Tech Radiopharmacy

download Tech Radiopharmacy

of 52

Transcript of Tech Radiopharmacy

  • 7/28/2019 Tech Radiopharmacy

    1/52

    The Radiopharmacy

    A Technologists Guide

    European Association of Nuclear Medici

    ne

    Produced with the kind Support o

  • 7/28/2019 Tech Radiopharmacy

    2/52

    2

    Contributors

    James R Ballinger, PhD

    Chie Radiopharmaceutical Scientist

    Guys and St Thomas NHS Foundation Trust

    London, United Kingdom

    Clemens Decristooro, PhD (*)

    Chair o the EANM Radiopharmacy Committee

    Clinical Department o Nuclear Medicine

    Medical University Innsbruck

    Innsbruck, Austria

    Brit Farstad

    Member o the EANM Radiopharamcy Committee

    M.Pharm/Radiopharmacist

    Department Head, Isotope Laboratories

    Institute or Energy Technology

    Kjeller, Norway

    Brendan McCoubrey

    Radiation Saety Ocer

    Dept. o Diagnostic Imaging

    St. Jamess Hospital

    Dublin, Ireland

    Geraldine OReilly, PhD

    Radiation Protection Advisor

    Dept. o Medical Physics and BioEngineeringSt. Jamess Hospital

    Dublin, Ireland

    Helen Ryder

    Clinical Specialist Radiographer

    Dept. o Diagnostic Imaging

    St. Jamess Hospital

    Dublin, Ireland

    Tanja Gmeiner Stopar, PhD

    Member o the Education Board,

    EANM Radiopharmacy Committee

    Radiopharmacy, Head

    University Medical Centre Ljubljana

    Department or Nuclear Medicine

    Ljubljana, Slovenia

    Wim van den Broek

    Chair o the EANM Technologist Committee

    Chie Technologist

    Dept o Nuclear Medicine

    University Medicine Centre

    Nijmegen, The Netherlands

    Editors

    Suzanne Dennan

    Vice-Chair o the EANM Technologist Committee

    Acting Radiographic Services Manager

    Dept. o Diagnostic Imaging

    St. Jamess HospitalDublin, Ireland

    Clemens Decristooro, PhD *

    This booklet was sponsored by an educational grant from Lantheus Medical Imaging . The views expressed are those

    of the authors and not necessarily of Lantheus Medical Imaging.

  • 7/28/2019 Tech Radiopharmacy

    3/52

    EA

    NM

    3

    Contents

    Foreword

    Wim van den Broek . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

    Introduction

    Clemens Decristooro . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

    Chapter 1 - Radiopharmacy Technology

    Brit Farstad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Chapter 2 - Radiopharmacy Design

    James Ballinger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12

    Chapter 3 - Radiopharmacy: Preparing & Dispensing Radiopharmaceuticals

    Geraldine OReilly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16

    Chapter 4 - Radiopharmacy: Kits & Techniques

    Helen Ryder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22

    Chapter 5 Radiopharmacy: Blood Labelling

    Tanja Gmeiner Stopar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33

    Chapter 6 - Radiopharmacy: Record Keeping & Administration

    Brendan McCoubrey . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41

    Reerences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49

    Imprint. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51

  • 7/28/2019 Tech Radiopharmacy

    4/52

    4

    ForewordWim van den Broek

    should be provided or all sta working in ra-

    diopharmacy departments in the aspects o

    quality assurance in which they are involved.

    This includes: preparation, release, quality con-

    trol and analytical techniques, cleaning, trans-

    portation, calibration o equipment (especially

    or the measurement o radioactivity), working

    practices in the radiopharmacy, preparation o

    the individual doses, documentation, hygiene,pharmaceutical microbiology, and microbio-

    logical monitoring. Oten a Nuclear Medicine

    Technologist is the person who is involved

    in the preparation and quality control o the

    radiopharmaceuticals.

    I am grateul or the eort and hard work o all

    the contributors, who are the key to the con-

    tents and educational value o this booklet.The most essential and relevant aspects o ra-

    diopharmacy in daily practice are emphasised

    here. This booklet is prepared in cooperation

    with the Radiopharmacy Committee o the

    EANM. This Committee is very active and criti-

    cal in the eld o regulations and guidelines or

    the production o radiopharmaceuticals and

    constantly proposes practical solutions. Many

    thanks to Suzanne Dennan who coordinated

    this project.

    With this new booklet, the EANM Technolo-

    gist Committee oers to the NMT community

    again a useul and comprehensive tool that

    may contribute to the advancement o their

    daily work.

    Since it was ormed, the EANM Technologist

    Committee has been devoted to the improve-

    ment o nuclear medicine technologists

    (NMTs) proessional skills. Publications that will

    assist in the setting o high standards or NMTs

    work have been developed and since 2004

    a series o brochures, Technologists Guides,

    have been published yearly. This booklet

    about radiopharmacy is already the th vol-ume. The new and stricter regulations in the

    eld o preparation o radiopharmaceuticals

    changed the daily practice in the radiophar-

    macy in the last 5 years.

    Nuclear medicine is a multidisciplinary special-

    ty in which medicine, physics and pharmacy

    are involved. The Radiopharmacy is an integral

    part o a nuclear medicine department and itsprime responsibility is the preparation o high

    quality radiopharmaceuticals, the base or a

    high quality nuclear medicine examination.

    The majority o these radiopharmaceuticals

    is mainly used or diagnostic imaging, which

    is the main activity o nuclear medicine. Ra-

    diopharmaceuticals are medical products

    dened in the European directive 2004/27/

    EC amending the directive 2001/83/EC. As in

    other disciplines the complex changes driven

    by European legislation had their impact on

    everyday practice in the preparation o radio-

    pharmaceuticals.

    Only trained people should be responsible or

    and participate in the preparation and qual-

    ity control o radiopharmaceuticals. Training

  • 7/28/2019 Tech Radiopharmacy

    5/52

    EA

    NM

    5

    I want to congratulate the Technologists Com-

    mittee o the EANM or this excellent Tech-

    nologists Guide on the Radiopharmacy. Issues

    o quality assurance especially in the eld o

    pharmaceutical preparations are becoming

    increasingly important. The Radiopharmacy

    Committee o the EANM thereore recently

    has issued general guidelines or Current

    Good Radiopharmacy Practice describingthe quality standards in the preparation o

    conventional and PET radiopharmaceuticals

    (http://www.eanm.org/scientic_ino/guide-

    lines/gl_radioph_cgrpp.php) . These serve as

    a general reerence standard or radiophar-

    maceutical preparation as radiopharmacy

    practice still shows a great variability all over

    Europe.

    Technologists in many countries are the

    backbone or radiopharmacy services within

    nuclear medicine departments. This is espe-

    cially true or the preparation and handling

    o conventional radiopharmaceuticals includ-

    ing eluting radionuclide generators, prepara-

    tion o99mTc-radiopharmaceuticals rom kits,

    dispensing and cell labelling. Thereore the

    current issue o the Technologists Guides is

    dedicated to radiopharmacy practice.

    The Technologists Committee o the EANM

    has been very active in promoting proes-

    sional skills o technologists and to support

    high quality standards in daily practice. The

    series o Technologists Guide booklets by the

    Eductional Sub-Committee has been a valu-

    able part o these initiatives. The current issue

    o this series intends to provide guidance or

    a good radiopharmacy practice, to describequality standards and to bring radiopharmacy

    practice to equal standards throughout Eu-

    rope.

    This booklet contains chapters o all relevant

    topics o daily radiopharmacy practice o tech-

    nologists such as radiopharmacy design, prep-

    aration and dispensing as well as documenta-

    tion written by European experts in the eld,both radiopharmacists and technologists.

    I am very condent that this booklet will not

    only provide valuable inormation and quick

    reerence or problems arising in daily prac-

    tice, but also will help to continuously improve

    quality standards o radiopharmacy practices

    in nuclear medicine.

    IntroductionClemens Decristooro, PhD

  • 7/28/2019 Tech Radiopharmacy

    6/52

    6

    Chapter 1 Radiopharmacy TechnologyBrit Farstad

    A radiopharmaceutical can be as simple as a

    radioactive element such as 133Xe, a simple salt

    such as 131I-NaI, or a labelled compound such

    as 131I-iodinated proteins and 99mTc-labeled

    compounds.

    Usually, radiopharmaceuticals contain at least

    two major components:

    A radionuclide that provides the desired

    radiation characteristics.

    A chemical compound with structural or

    chemical properties that determine the in

    vivo distribution and physiological behav-

    iour o the radiopharmaceutical.

    Radiopharmaceuticals should have severalspecic characteristics that are a combination

    o the properties o the radionuclide used as

    the label and o the nal radiopharmaceutical

    molecule itsel.

    Decay o radionuclides

    Radionuclides are unstable nuclei that are sta-

    bilised upon radioactive decay. Approximately

    3000 nuclides have been discovered so ar;

    most o these are unstable, but only about 30

    o these are routinely used in nuclear medicine.

    Most o these are articial radionuclides, which

    may be produced by irradiation in nuclear re-

    actors, cyclotrons, or large linear accelerators.

    A radionuclide may decay by emitting dier-

    ent types o ionising radiation: alpha (), beta(-), positron (+) and gamma () radiation.

    Radiopharmacy

    Radiopharmacy encompasses studies related

    to the pharmaceutical, chemical, physical,

    biochemical, and biological aspects o ra-

    diopharmaceuticals. Radiopharmacy com-

    prises a rational understanding o the design,

    preparation and quality control o radiophar-

    maceuticals, the relationship between the

    physiochemical and biological properties oradiopharmaceuticals and their clinical appli-

    cation, as well as radiopharmaceuticals chem-

    istry and issues related to the management,

    selection, storage, dispensing, and proper use

    o radiopharmaceuticals.

    Characteristics o radiopharmaceuticals

    A radiopharmaceutical is a pharmaceutical

    that, when ready or use, incorporates oneor more radionuclides (radioactive isotopes).

    Radiopharmaceuticals are used or diagnosis

    or therapeutic treatment o human diseases;

    hence nearly 95% o radiopharmaceuticals are

    used or diagnostic purposes, while the rest is

    used or therapy.

    Radiopharmaceuticals usually have no phar-

    macologic eects, as they are used in tracer

    quantities. There is no dose-response relation-

    ship in this case, which thus diers signicantly

    rom conventional drugs.

    Radiation is an inherent characteristic o all

    radiopharmaceuticals, and patients always re-

    ceive an unavoidable radiation dose. In the case

    o therapeutic radiopharmaceuticals, radiationis what produces the therapeutic eect.

  • 7/28/2019 Tech Radiopharmacy

    7/52

    EA

    NM

    7

    Chapter 1 Radiopharmacy Technology

    Depending on the radiation characteristics

    o the radionuclide, the radiopharmaceutical

    is used either or diagnosis or or therapy. Di-

    agnostic radiopharmaceuticals should decay

    by gamma emission or positron emission, and

    never emit alpha particles or even beta par-

    ticles. On the other hand, therapeutic radio-

    pharmaceuticals should decay by particulate

    decay (alpha or beta) since the intended eectis in act radiation damage to specic cells.

    Gamma radiation is characterised as electro-

    magnetic radiation. When used in diagnostic

    radiopharmaceuticals, the nally produced

    gamma rays should be powerul enough to

    be detected outside the body o the patient.

    The ideal energy or conventional (SPECT)

    nuclear medicine equipment is around 150keV. Normally, these radiopharmaceuticals are

    in such small quantities that the radiation dose

    received by the patient can be compared to

    that o a simple radiology investigation.

    Alpha decay is characterised by the emission o

    an alpha particle rom the nucleus. This particle

    is a helium ion containing two protons and two

    neutrons. In beta decay a negatively charged

    particle with the same charge and mass as an

    electron is emitted. Alpha emitters, which are

    monoenergetic and have a very short range

    in matter due to their mass, thus leaving much

    o its energy on a very small area (only a ew

    cell diameters), are used only or therapeutic

    purposes. Their clinical use is very limited, and

    they are mainly used or research purposes, orstill are in early phase clinical trials.

    Neutron rich radionuclides disintegrate by

    beta (-) decay. Beta emitters represent di-

    erent energy levels, and have dierent range

    in matter (40 100m) depending on their

    energy. Beta emitting radionuclides are also

    used in radiopharmaceuticals mainly or thera-

    peutic purposes.

    Positron (+

    ) decay occurs in proton rich nuclei.The range o a positron is very short in matter.

    At the end o the path o +-particles, positrons

    combine with electrons and are thus annihilat-

    ed, giving rise to two photons o 511 keV that

    are emitted in opposite directions. Positron

    emitters are used to label radiopharmaceuti-

    cals or diagnostic purposes by imaging.

    Radioactivity unitsRadioactivity is expressed in Becquerels (Bq)

    as the SI-unit. One Becquerel is dened as one

    disintegration per second (dps). Normally, ac-

    tivities used in radiopharmacy are in the range

    o megabecquerels (MBq) or gigabecquerels

    (GBq). There is a non-SI-unit or radioactivity

    called Curie (Ci), which is used in some occa-

    sions. One Ci represents the disintegration o

    one g o radium. The equivalence between

    the Bq and the Ci is as ollows:

    1 Bq = 2,7 x 10-11 Ci

    1 Ci = 37 GBq

    Every radionuclide is characterised by a hal-lie,

    which is dened as the time required to reduce

    its initial activity to one hal. It is usually denotedby t

    , and is unique or a given radionuclide.

  • 7/28/2019 Tech Radiopharmacy

    8/52

    8

    Principles o radiation protection

    Production, transportation and use o radio-

    pharmaceuticals, as radioactive products, are

    governed by regulatory agencies dealing with

    radiation protection and nuclear saety. Only

    licensed personnel in an authorised acility

    are authorised to handle and use radiophar-

    maceuticals.

    The general principles oradiation protec-

    tion are:

    Justication: All procedures involving ra-

    dioactive material must be justied.

    Optimisation: The radiation exposure to

    any individual should be as low as reason-

    ably achievable. This principle is the widelyknown ALARA concept (as low as reason-

    able achievable).

    Limitation: The radiation dose received

    by the personnel handling radioactive

    material will never exceed the legally es-

    tablished dose limits.

    When planning acilities and procedures or

    handling o radioactive materials according

    to the ALARA principle, it is important to keep

    in mind the basic principles or reduction o

    radiation doses:

    Time: The shorter the time o exposure to

    radiation, the lower the dose to the op-

    erator.

    Distance: The radiation dose decreases

    with a actor equal to the square root o

    the distance rom the radiation source. The

    operators distance rom the source can

    be increased by using orceps, tongs, ormanipulators in handling the radioactive

    material.

    Shielding: The radiation dose can be re-

    duced by placing shielding material be-

    tween the source and the operator. For

    protection against gamma radiation, walls

    made o heavy concrete or lead bricks are

    used. For transport containers, materialsuch as tungsten may be used or higher

    energy gamma irradiation radionuclides,

    giving a higher shielding per weight unit

    when compared to lead.

    Formulation and production o

    radiopharmaceuticals

    When designing a radiopharmaceutical, one

    should have in mind the potential hazard the

    product may have to the patient. The goal

    must be to have a maximum amount o pho-

    tons with a minimum radiation exposure o

    the patient.

  • 7/28/2019 Tech Radiopharmacy

    9/52

    EA

    NM

    9

    Chapter 1 Radiopharmacy Technology

    The unction o the carrier molecule in a ra-

    diopharmaceutical is to carry the radioactiv-

    ity to the target organ, and to make sure the

    radioactivity stays there. The uptake o radio-

    activity should be as specic as possible, in

    order to minimise irradiation o other organs

    and parts o the body. This is particularly im-

    portant when using radiopharmaceuticals or

    therapy. But also or use in diagnostics, it isdesirable that the radiopharmaceutical is lo-

    calised preerentially in the organ under study

    since the activity rom non-target areas can

    obscure the structural details o the pictures

    o the target organ. It is thereore important

    to know the specic uptake in an organ or a

    potential chemical carrier, and also the rate o

    leaking out o the organ/organ system. Thus,

    the target-to-background activity ratio shouldbe large.

    In a radiolabelled compound, atoms or groups

    o atoms o a molecule are substituted by simi-

    lar or dierent radioactive atoms or groups

    o atoms. When a labelled compound is to

    be prepared, the rst criterion to consider is

    whether the label can be incorporated into the

    molecule to be labelled. This may be assessed

    rom knowledge o the chemical properties o

    the two partners. Furthermore, one needs to

    know the amount o each component to be

    added. This is particularly important in tracer

    level chemistry and in 99mTc-chemistry.

    As the radiolabelled substances emerge rom

    the laboratory to the clinics, there will be a

    need or scaling up the batch size o the prod-

    uct. This can be done either by increasing the

    total volume o the produced batches or by

    increasing the specic activity o the product,

    or both. When doing this, one has to consider

    two important aspects:

    The infuence on the stability o the prod-

    uct itsel due to possible radiolysis.

    The need or additional operator protec-

    tion due to handling o increased amounts

    o radioactivity.

    Manuacturing o radiopharmaceuticals is

    potentially hazardous. Both small- and large-scale production must take place on prem-

    ises designed, constructed, and maintained

    to suit the operations to be carried out. Ra-

    diation protection regulations stipulate that

    radionuclides must only be used in specially

    designed and approved radioisotope labora-

    tories. National regulations with regard to the

    design and classication o radioisotope labo-

    ratories must be ullled. Such laboratories are

    normally classied according to the amount

    o the various radionuclides to be handled at

    any time, and the radiotoxicity grading given

    to each radionuclide.

  • 7/28/2019 Tech Radiopharmacy

    10/52

    10

    Premises must be designed with two impor-

    tant aspects in mind:

    The product should not be contaminated

    by the operator.

    The operator and the environment should

    be protected rom contamination by the

    radioactive product.

    This is the basic principle o GRP Good Radio-

    pharmaceutical Practice.

    Quality considerations

    The key elements o GRP comprise a previ-

    ously dened manuacturing process known

    to lead to a radiopharmaceutical o the de-

    ned quality administered to the patient inthe prescribed dosage and orm. GRP is carried

    out and recorded by trained and qualied sta

    provided with the necessary acilities, includ-

    ing adequate premises, suitable equipment,

    correct materials and established procedures

    in written orm. Quality must also be main-

    tained during transportation and storage.

    Training and qualications should cover gen-

    eral principles o GMP (Good Manuacturing

    Practice) and radiation protection. All training

    must be recorded. Premises and equipment

    must have a layout and design that minimise

    the risks o errors by avoiding cross-contam-

    ination and build up o dust and dirt, as well

    as permit eective cleaning and maintenance.

    They must also be designed to give proper

    radiation protection to personnel and the

    environment. Documentation is an essential

    part o a quality system. Its purpose is to dene

    the control system, to reduce the risk o error

    that is inherent in oral communication and

    to ensure that detailed instructions are avail-

    able to personnel. The documentation system

    should allow tracking o use and disposal o

    any batch.

    Radiopharmaceuticals are a special orm o

    drugs that require much more handling im-

    mediately beore administration to the pa-

    tient, when compared to other drugs. Due

    to the short hal lie o the radionuclide, it is

    necessary or the nal preparation o many

    radiopharmaceuticals to take place shortly

    beore use. Only a minor proportion o all ra-diopharmaceuticals is delivered to hospitals in

    a ready-or-use-orm. Handling o radiophar-

    maceuticals in hospitals is thus an integral part

    o the system by which the quality o these

    pharmaceuticals is established.

    Quality control o radiopharmaceuticals

    All quality control procedures that are ap-

    plied to non-radioactive pharmaceuticals are

    in principle applicable to radiopharmaceu-

    ticals. In addition, tests or radionuclidic and

    radiochemical purity must be carried out.

    Furthermore, since radiopharmaceuticals are

    short-lived products, methods used or qual-

    ity control should be ast and eective. Still,

    some radiopharmaceuticals with very short

    hal-lives may have to be distributed and

  • 7/28/2019 Tech Radiopharmacy

    11/52

    EA

    NM

    11

    Chapter 1 Radiopharmacy Technology

    used ater assessment o batch documenta-

    tion even though all quality control tests have

    not been completed.

    Hospital departments dealing with radiop-

    harmaceuticals should have a programme

    or quality control o products beore admin-

    istration to the patient. The complexity o

    the quality control depends on whether theproduct is a ready-or-use-orm, or the product

    is labelled in the department prior to admin-

    istration (labelling kits).The specications and

    quality control testing procedures or most o

    the currently used radiopharmaceuticals are

    given in the European Pharmacopeia, or other

    Pharmacopeia (BP, USP etc.). For labelling kits,

    a simple quality control procedure should be

    stated in the package insert or the particularproduct. The quality control system should

    include a procedure which describes mea-

    sures to be taken i unsatisactory test results

    are obtained.

  • 7/28/2019 Tech Radiopharmacy

    12/52

    12

    Chapter 2 Radiopharmacy DesignJames R. Ballinger, PhD

    The radiopharmacy should be conveniently

    located or deliveries and shipments (i supply-

    ing external units). There should be arrange-

    ments or receipt o out o hours deliveries,

    such as a locked cupboard adjacent to the

    unit but not requiring direct access to the

    radiopharmacy by unauthorised personnel

    such as couriers.

    Layout

    Restricted access is important, rom both a

    radiation security and pharmaceutical manu-

    acturing point o view. Only persons with

    business in the radiopharmacy should be

    allowed access. Within the radiopharmacy,

    there will be urther restriction o access to

    the clean areas. The principle is moving rom

    dirty to clean areas with appropriate changeo apparel and sanitation o materials at each

    interace. The dirty areas include delivery and

    dispatch, an oce or preparation o paper-

    work, a supplies store, a waste store, and a

    QC laboratory. It is particularly important to

    unpack supplies in the dirty area where there

    is bulk storage; only minimal supplies are kept

    in the clean area. Some radiopharmacies may

    have a separate area or handling o131I. The

    clean areas include a 99mTc dispensing room,

    a support room, and a separate blood cell la-

    belling acility. The layout should enable an

    orderly fow o work, both within and between

    rooms.

    Facilities

    Although many o the principles o radiophar-

    macy design are universal, there may be di-

    erences between countries in how rigorously

    these principles are regulated. The radiation

    aspects are covered in the EC Euratom Direc-

    tive [1] while pharmaceutical manuacturing

    is controlled under EudraLex [2]. The EANM

    Radiopharmacy Committee has issued guid-ance on Good Radiopharmacy Practices [3]

    which addresses both aspects. With respect

    to both acilities and operations, there can be

    conficting requirements between radiation

    saety and aseptic processing. Another impor-

    tant consideration is complete segregation

    between radiopharmaceutical preparation

    (aseptic processing) and blood cell labelling,

    to minimise the risk o cross contamination.

    In general, the radiopharmacy should be at

    one end o a nuclear medicine/radiology de-

    partment rather than in the middle. Indeed,

    it is best i it is on an outside wall as there

    is less concern about shinethrough o radia-

    tion. Although high levels o radioactivity are

    handled, in most cases local shielding is used

    (e.g. around generators and individual vials)

    rather than extensive shielding in walls.

  • 7/28/2019 Tech Radiopharmacy

    13/52

    EA

    NM

    EA

    NM

    13

    Chapter 2 Radiopharmacy Design

    Figure 1 presents a sample layout o a large

    radiopharmacy with ideal separation o di-

    erent working areas required or handling

    and radiolabelling o99mTc and other SPECT

    radiopharmaceuticals.

    Figure 1: Sample layout o a radiopharmacy

    Equipment and ttings

    Standard equipment includes one or more

    dose calibrators (ionisation chambers), labora-

    tory equipment (e.g. balance, centriuge), and

    appropriate radioanalytical equipment (radio-

    chromatogram scanner, gamma counter; more

    advanced laboratories may have a multichannel

    analyser and radio-HPLC system). This equip-

    ment should be dedicated to use in the radiop-

    harmacy and not shared with outside users.

    Radiation survey meter(s) must be available to

    check or contamination within the unit and at

    the boundary o the radiation controlled area. I

    radioactive material is shipped, there must be

    a dose rate monitor (i.e. calibrated in Sv/h) to

    allow determination o the Transport Index.

    Rerigerators and other areas where sensitive

    supplies are stored should have temperature

    monitoring. This can be as simple as manual

    recording on a daily basis o minimum and

    maximum temperatures on a calibrated ther-

    mometer or an electronic output to a chart

    recorder or monitoring sotware.

    Volatile materials such as131

    I products and or-ganic solvents should be handled in a ume

    cupboard with a minimum infow o 0.5 m/s.

    The exhausted air is vented to the atmosphere

    and care must be taken in the location o the

    stack. Filters are not generally used as they

    could become radiation hazards themselves.

    The clean areas should be lined (foor, walls, and

    ceiling) with a smooth, continuous, impervious,non absorbent, cleanable material such as weld-

    ed sheet vinyl. Corners should be coved (curved)

    to minimise dirt collection. Light xtures should

    be recessed and fush with the surace. Benches

    must be made o impervious material (solid is

    preerred over laminate) and may require ad-

    ditional support or lead shielding.

    There should be transer hatches with inter-

    locking doors so supplies can be sanitised and

    passed into the next room without allowing

    direct contact. Entry/change rooms should

    have interlocking doors and a physical barrier,

    or at least a line on the foor, to demarcate the

    two sides. Changing on entry will involve, at a

    minimum: clean low-lint lab coat, shoe covers,

    hair cover, and gloves.

  • 7/28/2019 Tech Radiopharmacy

    14/52

    14

    Aseptic manipulations should be perormed

    either in a pharmaceutical isolator or a laminar

    airfow hood.

    The bulk o the work or the oreseeable uture

    will continue to be reconstitution o kits with99mTc pertechnetate. Thus, a single workstation

    is adequate even i there is occasional handling

    o other radionuclides (e.g. preparation o111

    Inpentetreotide or 90Y ibritumomab tiuxetan).

    However, i the usage o other radionuclides is

    more extensive, a separate workstation should

    be provided. In addition to minimising the po-

    tential o cross contamination o99mTc prod-

    ucts with longer lived and/or particle emitting

    radionuclides, it also reduces the risk that a

    major spill o one o these radionuclides could

    impede 99mTc dispensing or a number o days.In the uture, some products such as 90Y or177Lu labelled peptides might be prepared by

    automated synthesis units located in a sepa-

    rate workstation.

    In general, radiation saety is maintained by

    local shielding (e.g. vial shields, syringe shields,

    bench top shields) rather than shielding in the

    walls. The waste store may require shielding, as

    may an area where high levels o radioactivity

    are handled i there is a signicant radiation

    eld in the adjacent room. The 99Mo/99mTc

    generator usually requires additional exter-

    nal shielding.

    It is simplest to dispose o all sharps into rigid

    biohazard containers behind a lead shield.

    Once the radioactivity has decayed to back-

    ground levels, the containers are disposed as

    biohazard waste. Radionuclides should be

    segregated by hal lie to minimise the build

    up o waste.

    The blood labelling suite will require a vari-able speed centriuge capable o accepting a

    range o tube sizes. Ideally the centriuge will

    be located within the workstation, to minimise

    the number o transers out o the Grade A

    environment.

    Area designation

    A radiation controlled area is one in which a

    ull time worker might receive an exposure o6 mSv/yr whole body or 150 mSv/yr extrem-

    ity. For a radiation supervised or monitored

    area, the exposure limits are 1 mSv/yr whole

    body or 50 mSv/yr extremity. Because o the

    presence o generators and the amount o

    radioactivity handled, the hot lab will be des-

    ignated a controlled area. Other areas may be

    classied as supervised.

    Radiation designated areas must be physically

    demarcated and have signs indicating the

    type o hazard. There should be washing and

    changing acilities available at the perimeter;

    and radiation monitoring or contamination

    must be perormed. Eating and drinking is

    prohibited in designated areas.

  • 7/28/2019 Tech Radiopharmacy

    15/52

    EA

    NM

    EA

    NM

    15

    Chapter 2 Radiopharmacy Design

    Conicts between radiation and aseptic

    regulations

    As noted above, there can be conficts be-

    tween the requirements o dierent regula-

    tory systems. However, there are also areas o

    agreement. For example, segregation o ac-

    tivities, with change o apparel and dedicated

    equipment, minimises cross contamination

    with both radioactivity and microbes, as doesa separate air handling system and the use

    o containment hoods/glove boxes. Special-

    ist trained sta are required and meticulous

    record keeping is important.

    However, even within these areas, there are

    conficts. For radioprotection, the production

    area should be at negative pressure relative

    to the outside world (containment o gaseousor aerosol discharge), while pharmaceutical

    aseptic units are at positive pressure to mini-

    mise ingress o microbes. The compromise is

    a negative pressure isolator within a positive

    pressure room. From a pharmaceutical point

    o view, there should be a minimum number

    o trained sta, whereas or radioprotection

    there should be a rotation o sta to share

    the radiation dose. Radioprotection requires

    handwashing acilities at the perimeter o

    the controlled area, while the medicines in-

    spectors dont want a sink anywhere near a

    cleanroom. Radiopharmacy managers must

    nd a delicate balance to satisy both sets o

    inspectors.

  • 7/28/2019 Tech Radiopharmacy

    16/52

    16

    Chapter 3 RadiopharmacyPreparing & Dispensing Radiopharmaceuticals

    Geraldine OReilly, PhD

    Most radiopharmaceuticals are administered

    by IV injection; so good pharmaceutical prac-

    tice is an important consideration in their

    preparation. All manipulation o radioactive

    materials should be carried out, using asep-

    tic techniques, within the shielded contained

    workstation or laminar fow cabinet (LAFC)

    (Figure 1). No ood or drink, cosmetic or smok-

    ing materials, crockery or cutlery should bebrought into an area where unsealed radioac-

    tive substances are used.

    Figure 1: Laminar fow unit or preparation o

    99mTc radiopharmaceuticals

    Personnel monitoring

    All sta classied as radiation workers must

    wear a personal dosimeter (TLD, lm badge,

    electronic dosimeter). In addition to their regu-

    lar whole body dosimeter, sta preparing and

    handling radioactive materials should wear a

    nger TLD to monitor extremity dose. Prior to

    each use, the TLD should be wiped, using an

    alcohol wipe, and worn inside the glove. Upon

    leaving the preparation area, the nger TLDshould be removed and appropriately stored.

    Introduction

    The radiopharmacy would normally be desig-

    nated as a controlled area; and access will be

    restricted. Only properly trained sta should

    be permitted to work in the radiopharmacy;

    and strict adherence to work procedures is

    essential. There are three undamental pa-

    rameters that aect sta doses in the radio-

    pharmacy:

    1. the distance between the sta member

    and the source,

    2. the time spent manipulating the source

    and

    3. the amount o shielding used to reduce

    the dose rate rom the source.

    Sometimes there is a trade o between theseparameters as using more shielding might in-

    crease handling time. With this in mind, care-

    ul design o procedures should optimise the

    workfow. Skill and expertise o the sta carry-

    ing out the procedures are also important ac-

    tors. Thus, it is crucial that sta are adequately

    trained.

    Work practices in the radiopharmacy should

    be standardised and incorporated in standard

    operating procedures (SOPs). These proce-

    dures should be documented and made

    readily available to those working in the ra-

    diopharmacy. This will ensure harmonisation

    o practice and maintenance o standards. Ac-

    curate and comprehensive record keeping is

    an essential part o good work practice in aradiopharmacy.

  • 7/28/2019 Tech Radiopharmacy

    17/52

    EA

    NM

    17

    Chapter 3 Radiopharmacy: Preparing & Dispensing Radiopharmaceuticals

    Protective clothing

    Prior to commencing work in the radio-

    pharmacy, sta should ensure that they wash

    their hands thoroughly. Beore a person en-

    ters an area where radioactive substances are

    handled, any cut or break in the skin should

    be covered. Dressings should incorporate a

    waterproo adhesive strapping. Protective

    coats or gowns should be worn or prepara-tion and dispensing o radiopharmaceuticals.

    Disposable gowns oer benets in terms o

    maintaining sterility. Gloves worn in the LAFC

    or contained workstation must be powder ree

    in order to prevent clogging o the air lters

    within the cabinet. Alcohol rub should be

    rubbed onto gloves and allowed to evapo-

    rate beore entering the LAFC. Ater handling

    radioactive materials, gloves must alwaysbe removed and disposed o as radioactive

    waste beore handling/touching any other

    materials/suraces within the radiopharmacy.

    Hands should be washed again ater removal

    o gloves. Upon leaving the radiopharmacy,

    disposable gowns should be removed. Prior to

    disposal, they should be stored as radioactive

    waste until monitoring conrms that they are

    at background radiation levels.

    Protective equipment

    The use o protective equipment, when han-

    dling radioactive materials, can have a signi-

    cant impact on reducing sta dose. Laborato-

    ries and other work areas or manipulation o

    unsealed radioactive substances should be pro-

    vided with equipment kept specically or thispurpose, and should include the ollowing:

    1. tools or maximising the distance rom the

    source, or example tongs and orceps,

    2. syringe shields,

    3. vial shields,

    4. drip trays or minimising the spread o con-

    tamination in the case o spillage,

    5. shielded syringe carriers and

    6. decontamination kit

    Unshielded syringes or vials should never be

    used during manipulation o radiopharma-

    ceuticals. Equipment should be stored outside

    the laminar fow cabinet when not in use and

    should be cleaned regularly in accordance

    with local recommendations.

    Work procedures

    GeneralBeore starting the preparation and dispensing

    o radiopharmaceuticals, all o the materials

    required should be assembled and placed in

    or close to the contained workstation/LAFC

    (Figure 2). All vials containing radioactive

    materials must be shielded while handling;

    and vials should only be removed rom their

    shields or assay, inspection or disposal. All sy-

    ringes containing radioactive liquids must be

    shielded while handling, except during an as-

    say. Unshielded vials or syringes should not be

    handled directly. Long handled tongs should

    be used to place and remove unshielded ma-

    terials in the dose calibrator.

  • 7/28/2019 Tech Radiopharmacy

    18/52

    18

    Work procedures should be designed so as

    to minimise exposure rom external radiation

    and contamination. Care must be taken to pre-

    vent spillage rom occurring. All manipulation

    or dispensing radioactive materials should

    be carried out over a drip tray, in order to

    minimise the spread o contamination due

    to breakages or spills. Should a spill occur then

    it should be cleaned up beore proceedingany urther. All items that might be contami-

    nated should be removed rom the aected

    area and stored saely. Care should be taken

    doing this, in order to minimise the spread

    o contamination. As with all spills, it is more

    convenient to allow natural decay to take care

    o the contamination, i the items are not re-

    quired immediately. For those items that are

    needed, they should be cleaned with alcoholswabs taking care not to spread the contami-

    nation. Using multiple swabs which are then

    disposed is the most eective way to remove

    contamination.

    Figure 2: Preparation o work area

    (courtesy o VirRAD)

    Reconstitution o pharmaceuticals

    The manuacturers recommendations should

    be ollowed closely as many pharmaceutical

    kits have specic reconstitution instructions in

    terms o the activity and volume to be added

    to the kit. Recommended incubation times

    also vary and must be adhered to. Some ra-

    diopharmaceuticals must be rerigerated ater

    preparation. Thereore consideration should begiven to the provision o suitably shielded re-

    rigeration acilities. Most radiopharmaceuticals

    are reconstituted with 99mTc; and this assump-

    tion applies to the ollowing paragraphs.

    Protective caps should be removed rom the

    pharmaceutical vials; and the vials should be

    placed in the appropriate labelled vial shields.

    The rubber septum o each pharmaceutical vialshould be swabbed with alcohol; and the alco-

    hol should be allowed to evaporate (Figure 3).

    Shielded 10ml or 5ml syringes capped with

    21G needles are generally used to recon-

    stitute the pharmaceuticals.

    The appropriate activity o 99mTc solution

    should be added to each shielded phar-

    maceutical vial; and the pharmaceutical

    should be allowed to incubate or the

    specied length o time.

    Having introduced 99mTc solution to a pharma-

    ceutical vial, the needle should not be placed

    back into the shielded elution vial. In the event

    that additional

    99m

    Tc solution is required, a newsyringe and needle must be used.

  • 7/28/2019 Tech Radiopharmacy

    19/52

    EA

    NM

    19

    Chapter 3 Radiopharmacy: Preparing & Dispensing Radiopharmaceuticals

    When introducing 99mTc solution or saline

    to a vial, it may be necessary to equalise

    pressure by withdrawing an equivalent

    volume o air at the same time. This can

    be done gradually as the solution is added.

    In general, breather needles are not recom-

    mended and should not be used unless

    specically recommended by the manu-

    acturer o the pharmaceutical.

    The activity and volume o 99mTc solution

    added to each pharmaceutical should be

    recorded in the laboratory log book.

    Figure 3: Disinection o a kit beore use

    (courtesy o VirRAD)

    Preparation o patient injections

    Ater the recommended incubation time

    has elapsed, patient activities are with-

    drawn using shielded 2ml syringes capped

    with 21G.

    Each patient activity must be measured

    and recorded in the radiopharmacy log.

    The activity withdrawn or each patientmust be within 10% o the required activ-

    ity at the specied injection time.

    Patient injections are usually prepared in

    a volume o 1ml. There are exceptions to

    this: the manuacturers instructions on the

    volume should be ollowed. Saline may be

    used to increase the volume i the volume

    in which the required activity is obtainedis below 1ml.

    Once the patient injection is prepared,

    the green needle must be replaced with a

    needle o the appropriate gauge, and the

    air in the syringe must be expelled. When

    expelling the air, ensure that the needle

    is capped.

    When changing needles, withdraw the

    plunger suciently so as to pull liquid rom

    the syringe tip.

    I, at any time, there is a droplet o liquid vis-

    ible in the needle cap, replace the needle

    and cap.

  • 7/28/2019 Tech Radiopharmacy

    20/52

    20

    Each patient injection must be labelled

    with an appropriate label detailing the

    patient name, scan type, activity to be ad-

    ministered, date and time o injection.

    Waste management procedures

    Non-radioactive waste should be separated

    rom radioactive waste to minimise storage

    requirements; and it should be disposed oas normal hospital waste. Shielded waste bins

    should be lined with plastic liners that can be

    easily removed when ull.

    Technetium 99m is the main isotope in use

    in the radiopharmacy: the duration o stor-

    age will be determined by its hal lie o 6.02

    hours. Longer lived waste should be stored

    separately. Radioactive waste generated dailywithin the radiopharmacy includes syringes,

    elution vials, pharmaceutical vials, needles and

    swabs. Waste arising rom the preparation and

    dispensing o radiopharmaceuticals should

    be primarily disposed in the waste bin built

    into the contained workstation/LAFC. Some

    bulky items such as paper waste and gloves

    may be disposed in a shielded waste bin in the

    pharmacy, as long as there is no risk o con-

    taminating the room by removing them rom

    the cabinet. Radioactive waste contaminated

    by blood (e.g. syringes ollowing cell labelling

    procedures) should not be let in the worksta-

    tion but removed to a shielded bin.

    The waste container in the workstation should

    not be allowed to overfow and should be

    emptied regularly. The bin is best emptied

    beore starting work in the cabinet, when the

    waste in the bin has decayed over night.

    Segregation o waste according to hal-lie

    is good practice and can reduce the length

    o time that waste arising rom shorter livedisotopes has to be stored.

    Paper waste

    Any gloves used in the cabinet or used to

    handle blood or isotopes will be considered

    to be contaminated. Paper tray liners in the

    cabinet or paper used to clean suraces in

    the cabinet are also considered to be con-

    taminated. Contaminated gloves and papershould be disposed in a shielded bin in the

    pharmacy, as long as there is no biological

    contamination (blood or plasma). Otherwise

    this waste should be placed in a sharps bin, us-

    ing tongs. For long term storage o waste, the

    waste should be removed rom the shielded

    bin, labelled with details o the contents and

    stored as radioactive waste in a designated

    store.

    Sharps bins

    Syringes, needles, butterfies etc. should be

    disposed o ater use to shielded sharps bins.

    Bins should not be allowed to overll. Full

    sharps bins should be closed, marked radio-

    active, dated and removed to the radioactive

    waste store.

  • 7/28/2019 Tech Radiopharmacy

    21/52

    EA

    NM

    21

    Chapter 3 Radiopharmacy: Preparing & Dispensing Radiopharmaceuticals

    Disposal o waste

    All radioactive waste - sharps bins, paper

    waste, ventilation kits - should be securely

    stored and monitored regularly. Waste should

    be checked by using a suitable meter in a low

    background environment and should be dis-

    posed o, once it has decayed to background

    level. Any items above background should

    be retained or a urther period o decay instorage. All radioactive warning labels should

    be removed rom waste, prior to disposal in

    hospital waste. The hospital waste disposal

    policy should be adhered to.

    Dose calibrator quality assurance

    The accuracy and constancy o the dose

    calibrator should be checked regularly with

    a reerence source. Isotopes such as Co-57,Ra-226 or Cs-137 with a relatively long hal lie

    are suitable. The use o more than one source

    will allow checking o the calibrator over a

    range o energies. A daily check o system

    operation, accuracy and constancy should

    be carried out. Linearity tests may be carried

    out less requently.

    Beore starting the QA, remove all radioactive

    sources rom the vicinity o the calibrator and

    record the background reading. The isotope

    selected on the dose calibrator should be that

    o the reerence source. The accuracy test en-

    sures that the activity is within 10 percent o

    a given calibrated reerence source o known

    activity (within 5%). At least two sealed sourc-

    es with dierent principal photon energies,

    one o which has a principal energy between

    100 keV and 500 keV, should be used to deter-

    mine accuracy upon installation, and at least

    annually thereater. For the routine accuracy

    check, place the reerence source in the dose

    calibrator and record the activity measured.

    The value recorded should be within the al-

    lowable tolerance (typically 10%).

    The constancy test looks at reproducibility in

    measuring the activity o a known source over a

    long period o time. The dose calibrator should

    be checked daily or constancy at the setting o

    the most requently used isotope. To carry out

    this test, the reerence source is let in place; but

    the setting is changed to Tc-99m and the value

    indicated is recorded. This can also be done

    or other isotopes that are selectable on thedose calibrator. The ratio o the values indicated

    to that o the activity o the reerence source

    should be constant over time.

    The linearity test ensures that the dose calibrator

    can indicate the correct activity over the range

    o use o administered or measured activities.

    The dose calibrator should be tested or linearity

    upon installation and at least quarterly therea-

    ter. Technetium-99m is most requently used or

    the linearity test because o its availability and

    short hal-lie. The variation between indicated

    and known activity should not exceed 10%.

    The results o the routine QA should be docu-

    mented and stored as part o the radiophar-

    macy records.

  • 7/28/2019 Tech Radiopharmacy

    22/52

    22

    Chapter 4 RadiopharmacyKits & Techniques

    Helen Ryder

    patient movement; or too short a time per

    image resulting in poor count statistics. I

    the hal-lie is too long, then there may be

    excessive radiation exposure to the patient

    over its period o decay.

    3. High target - background ratio

    The radionuclide is only o use i it accu-

    mulates within the target organ, and this

    is oten enhanced by binding the radionu-clide onto a tracer that will take it to the

    organ under study. The binding should be

    sucient that little radionuclide is let ree

    in the body tissues, thus enhancing the

    target-to-background ratio.

    4. Low dose rate to both patient and per-

    sonnel

    To avoid excessive received radiation dose,

    it is necessary to avoid those radionuclideswhich have signicant particulate emis-

    sions i.e. alpha and beta particles. The short

    range o emission means that these are

    absorbed within the patient, adding to the

    radiation dose with no increase in image

    quality. Because o the reality o radiation

    decay, that is the attempt to balance the

    particles in the nucleus, beta rays are oten

    ound as a product o decay. The rate o

    emission o these rays can be signicantly

    lower in two particular decay processes:

    isomeric transition and electron capture.

    5. Non-toxicity of radiopharmaceuticals

    Most radiopharmaceuticals must be inject-

    ed, with a small amount being ingested or

    inhaled. Thus they must be non-toxic in

    nature, sterile and pyrogen-ree.

    Radiopharmaceuticals have been dened as

    radioactive drugs that, when used or the

    purpose o diagnosis or therapy, typically elicit

    no physiological response rom the patient.

    In the main this is true, and though some

    radiopharmaceuticals have been known to

    cause minor side eects (such as urticaria,

    or changes in blood pressure), these are notcommonly seen in practice.

    Ideal radionuclide

    Properties o the ideal diagnostic radionuclide

    include:

    1. Pure gamma emitter, with a gamma en-

    ergy within the range of 100 - 250 keV, to

    match the optimum scanning range of agamma camera.

    The radiation emitted by a radionuclide

    should be sucient to be detected outside

    the patient or imaging purposes, thus lim-

    iting the choices to X-rays or gamma rays.

    However, too high an energy will result in

    the gamma ray penetrating the detector o

    the imaging device without being stopped

    and hence recorded.

    2. A half-life which is suitable for diagnostic

    use i.e. 1.5 X test duration

    The hal-lie o a radionuclide determines

    the rate o radioactive decay. I the hal-lie

    is very short, then the activity may have

    decayed to a very low level beore imag-

    ing can be started. This can result in either

    a long scanning time, where there may be

  • 7/28/2019 Tech Radiopharmacy

    23/52

    EA

    NM

    23

    Chapter 4 Radiopharmacy: Kits & Techniques

    6. Chemical stability during use

    Not all radioisotopes are suitable or binding

    to tracers. Technetium 99m has the ability to

    readily bind to a wide variety o compounds

    under physiological conditions [1] without

    causing physiological changes in the patient.

    7. Inexpensive and readily available

    Many suitable radionuclides are obtainable

    rom a generator, which may be deliveredto a nuclear medicine acility and eluted as

    required. This results in economical use o

    the radionuclide.

    Other radionuclides are produced in cyclo-

    trons as specialised units and are shipped

    ready or use. These are decaying rom the

    moment they are manuactured and thus

    must be ordered or use on a particular

    day, either as the resultant product or orpreparation with other tracers. This can be

    expensive and uneconomical or everyday

    requirements.

    8. Ease of preparation and appropriate

    quality control

    Preparation requirements o more than three

    steps do not usually meet the denition o

    ease o preparation. Nor should a complex

    variety o equipment be required. Quality

    control procedures should be available to

    check each batch o the radiopharmaceuti-

    cal reconstituted in the working laboratory.

    This ensures that the preparation received by

    the patient will result in high-quality images

    without detrimental eect on the patient.

    Radionuclide generator

    A radionuclide generator is a source o radio-

    nuclides or the preparation o radiopharma-

    ceuticals. It is based on the separation o a

    short lived radionuclide (daughter) rom a

    long lived radionuclide (parent). Examples o

    radionuclide generators are provided in Table

    1. The most commonly used generator in Nu-

    clear Medicine is the99

    Mo/99m

    Tc generator.

    Table 1: Radionuclide generators

    Molybdenum/Technetium (99Mo/99mTc)

    generator

    The molybdenum/technetium generator con-

    sists o an alumina-lled column onto which is

    absorbed 99Mo. The 99Mo is present as 99MoO4

    2-,

    which decays to its daughter radionuclide 99mTc

    as pertechnetate 99mTcO4

    - (Figure 1). 99mTc is re-

    moved rom the columns as 99mTcO4

    - by draw-

    ing over a solution o sodium chloride (NaCl)

    0.9% w/v across the column (Figure 2). This

    process is known as eluting the generator and

    the resultant eluate is used to compound the

    radiopharmaceuticals (Figure 3).

  • 7/28/2019 Tech Radiopharmacy

    24/52

    24

    Figure 1: Decay scheme or 99Mo and 99mTc

    (courtesy o VirRAD)

    Figure 2: Molybdenum/Technetium generator

    Figure 3: Eluting the generator (courtesy o VirRAD)

    The time o maximum yield o99mTechnetium is

    23 hours, ater which the 99mTc appears to decay

    with the hal-lie o99Mo (66hrs). This time o

    almost one day is thereore eminently suitable

    or the requirements o the Nuclear Medicine

    department (Figure 4).

    Attach Saline Vial Insert Elution vial

    in lead shielding

    Attach Elution vial and wait

    until elution is completed

    Remove elution vial and attach

    protecting vial/Cap

  • 7/28/2019 Tech Radiopharmacy

    25/52

    EA

    NM

    25

    Chapter 4 Radiopharmacy: Kits & Techniques

    Figure 4: Plot o logarithm o99Mo and 99mTc

    activities against time showing transient

    equilibrium. The generator is eluted daily and

    the 99mTc activity grows again until transient

    equilibrium is reached (courtesy o VirRAD).

    Preparation o99mTc-

    radiopharmaceuticals rom kits

    Categories o radiopharmaceuticals:Ready-to-use radiopharmaceuticals

    Instant kits or preparation o 99mTc products

    Kits requiring heating

    Products requiring signicant manipulation

    Reconstitution o proprietary kits:

    The basic steps involved in the reconstitution

    o99mTc radiopharmaceuticals rom generators

    and kits are outlined in Figure 5. Appendix 1

    urther details the procedure or reconstitu-

    tion o proprietary kits. An example o how

    to calculate the required activity is provided

    in Appendix 2.

    Figure 5: Reconstitution o99mTc

    radiopharmaceuticals rom generators and kits

    Appendix 1: Reconstitution o proprietary

    kits with 99mTc

    Record the ollowing inormation in the

    injection log:

    The batch number and expiry date o

    the elution vial, pharmaceutical vials

    and saline bottles.

    The activity and volume o the eluate.

    The activity and volume o 99mTc solu-

    tion introduced to each pharmaceutical

    vial during labelling.

    The generator is eluted

    every morning and used

    or reconstitution o the

    kits. Some departments

    may have a high genera-

    tor and a low generator.

    The low generator is

    useul or lung perusion

    scanning, Meckels diver-

    ticulum and thyroid scintigraphy.

  • 7/28/2019 Tech Radiopharmacy

    26/52

    26

    Remove the fip/

    oil cap rom the

    pharmaceutical

    vial and place

    the vial in a la-

    belled shielded

    container. Swab

    the rubber septum with alcohol, and al-

    low the alcohol to evaporate.

    When reconstituting the pharmaceu-

    tical use a shielded 5ml syringe

    capped with a green needle.

    10ml and 2ml syringes may also

    be used, depending on the re-

    quired volume.

    The required activity o 99mTc solution mayneed to be topped up to the required vol-

    ume with saline.

    Patient injections are typically made

    up to a volume o 1ml. The re-

    quired activity o radio-labelled

    pharmaceutical may need to

    be topped up to 1ml with

    saline.

    All patient injections are to be labelled with

    an appropriate label, detailing the patient

    name, date and time o procedure, proce-

    dure type and activity to be administered.

    99mTc-Cardiolite

    Act ............MBq Date ...............Pt. Name

    Appendix 2: Calculation o activity

    Example calculation o activity to add to vial

    Example: What is the minimum activity o99m

    Tc eluate needed to be added to a vial o99mTc-

    Cardiolite at 8a.m., to obtain ve injections o

    740Mbq each 2 injections at 8a.m., 3 injec-

    tions at 11a.m.?

    The equation to calculate radioactive decay is:

    At= A

    oe-t

    1. Calculating

    is the decay constant which or any ra-

    dionuclide is dened as

    2. Where Ao = 740MBq, t= -3

    (When calculating an activity *prior* to the

    time o known activity, the value o t is nega-

    tive. To calculate the decay o a radionuclide

    *rom* a known activity to a later time then t

    has a positive value.)

    At= A

    oe-(0.115)(t)

    At= A

    oe-(0.115)(-3)

    At= 740 x e0.346

    At = 740 x 1.413 = 1045.6MBq

  • 7/28/2019 Tech Radiopharmacy

    27/52

    EA

    NM

    27

    Chapter 4 Radiopharmacy: Kits & Techniques

    Thereore the required activity to be drawn up

    per injection at 8a.m. is 1045.6MBq.

    The total activity o99mTc to be added to the

    vial o Cardiolite is:

    a. 2 x 740MBq or two injections at 8a.m.

    =1480MBq

    b. 3 x 81045.6MBq or three injections at11a.m. = 3136.9MBq

    c. Add (a) to (b) to get total o 4616.9MBq to

    add to vial o HDP

    Quality control in the radiopharmacy

    Rationale or perorming quality control

    I a radiopharmaceutical is improperly pre-

    pared it can result in a non-diagnostic study.

    The procedure must then be repeated result-

    ing in increased radiation dose and patient

    discomort. Many dierent classications o

    impurities aecting the imaging process canbe distinguished, as outlined in Table 2. Meth-

    odologies or detecting these impurities are

    briefy outlined in Table 3.

    Table 2: Classication o impurities

    Type o Impurity Example Efect

    Radionuclidic 99Mo Increased radiation dose; poor image quality

    Radiochemical Free 99mTc-pertechnetate Poor image quality; increased radiation

    Chemical Al3+ Poor image quality due to labelling problems

    Table 3: Detection o impurities

    Method Type o Impurity

    Dose calibrator/multichannel analyzer Radionuclidic

    Thin layer chromatography Radiochemical

    Colorimetric Chemical

    QC test procedures or Mo/Tc generator

    1. Mo-99 breakthrough:

    Mo-99 is assayed directly in the special lead pig supplied by the manuacturer o the dose calibra-

    tor. 99mTc is then assayed directly in the plastic sleeve in the dose calibrator. The activity o99Mo

    must not exceed 0.1% o the total 99mTc-activity. Dose calibrators have dierent methods how

    to determine whether this amount is exceeded dependent on manuacturer and age.

  • 7/28/2019 Tech Radiopharmacy

    28/52

    28

    The timetable o required testing is shown in Table 4, and or optional testing in Table 5.

    Table 4: Required QC testing o a 99Mo/99mTc generator

    Test Frequency Specications

    Mo breakthrough Every elution

  • 7/28/2019 Tech Radiopharmacy

    29/52

    EA

    NM

    29

    Chapter 4 Radiopharmacy: Kits & Techniques

    Radiochemical purity, paper and thin layer

    chromatography

    The radiochemical purity (RCP) o a radio-

    pharmaceutical is the ratio o the radionuclide

    present in a bound orm (i.e. as a bound radio-

    pharmaceutical) to the radionuclide present in

    its unbound orm (i.e. reeradionuclide).

    It can happen that, during the binding o radio-nuclide to pharmaceutical, the complete re-

    duction o pertechnetate does not occur and

    some ree or unbound technetium is present

    in the resulting solution. This can also happen

    i the reduced radionuclide becomes oxidised

    again. A result can be poor imaging o the

    radiopharmaceutical and increased radiation

    dose or the patient. To ensure radiochemical

    purity, it is important to perorm regular qual-ity control on the radiopharmaceuticals.

    General principles o planar chromatography

    or radiochemical purity testing

    The test strip or chromoplate may be obtained

    as a commercial product, and dierent types

    are produced or dierent pharmaceuticals.

    Generally a ew microlitres o the product to

    be tested is applied near the bottom (origin)

    o the chromoplate. The chromoplate is then

    placed in a solvent, ensuring that the bottom

    (origin) point is not immersed. The solvent is

    then allowed to rise or migrate up along the

    plate. The dierent chemical species separate;

    the most soluble product moving the urthest

    distance along the chromoplate. (Figure 7)

    Figure 7: A chromatography strip with the

    position o the origin and solvent ront

    shown. The sample is placed at the point

    marked on the origin.

    When the solvent has moved the required

    distance along the chromoplate (the solvent

    ront), then it is removed rom the solvent and

    allowed to dry. As a next step, this is read by

    scanning the radiochromatogram, either by

    passing it under a scintillation detector or on

    the surace o a gamma camera. The distance

    travelled along the chromoplate by each o

    the radiochemical species is expressed as a

    raction o the distance travelled by the sol-

    vent ront (Rvalue). From these R

    values can

    be determined the quantity o activity in each

    portion and thus the respective binding qual-

    ity o the radiopharmaceutical.

  • 7/28/2019 Tech Radiopharmacy

    30/52

    30

    Radiochemical impurities in 99mTc

    radiopharmaceuticals

    Common impurities in 99mTc-kits are 99mTc-

    pertechnetate and 99mTc-colloid. However a

    number o 99mTc-preparations may contain

    dierent impurities that have to be covered

    by respective validated tests (e.g. 99mTc-Tartrate

    in 99mTc-MAG3). A number o recommended

    chromatography methods are shown belowin Table 6 and Table 7.

    Separation o radiopharmaceutical and

    impurity

    Radiochemical purity testing can be based

    on thin layer or paper chromatography (TLC),

    solid phase extraction methods based on

    cartridges (SPE), ltration methods, HPLC or

    electrophoresis. Methods may derive romthe European Pharmacopeia, the SPC o the

    labelling kit or validated literature methods. In

    many cases a variety o methods are available

    or one particular radiopharmaceutical. TLC

    methods and SPE are most commonly used

    and recommendations in this guidance are

    based on these methods.

    Equipment and location:

    The determination o the radiochemical purity

    in the hospital with TLC can be perormed with

    little expenditure o material and equipment.

    It should be perormed in a dedicated area

    with radiation protection and proper ventila-

    tion (organic solvents).

    Quantication:A variety o methods may be used to quantiy

    impurities including cut and count in the dose

    calibrator, using dedicated scanner, gamma

    cameras and others. The dierences are in

    sensitivity, resolution, linear range, speed,

    availability and practicability and should

    be chosen dependent on available space,

    throughput and general organisation o the

    radiopharmacy. Equipment used or qual-ity control should be regularly calibrated.

    Generally, the quantication o the radio-

    chemical purity is based on the equation

    below.

    In case the radiochemical purity limit is not

    reached, the preparation has to be discarded

    and clearly labelled that it may not be used

    or patient application.

  • 7/28/2019 Tech Radiopharmacy

    31/52

    EA

    NM

    31

    Chapter 4 Radiopharmacy: Kits & Techniques

    Table 6: Recommended methods or quality control based on TLC

    Solid Phase / Mobile PhaseR Radio-

    pharmaceuticalR Impurity

    Pertechnetat ITLC-SG/0.9% NaCl Front Start

    99mTc-DMSA ITLC-SG/ 2- Butanone Start Front

    99mTc-Diphosphonates

    MDP, DPD, HEDP

    A) ITLC-SG/ 1M NaAcetate

    B) ITLC-SG/ 2-Butanone

    Front

    Start

    Start

    Front

    99mTc-DTPAA) ITLC-SG / NaCl

    B) ITLC-SG / 2-Butanone

    Front

    Start

    Start

    Front

    99mTc-ECD Ethylacetate / Baker Silica gel Front Start

    99mTc-HMPAOA) ITLC-SG/ 2-Butanone

    B) ITLC-SG/ 0.9% NaCl

    Front

    Start

    Start

    Front

    99mTc-IDA-Derivates

    A) saturated saline solution /

    ITLC-SG

    B) 50% Acetonitril / ITLC-SG

    Start

    Front

    Front

    Start

    99mTc-Colloids ITLC-SG / 2-Butanone Start Front

    99mTc-MAA ITLC-SG / 2-Butanone Start Front

    99mTc-ECA) ITLC-SG/2-Butanone

    B) ITLC-SG/ 0.3M Acetic acid

    Start

    Front

    Front

    Start

    99mTc-MIBI Ethanol /Baker Aluminiumoxide Front Start

    99mTc-Depreotide

    A) ITLC-SG / saturated NaClB) / ITLC-SG

    Methanol/1Ammonacetate

    (50/50)

    Start

    Front

    Front

    Start

    99mTc-TetroosminITLC-SG / Aceton :

    Dichloromethane =35:65 /Middle Start/Front

    99mTc-monoclonal Anti-

    bodies; HIG, ZevalinITLC-SG/ 0.9% NaCl Start Front

    111

    In Octreotide 0.1M Na-Citrat pH5 / ITLC-SG Start Front

  • 7/28/2019 Tech Radiopharmacy

    32/52

    32

    Table 7: Recommended methods or quality control based on SPE

    Cartridge / SolventRadiopharma-

    ceutocalImpurity

    99mTc-MAG3SEPPAK C18 light1.) 0.001N HCl,2.) Ethanol/0,9% NaCl (50% 50%)

    Eluate 2 column/ eluate 1

  • 7/28/2019 Tech Radiopharmacy

    33/52

    EA

    NM

    33

    Several blood cellular elements can be ra-

    diolabelled with dierent radionuclides and

    radiolabelling approaches or varies clinical

    applications (Table 1). Regardless o the na-

    ture o the blood cells, o the radionuclide

    used or o the clinical application, it is neces-

    sary to maintain both cell viability and steril-

    ity and to avoid the operators exposure to

    biological and radiation hazard during cellmanipulation and radiolabelling. Manipula-

    tions and radiolabelling procedures o cells

    require strict aseptic conditions and should

    ollow the EANM guidelines on current Good

    Radiopharmacy Practice (cGRPP) in the Prepa-

    ration o Radiopharmaceuticals [1]. The use o

    an open or closed vial system on a laboratory

    workbench is not an alternative to a controlled

    sterile environment. Cross contamination ormix-up o blood should be prevented at all

    times. Preparation o radiolabelled cells must

    be perormed successively or by dierent peo-

    ple in dierent locations. All suraces should

    be properly cleaned, decontaminated and dis-

    inected prior to use, ater all procedures are

    completed, and whenever suraces are overtly

    contaminated.

    Chapter 5 RadiopharmacyBlood Labelling

    Tanja Gmeiner Stopar, PhD

    During radiolabelling, approved written proce-

    dures should be ollowed at all times. All mate-

    rials used should be identied and certied or

    human use. Whenever possible radiochemical

    purity should be checked and radiolabelling

    eciency (percentage o radiolabelling o

    cells) calculated. Beore release, radiolabelled

    blood should be checked or aggregation o

    clumping o cells and or presence o particu-late contamination. At regular intervals, the

    integrity o the cells should be ascertained

    by use o suitable procedures i.e. using trypan

    blue. Prior to administration, control o patient

    identity should be perormed [1-3].

    Red blood cells

    Red blood cells (RBCs) are the most common

    type o blood cell. Their main unction is de-livering oxygen rom lungs to body tissues.

    Erythrocytes are continuously being produced

    in the red bone marrow o large bones. They

    develop rom committed stem cells through

    reticulocytes to mature erythrocytes and

    live a total o about 120 days. The aging o

    erythrocyte undergoes changes in its plasma

    membrane making it susceptible to recogni-

    tion by phagocytes and subsequently they

    are destroyed in the spleen, liver and bone

    marrow [4].

  • 7/28/2019 Tech Radiopharmacy

    34/52

    34

    Table 1: Clinical applications o radiolabelled cells and radiolabelling approach

    Clinical application Blood cellular elementRadionuclide

    /radiolabel

    Radiolabelling

    approach

    Cardiac and vascular

    imagingRed cells 99mTc-pertechnetate

    In vivo

    In vivo/in vitro

    Gastrointestinal

    bleedingRed cells 99mTc-pertechnetate

    In vivo

    In vivo/in vitro

    Spleen imaging Denaturated red cells 99mTc-pertechnetateIn vitro

    In vivo/in vitro

    Blood volume and red

    cell volumeRed cells

    99mTc-pertechnetate51Cr-chromate

    In vitro

    Red cell survival Red cells 51Cr-chromate In vitro

    Site o red blood cell

    destruction Red cells51Cr-chromate In vitro

    Inection and

    infammationWhite blood cells

    111In-oxine111In-tropolone99mTc-HMPAO

    In vitro

    Abnormal platelet

    depositionPlatelets

    111In-oxine111In-tropolone99m

    Tc-HMPAO

    In vitro

  • 7/28/2019 Tech Radiopharmacy

    35/52

    EA

    NM

    35

    Chapter 5 Radiopharmacy: Blood Labelling

    Radiolabelling with 99mTc

    Radiolabelling approaches may by in vivo, in

    vitro and combined in vivo/in vitro. The basic

    mechanism o the radiolabelling o RBCs with99mTc is in all approaches the same. The RBCs

    are pre-tinned by stannous ions or 10-30

    min beore 99mTc-pertechnetate is added to

    the cells. The stannous ions diuse into the

    cell and are rmly bound to cellular compo-nents. The pertechnetate diuses reely across

    the cell membrane and becomes reduced in

    the presence o stannous ions in the cell and

    subsequently binds to the beta chain o hae-

    moglobin. At physiological pH, stannous ions

    are likely to hydrolyse and precipitate and are

    rapidly removed rom the blood by the reticu-

    loendothelial system. To prevent hydrolysis

    and precipitation, stannous ions are thus usu-ally in a complex o a weak chelator, such as

    pyrophosphate or medronate. The amount o

    stannous ions required or optimal radiolabel-

    ling is in the range o 10-20g per kilogram o

    body weight.

    In vitro radiolabelling

    In vitro radiolabelling o RBCs gives by ar the

    highest labelling eciency and the most sta-

    ble labelling over time. It may be used or the

    determination o red cell and blood volume.

    It may also be employed in patients who are

    taking drugs which may interere or inhibit

    stannous ion transport through the cell mem-

    brane such as heparin or hydralazine, resulting

    in lower labelling eciency.

    A small volume o anticoagulated blood (hep-

    arin or ACD) is incubated with an aliquot o a

    reconstituted stannous agent. Any excess o

    Sn2+ is oxidised by addition o 0.1% sodium

    hypochlorite and may be removed by cen-

    triugation. The cells are separated and incu-

    bated with 99mTc-pertechnetate or 5-20 min-

    utes with occasional mixing. Ater incubation,

    unbound activity is washed away by additiono a ew mls o saline and centriugation. The

    cells are separated and re-suspended in saline

    beore re-injection.

    In vivo radiolabelling

    This is the simplest and least time consum-

    ing radiolabelling technique. An injection o

    a reconstituted solution o stannous agent is

    ollowed by injection o 99mTc-pertechnetate20-30 min later. The major disadvantage o

    this radiolabelling approach is a generally

    lower and more variable labelling eciency.

    This may be due to insucient Sn2+ incorpora-

    tion into the cells which results in reduction o99mTc-pertechnetate outside the RBC. Reduced99mTc is then not able to diuse across the red

    cell membrane, resulting in a high background

    activity. Low labelling yields may also be a con-

    sequence o low haemoglobin concentration

    and/or low haematocrit.

    In vivo/in vitro radiolabelling

    More variations o the in vivo/in vitro radiola-

    belling approach are in use. In all approaches,

    the intravenous administration o a stannous

    agent is ollowed by withdrawal o an aliquot

  • 7/28/2019 Tech Radiopharmacy

    36/52

    36

    o pre-tinned blood 15-30 min ater applica-

    tion. The excess o Sn2+ not incorporated into

    cells may be removed by centriugation beore99mTc-pertechnetate is added to the cells. Al-

    ternatively blood may be taken into a shielded

    syringe containing an anticoagulant and the

    required amount o99mTc-pertechnetate.The

    blood is then mixed with the 99mTc-pertech-

    netate and allowed to incubate or 5-20 min atroom temperature with occasional mixing. The

    unbound activity is washed away by centriu-

    gation beore reinjection. Radiolabelled blood

    may alternatively also be re-injected without

    removal o unbound activity. With the later ap-

    proach in which no washing step is involved

    one can expect lower radiolabelling eciency

    and higher background activity, depending

    on the complexity (number o washing stepsavoided) o the procedure.

    Heat-damaged RBCs

    When radiolabelled RBCs are damaged by

    heat, they will be recognised by phagocytes

    and subsequently destroyed in the spleen,

    liver and bone marrow. This mechanism en-

    ables the use o heat-damaged 99mTc-RBCs or

    spleen imaging studies. Radiolabelled RBCs

    are damaged by incubation in water bath at

    49.5 C or 15 min beore reinjection. To su-

    ciently denature RBCs but prevent bursting

    them, the temperature and incubation time

    are critical. Localisation in liver indicates pres-

    ence o cell destruction ormatting bursting

    ragments.

    Radiolabelling o RBCs can be aected by pa-

    tient medication. Drugs may interere directly

    by reaction with Sn2+ by preventing accumula-

    tion o Sn2+ in the cells or indirectly by aect-

    ing the RBC membrane or reducing the hae-

    matocrit and/or haemoglobin concentration.

    For more detailed inormation see [2,5-7].

    The usual administered activity o99m

    Tc-RBCsor adult patients is within the range o 500

    1050 MBq. For children the activity may be

    adjusted according to body weight, with a

    minimum activity o 80 MBq in order to ob-

    tain images o sucient quality [8]. Only when

    in vivo radiolabelling approach is employed,

    breast eeding should be interrupted and the

    expressed eeds discarded due to the pres-

    ence o ree 99mTc-pertechnetate which con-centrates in the mammary gland. The total

    ractional activity ingested by the baby rom in

    vivo radiolabelled RBCs is estimated to be 3 5

    times higher than the activity in milk rom in

    vitro radiolabelling.Ater in vitro radiolabelling

    breast eeding can be continued [9].

    General recommendations or application o

    radioactive drugs are applicable or adminis-

    tration o radiolabelled RBCs. Use o a tefon

    catheter or cannula should be avoided or ad-

    ministration o Sn2+ compounds because Sn2+

    can react with the catheter [10]. To prevent

    reaction o 99mTc-pertechnetate with traces

    o Sn2+ the stannous agent in the cannula

    and 99mTc-pertechnetate should not be given

    through the same system.

  • 7/28/2019 Tech Radiopharmacy

    37/52

    EA

    NM

    37

    Chapter 5 Radiopharmacy: Blood Labelling

    Radiolabelling with 51Cr

    Radiolabelling o RBCs with 51Cr is carried out

    in vitro: 51Cr in the orm o sodium chromate

    is incubated with whole blood containing

    ACD or approximately 10 -15 min. Chromate

    ion reely diuses into the RBCs, where it is

    reduced to chromic ion (Cr3+). Cr3+ bound

    to beta globin chain o the haemoglobin

    molecule is retained in the cell. The labellingprocess is stopped by adding ascorbic acid

    which reduces the chromate outside the cells

    to chromic ion. Alternatively, ree chromate is

    washed away by centriugation.

    Non imaging studies

    The basic principle in all non-imaging studies

    using radiolabelled RBCs is the same: a known

    quantity o radiolabelled RBCs is added to anunknown volume o blood in the body. Radio-

    labelled blood is allowed to mix, and a known

    quantity o the mixture is then removed and

    quantitated (counted in a gamma counter).

    The ratio o the quantity measured ater mix-

    ing (number o counts per mass) to the quan-

    tity added is the base or calculation o the

    unknown volume [2,11,12].

    RBC mass and volume determination

    For RBC mass and volume determination, one

    part o the radiolabelled blood is injected to

    the patient and the other part is used as a stan-

    dard or urther calculations. When complete

    mixing o re-injected radiolabelled blood has

    taken place (ater 30 min or longer i the pa-

    tient has splenomegaly), blood samples are

    taken or counting in a gamma counter. The

    red cell mass and plasma volume are calcu-

    lated by using the dilution equations. The true

    blood volume is then obtained by summing

    up the red cell mass and the plasma volume.

    The total body haematocrit can be calculated

    by dividing the red cell mass by the true blood

    volume.

    RBC survival

    For RBC survival and sequestration studies,51Cr-RBCs are reinjected to the patient and

    blood samples are obtained 3 times a week

    or 2 to 4 weeks ater injection. To eliminate

    the need or decay correction, the samples are

    counted at the end o the study. For the 51Cr

    survival curve, activity o the blood samplesover time can be plotted on a semilog pa-

    per to linearise the curve. The patients RBC

    eective hal-lie is calculated rom the curve,

    the normal being 25 to 35 days. In case o a

    signicant splenic sequestration o RBCs, a sig-

    nicant and progressive increase in splenic

    uptake relative to the other expected site o

    RBCs, like blood pool or liver, may be present.51Cr-RBC uptake is monitored using thyroid

    uptake probe positioning over spleen, liver

    and heart.

    The usual administered activity o51Cr-RBC or

    RBC mass and volume determination is 0.8

    1.5 MBq and 2.0 - 4.0 MBq or survival and

    sequestration studies.

  • 7/28/2019 Tech Radiopharmacy

    38/52

    38

    Leucocytes and platelets

    White blood cells (WBCs) or leukocytes are

    cells o the immune system deending the

    body against both inectious disease and or-

    eign materials. Several dierent and diverse

    types o WBC exist. They are all produced and

    derived rom a multipotent hematopoietic

    stem cell in the bone marrow. WBCs are ound

    throughout the body, including the blood andlymphatic systems.One primary technique

    to classiy WBCs is to look or the presence

    o granules, which allows the dierentiation

    o cells into two categories: granulocytes

    (neutrophils, basophils and eosinophils) and

    agranulocytes (lymphocytes, monocytes and

    macrophages) [4].

    Platelets or thrombocytes are blood cells involvedin the cellular mechanisms o primary haemosta-

    sis leading to the ormation o blood clots [4].

    Radiolabelling o WBCs and platelets

    Non-selective lipophilic 111In and 99mTc complex-

    es, which label all cells indiscriminately are used

    or radiolabelling o WBCs and platelets. Gen-

    erally radiolabelling eciency is higher when111In complexes are used (80-90%) compared

    to 99mTc-HMPAO, where radiolabelling eciency

    is normally 50-80%. Availability, low radiation

    exposure and better imaging characteristics are

    major advantages when 99mTc-HMPAO is used

    or radiolabelling. The clinical results using 111In

    or 99mTc-HMPAO WBC are similar. However when99mTc is used or radiolabelling, scans can be

    completed sooner, because imaging takes placeat 1 hour and at between 4-6 hours [2,12].

    Neutral lipophilic complexes rapidly diuse

    across cell membranes. Once inside the cell,

    the complex either dissociates and allows 111In

    to bind to intracellular proteins (111In-oxine) or

    breaks down to a more hydrophilic complex

    which is unable to cross the cell membrane

    and is thus trapped in the cell (99mTc-HMPAO).

    In the case o radiolabelling with 111In-oxine,111

    In also binds to transerrin present in theplasma. Thereore the cells have to be washed

    thoroughly to remove plasma beore radiola-

    belling. In the case o radiolabelling with either111In-tropolone or 99mTc-HMPAO, radiolabel-

    ling can take place in the presence o small

    amounts o plasma.

    Because o the non-selective approach, the

    cells need to be separated prior to radiolabel-ling. The separation o WBCs and platelets rom

    RBC in the blood can be achieved by sedimen-

    tation o anticoagulated blood. 30-50 ml o

    blood is taken into a 60 ml syringe containing

    an anti-coagulant. The anti-coagulant o choice

    is acid-citrate-dextrose (ACD) in concentration

    1.5 parts o ACD to 8.5 parts o whole blood.

    Erythrocyte sedimentation may be accelerated

    by the use o sedimentation agents such as

    6% hydroxyethyl starch (Hespan). Alternatively

    6% dextran or methylcellulose may be used.

    Normally 3 ml o Hespan per 30 ml o whole

    blood is used. Hespan may be added to ACD

    in a syringe beore blood is taken. Sedimenta-

    tion time is generally 45-60 min and may be

    aected by dierent actors such as number

    o cells and certain disease states. Sedimenta-tion may be carried out in the syringe placed

  • 7/28/2019 Tech Radiopharmacy

    39/52

    EA

    NM

    39

    Chapter 5 Radiopharmacy: Blood Labelling

    upward in a suitable holder or in a universal

    tube. Ater sedimentation, the upper layer con-

    taining WBCs and platelets is transerred into a

    sterile universal tube. To separate WBCs rom

    platelets, the plasma is spinned at low speed

    e.g. 150 g. Platelet rich plasma is removed be-

    ore the radiolabelling agent is added to the

    pellet containing WBCs and small amounts o

    RBCs. The mixture is incubated at room tem-perature. Incub