TDO & RVG6X00 Kodak
Version draft
By Ryan Severud
6/26/2013
2
Change history
Date Comments EXE Version 09/22/08 Created TDO RVG6000 Kodak
document
01/06/09 Added Troubleshoot section
01/16/09 Added USB reference
01/29/09 Two or more sensors cannot be
connect into same computer
03/03/09 Added Kodak sporadically
disconnected in Troubleshoot section
4/23/09 Added picture of regedit in
Troubleshoot section
7/16/09 Added device button and sharpness
items – latest Kodak Integration 1.0.4
11/04/09 Added MS Vista info 4.6
03/04/2011 Added support to Kodak GIP 1.14. 2011.01.01
04/04/2011 New setup instructions
11/16/2012 New CSDAT instructions
4/11/2013 Updated Image Processing
instructions
6/26/2013
3
Table of Contents
Change history .................................................................................................................... 2 Table of Contents ................................................................................................................ 3 Glossary .............................................................................................................................. 4 Objective ............................................................................................................................. 5 1 SetupKodak.exe ............................................................................................................. 10 2 RVG6X00 drive installation .......................................................................................... 12 3 TDO Settings ................................................................................................................. 16 4 Troubleshooting ............................................................................................................. 20
4.1 Trophy ..................................................................................................................... 20 4.2 USB 1.1 ................................................................................................................... 20 4.3 Multiple Kodak sensors .......................................................................................... 20 4.4 Error Trophy CCX Timer ....................................................................................... 21 4.5 Kodak sporadically gets disconnected .................................................................... 21 4.6 First time Driver Installation ................................................................................... 22
6/26/2013
4
Glossary RVG6x00 Kodak X-ray Sensor – Series 6X00
TDO The Digital Office
CSDAT Carestream Digital Acquisition Tool
6/26/2013
5
Installing/Configuring CSDAT Integration for TDO
Objective
TDO now supports integration with Carestream’s CSDAT which will provide a much
better integration experience for our users.
The new integration supports new features such as being able to take multiple xrays and
having direct access to Carestream’s post processing tools.
Prerequisites
- TDO version 11.250a or newer is installed
- CareStreamIntegration.dll version 1.0.0.4 or newer
- Downloaded latest SetupRVGDriver.exe
- Downloaded latest SetupCSDAT.exe
* CareStreamIntegration.dll is located in C:\Program Files\tdoffice\dotnet folder
*Latest file can be found in U:\TDO Fixes\Kodak Files\Carestream Integration
6/26/2013
6
Setup
1. Make sure the sensor is unplugged
2. Run the SetupRVGDriver.exe
3. Click OK on the Installer Language prompt
4. DO NOT select demo mode, click next
5. Select the driver depending on the hardware they are using. Most cases with will
be RVG 5000/5100/6000/6100, click “Install”
6. Click OK to any prompts that come up during the install
7. Once the driver finishes it will launch the SetupCSDAT.exe
8. Click next through all the prompts and wait for the installer to finish
9. Plug the sensor in and let windows install the drivers for you.
a. If it cannot find the drivers you can point to this location for the drivers:
i. 32bit: C:\Program Files\Common
Files\Trophy\Acquisition\DriverRVG
ii. 64bit: C:\Program Files (x86)\Common
Files\Trophy\Acquisition\DriverRVG
6/26/2013
7
Configuring TDO for CSDAT
1. Open TDO and navigate to Administrative > Edit Control Table > Workstation
Settings
2. Under Digital Radiography select “Carestream 6000 series” from the drop down.
It should fill in the CSDAT folder path for you.
a. If it doesn’t the path is:
i. 32bit: C:\Program Files\Carestream\Digital Acquisition Tool
ii. 64bit: C:\Program Files (x86)\Carestream\Digital Acquisition Tool
6/26/2013
8
Acquiring X-rays for CSDAT
1. In TDO go to acquire, select an image category, and click on XRay. We no
longer require sensor activation through TDO. You should be met with the
familiar count down window.
6/26/2013
9
2. Once they expose the sensor the image will go into the new TDO Image
Processing window:
A: Gives the user the ability to have the image auto rotate or apply sharpness filter
B: The user can adjust the brightness/contrast to their liking. The values will be stored
and applied to all images from this point on.
C: Toggles the sharpness filter on and off
3. The user now has an option to either acquire another image by clicking on New
RVG or edit an image.
4. Once they are happy with the image, click the “Send to TDO” and it will
automatically save to the patients Image Organizer as the category that was
selected.
6/26/2013
10
Objective (OLD TDOKodak6000.exe)
Congratulations on your new Kodak purchase! This document presents
the full installation and integration of RVG6000 Kodak x-ray sensor and
TDO.
Follow the next 3 steps:
SetupKodak.exe
RVG6000 drive installation and
TDO Settings
1 SetupKodak.exe
Make sure to unplug the RVG6X00 sensor during this step. And also,
beforehand check the MS and Office’s updates! Do not install Kodak RVG
sensor if you don’t have these items correctly installed!
Beware that Trophy v5.0.5.0 is incompatible with RVG6000. Please, uninstall
it before continuing this installation!
Download “SetupKodak.exe” from the TDO Builds ftp server. Run it.
6/26/2013
11
PS: NEVER CHECK THIS CHECKBOX “Demo mode”!!!!
If there is any message about “Windows can’t verify the publisher of this
driver software”, please accept and install the driver:
6/26/2013
12
2 RVG6X00 drive installation
Connect RVG6000 sensor in the USB port. Let Windows to recognize the
new hardware.
6/26/2013
13
6/26/2013
14
Windows will install it twice, because of 5000 series.
6/26/2013
15
And then finally:
*** If for whatever reason Windows cannot find the drivers. You can
manually point to where the drivers reside which is in:
C:\Program Files\Common Files\Trophy\Acquisition
Or if a 64bit workstation
C:\Program Files (x86)\Common Files\Trophy\Acquisition
6/26/2013
16
3 TDO Settings
In TDO, open Administrative Edit Control Table Workstation Settings.
Select “Kodak 6000 Series” in Radiography Software. Note that
Path/Source is set as “C:\Program Files\TDOffice\Kodak” or if the
workstation is 64 bit, the path would be: “C:\Program Files
(x86)\Tdoffice\Kodak”
6/26/2013
17
Note the “Use old Kodak sharpness filter” option. If this is checked, the
next xray acquired by Kodak 6000 series will have the old sharpness filter.
Close it.
Open “Acquire”
And hit “X-Ray” button. For the very first time only, the RVG6000 needs to
be registered. Therefore, you will be prompted with this message box:
6/26/2013
18
Once registered, hit ‘x-ray’ button again and it is ready to take an x-ray.
6/26/2013
19
Once the x-ray is taken, apply the filters, if necessary, and then click on
“Send to TDO” button.
On the other hand, when cancelling the x-ray, this is the message:
For multiple xrays, the device button can be useful: open xrays in TDO,
and it will automatically be waiting for the first xray. Once it has been
taken, click on the device button to take the next one, without the need
to click on the xray button again in TDO.
6/26/2013
20
4 Troubleshooting
4.1 Trophy
Beware that some Trophy software versions do not provide the “uninstall”
feature. Delete them only will not allow TDO acquires x-rays.
Solution: rename the folder “TW” to “TW-bak” and reboot the computer.
Also check the HKEY_CLASSES_ROOT\TROPHY\RVG that should be
something like this:
4.2 USB 1.1
Kodak sensor only runs on USB 2.0.
Solution: their IT team has to change to USB2.0.
4.3 Multiple Kodak sensors
Kodak driver does support only ONE sensor connected into the computer.
If, two or more sensors are connected into the same computer, TDO
keeps requesting the activation code.
Solution: Do NOT connect two or more sensors into the same computer.
“Acquire” must be closed before unplugging/plugging a different sensor.
6/26/2013
21
4.4 Error Trophy CCX Timer
If while installing Kodak sensor, the “Check to enable synchronization link”
checkbox is marked, then x-ray will not be acquired.
Solution: set HKEY_CLASSES_ROOT\TROPHY\RVG\Synchro from zero to 2.
Open Demo_GIPSDK.exe and run Load GIP Toolkit menu option. Then
run RVG Open the driver menu option and RVG Acquisition menu
option. Check that it successfully acquired an image. Rename
Demo_GIPSDK.exe and restart TDO. Check that now TDO is able to
acquire successfully.
4.5 Kodak sporadically gets disconnected
If Kodak sensor gets disconnected, it could be caused by the power
management that sets the USB to sleep mode. It is not possible to
determine the exact USB Root Hub Kodak sensor is connected. Therefore,
the solution must be applied for ALL USB Root Hub.
Solution: From the “Device Manager”, for each USB Root Hub, uncheck
the option “Allow the computer to turn off this device to save power” in
Power Management tab, as shown below:
6/26/2013
22
4.6 First time Driver Installation
When installing the drivers for the very first time, you might receive an error
message in TDO that the sensor is not connected. Usually this message is
displayed after a “Initializing Sensor” message is displayed. If this happens,
just let the TDOKodak9000 windows close normally and click on TDO
Acquire XRay button again and everything will work properly.