takaya hardware
-
Upload
villanegra -
Category
Documents
-
view
351 -
download
7
Transcript of takaya hardware
-
8/17/2019 takaya hardware
1/40
Fixtureless Tester
APT 9400CE/CJ
Operator s guide
TAKAYA CORPORATION
-
8/17/2019 takaya hardware
2/40
-
8/17/2019 takaya hardware
3/40
1
Preface
Takaya Fixtureless tester APT-9400 is the in-circuit tester (MDA) with four indendently moving
probes that does not require any pin board jig (Bed-of-nails). It is designed as the high performance
tester with various useful functions so that it can be used for testing production board from a small
to medium lot, a wide range of sizes, prototype and SMD board tests. It can also be used for
repairing.
This operator s guide, which is separated into two parts (Hardware section, Software section),
contains all information needed for an optimal use of your tester. Please read this operator s guide
before using your tester. Then keep this operator s guide handy for answers to any questions youmay have.
If you have any questions or thoughts you would like to share with us ! we would like to hear from
you.
(NOTE)
The design of the tester and software are under constant review and while every effort is made to
keep this operator s guide up to date, the rights is reserved to change specifications and equipment
at any time without prior notice.
And this operator s guide assumes that you have a standard knowledge about your computer and
the Windows NT operating system. If you have questions, please consult your computer and
Windows NT reference documentation.
Copyright 1999, 2001 TAKAYA CORP. All rights reserved. No portion of the contents of this
publication may be reproduced or transmitted in any form or by any means without the expresswritten permission of TAKAYA CORP.
Microsoft Windows and Windows NT" is the trademark of Microsoft Corporation.
-
8/17/2019 takaya hardware
4/40
2
Safe Symbols & Labels
Pay special attention to the symbols described in this operator s guide.
Symbol Explanation
Calls attention to a procedure, practice, or condition that could possibly cause
serious accident or death.
Calls attention to a procedure, practice, or condition that could possibly cause
bodily injury or damage to the tester.
Calls attention to a procedure, practice, or condition that could possibly
damage to the tester.
Calls attention to general instruction.
Failing to follow this could loss of data stored on disks, cause possibly misjudge
the unit under test, or damage to the tester.
Calls attention to especially key operation. It navigates you to operate the
process.
Calls attention to #One-point advice$ which should be useful when you are at a
loss to operate the tester.
Also, the tester is labeled with some important messages that invite your utmost attention.
Hand hazard
This label calls attension to possible crushing/punching
hands or fingers. Please never put any part of your body
nearby the moving parts/area to avoid the serious
accident.
Electrical shock hazard
This label calls attension to the possibility of electric shock.
Please never open the panel to avoid the serious accident.
This label calls attension to the possibility of accident or
trouble. Please never open, disassemble or modify this
part to avoid the serious problem.
(IMPORTANT)
1) Do not peel these seals off.
2) Do not write any contents on these seals.
3) Do not shift these seals to another location.
3) Do not remodel these seals.
-
8/17/2019 takaya hardware
5/40
3
Precautions
All instructions below are very important !
Do not attempt to install and/or use the tester unless you could understand any of
them.
(1) Like any other delicate electronic device, the tester should be put in a suitable location. This
location should be dry and kept away from direct sunlight or water. Otherwise it may possibly
cause an electric shock or other bodily accident.
(2) Place the tester on a flat, sturdy floor or position where can enough sustain the weight of the
tester. Otherwise it may possibly cause such serious accident like a falling floor. After placed
the tester, fix it tightly with adjuster feet.
(3) Never modify or repair any part of the tester by yourself. Otherwise it may possibly cause an
electric shock or other bodily accident.
(4) For safety reason, it is recommended that only one person operate the tester. If not, the tester
may cause an unthought of action and a serious accident as the result.
(5) Never attempt to put a part of your body in the working area of the tester. Otherwise it may
possibly cause a serious accident.
(6) When the tester is operated, always close the Safety cover. Failure to observe this warning
may result in a serious bodily accident.
(7) Once some smoke or bad smell came from the tester, it may possibly cause a burning or
electric shock and is very dangerous. In such cases happened, switch the power off and stop
operating the tester on the spot, then consult to our local distributor for the solution. Do not use
the tester until safe operation can be verified by service-trained personnel.
(8) Once some liquid or foreign material was put in the tester, it may possibly cause a burning or
electric shock and is very dangerous. In such cases happened, unplug the tester and stop
operating the tester on the spot, then consult to our local distributor for the solution. Do not use
the tester until safe operation can be verified by service-trained personnel.
(9) Once you felt something strange during the operation (i.e. noise, smell, malfunction, etc.),
switch the power off, unplug the tester and stop operating the tester on the spot, then consult
to our local distributor for the solution. Do not use the tester until safe operation can be verified
by service-trained personnel.
-
8/17/2019 takaya hardware
6/40
4
(10) Make sure that the tester must use power source shown on the rated label. Otherwise it may
possibly cause a burning or electric shock.
(11) The tester power must be OFF before plugging in/out.
(12) The tester power must be OFF and unplugged before relocating the tester.
(13) The tester power must be OFF and unplugged when the tester is serviced for maintenance or
cleaning. Needless to say, these services must be made by service trained personnel only
who are familiar with electronic circuitry and understand the hazard involved.
(14) The tester power must be OFF and unplugged when either the peripheral devices or internal
PC boards (i.e. Measuring card, Scanner card) are connected or removed
(15) Make sure that the tester is earthen using a thick wire (0.75mm or more).The GND terminal
should not be grounded to such location (Ex. a conveyor rail) where may be not grounded
properly. Otherwise it may cause serious electric shock and is very dangerous.
(16) Unplug unless the tester would be powered for a long period of time.
(17) Do not store disks near motors, telephones or other magnetic sources. If the data is critical,
always make sufficient backup copies under your responsibility.
Also, the manufacturer has no responsibility for disappearance of test data resulting from
interruption of electric service or other accident.
(18) The manufacturer has no responsibility for such defect resulting from misuse or unauthorized
modification by user, as well as operation outside of the environmental specifications for the
tester.
-
8/17/2019 takaya hardware
7/40
-
8/17/2019 takaya hardware
8/40
6
a
b c d e f
g h i j k
l m n o p
q r s t
-
8/17/2019 takaya hardware
9/40
7
Size and Weight
3 8 0
Front conveyor rail
1 2 3 0
1 6 5
1 9 5
8 7 0
Fig. 1 Top view
9 5 0
1 8 1 5
3 3 0
5 3 5 675 1325
F L
9 0 0
Conveyor level
2000
Fig. 2 Front view
Dimension: W 2,000mm ! D 1,230mm ! H 1,815mm (including Signal towe)
(Excluding the Monitor/Keyboard arm)
Weight: Approx. 1,100 Kgs
-
8/17/2019 takaya hardware
10/40
8
Name of Components
Safety cover
Emergency stop SW
Power ON/OFF SW
Power ON lamp
Operation panel
Reset SW
Tester terminal
Signal tower
Operation panel(Conveyor/Supporter)
Emergency stop SW
Safety cover
Fig. 3
Supporter SW
Conveyor run SW
Reset SW
TRAY OUT SW
TRAY IN SW
AUTO RUN SW
AUTO STOP SW
Tray operation & AutoRum/Stop Operation oanel (Conveyor/Supporter)
Fig. 4
Rear panel
Earth terminal
Rear panel
regulatorAir filter
Air IN
Relief valve
AC IN
Safetycover SW
Safetycover SW
Safetycover SW
Fig. 5
-
8/17/2019 takaya hardware
11/40
9
Safe System
This section contains the important information to use the tester safely. Please read
this section carefully prior to your operation. If you mis-operate the tester, it may
cause serious accident.
Various safety functions are built in the tester to avoid any serious accident. Never
modify or makeover those safety function by yourself. If you done, we get off any
responsibility of the accident.
The APT-9400 is equipped with safety system to suspend their moving mechanism on the spot for
an accidental trouble. Once Emergency Stop SW and/or Safety Cover SW operated, it suspends
every mechanical movement.
Fig. 6 Safe system
Emergency stop SWPressing the Emergency Stop SW, the power supply to the moving mechanism is forcibly cut and
the tester stops every operation on the spot. During this stop, it shows #Device is now suspended by
emergency stop!$ error on the display.
To release the Emergency Stop, turn the Emergency Stop SW clockwise until it sounds click, then
depress the Reset SW. (No cancellation of the emergency stop is possible while the safe cover is
open)
Be sure to release the Emergency stop SW while the safe cover is closed. Otherwise
it cannot be released.
Emergency stop SW
Signal tower
Safety cover
Emergency stop SW
Power ON/OFF SW
Power ON lamp
Safety cover
Reset SW
-
8/17/2019 takaya hardware
12/40
10
Safety cover
The tester is designed not to operate while the safety cover is open. Once the safety cover opened
during the operation, the power supply to the moving mechanism is forcibly cut and the tester stops
every operation on the moment. During this stop, a message #Cover is open!$ appears on the
display.
After the tester was suspended by opening the safe cover, close the safe cover again, then depress
the Reset SW.
Reset SW
The Reset SW enables to release the supension of the tester caused by inputting the Emergency
stop SW and/or opening the safe cover. After the suspension was released, all probes moves back
to the home position.
Be sure to press the Reset SW after the safe cover was closed and the Emergency
stop SW was released.
Disposal of emergency
In case of the emergency, depress the Emergency Stop SW to stop the operation immediately. (In
addition, power off the tester if you judged it necessary.)
To release the Emergency Stop, first remove all cause which suspended the tester, next turn the
Emergency Stop SW clockwise until it sounds click, then depress the Reset SW.
When the tester was powered off, next turn the Emergency Stop SW clockwise until it sounds click,
then power on the tester again.
If you judges the tester could not run in succession, power off the tester, and unplug
the AC cable as soon as possible. Then contact your local distributor.
-
8/17/2019 takaya hardware
13/40
-
8/17/2019 takaya hardware
14/40
12
Rail height FL=900mm (-15mm / +65mm)
Positioning PCBs Stoppers
Reform PCB warp By automatic side clamp mechanism
Belt Synthetic rubber belt
Signal tower 3 colors lamp (red/green/yellow) with buzzer
LED (red/blue) light source for the camera systemMain function
Printer (40 column red/black dot impact type, fail ticket width: 114.3mm / 4.5$)
Power su l AC 200 220 240V / 50/60Hz / 2.5KVA
Operation sound Maximum 66 dB
Air supply 0.5~0.7 Mpa, dry and clean air, 11L/min (at usual operation)
Environmental requirementsTemperature: 15 ~ 30 degree centigrade
Humidity: 35 ~ 70% / Altitude: lower than 1,000m
Preservation/Transportation Temperature !25 ~ 55 degree centigrade / Humidity 20 ~ 80% (no condensation)
Dimension / weight 2,000mm(W) ´ 1,230mm(D) ´ 1,830mm(H) / 1,100Kgs (without any option system)
System Requirements
Computer type PC/AT compatible PC (IBM clone), Tower type
CPU Pentium III 500MHz or higher
Memory 128MB or more (256MB recommended)
VRAM 4MB or more (16MB recommended)
FDD 3.5 inch (2HD) drive
CD-ROM 1 drive for System software install (CD-RW system recommended)
HDD 4GB or more (Free space needs approx. 30MB for system installation)
Operating system Microsoft WindowsÒ NT 4.0 (SP6)
Expansion slots 2 PCI slots or more
Monitor 17$ CRT XGA (1024 x 768 or higher resolution) High (65,535) colors or more
Keyboard Full type
Mouse PS/2 type
Centronics compatible port (for printer, LPT port)
RS-232C serial port (COM port)
Keyboard port (PS/2)Interface
Mouse port (PS/2)
More than 170 mm (W) ´ 300 mm (D) ´ 280 mm (H)Dimension (Computer)
Less than 220mm (W) ´ 500mm (D) ´ 500mm mm (H)
D300 D500
H 5 0 0
W220W170
H 2 8 0
Fig. 7 Computer Dimension (Minimum/Maximum)
-
8/17/2019 takaya hardware
15/40
13
Installation
This section contains information to aid in your installing the APT-9400.
When the tester is relocated, be sure to deal it with two persons or more. Also, pay your
attention not to squeeze a part of your body into the casters and the level pads.
Never fail to check that the AC power cable is unplugged and the tester is powered off
before you start installing the tester.
Be sure to select a flat and hard floor where can sustain the weight of this tester.
Otherwise it may cause such serious accident like a falling floor.
How to Install
1. Unpack your tester, and bring up four level pads to ground the casters.
2. Move the tester to a suitable location where is flat and sturdy, then place it with the level pads.
Use the hand-lift whenever the casters are useless to move the tester.
3. Unscrew all hexagon socket head cap screws at position (1) ~ (3) in Fig.5 that fix the moving
mechanisms not to make them interfere each other during the transportation.
4. Cut a vinyl band which is installed at position (4) in Fig.5 by using a diagonal clipper, etc.
AA
SA940-35249W0Fitting for X-armFixing point for X-arm( 1 )
Fitting for Y-armSA940-45969W0 Viewed from arrow-A
Fixing point for Y-arm ( 3 )
Here is tied with a vinylband.
Fixing point for Z-arm( 4 )
Here is tied with avinyl band.
Fitting for Y-armSA940-45969W0
Fixing point for Y-arm( 3 )
Fixing point for Z-arm( 4 )
M5X45
Hexagon scoket head
Fitting for supportc.screw( 2 )
Fig. 8 Top View
In cutting the vinyl band, care should be taken that the other wires are not cut,
stretched or loaded.
-
8/17/2019 takaya hardware
16/40
14
5. Set FL (=the height from the floor to the conveyor rail) for #900mm$ by four adjuster bolts as
per Fig.9.
Machine base
F L
:
9 0 0
Adjuster bolt
Level pad
Conveyor rail
Water Level
Fig. 9 Side View
2
3
Water Level
Water Level
4 Water Level
1 Water Level
Fig. 10 Top View
6. Put a water level on the position (1), (2) in Fig.10, and fine-tune the adjuster bolts while
checking the water levels if the tester is placed horizontally with accuracy of 0.2 mm / 1,000
mm or less. (Refer to Fig.11 ,12)7. Put a water level on the position (3), (4) in Fig.10, and fine-tune the adjuster bolts while
checking the water levels if the tester is placed horizontally with accuracy of 0.2 mm / 1,000
mm or less. (Refer to Fig.11 ,12)
8. Put a water level on the position (1) ! (4) in Fig.10, and check the water levels if the tester is
placed horizontally with accuracy of 0.2 mm / 1,000 mm or less. If 0.2 mm / 1,000 mm or less
was not achieved, repeat the above procedure.
9. Make sure again if the FL is 900 mm. If not 900 mm, repeat the above procedure.
10. All that you need to do for the installation finished. Next, setup your PC peripheral devices.
-
8/17/2019 takaya hardware
17/40
15
How to increase FL
Insert an iron rod (f10) in the adjuster bolt, and turn it clockwise while fastening the nut by hand.
Rod (accessory)
Increase
Nut
Adjuster bolt
Increase
Level pad
Fig. 11 Increase FL
How to decrease FL
Insert an iron rod (f10) in the adjuster bolt, and turn it counterclockwise while fastening the nut by
hand.
Decrease
Decrease
Fig. 12 Decrease FL
It may be necessary to adjust this adjuster bolt more than once.
-
8/17/2019 takaya hardware
18/40
16
Installing Signal tower
This section contains information to aid in the installing a signal tower.
WARNING Never fail to check that the AC power cable is unplugged and the tester is powered off before
you start installing the tester.
How to Install Signal tower
1. Remove the rear panel.
2. Install the signal tower on the cover and fasten it with the hexagon nut.
3. Connect the relay cable and wire the cable not to touch any moving part.
4. Install the rear panel.
5. Fix the hexagon nut tightly by using the spanner or screw wrench.
retsetehtfoedisraeR
srotcennoC
tunnogaxeH
tunnogaxeH
lenaPraeR
rewotlangiS
924-SHAYAKATelbacyalerrewotlangiS
Fig. 1 Signal tower
-
8/17/2019 takaya hardware
19/40
17
Installing PC peripherals
This section contains information to aid in your installing the PC peripherals.
Never fail to check that the AC power cable is unplugged and the tester is powered off
before you start installing the tester.
How to Install PC peripherals
1. Put the CRT on the monitor plate. (Refer to Fig.14, 15)
Keyboard
Mouse
Monitor Plate
PC rack
Fig. 2 Overview
Monitor plate
Monitor
Fig. 15 Monitor set
-
8/17/2019 takaya hardware
20/40
18
Mouse
Keyboard
Keyboard base
Keyboard attachement
Fig. 3 Keyboard & Mouse
2. Install both the keyboard and the mouse. (Refer to Fig.16)
3. Install the personal computer in the PC rack. (Refer to Fig. 17)
CoverPersonal Computer
Fig. 4 PC installation
4. Fix the personal computer with the standard fittings as per Fig.18.
D500D300
H 2 8 0
W170W220
H 5 0 0
Fig. 5 PC dimension (Minimum and Maximum size)
-
8/17/2019 takaya hardware
21/40
-
8/17/2019 takaya hardware
22/40
20
Earthing / Air supply
This section contains information to aid in the connection of GND, the air piping and the powercable.
WARNING
Never fail to check that the AC power cable is unplugged and the tester is powered off before
you start installing the tester.
Air inlet
GND terminalPower cable
Relief valve Air filter / Regulator
Fig. 7 Rear panel
1. It is always necessary to earth the GND terminal on the rear tester using a thick wire (0.75mm
or more), with the except that the tester is properly grounded through the power cable in which
the earth lead is included as standard.
WARNINGThe GND terminal should not be grounded to such location (ex. your conveyor)where may be not grounded properly.Otherwise it may cause serious electric shock and is very dangerous.
2. Plug in the standard air hose to # Air inlet$ on the rear tester, and plug in another side of the air
hose to your air source. (Refer to Fig.20)
3. Then check the air pressure gauge to see if the air pressure is 0.5 MPa while supplying the air
to the tester. If it was out of rating value, verify if the air pressure from the air source is 0.5 MPa
-0.51 MPa. (Refer to Fig.21)
In case that the correct air pressure could be verified, pull up the adjustable knob of the air
-
8/17/2019 takaya hardware
23/40
21
filter regulator to readjust the air pressure to be 0.5 MPa. After adjusted, presses down the
adjustable knob to the original position.
Fig. 21 Air configuration
4. On the tag plate on the rear tester, Power requirement, Maximum current consumption and
Maximum voltage consumption are listed.
WARNING
Be sure to use clean dry air which excludes oil, foreign body and moisture.Otherwise the pneumatic actuators may be damaged.
5. Be sure that your local power source complies with our power specification before you plug inthe power cable.
WARNING
Never supply wrong power voltage to the tester. Otherwise it may cause seriouselectric shock or fire and is very dangerous.
Pressure meter
Adjustable kncb
Air filter/Regulator
" # $ %
8
10
&
'( )*+,
%
1
0.4 0.6
2
6
4
0.2 0.8
-
8/17/2019 takaya hardware
24/40
22
Operation panel
Arrow switches. (Probe moves in direction of arrow)
Used for inputting and correcting X/Y coordinates.
PROBE DOWN Set probes in the #down$ position. Once the SW is pressed, the probe
remains down until any other key is pressed.
FAST Increase the moving speed of probes.
SLOW Decrease the moving speed of probes.
STEP Decrease the moving speed of probes to 2.5um/step movement.
PROBE 1-4 Verify and/or correct X, Y coordinates at Ste data review menu.
ENTER Enter XY coordinates.
PRINT Output fail report to the printer.
TEST START Move probes or start test run.
TEST STOP Stop test run.
RETEST Retest fail steps only (with printout).
F1 Restart the test run that was broken off.
F2 Retest fail steps only step by step (without printout).
F3 Output #all test results$ on the printer.
F4 Change the conveyor rail width.
F5 Move the conveyor rail to home position.
-
8/17/2019 takaya hardware
25/40
23
Conveyor operation
TRAYOUT
TRAY AUTO AUTO
RUN STOPIN
Side Desk Switches
RESET
SWITCH
CONVEYOR RUN
SUPPORTER
Conveyor Switches
Fig. 22
TRAY OUT Used for pull out the tray from the test position to the side desk.
TRAY IN Used for put in the tray from the side desk to the test position.
AUTO RUN Used for starting the test (including conveyor movements).
AUTO STOP Used for stopping the test (including conveyor movements).
RESET SWITCH Used for reset the emergency stop signal.
Left Used for pull out the PCB from the test position to the side desk.CONVEYOR RUN
Right Used for put in the PCB from the side desk to the test position.
Up Used for supporting the PCB.SUPPORTER
Down Used for releasing the PCB support.
-
8/17/2019 takaya hardware
26/40
24
Maintenance
Replacing Probe-1, 4
When you need to replace the probe, be sure to perform as follows:
Particular care should be paid to the keen-edged head of the probe pin to be
replaced. Otherwise it may be an injury to your body.
There is somespace
space
Probe socket
There is some
Viewed from arrow-A
Probe holder
Probe pin
Probe socket
Probe holder
Probe pin
Probe lead wire
A
Hex. bolt (A)
Hex. bolt (B)
Fig. 23 Probe replacement 1 or 4Name Type
Probe (Standard) XX-00222 (Crown type)
Probe (Option) XX-00221 (Needle type)
Probe socket SA941-46616P0
1. Select [Tool]ð [Self Diagnosis] ð [Adjust Probe] ð [Probe 1] (or [Probe 4]) on the APT menu.
2. Press [TEST START] SW on the operation panel.
3. Open the Safe cover, after the Z-1 axis (or the Z-4 axis) moved in the front side.
4. Release or loose the Hex bolt (A) to detach the probe holder unit from the Z-1 block (or the Z-4
block) as per Fig.23.
5. Pull out the probe lead wire straight from the probe.
6. Loosen the Hex bolt (B), and pull out the probe as per Fig.23.
7. Put a new probe in the probe socket and put them together into the probe holder tightly as per
Fig.23.
8. Fasten the probe with the Hex bolt (B).9. Connect the probe lead wire to the probe.
-
8/17/2019 takaya hardware
27/40
25
10. Install the probe holder unit on the Z-1 block (or the Z-4 block) and fasten it with the hexagon
socket head cap screw.
11. Close the safety cover, and depress [RESET] SW then select #Continue$ button on the screen.
12. Depress [TEST STOP] SW.
13. Set the Machine Reference Point at the Master mode again.
14. After replaced the probe-4 block, be sure to readjust the Camera offset at the Master mode.
(NOTE)
Please detach the probe holder whenever you would like to replace the probe. If not, it may possibly
break the probe holder itself.
The probe must be tightly inserted so as the flange is in contact with the probe holder.
-
8/17/2019 takaya hardware
28/40
26
Replacing Probe-2, 3
When you need to replace the probe, be sure to perform as follows.
Particular care should be paid to the keen-edged head of the probe pin to bereplaced. Otherwise it may be an injury to your body.
Probe lead wire
Probe holder
Probe pin
Hex. bolt (A)
A Probe holder
Probe pin
Probe socket
Hex. bolt (B)
Probe socket
Viewed from arrow-A
There is some
space
spaceThere is some
Fig. 24 Probe replacement 2 or 3
Name Type
Probe (Crown type) XX-00222(Crown type)
Probe (Needle type) XX-00221(Needle type)
Probe socket SA941-46616P0
1. Select [Tool]ð [Self Diagnosis] ð [Adjust Probe] ð [Probe 2] (or [Probe 3]) on the APT menu.
2. Press [TEST START] SW on the operation panel.
3. Open the Safe cover, after the Z-2 axis (or the Z-3 axis) moved in the front side.
4. Release or loose two Hex bolts (A) to detach the probe holder unit from the Z-2 block (or Z-3
block) as per Fig. 24.
5. Pull out the probe lead wire straight from the probe.
6. Loosen the Hex bolt (B), and pull out the probe as per Fig. 24.
7. Put a new probe in the probe socket and put them together into the probe holder tightly as per
Fig. 24.
8. Fasten the probe with the Hex bolt (B).
9. Connect the probe lead wire to the probe.
10. Install the probe holder unit on the Z-2 block (or the Z-4 block) and fasten it with the hexagon
-
8/17/2019 takaya hardware
29/40
27
socket head cap screw.
11. Close the safety cover, and depress [RESET] SW then select #Continue$ button on the screen.
12. Depress [TEST STOP] SW.
13. Set the Machine Reference Point at the Master mode again.
(NOTE)
Please detach the probe holder whenever you would like to replace the probe. If not, it may possibly
break the probe holder itself.
The probe must be tightly inserted so as the flange is in contact with the probe holder.
-
8/17/2019 takaya hardware
30/40
28
Replacing Probe holder unit (Probe-1 or 4)
When you need to replace the whole probe holder unit, be sure to perform as follows.
There is somespace
space
Probe socket
There is some
Viewed from arrow-A
Probe holder
Hex. bolt
Probe pin
Probe socket
Probe holder
Probe pin
Probe lead wire
Hex. bolt
A
Fig. 25 Fig. 26
1. Select [Tool] ð [Self Diagnosis] ð [Adjust Probe] ð [Probe 1] on the APT menu. (Refer to
Fig-27)
2. Press [TEST START] SW on the operation panel.
3. Open the Safe cover, after the Z1 axis moved in the front side.
4. Unscrew a hex. bolt, and remove the existing probe holder unit from the Z1 axis. (Refer to
Fig-25)
5. Pull out the probe lead wire straight.
6. Install a probe pin in the probe socket of the new probe holder unit. (Refer to Fig-26)
7. Connect the probe lead wire to the probe pin.
8. Install the new probe holder unit on the Z1 axis and fix it with the hex. bolt.
9. Close the Safe cover and press [Reset] SW, then click on [Continue] button.
10. Click [Probe 4] button on the Adjust Probe menu.
11. (The Z1 axis goes back, and then the Z4 axis comes in the front side.)
12. Repeat the same process 3 to 9 above for the Z4 axis.
Fig.27 Adjust Probe menu- - -To be continued to the process 13 in the next page - - -
-
8/17/2019 takaya hardware
31/40
29
Replacing Probe holder unit (Probe-2 or 3)
A Probe holder
Hex. bolt
Probe pin
Probe socket
Probe socket
Viewed from arrow-A
There is somespace
spaceThere is some
Probe lead wire
Probe holder
Probe pin
Hex. bolt
Fig. 28 Fig.29
13. Click [Probe 2] button on the Adjust probe menu.
14. (The Z4 axis goes back, and then the Z2 axis comes in the front side.)
15. Open the Safe cover.
16. Unscrew two hex. bolts, and remove the existing probe holder unit from the Z2 axis. (Refer to
Fig-28)
17. Pull out the probe lead wire straight.
18. Install a probe pin in the probe socket of the new probe holder unit. (Refer to Fig-29)
19. Connect the probe lead wire to the probe pin.
20. Install the new probe holder unit on the Z2 axis and fix it with the hex. bolts.
21. Close the Safe cover and press [Reset] SW, then click on [Continue] button.
22. Click [Probe 3] button on the Adjust probe menu.
23. (The Z2 axis goes back, and then the Z3 axis comes in the front side.)
24. Repeat the same process 15 to 20 above for the Z3 axis.
25. Close the Safe cover and press [Reset] SW, then click on [Continue] button.
26. Select [Close] button on the Adjust Probe menu.27. Press [TEST STOP] SW on the operation panel.
28. Set the Machine Reference Point again. (Refer to Operation manual)
29. Then, next readjust the Camera offset according to the menu guidance.
30. Finish
Particular care should be paid to the keen-edged head of the probe pin to be
replaced. Otherwise it may be an injury to your body.
-
8/17/2019 takaya hardware
32/40
30
Replacing Bottom probes
When you need to replace the probe (bottom probe), be sure to perform as follows.
Pay attention to handle the tip of the probe not to be injured somebody.
Probe pin
Socket
Bottom probe
Probe pin
Socket
Needle nose plier
Fig. 30 Bottom probe replacement
Name Type
Probe (Crown type) C-2SLL-HD
1. Pull the probe pin out from the socket, using a needle nose plier, so on.
2. Insert a new probe pin into the socket.
(NOTE)
The probe must be tightly inserted so as the flange is in contact with the probe socket. The probe
socket must be tightly inserted so as the flange is in contact with the bottom probe.
-
8/17/2019 takaya hardware
33/40
31
Replacing Sensor probes (IC Open test system)
When you need to replace the sensor probe (IC Open test system), be sure to perform as follows.
X-direction
AA
blockblock
Sensor probe
Connector
Antenna driveAntenna drive
Sensor probe
Hex.bolt Hex.bolt
Connector
Fig. 31 Sensor probe replacement
Name Type
Sensor Probe CS9000-A23
Sensor Plate CS8000-S21
1. Open the safety cover.
2. Loosen the Hex bolt and detach the sensor probe from the antenna drive block as per Fig. 31..
3. Disconnect the connector.
4. Replace the sensor probe with new one.
5. Connect the connector and fix the Hex bolt tightly.
6. Set the Sensor Probe/Probe offset at IC Open menu in Option mode.
There issome space
Holder
Sensor Probe
X-direction
A border line between green sideand white side should be in parallelwith the X-direction.
Fig. 32 Remark point of sensor probe replacement
-
8/17/2019 takaya hardware
34/40
32
Replacing Probe lead wire (Probe-1 or 4)
When you need to replace the probe lead wire, be sure to perform as follows.
Pay attention to handle the tip of the probe not to be injured somebody.
Probe holder
Hex. bolt
Probe lead wire
Relay PCB
Screw
Fig. 33 Layout of the probe lead wire
Cable lock
Mark-tube Cable lockMark-tube
Mark-tube
Cable lock
Fig. 34 Connection diagram
Name Type
Probe lead wire TAKAYA HS-432
-
8/17/2019 takaya hardware
35/40
33
1. Proceed #Tool$ ð$Self diagnosis$ ð$ Adjust probe$ on the system menu, then select either
Probe-1 or 4 and depress [TEST START] SW.
2. Open the safety cover.
3. Release the Hex bolt, and detach the probe holder. (Refer to Fig.33)
4. Pull out the probe lead wire straight.
5. Unscrew the screw on the Probe Relay PCB, and remove the probe lead wire.
6. Install a new probe lead wire to the Probe Relay PCB.
7. Wire and fix the probe lead wire with the cable lock as per Fig. 34.
(NOTE) In this case, be sure that #Mark-tubes$ locates as per Fig. 34.
8. Connect the probe lead wire to the probe.
(NOTE) The probe must be tightly inserted so as the flange is in contact with the probe holder.
9. Install the probe holder on the Z-1 or 4 block and fasten it with the hexagon socket cap screw.
10. Close the safety cover, and depress [RESET] SW then select #Continue$ button on the screen.
11. Depress [TEST STOP] SW.
12. Set the Machine Reference Point at the Master mode again.
13. After replaced the probe-4 block, be sure to readjust the Camera offset at the Master mode.
-
8/17/2019 takaya hardware
36/40
34
Replacing Probe lead wire (Probe-2 or 3)
When you need to replace the probe lead wire, be sure to perform as follows.
Pay attention to handle the tip of the probe not to be injured somebody.
Mark-tube
Cable lock
Mark-tube
Cable lock
Hex. bolt
Probe lead wire
Probe holder
Relay PCB
Screw
Bead to 45 degrees
Terminal
Fig. 35 Replacing probe lead wire
Name Type
Probe lead wire TAKAYA HS-433
1. Proceed #Tool$ ð$Self diagnosis$ ð$ Adjust probe$ on the system menu, then select either
Probe-2 or 3 and depress [TEST START] SW.
2. Open the cover.
3. Release the Hex bolts, and detach the probe holder. (Refer to Fig. 35)4. Pull out the probe lead wire straight.
5. Unscrew the screw on the Probe Relay PCB, and remove the probe lead wire.
6. Install a new probe lead wire to the Probe Relay PCB.
7. Wire the probe lead wire with the cable holder as per Fig. 35.
(NOTE) In this case, be sure that #Mark-tubes$ locates as per Fig. 35.
8. Connect the probe lead wire to the probe.
(NOTE) The probe must be tightly inserted so as the flange is in contact with the probe holder.
9. Install the probe holder on the Z-2 or 3 block and fasten it with the hexagon socket cap screw.
10. Close the cover, and depress [RESET] SW then press #Continue$ button.
11. Depress [TEST STOP] SW.
12. Set the Machine Reference Point at the Master mode again.
-
8/17/2019 takaya hardware
37/40
35
Printer
Self-diagnosis function is equipped with the printer as standard.
1. Turn off the printer by using the Power SW.
2. Turn on the Power SW again with pressing [Feed SW].
3. After above operations, the self-diagnosis function is available and it automatically starts
printing out.
(All letters can be output by read and black.)
Power SW
Feed SWAlarm lamp
Fig. 36 Printer
If the printer does not work correctly after executes the self-diagnosis, the following situations
happen.
l The self-diagnosis function cannot be used and the printer doesn t work.
l After alarm sounds, the self diagnosis function is suspended.(The ALARM lamp (3)
lights.)
l The letter is not printed out by two colors (Red/Black).
l Even though the letter can be printed out by two colors (Red/Black), but the letter is
not completely.
-
8/17/2019 takaya hardware
38/40
36
Feeder
The lubricating grease that is made from lithium soap, and is waterproof and heat-resident, is
applied to such mechanical parts as the ball screws and the slide rails before our delivery. So there
is no need to apply the grease within one year after being delivered. After this period, you should
check to see if the grease film is properly formed on the surface of both the ball crews and the slide
rails periodically (every half a year at least). When you found no grease film there, you should take
well care of applying the grease to these mechanical parts directly or through the grease nipple.
IMPORTANT
If the grease film was exhausted, the frictional resistance increases. As the result the test speed
badly goes down, along with the probe contact accuracy.
The lubricating grease can be applied as mentioned below.
Parts Position Method
Ball screw (Refer to Fig. 37 A) direct
Rail (Refer to Fig. 37 B) directLM guide unit
Rotation holder (Refer to Fig. 37 C) through greaseX arm
LM guide Rail (Refer to Fig. 37 D) direct
Ball screw (Refer to Fig. 37 E) direct
Rail (Refer to Fig. 37") directY arm LM guide unit
Rotation holder (Refer to Fig. 37G) through grease
Z arm LM guide Rail (Refer to Fig. 37 H) direct
P.C.B Unit LM guide Rail (Refer to Fig.38 I) direct
LM guide (traveling) Rail (Refer to Fig.38 J) direct
LM guide (up/down) Rail (Refer to Fig.38 K) directSupport unit
Gear/ LM guide Gear/Rail (Refer to Fig.38 L) direct
Stopper unit LM guide (traveling) Rail (Refer to Fig.38 M) direct
-
8/17/2019 takaya hardware
39/40
37
Location of grease nipple -1
A B A
B
C
D
E E E E
F F F F
G G
X1 arm
BA A CY1 arm Y2 arm Y3 arm Y4 arm
B
X4 arm
X2 arm
X3 arm
Z2 ,3 a rmZ1 ,4 a rm
H H
Fig. 37 Lubrication map 1
-
8/17/2019 takaya hardware
40/40
Location of grease nipple "2
Viewed from arrow-B
Viewed from arrow-A
BA
I
K L
M
LL K K
J J
I
KJL
Fig. 38 Lubrication map 2