Temperature and Dimensional Measurements

13
www.mitutoyo.com/education © 2018 Mitutoyo America Corporation Jim Salsbury, Ph.D., Corporate Metrology Mitutoyo America Corporation Temperature and Dimensional Measurements www.mitutoyo.com/on-demand 2 Scroll down to “Technical Presentations”

Transcript of Temperature and Dimensional Measurements

Page 1: Temperature and Dimensional Measurements

www.mitutoyo.com/education

© 2018 Mitutoyo America Corporation

Jim Salsbury, Ph.D., Corporate Metrology

Mitutoyo America Corporation

Temperature and Dimensional

Measurements

www.mitutoyo.com/on-demand

2

Scroll down to “Technical Presentations”

Page 2: Temperature and Dimensional Measurements

www.mitutoyo.com/education

© 2018 Mitutoyo America Corporation

Mitutoyo America Calibration Lab

20±0.1°C, 3500 sq. ft.0.03°C/hour, 0.03°C/meter

Environmental specifications based on accuracy requirements

Temperature Impacts all Measurements

• Listen and understand. Temperature is out there. It can’t be bargained with. It can’t be reasoned with. It doesn’t feel pit, or remorse, or fear. And it absolutely will not stop, ever, until all your measurements are screwed up.

4

Page 3: Temperature and Dimensional Measurements

www.mitutoyo.com/education

© 2018 Mitutoyo America Corporation

Example: Long Gage Block Calibration

5

Master and test block soak overnight

Heat shields to block operator heat

0.03°C/meter

Match materials

Example: Step Gage Calibration on CMM

20±0.1°C0.03°C/hourSeparate room

to reduce traffic

Automated to remove operator

body heatCMM controller in adjacent room

Temperature compensation

Page 4: Temperature and Dimensional Measurements

www.mitutoyo.com/education

© 2018 Mitutoyo America Corporation

Example: High Accuracy CMM Calibration

20±0.1°C

Use gage blocks of low thermal

expansion material (about 500X less

than steel)

Example: Ultra Accuracy Length Calibration

NIST Moore CMM in 20 ± 0.01°C room

Separate control room

“Whoever has the best thermometer wins” - Dr. Doiron, NIST

No people in room for many hours prior to measurements

Page 5: Temperature and Dimensional Measurements

www.mitutoyo.com/education

© 2018 Mitutoyo America Corporation

Justify Expense for Temperature Control

• CMM specifications in accordance to ISO 10360-2.

• Calibrate with Checkmasters.

• E0 test run in 7 positions. In each position, 5 lengths and 3 repeats, for a total of 105 measurements.

• We need Checkmasterscalibrated with adequate uncertainty.

All manufacturers specify CMM accuracy following the ISO 10360 standards

Page 6: Temperature and Dimensional Measurements

www.mitutoyo.com/education

© 2018 Mitutoyo America Corporation

E0,MPE = 1.9 + 3L/1000 m, with L in mmE0,MPE = 75 + 3L inches, with L in inches

0 8 16 24 32 40 48 56 64

-0.8

-1.6

-2.4

-3.1

0.8

3.1

2.4

1.6

0

Measured Length L (inches)

Measured Length L (mm)

Eo,MPE

(m)E0,MPE

(0.0001”) At 40” (1000 mm), E0,MPE 0.0002” (5 m)

Legex CMM – Length Calibration

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1

0 250 500 750 1000

Sandia

Oak Ridge

NIST

MACMJ

Uncertainty comparison

To ensure quality in CMM calibration, Mitutoyo America

developed methods that are the most accurate of any commercial

laboratory in the U.S.

Page 7: Temperature and Dimensional Measurements

www.mitutoyo.com/education

© 2018 Mitutoyo America Corporation

13

Fight Temperature to the Level You Need

• Temperature stability (vs 68°F).• Match materials between gage and part.• Allow proper thermal soak-out time.• Reduce handling and operator influences.• Reduce traffic and large doors.• Reduce heat sources (lights/people/equipment).• Eliminate direct blowing of air.• Monitor the environment (with alarms).• Keep the lights on all the time.• Temperature compensation.

Understanding Temperature Issues

• Thermal expansion:L = (L)()(T-20)

where is the coefficient of thermal expansion, or CTE, of the material.

• 20” gage block at 21°C.L = (20)(10.8)(21-20)

L 0.0002 inches

• 20” gage block, Grade 00, tolerance is 10x smaller.

14

Long gage block calibration

Page 8: Temperature and Dimensional Measurements

www.mitutoyo.com/education

© 2018 Mitutoyo America Corporation

Coefficient of Thermal Expansion (CTE)

• 10.5 ± 1.5 x 10-6/°C:– Most steels

– Steel gage blocks

– Ceramic gage blocks

– Calipers

– Carbon fiber calipers

– Micrometers

– Micrometers standards

15

Use of similar CTE reduces thermal errors

20

CTE (10-6/°C)

10

0

5

15

- Aluminum

- Steel- Ceramic (ZrO2)

- Tungsten Carbide

- Glass scales

- Zero CERA

Matching Materials (similar CTE) is Good

• Similar CTE materials at same temperature (even not 68°F) will reduce thermal errors

Calibrate in lab and use on shop floor relies on similar materials

between micrometer and standard

Page 9: Temperature and Dimensional Measurements

www.mitutoyo.com/education

© 2018 Mitutoyo America Corporation

Thermal Error – Aluminum Part, Steel Micrometer

70°F

72°F

74°F

76°F

78°F

0.0000

0.0002

0.0004

0.0006

0.0008

0.0010

0.0012

0.0014

0.0016

0 2 4 6 8 10 12 14 16 18 20 22 24

The

rmal

Err

or (

inch

es)

Measured Length (inches)

To reduced this error:(1) Manually compensate(2) Adjust zero point

High Accuracy Height Gage: Mitutoyo LH-600

18

Calibration at 20°C Used at 23°C

High accuracy equipment often requires temperature compensation to fully utilize the accuracy for which you paid

Use built-in temp comp

Page 10: Temperature and Dimensional Measurements

www.mitutoyo.com/education

© 2018 Mitutoyo America Corporation

Steel Part on LH-600 without Temp Comp

LH-600 Spec

70°F

72°F

74°F

76°F

78°F

0.00000

0.00005

0.00010

0.00015

0.00020

0.00025

0.00030

0.00035

0.00040

0 2 4 6 8 10 12 14 16 18 20 22 24

The

rmal

Err

or (

inch

es)

Measured Length (inches)

LH-600 Temperature Compensation

• LH-600 temperature compensation is built-in and is easy to use.

• Historical good practice of “thermal equilibrium” needs to be updated for modern equipment.

20

Page 11: Temperature and Dimensional Measurements

www.mitutoyo.com/education

© 2018 Mitutoyo America Corporation

Aluminum Part on LH-600 without Temp Comp

LH‐600 Spec

70°F

71°F

72°F

73°F

69°F

0.00000

0.00010

0.00020

0.00030

0.00040

0.00050

0.00060

0.00070

0.00080

0.00090

0.00100

0 2 4 6 8 10 12 14 16 18 20 22 24

Thermal Error (inches)

Measured Length (inches)

Built-In Real-Time Temperature Compensation

• Temp comp is standard on many CMMs today.

• CMM accuracy relies on use of temp comp.

• If not used properly, you may create a false temperature differential that results in big errors.

22

Workpiece temperature sensor

Page 12: Temperature and Dimensional Measurements

www.mitutoyo.com/education

© 2018 Mitutoyo America Corporation

Steel part on temp comp CMM when NOT correctly using the temperature compensation

CMM spec

70°F

72°F

74°F

76°F

78°F

0.0000

0.0002

0.0004

0.0006

0.0008

0.0010

0.0012

0.0014

0.0016

0 2 4 6 8 10 12 14 16 18 20 22 24

The

rmal

Err

or (

inch

es)

Measured Length (inches)

CMM Accuracy across Temperature Range

• CMMs are generally specified to be accurate across a range of temperatures.

• Example for Mitutoyo Crysta-Apex S:– 16–26°C (60.8–78.8°F)

• Specification assumes proper use of temperature compensation.

24

Workpiece temperature sensors in use during calibration

Page 13: Temperature and Dimensional Measurements

www.mitutoyo.com/education

© 2018 Mitutoyo America Corporation

Temp Comp on CMMs Made Easy

• Part temperature can be estimated by keeping workpiece temperature sensors attached to nearby fixturing.

• Must still use correct CTE of measured part.

• Also helps eliminate wear and tear on the temperature sensor.

25

Thank You