Surface Properties of Bio Materials

54
Surface Properties of Biomaterials

Transcript of Surface Properties of Bio Materials

Page 1: Surface Properties of Bio Materials

Surface Properties of Biomaterials

Page 2: Surface Properties of Bio Materials

Introduction1.1 Concepts in Surface Chemistry and Biology in this presentation I will focus on surface properties of biomaterials. This includes a description of both physicochemical and biological means to modify solid surfaces, as well as processing techniques to form patterned biomaterial surfaces.

• The atoms at the surface are not bonded on all sides to other atoms.

• Extra energy associated with this region due to unfilled valence shells.

• The surface tension (y) is unstable• Adsorption is the adhesion of molecules to a solid surface.

a. The surface of a material can be considered a type of planar defect.

1.1.1 Protein Adsorption and Biocompatibility

Page 3: Surface Properties of Bio Materials

1.1.2 Surface Properties Governing Protein Adsorption There are two important surface properties have the largest effect on the favourability of adsorption. The properties below affect the thermodynamics of protein adsorption to biomaterials

• Surface hydrophobicity It describes how it responds to the presence of water To quantify hydrophobicity, a biomaterial surface is subjected to contact angle analysis.

• Surface charge The effects of surface charge do not completely separate from the effects of hydrophobicity

A significant surface charge can have the additional effect of attracting or repelling charged areas of proteins.

Surface charge occurs via dissociation of ionisable surface groups or though specific adsorption of ions from the solution

Page 4: Surface Properties of Bio Materials

The physical characteristics of the biomaterial surface

Steric concerns

For example adding (PEG) to a biomaterial surface will result in a reduce in protein adsorption. This because of quick moving of bulky chain, this allows the proteins to adsorb to them and they form a type of wall.

Surface roughness A surface with a high degree of roughness may promote protein

adsorption in certain areas by physically "trapping" the proteins in the valleys on the surface.

Note:the hydrophobicity, charge, steric hindrance and/or roughness of a biomaterial surface can be altered during formation or processing to change the protein adsorption profile of the final material.

Page 5: Surface Properties of Bio Materials

1.2 Physicochemical Surface Modification Techniques

1.2.1 Introduction to Surface Modification TechniquesThere are a wide variety of surface modification

techniques designed to treat all types of biomaterials. These include naturally based and synthetic polymers, metals, and ceramics .

Techniques characteristics1. Thin (to minimize effects on bulk properties)2. Resistant to delamination3. Simple and robust (to promote commercialization)

1.2.2 Physicochemical Surface Coatings: Covalent Surface Coatings

Physicochemical surface treatments use two main principles to alter the surface composition of the sample Physical principles Chemical reactions

Page 6: Surface Properties of Bio Materials

The below table shows various Surface Modification Techniques and the Materials

Polymer

M e ta l

Ceramic

Glass N o n -co v a len t

Solvent coating

V

v

V

V L an g m u ir-B lo d g e tt film d ep o sitio n

^

^

V

V Surface-active additives

V

V

V

,/ V a p o r d e p o s it io n o f c a rb o n s a n d m e ta ls "

V

V

V

V Vapor deposition of parylene (/?-xylylene)

V

V

V

V C o v a len tly a tta ch ed co a tin g s

R ad ia tio n g ra ftin g (e le c tro n a c c e le ra to r an d g a m m a )

V

— Photografting (UV and visible sources)

V

V Plasma (gas discharge) (RF, microwave, acoustic)

V

V

V

V G a s -p h a se d e p o si tio n

• Io n b ea m sp u tte rin g

V

V

V

V • Chemical vapor deposition (CVD)

V

v

V • F la m e sp ra y d ep o sitio n

V

V

V C h e m ic a l g ra ftin g (e .g ., o zo n a tio n + g ra ftin g )

v

V

V

v

Silanization

V

V

V

v

Biological modification (biomolecule immobilization) V

•V

V

V M o d ific a tio n s o f th e o rig in a l su r fac e

Io n b e a m e tch in g (e .g ., a rg o n , x e n o n )

V

V

V

V Io n b e a m im p la n ta t io n (e .g ., n i tro g e n )

V

V

V P la s m a e tc h in g (e .g ., n itro g e n , a rg o n , o x y g e n , w a te r v a p o r)

V

V

V

V Corona discharge (in air)

V

V

v

V Ion exchange

/

V

V

V U V irrad ia tio n

V

V

v

V C h e m ic a l re ac tio n

• N o n sp e c ific o x id a tio n (e .g ., o zo n e )

V

V

V

V • F u n c tio n a l g ro u p m o d ific a tio n s (o x id a t io n , red u c tio n )

V

— • A d d itio n re a c t io n s (e .g ., a c e ty la t io n , c h lo r in a tio n )

V

— C o n v e rs io n co a tin g s (p h o sp h a tin g , a n o d iz a tio n )

V

— M e ch an ica l ro u g h en in g an d p o lish in g

V

V

V

V

Page 7: Surface Properties of Bio Materials

The table above showed only Physicochemical modifications that only using coating methods some of these methods, which include plasma discharge, chemical or physical vapor deposition, radiation or photografting, and self-assembled monolayers, have been used to create biological coatings as well.

1.2.2.1 plasma treatment Includes several methods of physicochemical surface

modification. some of methods result in surface coatings, while others

do not. Exposure to a plasma environment can be used as a pre-

treatment. Refers to an assembly of species in an

atomically/molecularly dissociated gaseous environment.

The advantages of Plasma discharge treatments 1. Conformal.2. Free of voids/pinhole defects.3. Easily prepared.4. Sterile when removed from the reactor.5. Produce a low amount of leachable substances.6. Demonstrate good adhesion to substrate.7. Allow unique film chemistries to be produced.8. Can be characterized relatively easily.

Page 8: Surface Properties of Bio Materials

The disadvantages of plasma-based treatments

1. Can be ill defined within the reactor.2. Expensive equipment.3. Uniform reaction may be difficult (in long and narrow pores).4. Needs care to prevent contamination during or after processing.1.2.2.2 Chemical Vapor Deposition (CVD)

surface treatment in which a mixture of gases is exposed to a sample at a high temperature.

CVD techniques are most commonly used in biomaterials applications

PVD techniques employ successfully for applications such as orthopedic implants, surgical tools, and orthodontic appliances

results in surface coating via deposition of atoms generated through physical processes onto the sample

Uses mainly to increase wear resistance of metallic hip implants.

includes sputtering and thermal evaporation.

1.2.2.3 Physical Vapor Deposition (PVD)

Page 9: Surface Properties of Bio Materials

1.2.2.4 Radiation Grafting/Photografting radiation grafting and photo grafting form well-bonded

surface coatings via similar mechanisms. Radiation grafting one type of radiation grafting (mutual irradiation), the

biomaterial substrate (I) placed in a monomer solution and then irradiated by electrons or gamma rays to (I) to produce the polymerized coating

Another type of radiation grafting occurs in air rather than in an inert environment

Photografting the radiation is UV or visible light. A number of photoresponsive chemical moieties develop to

facilitate this type of surface modification. Two methods involve phenyl azide or benzophenone

chemistry. excites by exposure to the light and form free radicals or

other reactive species. activated molecules participate in reactions at the

substrate surface resulting in covalent linkage of the coating to the underlying biomaterial.

Page 10: Surface Properties of Bio Materials

1.2.2.5 Self-Assembled Monolayers

Surface treatment using self-assembled mono-layers (SAMs).

The molecules composing the coating are designed so that it is thermodynamically favourable for them to align on and form covalent bonds with the surface of the biomaterial.

no specialized equipment is required. the modification can be carried out at room temperature

and under normal atmospheric pressure. have both hydrophilic (polar) and hydrophobic (nonpolar)

areas. Advantages of Self-Assemled monolayers Ease of their formation Stability of the coating is greater. SAMs can also be used to form molecularly smooth

surfaces.

Page 11: Surface Properties of Bio Materials

1.2.3 Physicochemical Surface Coatings: Non-Covalent Surface Coatings

1.2.3.1 Solution Coatings

Surface modification procedures producing not covalently bound coatings

the substrate is dipped in a solution containing left to dry as the solvent evaporates, the coating is deposited on the

surface.1.2.3.2 Langmuir-Blodgett Films the molecules are amphiphilic with two regions(a

hydrophilic head and a hydrophobic tail). a piece of equipment are used called a Langmuir trough. molecules can be transferred to a biomaterial surface the substrate is placed into aqueous media and the

amphiphilic molecules are added then the polar head groups interact with the water and the

remainder of the molecule rests in the air At the critical area, the surface pressure increases. By maintaining a surface pressure corresponding to the

critical area as the material to be coated is slowly removed from the trough, a homogenous, well-orientated coating can be deposited.

Page 12: Surface Properties of Bio Materials

Process of Langmuir-Blodgett Films

Solid phase

Area per molecule

------------------------------------------------------------------------------------------------

----------------------------------------------------------------------------------------------------

(A)

(B)

(C)

Gaseous phase

Area per molecule

Page 13: Surface Properties of Bio Materials

1.2.3.3 Surface-Modifying Additives (SMAs) atoms or molecules added in material bulk, will

spontaneously rise to the surface. Depends on the reduction of the surface free energy. SMA treatment is not a post-fabrication procedure. a part of the formation/synthesis of the biomaterial. Examples of SMAs for metals include copper in gold

alloys. Easy to be designed

Page 14: Surface Properties of Bio Materials

1.2.4 Physicochemical Surface Modification Methods with No Overcoat

The below are physicochemical modification methods that alter the surface properties of biomaterials without the formation of a separate coating.

1.2.4.1 Ion Beam Implantation accelerate ions with high energies are directed at the

surface of a biomaterial. employs for metals and ceramics and some with polymers.

Schematic representation of ion beam implantation. Ions with high energy states are projected at the surface of the material at high speed, where they break the surface. Once within the material, some atoms from the material are ejected, while others collide with each other and create vacancies and changes within the crystal structure. Using different ions at various concentrations can alter the degree of modification. (Adapted with permission from

Page 15: Surface Properties of Bio Materials

1.2.4.2 Plasma Treatment plasma discharge processes can result in deposition of a

coating. can alter the surface properties through etching and

cleaning processes. the plasma is created using an inert gas and the energy

of the plasma species causes ablation as the atoms/ions strike the biomaterial surface.

1.2.4.3 Conversion Coatings

It is not an overlay coating It is a modification of atoms at the surface of a metallic

implant to form an oxide layer.

Page 16: Surface Properties of Bio Materials

1.2.4.4 Bioactive Glasses

It is a type of ceramic used in orthopedics. There are many classes of biological response to these

glasses to complete material dissolution. depending on the ratio of CaO, Na2O, and SiC>2 used in the

fabrication. The 1% index included on the diagram is a measurement of the bioactivity of these materials.

Certain compositions of bioactive glasses cause coating of the surfaces with layers of CaCM^Oj and SiO2 under physiological conditions.

A phase diagram depicting the range of compositions for bioactive glasses, often used as orthopedic biomaterials. The IB index is a measurement of the bioactivity of these materials, with a higher number indicating that the material integrates with the surrounding bone more quickly. (Adapted with permission from [4].)

Page 17: Surface Properties of Bio Materials

1.2.5 Laser Methods for Surface Modification focusing a high-powered light beam at the sample. a large amount of energy can be quickly deposited in a

small area, which facilitates reactions at the surface. It must be considered whether a pulsed or continuous

wave beam would be more appropriate and the effects of possible surface heating by the laser.

Much of this depends on the degree of laser energy absorption by the substrate and the amount of interfacial reflection and scattering.

• The advantages of laser methods

performed at atmospheric conditions and with a wide range of very precise wavelengths.

there is specific control of reaction time as well as the spatial location of the excitation beam.

Lasers also allow the use of a combination of heat and light-induced excitation to initiate the desired reaction.

Page 18: Surface Properties of Bio Materials

1.3 Biological Surface Modification Techniques involve attachment of biologically active molecules to a substrate

through a variety of means. The attached molecules are then free to interact with specific

target areas on cells . the molecule of interest remain attached while maintaining its

biological activity. paid attention to the orientation and rotational ability of individual

molecules after coating. A list of the types of biomolecules used in surface modifications and

their potential applications is found in the Table below.

Applications

Enzymes Bioreactors (industrial, bio-medical) Bioseparations Biosensors Diagnostic assays Biocompatible surfaces

Antibodies, peptides, and other Biosensors affinity molecules Diagnostic assays

Affinity separations Targeted drug delivery Cell culture

Drugs Thrombo-resistant surfaces Drug delivery systems

Lipids Thrombo-resistant surfaces Albuminated surfaces

Nucleic acid derivatives and DNA probes nucleotides Gene therapy

Page 19: Surface Properties of Bio Materials

1.3.1 Covalent Biological Coatings impart additional stability , so preferred in many

applications. required the presence of a reactive substrate surface,. often

containing hydroxyl (OH), carboxyl (COOH), or amine (NH2) groups.

If these groups are not found can be modified (via plasma discharge).

the molecule can be bound to the substrate directly or by a spacer arm.

Below Figure depicts several schemes for covalent attachment of biomolecules

7.3 B io logical Surface M odification Techniques 245

a) Pre-activation of support

1] Binding Support agent

b) Direct coupling to suppport

Biomolecule

• 7T

c) Pre-activation of biomolecule

If

d) Conjugation followed by copolymerization

Monomer conjugated biomolecule

Activated monomer

e) Direct attachment to pre-activated polymer, gel or graft copolymer

2 ) G e l

Page 20: Surface Properties of Bio Materials

1.3.2 Non-Covalent Biological Coatings

usually involves adsorption of the biomolecule to the biomaterial, and then possible crosslinking to improve the coating stability.

a given surface is often dictated by hydrophobic and electrostatic interactions. TAn example of this method is one coating heparin

on a hydrophobic surface is the addition of a hydrophobic region to the heparin.

interaction of the hydrophobic regions of the biomolecule and the substrate in an aqueous environment will result in extension of the heparin portion away from the surface.

(b )

Hydrophobic Hydrophobia material / moiety

I

Positively charged material

Page 21: Surface Properties of Bio Materials

1.3.3 Immobilized Enzymes

a subclass of proteins that act to promote specific chemical reactions involving other biomolecules.

Regardless of the application, because the function of the device depends on the action of the enzyme, the bioactivity of this molecule is of utmost importance.

the geometry of the carrier is crucial to allow a sufficiently large surface area for enzyme contact.

activity of the enzyme and to the availability of substrate continue to be optimized to produce more efficient biomedical devices, such as glucose sensors.

1.4 Surface Properties and Degradation

the surface treatment is designed as a method to decrease unwanted degradation (corrosion) of the underlying material.

control degradation of the bulk material is desired, such as for biodegradable polymers,.

there are Sometimes significant deleterious effects on the surface coating.

Page 22: Surface Properties of Bio Materials

1.5 Patterning Techniques for Surfaces

It is a term that encompasses several methods developed to alter the surface properties of biomaterials in a controlled manner.

can be used with a variety of surface-active molecules, and common substrates include both metals and polymers.

most widely used patterning techniques are microcontact printing and microfluidics.

S u rfa c e su b s tra te p a t te rn in g c a n b e u se d to a l te r th e su r fa c e p ro p e rtie s o f a m a te r ia l in c o n tro lle d , w e ll -d e fin e d a re a s , a s se e n in th e c irc u la r p a tte rn in th e p h o to . T h e in se t p resen ts a lo w er m agn ifica tio n v ie w o f th e sa m e su rfac e . (R ep rin ted w ith p e rm issio n fro m [7 ] .)

Page 23: Surface Properties of Bio Materials

1.6 Techniques: Introduction to Surface Characterization to determine the quality of the surface treatments.

to provide information about the extent of protein adsorption to a material.

Characterization Methods for Biomaterial Surfaces

M e th o d

Principle

D ep th A nalyzed

S p atia l R e so lu tio n

Analytical Sensitivity Costc

C o n ta c t an g les

Liquid wetting of surfaces is used

3 -2 0 A

1 mm

L o w o r h ig h d ep e n d in g $

to estimate the energy of surfaces

o n th e c h e m is try ESCA (XPS)

X -ra y s in d u c e th e e m iss io n o f

10-250 A

10-150 ixm

0 .1 a to m % $ $ $

e le c tro n s o f ch ara c te ris tic en erg y

A u g er e lec tro n

A focused electron beam stimulates

5 0 -1 0 0 A

100 A

0.1 atom % . $$$ sp ectro scop y^

the emission of Auger electrons

SIMS

Io n b o m b a rd m e n t sp u t te rs

l O A - 1 |j,m *

1 0 0 A

Very high $$$

se co n d ary io n s fro m th e su r fa c e

FTIR-ATR

IR rad ia tio n is ad so rb e d an d e x c ite s

1 -5 (x m

10 ujn

1 m o l % $ $

m o le cu la r v ib ra tio n s

STM

M e a su re m e n t o f th e q u a n tu m

5A

1 A

Single atoms $$

tu n n e lin g cu rren t b e tw e e n a m e ta l

tip and a conductive surface

SEM

Secondary electron emission

5 A

40 A, typically

H ig h , b u t n o t $ $

induced by a focused electron

q u a n tita tiv e

beam is spatially imaged

Page 24: Surface Properties of Bio Materials

1.6.1 Contact Angle Analysis1.6.1.1 Basic Principles

often used to provide overall information about the hydrophobicity of a surface.

The surface free energy or surface tension (y) of a material can be defined thermodynamically as the work of making a unit area of new surface.

a system with three interfaces (and corresponding y values) is created

M eth o d

Principle

D ep th A nalyzed

S patial R e so lu tio n

Analytical Sensitivity Costc

C o n ta c t an g les

Liquid wetting of surfaces is used

3 -2 0 A

1 mm

L o w o r h ig h d ep e n d in g $

to estimate the energy of surfaces

o n th e c h e m is try ESCA (XPS)

X -ra ys in d u c e th e e m issio n o f

10-250 A

10-150 ixm

0 .1 a to m % $ $ $

e le c tro n s o f ch ara c te ris tic en erg y

A u g er e lec tro n

A focused electron beam stimulates

5 0 -1 0 0 A

100 A

0.1 atom % . $$$ spectro scop y^

the emission of Auger electrons

SIMS

Io n b o m b a rd m e n t sp u tte rs

l O A - 1 |j,m *

1 0 0 A

Very high $$$

se co n d ary io n s fro m th e su r fa c e

FTIR-ATR

IR rad ia tio n is ad so rb e d an d e x c ite s

1 -5 (x m

10 ujn

1 m o l % $ $

m o le cu la r v ib ra tio n s

STM

M e a su re m e n t o f th e q u a n tu m

5A

1 A

Single atoms $$

tu n n e lin g cu rr en t b e tw e e n a m e ta l

tip and a conductive surface

SEM

Secondary electron emission

5 A

40 A, typically

H ig h , b u t n o t $ $

induced by a focused electron

q u a n tita t iv e

beam is spatially imaged

Characterization Methods for Biomaterial Surfaces

Page 25: Surface Properties of Bio Materials

Mostly ,the liquid chosen for testing of biomedical materials is water. the water droplet assembled to assume a particular shape. by accurately measuring the angle between the drop and the solid surface (the contact angle, 0), the surface tension can be calculated using certain equation which represents a force balance between the horizontal components of three surface tensions (Young's equation. Jsv ~ JSL ~ JLV cos 0 = 0

S O L ID

(a)

(b)

Tsv

JLV

fsv

(a) Schematic of contact angle testing. Around the water droplet, there are three important interfaces: liquid vapor (TLV)I solid-liquid (JSL), and solid vapor (ysv)- (Adapted with permission from [11].)(b) Change in wettability via surface modification. The water droplet spreads more on the modified surface because the modification decreases the surface tension of the liquid/solid interface, thus reducing the contact angle as calculated using Young's equation.

Page 26: Surface Properties of Bio Materials

EXAMPLE:

Guided by the following image, derive Young's equation for surface tension:

Solution: The forces in the diagram must balance in the plane of the solid material for the droplet to be in static equilibrium. As a result, the sum of y$L and the projection of ji\r on the plane of the solid must equal y^v f°r the net force in the plane of the material to be 0. The projection of yLV on the plane of the solid can be determined by drawing a perpendicular from the material surface to the end of the yLV vector to form a right triangle. The projection (p) of JLV on the plane of the material can then be calculated as follows: adjacentcos = -—————— hypotenusepcos d = —— JLVP = JLV cos 0 Consequently, the vectors can be balanced as follows to yield Young's equation:TSL + TLV cos 6 = jsv Jsv ~ VSL ~ JLV cos 0 = 0

Page 27: Surface Properties of Bio Materials

1.6.1.2 Instrumentation the output of contact angle analysis is a single number (9 or

yc), rather than a plot. The instrumentation for these measurements is fairly

straight-forward. All types of contact angle instrumentation have these components in common:1. Holder for solid sample2. Holder for liquid3. Means to determine contact angle (may be automated)

Different experimental setups for determining the contact angle: (a) sessile drop, (b) cap-tive air bubble, (c) capillary rise method, and (d) Wil-helmy plate method. The cir-cles indicate where the contact angle is measured. (Adapted with permission from [10].)

Page 28: Surface Properties of Bio Materials

1.6.1.3 Information Provided

contact angle data technique cannot provide detailed information about the chemical composition of the surface, so it is often used as a first step in surface characterization.

Dynamic contact angle measurements used to measure contact angle hysteresis.

water is slowly added to an area of the surface with a syringe and the advancing contact angle is measured.

Water removed via the same mechanism and the receding contact angle is recorded.

Page 29: Surface Properties of Bio Materials

1.6.2 Light Microscopy1.6.2.1 Basic Principles

simple technique that is used as a first approach to gain primarily qualitative information about surface topography.

one type of light microscope, the compound microscope, a white light source is projected through the sample.

The light is reflected by an objective lenses. A diagram of the light path in such a microscope is found in the figure below. For opaque samples, the light source can be located above rather than below the

sample. Can not distinguish the resolution that are less than 0.2 | xm.

Eyepiece 1— Eyepiece Lens

Magnified Image

Objective Slide with Specimen

Light Source

Virtual Image

Light Source (a)

(a) Gross image of a compound microscope. On a compound microscope, the light source is placed below the specimen and the specimen is viewed from above, (b) Light path of a compound microscope. The objective forms a magnified image of the object that is larger than the original object. This image is then magnified many times by the eyepiece to form the large (inverted) virtual image.

Page 30: Surface Properties of Bio Materials

Light microscopy images, like that found in Fig below are widespread and have a variety of uses.

there are four basic components to a light microscope:

1.6.2.2 Instrumentation

1. Source—produces white light.2. Lenses—glass lenses focus light beam and/or magnify image of sample.3. Sample stage—holds sample securely.4. Detector (camera or human eye)—views and captures resulting image.

Sample of a light microscopy image, in this case, a fibroblast attached to a substrate.

Page 31: Surface Properties of Bio Materials

Block diagram of the compo-nents of a light microscope, which include the source, lens-es, sample stage, and detector (camera or human eye).

Surface topographical ridges as seen on a polymer sur¬ace. The sample is viewed via a particular type of light microscopy called fluorescence microscopy. Scale bar is 200 |xm.

Page 32: Surface Properties of Bio Materials

1.6.2.3 Information Provided Light microscopy is used exclusively for imaging, and provides in

itself only qualitative assessment. Images can be analyzed using specialized software to obtain

semiquantitative measures of certain colors. both the biomaterial and surrounding tissue after in vivo

implantation, are taken, stained, and viewed to determine the extent of the inflammatory response to the implant.

vacuum is not required to view samples. it is the first approach to imaging, particularly for

sample surface features, due to its limited spatial resolution.

The advantages of Light microscopy

Page 33: Surface Properties of Bio Materials

1.6.3 Electron Spectroscopy for Chemical Analysis (ESCA) or X-ray Photoelectron Spectroscopy (XPS)

1.6.3.1 Basic Principles

X-rays are a type of very high energy electromagnetic radiation used as a source in ( XPS) analysis of material surfaces.

X-ray absorption causes the removal of an electron from one of the innermost atomic orbitals.

The kinetic energy of the emitted electron is then recorded.

From the kinetic energy (E^), the binding energy (E^) of the electron can be calculated as Eh = hv —

where v is the frequency h is the Planck's constant (6.6 X 1CT34 J-s).

Page 34: Surface Properties of Bio Materials

Summary of the Differences Between X-ray

Techniques

Interaction between X-rays and electrons of the sample material causes X-ray scattering. The angle of scattering gives information about the crystal structure of the material.

Exposure of a sample to X-rays causes the removal of core electrons with a certain kinetic energy. From this, the binding energy, which gives information about the chemical composition of the material, can be calculated. The kinetic energy of ejected Auger electrons maybe detected as an artifact.

Exposure of the sample to an electron beam or X-ray causes the removal of a core electron and formation of an ion. The ion can return to its ground state as one of the outer electrons falls into the vacancy, and, in the process, X-rays of a certain wavelength that is characteristic for that atom are emitted. It is often used in biomaterial analysis as an addition to electron microscopy to determine the composition of a sample that has been imaged.

Exposure of the sample to an electron beam or X-ray causes the removal of a core electron and formation of an ion. The ion can return to its ground state as one of the outer electrons falls into the vacancy, and, in the process, a second electron (Auger electron) with a kinetic energy characteristic for that atom is emitted. Auger emission and X-ray fluorescence are competing events, with atoms of higher atomic number favoring relaxation through X-ray fluorescence.

X -ra y d iffra c tio n

Electron sp ec tro sco p y fo r chem ica l analysis (E S C A )

X-ray fluorescence

Auger electron spectroscopy

Page 35: Surface Properties of Bio Materials

1.6.3.2 Instrumentation

ESCA spectra are generally plotted as electron count (y-axis) as a function of binding energy (x-axis).

there are four basic components to an electron spectrometer:

Source—produces X-rays with known wavelength. Electron analyzer—uses an electrostatic field to

separate electrons based on kinetic energy2. Detector—converts impact by separated electrons into

an electrical signal. 4Processor (computer)—translates the signal from the

detector into the appropriate spectrum.

Page 36: Surface Properties of Bio Materials

1.6.3.3 Information Provided

ESCA methods are extremely sensitive can detect all elements except hydrogen and helium. ESCA spectrum also provides information about neighbouring

atoms to which an element is bonded.

1.6.4 Attenuated Total Internal Reflectance Fourier Transform—Infrared Spectroscopy (ATR-FTIR)

1.6.4.1 Basic Principles

ATR-FTIR is an analysis technique that has developed to probe particular parameters at the material surface.

The major characteristic of ATR-FTIR is the addition of a specialized probe made of a high-refractive index crystal.

When a beam of electromagnetic radiation passes from a medium that is more dense to one that is less dense, reflection occurs. Upon reflection, the beam penetrates a small distance into the less dense medium and this penetrating radiation is called an evanescent wave.

then the sample may absorb the evanescent beam due to the vibration frequency of bonds found within the material.

Page 37: Surface Properties of Bio Materials

mainly provides information about the chemical structure of the material.

The signal can be achieved from relatively deep within the sample by surface analysis standards (1-5 u,m).

3 Sample introduction nd preparation chamber

X-ray source

Electrostatic hemispherical

electron energy analyzer

Acceptance lens Retardation lens

Schematic of ESCA equipment. First, the sample is bombarded with X-rays. The resulting emitted electrons then enter the analyzer chamber. Because of the difference in voltage between the two walls and the geometry of the analyzer, only electrons with a certain kinetic energy can be collected by the detector, with the remaining electrons striking non-detectable areas. By altering the voltage difference between the walls in a controlled manner, the electrostatic field is altered to permit the detector to record the amount of electrons having various kinetic energies. At the end of the voltage sweep, the entire spectrum is plotted for that sample.

Page 38: Surface Properties of Bio Materials

1.6.4.2 Instrumentation A typical IR spectrum of poly(dimethyl siloxane) Like other IR

spectra, ATR-FTIR data are generally plotted by transmittance (or absorbance) (y-axis) as a function of wavelength (or wavenum-ber) (x-axis).

The basic instrumentation for ATR-FTIR spectroscopy includes an IR source, an intetferometer, a detector and a processor.

Fourier-transform techniques are required for this application because they significantly increase the signal strength.

S ch e m atic o f th e lig h t p a th w ith in an A T R -F T IR p ro b e . In o rd e r to in c rea se th e lev e l o f ab so rp tio n , th e p ro b e is d es ig n ed so th a t, a t ce r ta in an g les o f in c id en t rad ia tio n , re f lec tio n o f th e IR b ea m is co m p le te . T h ere fo re , m u ltip le re f lec tio n p ro c esses an d m u ltip le a tten u a tio n s o ccu r as th e b e am p a sses c lo se to th e sam p le , in c reasin g th e signal.

High-Refractive Index Crystal

Source

IR Detector

Evanescent Wave

Page 39: Surface Properties of Bio Materials

Two types of ATR sample cells, one for solids and one for the solid/liquid.

Cells allowing for the presence of liquid have particular application in the study of protein adsorption via ATR-FTIR techniques.

ATR crystal ^ ^ ^ ^ m tM U $ M z & I M [Solid sample!

— — — p $ m f$ 8^ - (b )

Two types of ATR sample cells, (a) one for solids and (b) one for the solid/liquid interface.

Page 40: Surface Properties of Bio Materials

1.6.5 Secondary Ion Mass Spectrometry (SIMS)1.6.5.1 Basic Principles

SIMS is a derivation of a bulk analytical technique. SIMS does not involve the absorption of electromagnetic radiation. A minor difference in the surface version is the method of sample

ionization, which involves the use of primary and secondary ions (thus giving SIMS its name).

Ionization begins when primary ions, such as O2+, Ar+, Xe+, or Cs+, are ejected from an ion gun and strike the sample surface. This causes the surface layer of atoms to be stripped off, both as neutral species and as ions. These emitted ions are called secondary ions and are drawn into the analyser for separation by mass in a similar manner to bulk mass spectroscopy.

SIMS is considered a surface analysis technique since the energy of the incident ions generates collision cascades only in the surface region of the sample, and, of these collisions, only those that occur in the outermost layers produce secondary ions with sufficient energy to escape from the surface.

Static SIMS uses a relatively low ion dose (less than 1013 ions/cm2) and induces minimal surface damage. In contrast,

dynamic SIMS bombards the sample with a much larger ion dose. In this case, so much material is sputtered that the surface erodes while the experiment is being performed. This allows depth profiling of specimens.

Page 41: Surface Properties of Bio Materials

1.6.5.2 Instrumentation

SIMS spectra are plotted as relative intensity (y-axis) as a function of mass (%-axis).

the instrumentation required for SIMS is very similar to that for bulk mass spectroscopy.

SIMS analysis is completed under vacuum. it includes four main components:

an ionization chamber containing the sample. a mass analyzer. an ion detector. a processor/computer to translate signals from the detector

into the appropriate spectrum. Like other types of mass spectroscopy.

lonization chamber

B lo c k d ia g ra m o f in s tru m e n ta tio n fo r S IM S , w h ich in c lu d e s fo u r m a in co m p o n e n ts : an io n iz a t io n ch a m b er c o n ta in in g th e sa m p le an d io n g u n , a m a ss a n a ly z e r (m a ss fi lte r ) , an io n d e te c to r, an d a p ro c esso r/co m p u te r to tran sla te s ig n a ls fro m th e d e te c to r in to th e ap p ro p ria te sp e c tru m . S IM S a n a ly s is is co n d u c te d u n d e r u l tra -h ig h v ac u u m (U H V ).

Page 42: Surface Properties of Bio Materials

1.6.5.3 Information Provided provides information about the structure and composition of

the outermost few A of both inorganic and organic materials. the accuracy of quantitative methods for this type of

spectroscopy is limited. With dynamic SIMS, composition as a function of depth can

be recorded.

1.6.6 Electron Microscopy: Transmission Electron Microscopy (TEM) and Scanning Electron Microscopy (SEM)1.6.6.1 Basic Principles accelerated electrons have wave-like properties. Electron microscopy techniques utilize this property of

electrons to form images, much like light microscopy. The most direct corollary to light microscopy is the

transmission electron microscope. TEM requires very thin samples (20-200 jxm thick) . It is less often used in biomaterials research than SEM. TEM is not strictly a surface analysis technique. SEM is based on a different mechanism of imaging.

Page 43: Surface Properties of Bio Materials

In SEM the surface of a sample is scanned with an electron beam. The electrons from the beam undergo elastic and inelastic

scattering as they collide with atoms in the sample. Elastic scattering results in alteration of the trajectory of the

electron, but not its energy. The atom then emits secondary electrons, Auger electrons.

SEM images are produced by recording the production of secondary electrons after an area is bombarded with the primary electron beam.

the intensity of these electrons is dependent on the surface topography of the sample.

SEM is considered a surface imaging technique

V

S O U R C E O F E L E C T R O N S

L I G H T S O U R C E

C O N D E N S E R L E N S

S P E C IM E N

C O N D E N S E R L E N S

S P E C I M E N

P R O JE C T O R L E N S

E Y E P IE C E L E N S

O B JE C T IV E L E N S

O B JE C T I V E L E N S

Comparison of Energy path in (a) light microscopy and (b) transmission electron mi-croscopy, which is designed very similarly to a compound microscope but uses magnetic lenses, as opposed to glass. (Adapted with permission from [13].)

Page 44: Surface Properties of Bio Materials

1.6.6.2 Instrumentation A typical SEM image of osteoblasts on a titanium mesh is found

in Fig below This image shows the interaction of cells and a biomaterial

scaffold for bone tissue engineering.

there are five basic components to a SEM:

1. Source—produces accelerated electrons.2. Lenses—magnetic coils focus electron beam and reduce spot size.3. Sample holder—holds sample securely. 4. Detector—records the spatial position of secondary electron impact and converts this information into an electrical signal.5. Computer—translates the signal from the detector to produce an image.

[Sample Holder

| Detector

Computer Figure 7.37 Block diagram of the compo-nents of a SEM. There are five basic components to a SEM: the accelerated electron source, lenses, the sample holder, the detector, and a computer, which produces the final image.

Page 45: Surface Properties of Bio Materials

optimal imaging, such as polymers, must be pre-coated with a thin layer of conductive material (metal) to reduce charge build-up during scanning.

This is accomplished via physical sputtering from a metallic target onto the sample before imaging.

the secondary electrons are produced from the coating only. As illustrated in Fig. below, the (coated) sample is placed into the holder and the electron beam is scanned over the surface in a raster motion. (The beam scans each line from side to side and all the lines from top to bottom.) The secondary electron detector is positioned so that the locations of the emitted electrons are recorded. The signal from the detector is then processed using appropriate software to produce a three-dimensional image.

Variable high-voltage power supply

Electron beam -

Electron gun

Magnetic condenser lens Magnetic objective lens

X-ray detector /

Sample chamber

To vacuum -

Page 46: Surface Properties of Bio Materials

1.6.6.3 Information Provided SEM is commonly used to visualize the surface topography of a

biomaterial. the requirement of imaging in vacuum prevents full analysis of

biomaterials as they would be found in vivo. recent development of the environmental SEM allows imaging of

partially hydrated samples. The combination of SEM and EDXA provides information about

chemical composition of the sample. the ability to distinguish surface chemistry is limited.

Page 47: Surface Properties of Bio Materials

1.6.7 Scanning Probe Microscopy (SPM): Atomic Force Microscopy (AFM)1.6.7.1 Basic Principles

Atomic force microscopy (AFM). Like SEM, AFM can provide three-dimensional images of material surfaces with A to nm spatial resolution.

Limited analytical capabilities to the uppermost atomic layer of a sample because its operation is based on interactions with the electron clouds of atoms at the surface.

in AFM, a small tip is attached to a cantilever. When the tip encounters the material surface, van der Waals and electrostatic interactions between atoms in the tip and those on the surface create a characteristic force profile.

then cause eventual attraction of the tip to the surface, the bending the cantilever.

To obtain an image, the cantilever/tip assembly is rastered across the material in a controlled manner.

the stage containing the sample is moved up or down in response to the bending of the cantilever.

The record of the change in stage position required to achieve this constant contact forms the basis of the height data displayed in the three-dimensional image.

Page 48: Surface Properties of Bio Materials

1.6.7.2 Instrumentation AFM techniques can be utilized to image pure biomaterial

surfaces as well as those including adsorbed proteins.

there are four basic components to an AFM:

1. Cantilever/tip—bends in response to forces between tip and sample.2. Easer/detector—a laser beam is bounced off the cantilever and directed toward a photodiode detector. 3. Sample stage—holds sample securely. 4. Computer—translates the signal from the photodiode detector and provides feedback to control stage position.

S u rfa ce to p o g rap h y (v ia A F M ) o f p o ly(D ,L -lac tic ac id ) -p o ly(e th y len e g ly co l)-m o n o m eth y l e th e r d ib lo ck co p o ly m e r (M e .P E G -P L A s— n u m b e rs in d ica te th e w e ig h t o f th e co rresp o n d in g p o lym e r b lo ck s in k D a) film s co m p ared to a P L A film : (a ) P L A , ( b ) M e .P E G 5 -P L A 4 5 , (c ) M e.P E G 5 -P L A 2 0 , ( d ) M e .P E G 5 -P L A 1 0 as o b se rv ed b y A F M . T h e P L A film h as an a lm o st sm o o th , n o n stru c tu red su rfa ce . W ith in c re as in g M e.P E G co n ten t o f th e p o ly m e rs , th e d en sity o f p a rticu la te s tru c tu res (p red o m in an tly c ry s ta llized P E G ) in c rea se s tre m en d o u sly .

Page 49: Surface Properties of Bio Materials

imaging begins when the sample is placed on the stage the tip is lowered until it contacts the surface. cantilever indicates that the cantilever is bent toward the sample. The tip/cantilever are then moved across the sample surface in a

controlled manner the cantilever deflection is monitored. the stage is moved up and down to maintain contact between the

sample and the tip. Alterations in stage position are recorded and processed by the

appropriate software to form a three-dimensional image.

Surface and tip are out of the interactive range Piezo driver moves the specimen under computer

Surface approaches the tip

S c h e m a tic o f A F M in stru m e n ta tio n . Im a g in g b e g in s w h en th e sa m p le is p la ce d o n a s ta g e an d th e tip is lo w e red u n ti l it c o n ta c ts th e su r fa c e . W h en th is o c cu rs , re fle c tio n o f a la se r o ff th e c an ti le v e r in d ic a tes th a t th e c an ti le v e r is b en t to w a rd th e s a m p le . T h e tip /c an tile v e r a re th e n m o v e d a c ro ss th e sa m p le su rfa c e an d th e c a n tile v e r d e flec tio n is m o n ito red . In re sp o n se to ch a n g es in d e fe c tio n , th e s ta g e is m o v e d u p an d d o w n to m a in ta in co n ta c t b e tw e e n th e sa m p le an d th e tip . A lte ra tio n s in s ta g e p o s itio n a re re c o rd ed an d p ro c e sse d b y th e ap p ro p ria te so ftw a re to fo rm a th re e -d im e n sio n a l im a g e .

Page 50: Surface Properties of Bio Materials

Summary The adsorption of proteins onto a biomaterial surface is governed

by thermo-dynamic principles that are strongly influenced by two surface properties: surface hydrophobicity and surface charge physical properties of the material surface such as surface

roughness and steric hindrance and influence protein adsorption

It is very important to control Protein absorption Controlling protein adsorption to the biomaterial surface because it is to this coated surface, rather than the original implant, that the body reacts.

Physicochemical surface treatments of biomaterials use of physical principles or chemical reactions, but they do not involve the attachment of biological molecules to the material.

Biological surface modification techniques are used to attach biologically active molecules to a material surface.

Physicochemical methods can involve a coating of the surface can not.

Surface modifications that result in surface coatings can involve covalent or non-covalent bonding to the surface.

Page 51: Surface Properties of Bio Materials

Covalent surface coatings can be achieved through: plasma discharge chemical vapor deposition physical vapor deposition techniques radiation grafting techniques self-assembled monolayer deposition.

Non-covalent surface coatings can be formed through: solution coating deposition of Langmuir-Blodgett films the use of surface-modifying additives

Techniques of physicochemical surface modification that do not produce an overcoat include:

ion beam implantation plasma treatment, conversion coatings (despite the

implications of the name). the use of bioactive glasses.

Page 52: Surface Properties of Bio Materials

Biological surface modifications can be either covalent or non-covalent in nature

Surface characteristics of a material can change with time. Patterning techniques for altering the surface properties of

biomaterials in a controlled manner, resulting in a geometric design of well-defined dimensions, include microcontact printing and microfluidics.

Contact angle analysis provides information about hydrophobicity and it is typically an initial step in material surface characterization.

Microscopy techniques including:

light microscopy electron microscopy, scanning probe microscopy primarily provide information about the appearance and

topology of the biomaterial.

Page 53: Surface Properties of Bio Materials

spectroscopic techniques involve:

the absorption of electromagnetic radiation by the material. include electron spectroscopy for chemical analysis (ESCA) . An attenuated total internal reflectance Fourier-transform

infrared spectroscopy (ATR-FTIR). These methods. a modification of mass spectrometry for surface analysis. secondary ion mass spectrometry (SIMS). record information about the structure and composition of

the outermost few layers of both organic and inorganic materials.

Page 54: Surface Properties of Bio Materials

THANK YOU