Standby Power & IEC62301
description
Transcript of Standby Power & IEC62301
www.voltech.com
1106-046/1
Standby Power & IEC62301
www.voltech.com
2106-046/1
Standby Power
• Legislative drive to reduce standby power consumption
• Increased consumer awareness of cost to pocket and the environment
With the political and market forces of climate change and rising energy costs
their has been an International Agreement to reduce the energy
consumption of electrical equipment in the “ standby mode “, this is known as
the IEA “1-watt Plan “
The agreed measurement standard and accuracy is IEC 62301
www.voltech.com
3106-046/1
Who Does This Affect ?
All manufacturers of domestic products
Every country can apply its own limits by directive or legislation.
In Europe the generally accepted directive is the “ Energy Star” but Germany also has “Blue Angel” and Scandinavian countries have adopted the “Nordic Swan”
The measurement standard for all the world wide directives is IEC62301
www.voltech.com
4106-046/1
Measurement Overview
www.voltech.com
5106-046/1
Check Test Environment
The Voltage source must be verified through the test to comply with :
Total Harmonic Content (THC) of the supply voltage must be less than 2%
The THC is up to and including the 13th harmonic
Voltage Crest Factor shall be between 1.34 and 1.49 during the measurement
www.voltech.com
6106-046/1
Power Measurement Accuracy
Power Measurement Accuracy
The measurement standard requires a Watts measurement resolution of:
0.01W Load 10W or less*
0.1W Load 10-100 W
1W Load greater than 100W
* Requires an instrument with a minimum current range of 10mA
www.voltech.com
7106-046/1
Measurement Requirements
The product should first be allowed to stabilize for a minimum of 5 minutes
Then run a measurement test for 5 minutes
Recording every second
Vrms, Vcf, Vthd, V Harmonics ( up to 13th )
Arms, Acf,
Wrms
Freq
www.voltech.com
8106-046/1
Measurement Requirements
Standby power is very small, with short bursts of power.
The PM1000+ has a 1MHz bandwidth and high sample rate to capture these distorted waveforms
www.voltech.com
9106-046/1
The Measurement System
Key Points
Universal Plug Socket for UUT
External 1 ohm Shunt for low power
10W< resolution
Direct shunt connections for measurement of asymmetric waveforms which include a DC component (H0)
www.voltech.com
10106-046/1
Power Stability
Less than 5% deviation from the maximum value observed
Measurements are stable and the Watts reading can be directly reported
www.voltech.com
11106-046/1
Stable Power ? No
If the power measure varies by more the 5% then either of the following options may be chosen
Average Power – calculate the average of all the power readings over the test period
Accumulated Energy- measure the Whrs consumed by the load.
To calculate the Average Power, divide the energy consumed in Whrs, by the measurement period.
www.voltech.com
12106-046/1
The Compliance Test Report
Product Details
Test Parameters
Measured Data
Laboratory Details
www.voltech.com
13106-046/1
Voltech SolutionPM1000+
Universal Break Out Box
IEC62301 Software
www.voltech.com
14106-046/1
Benefits to the Manufacturer
Complete Solution
Ease of use
Provides full Test Report
Provides full Test Data
PM1000+ provides both accurate and stable readings due to superior bandwidth and high sample rates
Provides confidence