TFA-1™ Disposable Forehead Sensors
> The forehead site may provide oxygen saturation measurements that are less susceptible to peripheral perfusion changes
> In a study comparing response rate, forehead sensors were found to have faster detection of desaturation and resaturation compared to digit sensors1
> Offers easy access during surgery, resuscitation, and in patients with finger deformities, or when digit access is not possible
© 2017 M
asimo. A
ll rights reserved.
6022
0/PL
M-1
0380
B-01
17
Masimo InternationalTel: +41 32 720 [email protected]
Masimo U.S.Tel: 1 877 4 Masimo [email protected]
LNCS TFA-1 Sensor
The TFA-1 Sensor offers:
> Faster detection of desaturation events> Faster indication of resaturation2
A
B
0 20 6040 80
Oxy
gen
Satu
ratio
n (%
)
Time (seconds)
TFA-1 Sensor with Masimo SET®
DCI® Digit Sensor with Masimo SET®
85
86
87
88
89
84
82
81
80
83
79
TFA-1 Disposable Forehead Sensors
Specifications
ACCURACY (ARMS)3, 4
Oxygen Saturation (SpO2%) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70%—100%No Motion (Adults, Pediatrics) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2%Low Perfusion (Adults, Pediatrics) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2%
Pulse Rate (bpm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25—240 bpm
No Motion (Adults, Pediatrics) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 bpm Low Perfusion (Adults, Pediatrics) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 bpm
COMPATIBILITY
Masimo or OEM monitors with Masimo SET® or rainbow SET™
WEIGHT RANGE
LNCS TFA-1 (Adult/Pediatric) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . >10 kg, Forehead Application
ORDERING INFORMATION
Single-patient-use / Non-sterile / Does not contain natural rubber latexPackaged 10 per box / Headband Included
TFA-1 Disposable Sensors Part NumberLNCS TFA-1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3858
Caution: Federal (USA) law restricts this device to sale by or on the order of a physician . See instructions for use for full prescribing information, including indications, contraindications, warnings, and precautions .
1 Redford DT et al. Anesth Analg, 2004;98(2S):S-94. 2 Data on file. 3 ARMS accuracy is a statistical calculation of the difference between device measurements and reference measurements. Approximately two-thirds of the device measurements fell within ± ARMS of the reference measurements in a controlled study. 4 SpO2 accuracy has been validated on healthy adult male and female volunteers with light to dark skin pigmentations in the range of 70 to 100% against a laboratory CO-Oximeter. Pulse Rate accuracy has been validated over the range of 25–240 bpm in bench top testing against a Biotek Index2 simulator. Low perfusion SpO2 and pulse rate accuracy has been validated with a signal strength of 0.02% over a range of 70% to 100% SpO2 in bench top testing against a Biotek Index2 simulator.
> Available with the LNCS® connector and a flat cable for improved patient comfort
> Weight range allows use on both pediatric and adult patients
Faster Response to Oxygenation Changes
Desaturation of a Healthy Subject