The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed....

29
The newest full compliance CE mark test generator July 2014 EMC PARTNER - Largest range of impulse test equipment up to 100kV and 100kA

Transcript of The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed....

Page 1: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

The newest full compliance CE mark test generator

July 2014

EMC PARTNER -

Largest range of impulse test equipment up to 100kV and 100kA

Page 2: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

EMC PARTNER AG

Founded in 1994 Swiss private company, headquarters in Laufen (CH)Largest choice of impulse generatorsMarket leader, reputed worldwideDevelopment, production and testing in houseGlobal representative network

CDN-UTP8 Ed. 3 - Universal CDN for data and telecom lines surge

Largest range of impulse test equipment up to 100kV and 100kA

The Swiss company EMC Partner AG

Page 3: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

Introduction

Application range

Technical specification according to norms

Technology used in IMU 4000

Unique advantages for IMU 4000 users

Full compliance plus benefits

Page 4: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

Introduction

EMC Partner provides

conducted

immunity

test solutions

for

a broad

range

of sectors:

Industry Household

Components

Renewable energy

Avionics

Military

Telecom

Page 5: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

Concise approach: only useful information is presented

Your time will be used successfully to learn about:

• Precise technical specification of IMU 4000 generators

• Benefits of newest technology from the market

• Unique advantages of IMU 4000 users

Page 6: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

Application range of IMU 4000

Application Norm ComplianceESD IEC 61000-4-2 full compliantBurst IEC 61000-4-4 full compliant

Surge 1.2/50 µs IEC 61000-4-5 full compliantMagnetic Field IEC 61000-4-8 full compliantMagnetic Pulse IEC 61000-4-9 full compliant

AC dips, drops, interruptions IEC 61000-4-11 full compliantCommon mode disturbances IEC 61000-4-16 full compliant

Differential mode disturbances IEC 61000-4-19 full compliantDC dips, drops, interruptions IEC 61000-4-29 full compliant

The power of innovation:

9 basic standards to be tested with only one generator

Page 7: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

Equipment to be tested

EUT Type Parameters

AC supplied equipment

1 phase, 250V 50Hz, up to 16A1 phase, 250V 60Hz, up to 16A1 phase, 115V 400Hz, up to 16A

3 phase, 400V, up to 100A3 phase, 480V, up to 200A3 phase, 690V, up to 200A

DC supplied equipment 1000V DC, up to 100A

I/O lines

Sym. low speed data linesUnsym. low speed data lines

High speed telecom linesBus systems

Gigabit Ethernet lines

Page 8: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

ESD - Electrostatic discharge according to IEC 61000-4-2 Ed. 2

Norm requirement IMU 4000 with ESDAir discharge up to 15 kV Air discharge up to 16 kVContact discharge up to 8 kV Contact discharge up to 10 kV

Repetition rate up to 20 Hz

Positive and negative polarity

Pulse or discharge count

Page 9: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

ESD - Electrostatic discharge according to IEC 61000-4-2 Ed. 2

Customer advantages

-Easy programming through large touch screen

-Test report integrated with other norms

-Unbeatable cost of the entire test system.

Page 10: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

Burst – Electrical fast transients according to IEC 61000-4-4 Ed. 3

Norm requirement IMU 4000 with EFTEFT voltage up to 4 kV EFT voltage up to 4.4 kV or

5.1 kV

50 Ω

output

1 MHz burst

frequency

Positive and negative polarity

Page 11: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

Surge – 1,2/50µs and 8/20µs waves according IEC 61000-4-5 Ed. 3

Norm requirement IMU 4000 with CWGSurge voltage up to 4 kV Surge voltage up to 4.1kVSurge current up to 2000 A Surge current up to 2050 A

2 Ω

and 12 Ω

output

Up to 60 pulses/minute

Positive and negative polarity

Page 12: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

Surge – 1,2/50µs and 8/20µs waves according IEC 61000-4-5 Ed. 3

Customer advantages

- IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1 kV solutions

- Unbeatable repetition rate reduces test time and costs.

- Optimal range of external CDNs

covers newest test requirements.

3 phase CDN for EUTs

up to 3 x 690V / 3 x 200A

Page 13: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

Magnetic field – 50/60 Hz magnetic field according to IEC 61000-4-8

Norm requirement IMU 4000Continuous m. field up to 100 A/m Continuous m. field up to 100 A/mShort term m. field up to 1000 A/m Short term m. field up to 1000 A/m

Health and safety

first:

no high currents

from

generator

to the

antenna

Page 14: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

Magnetic field – 50/60 Hz magnetic field according to IEC 61000-4-8

Customer advantages

- Norm levels achievable with the internal 5A variac

- No high current cables from external transformer to the antenna

- Improved safety

- Limitation of human body exposure to magnetic fields during test.

Page 15: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

Impulse Magnetic field according IEC 61000-4-9

Norm requirement IMU 4000

Pulse magnetic field 1000 A/m Pulse magnetic field up to ~2000 A/m

Same antenna

as for

IEC 61000-4-8

Page 16: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

AC dips, interruptions and variations according IEC 61000-4-11

Norm requirement IMU 4000Dips 40, 70 and 80% Dips 40, 70 and 80%Interruptions

0% Interruptions

0%Variations 40, 70% (optional) Variations 40, 70% (optional)

Page 17: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

AC dips, interruptions and variations according IEC 61000-4-11

Customer advantages

-Unique internal variac

for equipment with consumption ≤

5 A.

- 5 A variac

integrated in generator’s case

- 16 A available if required.

Page 18: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

Common mode disturbances up to 150 kHz according IEC 61000-4-16

Norm requirement IMU 4000Frequencies DC, 16.7, 50, 60 Hz Frequencies DC, 16.7, 50, 60 HzTest levels 30V cont. and

300V short Levels 35V cont.

and

330V shortSweep 15 Hz -

150 kHz

with 30 V Sweep 15 Hz -

150 kHz

with 35 V

Page 19: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

Differential mode disturbances 2 - 150 kHz according IEC 61000-4-19

Norm requirement IMU 4000 SMARTDiff. voltage up to 20 V Diff. voltage up to 22 VDiff. current up to 4 A Diff. current up to 5 A

Automatised

test setup, all accessories built in

Reference current source available

Modulation frequencies

selectable

Page 20: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

Differential mode disturbances 2 - 150 kHz according IEC 61000-4-19

Customer advantages

-The first and only full compliance solution available on the market (since May 2014)

-

Solution optimized and tested with electricity meters from worldwide reputed companies.

Page 21: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

DC dips, interruptions and variations according IEC 61000-4-29

EMC Partner is able to deliver a calibrated test system together with EMC Partner DC source(s)

Norm requirement IMU 4000Dips 40 and 70% Dips 40 and 70%Interruptions

0% Interruptions

0%Variations 80, 120% Variations 80, 120%

Page 22: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

Benefits overview

Have you ever seen a test generator capable to meet all these basic standards?

Page 23: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

Benefits overview

We did, the IMU 3000.

Otherwise,

no generator in the world meets so many basic standards.

Page 24: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

IMU 4000 market leading technology

Leading technology means:

Proprietary OS: EMC PARTNER Operating System developed in house

Benefits:

• rapid implementation of new features and requirements

• advanced support capability

• strong feedback response possibility

Page 25: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

IMU 4000 market leading technology

Leading technology means:

Modern GUI optimized for touch screen

Benefits:

• intuitive operation (assisted)

• quick learning and accommodation process

• comfortable and comprehensive operation

Page 26: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

IMU 4000 market leading technology

Leading technology means:

Fully modular: single purpose generators are history

Benefits:

up to 9 basic norms can be tested with IMU 4000 and up to 10 with IMU 3000 -> price advantage.

There is no other generator on the market to achieve this performance.

• Hardware upgrades on site, modules are not screwed.

• Modules can be sent separately for calibration.

Firmware updates can be downloaded directly from the web and installed on site: no unscrewing, no memory cards.

Page 27: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

IMU 4000 market leading technology

Leading technology means:

Latest innovations are now accessible

Pulse synchronisation is performed without an additional wire for 1 phase or 3 phase systems (old generators need an additional input)

Environment parameters (temperature, humidity, air pressure) are measured and included in the test report.

• Up to 7 languages are available for generator’s GUI

Page 28: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

• Up to 9 basic norms with one generator, newest versions always.

• EUT power/voltage

monitoring in the menu

• True on site hardware, firmware and software update

• Built-in or external environment parameters’

measurement

• ESD module as per IEC 61000-4-2

• Common Mode module as per IEC 61000-4-16

• Differential Mode module as per IEC 61000-4-19

Internal variac

for voltage variations up to 5 A and magnetic field up to 1000 A/m

Magnetic field antennas with current transformer attached. Advantages: safe operation and reduction of user’s exposure to magnetic fields.

Unique advantages for IMU 4000 users

Page 29: The newest full compliance CE mark test generator · 2014. 10. 8. · according IEC 61000-4-5 Ed. 3. Customer advantages - IMU series allows choosing between 4.1 kV, 6.6 kV and 8.1

Thank you for AttentionThank you for AttentionEMC PARTNER AGEMC PARTNER AG

Find your

local

partner: www.emc-partner.com/reps

Checkout all products on our website www.emc-partner.com