Stir-Pak & Servodyne Mixer Manual Approved small Rev 2

21
-- 1 -- OPERATING MANUAL Cole-Parmer ® Stir-Pak ® Mixer Head Model Numbers 50007-10, 50007-15, 50007-20, 50007-25 50007-30, 50007-35, 50007-40, 50007-45 & Mixer Controller Model Numbers 50007-00, 50007-05 Cole-Parmer ® ServoDyne™ Mixer Head Model Numbers 50008-10, 50008-20, 50008-30, 50008-40 & Mixer Controller Model Number 50008-00 Cole-Parmer 625 East Bunker Court Vernon Hills, Illinois U.S.A. 60061-1844 Toll Free: 800-323-4340 Phone: 847-549-7600 Fax: 847-247-2929 www.coleparmer.com E-mail: [email protected] Version 11Nov08 Rev1

Transcript of Stir-Pak & Servodyne Mixer Manual Approved small Rev 2

-- 1 --

OPERATING MANUAL

Cole-Parmer® Stir-Pak® Mixer Head Model Numbers

50007-10, 50007-15, 50007-20, 50007-25 50007-30, 50007-35, 50007-40, 50007-45

& Mixer Controller Model Numbers 50007-00, 50007-05

Cole-Parmer® ServoDyne™ Mixer Head Model Numbers

50008-10, 50008-20, 50008-30, 50008-40 &

Mixer Controller Model Number 50008-00

Cole-Parmer 625 East Bunker Court

Vernon Hills, Illinois U.S.A. 60061-1844 Toll Free: 800-323-4340 Phone: 847-549-7600

Fax: 847-247-2929 www.coleparmer.com

E-mail: [email protected]

Version 11Nov08 Rev1

-- 2 --

SAFETY PRECAUTIONS

• High voltages exist and are accessible in the controller. Contact Cole-Parmer for servicing.

• Do not operate mixers where explosive vapors or flammable materials exist. Death or serious injury could result. Check NEC and local codes before installing.

• Never clean the controller or mixer head while power is applied to the system.

• Use only those mixer shafts that are properly sized for the mixer head being used. Refer to mixer head specifications for the proper shaft diameter or range to be used. Sizes other than those specified can cause injury to personnel or damage to equipment.

• Disconnect the AC power input line cord before connecting the mixer head cable.

• Exercise extreme care when adjusting mixer height and/or position to avoid personal injury. Never make these adjustments while the mixer shaft is rotating.

• Never insert or remove the mixer shaft/impeller while the mixer shaft is rotating.

• Be sure speed control is fully counterclockwise before the POWER switch is placed in the FWD or REV positions. (50007-00 and 50007-05 Controller)

• Check that all cords are clear of any moving parts and not subject to splashes or spills.

• Use care when setting the operating speed of the mixer. Operation at high speeds in small mixing vessels could result in hazardous splashing on the operator and equipment.

• When running at full speed and load at higher ambient temperatures, the mixer head may get very hot to the touch.

• Use of mixing shafts and impellers larger than those recommended could result in poor product performance and shortened service life.

• Mixer shafts must be inserted into high-speed through-shaft mixers at least 3 inches (8 mm) and the collet hand-tightened firmly to ensure adequate shaft retention. Keep mixer shafts as short as possible to reduce resonant vibrations caused by the shaft flexing at higher speeds.

• The mixer head is to be used only for mixing. It is not intended for any other purpose.

-- 3 --

INTRODUCTION The Stir-Pak 50007-series heavy-duty modular mixing system covered in this manual consists of eight different mixer heads and two analog controllers for 115VAC and 230VAC operation.

The ServoDyne 50008-series digital modular mixing system consists of four different mixer heads and a precision digital controller for 115VAC operation.

The following chart cross-references the model number, operating speed range and torque rating of the mixer heads and the compatible controller.

MIXER HEAD CONTROLLER OPERATING VOLTAGE

SPEED RANGE, rpm

MAXIMUM TORQUE, oz-in (kg-cm)

50007-10 50007-00 115 VAC, 50/60 Hz 2 to 180 340 (24.5) 50007-20 50007-00 115 VAC, 50/60 Hz 9 to 900 70 (5) 50007-30 50007-00 115 VAC, 50/60 Hz 23 to 2300 45 (3.2) 50007-40 50007-00 115 VAC, 50/60 Hz 60 to 6000 17 (1.2) 50007-15 50007-05 230 VAC, 50 Hz 2 to 180 340 (24.5) 50007-25 50007-05 230 VAC, 50 Hz 9 to 900 70 (5) 50007-35 50007-05 230 VAC, 50 Hz 23 to 2300 45 (3.2) 50007-45 50007-05 230 VAC, 50 Hz 60 to 6000 17 (1.2) 50008-10 50008-00 115 VAC, 50/60 Hz 3 to 180 340 (24.5) 50008-20 50008-00 115 VAC, 50/60 Hz 20 to 900 70 (5) 50008-30 50008-00 115 VAC, 50/60 Hz 60 to 2300 45 (3.2) 50008-40 50008-00 115 VAC 50/60 Hz 150 to 6000 17 (1.2)

-- 4 --

50007-00, -05 STIR-PAK CONTROLLER

The variable-speed controller, Figure 1a and 1b, is used to set the speed of the mixer heads and provides a 100:1 adjustment range.

The controller ensures accurate, dependable motor speed (mixing) control, with “soft-start” and smooth, gradual adjustment to the speed setting. Constant speed with a 100:1 adjustment range is maintained electronically via automatic compensation for changes in load torque demands at both high and low speeds. Built-in motor current compensation minimizes speed drift with changes in load. Motor current limiting protects the system in case of a locked rotor condition. Output circuit protection protects against inadvertent output shorting. Mixer rotation is reversible.

Figure 1b: Controller Rear Panel (115VAC model shown)

MOTOR CONNECTOR

LINE CORD

FORWARD-OFF-REVERSE SWITCH

POWER ON INDICATOR

SPEED CONTROL

Figure 1a: Display and Controls

-- 5 --

50008-00 SERVODYNE CONTROLLER/DISPLAY FUNCTIONS

Four-digit LED display shows selected mixing

time, mixing time remaining, rpm, and

torque values.

Displays last countdown setting.

Starts counting down when you

press the START key.

Displays the last mixer shaft speed.

Displays current torque value (in-oz for 115VAC

model; cm-kg for 230VAC model) and high and low

torque limits. Mixer automatically stops when preset limits are reached.

Zeroes the torque when impeller is

running in free air or in a reference fluid.

Starts mixer. Stops when countdown

reaches zero.

Press at any time to stop mixer.

Decreases value of parameter.

Increases value of parameter.

LEDs indicate the function being displayed (time, rpm, or torque). A flashing time or torque light and audible alarm

indicate that a limit condition has been

reached.

Figure 2a: Controller Functions

IEC 320 power entry

module

POWER SWITCH All settings are

retained in memory.

FUSE T3.15A (115VAC) T1.6A (230VAC)

MIXER HEAD PLUG RECEPTACLE

Multi-pin plug enables microprocessor to identify

connected mixer head and properly control speed and torque.

IN-RS-232C IN-Connect cable

from host computer. OUT-RS-232C

OUT-Connect cable to next mixer.

Figure 2b: Controller Rear Panel

-- 6 --

INSTALLATION AND SETUP PROCEDURES The mixer heads are equipped with an adjustable-tilt mounting assembly. The mounting assembly is designed to be attached to an optional mixer support stand or tank clamp. Components supplied as part of the mixer package include the following:

• Mixer head with 6-ft cable • Precision controller with 6-ft line cord • 3-jawed chuck on high-torque mixers • ⅜-inch collet on high-speed mixers • Adjustable-tilt mounting assembly with support rod

Installation of Adjustable-Tilt Mounting

Perform the following steps:

1. Attach mixer head to adjustable-tilt mounting assembly as follows:

a. For Stir-Pak models 50007-10, -15, -20, -25 and ServoDyne models 50008-10, -20, attach mounting assembly with four screws and lock washers. (See Figure 3a)

Figure 3a: 50007-10, -15, -20, -25 and 50008-10, -20 mixer head mounting

b. For Sir-Pak models 50007-30, -35, -40, -45 and ServoDyne models 50008-30, -40, attach mounting assembly with two screws and lock washers. (See Figure 3b)

Figure 3b: 50007-30, -35, -40, -45 and 50008-30, -40 mixer head mounting

Note: The mixer is now ready to be attached to a support stand or to a tank clamp (not supplied). Recommended stands, tank clamps, and a double (rod) clamp are available through Cole-Parmer. Part numbers are listed in the ACCESSORIES section.

• Never insert or remove the mixer shaft/impeller while the mixer shaft is rotating.

-- 7 --

Installation Using a Support Stand 1. Set up support stand in accordance with the manufacturer’s instructions.

2. Attach a double (rod) clamp to the vertical rod on the support stand and tighten clamp screw. (See Figure 4)

3. Slide rod of adjustable-tilt mounting assembly into clamp and secure with clamp screw.

4. Select an impeller and shaft for the required operation and install as follows:

a. Place impeller on shaft and tighten set screw.

b. For high-torque mixers head, slide the shaft of the stirring device into the three-jawed chuck and tighten chuck.

c. For high-speed mixer head, slide the shaft through the collet at least 3 inches and hand-tighten collet.

5. Place support stand legs around vessel. Tilt angle is adjustable from 0 to 30 degrees.

6. Position the height of the mixer head by moving the clamp up or down on the support stand or, on high-speed models, by loosening the collet and positioning the shaft.

• Exercise extreme care when adjusting mixer head height and/or position to avoid personal injury. Never make these adjustments while the mixer shaft is rotating.

• Mixer shafts must be inserted into high-speed through-shaft

mixers at least 3 inches and the collet hand-tightened firmly to ensure adequate shaft retention. Keep mixer shafts as short as possible to reduce resonant vibrations caused by shaft flexing at higher speeds.

-- 8 --

Figure 4: Typical mixer setup using a support stand and pivot mount (ServoDyne system shown)

Note: The controller can be placed on any flat surface within 6 feet of the mixer head or can be mounted on a support platform (not supplied) or a pivot mount (not supplied). Refer to the ACCESSORIES section for part numbers.

To attach the support platform proceed with step 7. To attach the pivot mount proceed with step 8. If controller is placed elsewhere proceed with step 9.

7. To install the support platform proceed as follows:

a. Attach a double (rod) clamp to the vertical support rod and position so that platform will be clear of mixer head.

b. Slide the mounting rod of the support platform into the double (rod) clamp and tighten clamp.

c. Place controller on platform and proceed to step 9.

8. To install the pivot mount proceed as follows:

a. Remove the two cap nuts from the mounting studs on bottom of controller (see Figure 4).

b. Position pivot mount bracket over two studs, reinstall two cap nuts and tighten.

c. Attach a double (rod) clamp to the vertical support rod on the stand and position so that pivot mount bracket will be clear of mixer head.

d. Slide the mounting rod of the pivot mount bracket into the double (rod) clamp and tighten clamp.

e. Adjust position of controller by first loosening knob on pivot mount bracket, adjust position and then retighten knob.

9. Connect the 6-foot cable between the mixer head and the controller rear panel.

10. Check that the controller is not powered.

a. For Stir-Pak controller, ensure that the FWD-OFF-REV switch is in the OFF position.

b. For the ServoDyne controller, check that the rear panel ON-OFF switch is in the OFF position.

11. Connect the 6-ft line cord from the controller to the appropriate 3-wire, grounded power source.

12. Place the red safety switch on the top of the mixer head in the On ‘1’ position.

The mixer is now ready to operate.

• Check that all cords are clear of any moving parts and not subject to splashes or spills.

• Be sure the SPEED control is fully counterclockwise before the POWER switch is placed in the FWD or REV positions (Stir-Pak models only).

-- 9 --

Installation Using a Tank Clamp The tank clamp is attached to the side of a tank and supports both the mixer head and the controller (see Figure 5).

1. Position the tank clamp over the edge of the tank with large knob to the outside of tank.

2. Allow clamp to rest on edge of tank for added stability and tighten knob.

3. Position mixer head over material to be mixed and insert mixer mounting shaft into upper hole in tank clamp, tighten locking knob securely.

4. Select an impeller and shaft for the required operation and install as follows:

a. Place impeller on shaft and tighten set screw.

b. For high-torque mixer head, slide the 5/16 inch to 3/8 inch shaft of the stirring device into the keyless chuck and tighten.

c. For high-speed mixer head, slide the 0.375 inch mixer shaft through the collet and tighten collet. Be sure only 0.375 (+/-0.0005) inch shafts are used. Contact Cole-Parmer or your dealer for obtaining the proper shaft size. Refer to the ACCESSORIES section for the part numbers.

• Exercise extreme care when adjusting mixer head height and/or position to avoid personal injury. Never make these adjustments while the mixer shaft is rotating.

• Mixer shafts must be inserted into high speed through-shaft mixers at least 3 inches and the collet hand-tightened firmly to ensure adequate shaft retention. Keep mixer shafts as short as possible to reduce resonant vibrations caused by the shaft flexing at higher speeds.

• Do not operate mixers where explosive vapors or flammable materials exist. Death or serious injury could result. Check NEC and local codes before installing.

-- 10 --

Figure 5: Typical mixer setup using a tank clamp (ServoDyne controller shown)

5. Position the height of the mixer head on high-speed models by loosening the collet and positioning the shaft.

Note: The controller can be placed on any flat surface within 6 feet of the mixer head and can be mounted on a support platform (not supplied) or a pivot mount (not supplied). Refer to ACCESSORIES for part numbers. To attach the support platform, proceed with step 6. To attach the pivot mount, proceed with step 7. If controller is placed elsewhere, proceed with step 8.

6. To install the support platform proceed as follows:

a. Slide the mounting rod of the support platform into the lower hole on the tank clamp and tighten clamp.

b. Place controller on platform and proceed to step 8.

7. To install the pivot mount proceed as follows:

a. Remove the two cap nuts from the mounting studs on bottom of controller.

b. Position pivot mount bracket over two studs, reinstall two cap nuts and tighten.

c. Slide the mounting rod of the pivot mount bracket into the lower hole on the tank clamp and tighten clamp.

d. Adjust position of controller by first loosening knob on pivot mount bracket, adjusting position and then retighten knob.

8. Connect the 6-foot cable between the mixer head and the controller rear panel.

9. Check that the rear panel ON-OFF switch is in the OFF position.

-- 11 --

10. Connect the 6-ft line cord from the controller to the appropriate power source, either a 90-130VAC 50/60Hz 3-wire, grounded power source or a 200-260VAC 50/60Hz 3-wire, grounded power source.

11. Place the red safety switch on the top of the mixer head in the On ‘1’ position.

The mixer is now ready to operate.

• Check that all cords are clear of any moving parts and not subject to splashes or spills.

• Be sure the SPEED control is fully counterclockwise before the POWER switch is placed in the FWD or REV positions (Stir-Pak models only).

• Always wear appropriate eye protection when mixing any kind of fluid.

• Avoid wearing loose clothing that may be caught by a rotating mixer shaft.

• Use care when setting the operating speed of the mixer. Operation at high speeds in small mixing vessels could result in hazardous splashing on the operator and equipment.

• Be sure the SPEED control is fully counterclockwise before the POWER switch is placed in the FWD or REV positions (Stir-Pak models only).

-- 12 --

50007-00, -05 STIR-PAK CONTROLLER OPERATION

FWD-OFF-REV Switch In the FWD position, the shaft will rotate in a counterclockwise direction when viewed from the impeller.

The center or OFF position removes power from both the controller and mixer head.

In the REV position, the shaft will rotate in a clockwise direction when viewed from the impeller.

POWER Indicator

The green LED indicator illuminates whenever the FWD-OFF-REV switch is in either the FWD or REV position.

SPEED Control The SPEED control precisely sets the output speed of the shaft. The controller has a range of 100:1 depending on the model; shaft speed can be as low as 2RPM or as high as 6000RPM.

Slowly adjust the SPEED Control from the 0 position to a setting that provides the desired mixing speed.

Always return the speed control to the 0 position when mixing is complete.

Note: With new mixing heads, allow approximately 20 minutes run-in time to allow warm-up drift to become minimal.

• To stop the mixer head in an emergency, place the FWD-OFF-REV switch in the OFF position or remove the power plug.

• Be sure the SPEED control is fully counterclockwise before the POWER switch is placed in the FWD or REV positions.

• Use care when setting the operating speed of the mixer. Operation at high speeds in small mixing vessels could result in hazardous splashing on the operator and equipment.

• When running at full speed and load at higher ambient temperatures, the mixer head may get very hot to the touch.

• Never insert or remove the mixer shaft/impeller while the mixer shaft is rotating.

FORWARD-OFF-REVERSE SWITCH

POWER ON INDICATOR

LIGHT SPEED CONTROL

-- 13 --

50008-00 SERVODYNE CONTROLLER OPERATION Power On

An EEPROM permits storage of all previously set parameters including speed, torque, time and calibration constant. Turning the rear-panel mounted ON/OFF switch to the “ON” position energizes the display without running the motor. The display will come up in the last mode operated. Depressing the START button causes the mixer motor shaft to rotate at the previously programmed speed. Depressing the STOP button stops the motor. Changing Display Modes (TIME, RPM, TORQUE) is accomplished by pressing the desired mode button once. Additional operation of the mode button advances operation to the set mode.

Set Run Time The use of the TIME button in conjunction with the INC/DEC buttons permits entry of a desired mixing time. Pressing the TIME button once when already in the time display mode causes the left (minutes) digit to flash, pressing the TIME button once again causes the right two (seconds) digits to flash. During this period the time can be set with the INC/DEC buttons. Pressing the TIME button a third time starts the countdown provided the motor is running; otherwise the display shows the set time. Maximum programmable time is 99 minutes 59 seconds. A “0” programmed time results in continuous motor operation.

Set Motor Speed Depressing the RPM button when in the RPM mode causes the display to flash and read the set motor speed. Use of the INC/DEC buttons changes the desired speed.

Set Motor Torque Limit The TORQUE limit feature provides for the setting of both HI and LO torque limits. Depressing the TORQUE button when in the TORQUE mode causes the first digit of the display to read either on “H” or “L” (for High or Low limit) and the last three digits to flash and read the set limit. Pressing the INC/DEC buttons changes the set torque limit.

Pressing the Torque button once again while the digits are flashing causes the H or L to flash while the set limit digits stop flashing. It is now that the High or Low limit can be selected by pressing either the INC/DEC buttons. Pressing any button other than INC/DEC buttons returns the display to actual torque. If the motor is running while the torque limit is being modified, the previously programmed limit will be in effect until the torque limit mode is exited; then the new limit will take over.

Regardless of the set torque limit, there is always a maximum torque limit check to guard against motor overheating. If this limit is exceeded, the motor stops, the alarm sounds, the display flashes “HI” and the in-oz LED flashes. Pressing any key causes the display to return to what was previously displayed before the alarm. The error must be corrected or the over-limit warning will sound again when the motor is restarted.

The TORQUE ZERO button is used to zero the torque display so that the actual torque necessary to mix something can be read directly on the display. To use this feature, the TORQUE ZERO button is pressed while the mixer head shaft is running in free air; the display will zero. The mixer shaft is then placed into the solution being mixed, the reading on the display will now indicate the torque necessary to mix the solution. This number can now be used to determine the proper shut off or torque limit when mixing several batches of the same consistency.

When the TORQUE ZERO button is operated, the maximum torque limit that can be set is adjusted so that the absolute torque does not exceed the continuous torque of the mixer head being used.

Example: The operator has set a high torque limit of 50 oz-in for a motor rated for 65. The motor is then running with an actual torque reading of 30 oz-in. When the TORQUE ZERO button is pressed, the maximum torque that can be set would be 35 (65-30). Since the set limit of 50 is above this, it will automatically be changed to 35. The operator will see this when the set mode is entered, which will prevent a setting over 35.

An audible alarm will sound and the LED that corresponds to the limit that has been reached will flash when either the preset time or the torque limit is reached. Both the LED and alarm will cycle on and off until turned off by pressing any button other than the INC/DEC buttons.

-- 14 --

• Never insert or remove the mixer shaft/impeller while the mixer shaft is rotating.

• When running at full speed and load at higher ambient temperatures, the mixer head may get very hot to the touch.

-- 15 --

Specifications – Mixer Head Direction of Rotation:

50007 Stir-Pak Series Bi-directional controlled with the FWD-OFF-REV switch 50008 ServoDyne Series CCW (Viewed from Impeller end)

Speed Range: 50007-10 and 50007-15 2 to 180 rpm 50007-20 and 50007-25 9 to 900 rpm 50007-30 and 50007-35 23 to 2300 rpm 50007-40 and 50007-45 60 to 6000 rpm 50008-10 3 to 180 rpm 50008-20 20 to 900 rpm 50008-30 60 to 2300 rpm 50008-40 150 to 6000 rpm

Torque Output, max: CONTINUOUS INTERMITTENT 50007-10, 50007-15, and 50008-10 340 in-oz 510 in-oz 50007-20, 50007-25, and 50008-20 70 in-oz 105 in-oz 50007-30, 50007-35, and 50008-30 45 in-oz 65 in-oz 50007-40, 50007-45, and 50008-40 17 in-oz 25 in-oz

Mixer Shaft Interface:

50007-10, -15, -20, -25 50008-10, -20

3-jawed type chuck with key accepts 5/16 in. to 3/8 in. dia shaft

50007-30, -35, -40, -45 50008-30, -40

Collet accepts 0.375 (± 0.0005) inch diameter shaft

Input Voltage, max: 50007-10, -20, -30, -40 50008-10, -20, -30, -40 90 Volts DC

50007-15, -25, -35, -45 180 Volts DC

Input Current, max: 50007-30, -40 50008-30, -40

2.3 Amps DC

50007-35, -45 1.2 Amps DC 50007-10, -20 50008-10, -20

1.5 Amps DC

50007-15, -25 0.8 Amps DC

Installation Category:

Installation Category I per IEC 664 (Signal level–parts of equipment)

Construction: Dimensions (L x dia) 11.5 in. x 3.6 in. (292 x 92 mm) Weight 8 lb (3.5 kg) Enclosure rating IP 24 per IEC 34-5

-- 16 --

Environment: Operating temperature 0 to 40°C (32 to 104°F) Storage temperature Storage: −45 to 65°C (−49 to 149°F) Humidity (noncondensing) 10% to 90% Altitude Less than 2000 m Degree of pollution Pollution Degree 2 per IEC 664 (Indoor Usage–lab, office)

Chemical Resistance:

Exposed material is painted aluminum, plastic, and vinyl

Compliance: 115VAC CAN/CSA C22.2, No.61010-1 2nd Edition

CAN/CSA-C22.2 No. 61010-2-051-04 ANSI/UL 61010-1 2nd Edition UL61010A-051 1st Edition

230VAC EN 61010-1 EN 61010-2-051

-- 17 --

Specifications – Controller Output:

Model Maximum Voltage, V

Maximum Current, Amps

Maximum Power, Watts

50007-00 90 2.3 207 50007-05 180 1.2 216 50008-00 90 2.3 207

Speed regulation: Model Line Load Drift

50007-00, -05 ±1% F.S. ±2% F.S. ±10% F.S. 50008-00 ±0.25% F.S. ±0.25% F.S. ±0.25% F.S.

Display Four-digit, seven-segment LED Maximum settable countdown timer 99 minutes, 59 seconds Time accuracy ±1 second

Input: Model Voltage Maximum

Current, Amps Maximum

Power 50007-00 90 to 130VAC RMS, 50/60 Hz 3 Amps 360 VA 50007-05 200 to 260VAC RMS, 50 Hz 1.6 Amps 370 VA 50008-00 90 to 130VAC RMS, 50/60 Hz 2.3 Amps 280 VA

Installation Category: Installation Category II per IEC 664 (Local level – appliances, portable equipment, etc.)

Computer Control: RS-232-C (Full Duplex)

Construction: 50007-00, -05 50008-00

Dimensions (L x W x H) 7 x 5 x 3½ inches (178 x 127 x 89 mm)

7¾ x 6½ x 3½ inches (197 x 165 x 89 mm)

Weight 1.4 lb (0.62 kg) 3 lb (1.4 kg) Enclosure rating IP 22 per IEC 529 IP 22 per IEC 529

Environment: Operating temperature 0 to 40°C (32 to 104°F) Storage temperature -45 to 65°C (-49° to 149°F) Humidity (noncondensing) 10% to 90% Altitude < 2000 m Degree of pollution 2 per IEC 664, (indoor usage – lab, office)

Chemical Resistance: Exposed material is painted aluminum, plastic, and vinyl

Compliance: 115VAC CAN/CSA C22.2, No.61010-1 2nd Edition

CAN/CSA-C22.2 No. 61010-2-051-04 ANSI/UL 61010-1 2nd Edition UL61010A-051 1st Edition

230VAC EN 61010-1 EN 61010-2-051

-- 18 --

MAINTENANCE AND TROUBLESHOOTING Maintenance is limited to replacement of motor brushes on the mixer head, replacing the fuse on the controller, and cleaning the unit. If a fault is detected when troubleshooting and cannot be corrected by the suggested remedy, contact Cole-Parmer or your dealer for service.

Motor Brush Check/Replacement Note: Brushes should be checked every 6 months or 1000 operating hours or if erratic operation

occurs. Erratic operation may occur immediately after brush replacement. Allow motor to run up to an hour to allow brushes to seat.

1. Turn the controller off

a. For the Stir-Pak controllers, place the FWD-OFF-REV switch in the OFF position

b. For the ServoDyne controller, place the rear panel ON-OFF switch in the OFF position.

2. Disconnect the AC line cord from the AC receptacle.

3. Disconnect the mixer head cable from the receptacle on the rear of the controller.

4. Carefully unscrew each brush holder cap. Withdraw the brush and examine it for wear.

Note: Replace both brushes if either brush is less than 0.300 inch (8 mm) long. The replacement brushes (set of 2) part number is listed under REPLACEMENT PARTS.

5. Insert brushes into brush holder on each side of motor as shown in Figure 6 and screw in brush holder caps.

6. Reconnect mixer head to rear-panel connector on controller.

7. Reconnect controller to primary power source.

Figure 6: Motor brush orientation

ServoDyne Controller Fuse Replacement

1. Place the power switch in the OFF position.

2. Disconnect the AC power input line cord from the AC receptacle.

3. Remove and check the fuse and replace if defective.

4. Reconnect the AC power input line cord to the AC receptacle.

-- 19 --

Cleaning the Mixer/Controller System Keep the mixer housing and controller enclosure clean with mild detergents. Never immerse nor use excessive fluid.

Replacement Parts No user-serviceable parts are inside of this instrument. Refer servicing to Cole-Parmer or your dealer.

The following parts are user-replaceable:

Description For model(s) Part Number Speed Control Knob 50007-00, -05 04552-91 Fuse, T3.15A 50008-00 77500-25 Fuse, T1.6, 5x20mm 50007-05 77500-11 Motor Brushes (2) All 07520-04 3-Jawed Chuck Assembly 50007-10, -15, -20, -25

50008-10, -20 04552-96

Collet Nut 50007-30, -35, -40 ,-45 50008-30, -40

04552-98

Pivot Assembly Knob (1) All 04552-99 Cordset–USA (115VAC) 50001-68 Cordset–European (230VAC) 50001-70 Cordset–British (230VAC) 50001-72 Cordset–Swiss (230VAC) 50001-74 Cordset–Italian (230VAC) 50001-76 Cordset–NEMA (230VAC) 50001-78

Contact Cole-Parmer or your dealer if you have service needs.

• Never clean the Controller or Mixer Head while power is applied to the system.

-- 20 --

Troubleshooting – Stir-Pak

Symptom Cause Remedy Mixer shaft does not rotate when controller is switched to FWD or REV regardless of SPEED control setting. POWER indicator does not glow.

• Disconnected or defective line cord

• Defective fuse

1. Check that the unit is plugged into a live line.

2. For 115VAC unit: Return for service (fuse is not user replaceable)

3. For 230VAC unit: Replace fuse

Mixer shaft does not rotate when controller is switched to FWD or REV regardless of SPEED control setting. POWER indicator glows.

• Defective mixer head or controller

1. Check mixer cable connections at rear of the controller - make sure connector is fully inserted into the receptacle

2. Check motor brushes (see motor brush check/replacement for details)

3. Replace mixer head or controller with known working unit, if available, to determine source of the problem

4. Return for service 5. Check that the red safety switch

on the mixer head in the On ‘1’ position.

Troubleshooting – ServoDyne The controller microprocessor detects and indicates system faults. The display alternately shows the error number and a “FAIL” message. The following table summarizes the faults and suggested corrective action. The system cannot be operated until the fault is corrected.

Error Code Fault Corrective Action

Err1 No mixer head connection Plug mixer head into controller. Err2 Invalid mixer head Bad motor cable - return mixer head for repair.

Err3 Safety Switch in Off ‘0’ position No encoder pulses

Check that the red safety switch on the mixer head in the On ‘1’ position. Return mixer head and controller for repair.

Err4 Motor speed is excessive Return mixer head and controller for repair. Err5 No torque information display Return controller for repair. Err6 Bad PROM check Return controller for repair. Err7 Motor overload Check mixer for obstructions. Turn off unit to clear error. Err8 Zero cross failure Return controller for repair. Err9 Bad EEPROM data Return controller for repair.

“HI” Limit Message

A flashing “HI” limit message will be displayed and the mixer head will stop if the maximum rated torque of the mixer head is exceeded when the low torque limit mode is being used. To protect the mixer head motor from overheating, it cannot be restarted until the fault is corrected.

Note: Before contacting Cole-Parmer or your dealer for repair, try to determine whether the fault is in the controller or mixer head. The best way is to substitute a known good unit, if available, and recheck.

-- 21 --

ACCESSORIES The accessories listed below, used for mounting the mixer head and controller, are available from Cole-Parmer or your dealer. Many other accessories such as shafts and impellers are also available. Contact your dealer for the full line of available accessories.

Part Name Part Number Tank Clamp 04552-65 Controller Support Platform 04552-75 Pivot Mount Controller Support 04552-70 Support Stand 04552-80 Double (Rod) Clamp 08041-22 Software 07550-74 ServoDyne RS-232-C Interface Cable 07550-64 Satellite Cable 07550-66 Shaft, 12 inches (305 mm) 04552-25 Shaft, 18 inches (457 mm) 04552-35 Shaft, 24 inches (610 mm) 04552-00 Shaft, 30 inches (762 mm) 04552-05

WARRANTY

The manufacturer warrants this product to be free from significant deviations from published specifications. If repair or adjustment is necessary within the warranty period, the problem will be corrected at no charge if it is not due to misuse or abuse on your part as determined by the manufacturer. Repair costs outside the warranty period, or those resulting from product misuse or abuse, may be invoiced to you.

The warranty period for this product is noted on the Warranty Card.

PRODUCT RETURN To limit charges and delays, contact Cole-Parmer or manufacturer for authorization and shipping instructions before returning the product, either within or outside of the warranty period. When returning the product, please state the reason for the return. For your protection, pack the product carefully and insure it against possible damage or loss. Any damages resulting from improper packaging are your responsibility.

TECHNICAL ASSISTANCE If you have any questions about the use of this product, contact Cole-Parmer or your local dealer.

SSMPK027 Rev 1.02