Test Report issued under the responsibility of:
TEST REPORT IEC 61010-031
Safety requirements for electrical equipment for measurement, control, and laboratory use
Part -031: Safety requirements for hand-held probe assemblies for electrical measurement and test
Report Reference No......................: GZ11051010-1
Date of issue.....................................: 27 May 2011
Total number of pages....................... 31
CB Testing Laboratory...................: Intertek Testing Services Shenzhen Ltd. Guangzhou Branch
Address.............................................: Block E, No.7-2 Guang Dong Software Science Park, Caipin Road, Guangzhou Science City, GETDD, Guangzhou, China
Applicant’s name ............................: Precision Mastech Enterprises(Hong Kong) Limited
Address.............................................: Unit 1901, Yen Sheng Centre, 64 Hoi Yuen Road, Kwun Tong, Kowloon, Hong Kong
Test specification:
Standard ...........................................: EN 61010-31:2002 (First Edition) + Amd 1:2008
Test procedure .................................: LVD
Non-standard test method…………..: N/A
Test Report Form No......................: IEC 61010_031C
Test Report Form(s) Originator ........: KTL (Korea Testing Laboratory)
Master TRF.......................................: 2008-08
Copyright © 2008 IEC System for Conformity Testing and Certification of Electrical Equipment (IECEE), Geneva, Switzerland. All rights reserved. This publication may be reproduced in whole or in part for non-commercial purposes as long as the IECEE is acknowledged as copyright owner and source of the material. IECEE takes no responsibility for and will not assume liability for damages resulting from the reader's interpretation of the reproduced material due to its placement and context.
If this Test Report Form is used by non-IECEE members, the IECEE/IEC logo and the reference to the CB Scheme procedure shall be removed.
This report is not valid as a CB Test Report unless signed by an approved CB Testing Laboratory and appended to a CB Test Certificate issued by an NCB in accordance with IECEE 02.
Test item description .................... : Test Lead
Trade Mark .......................................: Mastech
Manufacturer.....................................: Precision Mastech Enterprises(Hong Kong) Limited
Model/Type reference.......................: T3030C
Ratings..............................................: 1000 V CAT III 600 V CAT IV, MAX.10 A
Page 3 of 31 Report No. : GZ11051010-1
TRF No. IEC61010_031C
Summary of testing:
Tests performed (name of test and test clause):
All applicable test items
Testing location:
Block E, No.7-2 Guang Dong Software Science Park, Caipin Road, Guangzhou Science City, GETDD, Guangzhou, China
Summary of compliance with National Differences:
None
Copy of marking plate
The following Markings are molded in the probe body:
Page 4 of 31 Report No. : GZ11051010-1
TRF No. IEC61010_031C
Test item particulars.................................................. :
Type of item tested ...................................................... : Measurement
Description of equipment function............................... : The unit is only test probe for measurement
Classification ................................................................ : Type A
Protection class............................................................ : II
Measurement category................................................ : III, IV
POLLUTION DEGREE ....................................................... : 2
Environmental rating..................................................: standard
Operating conditions..................................................: continuous
Overall size of the equipment (W x D x H) ................: 1220 mm
Mass of the equipment (kg) .......................................: 0.075
Marked degree of protection to IEC 60529 ...............: Ordinary equipment
Possible test case verdicts:
- test case does not apply to the test object................. : N/A
- test object does meet the requirement ...................... : P (Pass)
- test object does not meet the requirement ................ : F (Fail)
Testing.......................................................................... :
Date of receipt of test item............................................ : 18 May 2011
Date (s) of performance of tests................................... : 18 May 2011 – 25 May 2011
General remarks:
The test results presented in this report relate only to the object tested. This report shall not be reproduced, except in full, without the written approval of the Issuing testing laboratory. "(see Enclosure #)" refers to additional information appended to the report. "(see appended table)" refers to a table appended to the report. Throughout this report a point is used as the decimal separator.
Page 5 of 31 Report No. : GZ11051010-1
TRF No. IEC61010_031C
General product information:
The apparatus is a component, it shall be used with relevant measure apparatus.
TABLE: 1 - Documents attached to this report
Document No. Document description Page Numbers
Attachment Photo 2
Page 6 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
TABLE: 3 - List of components and circuits relied on for safety P
Unique component reference or location
(including drawing reference if required)
Application/Function Manufacturer (NOTE 1)
Part number RATING (NOTE 2)
Evidence of acceptance
(NOTE 3)
Plastic enclosure of probe body
Probe body SILVER AGE ENGINEERING PLASTICS (DONGGUAN) CO LTD
730 PVC, V-0, 50℃ UL and tested with appliance
Lead wire Cable Dongguan Huayi Mastech Co., Ltd.
1803 18AWG, 80 ℃, 2000V, PVC insulation
E 229772 and tested with appliance
NOTE 1 - List all manufacturers concerned. NOTE 2 - Electrical, mechanical, flammability, etc. NOTE 3 - Licence number, file number or other documentary evidence of acceptance
Page 7 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
5 MARKING AND DOCUMENTATION P
5.1 Markings P
5.1.1 Markings applicable for whole probe assembly not located on operator removable parts
Markings molded in the probe body
P
Letter symbols (IEC 60027) used P
Graphic symbols (Table 1) used; or Symbol used
P
if other symbol used; explained in accompanying documentation
N/A
In case of less space for required markings: N/A
- symbol 10 of table 1 used N/A
- all necessary information included in documentation
N/A
5.1.2 Identification P
5.1.2 a) Name or registered trademark Mastech P
5.1.2 b) For type B and C, also model no. or similar Type A N/A
If designed for use with specific model this is made clear and
N/A
model identified by marking or in documentation N/A
5.1.3 Fuses No fuse employed N/A
All details necessary for fuse replacement N/A
Includes rated voltage and current breaking capacity
N/A
If selected according to particular application; marked with symbol 10 and information in documentation
N/A
5.1.4 Necessary identification for TERMINALS, connectors etc
N/A
5.1.6 Rating P
Maximum RATED voltage to earth 1000V CATIII 600V CATIV P
(CAT I) Symbol 10 used N/A
(CAT II-IV) Category marked CAT III, CAT IV P
Nature of voltage (ac, dc etc.) Applicable to both r.m.s and dc
N/A
Reference connector intended for connection to voltages exceeding the values of 6.3.1.1
No reference connector N/A
Page 8 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
For type A and type D only, the maximum RATED current unless specified for high impedance inputs
MAX. 10 A molded in the probe body
P
5.2 Warning markings P
Visible when ready for NORMAL USE P
If necessary marked with symbol 10 P
Near or on particular parts of the PROBE ASSEMBLY P
Advise to disconnect or isolate during access to HAZARDOUS LIVE parts or
N/A
marked with symbol 10 and information in the instruction manual
N/A
Easily touched heated parts, if not self-evident, marked with symbol 9
N/A
5.3 Durability of markings P
The required markings are clear and legible (NORMAL USE)
see Form A.3 P
Resist cleaning (clear, legible and not worked loose)
P
5.4 Documentation N/A
5.4.1 General N/A
5.4.1 a) Technical specification N/A
5.4.1 b) Instructions for use N/A
5.4.1 c) Name and address of manufacturer or supplier N/A
5.4.1 d) The information specified in 5.4.2 to 5.4.4 N/A
A clear explanation of warning symbols is in the documentation or
N/A
Information is durably and legibly marked on the equipment
N/A
Statement that symbol 10 means documentation needs to be consulted
N/A
5.4.2 Ratings N/A
Maximum voltage RATING N/A
Maximum current RATING N/A
Statement of the range of environmental conditions N/A
Page 9 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
5.4.3 Operation N/A
5.4.3 a) Identification of operating controls N/A
5.4.3 b) Interconnection requirements N/A
Specification of accessories, materials etc N/A
5.4.3 c) Specification of intermittent operation limits N/A
5.4.3 d) Explanation of required and used symbols N/A
5.4.3 e) Replacement of consumables N/A
5.4.3 f) Definition of measurement category (if marked with CAT)
N/A
5.4.3 g) If marked CAT I, a warning not to use in other CAT N/A
5.4.3 h) Cleaning if necessary N/A
5.4.3 i) Warning for the lower CAT of a combination of a PROBE ASSEMBLY and an accessory
N/A
A statement against use in a manner not specified by the manufacturer
N/A
5.4.4 Maintenance N/A
Sufficient preventive maintenance and inspection for RESPONSIBLE BODY
N/A
Parts to be supplied or examined by the manufacturer only
N/A
RATING and characteristics of fuses (see 5.1.3) N/A
6 PROTECTION AGAINST ELECTRIC SHOCK P
6.1 General see Form A.4 P
6.1.1 Exceptions P
6.1.1 a) Parts intended to be replaced by the operator (for example, fuses), but only if they have a warning marking according to 5.2
N/A
6.1.1 b) PROBE TIPS, provided that they meet the requirements of 6.4.4
Refer to clause 6.4.4 P
6.2 Determination of ACCESSIBLE parts N/A
According to figure 3 Obvious to determine the accessible parts
N/A
6.3 Permissible limits for ACCESSIBLE parts P
Measurements performed according to figure 4 P
6.3.1 Values in NORMAL CONDITION see Form A.6 P
6.3.2 Values in SINGLE FAULT CONDITION see Form A.7 P
Page 10 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
6.4 Insulation requirements for protection against electric shock P
6.4.1 Connectors N/A
6.4.1 a) Connectors in fully mated position: No such part N/A
i) Connecting probe to measuring equipment insulated by at least basic insulation
N/A
ii) Intended to be HAND-HELD insulated by DOUBLE or REINFORCED INSULATION
N/A
6.4.1 b) Connectors in partially mated position: N/A
insulated by at least BASIC INSULATION N/A
Voltage test with test finger (B.1) N/A
6.4.1 c) Connectors in unmated position: N/A
Except for locking or screw-held type connectors or limited current by PROTECTIVE IMPEDANCE:
N/A
i) HAZARDOUS LIVE parts not ACCESSIBLE N/A
Up to 1 kV a.c. or 1.5 kV d.c., not ACCESSIBLE N/A
Above 1 kV a.c. or 1.5 kV d.c., voltage test with test finger
N/A
ii) Stackable connectors N/A
HAZARDOUS LIVE parts separated by BASIC
INSULATION from ACCESSIBLE parts N/A
CLEARANCE and CREEPAGE meet the requirements for BASIC INSULATION
N/A
Voltage test in acc. to 6.6 N/A
6.4.2 HAND-HELD parts other than connectors P
HAZARDOUS LIVE parts separated by DOUBLE or REINFORCED INSULATION from ACCESSIBLE parts
see Form A.4 P
CLEARANCE and CREEPAGE meet the requirements for DOUBLE or REINFORCED INSULATION
see Form A.9 P
Voltage test in acc. 6.6 (specify parts) see Form A.10 P
REFERENCE CONNECTOR N/A
6.4.3 Cables P
RATED for maximum voltage and current P
DOUBLE or REINFORCED INSULATION based on voltages (min 125 V/500 V) according to type of PROBE ASSEMBLIES ..................................................:
N/A
or for maximum RATED voltage ...............................: Rated 1000V CATIII 600V CATIV
P
Page 11 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
Voltage test in acc. 6.6 (specify parts) see Form A.10 P
6.4.4 PROBE TIPS P
BARRIER providing safe distance: P
- CLEARANCE and CREEPAGE meet the requirements for REINFORCED INSULATION
see Form A.9 P
Spring-loaded squeeze PROBE ASSEMBLIES: (rated for WORKING VOLTAGE 1 kV)
No such part N/A
a) Actuation prevents touching HAZARDOUS LIVE parts
N/A
b) Additional protective distance of 45 mm longer than for barrier
N/A
Crocodile clips and similar without barrier: (rated for CAT I or II)
N/A
- have tactile indication N/A
6.4.5 DOUBLE INSULATION and REINFORCED INSULATION P
See 6.5, 6.6 and 6.7.2 P
6.4.6 PROTECTIVE IMPEDANCE N/A
Appropriate HIGH-INTEGRITY single component used for protection (see 12.3)
No such component N/A
Components, wires and connections are suitably RATED even for SINGLE FAULT CONDITION
N/A
6.5 CLEARANCES AND CREEPAGE DISTANCES P
CLEARANCES and CREEPAGE DISTANCES between circuits and parts
see Form A.4 and Form A.9 P
6.6 Voltage tests P
Humidity pre-conditioning (6.6.2) conducted P
Test voltages (6.6.4) see Form A.4 and Form A.10 P
6.7 Constructional requirements P
6.7.1 General N/A
6.7.1 a) Security of soldered wiring connections N/A
6.7.1 b) Screws securing removable covers are captive if their length affects isolation distances
N/A
6.7.1 c) Accidental loosening N/A
The following is not used for safety purposes: P
1) Materials which can be easily damaged (enamel etc)
P
Page 12 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
2) Non-impregnated hygroscopic materials P
6.7.2 ENCLOSURES of PROBE ASSEMBLIES with DOUBLE or REINFORCED INSULATION P
ENCLOSURE which surrounds all metal parts P
Small metal parts are separated from HAZARDOUS
LIVE voltages by DOUBLE or REINFORCED INSULATION
No such part N/A
ENCLOSURES or parts made of insulating material fulfil requirements for DOUBLE or REINFORCED
INSULATION.
see Form A.4 and Form A.9 P
Protection for metal ENCLOSURES or parts is provided by one of the following:
N/A
a) provision of an insulating coating or BARRIER on the inside of the ENCLOSURE
N/A
b) CLEARANCES and CREEPAGE DISTANCES cannot be reduced by loosening of parts or wires
N/A
6.7.3 Corona and partial discharge N/A
No corona or partial discharge while operating at maximum voltage
N/A
6.7.4 Cable attachment P
Withstand forces likely to be encountered P
6.7.4.1 Pull test see Form A.11 P
6.7.4.2 Flexing/pull test see Form A.11 P
6.7.4.3 Rotational flexing test see Form A.11 P
6.7.5 Insulation of a probe cable P
Probe cable with a wear indicator provide DOUBLE or REINFORCED INSULATION when new, and at least BASIC INSULATION when the wear indicator is reached
Wear indicator employed P
PROBE CABLE without a wear indicator provide DOUBLE or REINFORCED INSULATION
N/A
Voltage test in acc. 6.6 (specify parts): see Form A.10 P
- REINFORCED INSULATION: one unconditioned sample before cycling treatment
P
- BASIC INSULATION: contrasting colour became visible during the cycling treatment
P
- REINFORCED INSULATION: 250 cycles treatment without contrasting colour becoming visible.
N/A
7 PROTECTION AGAINST MECHANICAL HAZARDS P
Handling during normal use shall not lead to hazard P
Page 13 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
8 MECHANICAL RESISTANCE TO SHOCK AND IMPACT P
Withstand shock and impact likely to occur in NORMAL USE
P
8.1 Rigidity test P
20 N applied three times P
8.2 Drop test P
Three samples dropped P
8.3 Impact swing test P
Probe subjected to impact against a hardwood board
P
After the tests of 8.1 to 8.3: P
Voltage tests in acc. to 6.6 (see Form A.10) P
Inspections: P
8a) HAZARDOUS LIVE parts not accessible P
8b) ENCLOSURE shows no cracks (hazard) P
8c) CLEARANCES not less than their permitted values (see Form A.9) P
8d) BARRIERS not damaged or loosened P
8e) No damage which could cause spread of fire P
9 TEMPERATURE LIMITS AND PROTECTION AGAINST THE SPREAD OF FIRE P
9.1 General P
Any heating does not cause a HAZARD in NORMAL
CONDITION nor in SINGLE FAULT CONDITION P
No spread of fire outside the PROBE ASSEMBLY P
Easily touched surfaces not exceeding the following limits in NORMAL CONDITION :
P
- metal less than 55 C No such part N/A
- non-metallic less than 70 C P
- wires and cables less than 75 C P
Temperatures in SINGLE FAULT CONDITION less than 105 °C
N/A
Easily touched heated surfaces recognizable or marked with symbol 9 of table 1 (s. 5.2), if necessary for functional reasons
No such part N/A
Circuits separated by at least by BASIC INSULATION, if protection depends on separation of circuits
N/A
9.2 Temperature tests see Form A.12 P
Page 14 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
10 RESISTANCE TO HEAT P
10.1 Integrity of CLEARANCES and CREEPAGE DISTANCES P
Requirements of 6.5 are met at an ambient temperature of 40 °C of maximum RATED ambient temperature (if higher)
see Form A.9 P
10.2 Resistance to heat P
Probe assemblies with non-metallic ENCLOSURES are resistant to elevated temperatures:
see Form A.13 P
11 PROTECTION AGAINST HAZARDS FROM FLUIDS N/A
11.1 General N/A
OPERATOR and surrounding area are protected against HAZARDS from fluids if PROBE ASSEMBLIES
containing or intended to be used with fluids
No fluid employed N/A
11.2 Cleaning N/A
Cleaning procedure applied three times to the PROBE ASSEMBLY
N/A
11.3 Specially protected PROBE ASSEMBLIES N/A
Where the equipment is RATED or marked by the manufacturer the requirements of IEC 60529 are fulfilled
N/A
After the tests of 11.1 to 11.3: N/A
Accessible parts do not exceed the limits of 6.3.1 N/A
Voltage tests in acc. to 6.6 N/A
12 COMPONENTS P
12.1 General P
Safety components operated within their specified RATINGS
see Table 3, probe body and cable
P
Components approved by a recognized testing authority for conformity
see Table 3 P
Those components comply with one of the following :
N/A
12.1 a) comply with all applicable safety requirements in relevant IEC standards
N/A
and subjected to the tests of this standard if necessary for application
N/A
12.1 b) comply with all relevant requirements of this standard
N/A
Page 15 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
and subjected to the tests of relevant IEC component standard if necessary for application
N/A
12.1 c) comply with all relevant requirements of this standard only if there is no relevant IEC standard
P
12.2 Fuses N/A
Voltage RATING No fuse N/A
Breaking capacity and current rating................... : N/A
12.3 HIGH-INTEGRITY components N/A
Positions of use No such component N/A
Evaluated to IEC Publications N/A
A single electronic device which employs electron conduction in a vacuum, gas or semiconductor is not used as HIGH-INTEGRITY component
N/A
12.3.1 Resistors used in PROTECTIVE IMPEDANCE N/A
12.3.1 a) Withstand twice the dissipation at RATED voltage N/A
12.3.1 b) Withstand twice the RATED voltage for 1 s N/A
12.3.1 c) Distance across resistor or assembly: N/A
fulfil requirements for DOUBLE or REINFORCED
INSULATION see Form A.9 N/A
If heating occurs at maximum working voltage, CLEARANCE complies with temperature corrected value
N/A
13 Prevention of HAZARD from arc flash and short-circuits P
13.1 General P
PROBE TIPS and crocodile clips are constructed to mitigate the risk of arc flash and short-circuits.
P
13.2 Exposed conductive parts P
13.2. a) PROBE ASSEMBLIES RATED for CAT III or IV, the exposed conductive part of a PROBE TIP 4 mm.
3.80 mm P
13.2. b) Special applications within CAT I where the energy levels not support arc flash or fire, the exposed conductive part of a PROBE TIP 80 mm
N/A
13.2. c) Other PROBE ASSEMBLIES, the exposed conductive part of a PROBE TIP 19 mm.
N/A
13.2. d) The outer surfaces of the jaws of crocodile or similar clips RATED for CAT II, III, or IV are not conductive.
N/A
Page 16 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
HAZARDOUS LIVE parts are not ACCESSIBLE when closed
N/A
Page 17 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
4.4.2 TABLE: Summary of SINGLE FAULT CONDITIONS Form A.1 N/A
Subclause Title Does not apply
Carried out
Comments
4.4.2.1 Equipment or parts for short-term or intermittent operation
4.4.2.2 Outputs of type B and type C PROBE
ASSEMBLIES
4.4.2.3 Insulation between circuits and parts
4.4.2.4 Components of type B and type C PROBE ASSEMBLIES
see Form A.2
List below all SINGLE FAULT CONDITIONS not covered by 4.4.2.1 to 4.4.2.4:
Supplementary information:
(see Form A.2 for details of tests)
Page 18 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
4.4 TABLE: Testing in single FAULT CONDITION – Results Form A.2 N/A
Test sub clause
Fault No.
Fault description Td 4.4.3 (NOTE)
How was test terminated Comments
Meets 4.4.4
NOTE Td = Test duration in h:min:s Record voltage test on Form A.10 and temperature tests on Form A..12 Record in the comments column for each test whether carried out during or after SINGLE FAULT CONDITION.
Supplementary information:
Page 19 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
5.3 TABLE: Durability of markings Form A.3 P
Marking method (see NOTE) Agent
1) A Water
2) B Isopropyl alcohol
3) C (specify agent)
4) D (specify agent)
5) E (specify agent)
NOTE – Where applicable include print method, label material, ink or paint type, fixing method, adhesive and surface to which marking is fixed.
Marking location Marking method (see above)
Identification (5.1.2) Molded
Fuses (5.1.3) N/A
TERMINALS and operating devices (5.1.4) N/A
DOUBLE/REINFORCED equipment (5.1.5) Molded
Rating (5.1.6) Molded
Warning marking (5.2) Molded
Method Test agent Remains legibleVerdict
Label loose Verdict
Curled edges Verdict
Comments
molded B Yes Yes Yes
Supplementary information:
Page 20 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
6 TABLE: Protection against electric shock - Block diagram Form A.4
POLLUTION DEGREE... : 2 Measurement category (overvoltage category).. : CAT IV 600 V
CAT III 1000 V
Location or Insulation type
Maximum working
CREEPAGE DISTANCE (NOTE 3)
CLEARANCE(NOTE 3)
Test voltage
Comments
description (NOTE 1) voltage (NOTE 2)
PWBmm
CTI Othermm
CTI mm (NOTE 2) V
Probe tip to barrier
RI CAT IV 600 V CAT III
1000 V
-- -- 31.5 IIIa-b 31.5 6880
NOTE 1 – Type of insulation: NOTE 2 - Types of voltage NOTE 3 - MEASUREMENT CATEGORY (OVERVOLTAGE CATEGORIES)
BI = BASIC INSULATION Peak impulse test voltage (pulse) or POLLUTION DEGREES which differ from DI = DOUBLE INSULATION r.m.s. these should be shown under "Comments". PI = PROTECTIVE IMPEDANCE d.c. RI = Reinforced INSULATION peak SI = Supplementary INSULATION
Supplementary Information:
Limits: pollution degree 2, material group IIIa-b
Cr=20.0 mm(RI), CL=14.3 mm(RI)
Probe tip
Probe
Probe barrier
Page 21 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
6.2 TABLE: List of ACCESSIBLE parts Form A.5 P
6.1.1 Exceptions —
6.2 Determination of ACCESSIBLE parts —
Item Description Determination method (NOTE 5)
Exception under 6.1.1 (NOTE 4)
Probe body
-- Visual --
NOTE 1 – Test fingers and pins are to be applied without force unless a force is specified (see 6.2.1) NOTE 2 – Special consideration should be given to inadequate insulation and high voltage parts (see 6.2) NOTE 3 – Parts are considered to be ACCESSIBLE if they could be touched in the absence of any covering which is not considered to
provide suitable insulation (see note to paragraph 1 of 6.4). NOTE 4 – Capacitor test may be required NOTE 5 – The determination methods are: visual; rigid test finger; jointed test finger; pin 3 mm diameter.
Supplementary information:
Page 22 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
6.3.1 TABLE: Values in NORMAL CONDITION (see NOTE 1) Form A.6 P
6.1.1 Exceptions 11.1 General
6.3.1 Values in NORMAL CONDITION 11.2 Cleaning
11.3 Specially protected PROBE ASSEMBLIES
Item Voltage Current Capacitance 10 s test Comments
(see Form A.5)
V r.m.s.
V peak
V d.c.
Test circuit
A1/A2/A3
mA r.m.s.
mA peak
mA d.c.
C
mJ
V
C
mJ
Probe body
117.0 165.4 -- A1 0.15 0.48 -- -- -- -- -- --
NOTE 1 – The requirements of 6.3.1 include drying out (if specified).
Supplementary information:
Page 23 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
6.3.2 TABLE: Values in SINGLE FAULT CONDITION Form A.7 N/A
Item Sub clause and
Voltage Transient (see NOTE)
Current Capacitance
(See Form A.4)
fault No. (see FormA.2)
V r.m.s.
V peak
V d.c.
V
s
Test circuit
A1/A2/A3
mA r.m.s.
mA peak
mA d.c.
F (NOTE)
Comments
NOTE – Transient voltages must be below the limits given from Figure 1 and the capacitance below the limits from figure 5 of IEC 61010-031.
Supplementary information:
Page 24 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
6.4.6 TABLE: PROTECTIVE IMPEDANCE Form A.8 N/A
A high INTEGRITY single component
Component Location Comments
A combination of components
Component Location Comments
A combination of BASIC INSULATION and a current or voltage limiting device
Component Location Comments
Supplementary information:
Page 25 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result – Remark Verdict
TRF No. IEC61010_031C
6.5 TABLE: CLEARANCES and CREEPAGE DISTANCES Form A.9 P
6.4 Insulation requirements for protection against electric shock P
6.7.2 ENCLOSURES of PROBE ASSEMBLIES with DOUBLE or REINFORCED INSULATION P
8 Mechanical resistance to shock and impact P
10.1 Integrity of CLEARANCES and CREEPAGE DISTANCES P
Location Measured (initial)
Verdict Mechanical tests (note) Test at max.
Measured after test(if required)
Verdict
(see Form A.4)
CREEPAGE
DISTANCE
CLEARANCE Appliedforce
Rigidity Drop Impact swing
RATED ambient
CREEPAGE
DISTANCE
CLEARANCE
mm mm N (8.1) (8.2) (8.3) (10.2) mm mm
Probe tip and barrier
31.5 31.5 P 30N 20N 1 m 0.37m 70 31.5 31.5 P
NOTE – Refer to Form A.10 for voltage tests following the above tests.
Supplementary information:
Page 26 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result - Remark Verdict
TRF No. IEC61010_031C
6.6 TABLE: Voltage tests Form A.10 P
4.4.4 Conformity after application of fault conditions1 N/A
6.4 Insulation requirements for protection against electric shock P
6.7.2 ENCLOSURES of PROBE ASSEMBLIES with DOUBLE or REINFORCED INSULATION P
6.7.5 Insulation of a probe cable P
8 Mechanical resistance to shock and impact P
11 Protection against hazards from fluids N/A 1 Record the fault, test or treatment applied before the voltage test
Test site altitude ................................................: m —
Test voltage correction factor (see Table 10)....: —
Location or references from
Forms A.2 and A.4
Clause or sub-clause
Humidity
Yes/No
Working voltage
V
Test voltager.m.s./peak/d.c
V
Comments Verdict
Probe tip to probe body
6.4 No 6880 Vrms
-- P
Probe tip to probe body
6.7.2 No 6880 Vrms
-- P
Probe tip to probe body
6.7.5 No 4300 Vrms -- P
Probe tip to probe body
8 No
CAT III
1000 V
CAT IV 600 V
6880 Vrms
-- P
Supplementary information:
Page 27 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result - Remark Verdict
TRF No. IEC61010_031C
6.7.4 TABLE: Cord anchorage of cable attachment Form A.11 P
Location Pull N
Verdict
Flexing/ pull
Verdict Rotational flexing
Verdict Comment
Cable to probe body 36 P 8.1 P 500 P
Cable to equipment 36 P 8.1 P 500 P
Supplementary information:
Page 28 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result - Remark Verdict
TRF No. IEC61010_031C
9. TABLE : Temperature Measurements Form A.12 P
9.1 Surface temperature limits - NORMAL CONDITION and / or SIGNLE FAULT CONDITION
Operating conditions:
Frequency ............... : -- Hz Test room ambient temperature (ta) .....: 23.1°C
Voltage .................... : -- V Test duration ........................................: 2 h 30 min
Part / Location tm
C tc
C tmax
C Verdict Comments
Probe body 27.9 44.8 70 P
Cable 26.9 43.8 75 P
Connector (To equipment) 27.2 44.1 70 P
NOTE 1 - tm = measured temperature tc = tm corrected (tm–ta+ 40 C or max. RATED ambient) tmax = maximum permitted temperature NOTE 2 - See also 12.1 with reference to component operating conditions NOTE 3 - Record values for NORMAL CONDITION and / or SINGLE FAULT CONDITION in this Form use additional form if necessary NOTE 4 - The tests of 6.7.4.1 to 6.7.4.3 are performed before temperature tests.
Supplementary information:
Page 29 of 31 Report No. : GZ11051010-1
IEC 61010-031
Clause Requirement + Test Result - Remark Verdict
TRF No. IEC61010_031C
10.2 TABLE: Resistance to heat of non-metallic enclosures Form A.13 P
Test method used: —
Non operative treatment .........................................: [V ]
Empty ENCLOSURE...................................................: [ ]
Operative treatment................................................: [ ]
Temperature during tests .......................................: —
ENCLOSURE samples tested were ...........................: —
Description Material Comments Verdict
Probe assembly PVC P
Voltage test (6.6) ....................................................: 6880 V r.m.s./peak/d.c. r.m.s
Supplementary information:
Page 30 of 31 Report No. : GZ11051010-1 Attachment
TRF No. IEC61010_031C
Probe assembly
Probe body
Page 31 of 31 Report No. : GZ11051010-1 Attachment
TRF No. IEC61010_031C
Connector to equipment
Wear indicator