Standard Operating Procedure

16
1 STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument: The Renishaw inVia Raman Spectrometer is an instrument used to analyze the Raman scattered light from samples to infer the chemistry and structure of the material of interest. Location: 325 Hodges Hall Primary Staff Contact: Weiqiang Ding (304) 6851938 cell [email protected] Secondary Staff Contact: Harley Hart (412) 4431514 cell [email protected] The Shared Research Facilities are operated for the benefit of all researchers. If you encounter any problems with this piece of equipment, please contact the staff member listed above immediately. There is never a penalty for asking questions. If the equipment is not behaving exactly the way it should, contact a staff member. WARNING: This system uses a Class IIIb laser and emits visible and invisible radiation. Laser safety goggles MUST be worn when operating the laser as an eye hazard exists. START UP NOTE: The RAMAN spectrometer, microscope, and laser should be powered down when not in use. If the power has been left on to any of these items, please inform a MFCF staff member. 1. Turn on the computer if it is not already on. Log in as Raman2 without password. 2. Log in your session on the FOM. Write down your name and sample information on the log book located on the computer cart. 3. Place the DANGER: LASER IS OPERATING sign (stored on the workbench by the door) on the outside of the door to warn any individuals who need to enter the lab. 4. Turn ON the orange microscope power switch on the leftside of the optical microscope base (Figure 1a). 5. Turn ON the power toggle switch on the frontside of the PRIOR control box located to the left of the microscope (Figure 1b).

description

Renishaw InVia Spectrophotometer

Transcript of Standard Operating Procedure

Page 1: Standard Operating Procedure

 1

STANDARD OPERATING PROCEDURE: RENISHAW INVIA RAMAN SPECTROMETER Purpose of this Instrument:  The Renishaw inVia Raman Spectrometer is an instrument used to analyze the Raman 

scattered light from samples to infer the chemistry and structure of the material of interest. 

Location: 325 Hodges Hall 

Primary Staff Contact:        Weiqiang Ding    (304) 685‐1938 cell     [email protected]  

Secondary Staff Contact:    Harley Hart          (412) 443‐1514 cell              [email protected]   

The Shared Research Facilities are operated for the benefit of all researchers.  If you encounter any problems with 

this piece of equipment, please contact the staff member  listed above  immediately.   There  is never a penalty for 

asking questions.  If the equipment is not behaving exactly the way it should, contact a staff member. 

WARNING: This system uses a Class IIIb laser and emits visible and invisible radiation.  

Laser safety goggles MUST be worn when operating the laser as an eye hazard exists. 

 

 

 

START‐UP  

NOTE: The RAMAN spectrometer, microscope, and laser should be powered down when not in use. If the power has 

been left on to any of these items, please inform a MFCF staff member. 

1. Turn on the computer if it is not already on. Log in as Raman2 without password. 

2. Log in your session on the FOM. Write down your name and sample information on the log book located on the computer cart. 

3. Place the DANGER: LASER  IS OPERATING sign (stored on the workbench by the door) on the outside of the door to warn any individuals who need to enter the lab.   

4. Turn ON the orange microscope power switch on the left‐side of the optical microscope base (Figure 1a). 

5. Turn ON  the power  toggle switch on  the  front‐side of  the PRIOR control box  located  to  the  left of  the microscope (Figure 1b). 

Page 2: Standard Operating Procedure

 2

(a)        (b)   

Figure 1. (a) Optical microscope power; and (b) motor control box power. 

(a)     (b)   

Figure 2.  (a) Laser and Power Supply Positions; (b) . Raman spectrometer power switch (marked). 

6. Check and remove any  item placed  in the path of the  laser beam on the optical table  (Figure 2a). Such 

item will cause dangerous reflections of the laser beam that poses a serious eye hazard.  

7. Toggle  the power switch on  the  right‐hand side of  the RAMAN spectrometer set‐up  to ON  (Figure 2b). Allow 20 minutes for the CCD camera to reach operating temperature. 

8. Switch  the 532 nm  laser power  supply  toggle  switch  to  the ON position  (located on  the optical  table, Figure 3a).  The power light indicator on the front of the power supply will light up. 

9. WAIT 30 SECONDS before turning on the 532nm laser key switch to the ON position to allow the power supply to stabilize.  Turn the laser power supply key to the ON position.  The laser light indicator will light up (Figure 3b).  Allow 3 minutes for the laser to stabilize. 

(a)        (b)   

Figure 3. 532nm laser power supply: (a) power switch in the back; (b) laser in the ON position. 

Page 3: Standard Operating Procedure

 3

10. OPEN the WIRE 3.2 software on the computer desktop.  

11. The  software  will  give  options  as  to  which  motors  to  reference  (Figure  4).    Select  REFERENCE  ALL MOTORS and then choose START.  The tool will initialize all motors of the spectrometer.  If the motors fail to reference, contact a MFCF staff member for help. 

 

Figure 4: Motor Reference Options window.  

SYSTEM  CHECK    

 1. Turn the objective turret so that the scope objectives are in the rear (Figure 5).  This is to prevent accidental 

crashing of the objectives into the sample. 

2. Lower the stage of the microscope using the Z‐position adjustment knobs on the side of the microscope. Pull back the sample clip on the stage, and place the silicon reference standard (stored in the black box on the computer cart) on the stage.  Release the stage clip to secure the glass slide.  

 

 

Figure 5. Lowered sample stage with objectives in the rear. 

3. Move the 5x objective into position so that it is above the silicon reference sample.  

NOTE:   Whenever the sample objective  is changed on the turret, the objective setting needs to be changed  in the Sample Review window.  This will adjust the scale on the video window (see Figure 6b). 

4. On the filter wheels of the microscope, adjust the filter wheels to the following positions (Figure 6a): 

Page 4: Standard Operating Procedure

 4

Upper filter wheel =  2 

Lower filter wheel = 1   This allows for viewing of the sample using the white light source of the microscope and video camera (Figure 6b).      

(a)       (b)   

Figure 6. (a) Filter wheels in positions 2 and 1; (b) Focused silicon reference sample in video window. 

5. Focus on the surface of the sample using the stage height adjustment knob on the side of the microscope.  It may be necessary  to adjust  the F‐stop on  the bottom  left of  the microscope  in order  to  focus on  the surface of clear or highly reflective samples. 

6. Once the sample is in focus, rotate the objective turret to the 20x position and focus the sample using the stage height adjustment knob.  Then rotate the turret to the 50x objective and focus the sample using the stage  height  adjustment  knob.   WARNING:    Adjust  the  stage  height  SLOWLY  so  as  NOT  to  crash  the objective onto the sample, which will damage the objective.  

7. In the software program, select Measurement > NEW Measurement > 532nm‐1800 set‐up.   This set‐up  is the saved measurement set‐up for the silicon reference sample. 

8. On the filter wheels of the microscope, adjust the filter wheels to the following positions to activate Raman collection channel (Figure 7a): 

Upper filter wheel =  ARROW 

Lower filter wheel = 4 

The laser spot will appear in the crosshairs (Figure 7b).  Using the stage height adjustment knob, focus the laser spot. 

(a)        (b)  

Page 5: Standard Operating Procedure

 5

Figure 7. (a) Filter wheels positions for measurement; (b) Centered laser spot in the Video window. 

NOTE:  If the laser spot is outside of the crosshairs, DO NOT ADJUST THE CROSSHAIR POSITION.  Contact a MFCF staff for help in adjusting the laser position. 

9. Select Tools > System Health > System Health Check.   The  tool will check  the alignment position of  the laser,  slits,  and CCD  camera.    If  the  System Health Check PASSES,  then  continue.    If  the  System Health Check FAILS, then an alignment of the system needs to be performed and a Shared Facilities staff member needs to be contacted.  Record this information in the log book. 

10. Choose the RUN icon ‐   ‐ on the toolbar to run the measurement.  The acquisition can be stopped at any time by pressing the ABORT icon.  The acquisition scan will appear in the Spectrum Viewer window. 

11. Select Tools > Calibration > Quick Calibration and the tool will set the offset to the silicon reference peak to 521 l/mm (Figure 8). 

 

Figure 8. Silicon reference sample spectral scan. 

12. After the calibration, lower the stage height using the adjustment knobs on the side of the microscope and rotate the objective turret so that an empty objective position is above the sample.   

13. Release sample clip and remove the silicon reference sample and place  it back  into the storage box.   The tool is calibrated for the processing of samples. 

 

SPECTRAL  ACQUISITION    

 The various types of measurements available for users are as follows: 

Spectral Acquisition – standard method for spectral acquisition 

Image Acquisition – collection of filter spectra.   

Mapping Acquisition  –  collection  of  spectral  acquisitions  over  an  area  of  the  sample.    If  the mapping technique is desired to be used, contact a Shared Facilities staff member. 

NOTE:  The  following  SOP  is  for  a  Spectral Acquisition.    For  instruction on  the Mapping  and  Image Acquisition techniques, please contact a MFCF staff member. 

1. Place sample onto a glass microscope slide located on the sample prep table. 

Page 6: Standard Operating Procedure

 6

2. Turn the objective turret so that scope objectives are in the rear.  This is to prevent accidental crashing of the objectives into the sample. 

3. Lower the stage of the microscope using the stage height adjustment knobs on the side of the microscope. Pull back  the  sample  clip on  the  stage, and place  the glass  slide on  the  stage.   Release  the  stage clip  to secure the glass slide. 

4. Move the 5x objective into position so that it is above the sample. 

NOTE:  Whenever the sample objective is changed on the turret, then the objective setting needs to be changed in the Sample Review window.  This will adjust the scale on the video window. 

5. On the filter wheels of the microscope, adjust the filter wheels to the following positions: 

Upper filter wheel =  2 

Lower filter wheel = 1 

This allows for viewing of the sample using the white light source of the microscope and video camera.   

6. Using the stage height adjustment knob on the side of the microscope, focus on the surface of the sample.  It may be necessary  to adjust  the F‐stop on  the bottom  left of  the microscope  in order  to  focus on  the surface of clear or highly reflective samples. 

7. Once the sample is in focus, rotate the objective turret to the 20x position and focus the sample using the stage height adjustment knob.  Then rotate the turret to the 50x objective and focus the sample using the stage height adjustment knob.  

CAUTION:   Adjust  the  stage height  slowly  so  as  to NOT  crash  the objective  into  the  sample  as damage  to  the objective can occur.  

8. Select Measurement  > NEW  >  Spectral  Acquisition  and  set  the  scan  to  the  desired  parameters  in  the Spectral Acquisition Setup window: 

NOTE: The APPLY button must be pressed to save any changes made to the acquisition parameters. 

       1). RANGE TAB 

(a)  

Page 7: Standard Operating Procedure

 7

(b)  

Figure 9. Range Tab windows for (a) Static scan and (b) Extended scan. 

Grating Scan:  

i. Static Scan: The user sets the center point of the Spectrum Range and the system will set the scan range 500 cm‐1 on the center point.  This scan can have an exposure time of less than one second per accumulation. 

ii. Extended Scan: The user sets  the upper and  lower  limit of  the Spectrum Range and  the spectrometer will scan continuously over the range.  This scan is limited to exposure times of 10 seconds or greater. 

Confocality: sets the sample volume of the sample collected 

i. Standard: Uses a larger volume to increase signal strength. 

ii. High: Reduces the volume and signal strength to increase depth resolution. 

Configuration: 

i. Laser name: 532nm notch 

ii. Grating Type: 1800 l/mm (vis) 

 

2) ACQUISITION TAB 

 

Page 8: Standard Operating Procedure

 8

 

Figure 10. Acquisition Tab window. 

Title and Description:   Name the acquisition and give a description of the scan.   This  information will be saved with the acquisition. 

Exposure Time:  This is the amount of exposure time at the detector where longer exposure times will improve the signal‐to‐noise ratio. 

Accumulations: This  is  the number of scan  repetitions added  together  to  improve signal‐to‐noise ratios and  for  the  removal of  cosmic  rays.    If  the  cosmic  ray  removal box  is  selected, then two additional scans will be processed in addition to the number of accumulations selected 

Objective:  This is the objective selected in the Sample Review window and must be changed every time the user changes the objective manually on the microscope.  

Laser Power:   This  is the percentage of  laser power being used for the scan where higher powers translate into better signal‐noise ratios.  Start with lower laser powers on initial scans if the sample is susceptible to being damaged by the laser. 

Cosmic Ray Removal:  Select this box to have the spectrometer take two additional scans to remove random cosmic peaks 

Close Shutter on Completion: Select this box to close the laser at the end of the acquisition and limit laser exposure on the sample. 

 

3) FILE TAB  

Autosave:    Select  this  box  so  that  the  acquisition  will  be  saved  to  the  folder  with  filename designated by the user.   Select the AUTO  INCREMENT box to save every scan acquisition with a sequential numbered filename.  

 

Page 9: Standard Operating Procedure

 9

 

Figure 11. File Tab window. 

4)  TIMING TAB 

Sample Bleaching: This is used to decrease fluorescence from the sample by exposing the sample to the sample to the  laser prior to taking a measurement.   Select this box and set the time if this issue is occurring on the sample.  

Time Series Measurements:  This allows for the user to set a specified time in between scans. 

 

Figure 12. Timing Tab window. 

5) TEMPERATURE 

This is currently NOT an option to users as a temperature stage is required. 

6) FOCUS TRACK TAB 

This is used to maintain the laser focus position for time and mapping acquisitions.   See Module #7 of user manual  (located on the computer desktop) and/or contact a MFCF staff member  for additional information.  

7) ADVANCED TAB 

Scan Type:  When an Extended Scan has been selected in the RANGE TAB, then the user has the option of using a Step Scan or a Synchroscan (or Continuous Scan).  The recommended 

Page 10: Standard Operating Procedure

 10

option  is the Synchroscan, but  if saturation of the detector  is occurring, then the Step option should be selected so exposure time under 10s can be used. 

Camera Gain:  Select the HIGH sensitivity option. 

Pinhole:  The system has a motorized pin‐hole assembly and this function is used to for improving the profile of the laser beam.  The option IN is recommended. 

Laser  Focus:    This  option  controls  the  beam  expander.    A  setting  of  100%  is  a  completely defocused beam, while 0% is a tightly focused beam.  If the sample being used is subject to damage by the  laser, then a highly defocused beam  is recommended to reduce the power density at the sample.    

Image Capture:  This feature can be used to have white light images of the sample taken before or after an acquisition.    

 

Figure 13. Advanced Tab window. 

9. On the filter wheels of the microscope, adjust the filter wheels to the following positions: 

Upper filter wheel =  ARROW 

Lower filter wheel = 4 

The laser spot will appear in the crosshairs (Figure14).  Using the stage height adjustment knob, focus the laser spot. 

 

Figure 14. Centered laser spot in Video window. 

Page 11: Standard Operating Procedure

 11

NOTE:  If the laser spot is outside of the crosshairs, DO NOT ADJUST THE CROSSHAIR POSITION.  The laser position needs adjusted and a Shared Facilities staff member needs to be contacted. 

10. Turn off the room light. 

11. Choose the RUN icon   on the toolbar to run the measurement.  The acquisition can be stopped at any time by pressing the ABORT icon.  

12. If  another  acquisition  is  required  and  spectral  acquisition  settings  need  to  be  changed,  select Measurement > Set‐up Measurement option.  The scan parameters from the previous acquisition will still be stored in the Spectral Acquisition Set‐up window.  If the scan parameters need to be saved for later use, select Measurement > Save Measurement.  

13. To remove the sample from the microscope stage, lower the stage height using the adjustment knobs on the side of the microscope and rotate the objectives turret so that an empty objective position is above the sample.    

14. Release the sample clip and remove the microscope slide from the stage.  

15. See Module #3 of user manual  for data manipulation  techniques under  the Analyze  tab of  the Wire 3.2 software.  

16. If  another  sample  is  to  be  processed,  then  proceed  back  to  step  #1  of  this  section.    If  all  necessary acquisitions, have been acquired, then proceed to the SYSTEM SHUT DOWN section of the SOP.  

NOTE:    Users  should make  sure  that  they  copy  their  data  off  the  computer  in  a  timely manner.  The  Shared Research Facilities is a multi‐user facility and therefore cannot guarantee that the saved file will not be modified or deleted. 

 

DATA  OPTIMIZATION      

1. To improve signal to noise ratio 

a. Increase exposure time 

b. Increase number of accumulations 

2. To eliminate strong background due to fluorescence 

a. Decrease laser power  

b. Quench fluorescence by exposing sample to incident laser light for a period of time 

c. Change excitation wavelength 

3. To avoid saturated signal 

a. Reduce laser power 

b. Reduce exposure time 

4. To avoid laser ablation on the sample 

a. Use a lower magnification objective lens 

b. Defocus the laser spot 

Page 12: Standard Operating Procedure

 12

c. Reduce laser power 

DATA  ANALYSIS    

NOTE:  Below  lists  two  basic  data  processing  procedures.  Please  refer  to M3‐  Data manipulation module  (on computer desktop) for additional information or contact a MFCF staff member. 

1. Baseline subtraction 

1) Open the spectrum in the Viewer. 

2) Select Processing > Subtract Baseline. A new Viewer opens with the spectrum  in the top half and the results of any baseline subtraction in the lower (Figure 15). 

3) The default baseline is fitted between the two end points of the spectrum. Use left mouse click to add more points to the baseline. 

4) Use right mouse click to get the context menu. Select Properties > Cubic Spline Interpolation from the menu. 

5) Select Accept from the context menu. 

6) You will be asked to keep the correction. NOTE: If you select Yes, you will overwrite the original file. You may need to keep a backup or rename the file. 

 

Figure 15. Baseline subtraction example. 

2. Smoothing 

1) Select Processing > Smooth. 

2) A new window will open with the sample spectrum at the top and the result spectrum below  (Figure 16). 

3) To modify the degree of smoothing, select Properties from the context menu (right mouse click) to see the Smooth Properties window. 

4) You  can  save  the  resulting  smoothed  spectrum  by  select  Processing  >  Smooth  >  Accept  >  Current dataset. 

Page 13: Standard Operating Procedure

 13

 

Figure 16. Curve smoothing example 

 

       3.  Zap 

  Use Zap function to remove stray bands such as cosmic rays from the spectrum. 

1) Open the spectrum of interest. 

2) Select Processing > Zap from the menu. A new Viewer will open with the sample spectrum at the top and the result spectrum below (Figure 17). 

3) The upper spectrum has a zap region between two vertical black lines. Grab each line in turn and adjust the position of the zap region to just enclose the band to be removed. You may need to use the zoom function to isolate the band. 

 

Figure 17. Zap function example. 

Page 14: Standard Operating Procedure

 14

        4.  Curve‐fitting 

1) Select Analysis > Curve Fit to open the curve fit window. 

2) Zoom into a region that contains the band and some baseline data at either side. 

3) Use the mouse to position the approximate centre of the band. 

4) Click to add the band and repeat for the centers of other bands if there are multiple bands. 

5) You can adjust the center or width markers for each band using the mouse and cursor to get a better rough fit. 

6) Use mouse right click to get the menu and select Start Fit to fit the peak (Figure 18). 

7) You can save the curve fit file from the context menu. 

 

Figure 18.  Curve fitting example. 

 

5. Peak pick 

1) Select Analysis > Peak pick to open the curve fit window (Figure 19). 

2) You may adjust the thresholds by selecting Analysis > Autoset thresholds.  

 

Page 15: Standard Operating Procedure

 15

  

Figure 19.  Peak pick example.    

SYSTEM  SHUT  DOWN  

1. Turn the laser power supply key to the OFF position.  The laser light indicator will turn off. 

2. Switch the laser power supply toggle switch to the OFF position.  The power light indicator on the front of the power supply will turn off.  

3. Toggle the power switch on the right side of the RAMAN spectrometer set‐up to OFF. 

4. Turn OFF the power switch on the front‐side of the PRIOR control box. 

5. Turn OFF the power switch on the left‐side of the white light source for the microscope. 

6. CLOSE the WIRE 3.2 software on the computer desktop. 

7. Check that all required parameters are recorded in the log book. Sign out on the log book. 

8. Clean up work area by doing the following:  

a. Safety glasses are returned to the proper case on top of the computer cart. 

b. The silicon reference sample  is placed back  in the storage case  located on top of the computer cart. 

c. Remove the WARNING: LASER IN USE sign from the front door of the lab. 

9. Properly dispose of any used materials – i.e. alpha wipes, glass slides, etc. 

10. Sign out of the FOM. 

   

Page 16: Standard Operating Procedure

 16

EMERGENCY  PROCEDURES  

If no one is available and the machine is not acting as expected, the user should do the following: 

Press ABORT on the WIRE 3.2 software. 

Turn the key to the laser power supply to the OFF position. 

Turn off the x‐rays by pressing the red OFF button on the power controller.  

Do not  leave  the machine  running  in an abnormal  state.    If  the machine cannot be placed  in  the default  state, immediately contact: 

Primary Staff Contact:   Weiqiang Ding      (304) 685‐1938    [email protected]  

Secondary Staff Contact:   Harley Hart      (412) 443‐1514  [email protected]  

Then,  if possible,  the user should stay by  the  tool while  trying  to contact a Shared Facilities staff member.    If  it 

becomes necessary to leave the tool then the user should leave a large, legible note at the RAMAN stating the tool 

is DOWN.  

The user should then complete a Tool Issue Report form and submit this form to ESB G75D.  The user should also 

add comment using the FOM software indicating the tool status. 

If a dangerous situation is evident (smoke, fire, sparks, etc), ONLY if it is safe to do so, the user should turn off 

power to the system by switching the toggle switch on the right side of the tool and the laser power supply key 

and toggle switch to the OFF position.  The user should then notify all other persons within the Hodges Hall 325 

lab to evacuate and leave the lab immediately. The user should then contact proper emergency personnel from a 

safe place.