Temperature Rise of the Post & on the Root Surface During Ultrasonic Post Removal

download Temperature Rise of the Post & on the Root Surface During Ultrasonic Post Removal

of 7

Transcript of Temperature Rise of the Post & on the Root Surface During Ultrasonic Post Removal

  • 8/9/2019 Temperature Rise of the Post & on the Root Surface During Ultrasonic Post Removal

    1/7

    Temperature rise of the post and on the root surface

    during ultrasonic post removal

    J. C. Budd, D. Gekelman & J. M. White

    Department of Preventive and Restorative Dental Sciences, School of Dentistry, University of California, San Francisco, CA, USA

    Abstract

    Budd JC, Gekelman D, White JM. Temperature rise of the

    post and on the root surface during ultrasonic post removal.

    International Endodontic Journal, 38, 705711, 2005.

    Aim To determine the temperature rise on the root

    surface caused by ultrasonic post removal using

    different devices and techniques in a laboratory setting.

    Methodology Two ultrasonic devices, one piezo-

    electrical (Pi) and one magnetostrictive (Ma), were

    investigated. A serrated titanium post was placed into

    the distal root canal of a human mandibular first molar.

    Four coolant parameters were utilized: no air, no water,

    no evacuation (NN), air only with high-speed evacu-

    ation (A), 15 mL min)1 water coolant with high-speed

    evacuation (W15) and30 mL min)1 water coolant with

    high-speed evacuation (W30). Five simulated post

    removals were measured at two locations, the post (P)

    and the root (R), for each coolant parameter. Tempera-

    ture rise was measured at 30, 60, 90and 120 s intervals

    using calibrated infrared thermography (n 80). Tem-

    peratures were recorded at 45 ms intervals. Data were

    analysed using repeated measures anova with the

    Scheffe post hoc test (P 0.05).

    Results The overall mean pooled effect showed

    that temperature rise for P 20.1 27.9 C and

    R 10.9 7.9 C were significantly different. Signifi-

    cant differences in temperature rise were: Pi > Ma,

    P > R, NN > A W15 W30 however, A > W30.

    Conclusions There were significant differences in

    temperature rise as a function of ultrasonic device,

    location on the tooth and cooling method utilized for

    post removal.

    Keywords: post removal, root, temperature, ultra-

    sound.

    Received 17 August 2004; accepted 16 May 2005

    Introduction

    Ultrasonic devices might be used for intraradicular post

    removal. Clinicians utilize a range of techniques,

    including no coolant, to improve visibility and air/

    water coolant to remove debris (Cohen & Burns 2001).

    One concern with these devices is the heat that they

    generate, which could penetrate into periradicular

    tissues and cause damage (Atrizadeh et al. 1971).

    The ultrasonic energy utilized in endodontic devices

    is generated by one of two types of ultrasonic

    transducers that convert one form of energy into

    another. Piezoelectrical (Pi) transducers produce ultra-

    sonic energy by transforming electricity into ultrasonic

    vibrations. Crystals within the transducer (usually

    made of quartz) are vibrated by the electricity flowing

    through them. By applying an alternating electrical

    field across the crystals, the quartz is compressed and

    released producing vibration of the tip. Magnetostric-

    tive (Ma) transducers use ferromagnetic materials and

    certain nonmetals called ferrites. A change in dimen-

    sion occurs when a rod or bar of this material is

    subjected to an alternating magnetic field producing

    vibration of the tip. Pi and Ma devices also produce

    noise and heat.

    Research has focused on the efficiency of different

    ultrasonic devices by measuring the time and force

    required to achieve post removal. Dixon et al. (2002)

    compared the time required to remove a 16 mm No. 5

    (0.050 in) Para-post cemented with zinc phosphate

    Correspondence: Joel M. White, DDS, MS, Professor, Division of

    Biomaterials and Bioengineering, Department of Preventive

    and Restorative Dental Sciences, 707 Parnassus Avenue

    Box 0758, San Francisco, CA 9414-0758, USA (Tel.: +1 415

    476 0918; fax: +1 415 476 0858; e-mail: whitej@

    dentistry.ucsf.edu).

    2005 International Endodontic Journal International Endodontic Journal, 38, 705711, 2005 705

  • 8/9/2019 Temperature Rise of the Post & on the Root Surface During Ultrasonic Post Removal

    2/7

    cement using the Piezo-Ultrasonic (Spartan USA,

    Fenton, MO, USA) and the Enac OE-50 (Osada Inc.,

    Los Angeles, CA, USA) at their highest intensities. The

    results showed that both devices were effective with

    typical post removal times of

  • 8/9/2019 Temperature Rise of the Post & on the Root Surface During Ultrasonic Post Removal

    3/7

    was removed at the level of the cemento-enamel

    junction. The removal of the crown also eliminated

    the possibility of heat dissipating into the crown.

    Temperature rise was measured at two locations.

    The first location was at the coronal aspect of the post

    root interface (P). The second location was at the distal

    root surface 3 mm from the apex of the root (R) at the

    terminus of the post (Fig. 1). The ultrasonic tip supplied

    with each device was applied to the post at the junction

    of the post and tooth surface for a duration of 120 s

    using a circumferential technique around the post. The

    force applied to the toothpost interface was designedto simulate the range of force used clinically during

    post removal.

    The experimental setup with evacuation and ultra-

    sonic tip is illustrated in Fig. 1. The tooth was held

    stationary by clamping the mesial root. Temperatures

    were measured using an infrared camera (Model

    TH-5104; Mikron Infrared Inc., Oakland, NJ, USA).

    The camera measures temperatures by performing a

    scan of the field of focus every 45 ms. The accuracy of

    the camera was verified to within 0.5 C using heated

    water at 51 C and cooled water at 19 C. The infrared

    camera has a mercurycadmiumtellurium linear

    array detector which enables the camera to detect

    changes in temperature. The operator distinguished

    changes in temperature using two different methods.

    First, the camera was programmed to distinguish

    variations in temperature by producing a different

    coloured image on the monitor for every 5 C change

    in temperature. Secondly, numeric point values to one

    significant digit were displayed on the screen in C.

    A measurement of room temperature was simulta-

    neously recorded as a baseline for temperature rise. An

    example of an image produced by the camera is

    provided in Fig. 2.

    Following each trial, the videotape was used to

    record temperatures at 30, 60, 90 and 120-s intervals.

    Five repetitions were recorded for each location for

    each cooling regimen (n 80). Five repetitions were

    chosen based on a sample size calculation using

    repeated measures analysis of variance (anova) design

    with power 0.8, minimum detectable difference of

    5 C and SD of residuals of five for the eight treatmentgroups over the four time points measured. The

    temperature rise was calculated by taking the maxi-

    mum temperature at each interval and subtracting the

    baseline room temperature. The temperature of the

    toothpost system was allowed to return to room

    temperature in between repetitions. All data were

    obtained and recorded by a single operator. Independ-

    ent variables were as follows: device type, location on

    the tooth and cooling regimen. The dependent variable

    was temperature rise. Results were analysed for statis-

    tical significance using repeated measures anova with

    the Scheffe post hoc test (P 0.05).

    Results

    The overall mean pooled effect for temperature rise

    following instrumentation with each ultrasonic device

    is displayed in Table 2. Temperature rises were higher

    for the Pi, especially with no air and no water coolant.

    Figure 1 Experimental setup showing the ultrasonic hand-

    piece, high volume evacuation (E), the lens of the infrared

    camera (L) and the tooth with post. Temperature measure-

    ments were taken at the posttooth interface (P) and at the

    apical root surface (R).

    +22.6

    59.124.55

    25.731.0

    +

    (200.0)50.0

    45.0

    40.0

    35.0

    30.0

    25.020.0

    15.0

    10.0< (10.0)

    P 0

    +

    ++

    Figure 2 Infrared camera image showing temperature scale

    on the right, a baseline temperature of 22.6 C with a

    temperature of 59.1 C at the toothpost interface and 31 C

    at the surface of the root during piezoelectric post removal

    with coolant parameter air only (treatment condition A).

    Budd et al. Temperature from post removal

    2005 International Endodontic Journal International Endodontic Journal, 38, 705711, 2005 707

  • 8/9/2019 Temperature Rise of the Post & on the Root Surface During Ultrasonic Post Removal

    4/7

    Table 3 shows the temperature rise as a function of

    device type and distinguishes between the locations on

    the tooth. As expected, the temperature at the post

    tooth interface was higher than at the surface of the

    root. The posttooth interface and root surface heat

    profiles were uniform and followed the outline of the

    tooth and we did not visualize any change in heat

    profile at the slight concavity of the mesial aspect of the

    distal root used in the study. Figures 3 and 4 show

    temperature rise as a function of device, location and

    cooling regimen. Temperature rise was inversely pro-

    portional to air and water coolant. Statistically signi-

    ficant differences were found for temperature rise at

    the post location using the Pi device where

    NN > A W15 W30. There were no statistically

    significant differences in temperature rise at the post

    location using the Ma device where NN A W15 W30. At the root, results for the Pi device were

    NN > A W15 W30. At the root, results for the

    Ma device were NN A W15 > W30. The mean

    pooled results were significantly different between

    device type (Pi > Ma) and location on the tooth

    (P > R). Statistical significance regarding temperature

    rise between cooling regimens was as follows: NN > A,

    A W15, W15 W30 and A > W30 for both

    devices tested. Temperature rise as a function of time

    and location is listed in Table 4. Overall, the majority of

    the temperature rise at the post occurred within the

    first 30 s whereas the temperature rise at the root

    surface continued to rise up to 90 s.

    Discussion

    The ultrasonic devices used in this study were selected

    to compare the differences in the temperature rise using

    different ultrasonic transducers. The four cooling reg-

    imens used in this study were selected to include

    common cooling techniques currently used by many

    practitioners. The NN parameter was selected as this

    technique provides the best visualization for the clini-

    cian, although with the highest temperature rise.

    Temperature rise at both the crown and the root was

    measured in order to determine the amount of heat

    conducted down the post to the root. The distal root of

    the mandibular molar was chosen as this represents a

    common site of post placement. The temperature rise at

    the distal surface of the root was measured to facilitate

    imaging with the infrared camera. The 120 s duration

    of instrumentation was chosen to simulate a clinical

    setting in which the progress of post removal would be

    Table 3 Temperature rise as a function of ultrasonic deviceand location

    Device/location Temperature rise (C) 1 SD

    Piezoelectric/post 27.9 34.2

    Piezoelectric/root 12.3 16.6

    Magnetostrictive/post 1 3.9 8.9

    Magnetostrictive/root 7.9 5.3

    NN A W15 W30 NN A W15 W300

    10

    20

    30

    40

    50

    60

    70

    80

    90

    100

    110

    Temprise(C)

    Piezoelectrical Magnetostrictive

    Figure 3 Temperature rise at the post as a function of

    ultrasonic device and cooling regimen.

    Table 4 Mean temperature rise (C) as a function of location

    and time interval

    30 s 60 s 90 s 120 s

    P os t 23.8 31.0 29.6 35.8 31.0 36.5 27.1 35.3

    Root 6.7 7 .8 12.0 15.3 14.8 1 8.9 15 .6 20.9

    Table 2 Temperature rise as a function of ultrasonic device

    Device Temperature rise (C) 1 SD

    Piezoelectric 20.1 27.9

    Magnetostrictive 10.9 7.9

    NN A W15 W30 NN A W15 W30

    0

    10

    20

    30

    40

    50

    60

    70

    80

    90

    100

    110

    Temprise(C)

    Piezoelectrical Magnetostrictive

    Figure 4 Temperature rise at the root as a function of

    ultrasonic device and cooling regimen.

    Temperature from post removal Budd et al.

    International Endodontic Journal, 38, 705711, 2005 2005 International Endodontic Journal708

  • 8/9/2019 Temperature Rise of the Post & on the Root Surface During Ultrasonic Post Removal

    5/7

    verified periodically and some form of coolant and

    evacuation applied to wash the area.

    Heat is produced from ultrasonic devices through

    three different mechanisms. First, via friction created

    between the titanium post and the ultrasonic tip.

    Secondly, via the temperature of the coolant flowing

    through the handpiece. Thirdly, via acoustic energyabsorption of ultrasound transmitted to the tooth

    (Bergeron et al. 2001).

    The Pi transducer caused a significantly greater

    temperature rise than the Ma transducer when the NN

    cooling regimen was used. However, the Ma device

    produced a higher temperature using the W15 and

    W30 parameters at both the crown and the root than

    the Pi device. Although the posttooth system was

    allowed to cool to room temperature between repeti-

    tions, the heat produced and retained within the

    transducer most likely explains the above observation

    caused by the water coolant flowing through the

    handpiece. The increased temperature of the water

    coolant caused by the ferromagnetic rods might have

    contributed to the increased temperatures observed

    with the use of the Ma device.

    Repeated measure methods were used reducing the

    sample size estimate to five repetitions per treatment

    condition. This allowed for the determination of statis-

    tical differences between the two devices and four

    coolant regimens over time at the post and root surface.

    One post and root was utilized in order to limit the

    variability because of differences in remaining dentine

    thickness and root volume. By this method, it was

    possible to determine differences in heat produced as afunction of the devices and cooling regimen. The

    average length of the distal root of a lower first molar

    is 14 mm. In the cervical region, the mesial/distal

    width is generally equal to the buccal/lingual width,

    9 mm (Ash 1993). By comparison, the average root

    length for maxillary and mandibular canines is 16 mm

    with cervical mesial/distal width of 5.5 mm and

    cervical buccal/lingual diameter of 7 mm (Ash 1993).

    Anterior teeth with less root volume would be expected

    to have higher temperature rises than those that we

    reported and roots with larger volumes would be

    expected to have lower rises than those we reported.

    Further studies using different size roots would define

    the range of temperature rises that might be expected

    for all teeth. However, the sample size would be very

    large for each treatment group accounting for variation

    in root size, remaining dentine thickness and volume. A

    change in heat profile at the slight mesial concavity of

    the mesial aspect of the molar was expected but did not

    occur. The infrared images confirmed the majority of

    the heat diffused down through the post and into the

    dentine heating the root surface uniformly. It can be

    hypothesized that within the range of remaining

    dentine of the root (12 mm) that the temperature

    effects from the diffused heat were within 5 C because

    no difference was visualized in the infrared thermalprofile. The application of the ultrasonic device was

    limited to the post for 2 min to simulate a clinical

    application. Generally, clinicians use a device for this

    amount of time and then stop to inspect the area. Even

    2 min was also a good time as the temperature rise

    reached steady state within 90 s and did not continue

    to increase.

    Results from Eriksson & Albrektsson (1983) sug-

    gested that a temperature rise above 10 C can cause

    irreversible damage to the periodontal ligament and

    bone. Comparisons of temperature rise between cooling

    regimens were made to determine what regimen cooled

    the tooth sufficiently to consistently remain below the

    10 C threshold. The infrared camera used to measure

    temperature rise in this study allowed the measure-

    ment of temperature rise to be recorded over a broad

    area and to a degree of accuracy that is not achievable

    with the traditional thermocouples used in previous

    studies. Thermocouples allow temperatures to be

    recorded at only those locations between the thermo-

    couples where a circuit can be created. In contrast, an

    infrared camera allows temperatures to be recorded at a

    number of locations simultaneously with accuracy to

    one significant digit. Using infrared thermography, the

    area of highest temperature rise can be pinpointed andmeasured (McCullagh et al. 2002). Thermocouple

    accuracy can be negatively affected by a number of

    factors including how the thermocouple adheres to the

    material to be measured, alteration of the electrical

    circuit and specific operating temperature ranges rela-

    ted to thermocouple types. All of these factors are

    eliminated with the use of an infrared camera.

    Recently, Satterthwaite et al. (2003) reported tem-

    perature changes from ultrasonic vibration of ceramic

    and stainless steel posts. The results of their study

    indicated temperature rises lower than those in the

    present report. Their study utilized morphologically

    similar canine teeth mounted in a mounting jig

    (silicone rubber) that would have increased the vari-

    ability of the measurements and also reduced the

    temperature rise as compared with the present study.

    In their study, k-type thermocouples were mounted on

    the root surface, whereas infrared thermography was

    utilized in the present study. Satterthwaite et al. (2003)

    Budd et al. Temperature from post removal

    2005 International Endodontic Journal International Endodontic Journal, 38, 705711, 2005 709

  • 8/9/2019 Temperature Rise of the Post & on the Root Surface During Ultrasonic Post Removal

    6/7

    did not report the amount of water coolant and they

    used longer time intervals, up to 30 min, for post

    removal. The differences in the results of these studies

    are most likely because of differences in experimental

    technique and materials. Both utilized the tip of the

    ultrasonic device in contact with the post, creating heat

    by friction and both studies suggest temperature riseslikely to cause adverse thermal effects to adjacent

    tissues. In the present study, the majority of tempera-

    ture rise occurred at the posttooth interface within the

    first 30 s. The temperature rise continued to increase at

    the root surface up to 90 s before reaching a steady

    state. The difference in time to reach steady state at the

    crown and root surface is presumably because of the

    diffusion of heat through the tooth taking time to reach

    the root surface.

    In summary, this study shows that failure to provide

    some form of water coolant during ultrasonic post

    removal can result in temperature rises that exceed the

    10 C threshold. The results of the present study also

    illustrate the need of using the minimum time to reach

    the treatment objective; even 15 mL min)1 water

    coolant can permit a temperature rise that exceeds the

    10 C threshold. In this study, the temperature rise

    never exceeded the 10 C threshold when using

    30 mL min)1 water coolant. The results showed no

    statistical significance between 15 and 30 mL min)1

    water coolant. Nevertheless, we recommend that a

    minimum of 30 mL min)1 water coolant be used during

    ultrasonic post removal procedures as this regimen did

    not exceed the defined temperature rise threshold.

    Future studies should focus on determining differ-ences in temperature rise comparing different teeth,

    varying dentine and enamel thicknesses, a range of

    clinical techniques including a range of applied force

    and different coolant conditions. The intensity of the

    temperature rise, the frequency of the thermal insult,

    the size of the thermal mass and the duration of

    temperature rise are all important factors to be consid-

    ered in determining adverse thermal effects to tissue. In

    a clinical setting it would be expected that the

    temperatures would be lower given the larger body

    mass for heat dissipation and blood flow of surrounding

    tissues. These factors are likely to have an impact in

    reducing the amount of heat that is actually absorbed

    by periodontal tissues.

    Acknowledgements

    We would like to thank Kim Tran, SRA, for laboratory

    assistance, data management and graphics. We also

    thank Steve Barrabee, J.D. at Bradley, Curley, Asiano,

    Barrabee and Crawford, P.C. for collaboration on the

    design of this study and fruitful discussions on the

    results. This work supported in part by a grant from

    The Dentists Insurance Company of California.

    References

    Altshul JH, Marshall G, Morgan LA, Baumgartner JC (1997)

    Comparison of dentinal crack incidence and of post removal

    time resulting from post removal by ultrasonic or mechan-

    ical force. Journal of Endodontics 23, 6836.

    Ash MM. (1993) Wheelers Dental Anatomy, Physiology

    and Occlusion. Philadelphia, PA: W.B. Saunders Co.,

    pp. 27489.

    Atrizadeh F, Kennedy J, Zander H (1971) Ankylosis of teeth

    following thermal injury. Journal of Periodontal Research 6,

    15967.

    Bergeron BE, Murchison DF, Schindler WG, Walker WA III

    (2001) Effect of ultrasonic vibration and various sealer and

    cement combinations on titanium post removal. Journal of

    Endodontics 27, 137.

    Buoncristiani J, Seto BG, Caputo AA (1994) Evaluation of

    ultrasonic and sonic instruments for intraradicular post

    removal. Journal of Endodontics 20, 4869.

    Cameron JA (1988) The effect of ultrasonic endodontics on the

    temperature of the root canal wall. Journal of Endodontics 14,

    5549.

    Cohen S, Burns RC. (2001) Pathways of the Pulp, 8th edn. St

    Louis, MO: Mosby, pp. 899901.

    Dixon EB, Kaczkowski PJ, Nicholis JI, Harrington GW (2002)

    Comparison of two ultrasonic instruments for post removal.

    Journal of Endodonics 28, 1115.

    Eriksson AR, Albrektsson T (1983) Temperature thresholdlevels for heat-induced bone tissue injury: a vital micro-

    scopic study in the rabbit. Journal of Prosthetic Dentistry 50,

    1017.

    Garrido ADB, Braga NMA, Alfredo E (2003) Effectiveness

    of ultrasonic vibration on removal of radicular posts.

    Journal of Dental Research (Special Issue A), 87, Abstract

    1621.

    Kocher T, Plagmann HC (1996) Heat propagation in dentin

    during instrumentation with different sonic scaler tips.

    Quintessence International 27, 25964.

    McCullagh JJP, Setchell DJ, Gulabivala K et al. (2002) A

    comparision of thermocouple and infrared thermographic

    analysis of temperature rise on the root surface during

    continuous wave of condensation technique. International

    Endodontic Journal 33, 32632.

    Nicoll BK, Peters RJ (1998) Heat generation during ultrasonic

    instrumentation of dentin as affected by different irrigation

    methods. Journal of Periodontology 69, 8848.

    Satterthwaite JD, Stokes AN, Franke NT (2003) Potential for

    temperature change during application of ultrasonic vibra-

    Temperature from post removal Budd et al.

    International Endodontic Journal, 38, 705711, 2005 2005 International Endodontic Journal710

  • 8/9/2019 Temperature Rise of the Post & on the Root Surface During Ultrasonic Post Removal

    7/7

    tion to intra-radicular posts. European Journal of Prosthodon-

    tic and Restorative Dentistry 11, 516.

    Smith BJ (2001) Removal of fractured posts using ultrasonic

    vibration: an in vivo study. Journal of Endodontics 27,

    6324.

    Trenter S, Walmsley A (2003) Ultrasonic dental scaler: asso-

    ciated hazards. Journal of Clinical Periodontology 30, 95101.

    Verez-Fraguela JL, Vives Valles MA, Ezquerra Calva LJ (2000)

    Effects of ultrasonic dental scaling on pulp vitality in dogs: an

    experimentalstudy.Journal of Veterinarian Dentistry17, 759.

    Budd et al. Temperature from post removal

    2005 International Endodontic Journal International Endodontic Journal, 38, 705711, 2005 711