Structure and properties of polycaprolactone/ibuprofen ... · polycaprolactone matrix can reduce...
Embed Size (px)
Transcript of Structure and properties of polycaprolactone/ibuprofen ... · polycaprolactone matrix can reduce...
-
ORIGINAL PAPER
Structure and properties of polycaprolactone/ibuprofenrods prepared by melt extrusion for implantable drugdelivery
G. V. Salmoria1,2 • F. Sibilia2,3 • V. G. Henschel1,2 •
S. Fare3 • M. C. Tanzi3
Received: 3 August 2016 / Revised: 27 February 2017 /Accepted: 21 March 2017
� Springer-Verlag Berlin Heidelberg 2017
Abstract In this study, the structure and properties of polycaprolactone/ibuprofen(PCL/IBP) rods prepared by melt extrusion were investigated by infrared spec-
troscopy, X-ray diffraction, scanning electron microscopy, flexural tests, dynamic
mechanical analysis and drug release analysis. The crystallinity values for the PCL/
IBP rods were lower than that for the pure PCL rods. The PCL/IBP rods had higher
values for the flexural modulus compared with the pure PCL rods prepared using the
same processing temperature, suggesting that ibuprofen has a hardening effect when
dispersion in a PCL matrix. Rods prepared at a processing temperature of 130 �Chad the highest flexural modulus and glass transition temperature, probably due to
better drug dispersion in the PCL matrix at lower temperature. The surfaces of PCL/
IBP rods prepared at 150 �C had small particles and molten drug was deposited onthe surface. This is probably due to the low melt temperature of ibuprofen and thus
at high temperatures the ibuprofen phase migrates from the PLC matrix to the rod
surface. The PCL/IBP rods prepared using different processing temperatures pro-
vided different drug release behaviors, with fast or slow drug release depending on
the ibuprofen distribution. This feature is of great interest in relation to producing
implantable drug delivery rods for acute inflammatory crisis or therapeutic treat-
ments via controlled release.
Electronic supplementary material The online version of this article (doi:10.1007/s00289-017-1999-x)contains supplementary material, which is available to authorized users.
& G. V. [email protected]
1 CIMJECT Laboratory, Department of Mechanical Engineering, Federal University of Santa
Catarina, Florianópolis, SC 88040-900, Brazil
2 Biomechanics Engineering Laboratory, University Hospital (HU), Federal University of Santa
Catarina, Florianópolis, SC 88040-900, Brazil
3 Laboratorio di Biomateriali, Politecnico di Milano, Via Golgi 39, 20133 Milan, LO, Italy
123
Polym. Bull.
DOI 10.1007/s00289-017-1999-x
http://dx.doi.org/10.1007/s00289-017-1999-xhttp://crossmark.crossref.org/dialog/?doi=10.1007/s00289-017-1999-x&domain=pdfhttp://crossmark.crossref.org/dialog/?doi=10.1007/s00289-017-1999-x&domain=pdf
-
Keywords Extruded rods � Polycaprolactone/ibuprofen � Drug delivery
Introduction
In recent years the melt extrusion (ME) process has found widespread application as
a viable drug delivery option in the drug development process. It is mainly used to
improve the dissolution of insoluble drugs and to obtain controlled-release
formulations and homogeneous mixtures. During the melt extrusion of a drug
within a polymer binary mixture the drug can remain molecularly dispersed within
the polymer or it can be present as an amorphous or crystalline phase [1, 2]. Hot-
melt extrusion offers some advantages, including ease of use and lack of residual
solvents, decreasing the environmental hazards and costs, and it allows continuous
processing [1–6]. Implantable medical devices are widely used in the pharmaceu-
tical field and these are generally prepared by melt extrusion. Considering the local
delivery and controlled-release profiles, these implants could be used as
implantable drug carriers and functional materials in medical devices [7–9].
Ibuprofen, obtained from isobutyl-phenyl-propanoic acid, is a nonsteroidal anti-
inflammatory drug used to treat fever, mild-to-moderate pain after surgery, painful
menstruation, osteoarthritis, dental pain, headaches and pain from kidney stones
[10–13]. It is also effective in cases of inflammatory diseases, such as juvenile
idiopathic arthritis and rheumatoid arthritis. In addition, it has been given to patients
with pericarditis and patent ductus arteriosus. Ibuprofen is an inhibitor of
cyclooxygenase, an enzyme involved in prostaglandin synthesis via the arachidonic
acid pathway. Its pharmacological effects are believed to be due to the inhibition of
cyclooxygenase-2 (COX-2), which decreases the synthesis of prostaglandins
involved in mediating inflammation, pain, fever, and swelling. Antipyretic effects
may be due to action on the hypothalamus, resulting in an increased peripheral
blood flow, vasodilation, and subsequent heat dissipation [11–13]. Inhibition of
COX-1 is thought to cause some of the side effects of ibuprofen including
gastrointestinal ulceration. The toxic effects of ibuprofen can be severe at doses
above 400 mg/kg or 3200 mg per day [11].
Polyesters are polymers used in biomedical implants and drug delivery systems
with a biodegradable character [13–15]. The best-known examples are polycapro-
lactone (PCL), poly-lactic acid (PLA) and copolymers of lactic acid and glycolic
acid [4, 7, 13–15]. Preliminary, the production of PCL/IBU tubes presented
interesting mechanical properties and ibuprofen release profile for nerve regener-
ation [16]. The general use of an implantable drug delivery system based on a
polycaprolactone matrix can reduce the risk of ibuprofen overdose compared to oral
administration. In this study, the structure and properties of polycaprolactone/
ibuprofen rods prepared by melt extrusion for general drug delivery use were
investigated employing infrared spectroscopy, X-ray diffraction, scanning electron
microscopy, flexural tests, dynamic mechanical analysis and drug release analysis.
The structure and properties of the rods were correlated with the processing
temperature.
Polym. Bull.
123
-
Experimental
Materials
In this study poly(e-caprolactone) reference 440744 (Sigma-Aldrich) was used. Itwas acquired in the form of pellets with 3 mm of diameter and average molecular
weight (Mn) in the range of 70,000–90,000 g/mol, determined by the GPC
technique. The polydispersity value was\2. The ibuprofen (IBP) was obtained fromPharmanostra (São Paulo, Brazil) and it had the appearance of a white crystalline
powder with a density of 1.175 g/cm3. The ibuprofen melt temperature was
determined by DSC as 77.4 �C.
Melt extrusion process
The extrusion equipment used was an Axplastic mono-screw extruder (model LAB
AX-14) with an L/D ratio of 20. The extruder was instrumented with an extrusion
die for small-diameter rods (diameter 5 mm). The three different zones of the
extruder barrel were pre-set to the same processing temperature. A composed
cooling system with cold air (10 �C) and chill ceramic plate (15 �C) was used forthe downstream processing of extrudates. Figure 1 shows a diagram of the extrusion
system used in the preparation of the PCL/IBP rods. The pure PCL was processed
using the intermediate values for the extrusion parameters (i.e., melt temperature of
140 �C and screw speed of 50 rpm). The processing conditions used in thefabrication of the PCL/IBP rods with 10% of ibuprofen content are shown in
Table 1.
Fig. 1 Diagram of extrusion system
Polym. Bull.
123
-
Scanning electron microscopy (SEM)
Philips XL 30 and JEOL JSM-6390LV scanning electronic microscopes were used in
the microstructural analysis. For fracture analysis, the samples were mechanically
fractured in liquid nitrogen (cryogenic fracture). Prior to the analysis by SEM, all
samples were coated with a thin layer of gold in a sputter diode D2 sputtering system.
X-ray diffraction and Fourier transform infrared (FTIR) analysis
X-ray diffraction analysis was carried out on a Philips X’Pert diffractometer. The
scan was obtained at 2h = 5�–40� directly from the surface of the sample rodpositioned horizontally. The specimen crystallinity was determined by the area
measurement method. Fourier transform infrared spectra of the pure material and
the rods were obtained using a Perkin-Elmer Frontier MIR/NIR spectrophotometer
in attenuated total reflectance (ATR) mode scanning from 4000 to 450 cm-1.
Mechanical analysis
A DMAQ800 analyzer (TA instruments) with a single cantilever clamp was used for
mechanical tests carried out on rods with 2.2 mm diameter and 30.0 mm length. A
force rate of 2 N/min from 0 to 18 N was applied in the quasi-static tests. Dynamic
mechanic analysis (DMA) was conducted to obtain the values for the storage modulus
(E0), loss modulus (E00) and tan delta (d) at a frequency of 1 Hz within the temperaturerange of -80 to 80 �C, using a heating rate of 3 �C/min and strain of 0.3%.
Recovery of ibuprofen incorporated in the rods
Rod segments obtained from three different portions of the total material produced
under each condition were collected. Each piece was weighed and cut into smaller
pieces to increase the surface area in contact with the solvent added (10 mL of
methanol; in triplicate). Samples were kept in an ultrasonic bath for 2 h and then
analyzed by UV–vis spectrophotometry, at kmax of 269 nm, on a Hitachi 2010double-beam UV–visible spectrophotometer.
Drug-release test
Specimenswith knownmasswere placed in sealedvialswith 20 mLofphosphate buffer
solution (pH = 7.4) and 0.01 M cetyltrimethyl ammonium bromide (to maintain sink
Table 1 Processing conditions used in the production of the PCL/IBP rods
Factors Low Inter-mediate High
Temperature (�C) 130 140 150Screw speed (rpm) 50 50 50
Drug content (%) 10 10 10
Polym. Bull.
123
-
solubilization conditions). The vials were shaken horizontally in a Dubnoff bath (Faalk,
Brazil)maintained at a temperature of 37.0 ± 0.5 �Cat a rate of 60 rev/min tominimizethe boundary effect. Total buffer solutionswere collected periodically (at predetermined
time intervals). After suitable dilutionwith the buffer solution, the total drug releasewas
determined, based on a previously obtained calibration curve (five dilutions of
0.050–0.40 mg/mL), using UV–vis spectrophotometry, at kmax of 269 nm, on a Hitachi2010 double-beam UV–visible spectrophotometer.
Results and discussion
Figure 2 shows the cryogenic fracture surfaces of the PCL/ibuprofen rods and
ibuprofen particles. Measurements revealed an average value for the diameter of the
PCL/IBP rods of 2.2 (±0.2) mm. The cryogenic fracture surface morphologies of
the PCL/IBP rods showed similar brittle characteristics. The fracture surface of
PCL/IBU rod prepared at 130 �C shows the distribution of ibuprofen particles intothe PCL matrix (Fig. 2a–c). The ibuprofen particles present at fracture surface have
smaller average size (50 lm) than the pure ibuprofen particles (200 lm, Fig. 2f)used, suggesting the partial dissolution of ibuprofen into PCL matrix. The PCL/IBU
rod processed at 150 �C presents the lower quantity of ibuprofen particles at thefracture surface (Fig. 2g–i).
Fig. 2 Images of cryogenic fracture surfaces of PCL/ibuprofen rods extruded using different processtemperatures: 130 �C (a–c), 140 �C (d, e) and 150 �C (g–i), and of ibuprofen particles (f)
Polym. Bull.
123
-
The images in Fig. 3 shows the surfaces of the PCL/IBP rods where low surface
irregularities can be observed. The PCL/IBP rods presented small particles and
molten drug deposited on the surface, particularly in the case of PCL/IBP rods
prepared using 150 �C (Fig. 3c), probably due to the low melt temperature of theibuprofen which leads to migration of the ibuprofen phase from the PLC matrix to
the surface of the extrude rods. The growth of ibuprofen particles can be
significantly affected by the presence of polymer matrix. This phenomenon was also
observed in other studies with ibuprofen and polymers [16, 17]. The use of the EDS
technique confirmed the presence of small particles of ibuprofen on the surfaces of
the PCL/IBP rods, using the composition analysis to compare the oxygen contents
of the ibuprofen particles and the PCL matrix (Fig. 3d).
Figure 4 shows the X-ray diffractograms for the pure PCL and PCL/IBP rods
prepared at 140 �C. The diffractogram for pure PCL shows two peaks (at 2h 21.6�and 23.8�). These peaks are indexed to the (110) and (200) planes of anorthorhombic crystalline structure, respectively [18, 19]. The diffractograms of the
PCL/IBP rods prepared using different temperatures showed the same peaks at 2h21.6� and 23.8�. Peaks related to the crystalline phase of ibuprofen at 11.9�, 16.5�,19.3� and 22.5� [20] were not detected, due to the fact of ibuprofen was the minorcomponent (10 wt.%) and it was present as kinetically stabilized amorphous phase
[21].
Fig. 3 Scanning electronic micrographs of the surface of PCL/ibuprofen rods prepared applyingprocessing temperatures of 130 �C (a), 140 �C (b) and 150 �C (c) and EDS results for carbon and oxygencontents in ibuprofen particles and PCL matrix (d)
Polym. Bull.
123
-
The crystallinity values for the pure PCL and PCL/IBP rods are shown in
Table 2. It can be observed that the crystallinity values for the PCL/IBP rods are
lower than that for the pure PCL rod, probably due to interaction of the drug with
the polymer chains and the processing conditions. According to Pant [22] and
collaborators, the occurrence of this phenomenon is verified by the interaction of
components at the molecular level, which suggests the partial dissolution of
ibuprofen in the PCL matrix. Lower processing temperatures led to a decrease in the
crystallinity of the PCL/IBP rods, probably due to a rapid solidification and better
dispersion of ibuprofen in the PCL matrix.
Figure 5 shows the FTIR spectra for the ibuprofen (a), pure PCL (b) and PCL/
IBP rods (c–e) prepared using different processing temperatures. The pure PCL
shows peaks at 2940, 2863, 1721, 1294, 1164, 1157 and 717 cm-1 related to CH2asymmetric stretching, CH2 symmetric stretching, C=O carbonyl stretching, C–O
stretching in the crystalline phase, asymmetric C–O–C stretching, C–O stretching in
the amorphous phase and CH2 rocking, respectively (Fig. 5b) [23, 24]. The peaks
present in the ibuprofen spectrum at 2955, 1721 and 1240 cm-1 are assigned to
asymmetric CH3 stretching, C=O stretching and C–O stretching, respectively
(Fig. 5a) [25], while the peaks at 668 and 580 cm-1 are related to the aromatic ring
Fig. 4 Diffractograms of pure PCL and PCL/IBP rods produced using a processing temperature of140 �C
Table 2 Crystallinity by X-raydiffraction to pure PCL and
PCL/IBP rods prepared with
different extrusion temperatures
Samples Crystallinity (%)
Pure PCL 140 �C 56 ± 2PCL/IBP 130 �C 48 ± 2PCL/IBP 140 �C 50 ± 2PCL/IBP 150 �C 51 ± 3
Polym. Bull.
123
-
vibration in the ibuprofen structure. The main peaks in the FTIR spectra for the
PCL/IBP rods (Fig. 5c–e) were the same as those observed in the spectra for the
PCL rods and the peaks at 668 and 580 cm-1 are related to the aromatic group in the
ibuprofen structure, verifying the presence of the drug. The interpretation on the
intermolecular interaction was from the peaks related to PCL, the main component
(90 wt.%). The C=O stretching peak of PCL (1721 cm-1) overlapped too much with
that of IBU (1721 cm-1). However, the dislocation of the C=O stretching peak of
PCL to 1713 cm-1, suggested the existence of intermolecular interactions between
IBU and PCL. The appearance of a broad band from 3300 to 3100 cm-1,
corresponding to O–H bending, can be related to interactions between PCL and
ibuprofen by hydrogen bonds [26]. Also can be noted the displacement of the PCL
peaks 2940 and 2863 cm-1 corresponding to the stretching of CH2 groups in PCL/
IBU rods spectra.
DSC measurements were performed to characterize the thermal behavior the
ibuprofen, PCL and PCL/IBU rods. The DSC thermograms are presented in Fig. 6.
At ambient temperature, Ibuprofen is a white crystal. Under heating (at 10 �C/min),this crystalline form melts at 78.8 �C, which gives rise to an endothermic peak onthe DSC thermogram. This melting temperature is typical of the racemic mixture
[27]. The melting enthalpy is 99 J/g, in agreement with the value previously
reported [27, 28]. PCL showed a melting process located between 53 and 57 �Cwith a Tm = 56.3 �C, related to a crystalline phase, in agreement with the valuespreviously reported [29]. In all PCL/IBU rods, only an endothermic peak between
52 and 57 �C was observed, that is probably due to the melting of the crystallinephase of PCL matrix. The melting temperature of the PCL/IBU rods was found to
slightly decrease with the presence of ibuprofen (Table 3). The endothermic peak
Fig. 5 FTIR-ATR spectra for ibuprofen (a), pure PCL (b) and PCL/IBP rods prepared at processingtemperatures of 130 �C (c), 140 �C (d) and 150 �C (e)
Polym. Bull.
123
-
related to the melting of crystalline phase of ibuprofen was not detected. Previous
investigation showed that when ibuprofen is quenched faster than 20 �C/min,crystallization is avoided [27]. This ibuprofen behavior can explain the absence of
an endothermic peak related to the ibuprofen melting on DSC thermograms of PCL/
IBU rods fabricated by hot melt extrusion. Campbell and collaborators demon-
strated that ibuprofen was well dispersed and distributed throughout the PCL/clay
composite matrix. The glass transition of pure PCL increased by up to 16 �C foribuprofen composites. It was attributed to the constrained mobility of PCL chains
intercalated between the composites components and to the tethering of PCL chains
by hydrogen bonding with those. As a consequence, PCL crystallization was
inhibited. The fraction of PCL that was crystalline decreased by 15% on addition of
ibuprofen and nanoclay [30].
The DSC crystallinity calculated from the enthalpy of melting of 100%
crystalline PCL [29], for the pure PCL and PCL/IBP rods are shown in Table 3. It
can be observed that the DSC crystallinity values for the PCL/IBP rods are slightly
Fig. 6 DSC thermograms for ibuprofen (a), pure PCL (b) and PCL/IBP rods prepared at processingtemperatures of 130 �C (c), 140 �C (d) and 150 �C (e)
Table 3 Melting temperature and DSC crystallinity to pure PCL and PCL/IBP rods prepared withdifferent extrusion temperatures
Samples Melting temperature (�C) Crystallinitya (%)
Pure PCL 140 �C 56.3 62 ± 2PCL/IBP 130 �C 55.5 55 ± 2PCL/IBP 140 �C 55.7 58 ± 2PCL/IBP 150 �C 56.2 61 ± 2a Crystallinity values based in the enthalpy of melting of 100% crystalline PCL, 142.0 J/g [29]
Polym. Bull.
123
-
lower than that for the pure PCL rod, probably due to interaction of the drug with
the polymer chains and the processing conditions. Lower processing temperatures
(130 �C) led to a decrease in the crystallinity of the PCL/IBP rods, probably due to abetter dispersion of ibuprofen in the PCL matrix and faster solidification when
compared to PCL/IBU rod fabricated using higher processing temperature (150 �C).The flexural modulus for the PCL/IBP rods prepared at different processing
temperatures are shown in Table 4. The PCL/IBP rods fabricated at 140 �C hadhigher flexural modulus values compared with the pure PCL rods prepared using the
same processing temperature, suggesting that dispersion of the ibuprofen in the PCL
matrix has a hardening effect (Fig. 7). PCL/IBP rods prepared at 130 �C had thehighest flexural modulus value, probably due to better drug dispersion in the PCL
matrix at lower temperature.
Figure 8 shows the curves for the storage modulus (E0) as a function oftemperature for the PCL/IBP rods prepared using different processing conditions.
The storage modulus for the PCL/IBP rods decreased with the DMA test
temperature, particularly at temperatures close to -50 �C, which relates to thePCL glass transition temperature. The rods prepared using a higher processing
temperature had a lower storage modulus value, indicating the effect of temperature
Table 4 Flexural modulus forpure PCL and PCL/IBP rods
prepared by melt extrusion
Specimen Flexural modulus (MPa)
Pure PCL 140 �C 221 ± 36PCL/IBP 130 �C 257 ± 30PCL/IBP 140 �C 244 ± 27PCL/IBP 150 �C 230 ± 33
Fig. 7 Stress versus strain curves obtained in flexural test for PCL/IBP rods produced using differentmelt temperatures: 130 �C (a), 140 �C (b) and 150 �C (c), and pure PCL (d)
Polym. Bull.
123
-
on the ibuprofen dispersion which, in turn, influences the mechanical properties of
the PCL/IBP rods.
The determination of the maximum value for the loss tangent as a function of
temperature (Fig. 9) allowed the glass transition temperature (Tg) to be determined
for the PCL/IBP rods prepared using different processing temperatures. The PCL/
IBP rods had higher Tg values compared with the pure PCL rods (Table 5),
indicating that the dispersion of the ibuprofen in the PCL matrix had a hardening
Fig. 8 Storage modulus (E0) as a function of temperature for PCL/IBP rods produced using differentmelt temperatures: 130 �C (a), 140 �C (b) and 150 �C (c), and pure PCL (d)
Fig. 9 Values for loss tangent (tand) as a function of temperature for PCL/IBP rods prepared usingdifferent processing temperatures: 130 �C (a), 140 �C (b), 150 �C (c) and pure PCL (d)
Polym. Bull.
123
-
effect. Rods prepared using a processing temperature of 130 �C had the highestglass transition temperature, suggesting a greater dispersion of ibuprofen in the PCL
matrix. Similar effect was found by Campbell and collaborators on the glass
transition of ibuprofen composites [30].
The recovery of the ibuprofen incorporated into the rods showed an average of
96% (±1%) incorporation of the drug, presenting no significant differences in
relation to the processing temperature, which indicates that although there was drug
loss this was constant, contributing to the process reproducibility. The drug release
curves for PCL/IBP rods prepared applying different processing conditions are
shown in Fig. 10. The rods produced at a higher processing temperature (150 �C)showed a rapid release during the first 3 days and a release of 80% of ibuprofen
during the first 30 days due to the drug distribution mainly at the rod surface. In
previous study about the production thin wall tubes of PCL/IBU, we found a high
rate of ibuprofen release, adequate for nerve regeneration [16].
On the other hand, PCL/IBU rods produced at a lower processing temperature
(130 �C) showed only 10% of ibuprofen release during the first 30 days, showing acontrolled release of the dispersed ibuprofen from the polymeric matrix by
diffusion/erosion mechanism. The ibuprofen dissolution and initial burst effect
observed to the rods prepared at 150 �C, were retarded by the better dispersion of
Table 5 Tg values for purePCL and PCL/ibuprofen rods
obtained from the maximum loss
tangent in DMA
Sample Tg (�C)
Pure PCL 140 �C -40 ± 2PCL/IBP 130 �C -35 ± 2PCL/IBP 140 �C -36 ± 3PCL/IBP 150 �C -38 ± 2
Fig. 10 Ibuprofen release as a function of time for rods prepared applying different processingtemperatures: 130 �C (a), 140 �C (b) and 150 �C (c)
Polym. Bull.
123
-
ibuprofen into the PCL matrix of rods prepared 130 �C. Terebetski and Michniak-Kohn suggested that induction of intermolecular interactions between ibuprofen and
cellulose polymers were effective at inhibiting nucleation and maintaining
prolonged drug release [31]. Del Valle and collaborators showed a moderator
effect of PCL over ibuprofen release. Polymer-ibuprofen blends presenting a rapid
release could have the release delayed by the addition of polycaprolactone to the
mixture [32].
The PCL/IBP rods prepared at different processing temperatures provided varied
drug release behaviors (fast to slow release), which is of interest with regard to
producing implantable drug delivery rods for acute inflammatory crisis or
therapeutic treatments.
Conclusions
It has been demonstrated that the fabrication of PCL/IBU rods is possible, and it can
be done by hot melt extrusion with efficiency. PCL/IBP rods showed small particles
of ibuprofen deposited on the surface, especially in the case of PCL/IBP rods
prepared at 150 �C. The crystallinity values for the PCL/IBP rods were lower thanthat for the pure PCL rod, probably due to interaction of the drug with the polymer
chains. The rods prepared using 130 �C as the processing temperature had thehighest flexural modulus, probably due to better drug dispersion in the PCL matrix
at lower temperature. The rods prepared using a higher processing temperature had
a lower storage modulus value, indicating that the temperature had an effect on the
ibuprofen dispersion which, in turn, influenced the mechanical properties of the
PCL/IBP rods, as observed in the quasi-static tests.
Rods produced at a processing temperature of 150 �C showed a fast ibuprofenrelease during the first 30 days. The rods produced using the lowest processing
temperature (130 �C) showed a slower ibuprofen release, indicating greaterdispersion of the ibuprofen in the PCL matrix. The PCL/IBP rods prepared using
different processing temperatures showed different drug release behaviors, with fast
or slow drug release depending on the ibuprofen distribution. This is of interest with
regard to producing implantable drug delivery rods for acute inflammatory crisis or
therapeutic treatments applying controlled release.
Acknowledgements The authors would like to thank PRONEX/FAPESC, CNPQ and FINEP forfinancial support and the Center of Microscopy-UFSC for the micrographics.
References
1. Shah S, Maddineni S, Lu J, Repka MA (2013) Melt extrusion with poorly soluble drugs. Int J Pharm
453:233–252
2. Maniruzzaman M, Boateng JS, Snowden MJ, Douroumis D (2013) A review of hot-meltextrusion:
process technology to pharmaceutical products. International scholarly research network ISRN
pharmaceutics 2012, Article ID 436763, 9 pages doi:10.5402/2012/436763
Polym. Bull.
123
http://dx.doi.org/10.5402/2012/436763
-
3. Forster A, Hempenstall J, Tucker I, Rades T (2001) Selection of excipients for melt extrusion with
two poorly water-soluble drugs by solubility parameter calculation and thermal analysis. Int J Pharm
226:147–161
4. Santos DV, Casadei APM, Pereira RV, Aragones A, Salmoria GV, Fredel MC (2012) Development
of polymer/nanoceramic composite material with potential application in biomedical engineering.
Mater Sci Forum 727:1142–1146
5. Crowley MM, Fredersdorf A, Schroeder B, Kucer S, Prodduturi S, Repka MA, McGinity JW (2004)
The influence of guaifenesin and ketoprofen on the properties of hot-melt extruded polyethylene
oxide films. Eur J Pharm Biopharm 22:409–418
6. Crowley MM, Schroeder B, Fredersdorf A, Obara S, Talarico M, Kucera S, McGinity JW (2004)
Physicochemical properties and mechanism of drug release from ethyl cellulose matrix tablets
prepared by direct compression and hot-melt extrusion. Int J Pharm 269:509–522
7. Guo G, Fu SZ, Zhou LX, Liang H, Fan M, Luo F, Qian ZY, Wei YQ (2011) Preparation of curcumin
loaded poly(caprolactone)-poly(ethylene glycol)-poly(caprolactone) nanofibers and their in vitro
antitumor activity against Glioma 9L cells. Nanoscale 3:3825–3832
8. Zepon KM, Vieira LF, Soldi V, Salmoria GV, Kanis LA (2013) Influence of process parameters on
microstructure and mechanical properties of starch-cellulose acetate/silver sulfadiazine matrices
prepared by melt extrusion. Polym Testing 32:1123–1127
9. Zepon KM, Petronilho F, Soldi V, Salmoria GV, Kanis LA (2014) Production and characterization of
cornstarch/cellulose acetate/silver sulfadiazine extrudate matrices. Mater Sci Eng C 44:225–233
10. Bédouet L, Laurent B, Pascale F, Bonneau M, Wassef M, Laurent A (2011) In vitro evaluation of (S)-
ibuprofen toxicity on joint cells and explants of cartilage and synovial membrane. Toxicol In Vitro
25(8):1944–1952
11. Caviglioli G, Posocco V, Parodi B, Cafaggi S, Alzati A, Bignardi G (2002) Identification of
degradation products of ibuprofen arising from oxidative and thermal treatments. J Pharm Biomed
Anal 30(3):499–509
12. Kantor TG (1979) Ibuprofen. Ann Intern Med 91:877–882
13. Carreras N, Acuña V, Martı́ M, Lis MJ (2013) Drug release system of ibuprofen in PCL micro-
spheres. Colloid Polym Sci 291(1):157–165
14. Salmoria GV, Klauss P, Zepon K, Kanis LA, Roesler CRM, Vieira LF (2012) Development of
functionally-graded reservoir of PCL/PG by selective laser sintering for drug delivery devices.
Virtual Phys Prototyp 7(2):107–115
15. Salmoria GV, Klauss P, Zepon KM, Kanis LA (2013) The effects of laser energy density and particle
size in the selective laser sintering of polycaprolactone/progesterone specimens: morphology and
drug release. Int J Adv Manuf Technol 66(5):1113–1118
16. Salmoria GV, Paggi RA, Castro F, Roesler CRM, Moterle D, Kanis LA (2016) Development of PCL/
ibuprofen tubes for peripheral nerve regeneration. Procedia CIRP 49:193–198
17. Lee S.Y., Yu G., Kim W. (2013) Effects of Polymeric Additives on the Crystallization and Release
Behavior of Amorphous Ibuprofen. J Nanomate Article ID503069:7. doi:10.1155/2013/503069
18. Lee D, Lee S, Kim W (2012) Effects of humidity and surfaces on the melt crystallization of
ibuprofen. Int J Mol Sci 13:10296–10304. doi:10.3390/ijms130810296
19. Baji A, Wong S-C, Liu T, Li T, Srivatsan TS (2007) Morphological and X-ray diffraction studies of
crystalline hydroxyapatite-reinforced polycaprolactone. J Biomed Mater Res 81B:343–350. doi:10.
1002/jbm.b.30671
20. Monteiro MSSB, Chávez FV, Sebastião PJ, Tavares MIB (2013) 1H NMR relaxometry and X-ray
study of PCL/nevirapine hybrids. Polym Testing 32:553–566
21. Pattnaik S, Swain K, Rao JV, Talla V, Prusty KB, Subudhi SK (2015) Polymer co-processing of
ibuprofen through compaction for oral absorption. RSC Adv 5:74720–74725
22. Yu L (2001) Amorphous pharmaceutical solids: preparation, characterization and stabilization. Adv
Drug Deliv Rev 48:27–42
23. Pant HR, Neupane MP, Pant B, Panthi G, Oh H, Lee MH, Kim HY (2011) Fabrication of highly
porous poly(e-caprolactone) fibers for novel tissue scaffold via water-bath electrospinning. ColloidsSurf B 88:587–592
24. Paquet O, Krouit M, Bras J, Thilemans W, Belgacem M (2010) Surface modification of cellulose by
PCL grafts. Acta Mater 58:792–801
25. Mallick S, Pradhan SK, Chandran M, Acharya M, Digdarsini T, Mohapatra R (2011) Study of
particle rearrangement, compression behavior and dissolution properties after melt dispersion of
Ibuprofen, Avicel and Aerosil. Results in Pharma Science 1:1–10
Polym. Bull.
123
http://dx.doi.org/10.1155/2013/503069http://dx.doi.org/10.3390/ijms130810296http://dx.doi.org/10.1002/jbm.b.30671http://dx.doi.org/10.1002/jbm.b.30671
-
26. Carreras N, Acuña V, Martı́ M, Lis MJ (2013) Drug release system of ibuprofen in PCL-micro-
spheres. Colloid Polym Sci 291:157–165
27. Dudognon E, Danède F, Descamps M, Correia NT (2008) Evidence for a new crystalline phase of
racemic ibuprofen. Pharm Res 25(12):2853–2858. doi:10.1007/s11095-008-9655-7
28. Xu F, Sun LX, Tan ZC, Liang JG, Li RL (2004) Thermodynamic study of ibuprofen by adiabatic
calorimetry and thermal analysis. Thermochim Acta 412:33–37. doi:10.1016/j.tca.2003.08.021
29. Crescenzi V, Manzini G, Galzolari G, Borri C (1972) Thermodynamics of fusion of poly-b-propi-olactone and poly-e-caprolactone. Comparative analysis of the melting of aliphatic polylactone andpolyester chains. Eur Polym J 8(3):449–463
30. Campbell KT, Craig DQM, McNally T (2010) Ibuprofen-loaded poly(e-caprolactone) layered silicatenanocomposites prepared by hot melt extrusion. J Mater Sci 21(8):2307–2316. doi:10.1007/s10856-
009-3963-2
31. Terebetski JL, Michniak-Kohn B (2014) Combining ibuprofen sodium with cellulosic polymers: a
deep dive into mechanisms of prolonged supersaturation. Int J Pharm 475(1–2):536–546. doi:10.
1016/j.ijpharm.2014.09.015
32. Del Valle LJ, Roca D, Franco L, Puiggalı́ J, Rodrı́guez-Galán A (2011) Preparation and release study
of ibuprofen-loaded porous matrices of a biodegradable poly(ester amide) derived from L-alanine
units. J Appl Polym Sci 122(3):1953–1967. doi:10.1002/app.34017View/save
Polym. Bull.
123
http://dx.doi.org/10.1007/s11095-008-9655-7http://dx.doi.org/10.1016/j.tca.2003.08.021http://dx.doi.org/10.1007/s10856-009-3963-2http://dx.doi.org/10.1007/s10856-009-3963-2http://dx.doi.org/10.1016/j.ijpharm.2014.09.015http://dx.doi.org/10.1016/j.ijpharm.2014.09.015http://dx.doi.org/10.1002/app.34017View/save
Structure and properties of polycaprolactone/ibuprofen rods prepared by melt extrusion for implantable drug deliveryAbstractIntroductionExperimentalMaterialsMelt extrusion processScanning electron microscopy (SEM)X-ray diffraction and Fourier transform infrared (FTIR) analysisMechanical analysisRecovery of ibuprofen incorporated in the rodsDrug-release test
Results and discussionConclusionsAcknowledgementsReferences