Surgical instruments Catalogue
-
Upload
dio-implants -
Category
Documents
-
view
228 -
download
1
description
Transcript of Surgical instruments Catalogue
Sinus-T
Osteotome Technique by TR Sinus System
STD 3020LSTD 3525LSTD 4030L
SRD 3008L
ST 4006 ST 4007 ST 4008 ST 4009 ST 4010 ST 4011
SDG 2515RA 8011
SRD 3508LSRD 4008L
Trephine Drill Round Drill
Drill Stopper
RatchetAdapter
Depth Gauge
Ø3.0 / Ø3.5 / Ø4.0Ø3.0 / Ø3.5 / Ø4.0
Ø2.0 / Ø2.5 / Ø3.0
Ø2.5
Ø5.045
3
86
7
29
15
29
15
O-Ring
“TR Sinus System” contains trephine drills and round drills, will provide easy & safe sinus membrane lifting.
To overcome the short of bone length on maxillary molar area in case of sinus membrane lifting, osteotome technique is useful to obtain su�cient bone for placing implant.
Surg
ical In
strum
ents
Surgical Instruments Sinus-T
72
Order Code: SNT 01
Drilling (Trephine Drill)
Drilling (Round Drill)
Insertion (Fixture)
Push (Osteotome)
Sinus Protocol
1
2
3
4
Drilling Drilling InsertionPush
7mm
0.5~1mm
15~20rpm
1,000~1,500rpm
15~20rpm
1 2 3 4
Precise X-Ray reading is very importantto approach closely under the sinus �oor with trephine drill.
Choose correct length of stopper and put it on the trephine drill and then drill 0.5~1mm under the sinus �oor.(Drill Speed : 1,000 ~ 1,500 RPM)
After malleting at correct position of osteotome, do microfracture safely on sinus �oor.
Enter the osteotomy site with round drill, thereby lift the sinus membrane slowly.(Drill Speed : 15 ~ 20 RPM)
After checking the lift of the sinus membrane, place the �xture.
Putting stopper on the trephine drill reduce the risk of perforation of sinus �oor accidentally during the surgery.
Mallet 2 or 3 times to obtain a microfracture on sinus �oor.If the several malletings do not work for the microfracture, do drilling again with trephine drill toward sinus �oor.(Drill Speed : 15 ~ 20 RPM)
Round drill minimizes the risk of perforation of sinus membrane with rounded shape.
Surgical Instruments
73
When Torx Abutment Screw Head was damaged, you can use 0.3 Slot Driver instead of 1.7 Torx Driver.
0.3 Slot Driver
Machine Ratchet
MD 0322 HD 0315A
1.7 Torx Abutment Screw
Screw Tapping
Remove Bar
Reverse Cutting Drill
0.3 Slot Driver
External Guide Internal Guide
Dr.SOS
Immediate Action For EmergencyThe Kit for the easy removal of Broken Abutment ScrewRegeneration of damaged internal threads in the �xture
Surg
ical In
strum
ents
Surgical Instruments Dr.SOS
74
Order Code: DRS 00
It is connected to �xture and guides through Remove Bar toenter and support broken abutment screw.
Mores Taper
Short
Branemark Connection Branemark Connection Branemark Connection2.4 Hex 2.7 Hex 3.4 Hex
Hex
Morse Taper 6° Morse Taper 8°
*Option: ITI Guide
Morse Taper 11°
EXT 24S
EXT 24L
EXT 27S
EXT 27L
EXT 34S
EXT34L
Long
Internal Guide (Internal Taper Connection)
INT 6S Torx
INT 6L
INT 8S Torx
INT 8L
INT 11S Hex
*HEX Type
INT 11L
External Hex
Torx
Non-Torx
*Narrow Fix
Morse Taper 5.7°
INT 5.7S
External Guide (External Connection)It is connected to �xture and guides through Remove Bar toenter and support broken abutment screw.
Surgical Instruments
75
Connect the Reverse Cutting Drill to low speed handpiece,with 1,000~2,000 RPM in reverse rotation.Then, repeat drilling up and down to grind and changethe top of the broken screw.
*CAUTION): Be careful not to break it with overheating.
When internal threads in the �xture were damaged, use Screw Tappingto regenerate �xture internal threads.
ExternalFixture
Screw
Internal FixtureNon Submerged
Internal FixtureSubmerged SM
NarrowSM Regular
SM WideExtra Wide
(PSI, IFI)Internal Fixture
Narrow RegularT-Wide
Wide
SGT 1620A SGT1820A
RBEX 24S RBEX 24L IRBS IRBL RCD 36
RBH 10
External 2.4 HexØ1.25
External 2.7/3.4 HexInternal 6°, 8°, 11°
Screw Tapping
Ø1.6 Ø1.8 Ø2.0SGT2020A Ø2.5SGT2520A
Handle
DTW 0060
Torque Wrench
Remove BarUnder 20 RPM in reverse rotation
Surg
ical In
strum
ents
Reverse Cutting Drill
76
Ready
Guide Fixing
A) Remove Bar
B) Remove Bar + Handpiece
Explore
Screw Tapping
Tip)
After cleaning thoroughly broken screw area,repeat inserting oral-use oil to let the broken screw out.
Locate a proper guide from the "Dr.SOS Kit" onto a �xtureinside mouth and support it securely to make it vertical.
Connect Remove Bar to the Handle, reverse it by hand to removethe Broken Abutment Screw, and be careful not to drop o� theproduct into the mouth.
After connecting the Remove Bar into low speed handpiece,it is also possible to remove it with reverse rotation of low RPM.
Use explore to completely remove it after reaching certain height.
When �xture internal threads were damaged,use Screw Tapping to regenerate �xture internal threadsand connect Abutment.
TIP) When pulling, hand will feel it touches.If no touch on hands, lean guide and bur leftand right until getting feel.
*You can use ratchet to support the guide when it's hard to support the guide by hand.
*Con�rm the level of line on the Remove Bar.
*less than 25 RPM. *Remove Bar and drill extension connection enables a longer use.
The Kit for the easy removal of Broken Abutment ScrewRegeneration of dameged internal threads in the �xture
Immediate Action For Emergency, Dr. SOS
Surgical Instruments
77
0 1 2 3
4 5 6
7
8
9
10
11
Description
For Ratchet Wrench
For Contra Angle
Code Length Color CodeApical/Neck(Ø)
MSD 1218M 1.2 161.2mm Initial Drill
Torque Wrench DTW 0060
MHDC 2525Driver(long)
MHDC 2520Driver(short)
Driver(short) MHDR 2513
Driver(long) MHDR 2518
Open Wrench OW 002
1.0 / 1.6 17BEP 17161.0/1.6mm Bone Expander
1.3 / 2.3 171.3/2.3mm Bone Expander BEP 1723
1.7 / 3.1 171.7/3.1mm Bone Expander BEP 1731
2.4 / 3.7 172.4/3.7mm Bone Expander BEP 1737
3.0 / 4.4 173.0/4.4mm Bone Expander BEP 1744
3.5 / 5.1 173.5/5.1mm Bone Expander BEP 1751
0
1
2
3
4
5
6
7
8
9
10
11
(mm)
Bone Expander SetSurgical Instruments Bone Expander
78
Osteotomes have been utilized in a variety of techniques designed to prepare implant sites in maxilla, to elevate the �oor of the maxillary sinus and to expand the atrophic edentulous ridge. Despite its e�ectiveness, the surgical mallet is not well tolerated by patients.Using a drill to prepare implant sites in maxilla and mandibular also have a risk of maxillary perforation and penetration into the nerve mandibularis.As safe alternative and complementary instruments to the osteotomes and drills,Bone Expander can be used for preparation of implant sites, atraumatic ridge expansion and condensing of bone.
Order Code: BEP(M) 00
IFI55xxIFI48xx IFI65xx SFR45xx IFI65xx SFW53xxIFI40xxIFI35xx SFN38xx
In order to give depth and direction to desired surgical area, 1.2mm drill is used.(800 rpmn 1200rpm is the capable engine speed)
*Option: MSD 2016(Pic 3->)
To expand the implantation site, 1.6mm Bone Expander(Color Code: Yellow) is used (Check Laser Marking)
Use 3.1mm Bone Expander (Color Code: Yellow) to expand the desired surgical area.(Check Laser Marking)
Use 4.4mm Bone Expander (Color Code: Green) to expand the desired surgical area.(Check Laser Marking)
Use 3.7mm Bone Expander (Color Code: Red) to expand the desired surgical area.(Check Laser Marking)
Continue by using 5.1mm Bone Expander to expand the desired surgical area.(Check Laser Marking)Hardening the Bone has a risk of Ischemia which makes the Doctor to checkwhether su�cient bleeding is occurring.Give Scrape to the desired surgical area to increase bleeding - Bleeding help Osseo Integration.
Implant SFW 5312 to the expanded hole by Bone Expander
Use 2.3 Bone Expander (Color Code: Yellow) to expand the desired surgical area.(Check Laser Marking)
*Connect the Bone Expander with Adapter. Then �nish with Ratchet Wrench.
0 1 2 3 4 5 6
Pic 1
Pic 2
Pic 3
Pic 4
Pic 5
Pic 6
Pic 7
Pic 3Pic 2Pic 1 Pic 4 Pic 5 Pic 6 Pic 7
ProtocolDIO Implant Bone Expander hardens Bone density to erase di�culties caused fromOsseointegration of soft or Poor Bone.(This procedure shows the method of Implantation of SFW 5312)
79
Sinus Lift Kit
Custom designed kit comprised of 6 tools used in sinus procedures
Surg
ical In
strum
ents
Surgical Instruments Sinus Lift Kit
80
DSNK 02 Membrane Elevator DSNK 02 Membrane Elevator
DSNK 05 Membrane Elevator DSNK 05 Membrane Elevator
DSNK 04 Bone Graft PackerDSNK 04 Bone Graft Packer
DSNK 03 Membrane Elevator DSNK 03 Membrane Elevator
DSNK 01 Membrane Detach ElevatorDSNK 01 Membrane Detach Elevator
DescriptionCodeSinus Elevator & Membrane Lift Tips
Order Code: SLK 01
Osteotome KitSurgical Instruments Osteotome Kit
81
Order Code: OST 01
SM/ IFI/ FTN
Initial Drill
Compaction
Compaction
Compaction
Compaction
FTN 33xxB
SFN 38xx/ IFI 35xxPM
SFR 45xx/ IFI 40xxM
SFW 53xx/ IFI 48xxM/ FTN 50xxB
OSL 2028
OSL 2130
OSL 2133
OSL 3039
OSL 3546
Ø2.0
Ø2.8
Ø2.1
Ø3.0
Ø2.1
Ø3.3
Ø3.0
Ø3.9
Ø3.5
Ø4.6
Sinus ElevationOSE 2228Ø2.2
Ø2.80
1
2
3
4
5
Code ApplicationDiameter(L) Image
Narrow Regular Wide
Ø4.95 Ø5.65 Ø5.9
Protocol
Size & Code
Guide Drilling Initial Drilling Bone PlaningDrill through the cortical bone atthe selected site for the implantplacement.
Select an appropriate drill stopperaccording to the desired depth ofimplant placement.
Insert its guide part into a drilledosteotomy site and removethe uneven bone.
*Handpiece speed is recommended to be in a range between 400~600 rpm.
1 2 3
BP 5010 BP 5710 BP 5910
30
10
5.5
Bone Planer SetBone Planers are used to �atten the uneven bone during osteotomy reparation
82