1 Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC 1 Progress in calculations of k Q for TG-51 D. W. O. Rogers Carleton Laboratory for Radiotherapy Physics, Physics Dept, Carleton University Ottawa http://www.physics.carleton.ca/~drogers AAPM Charlotte,NC, Aug 2, 2012

Transcript of Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte...

Page 1: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

1Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

1

Progress in calculations of kQ for TG-51

D. W. O. RogersCarleton Laboratory for Radiotherapy Physics,

Physics Dept, Carleton University

Ottawa

http://www.physics.carleton.ca/~drogers

AAPM Charlotte,NC, Aug 2, 2012

Page 2: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

2Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

2

2/34

Acknowledgements

• much of this talk is based on the work of Bryan Muir, a graduate student, who has worked with me over the last few years

.

Page 3: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

3Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

3/34

TG-51 for photon beams

defines: chamber’s absorbed dose calibration coefficient

defines kQ: chamber specificbeam quality conversion

factor

-accounts for ND,w variation with Q

The addendum about to be published is exclusively about photon beams.

Work is proceeding on electron beams.

3

Page 4: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

4Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

4/34

Equation used by TG-51 for kQ

This eqn assumes (W/e)air (relating charge measured to dose to the air in cavity) is

independent of beam quality

For a detailed derivation, see Ch 9 in 2009 AAPM Summer School book

Ch 9 is available on my home page http://www.physics.carleton.ca/~drogers

4

Page 5: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

5Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

5/34

Status of TG-51 kQ photon calcs

• spr is OK, even in FFF beams• ratios of Pwall values wrong by up to 0.5%• Pfl value of unity assumed OK• ratios of Pgr values wrong by up to 0.2%• Pcel values wrong by up to 3% for high-Z

electrodes

Not bad except for higher-Z electrode chambers.They have other problems for reference dosimetry and are not recommended for use.

5

Page 6: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

6Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

6/34

Pcel: electrode correction

• TG-51 corrects for chambers with Al central electrodes

• more recent, more precise calculations agree with values TG-51 used for Al electrodes in filtered beams

• even more recent calculations for higher-Z electrodes show major effects • Pcel effects much larger• Pcel in FFF (flattening filter free) beams

even larger effects

SPC Silver Plated Copper Covered Steel Exradin A14,T14,A14SL,A16 or just steel PTW 31006, CC01

6

Page 7: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

7Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

7/34

Pcel: state-of-the-art

Muir and Rogers Med Phys 38(2011) 1081-1088

•high Z => much larger effect

•FFF beams => even larger effects

31010 Al electrode much larger fraction of chamber volume 12% vs 2% for NE2571 => much bigger effect even is same radius roughly.Note NE2571 small FFF effect is thereThe large effects for higher-Z⇒standard TG-51 kQ calculations do not work unless use these detailsThese results means high-Z electrodes must be used with care in all FFF beams.

SPC means sliver plated copper covered steel electrodes.

7

Page 8: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

8Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

8/34

Which path forward?

• rework TG-51 analytic calculations accounting for all the new data?

• base kQ values on measured values– McEwen published an extensive set of

values in 2010 (Med Phys 37(2010) 2179)

• do ab initio Monte Carlo calculations of kQ

-McEwen covered 27 different chamber models – how do we handle new chambers?-reworking analytic would be based on multiple MC calc– complex uncertainty analysis

8

Page 9: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

9Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

9/34

ab initio Monte Carlo calculations

• EGSnrc has been shown to calculate doses in an ion chamber within 0.1% relative to its own cross sections (Fano test)

• egs_chamber code of Wulff et al (Med Phys 35 (2008) 1328)– very efficient: correlated sampling– handles complex realistic geometries

A12 NE2571

Two significant advances in Monte Carlo since TG-51 written: EGSnrc and egs_chamber (+ much faster computers)

Sheath on NE2571 since not waterproof

9

Page 10: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

10Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

10/34

Calculating kQ with Monte Carlo

• definitions:

assume (W/e)is independent

of beam quality

-first done more than 5 yr ago but only with Wulff’s egs_chamber did statistical precision get sub -0.1%

The only important eqn on this slide is highlighted in yellow. This is what is used for calculations.-can use correlated sampling at each Q separately or in 4 separate runs (water calcs are common to all chambers)

10

Page 11: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

11Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

11/34

9 different “classes” of detectors

black:TG51gold: fit

labels:(wall/electrode)

Note large effects of

high-Z electrodes

The calculations are for WFF beams (i.e. with flattening filters) -FFF big issues for high-Z electrodesOverall good agreement for B-F chambers with TG-51 except for NE2571 gr/AlSPC means sliver plated copper covered steel

11

Page 12: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

12Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

12/34

fits to kQ

• rms deviation: less than 0.1 % for 10 WFF beams, except for 1 chamber (A14, 0.2%, very small volume, SPC electrode).

• a,b,c tabulated in paper and report (also as a function of TPR)

SPC sliver plated copper covered steel

12

Page 13: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

13Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

13/34

Uncertainties on calculated kQ

• EGSnrc is accurate to 0.1 % against its own cross sections (Kawrakow, Med Phys 27(2000) 499

• what are effects of cross section uncertainties?

• what is uncertainty on (W/e)air being constant?– TRS-398 says 0.5% but evidence for any value

is very thin

Page 14: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

14Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

14/34

standard error propagation, assuming uncorrelated

where u(xi) is uncertainty on cross section xiApproximate

where is change in kQ when cross section i is changed by . Calculate for acorresponding to u(xi).

Cross section uncertainties on kQ

The final eqn, which is all that matters, holds for absolute or percentage differences.

Derivation is in Muir and Rogers MP 37(2010) 5939.First applied this way by Wulff et al, PMB 55(2010) 4481

14

Page 15: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

15Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

15/34

NE2571 kQ uncertainty components

All correlated means that all photon cross sections are increased by 1%. in which case kQ does not change.

In related work by Ali and Spencer in my lab, it is shown that the 1% uncertainty on the cross sections is a very conservative estimate.

Page 16: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

16Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

16/34

NE2571 kQ uncertainties (cont)

The uncertainty shown here on W/e is the estimate from TRS-398. Below a better estimate is given as 0.25%, but this table is from a 2010 paper.

For comparisons to experiment discussed in next few slides, we used the correlated, no W/e uncertainty estimate and still get chi-squared values less than 1.

Page 17: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

17Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

17/34

Uncertainties on kQ for all chambers

Muir & Rogers Med Phys 37 (2010) 5939

worst case: 0.39% 0.86% 0.63% 0.99%

These are uncertainties on calculated kQ

17

Page 18: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

18Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

18/34

Experimental measurements of kQ

• many measurements done for one or two types of chambers

• McEwen measured kQ for 27 different types against Canadian primary standards of absorbed dose---->

(Med. Phys. 37 (2010) 2179)

• for “well-behaved” chambers uncertainty on kQwas 0.30%

• agreement with TG—51 values is excellent, typically 0.5% or better for “well-behaved”

Well-behaved means the chambers met the specifications for use with TG-51 as given in the addendum. These are related to Pion, Ppol, reproducibility and stability criteria. Basically, it excludes all very small volume chambers which are the ones with high-Z electrodes.

Page 19: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

19Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

19/34

Consistency of measured kQ

NE2571

Muir et al Med Phys 38 (2011) 4600

diamonds are from

standards labs (Stuckiet al, to be published) 0.3%

note good agreement between calns and fit to meas

19

Page 20: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

20Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

20/34

How well do calculations and measurements agree?

For 26 chambers in common,-χ2 /df < 0.65 for all chambers at 1 energy

-χ2 /df < 1 for all chambers vs energy except 1

Suggests, if anything, uncertainties are too largehttp://www.physics.carleton.ca/clrp/kQ

-df is number of degrees of freedom-s2

c is for correlated, no W/e uncertaintyhttp link has a report with plots of comparisons for each individual chamber

20

Page 21: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

21Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

21/34

Measured vs calculated kQ

26 chambers in common

shaded part is less precise

chambers

remarkable agreement

-ve calc are smaller than measuredWhat does this tell us?

-agreement remarkable-even single chambers representative-no massive change in W/e: MC assumes

none, meas make no assumptions

21

Page 22: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

22Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

22/34

• assume some variation of (W/e)air

• in this case:

• now we have

• calculate χ2 as before as function of α

Can we use this agreement to set a limit on the variation of (W/e)air ?

Page 23: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

23Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

23/34

What is the value of α ?

α = 1.0024±0.0011Conservatively one can say

W/e is constant within 0.29% (0.42%) with 68% (95%) confidence

chisqmin/df = 6.3/17 =0.38 for the reference chambers

23

Page 24: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

24Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

24/34

What about parallel-plate chambers?

• not allowed in TG-51 for photon beams because of a lack of Pwall data

• Muir et al (Med Phys 39(2012)1618) provides a complete set of calculated and measured kQvalues (which agree with rms deviation <0.36%)– unfortunately, unacceptable variations in ND,w

were observed although kQ was unchanging=> still not recommended (although with a cross

calibration technique, a protocol could be devised)

ND,w varied by up to 1.5%

Page 25: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

25Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

25/34

defines kR50: component of kQ which is independent of Pgr, the gradient at point of measurement.

What about electron beams?

defines kecal: chamber specific photon-electron conversion factor -Qecal an arbitrary e- energy

defines k’R50: chamber specific electron quality conversion factor

kecal accounts for ND,w variation between 60Co and Qecal

k’R50 accounts for ND,w variation between Qecal and R50

For e- beams Pgr varies for a given beam quality, R50,

=> must be explicitly found for each beam

Page 26: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

26Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

26/34

What about electron beams?

Much more complicated

Paper TU-A-BRB-11 (Muir et al) discussed an extensive set of calculations and measurements

for these electron beam factors.

Once primary standards in e- beams are in place, more direct measurements are possible.

In the meantime, we already know kecal factors need to change.

Page 27: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

27Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

27/34

Pwall for parallel-plate chambers in 60Co

EGSnrc results supersede EGS4 results used in TG-51

kecal values will decreasesince)

Full MC calculations may change this

Note the EGSnrc Pwall values are systematically up to 1% larger than the EGS4 values used in TG-51.

The simple formula for kecal makes use of the fitted eqn for stopping-power-ratios whereas a full Monte Carlo caln (similar to the kQ calns) would include this.

Full MC calns are needed for cylindrical chambers as well.The Pgr correction makes the analysis more complex.

Page 28: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

28Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

28/34

Monte Carlo calculated kR50

PTW Roos plane-parallel

Muir et al, COMP meeting 2012

preliminary

results

The ‘realistic’ calculations use a full BEAMnrcsimulation as the beam source.

The agreement with TG-51 for this simple chamber is quite good.

The `corrected’ TG-51 curve uses the know problems with Pwall, Prepl and spr from literature values to correct the original TG-51 curve.The next few figures come from a paper presented by Bryan Muir in the Young Investigators Symposium at the COMP meeting in Halifax in July, 2012.

Page 29: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

29Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

29/34

Monte Carlo calculated k’R50

Muir et al, COMP meeting 2012

preliminary

results

PTW Roos plane-parallel

This basically takes out the value of kecal

Page 30: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

30Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

30/34

k’R50 vs literature values

Muir et al, COMP meeting 2012

preliminary

results

PTW Roos plane-parallel

The Zink and Sempau results are other Monte Carlo calculations. The Cojocaru and McEwen results are measured values.

Page 31: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

31Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

31/34

k’R50 various chambers

Muir et al, COMP meeting 2012

preliminary

results

These are preliminary results but the good news is that TG-51 values look fairly good.

Page 32: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

32Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

32/34

Conclusions

• Monte Carlo calculations of kQ are feasible

• experimental agreement is exceptional– 0.13 % mean difference for 26 chambers– 0.31 % RMS deviation for 26 chambers

• uncertainty on calculated kQ values is between 0.40 and 0.49 % depending on wall material

• uncertainty on variation of (W/e)air from Co to 25 MV is 0.25 % (68 % limit)

Note the limit on W/e includes the data from Muir et al, Med Phys 39(2012)1618 which did a similar analysis as presented above, but including the calculations and measurements for the parallel-plate chambers.

32

Page 33: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

33Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

33/34

Conclusions (cont)

• results apply only to filtered beams- with low-Z electrodes, results still apply- with Z > 13 electrodes, values will not hold in

FFF beams (OK in WFF beams if reference quality)

• work on electron beams is on-going• it is more complex because of effective point

of measurement issues• changes are mostly expected to be small

final uncertainty assume correlated uncertainties in photon cross sectionsFFF=>Flattening Filter FreeWFF=> with flattening filter

33

Page 34: Th-D-BRCD-1 Rogers TG-51 Addendum 1 AAPM 2012 Charlotte NCamos3.aapm.org/abstracts/pdf/68-19868-234349-85790.pdf · 2012. 7. 28. · Th-D-BRCD-1 Rogers TG-51 Addendum 5 AAPM 2012

34Th-D-BRCD-1 Rogers TG-51 Addendum AAPM 2012 Charlotte NC

34/34

Acknowledgements

-as mentioned before, much of the work was done by Bryan Muir and the experiments in conjunction with Malcolm McEwen

-work supported by an NSERC CGS, OGSSTs, the CRC program, an NSERC DG, CFI and OIT