The displacement value equals to

    a negative incident wave number along the incident wave direction. Consequently,

    reverse Fourier transform of the measurement result leads to the object function. In

    THz CT, one needs to record THz field of an entire beam; however, in THz DT, THz

    wave distribution in the detection plane needs to be resolved. This can be achieved

    by using the 2D imaging technique discussed in Chapter 3. In THz DT set up, THz

    waves are generated from a point emitter via optical rectification process excited

    by high intensity fs laser pulses. The generated THz beam is collimated by a off-

    axis parabolic mirror and illuminates the entire target which is placed on a rotationstage. The forward scattered THz beam from the sample and the optical probe beam

    propogate collinearly through the EO crystal. This probe beam records the 2D dis-

    tribution of the THz diffraction formed on the EO crystal. The 2D image of the

    target is obtained for each rotation angle by recording the forward scattered wave

    detected using a large aperture EO crystal. Figure6.15shows a target made of three

    plastic rectangular bars used in THz DT. If the target modulates only the phase of

    THz wave, the first order Rytov approximation can be used to reconstruct the 3D

    image:

    u = u0ln

    us

    uo+ 1

    (10)

    .

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    13/22

    3Dimensional Imaging Using Fresnel Lens 139

    Fig. 6.15 Photo of the target in THz wave DT (left) and its top view illustration (right). The target

    consists of 3 plastic bars

    Figure 6.16a gives a reconstructed refractive index distribution of the sample

    shown in Fig.6.15and the Fig.6.16bshows the reconstructed image presented in

    3D. The image is reconstructed based on 0.2 THz component.

    The purpose of imaging is to obtain an image that closely resembles the target and

    provides the maximum structural information of the target. Therefore, the quality of

    image is evaluated by calculating similarity between the target and its reconstructedimage. The imaging quality of THz wave tomography is defined as a normalized

    cross correlation function between a target and its image [6, 7]:

    Q =M,N

    i,j=0[O(i,j) O][I(i,j) I]M,Ni,j=0

    O(i,j) OM,Ni,j=0 I(i,j) I , (11)where i,j are the pixel index, O is the object function of the target, O is mean of

    O,Iis the reconstructed image function and Iis its mean value. Figure6.17shows

    image quality of THz DT image presented in Fig.6.16as a function of the THz wave

    frequency. For low frequency wave ( 0.45

    THz), the imaging quality is limited by the lower Signal to noise ratio. Thus, the

    imaging quality of the THz wave diffraction tomography increases with frequency

    until 0.2 THz and has a plateau between 0.2 and 0.45 THz and further decreases

    with the frequency.

    3Dimensional Imaging Using Fresnel Lens

    A Fresnel lens is a Fresnel zone plate with phase or amplitude patterns formed

    by series of concentric circles. The Fresnel lens manipulates the optical wave via

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    14/22

    140 6 THz Wave 3D Imaging and Tomography

    Fig.

    6.1

    6

    Th

    ereconstructedTHzwaveDTimageofthetargetshowsinFig

    .6.1

    5(a)refractiveindexdistributionofthetarget,and(b)3D

    imageofthe

    refractiveindexdistribution

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    15/22

    3Dimensional Imaging Using Fresnel Lens 141

    Fig. 6.17 The image quality

    of THz wave DT as a function

    of THz wave frequency

    diffraction and interference. Fresnel lens can be considered as an example of diffrac-

    tion optical components. Compared to traditional optical lens, Fresnel lens is more

    flexible in design and fabrication. Further, it can be much thinner than refractive

    lens with same aperture and focal length and thus can replace large aperture lens

    in order to reduce the weight. Binary Fresnel lens are fabricated by photolithogra-

    phy and etching on a transparent disk to make phase modulation features.Ncircles

    of photolithography and etching gives L= 2 N levels of phase modulation depth.Figure6.18shows pictures of three THz wave binary Fresnel lens fabricated on an

    intrinsic silicon wafer. Levels of those lenses are 2, 4, and 8, respectively. Figure

    6.18illustrates the diffraction patterns of THz pulses focused by these three lenses.

    Diffraction of THz wave by Fresnel lens along z-axis (normal of the lens) is [8]

    u(z)

    = nAn expi2

    n

    r

    2

    p +

    1

    2z (x2 +

    y2) dxdy, (12)where n= 1,2,. . .., An= sin c(n/L). r2p is Fresnel zone of the lens with thedimensions of area and indicates a zone area where the phase delay is constant. For

    an incident plane wave, its focal spot is located at

    zn= r2p

    2n, n = 1,2,...... . (13)

    Equation (13) indicates that Fresnel lens has a very high chromatic dispersion and

    its focal length is inversely proportional to the wavelength of the incident wave.

    Equation (13) also shows that a Fresnel lens has multiple orders of focal spots for

    a monochromatic plane wave. The diffraction efficiency for its focal point of each

    order is

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    16/22

    142 6 THz Wave 3D Imaging and Tomography

    Fig. 6.18 THz wave Fresnel lenses made on silicon wafer with a level of 2 (a), 4 (c), and 8 (e),

    respectively. The corresponding THz wave intensity distribution at the focal plane ( b,d, andf)

    (n) |An|2 = sin c2(n/L). (14)

    Equation (14) indicates that the lower order of focal spot gives the higher diffraction

    efficiency. Table6.1summarizes the diffraction efficiency for the first order focal

    spot of those three silicon based Fresnel lenses. An aluminum zone plate is used as

    comparison.

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    17/22

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    18/22

    144 6 THz Wave 3D Imaging and Tomography

    Fig. 6.19 The first order

    focal length of a Fresnel lens

    as a function of THz

    frequency

    Fig. 6.20 The axial evolution of THz amplitude focused by a Fresnel lens

    length of 2.6 cm at 1 THz. By scanning the time delay between the THz and opti-

    cal probe beam, a temporal waveform of the THz wave at each pixel on the image

    plane was recorded using a CCD camera. Fourier transformation of the temporal

    waveforms provides the THz field amplitude (or intensity) distribution on the image

    plane at each frequency. The measured two-dimensional THz field distribution at

    each frequency provides images of the THz field transmission of a target at each

    corresponding position along the z-axis. Figure6.21a, b, and c show illustration of

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    19/22

    3Dimensional Imaging Using Fresnel Lens 145

    Fig. 6.21 Illustrations of three targets used in Fresnel lens based THz wave tomographic imaging

    (a), (b), and (c). Their images made with THz wave at different frequencies (d), (e), and (f)

    three objects used in the pulsed THz wave tomographic imaging system. The three

    targets (OU, OC and OT) were made from 2 mm thick polyethylene sheet with the

    dimensions of 60 by 40 mm with three different shapes carved on these three sam-

    ples. These targets were placed along the THz beam path with their distances tothe lens, corresponding to d0 in Equation (15), being 3, 7 and 14 cm, respectively.

    Images of patterns on the sensor plane at distance d1= 6 cm are measured at fre-quencies of 0.74, 1.24, and 1.57 THz, respectively. The corresponding images are

    shown in Fig.6.21d, e, and f. At each frequency, a Fresnel lens images a different

    plane section of a target object while images from other depths remain blurred. Each

    point in the different object planes along the z-axis is mapped onto a correspond-

    ing point on the image plane (sensor plane) with the magnification factor d1/d0 at

    their corresponding frequencies. Table6.2summarizes the properties of those three

    objects and their images.

    Figure6.22 shows target distance d0 as function of TLHz wave frequency for

    three different image distances of d1 equal to 4.6, 5.6 and 7.4 cm respectively

    and shows a strong agreement between the experimental (points) and the mea-

    sured (curves) values. Lateral spatial resolution of THz wave image created by a

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    20/22

    146 6 THz Wave 3D Imaging and Tomography

    Table 6.2 Characters of 3 targets and their images

    Object (Fig.6.22)

    Size (mm) 5 10 14

    Object distance (cm) 3 7 14Carrier frequency (THz) 0.75 1.24 1.57

    Image size (mm) 10 8 6

    Magnification (experimental) 2 0.8 4.2

    Magnification (calculation) 2 0.85 4.3

    Fig. 6.22 The object distance in Fresnel lens based THz wave tomographic imaging system as

    functions of THz frequency for three different image distances

    Fresnel lens is limited by diffraction of the imaging system. The longitudinal spatial

    resolution is determined by the spectral resolution of the THz wave detector. The

    relationship between longitudinal range of the object plane and the bandwidth of

    THz wave can be derived from Equation (16) as

    d0=

    d20

    z

    . (17)

    Equation (17) shows that the longitudinal resolution increases with higher THz wave

    frequency. The longitudinal resolution also quickly decays with the object distance.

    Figure 6.23 shows longitudinal spatial resolution in Fresnel lens induced tomo-

    graphic image as a function of THz frequency. The depth of field of the imaging

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    21/22

    3Dimensional Imaging Using Fresnel Lens 147

    Fig. 6.23 The longitudinal

    resolution in Fresnel lens

    based THz wave tomographic

    imaging system as a function

    of THz wave frequency

    (corresponding to the object

    distance)

    system, provided by Equation (21) of Chapter 3 also affects the longitudinal reso-

    lution of the system The distance between the two distinguishable planes should be

    larger than the depth of field for a better resolution.. If the spatial resolution on the

    imaging plane equals to the Airy disk size and is much smaller than aperture of the

    lens, depth of field of the imager is defined as

    L = 2.44 l2

    D2. (18)

    Equation (18) indicates that the depth of field is also proportional to square of the

    object distance and inversely proportional to the carrier wave frequency.

    Similar to THz DT, image quality of Fresnel lens induced THz tomographic

    imaging is also related to frequency of THz wave. The image quality is best between

    Fig. 6.24 Photo (a) and cw

    THz wave CT image of an air

    freshener

  • 8/12/2019 THz Wave 3D Imaging and Tomography

    22/22

    148 6 THz Wave 3D Imaging and Tomography

    0.5 and 1.25 THz and reduces at higher and lower frequency ends. At the lower fre-

    quency end, the lateral resolution is limited by diffraction of the carrier while the

    longitudinal resolution is limited according to Equation (17). It is also worth notic-

    ing that, since Fresnel lens creates image following the imaging formula in Equation

    (15), the magnification factors will be different for targets with different object dis-tances. This needs to be taken into account in order to correctly retrieve the 3D

    image.

    Above discussions are focused on THz wave tomography techniques utilizing

    pulsed THz wave. THz wave tomographic images can also be obtained using cw

    THz. Figure6.24shows a tomographic image (displayed in 2D) of an air freshner

    case obtained by using the CW Gas Laser at 1.63 THz with a power of 180 mW. The

    case is made of plastic, which is transparent to THz wave. The THz wave tomog-

    raphy image shows both the external pattern of the shell profile and the internal

    structure of the case.

    References

    1. S. Wang, and X.-C. Zhang, Pulsed terahertz tomography, J. Phys. D: Appl. Phys. 37, R1

    (2004).

    2. S. R. Deans,The Radon Transform and Some of its Applications. New York, John Wiley &

    Sons (1983).

    3. B. Ferguson, S. Wang, D. Gray, D. Abbot, and X. C. Zhang, T-ray computed tomography,

    Opt. Lett.27, 1312 (2002).4. M. Born, and E. Wolf,Principles of Optics. Cambridge, Cambridge University Press (1997).

    5. P. M. Morse, and H. Feshbach, Methods of Theoretical Physics. New York, McGraw-Hill

    (1953).

    6. J. C. Russ,The Image Processing Handbook. Michigan, CRC Press (1992).

    7. D. N. Bhat, and S. Nayar, Ordinal measures for image correspondence,IEEE Trans. Pattern

    Anal. Mach. Intell.20, 415 (1998).

    8. G. B. Jin, Y. B. Yang, and M. X. Wu,Binary Optics. Beijing, National Defense Industry Press

    (1998).

    9. S. Wang, T. Yuan, E. D. Walsby, R. J. Blaikie, S. M. Durbin, D. R. S. Cumming, J. Xu, and

    X. C. Zhang, Characterization of T-ray binary lenses, Opt. Lett.27, 1183 (2002).