The X’Celerator, how does it work?

An introduction to the workshop

Page 2: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

2

Introduction

• Since 2001, we shipped almost 1000 X’Celerators

• X’Celerators can be found in almost all countries in the world

• Used in a wide variety of applications …• … and more will surely come

Page 3: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

3

The X’Celerator• 128 parallel strips on a substrate• Real Time Multiple Strip (RTMS) technology• Speed & resolution

• Housing is designed for high 2Theta range

Page 4: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

4

The X’CeleratorModes of operation

• Modes of operation:– Scanning Mode– Receiving Slit Mode– (Static Mode)

• Parameters to be controlled:– Step size & counting time– Active length– (PHD levels)

Page 5: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

5

Scanning mode

active length

Each strip acts as an individual detector

Polycrystalline sample

Page 6: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

6

What is RTMS technology? (2)X’Celerator

Polycrystalline sample

Page 7: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

7

What is RTMS technology? (3)X’Celerator

Polycrystalline sample

Page 8: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

8

What is RTMS technology? (4)X’Celerator

Polycrystalline sample

Page 9: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

9

Step sizes in scanning mode

• Step size is determined by:– strip size– goniometer radius

• Minimum step size is ≈ 0.002º 2θ @ 240 mm radius

Page 10: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

10

Active length in scanning mode

• Software allows reduction of active length• Could reduce defocusing error• Effect is hard to see in practice

Page 11: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

11

Receiving slit mode

Results from strips are added up and give one 2Theta value

“receiving slit”

+

Polycrystalline sample

Page 12: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

12

Receiving slit mode

• Allows recording of pole figures (Phi and Psi scans)

• Useful for alignment of samples etcetera• Latest software versions allow display of count

rate on console display:– choose Stationary Measurement Program– choose appropriate measurement time

Page 13: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

13

Adjustment of the PHD level

• Pulse Height Discrimination is used in the X’Celerator

• Useful for suppression of – higher harmonics – sample fluorescence

Page 14: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

14

Other devices related to the X’Celerator

• Slots in housing for – Soller slits (0.08, 0.04, 0.02, 0.01 rad)– Beta filter

• Fixed anti-scatter slits• Programmable anti-scatter slits• Beam knives• X’Celerator monochromator

– for Cu or Co radiation

Page 15: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

15

The spectrum from a Cu tube

Inte

nsity

Wavelength

Cu Kα

W Lα

Cu Kβ

• Is a an X-ray tube monochromatic?

Cu Kα1 100%Cu Kα2 50%Cu Kβ 25%W Lα <1%

+ white radiation

Page 16: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

16

The beta filter

Inte

nsity

Wavelength

Cu Kα

W Lα

Cu Kβ

Ni filter

• Material with absorption matched to anode material

Ni for Cu tubeFe for Co tube….

Page 17: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

17

Example: NIST SRM 1976 Al2O3

20 30 40 50 60 70 80 90 100 110 120 130

NIST SRM 1976, Al2O340000

Counts

30000

20000

10000

0

Position [°2Theta]

Page 18: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

18

Example: NIST SRM 1976 Al2O3

25 30 35 40

NIST SRM 1976, Al2O3

Al2O3 Kα1 + Kα2

Mullite (Al6Si2O13)

Al2O3 Kβ

Ni absorption edge

40000

Counts

10000

0

Position [°2Theta]

Page 19: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

19

Instruments in which the X’Celerator is used• Diffractometer:

X’Pert PRO or CubiX FAST

X’Pert PRO CubiX FAST

Page 20: The X’Celerator, how does it work?old.vscht.cz/clab/RTG/dokumenty/panalytical/presentace... · 2015. 9. 18. · • Effect is hard to see in practice. 11 Receiving slit mode Results

20

Summary

• Introduction to the themes of the workshop • Overview of X’Celerator modes of operation

and options

• Questions?