Standard Data Analysis -Integration
ChemStation - Level 1 - Training
![Page 2: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/2.jpg)
Agilent Restricted
Integration locates the peaks in a signal and calculates their
size
What is Integration
![Page 3: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/3.jpg)
Agilent Restricted
The Integration Process consists of the following:
• construct initial baseline
• peak recognition – Start and end Time
• finds the apex of the peak
• baseline allocation (construction)
• peak area measurement
Integration Process
![Page 4: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/4.jpg)
Agilent Restricted
1 Slope and Curvature within limit
2 Slope and curvature above limit: peak?
3 Slope remains above limit: peak recognized
4 Curvature becomes negative: front inflection
5 Negative slope: Apex
6 Curvature becomes positive: rear inflection point
7 Slope and curvature within limit: end of peak?
8 Slope and curvature within limit: end of peak
5
1
2
3
4 6
7
8
Start, End and Apex of a Peak
![Page 5: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/5.jpg)
Agilent Restricted
Fit highest data point and those on each sideto a quadratic equation, solve for highest point
Determining Peak Apex
![Page 6: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/6.jpg)
Agilent Restricted
BB
BV
VB
BB
BV
VB
Baseline Tick Marks
The default baseline is constructed as a series of straight line segments between: the start of the baseline, the tick marks, and the end of the peak.
Default Baseline Construction
![Page 7: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/7.jpg)
Agilent Restricted
Standard or Enhanced Integrator The ChemStation includes two integrators:
the Enhanced Integrator (default) and the Standard Integrator (Genie)
New methods by default use the Enhanced Integrator and can not be down graded to the Standard Integrator
Old methods can be
updated to use the Enhanced Integrator ormaintained using the Standard Integrator so customers
can continue using validated methods.
The default method “defoldlc.m” has the Standard Integrator
The default method “def_lc.m” has the Enhanced Integrator
![Page 8: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/8.jpg)
Agilent Restricted
Both integrators co-exist so you may continueto use validated methods
The Enhanced Integrator offers better: baseline allocation on wandering baselines rejection of noise peaks through the use of an initial peak
height parameter control of integration start and stop marks peak allocation on noisy signals peak shoulder allocation through second derivative
calculations
Standard or Enhanced Integrator
![Page 9: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/9.jpg)
Agilent Restricted
Loading Signals
Select path, data file name and channel
Select
or ‘Load Signal’ from File Menu
Show/Hide the Signal Information
Choose to load all available or selected signals only
![Page 10: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/10.jpg)
Agilent Restricted
You may also integrate by:
•Selecting Integrate or Auto Integrate• from the Integration menu.•Selecting the Integration Tool.•Running a method where the Run• Time Checklist includes Data Analysis
When a Signal is Loaded, Integration May Occur Automatically
![Page 11: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/11.jpg)
Agilent Restricted
Initial Values
ThresholdPeak WidthArea RejectShoulders
Standard Integrator - Initial Values
![Page 12: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/12.jpg)
Agilent Restricted
What timed events could you use to fix the
integration below?
Standard Integrator - Timed Events
What timed events could you use for this integration?
![Page 13: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/13.jpg)
Agilent Restricted
Auto Integrate determines the optimum values for the initial threshold, peak width, and area reject by integrating the chromatogram twice.
Threshold - based on signal range and averaged baseline noise.Peak Width - proportional average of the peak widths of the first five peaks.Area Reject - 1/1000st of the greatest area found.
Auto Integration - Standard Integrator
![Page 14: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/14.jpg)
Agilent Restricted
Select Integration Tasks Integrate CurrentChromatogram
Auto Integrate
Integration Events
Manual IntegrationDraw Peak Baseline
Manual IntegrationNegative Peaks
Manual IntegrationTangent Skim
Manual IntegrationSplit Peaks
Manual IntegrationRemove Peaks
Integration Tools - Standard Integrator
![Page 15: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/15.jpg)
Agilent Restricted
The Enhanced Integrator offers better: baseline allocation on wandering baselines rejection of noise peaks through the use of an initial peak
height parameter control of integration start and stop marks peak allocation on noisy signals peak shoulder allocation through second derivative
calculations
Enhanced Integrator
![Page 16: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/16.jpg)
Agilent Restricted
You may update methods developed on previous revisions of the ChemStation to the new Enhanced Integrator. Once you have updated the new method, you cannot revert back to the old integrator.
Enhanced Integrator
![Page 17: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/17.jpg)
Agilent Restricted
Enhanced Integrator UI Define integration parameters just by selecting and clicking on the chromatogramThere are two sets of
integration events :
- intial events
- time-based
Use: Slope Sensitivity to define peak sensitivityPeakwidth to set an initial sampling interval for the
integrator to distinguish peaks from baseline noise
Area Reject to filter small peaksHeight Reject to set noise rejection
![Page 18: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/18.jpg)
Agilent Restricted
Advanced Baseline
12 13 14 15 16 17 18 19 20
Abso
rbance
Time [min.]
Options OffAdvanced Integration
Options OnAdvanced Integration
Peak
Baseline
Peak
AreaUnassigned
![Page 19: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/19.jpg)
Agilent Restricted
Using Initial Parameters
Peak Width
HeightReject
![Page 20: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/20.jpg)
Agilent Restricted
Add Timed Events
Delete a Timed Event
Enhanced Integrator - Timed Events
![Page 21: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/21.jpg)
Agilent Restricted
Solvent peak (Skim peak)
Tangent Peak
Straight Line Calculation
Exponential Calculation
Tangent Skim
![Page 22: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/22.jpg)
Agilent Restricted
Shoulders occur when two peaks are so close together that no valleyexists between them
Peak with shoulders
Tangent Shoulders Drop Line Shoulders
Shoulder Detection
![Page 23: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/23.jpg)
Agilent Restricted
Enhanced Integrator - Auto Integration
Examines beginning and end regions to estimate noise.
Assigns initial Slope Sensitivity and Height Reject.
Assigns temporary Peak Width value for first pass integration.
Sets Area Reject to zero.
Performs trial integration, may be repeated several times.
Calculates Peak Width based on early eluting peaks.
Refines Slope Sensitivity and Height Reject.
Computes Area Reject from initial Peak Width and noise level.
![Page 24: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/24.jpg)
Agilent Restricted
Integration TasksIntegrate CurrentChromatogram
Auto IntegrateIntegration Events
Set Baseline Now
Set Baseline Hold
Set Tail Tangent Skim
Set IntegrationSet Slope Sensitivity
Set Area Reject
Set Fixed Peak Width
Set Detect Shoulders
Enhanced Integrator - Tool Bar
Use Auto Integrate to find suitable integration events
Only visible with
integration events open
![Page 25: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/25.jpg)
Agilent Restricted
Save Events to Method
Save Integration Events as Part of a Method
![Page 26: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/26.jpg)
Agilent Restricted
Manual Integration
Manual integration events
![Page 27: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/27.jpg)
Agilent Restricted
Manual Integration
Use Cursor tools to perform manual integration by selecting the required action and dragging a line over or clicking on the chromatogram (depending on the task)
Zoom in using the zooming tools to better view your baseline
The new integration results are immediately visible in the user interface
If your chromatography is so demanding that you can’t use the the automatic integration process
Use the manual integration tool set
![Page 28: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/28.jpg)
Agilent Restricted
Option 1: Save Manual Integration to Method
Select Integration, Copy Manual Events to Method
Save Manual Integration - To the method
![Page 29: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/29.jpg)
Agilent Restricted
Step 1: Select View, Windows Functions, Save Window
Save Manual Integration - To the Raw Data File
![Page 30: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/30.jpg)
Agilent Restricted
Note: manual changes are stored in the raw data file directory.
Step 2: Click OK to use default name or rename
Save Manual Integration - To the Raw Data File
![Page 31: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/31.jpg)
Agilent Restricted
Select View, Windows Functions, Load Window
To Recall Manual Integration From Data File
![Page 32: Standard Data Analysis - Integration ChemStation - Level 1 - Training.](https://reader035.fdocuments.in/reader035/viewer/2022081506/56649c7e5503460f94933d41/html5/thumbnails/32.jpg)
Agilent Restricted
Lab Time !!!!! Lab 5 - Integration