The Absolute Calibration of the HiRes Detectors J.N. Matthews, S.B. Thomas, N. Manago, L. Perera, G....
-
Upload
phebe-norton -
Category
Documents
-
view
216 -
download
0
Transcript of The Absolute Calibration of the HiRes Detectors J.N. Matthews, S.B. Thomas, N. Manago, L. Perera, G....
The Absolute Calibration of the HiRes
Detectors
J.N. Matthews, S.B. Thomas, N. Manago,
L. Perera, G. Burt, and R. SnowFor the HiRes Collaboration
RXF Calibration
• The width/mean calibration
• Calibration of the Standard Candle
• Comparison of the two techniques
• Conclusion
Energy Spectrum - HiRes & AGASA
AGASA energy scaled by 0.79
Systematic Uncertainties• PMT calibration: 10%
• Fluorescence yield: 10%
• Unobserved energy: 5%
• Atmospheric absorption: most sensitive to vertical aerosol optical depth (VAOD)
– Mean VAOD = 0.04
– VAOD RMS = 0.02
– VAOD systematic is smaller.
– Modify MC and analysis programs to use VAOD = 0.02 and 0.06, reanalyze.
– J(E) changes by 15%
• Total systematic uncertainty = 21%
PMT Clusters/Cameras
16 x 16 PMT Array
•2 Sites – 12 km
•34 Buildings
•64 Cameras
•16,384 PMTs
PMT Calibration
• Roving Xenon Flasher (RXF) as a Standard Candle
• RXF sits at the center of the mirror and illuminates the entire cluster
• Can take it to each PMT cluster at both sites
• Very stable (< 2% over a night)
Current Statistical Method
• The measured QDC distribution is used to calculate the number of photo-electrons for each PMT:
Npe = α (μ/σ)2
• Applying the quantum efficiency then gives us the gain of the PMT or the number of ADC counts per
Applied Calibration Result
SigmaSigma
Problems with the Current Method
• The excess noise factor, α, is not well known. The value was determined by a small number of single photo-electrons measurements and atmospheric measurements (molecular edge). It is consistent with values reported in the literature and discussions with Photonis.
• The QE() is not well understood.
Calibration of the Standard Candle
• A two step approach using NIST calibrated silicon photo-diodes and hybrid photo-diodes (HPD)
• Use the NIST Si photo-diodes to determine the response of the HPD by measuring the single photo-electron peak
• Then use the HPD to measure the luminosity of the RXF (# /mm2 delivered at the mirror-cluster separation distance)
SPE Distribution
P(0)P(0)
Mean: 907.9 ; Events : 509991Mean: 907.9 ; Events : 509991 P(1) 1015.6; Events: 534870P(1) 1015.6; Events: 534870
P(1)P(1)
P(0)P(0)
A Gaussian A Gaussian fit is applied fit is applied to P(1) to to P(1) to obtain the obtain the mean.mean.
RXF Measurement with HPDMean: 907.5 Events: 2275 Mean: 907.5 Events: 2275 Mean: 8412 Events: 3203Mean: 8412 Events: 3203
CalculationsCalculations
(RXF signal – RXF background)(RXF signal – RXF background)
0.183* 0.68 *(SPE Peak – background)*Area0.183* 0.68 *(SPE Peak – background)*Area
HPD Calibration ConstantsHPD Calibration Constants
• 0.183 is the HPD efficiency0.183 is the HPD efficiency• 0.68 is a distribution correction factor for the HPD. It is 0.68 is a distribution correction factor for the HPD. It is the ratio of the SPE mean over the SPE peak. the ratio of the SPE mean over the SPE peak.
γγmmmm22 ==
HPD Systematics• Si diode current: I = 0.192*10-9 A ± 1% ± 2.6%• Si diode responsivity: β = 0.11325 A.s/J ± 5% ± 0%• Si diode area: A = 100mm2 ± 1% ± 0%• HPD count rate: μ = 16.22*103 counts/sec ± 5% ± 0%• Counting Efficiency: ε = 1.054 PE/count ± 3.5% ± 0%• Neutral density filter attenuation: η = 1.36*104 ± 1.5% ± 0%• Geometrical and other effects: ±5% ± 0%
• HPD Efficiency = 11.2 ± 9.6%Syst ± 2.6%Stat (# /pe/mm2)
A Test:Measure and compare the light output of the RXF via:
The HiRes detector width/mean methodThe HPD traceable to NIST
The Plan:• Measure RXF with the HPD in the University of
Utah lab• Take RXF to HiRes to calibrate the detector.
Used 20 cameras at HiRes-I for this test. Ten minutes (about 900 shots) of data are recorded at each detector.
• Return to the U and remeasure RXF in the lab.
Result:
0
2
4
6
8
10
12
Statistical Method:
8.97 /mm2 ± 10% (Syst)
HPD Method:
9.55 /mm2 ± 9.6% (Syst) ± 2.6% (Stat)
Comparison percentage is 6.5%.
Well within expected values.
Conclusions
• The NIST Si-PD and HPD calibration will allow us to absolutely calibrate the RXF and thus the HiRes PMT Clusters.
• Good agreement with the statistical method.
• We need to work on the uncertainties to get them down. Should be able to get to the ~5% level.
The High Resolution Fly’s Eye (HiRes) Experiment
• University of Utah• Rutgers University• Columbia University• University of Montana• University of New
Mexico• University of Adelaide• Los Alamos National
Laboratory (LANL)• University of Tokyo