TEAT-5050

download TEAT-5050

of 59

Transcript of TEAT-5050

  • 8/13/2019 TEAT-5050

    1/59

    Department of Electroscience

    Master of Science Thesis

    Niklas Aronsson & Daniel Askeroth

    April 2002

    A Comparative Study of

    Electromagnetic Dosimetric

    Simulations and Measurements

  • 8/13/2019 TEAT-5050

    2/59

    1

    Abstract

    In this master thesis a comparative evaluation between SEMCAD simulations and

    MapSAR measurements were made. SEMCAD is an electromagnetics simulation

    platform derived from theFinite Difference Time Domain method, FDTD. MapSAR

    is a measurements-tool for quick SAR assessment.

    Specific Absorption Rate, SAR, is a quantity expressing the amount of radiated power

    absorbed in a specific material. This is also the quantity in which health guidelines for

    Radio Frequency, RF, devices are set. The parameter used in this thesis was the

    maximum Peak SAR (i.e. not mass averaged SAR). SAR was the comparative

    quantity used since it is MapSARs primary readout and it is also a quantity that

    provides an easy comparison.

    A simplified cellular phone was used during this thesis, allowing a variety of test

    configurations and facilitating the import of the phone design into SEMCAD. The

    results of this thesis indicate that changes in SAR relative the reference phone-

    designs SAR was coherent between simulations and measurements (for most cases).

    However, most changes in SAR are within the level of uncertainty, hindering a more

    explicit analysis of the results. Another cause for concern is that the absolute readings

    from the two systems differ by a factor two. In this thesis, this is compensated for by

    looking at relative changes to understand the impact of change in phone design.

  • 8/13/2019 TEAT-5050

    3/59

    2

    Preface

    This master thesis was written during the winter 2001/2002 in Lund under the

    supervision of Daniel Sjberg, PhD, at the Department of Electroscience.

    We would also like to thank Ramadan Plicanic, MScEE. for assistance during theDASY3-measurements. Other people deserving our gratitude are Kenneth Hkansson,

    MScEE for assisting us during the tedious soldering jobs and to Andr da Silva

    Frazo, MScEE, Thomas Bolin, MScAP&EE, for their revision of our thesis. For

    assisting us in the design of test cases Zhinong Ying, Licentiate has earned our

    appreciation.

    Lund March 2002

    Niklas Aronsson & Daniel Askeroth

  • 8/13/2019 TEAT-5050

    4/59

    3

    ABSTRACT ............................................................................................................................ 1

    PREFACE...............................................................................................................................2

    1 INTRODUCTION.................................................................................................................. 6

    1.1 Background ................................................................................................... ....................................... 6

    1.2 SAR-Specific Absorption Rate ................................................................................... ....................... 6

    1.3 Depth of Penetration...................................................................................... ..................................... 7

    1.3 Antennas for Cellular Phones ............................................................................ ................................ 8

    1.3.1 The Patch Antenna ........................................................................... ............................................ 10

    1.4 Finite Difference Time Domain (FDTD), The Maxwell Equations.............................................10

    1.5 Yees Algorithm.............................................................................. ................................................... 12

    1.6 Numerical Stability ............................................................................... ............................................ 12

    1.7 Boundary Conditions .......................................................................................... .............................. 131.7.1 Absorbing Boundary Conditions ...................................................................................... ........... 13

    1.7.2 Conducting Boundary Condition...................................................................................... ........... 13

    1.7.3 Periodic Boundary Condition .......................................................................... ............................ 14

    1.8 Implementation ......................................................................................... ........................................ 14

    2. MEASUREMENTS AND SIMULATIONS........................................................................... 15

    2.1 The Box Phone............................................................................................. ...................................... 15

    2.2 Antenna Matching Procedure..................................................................................... ..................... 16

    2.3 Variations in the Box Phone Configuration ................................................................................... 172.3.1 Reference Layout ............................................................... .......................................................... 17

    2.3.2 Ground Positions................................................................................... ....................................... 17

    2.3.3 Resistors Layouts .................................................................................... ..................................... 17

    2.4 MapSAR Measurement Equipment.............................................................................. .................. 182.4.1 MapSAR Measurements ............................................................................................ .................. 18

    2.5 SEMCAD Introduction ........................................................................................ ............................ 19

    3 RESULTS .......................................................................................................................... 20

    3.1 70-m Copper Tape Layer.......................................................................... ..................................... 20

    3.2 4-m Aluminium Layer ........................................................................................... ......................... 22

    3.3 Display Opening .................................................................................... ............................................ 23

    3.4 Resistor Measurements ................................................................................. ................................... 24

    3.5 Reliability of the Measurement Procedure .................................................................. .................. 25

  • 8/13/2019 TEAT-5050

    5/59

    4

    4 DISCUSSION..................................................................................................................... 26

    5. REFERENCES..................................................................................................................28

    APPENDIX A: SOLIDS SPECIFICATIONS.......................................................................... 29

    A.1 Box Phone Specifications............................................................................ ..................................... 29

    A.2 Antenna and Carrier Specifications........................... .................................................................... 29

    A.3 Spherical Phantom ............................................................................................. .............................. 29

    A.4 Material Data............................................................................................... ..................................... 30

    APPENDIX B: A SEMCAD PRIMER .................................................................................... 31

    B.1 About...................................................................................................... ............................................ 31

    B.2 Model Design with SEMCAD .............................................................................................. ........... 31

    B.3 Sources and Sensors .................................................................................... ..................................... 32

    B.4 Grid Settings ......................................................................................... ............................................ 34

    B.5 Model Visualisation................................................................................... ....................................... 37

    B.6 Simulation Settings ............................................................................................... ............................ 37

    B.7 Starting a Simulation ........................................................................ ............................................... 40

    B.8 Post-Processing ............................................................................................ ..................................... 41

    APPENDIX C MAPSAR MANUAL........................................................................................ 44

    C.1 Introduction ................................................................................... ................................................... 44

    C.2 Description of the Graphical User Interface................................................................................. 44

    C.3 Step by Step Measurements Guide for MapSAR......................................................................... 47

    APPENDIX D: MEASUREMENTS RESULTS ...................................................................... 48

    D.1 Measurement Results .............................................................................................. ......................... 48

    D.2 Reference Designs: Near Field-Plots ............................................................................................. 49

    D.3 Ground Positions: Near Field-Plots ............................................................................................... 50

    D.4 Resistors: Near Field-Plots.................................................................................. ............................ 52

    D.5 Ground Positions with Display: Near Field Plots........ ................................................................. 53

    APPENDIX E: NUMERICAL RESULTS OF SEMCAD SIMULATIONS................................ 54

    E.1 Perfect Conducting Front Layer ......................................................................................... ........... 54

  • 8/13/2019 TEAT-5050

    6/59

    5

    E.2 Perfect Conducting and Non Perfect Layers Comparison..........................................................54

    E.3 Resistor Simulations without Display Opening ................................................................. ........... 54

    E.4 Resistor Simulations with Display Opening.................................................................................. 54

    E.5 No Front Layer ............................................................................................ ..................................... 54

    APPENDIX F: DASY3 MEASUREMENTS............................................................................ 55

    F.1 DASY3 Results ........................................................................................ .......................................... 55

    F.2 DASY3 SAR Distribution Plots......................................................................... .............................. 55

    APPENDIX G: ACRONYMS................................................................................................. 58

  • 8/13/2019 TEAT-5050

    7/59

    6

    1 Introduction

    Wire-less communication has become an important factor in everyday life for

    millions of people around the world. The widespread use of these radio frequency

    transmitters has made the safety concerns a hot topic and safety regulations have

    accordingly been set to limit the exposition. To help manufacturers comply with thesesafety standards it is vital to quickly assess the amount of radiation at any given stage

    of research and development without going through the hassle of certified testing.

    This thesis evaluates two different approaches, simulating antenna transmitters as well

    as using a simple small-scale workbench-measuring device.

    1.1 Background

    The cellular phone market has increased substantially during the last decade. Along

    with market growth the possible risks related to the use of cellular phones have

    become an issue. As a measure of exposure to electromagnetic radiation SAR,

    Specific Absorption Rate,has been used. SAR translates to the loss in transmitted

    energy that is absorbed in human tissues, usually brain tissue.

    To meet with consumer demands concerning safety, standardisation institutes such as

    IEEE,Institute of Electrical and Electronics Engineers, and ICNIRP,International

    Commission on Non-Ionizing Radiation Protection, have set guidelines for the

    maximum SAR to enable a comparison between the various cellular phone

    manufacturers and models. The SAR value has now become mandatory for the

    manufacturers to be included with the phone.

    All new models have to meet with guidelines and legal restrictions so all phone

    models have to be tested. To prevent a phone prototype from having to be partially

    redesigned after failing to meet with maximum SAR regulations the ability to

    simulate or assess the maximum SAR at an early stage has become essential.

    SAR-simulations of a realistic cellular phone interacting with a head phantom are

    very computation-intensive. Therefore simulation tools capable of meaningful

    simulations such as SEMCAD have not been available until recently. The simulation

    programs are derivatives of numeric methods such as MoM,Method of Moment, FEM

    Finite Element Method,and FDTD, theFinite Difference Time Domain. FDTD has

    become the most common method due to its ability to work with complex problems

    over a wide range of frequencies.

    1.2 SAR-Specific Absorption Rate

    SAR is a quantity that describes the amount of absorbed radiated effect for a specific

    material at a certain frequency. The quantity can be derived from either the

    temperature gain or from an electric field. The method used for measurements derives

    the radiated effect from the electric field since the difference in temperature is too

    small to measure in the cellular phone frequency band due to the low energies

    involved at these low frequencies (i.e. non-ionising radiation). SAR has become the

    most frequently used quantity involved when health issues are discussed. According

    to the ANSI/IEEE (American National Standard Institute/Institute of Electrical and

  • 8/13/2019 TEAT-5050

    8/59

    7

    Electronics Engineers) standard the maximum SAR averaged over 1 g should not

    exceed 1.6 W/kg and that the whole body mass averaged SAR should not exceed 0.08

    W/kg. ICNIRPs guidelines for SAR in the head average over a 10 g cube and should

    not exceed 2 W/kg.

    Definition:

    2absP E

    ==

    =

    tTcSAR (1.1)

    Where:

    t

    T

    - Changes in heat over time [K/s]

    c - Specific heat capacity [J(kg K)-1]

    - Conductivity [S/m]

    - Density [kg/m3]

    E - Electric field strength [V/m]

    Pabs - Absorbed power within the 1 or 10 g cube [W]

    1.3 Depth of Penetration

    Different media react differently to electromagnetic radiation. One material parameter

    that has been defined to classify these differences in various materials is the depth of

    penetrationorskin depth[m]. The penetration depth represents the distance atwhich the wave amplitude has been attenuated by a factor e-1 = 0.368, which translates

    to an exponential decay rate e-x/.

    Biological tissue is a very complex media due to that within each cell there are

    variations in electrical properties. However, the cell structures in biological tissue can

    be simplified by averaging its dielectric properties resulting in an approximate

    homogeneous lossy media with current conducting capabilities.

    The depth of penetration for a lossy material can be calculated from the relative

    permittivity r, conductivity and angular frequency as:

    ( )

    =

    +

    =

    r

    r

    04

    12

    2

    00

    arctan2

    1,

    )(sin

    1 (1.2)

    A good conductor is characterised by the limit , hence the penetration depth fora good conductor is:

    0

    1

    f= (1.3)

  • 8/13/2019 TEAT-5050

    9/59

    8

    Conductor (copper) Brain tissue (homogenous)

    Angular Frequency [rad/s] 2900106 2900106Permittivity 0[F/m] (36)-110-9 (36)-110-9Relative Permittivity r 1 42Permeability 0[H/m] 4107 4107Conductivity [S/m] 5.8107 1.1

    Attenuation Constant [m-1] 4.54105 31.0Depth of Penetration [m] 2.010-6 3.210-2Table 1.1 A comparison between conductor and a lossy material at 900 MHz

    As seen in the comparison in Table 1.1, the attenuation is very large in a good

    conductor, which makes it well suited to block undesired radiation. For more

    information on penetration depth in lossy media see [6] and [8].

    1.3 Antennas for Cellular Phones

    An antenna is a transmitting and/or receiving instrument for electromagneticcommunication. The antenna can be approached from two directions, either as a

    radiator in free space or as an electrical component in an electrical circuit.

    Every antenna has a specific radiation pattern describing its field properties. The

    radiation pattern is defined as a mathematical function or a graphical representation

    of the radiation properties of the antenna as a function of space coordinates. In most

    cases, the radiation pattern is determined in the far-field region and is represented as a

    function of the directional coordinates. Radiation properties include power flux

    density, radiation intensity, field strength, directivity, phase or polarization.[1, p.28]

    Figure 1.1 Spherical Coordinates

    Antenna performance parameters such as gain and directivity can be derived from the

    far-field radiation pattern. These two parameters are closely related where the relative

    gain parameter is defined as the ratio of the power gain in a given direction to the

    power gain of a reference antenna in its referenced direction.[1, p.58] The directivity

    is "the ratio of the radiation intensity in a given direction from the antenna to the

    radiation intensity averaged over all directions.[1, p.39]

  • 8/13/2019 TEAT-5050

    10/59

    9

    SWR1

    0,log20 1111

  • 8/13/2019 TEAT-5050

    11/59

    10

    1.3.1 The Patch Antenna

    Patch antennas in cellular phones are internal antennas. The patch antenna is a thin

    metal sheet etched on a dielectric carrier structure. The metal patch plane is parallel to

    the ground plane, see Figure 1.4. The metal patch can take on various shapes such as

    circular, rectangular or irregular.

    Another important parameter is the feed point that has to be considered since it playsa significant role to the characteristic impedance. The antenna feed structure might

    differ, but usually the patch is fed from a pogo pin (a spring-mounted conductor) or a

    micro strip.

    Figure 1.4.The Single-band Patch Antenna

    Simplified half-wave patch parameters from Figure 1.4 can be calculated, however

    since these parameters will be altered due to reflections from the cellular phone cover

    and to changes in load parameters no effort has been made to do so. The antenna feed

    placement will also alter the results, so to optimise the antenna performance the whole

    phone model has to be tested. For more information concerning antennas see [4] and

    [11].

    1.4 Finite Difference Time Domain (FDTD), The Maxwell Equations

    Electromagnetics is the study of electric charges or in this case electric fields, i.e.

    spatial distributions. There are three ways to predict electromagnetic fields and their

    effects on different materials; by experiment, analysis or computation. Most problems

    have the complexity to make a theoretical analysis impossible and experimentation

    can be prohibitively costly and time consuming. Several numerical methods are today

    available for electromagnetic computational analysis likeMethods of Moments

    (MoM),Finite Element Method(FEM) andFinite Difference Time Domain(FDTD).

    These applications have become more and more useful with the increasing

    accessibility of computer power that the last decades have brought. FDTD is wellsuited for calculating transients as well as single frequencies or continuous wave

    sources.

  • 8/13/2019 TEAT-5050

    12/59

    11

    The Maxwell equations describe the electromagnetic field behaviour completely as

    long as the data, such as the material properties and initial conditions, are known. In

    equation 1.14-17 the Maxwell equations are presented in their differential form.

    t

    =

    BE Faradays law of electromagnetic induction (1.14)

    t

    +=

    DJH Ampres circuital law (1.15)

    0= B (1.16)

    = D Gausss law (1.17)

    The four Maxwell equations are not independent since 1.16-17 can be derived from

    1.14-15 using the equation of continuity (t

    =

    J ). In order to solve equations

    1.14-15 the following two constitutive equations are needed, which describe thematerials behaviour.

    ED = Where 0 r= (1.18)

    HB = Where 0 r= (1.19)

    The symbols used above are defined and known as the following quantities.

    B is the magnetic flux density [Vs/m2or Tesla]

    Dis the electric displacement [As/ m2]

    Eis the electric field intensity [V/m]

    His the magnetic field Intensity [A/m]

    Jis the current density [A/m2]

    The permittivity 0[As/Vm] is for free space whereas ris the relative permittivity,and is dependent on the material in question, which can also be said for the

    permeability [Vs/Am]. For free space the permeability is 0and rhas materialdependencies. All relative constants are dimensionless by definition.

    If the E-field is discretised in space and time with a uniform step and only field values

    at these points are stored, this can be written as equation 1.20.

    ),,,(,,

    tnzkyjxiEEn

    kji = (1.20)

    Using the terminology below one can substitute derivatives with an approximatedifference. Equation 1.21 shows the finite difference formula for the first derivative.

    t

    EE

    tt

    EEE

    n

    kji

    n

    kji

    t

    kji

    t

    kji

    ttt

    =

    +

    2

    1

    ,,

    1

    ,,

    12

    ,,,,

    12

    012

    lim (1.21)

  • 8/13/2019 TEAT-5050

    13/59

    12

    1.5 Yees Algorithm

    Yee penned his algorithm in 1966 [9], since then extensive research have expandedand enhanced the original idea and applied it to a wide range of problems. The recentupswing in its popularity is undoubtedly thanks to the decreasing cost of computercalculations.

    The Yee algorithm uses a numerical solution to Maxwells curl equations using adiscretised differential form. The big novelty in Yees algorithm was his griddefinition that greatly decreased the amount of memory required. The E-field isdefined on one grid and the H-field on another in the time domain. Both grids areoffset in time and space. Each field is updated with a leapfrog scheme and a finitedifference form of the curl using the values of the surrounding cells. The electric fieldis solved at one moment in time and then the magnetic field is solved at the nextmoment (half a time step) similar to the game, leapfrog. This is repeated until thesimulation is done. For a general presentation of the FDTD method and Yeesalgorithm, see [5] and [7].

    Figure 1.5 The Yee Cell

    1.6 Numerical Stability

    To ensure a stable and converging solution the cell size and time step must be wiselychosen. Choosing a grid too fine the simulation will be penalised with an unnecessarylong computational time and in most cases not result in increased accuracy. It is clearthat the cell size should be shorter than the wavelength of interest. The cell sizesuggested by the Nyquist theorem (i.e. two samples per wavelength) will normally notsuffice due to the approximate nature of FDTD computation. A more realistic cell sizewould be somewhere between 1/10 and 1/20 of the wavelength. It is possible to use

    non-uniform cells with a finer cell size (grid) near complex areas, although these areasmay have to be interpolated both in time and/or space with the rest of the grid.

    For solids with penetrable materials the wavelength is shorter inside these for a givenfrequency thus requiring a smaller cell size. Cells also have to be small enough to givean accurate model of the actual geometry. Rounded objects will suffer from thestaircase effect when modelling them with rectangular cells, i.e. the problem ofbuilding circles and other non-rectangular objects with cubes.

  • 8/13/2019 TEAT-5050

    14/59

    13

    After the cell size has been set, next step is to determine a suitable time step t. TheCourant-Friedrich-Levy criterium sets the limit for a stable solution (Figure 1.26).

    222 )(

    1

    )(

    1

    )(

    1

    1

    zyxc

    t

    +

    +

    Courant-Friedrich-Levy criterium (1.26)

    Here x, y and z are the mesh size for the smallest cell to be found globally in thegrid for each axis. The parameter c is the speed of light within the cell material. The

    rule of thumb is that a propagating wave must not travel through more than one cell in

    one time step. Larger time steps than the one given with the above condition will

    quickly result in instability.

    1.7 Boundary Conditions

    Antenna models using FDTD are usually located in free space making a boundary

    condition necessary to make sure that the reflected fields do not bounce back into the

    computation space, since a infinite computational volume is impossible. Using an

    Outer Radiation Boundary Condition(ORBC) to absorb incoming fields at the outerlimits is the only solution, otherwise the calculations have to be aborted when the

    reflected wave reaches the outer computational limit. Also since the outer co-

    ordinates cannot be updated using the normal finite difference equations an ORBC is

    needed for simulations. Several types of boundary approximations are available to

    enclose the computational space. SEMCAD offers the three types of Boundary

    Conditions described in the following paragraphs [2].

    1.7.1 Absorbing Boundary Conditions

    Absorbing Boundary Conditions(ABC) simulates materials that absorbs outgoing

    waves or free space conditions. Ideally they should allow any outgoing wave to exit

    without any reflection. A popular technique that can be described as a material

    absorber is thePerfectly Matched Layertechnique (PML). This has been shown to be

    one of the most accurate algorithms. Other ABCs that SEMCAD has implemented

    are the first and second order Mur and the Higdon operator (up to the 4thorder) that

    offers better performance than PML but may be not as numerically exact.

    1.7.2 Conducting Boundary Condition

    The Conducting Boundary Condition(CBC) does simply what the name suggests.

    The computational space is truncated with a perfectly conducting plane that can beperfect electrically (PEC) or magnetically conducting (PMC). Incoming tangential E-

    or H-fields components are set to zero.

  • 8/13/2019 TEAT-5050

    15/59

    14

    1.7.3 Periodic Boundary Condition

    Periodic Boundary Condition can simulate periodic boundaries. Values from the other

    side of the grid are used for periodic computation.

    1.8 Implementation

    A FDTD simulation basically contains three steps. First the grid must be built and

    defined and the initial conditions set. The number of cells in each dimension must be

    determined as well as the size of each cell and the electrical properties of each cell.

    Other specifications necessary are where and what to be monitored and for how many

    time steps the computation shall proceed. After the time step has been set according to

    the smallest cell size using the Courant-Friedrich-Levy criterion the fields can be

    updated.

    After the initial specifications are set, the second part can commence. This is the heart

    of the simulation where the field is advanced step after step.

    When the criteria for terminating the simulation have been fulfilled the final part of a

    FDTD simulation can begin, the post processing. This is where the calculated data is

    processed and then presented in any manner desired by the user.

  • 8/13/2019 TEAT-5050

    16/59

    15

    2. Measurements and Simulations

    To make an evaluation of the MapSAR and the SEMCAD utilities possible a suitable

    comparison quantity had to be selected. For this thesis SAR was chosen since it is the

    best presented output from MapSAR. Different comparison scenarios were designed

    to alter the near field, aiming at making the results significant. This chapter isintended to clarify the methods and scenarios used.

    2.1 The Box Phone

    For this master thesis a simplified cellular phone was used, referred to as the Box

    phone from here on. The reason for using a simplified phone design is to minimise the

    number of cells needed during simulations in SEMCAD, thereby minimising the

    number of simplifications needed for the simulated model. Another concern for the

    simplified model was doubts about SEMCADs ability to simulate complex models,

    like a faithful representation of a real phone, and simply because it facilitates changes

    in the phone.

    The box phone is made of plastic materials, see Figure 2.1. For further details see

    Appendix A. It transmits with an internal dual band antenna (PIFA) fitted on a plastic

    carrier mounted on the antenna ground plane. Several fronts, each with different

    conductive layers, were tested with the phone. Different scenarios for interaction with

    these front layers will be discussed further on.

    Figure 2.1 The Box Phone

    The simulated Box phone differs geometrically from the actual one in two major

    aspects, it has no battery and no opening around the battery to allow an on/off switch.

    Other minor differences are that the simulated phone lacks assorted details such as

    screws and bolts to fit the antenna carrier onto the ground plane, the PCB.

  • 8/13/2019 TEAT-5050

    17/59

    16

    2.2 Antenna Matching Procedure

    An important step in the phone manufacturing process is tuning the antenna to the

    desired resonance frequency and making sure it covers the whole frequency band with

    an acceptable gain. As a visualisation of the antennas matching an S11-analysis was

    made, resulting in Figure 2.2. This S11-measurement was made with the Box phone

    taped to a head phantom. The Box phone antenna was designed using ie3d (a methodof momentsimulation tool) and then tuned, with copper foil and a scalpel until an

    acceptable resonance frequency at 926 MHz was measured. In SEMCAD an antenna

    with roughly the same dimensions as the corrected antenna displayed a simulated

    resonance at 922 MHz, see Figure 2.3.

    Figure 2.2 Measured Return Loss, S11 Figure 2.3 Simulated Return Loss, S11

    For frequencies outside the resonance dip the plot is not reliable, as seen in Figure 2.3.

    As the figure shows, strange behaviour is common for frequencies that have next to

    no power in the excitation signal as the sub 800 MHz area of the plot displays. In the

    area above 1 GHz the irregularities shown are mainly due to the fact that the grid used

    was designed for 900 MHz.

    The Box phone transmitter was set to 914.9 MHz and the simulations were done at

    916 MHz. The antenna can be seen in Figure 2.4 with the GSM component at the

    right and DCS to the left.

    Figure 2.4The Dual Band Patch Antenna

  • 8/13/2019 TEAT-5050

    18/59

    17

    2.3 Variations in the Box Phone Configuration

    To test SEMCADs ability to simulate complex models, different scenarios aiming to

    alter the electromagnetic characteristics of the Box phone were evaluated. These

    approaches involve alterations in the conductive shield and ground positions.

    2.3.1 Reference Layout

    Figure 2.5 shows the standard front layouts used. These will later be used as the basis

    of both measurements and simulations for the schemes discussed in the following

    paragraphs.

    a.) No conductive shield

    b.) 4 or 70-m thick conductive shieldc.) Display opening and a 2-mm gap

    d.) Display opening with a conductive

    display patch and a 2-mm gap

    Figure 2.5 Reference Layout

    2.3.2 Ground Positions

    All ground positions used are accounted for in Figure 2.6. Two different materials

    were used for the front layers, a 70-m thick copper-tape and a 4-m thick layer ofaluminium.

    Figure 2.6 Ground Positions

    2.3.3 Resistors Layouts

    An interesting feature for this thesis available in the SEMCAD simulation platform is

    electromagnetic models of electric circuits, such as resistors, capacitors and inductors.

    To test SEMCADs ability to handle this feature two rather basic schemes using a

    resistor at the centre of the phone over a 2-mm gap as well as two resistors parallel at

    each side over the same gap. These layouts can be seen in Figure 2.7.

  • 8/13/2019 TEAT-5050

    19/59

    18

    a.) Single resistor, either 5 or 10 .b.) Two resistors parallel, each 10 or 20 .c.) As b.) with display patch

    Figure 2.7 Resistor Configuration

    2.4 MapSAR Measurement Equipment

    SAR measurements were performed with MapSAR benchtop SAR assessment

    equipment, Figure 2.8, developed by Index SAR. It consists of a spherical phantom,

    an E-field probe with amplifier and a data converter. For the placement of the phone

    there is also positioning-rack, allowing movements in all directions, with a partial

    rotation, constructed in plastics to counteract reflections. All SAR readings measured

    with the MapSAR system in this thesis are not averaged over mass. Non-averaged

    readings from MapSAR are calculated from measured E-fields in the liquid.

    Figure 2.8 The MapSAR Equipment

    The probe movement is restricted to spherical coordinates with a fixed radius and

    origin of coordinates in the centre of the spherical phantom. A plastic arm is mounted

    on the sphere to render tracing of the probe movements through a drawing board. The

    drawing board has an underlay with a pattern suggesting the probe-movement. Usingcoordinates from the drawing board, the MapSAR software can determine the probe

    position. The software keeps track of the maximum SAR and plots a 2D graph of the

    SAR where each point is displayed on top of an imported image of the phone model.

    2.4.1 MapSAR Measurements

    Before starting a measurement the E-field probe is positioned at the start position;

    Figures 2.9-10 in the probes horizontally suspended position. The object that is to be

    measured is positioned in the phone holder touching the sphere.

    Figure 2.9 Figure 2.10

    Start-position from a lateral view Start-position from the top view

  • 8/13/2019 TEAT-5050

    20/59

    19

    The measurement procedure basically consists of six steps:

    i.) Placing the probe in the start position relative the phone (Figures. 2.9 & 2.10).

    ii.) Taking an initial SAR-reading at the start position for comparison with (v.).

    iii.) Scanning the phone surface using the mask design on the Wacom-board as a

    reference.

    iv.) Scanning extensively around the maximum SAR position(s) to determine the

    absolute maximum.v.) Comparing the final SAR-reading at the start position to the initial (ii.) to

    determine the drift i.e. changes in SAR-reading.

    vi.) Noting the maximum SAR-readout.

    The readings can be exported as raw data (text file) with coordinates, phone outline

    and conductivity for further analysis and the 2D plot produced in the main window

    can be saved as a windows bitmap file.

    To minimise errors the measurements for each phone configuration were made twenty

    times to avoid aberrations due to phone position and to minimise other possible

    sources of uncertainty. Another possible source of uncertainty could be differences in

    batteries; to be able to distinguish battery problems in the results two batteries withsimilar properties were selected and then replaced every ten measurements.

    For a more extensive guide to the MapSAR benchtop SAR assessment equipment see

    Appendix C.

    2.5 SEMCAD Introduction

    SEMCAD is an electromagnetics simulation platform using a FDTD solver. It is very

    well suited for simulating antenna designs, dosimetry, electromagnetic interference

    (EMI)/ electromagnetic compatibility (EMC) and optics involving complex objects.

    SEMCAD (the Simulation platform for EMC, Antenna Design and Dosimetry) isdeveloped by Schmid & Partner Engineering AG (SPEAG) in conjunction with

    several Swiss universities. It offers a graphical user interface for computer-aided

    design called ACIS made by Spatial coupled with SPEAGs FDTD implementation

    for a slick user-friendly interface. Please refer to Appendix B and the SEMCAD

    Reference Manual [2] for information on how to set up and run simulations.

  • 8/13/2019 TEAT-5050

    21/59

    20

    3 Results

    3.1 70-m Copper Tape Layer

    Copper tape (70-m thick) was put on the inside of the front piece. A short piece of

    conducting elastomer (conductive rubber) was used to ground the copper tape on thefront with the main board holding the carrier as well as the circuit board. The

    objective here is to see how placing a ground at various positions interacts with the

    copper layer. For maximum convenience the different ground positions is listed again

    in Figure 3.1.

    Figure 3.1 Ground Positions

    Ground positions C, D and E1+E2 are the only ones showing a significant SAR

    deviation compared to the case with floating copper layer according to the MapSAR

    measurements with respect to the margins of error. As Figure 3.2 shows, the highest

    SAR value comes with grounding at position C and the lowest values with grounding

    at E1+E2. SAR is decreased by almost 40 % comparing those two values.

    0.46

    0.46

    0.43

    0.47

    0.43

    0.53

    0.37

    0.42

    0.45

    0.33

    0.45

    0.46

    0.42

    0 0.1 0.2 0.3 0.4 0.5 0.6

    Floating

    A

    B1

    B2

    B1 + B2

    C

    D

    E1

    E2

    E1 + E2

    F1

    F2

    F1 + F2

    SAR (W/Kg)

    Figure 3.2 Measured SAR Values with 70-m Copper Layer for Various Ground

    Positions

  • 8/13/2019 TEAT-5050

    22/59

    21

    The SEMCAD simulations showed consistently less than half the value measured

    with the MapSAR set-up when using 93.2 mW as the phone input power in

    SEMCAD. The reason for this is hard to say, and if this is an error with SEMCAD or

    MapSAR is unknown. It could be a number of reasons like lack of MapSAR

    calibration, old fluid in the MapSAR bowl, the fact that the SEMCAD antenna has

    higher gain than the real life dito or something else. SEMCAD can show any given

    number by entering the right digits to use for the input power value if the absolutenumber is important. In this thesis normalised values are compared instead.

    Grid size and baselines have been more or less consistent for all SEMCAD

    simulations, but minor modifications have been made in a few cases. A few baselines

    have been manually added. No significant numerical difference (most likely less than

    1%) has been noticed after these grid alterations suggesting that the original model

    was sufficient.

    0

    0.2

    0.4

    0.6

    0.8

    1

    1.2

    1.4

    Ground Position

    MapSAR measurements SEMCAD simulation

    Figure 3.3 Normalised MapSAR and SEMCAD Comparison for 70-m Copper Layer

    Figure 3.3 shows a comparison with results obtained using MapSAR and SEMCAD.

    Both series have been normalised using the SAR value for the floating front layer

    from their respective run of numbers. As one can see both simulations and

    measurements are fairly consistent. The major differences are with grounds at B1+B2

    and E1+E2 where the SEMCAD values are considerably higher. Since the MapSAR

    probe has a fixed position 5 mm behind the glass bowl whereas with SEMCAD onecan get the SAR value anywhere inside the bowl and the maximum value is usually

    positioned at some depth inside the bowl it can be debated if it really is the same thing

    that is being measured. Using the probe to measure an object with more than one SAR

    hot spot can be difficult since it can only sample in one direction at a time. Several

    maximums (like using two ground points) could very well interfere and create a

    maximum at some depth inside the bowl, which could be impossible for the probe to

    pick up on. The SEMCAD results in Appendix E shows that for the majority of the

    simulations the maximum SAR values were at some depth inside the sphere.

  • 8/13/2019 TEAT-5050

    23/59

    22

    The MapSAR measurements readings are overall lower with a larger difference

    relative to the worst-case scenario with ground at position C. This may be due to the

    fact that locating the hottest spot with the probe can be difficult. Getting an exact

    reading of that actual value is almost impossible. Unless the phone is positioned

    wrong or the phone is put out of position during the measurement, finding a position

    with a higher SAR-reading than the actual absolute value is impossible.

    3.2 4-m Aluminium Layer

    The same measurements that were done with the copper tape inside the front were

    repeated with a 4-m thick aluminium layer. As Figure 3.4 shows, the ground

    positions give the same relative results as they did with the copper front. Here the

    ground positions that show the largest decrease are E1+E2 and D. Using the Al-front

    slightly lowers SAR for all ground positions, the average decrease is 0.0279 W/Kg

    calculated over all ground positions. Measurements varied with a decrease between

    3% and 20% with an average decrease of approximately 12%. It should be noted that

    the effect of alternating ground positions is less than with the copper front.

    0

    0.2

    0.4

    0.6

    0.8

    1

    1.2

    1.4

    Ground position

    Copper AluminumAluminium

    Figure 3.4 A Normalised Comparison of Measurements for Al and Cu

    All simulations have used an approximation of the copper layer asPerfect Electric

    Conductor(PEC), which is treated as a boundary condition in SEMCAD. Using the

    actual values for copper as well as aluminium during simulations showed no hint of

    the SAR decrease that have been measured with MapSAR. The actual SAR difference

    (if any) with an aluminium front may be lower than these measurements have shown.

    And the average 12% reduction is well within the margin of error.

  • 8/13/2019 TEAT-5050

    24/59

    23

    As a comparison between copper and aluminium fronts, measurements with no

    ground position for each case were made on a DASY3-system. For more information

    regarding the DASY3 system see [3]. Just as in SEMCAD and MapSAR the

    difference between copper and aluminium fronts was minimal and well within the

    margin of error. The DASY3 showed a 7% higher reading for aluminium than for

    copper. These measurements were performed for the left cheek position.

    3.3 Display Opening

    Measurements were done using the copper tape but with a rectangular opening where

    a display normally would be positioned. As earlier the effects of using a ground

    between the PCB and the copper front was investigated. There was some interesting

    resonance with ground positions D and E1+E2 as can be seen in Figure 3.5. Both

    SEMCAD and MapSAR readings shows the same behaviour but with a greater

    resonance shown with MapSAR. This may be due to the small differences in the

    position and size of the display opening between the actual phone and our SEMCAD

    model. For the non-resonant ground positions the results were very close.

    0

    1

    2

    3

    4

    5

    6

    Floating B1+B2 D E1+E2 F1+F2

    Ground Position

    MapSAR (Normalised with floating) SEMCAD (Normalised with floating)

    Figure 3.5 SAR Comparison with Display Opening

  • 8/13/2019 TEAT-5050

    25/59

    24

    3.4 Resistor Measurements

    For this measurement series the copper layer was separated by a 2 mm gap. The upper

    half of the copper layer still had the display opening. Resistors were soldered to the

    front to join the two parts in different configurations. All resistor configurations can

    be seen in Figure 3.6.

    Figure 3.6 Measurement Cases Using Resistors

    Case 1: A single 10 resistor.Case 2: Two 20 resistors parallel at each side.

    Case 3: A single 5 resistor in the centre.Case 4: Two 10 resistors parallel at each side.Case 5: As Case 2 with floating display patch.

    Case 6: No resistors.

    As Figure 3.7 shows the results were somewhat similar. Simulations showed no major

    numerical difference for the different cases and the measurements suggest the same

    tendency with one exception. Measured values for Case 1 is somewhat higher than

    simulated.

    1,1

    1

    0,8

    7

    1,0

    2

    0,8

    9

    0,8

    9 1,00

    0,9

    1

    0,9

    00,9

    5

    0,9

    00,9

    3 1,00

    0,0000

    0,2000

    0,4000

    0,6000

    0,8000

    1,0000

    1,2000

    Case 1 Case 2 Case 3 Case 4 Case 5 Case 6

    NormalisedSA

    Measured Simulated

    Figure 3.7 Simulation and Measurement Comparison with Resistors

  • 8/13/2019 TEAT-5050

    26/59

    25

    3.5 Reliability of the Measurement Procedure

    To examine the reliability of the MapSAR measurements a statistical survey of the

    results were made. The reliability of the SAR-measurements had a maximum

    deviation of between 4% and 10.5% for 20 measurements including one battery

    change on the same phone configuration. The standard deviation ranged between 2%

    and 6%, see Appendix D.

    Another factor of uncertainty is the antenna mismatch, which might result in up to

    15% bias. The mismatch factor of the antenna was derived via a comparison between

    the Gain parameter for SWR 2 and 3. A set of light emitting diodes show how well

    matched the phone is at any given moment, all measurements showed a SWR between

    2 and 3. To determine the gain more precisely than this is not possible without tests

    and it is this uncertainty that the before mentioned 15% comes from.

    The antenna mismatch is the result of changes in the impedance due to modifications

    of the phone. The worst-case scenario adds up to a 25.5% deviation between two

    single measurements.

  • 8/13/2019 TEAT-5050

    27/59

    26

    4 Discussion

    One thing that became clear during this thesis is that some tools are better suited than

    others for a given task. First of all the MapSAR system has a very small measurement

    area due to the spherical shape of the phantom. The spot on the phone with the

    maximum SAR should be positioned next to the probes resting position. Accreditedtesting uses a head and shoulder SAM phantom, approximating a head phantom with

    a sphere instead takes some consideration. For measurement positions with a tilted

    phone, some phone shapes gives two touch positions (i.e. where the phone is touching

    the SAM phantom). With a ground position that moves the peak SAR to the lower

    part of the phone, a MapSAR reading measuring the top of the phone will show a

    reduced SAR readout. A DASY3 reading where the lower part of the phone is right

    next to the cheek as well as the ear will not display the same value as the MapSAR

    where the probe is at a greater distance from the more transmitting part of the phone.

    The DASY3 probe will get much closer to the lower part of the phone as well and

    pick up a normal value. The MapSAR system should thus only be used for

    measurement around the maximum value, i.e. the area with the hotspot should be

    positioned next to the probe to make up for the differences between a spherical bowl

    and a head SAM where a larger area of the phone will be close to the SAM. This can

    be seen as worst-case scenario measurements, finding the maximum SAR anywhere

    on the phone regardless of its position and if it can be measured with a SAM

    phantom.

    SEMCAD simulations are ideal for visualising hotspots. Using the simulation post-

    processing is a good basis for MapSAR measurements since the user can easily

    determine the locations likely to show a high SAR in reality as well. However using

    MapSAR to find the location of maximum SAR on a phone surface can be done fairly

    quickly and to later use that position for a more thorough measurement of the absolute

    value as reference. The strength of the MapSAR is the short measurement time, whichmakes it ideal for comparative tests. This is one of the biggest selling arguments of

    the MapSAR system. Another benefit it has over a complex robotic set-up is that the

    liquid is sealed which gives it a longer life. Also since each measurement takes 2-3

    minutes at the most it is less likely that the batteries will drain during a measurement

    or that the characteristics of the measured object will change due to overheating.

    SEMCAD is a very useful tool when it comes to getting an idea of how complex

    structures such as models of cellular phones will interact with surrounding structures.

    Seemingly identical antenna geometrical dimensions in SEMCAD as the real world

    original design did result in slightly different antenna characteristics. Maybe the

    SEMCAD model requires a more detailed representation (like a cad file of the actual

    phone). The real antenna was cut out by hand and messed up with the soldering iron,which makes it impossible to model down to a certain degree of accuracy. Another

    difference is to which extent the electric properties given to solids in the SEMCAD

    simulations are accurate.

    While the differences in antenna performance may not be significant it may lead to

    somewhat different absolute numerical results but should be of no concern for relative

    comparisons between different simulation scenarios. Using SEMCAD as a serious

  • 8/13/2019 TEAT-5050

    28/59

    27

    development tool requires a fast computer, the faster the more realistic and rewarding

    simulations are possible.

    MapSAR usually reported a SAR value twice as high as the SEMCAD simulations.

    This may be due to the differences in antenna gain or perhaps the fact that the

    measured power from the phone had been measured without the mismatch that were

    present during all measurements. Due to the fact that MapSAR is designed foraveraged measurements (the probes position was set with this in mind) it may not be

    ideal to compare peak SAR. The MapSAR set up requires a corrective factor for

    comparisons with DASY3 measurements due to its 4mm thick bowl whereas SAM

    phantoms often use a 2mm thick shell. An uncorrected MapSAR value will thus be

    lower than a DASY3 reading. With careful calibration some MapSAR measurements

    could be fairly accurate, if the procedure is done carefully, and the differences

    between a spherical and a SAM phantom are kept in mind. This causes some practical

    issues regarding where on the test object measurements shall be done as discussed

    earlier. Using a complex head phantom in SEMCAD is very feasible. It is not a very

    big part of the head that interacts with the phone especially with the depth of

    penetration in mind. Figure 4.1 shows how the complex shape of a SAM phantom is

    approximated in a SEMCAD simulation. There is no real loss in accuracy with a morecrude head approximation, as shown in the picture to the right below, so the choice of

    tissue for SAR extraction should be made with respect to whatever measurement

    method one aims to simulate. With an efficient grid design using a head phantom

    must not imply more cells than a spherical phantom (Figure 4.2).

    Figure 4.1 Head Phantom Figure 4.2 Spherical Phantom

    Both SEMCAD simulation and MapSAR simulations are fine tools for relative studiesof antennas. As seen in chapter 3 the normalised results show the same tendencies for

    each test scenario with a few exceptions.

    The only really useful SAR measurement is done on accredited systems like the

    DASY3 but it should be possible with reference measurements from an accredited

    system to predict the actual value from results using SEMCAD and/or MapSAR.

    MapSAR with its spherical phantom makes it less versatile than SEMCAD and harder

    to compare with results using a head phantom.

  • 8/13/2019 TEAT-5050

    29/59

    28

    5. References

    [1.] Balanis, A. C., Antenna Theory: Analysis and Design, 2nd

    edition, John Wiley & Sons

    Inc. 1997[2.] SEMCAD Reference Manual. Bundled with SEMCAD and available at:

    http://www.semcad.com/downloads_free/SEMCAD_RefManual.pdf

    [3.] http://www.speag.com

    [4.] Fujimoto, K. and James, J. R., Mobile Antenna Systems Handbook, 2nd

    edition, Artech

    House Publishers, 2001[5.] Kunz, K. S., and Luebbers, R. J., The Finite Difference Time Domain Method for

    Electromagnetics, CRC Press 1999

    [6.] Cheng, D. K., Field and Wave Electromagnetics. 2nd

    edition, Addison-Wesley 1989[7.] A. Taflove. Computational electrodynamics: The Finite-Difference Time-Domain

    Method. Artech House, Boston, London, 1995.

    [8.] C. Gabriel, S. Gabriel and E. Corthout, The dielectric properties of biological tissues: 1-3,Phys. Med. Biol. 41, 1996. pp. 2232-2293.

    [9.] K.S Yee. Numerical Solution of initial boundary problems involving Maxwells equations

    in isotropic media. IEEE Trans. Antennas Propagat.,14. March 1966.[10.] Indexsar, Mapsar Sytem Manual. 2001

    [11.] Stutzman, W. L. and Thiele, G. A., Antenna Theory and Design, 2nd

    edition, John Wiley

    & Sons Inc. 1998[12.] SEMCAD Tutorial. Bundled with SEMCAD and available at:

    http://www.semcad.com/downloads_free/SEMCAD_Tutorial.pdf

  • 8/13/2019 TEAT-5050

    30/59

    29

    Appendix A: Solids Specifications

    A.1 Box Phone Specifications

    The Box phone has the following outer dimensions: 1005020 mm, see Figure A.1.All walls of the outer casing are 2 mm thick. It is a simplified model of a cellular

    phone with straight edges. Material characteristics used for SEMCAD simulations are

    listed in Table A.1.

    FigureA.1 Cross Section of the Outer Casing

    A.2 Antenna and Carrier Specifications

    A dual-band (900 and 1800 MHz) Micro strip Patch Antenna was designed for the

    box phone. This was fitted on a carrier made of Noryl. The carrier with mounted

    antenna was screwed on to the copper ground plane (the rectangular area in Figure

    A.2). Table A.1 lists the material properties.

    FigureA.2 Carrier and Antenna

    A.3 Spherical Phantom

    A spherical phantom (see Figure A.3) with the outer diameter of 200 mm and an inner

    diameter of 192 mm was used as a model of the MapSAR bowl (made of a glass-like

    plastic material called Pyrex) with the same dimensions. The bowl is filled with a

    complex mixture of fluids simply referred to in this thesis as the liquid.

    Electromagnetic properties for the liquid as well as Pyrex are described in Table A.1.

  • 8/13/2019 TEAT-5050

    31/59

    30

    FigureA.3 Spherical Phantom

    A.4 Material Data

    3m

    kg

    m

    S r r

    Plastic 1000 0 2.9 1

    Noryl 1000 0 2.7 1

    Copper (PCB) 8920 5.9 107 1 1

    Pyrex 2500 0.99 4.6 1

    Liquid 9990 1 41.5 0.97

    Table A.1. Box Phone Material Properties Used in SEMCAD

  • 8/13/2019 TEAT-5050

    32/59

    31

    Appendix B: A SEMCAD Primer

    B.1 About

    Here follows some hints for successful simulations with SEMCAD. This should be

    suitable for the novice user and hopefully be useful for the more experienced users aswell. It aims to cover most aspects of simulation settings for mobile phone type

    simulations. This text is written for SEMCAD 1.4 Final (build 38).

    B.2 Model Design with SEMCAD

    Modelling is fairly straightforward in SEMCAD. To start a new model design click on

    the New Project icon (the blank paper sheet at the far left) and the user will be

    prompted to input the name for this project. Everything will be saved under a folder

    with this particular name in the projects folder located is in the SEMCAD installation

    folder as default, this can be useful to know when making backups or migrating

    projects to other computers. The next dialog box is the global model properties

    (Figure B.1). SEMCAD uses dimensionless global units for its axes instead of anyreal length unit so enter the corresponding real life length in meters in the dialog box.

    This can be set more or less arbitrary since one isnt limited to the global unit length

    as the shortest length. Settings are always specified in model units.

    Figure B.1 Global Model Properties

    The tools for building solids should be more or less self-explanatory, but it can beuseful to keep reference points at corners and other locations of objects that can be of

    assistance when moving or resizing objects, see Figure B.1 how to set point

    coordinates. Point coordinates are also useful when building solids. Instead of

    entering coordinates for a solid one can use the mouse to drag and model using points

    entered earlier. Points are not attached to solids in any way they are merely a quick

    way to use that particular coordinate. Unless the point coordinates as well is selected

    (when performing an operation like moving or resizing) the actual coordinates used

    will not be affected.

    Figure B.2 The Point Coordinates Dialog Box

  • 8/13/2019 TEAT-5050

    33/59

    32

    By checking the box marked Relative in Figure B.2 the new coordinate will be

    relative to the one last entered. If the point (10,10,10) was the last entered coordinate,

    entering (-10, -10, 10) with the box Relative checked the resulting coordinate is at the

    absolute position (0,0,20).

    Chapter 6.2.1 of the SEMCAD reference Manual and the Tutorial do a good job

    describing the tools available for making solids of all shapes and sizes. Learning bydoing should be the best approach here and anyone with some CAD experience

    should feel at home in no time. Be sure to give every object an identifying name to

    help future editing. Solids saved in ACIS SAT or SAB and binary STL file formats

    can be imported. However it is not possible to edit imported solids in any way in

    SEMCAD. Some desirable operations that are impossible are uniting, subtracting or

    intersecting (gives the area common to two solids) an imported solid with an imported

    or non-imported solid.

    It is possible to group solids, points, sources and sensors into groups. Clicking the

    new group window as shown in Figure B.3 creates a new folder in your objects list.

    Grouping each solid and its point coordinates into a group or folder as the case may

    be of their own makes it much easier to perform certain operations or to hide all ofthem from the model view at once. To move any object into a group highlight it and

    choose cut using the right mouse button or use the standard windows short-cut

    Control-X, then double click the folder representing the group you want the object to

    be moved to. Then use Control-V to paste the object into its new group or select Edit |

    Paste from the menu.

    Figure B.3 New Group Icon

    B.3 Sources and Sensors

    For electromagnetic field excitation there are three sources available in SEMCAD.

    These are the Plane Wave Source, Wave-guide Source and Edge Source. Only the

    Edge Source (see Figure B.4 below) will be discussed and used in this thesis. The

    Edge Source applies an electric field on one single edge and requires two points in

    space and will appear as a line in the model. These points have to be perpendicular to

    the coordinate axes. There are a few options available for specifying the properties of

    the Edge Source, which will be discussed later. We will only use the Edge Source as a

    Voltage Source that have internal resistance taking energy out of the grid making the

    generated pulses decay faster and reaching the steady state earlier.

    Figure B.4 The Edge Source

  • 8/13/2019 TEAT-5050

    34/59

    33

    To record the fields during the simulation and extract data for presentation some kind

    of sensor has to be included. The sensors available in SEMCAD correspond to the

    different types of sources used and which aspects of the simulation that is of interest.

    With the Edge Source an Edge Sensor (see Figure B.5 below) is vital during antenna

    simulations to extract certain interesting transient behaviours (such as S11and SWR

    plots) that can be helpful when matching antennas. This sensor should be positioned

    at the exact place of the source and requires two coordinates that are parallel to one ofthe grid axes as well. Using an Edge Source in phasor mode during a harmonic

    simulation presents the gap impedance, voltage, current, power, SWR and S11. The

    four first mentioned dimensions are complex numbers. With an Edge Source in time

    mode only the voltage and current are recorded, both of which can be plotted for the

    whole duration of the simulation. These can be useful to verify that the simulation has

    not diverged since they both should display a periodic signal resembling the sinus

    wave that excites the model.

    Figure B.5 The Edge Sensor

    The field sensor (see Figure B.6) records B and H-fields and features such post

    processing calculation options as various SAR plots (if run in quasi-harmonic mode)

    and Standing Wave as well as other information. Recording area for the field sensor

    can be a box or a surface (SAR only available within a box model). Field sensors

    dont lend itself very well to time domain simulations due to the extreme storage

    space required.

    Figure B.6 The Field Sensor

    The far-field sensor (shown in Figure B.7) must be modelled as a box. The box sensor

    should enclose all components interacting in a simulation. A near-to-far-field

    transformation makes it possible to place the sensor nearby the radiating structures, at

    a distance of only a few cells. Far-field sensors can only be run in quasi-harmonic

    mode and presents the far-field patterns as well as several parameters that can be

    calculated. These include the maximum radiation intensity, effective angle,

    directivity, front to back ratio and the half power beam width. For more information

    read section 9.2.3 of the SEMCAD reference guide.

    Figure B.7 The Far-Field Sensor

  • 8/13/2019 TEAT-5050

    35/59

    34

    Other available sensors are port, voltage and current sensors that will not be used nor

    discussed here (instead see the SEMCAD reference manual). The port sensor is

    similar to the edge sensor but can be positioned as a rectangular loop instead of a

    straight line as the Edge Sensor requires. Voltage and Current sensors do just what

    their names suggest.

    Figure B.8 Lumped Elements

    It is possible to model circuit components in SEMCAD. Resistors, Inductors, and

    Capacitors are available and all are inserted into the model in the same way. Just as

    the Edge Source a lumped element requires a length of the minimum step size and

    two points aligned with one of the grid axes. A single edge of the field is updated

    according to the settings of the lumped element. How to set its characteristics will be

    described later on in this text. Figure B.8 shows the lumped element icon.

    B.4 Grid Settings

    After the model has been successfully designed it is time to set the simulation

    settings. Save the project and click the Simulations tab to move on (see Figure B.9).

    SEMCAD uses an automatic grid generator that adds base lines at material

    discontinuities and sources. It is designed to make most of the work independently but

    can be tweaked by the user for certain needs. The single most important factor for a

    successful simulation is the grid settings.

    Figure B.9 The Simulations Tab

    As mentioned earlier a minimum step less than one tenth of the wavelength is

    recommended. For GSM (890-915MHz) with a wavelength around 30 cm a

    theoretical maximum step size would be 3 cm (in free space, less for materials).

    However a global max step of 13 mm is more realistic and should be sufficient for

    most GSM band simulations. SEMCAD does not set these options in any way, the

    user has to carefully consider and set these options.

    On the top left part of the screen below the simulation folder is the default grid icon

    (Figure B.10). Highlighting that icon, right clicking and choosing properties brings upthe grid settings (Figure B.11). Here are the settings for the global maximum and

    minimum step size step as well as the ratio, extents and boundary conditions for all

    axes. The ratio is the maximum rate at which nearby cells can grow. Say that a cell is

    0.3 units in a given dimension the neighbouring cell can have a maximum width of

    1.3 times 0.3 units if we use the numbers in Figure B.11.

  • 8/13/2019 TEAT-5050

    36/59

    35

    Figure B.10 The Grid Icon Figure B.11 Grid Settings

    Under the next tab labelled Solids (Figure B.12) each modelled solid can be given

    unique properties. Perfect electric (PEC) or magnetic conductors (PMC) and dielectric

    are the available options (highlight each solid and choose properties to set its type).

    Dielectric materials will be discretised according to the grid setting and can be given

    arbitrary properties whereas PEC and PMC cells will be seen as edge conditions with

    special electric and magnetic rules. Unfortunately solids require a volume so making ainfinite thin layer using a flat two dimensional rectangle and specifying it as a PEC or

    PMC is impossible, it has to be modelled as a solid with some extent. Figure B.13

    also shows the options to limit the maximum grid step in the area of the solid. This

    step size has to be in the same range as the global step or it will simply be ignored.

    Normally SEMCAD refines the step size around material discontinuities. It

    automatically uses the lowest global step size at these discontinuities. To prevent

    SEMCAD from doing this uncheck the Refine at boundaries box.

    Figure B.12 The Solids Settings Figure B.13 Properties Dialog for Solids

    One can also to some at some degree determine to which extent SEMCAD shall

    automatically generate baselines. It is recommended to use the default setting with

    baselines at all extremes unless it is crucial to keep the amount of cells down. Set the

    base line generator as None for all solids and all baselines have to set manually.

  • 8/13/2019 TEAT-5050

    37/59

    36

    Continuity thresholds are useful for solids with complex shapes or solids that are not

    aligned with any grid axes. The continuity thresholds set the distance in grid lines that

    each cell can have and still get reconnected. This should only be used if a complex

    shape like a helix antenna looks fragmented or has missing edges.

    The next tab shows the base lines that have been automatically generated with respect

    to the dielectric solids that make up our model (Figure B.14). Here one can set theminimal and maximum step size that can be used between the intervals (on the right

    side of the given base line). Step sizes that are outside the global step sizes (as shown

    in Figure B.15) will be ignored.

    It is important to visualise that the model you have made will be discretised into

    single cells, atoms or voxels as SEMCAD labels them from the perfect shapes of your

    model. And it is these base lines that govern the size of each voxel. The result will not

    contain any perfect shapes like spheres and cylinders as we have modelled but

    approximations. Some complex shapes may have generated an uncalled amount of

    baselines and some shapes may have too few baselines to get a faithful representation

    when approximated with the rectangular voxels, so it is highly recommended that one

    check the model before starting simulations. The user can add and remove base lineswithout any limitations.

    When modelling thin layers or any small sized object it is wise to model them a few

    cells thick. If that depth is in the same magnitude as the minimal step size a few

    baselines should be added on the correct axis instead of lowering the global minimal

    grid-step, which could seriously increase the computation time.

    The Subgrids tab should be ignored until it has been implemented. It is unknown

    when this feature will be available.

    Figure B.14 Base Lines Figure B.15 Compute and View Voxels

  • 8/13/2019 TEAT-5050

    38/59

    37

    B.5 Model Visualisation

    Now that the grid settings have been set, choose Compute Voxels (highlight and right

    click the grid icon as shown in Figure B.15). It discretises our model and updates each

    cells (voxel) properties. This can be seen as a transformation from our geometrically

    perfect CAD model into a three-dimensional database with each cells properties.

    These voxels can be visualised and it is recommended that one check the resultsbefore simulating. To do this choose View Voxels (as Figure B.15 shows) and make

    sure that each part of the model has a faithful representation and that every solid

    really is connected.

    A certain amount of stair casing is for some of the more complex objects more or less

    impossible to get rid of, or as the case most often would be a waste of simulation

    time. Figure B.16 shows how a sphere can end up looking. In our case we have

    decreased the step sized on the part that interferes with the telephone. A more faithful

    representation of the backside of the bowl as well is probably not worth the extra time

    spent calculating.

    The easiest way to check the geometry is to show the voxels only and not togetherwith the original solids (uncheck Original Solids like Figure B.17 shows). Only the

    voxels that are included in the slice are now shown. Using a mouse with a scroll

    wheel makes it very easy to check each slice. Just highlight the box containing the

    coordinate that isnt frozen and scroll up and down all along the model using the

    scroll wheel. Everything should be ready for simulation if the approximated models

    look acceptable and each solid has a minimum depth of a few cells.

    Figure B.16 A Stair Cased Bowl Figure B.17 Voxel Viewer

    B.6 Simulation Settings

    Next up are the simulation settings. Highlight the light bulb icon and using the right

    mouse button and choose properties (as Figure B.18). This brings up the run settings

    and the global tab (Figure B.19), where one chooses if the simulation shall run inharmonic (single frequency) or transient mode (a span of frequencies with a gaussian

    pulse). In harmonic mode it is recommended to set the simulation time in periods

    (only possible in harmonic mode) and for transient in seconds. Ten periods is usually

    long enough for simulations in harmonic mode to reach the steady state. For transient

    mode the simulation time depends on the input source and should be decided based on

    the layout of the input signal, more about this later.

  • 8/13/2019 TEAT-5050

    39/59

    38

    Figure B.18 Run Settings Figure B.19 Global Settings

    Every simulation is a trade-off between simulation time and end result, which brings

    us to the next settings: the absorbing boundary dialog. The PML algorithm is better

    and preferred for simulations with less free space surrounding the model. We tested

    both Mur and PML with 25 to 30 cm of free space around the phone in all six

    directions. The PML computation took twice as long time as the one using second

    order Mur though both S11graphs came out identical. Having plenty of free space

    around the simulated object probably affects the result in a positive way more thanchoosing the most accurate boundary condition available.

    It is safe to say that one should leave the settings unaltered for PML and Higdon

    conditions unless you really know what you are doing. MUR can be set as first or

    second order only. Figure B.20 shows the Absorbing Boundaries settings.

    Figure B.20 Absorbing Boundaries Settings Figure B.21 Edge Source Dialog

    Figure B.21 shows the properties for the Edge Source expanded. As mentioned earlier

    it is recommended to run the Edge Source as a Voltage Source for these types of

    simulations. The source resistance must be set for a voltage source since the Voltage

    Source has an internal resistor taking energy out of the grid. With a Hard Source these

    waves would have been reflected instead of taken out of the grid at the source. TheAdded Source just adds a field-strength on top of the value already calculated with the

    Yee algorithm. The advantage with the Voltage Source is that the excitation signal

    decays faster thus reaching steady state earlier and saves simulation time. If a

    Gaussian Sine is chosen its characteristics can be set by pressing Properties as shown

    at the bottom of the screenshot on labelled Figure B.21. This will bring up Figure

    B.23 to your attention.

  • 8/13/2019 TEAT-5050

    40/59

    39

    Figure B.22 TPeak and TSigma Figure B.23 Signal Settings

    The simulation duration should be at least two times TSigma, to make sure that the

    simulation will reach the steady state. A suitable value for TPeak would be three

    times TSigma to ensure that the start of the pulse doesnt get too steep. Figure B.22

    shows how TPeak and TSigma describe the signal. These values also determine the

    frequency contents of the gaussian sine. If there is a wide range of frequencies

    included like one would require for simulating both GSM and DCS bands at the same

    time, the power of these frequencies may be too small since one has to set a centrefrequency. This can be seen in Figure B.23 the graph in the bottom plot. Here we have

    a preview of a gaussian sine with a fairly broad frequency content, with less power

    around the centre frequency. If TSigma were doubled we would have seen a narrower

    spectrum and more power around 900MHz. Using a centre frequency of say

    1400MHz to simulate both 900 MHz and 1800 MHz yields not enough power for the

    interesting frequencies according to our tests. Strange things usually happen on S11

    plots for frequencies with next to none power. Using two sources each transmitting on

    one band didnt get acceptable results either. Therefore we must recommend that each

    band should be simulated one at a time.

    Figure B.24 Lumped Elements Settings

    If the model has any lumped elements they will be listed under the sensors andlumped elements tab. Highlight the lumped element of your choice and press

    properties and the a box should pop up presenting the available settings (Figure B.24).

    Depending on the circuit type that is checked the two input boxes adapts and shows

    the correct units.

  • 8/13/2019 TEAT-5050

    41/59

    40

    The Sensors tab under Run Settings lists the different sensors in the current model.

    Usually the default settings for each type will suffice after they have been set for

    phasor or time mode. For time domain data collection the start and stop time for the

    recordings can be set, as well as the number of steps so the amount of data to be saved

    can be held on a manageable level. Under Phasor mode the sensor just presents what

    was in the air so to speak when the simulation is done and reached the steady state

    (if the simulation didnt diverge).

    Figure B.25 Dielectric Materials Settings

    The next tab, Materials, lists all solids. Choosing a Dielectric solids properties brings

    up a dialog box like the one shown in Figure B.25. All changeable characteristics can

    be set from this interface. For PEC solids only the density, heat capacity and thermal

    conductivity can be set. For dielectric materials the relative permittivity and

    permeability, electric and magnetic conductivity can be set as well as the parameters

    mentioned earlier for PEC solids.

    Be advised that parameters for 900 MHz is usually not applicable for 1800MHz as

    well. Most of the human tissue parameters shown in Figure B.25 have been imported

    from a parametric model by Camelia Gabriel that is included with SEMCAD. It

    contains permittivity, permeability and conductivity for over 30 body tissues. These

    are in a text file called Gabriel parametric model.txt and should be found in the

    SEMCAD install folder by default. They are imported using the import button that is

    shown at the bottom of Figure B.25. Then open the before mentioned text file and all

    materials are available. This ends the simulation settings and the simulation can now

    be started.

    B.7 Starting a Simulation

    To start a simulation, highlight the light bulb icon and right click. Choose Compute

    Fields as Figure B.26 suggests. You must have computed the voxels before a

    simulation is possible. When the simulation is running a box like Figure B.27 show

    the status of the simulation.

  • 8/13/2019 TEAT-5050

    42/59

    41

    If your computer accesses the hard drive a lot during the simulation, take that as a

    warning. Say for example a processor running at 500 MHz. A single clock cycle then

    has a duration of 2 nanoseconds. A standard IDE hard drive running at 7200 rpm has

    an average seek time somewhere around 10 milliseconds, whereas the seek time of

    standard SDRAM usually is advertised as somewhere around 10 nanoseconds or

    faster. Quite a few more clock cycles are lost waiting for the hard drive compared to

    fetching data from RAM, so the computation time can be increased by a large factorwith heavy hard drive access. Either get more RAM or decrease the number of cells

    used in the simulation.

    Figure B.26 Starting a Simulation Figure B.27 Almost Done!

    B.8 Post-Processing

    When the simulation is done the cancel box shown at the bottom right of Figure B.27

    will be labelled with close instead. Click close and you will find something like

    Figure B.28, where every sensor is presented and their respective post processing

    options. Please note that if you edit the model or grid settings in any way allsimulation results will be erased for that particular project.

    Figure B.28 Choosing Post Processor

  • 8/13/2019 TEAT-5050

    43/59

    42

    The field sensor used in this simulation presents a variety of outputs as Figure B.28

    shows (labelled Unnamed at the top in our case). For SAR presentations the

    recommended way to present the statistics is to overlay them in the model (Figure

    B.29). This is mainly because it is easier to check SAR at different positions and the

    statistics box will give you the global maximum values if you feel that that is enough.

    The Contour view shows a 2D graphical plot but is painfully slow if you have not

    beforehand settled on which slice to display. A surface plot is in 3D and can besomewhat difficult to navigate in.

    Figure B.29 Modes of Presentation

    Field quantities are displayed on top of the 3D model in a rectangular plane of your

    choice. There are two different SAR measures available for display. Peak SAR simply

    presents the peak SAR without any averaging for mass and the IEEE-1529 uses an

    algorithm for averaging over mass (for example 1 or 10 grams) according to the draft

    with the same name. Both can use source power as input for normalisation.

    The peak SAR post processor is useful for determining where local SAR peaks are

    located and presents an easy way to see how SAR decreases as one gets further and

    further away from the radiating source. Highlight Peak SAR from under the Edge

    Source, press the right mouse button and choose Overlay in model. The next dialog

    box lists the solids in your model and an input box for normalising the input power.

    To the left of each listed solid is a box, indication if crossed that that solid will be

    used for SAR computation.

    Press the OK button when done and after some processing a sliced view similar to the

    one available when checking voxels in Figure B.17 appears. Figure B.31 shows the

    information box that accompanies the standard 3D model view. In order to see the

    graphical overlay inside the solid you chose earlier switch to wire frame or facets

    mode instead of shaded. The real information is just a mouse click away though; press

    the statistics button shown at the top right of Figure B.31. Something resembling

    Figure B.30 should appear at the left side of the screen. This presents the top SAR

    value at the current slice as well as the global maximum complete with coordinates.To check the SAR value at different parts of the bowl choose an axis to slice and

    scroll your way around the bowl while looking at the statistics box and the value to

    the right of where it says Slice Maximum field value (current settings).

  • 8/13/2019 TEAT-5050

    44/59

    43

    Figure B.30 Statistical Peak SAR Information Figure B.31 Sliced SAR

    For IEEE-1529 averaged SAR there are a few things to think about. One must set thetype of each solid that is to be used for SAR computation. Each body mass must be

    set as body tissue or extremities tissue due to the difference in the compliance

    tests for extremities as hands, wrist, ankles and the external ear. When simulating

    handsfree sets or phones that is not positioned in a talk position specify extremities.

    When averaging, only the SAR from the type of tissue that the chosen mass belongs

    to is used for calculation. The same display possibilities that the peak SAR offers are

    available for averaged SAR as well, but with all values averaged.

  • 8/13/2019 TEAT-5050

    45/59

    44

    Appendix C MapSAR manual

    C.1 Introduction

    This manual will concentrate on the basics regarding the software included with the

    MapSAR bench top SAR assessment system. Only the basic functions regarding SARmeasurements are included here, for more information about calibration issues consult

    the MapSAR documentation bundled with the product [10].

    Figure C.1 MapSAR Graphical User Interface

    C.2 Description of the Graphical User Interface

    After starting the MapSAR software the programs interface appears on the top of the

    desktop. A screenshot is shown in Figure C.1. For SAR measurements the centre of

    attention will be at the upper left windows that displays the continuous and peak

    SAR-readings. Figure C.2 shows the two displays with the continuous readings at the

    top and the max SAR value below. The max SAR value is the highest encountered

    SAR reading during the measurement.

    Figure C.2 Continuous (top box) and Maximal (bottom box) SAR Readings

  • 8/13/2019 TEAT-5050

    46/59

    45

    The big window at the upper right with the grey background displays a plot of the last

    100 samples (Figure C.3). This graph can be useful for determining if the probe has

    been moved to fast by noting sudden jumps between samples.

    Figure C.3 Plot of the Last 100 Samples

    The lower right window displays the probe movement relative to the phone surface of

    the imported image. A number of phone models are included in the with the MapSAR

    software as well as a simple program to make simple 3D models of other phones

    using pictures of the phone as aid. Figure C.4 shows an example of this presentation

    with the simple rectangular box phone used for this thesis.

    Figure C.4 Past Probe Movement

    The window on the lower left side shows a three-dimensional plot of the SAR levels

    superimposed over the simple representation of the phones design (Figure C.5). This

    is a handy way to determine the position of the maximum SAR-value. Data presented

    in this plot can be exported for further analysis. Included in an exported text file arethe phone outline together with the coordinates, conductivity and frequency. Each

    graph can also be exported as a windows bitmap file in 2D, enabling analysis of the

    near field.

    Figure C.5 SAR Surface Plot

  • 8/13/2019 TEAT-5050

    47/59

  • 8/13/2019 TEAT-5050

    48/59

    47

    Figure C.9 shows an example of the averaged SAR presentation. The cheek factor

    correction factor is for comparison with MapSAR and other measurements methods.

    Due to the relatively high thickness of the sphere used to contain the liquid for

    MapSAR for comparisons with results acquired with other measurement systems with

    cheek that is not as thick. 2mm is a common shell thickness for a SAM phantom

    whereas the MapSAR pyrex bowl is 4mm thick. And it is to compensate for this

    difference that the SAR averages may need to be multiplied with this factor. At leasttheoretically this extra thick shell should result in a lower value for MapSAR

    measurements.

    Figure C.10 Cheek Correction Calibration

    The cheek factor needs to be calibrated and the settings