SPOT Actuator User Guide - Optalex
Transcript of SPOT Actuator User Guide - Optalex
SPOT Actuatori
Q U A L I T Y C U S T O M E R S O L U T I O N S
SPOT ACTUATORUSER GUIDEPUBLICATION NO 812250LANGUAGE: ENGLISH
Health and Safety InformationRead all of the instructions in this booklet - including all the WARNINGS and CAUTIONS - before using this product. If there is any instruction which you do not understand, DO NOT USE THE PRODUCT.
Safety SignsWARNINGIndicates a potentially hazardous situation which, if not avoided, could result in death or personal injury.
CAUTIONIndicates a potentially hazardous situation which, if not avoided, could result in minor or moderate injury to the user or users, or result in damage to the product or to property.
NOTEIndicates a potentially hazardous situation which, if not avoided, could result in damage or the loss of data.
Signs and Symbols used on equipment and Documentation
Caution, risk of electric shock.
Caution, attention to possibility of risk of damage to the product, process or surroundings. Refer to instruction manual.
Caution, hot surface.
Protective Conductor Terminal.
Observe precautions for handling electrostatic discharge sensitive devices.
Equipment OperationUse of this instrument in a manner not specified by Land Instruments International may be hazardous. Read and understand the user documentation supplied before installing and operating the equipment.The safety of any system incorporating this equipment is the responsibility of the assembler.
Protective Clothing, Face and Eye ProtectionIt is possible that this equipment is to be installed on, or near to, machinery or equipment operating at high temperatures and high pressures. Suitable protective clothing, along with face and eye protection must be worn. Refer to the health and safety guidelines for the machinery/equipment before installing this product. If in doubt, contact Land Instruments International.
Electrical Power SupplyBefore working on the electrical connections, all of the electrical power lines to the equipment must be isolated. All the electrical cables and signal cables must be connected exactly as indicated in these operating instructions. If in doubt, contact Land Instruments International.
StorageThe instrument should be stored in its packaging, in a dry sheltered area.
UnpackingCheck all packages for external signs of damage. Check the contents against the packing note.
Lifting InstructionsWhere items are too heavy to be lifted manually, use suitably rated lifting equipment. Refer to the Technical Specification for weights. All lifting should be done as stated in local regulations.
IMPORTANT INFORMATION - PLEASE READ
Health and Safety InformationRead all of the instructions in this booklet - including all the WARNINGS and CAUTIONS - before using this product. If there is any instruction which you do not understand, DO NOT USE THE PRODUCT.
Safety SignsWARNINGIndicates a potentially hazardous situation which, if not avoided, could result in death or personal injury.
CAUTIONIndicates a potentially hazardous situation which, if not avoided, could result in minor or moderate injury to the user or users, or result in damage to the product or to property.
NOTEIndicates a potentially hazardous situation which, if not avoided, could result in damage or the loss of data.
Signs and Symbols used on equipment and Documentation
Caution, risk of electric shock.
Caution, attention to possibility of risk of damage to the product, process or surroundings. Refer to instruction manual.
Caution, hot surface.
Protective Conductor Terminal.
Observe precautions for handling electrostatic discharge sensitive devices.
Equipment OperationUse of this instrument in a manner not specified by Land Instruments International may be hazardous. Read and understand the user documentation supplied before installing and operating the equipment.The safety of any system incorporating this equipment is the responsibility of the assembler.
Protective Clothing, Face and Eye ProtectionIt is possible that this equipment is to be installed on, or near to, machinery or equipment operating at high temperatures and high pressures. Suitable protective clothing, along with face and eye protection must be worn. Refer to the health and safety guidelines for the machinery/equipment before installing this product. If in doubt, contact Land Instruments International.
Electrical Power SupplyBefore working on the electrical connections, all of the electrical power lines to the equipment must be isolated. All the electrical cables and signal cables must be connected exactly as indicated in these operating instructions. If in doubt, contact Land Instruments International.
StorageThe instrument should be stored in its packaging, in a dry sheltered area.
UnpackingCheck all packages for external signs of damage. Check the contents against the packing note.
Lifting InstructionsWhere items are too heavy to be lifted manually, use suitably rated lifting equipment. Refer to the Technical Specification for weights. All lifting should be done as stated in local regulations.
IMPORTANT INFORMATION - PLEASE READ Contact UsUK - DronfieldLand Instruments InternationalTel: +44 (0) 1246 417691Email: [email protected]: www.landinst.com
USA - PittsburghAMETEK Land, Inc.Tel: +1 412 826 4444Email: [email protected] Web: www.ametek-land.com
ChinaAMETEK Land China Service Tel: +86 21 5868 5111 ext 122Email: [email protected]: www.landinst.com
IndiaAMETEK Land India ServiceTel: +91 - 80 67823240 Email: [email protected]: www.landinst.com
For further details on all AMETEK Land offices, distributors and representatives, please visit our websites.
Return of Damaged GoodsIMPORTANT If any item has been damaged in transit, this should be reported to the carrier and to the supplier immediately. Damage caused in transit is the responsibility of the carrier not the supplier.DO NOT RETURN a damaged instrument to the sender as the carrier will not then consider a claim. Save the packing with the damaged article for inspection by the carrier.
Return of Goods for RepairIf you need to return goods for repair please contact our Customer Service Department. They will be able to advise you on the correct returns procedure.Any item returned to Land Instruments International should be adequately packaged to prevent damage during transit.You must include a written report of the problem together with your own name and contact information, address, telephone number, email address etc.
Design and Manufacturing StandardsThe Quality Management System of Land Instruments International is approved to BS EN ISO 9001 for the design, manufacture and on-site servicing of combustion, environmental monitoring and non-contact temperature measuring instrumentation.
Approvals apply in the USA
Operation of radio transmitters, telephones or other electrical/electronic devices in close proximity to the equipment while the enclosure doors of the instrument or its peripherals are open, may cause interference and possible failure where the radiated emissions exceed the EMC directive. The protection provided by this product may be invalidated if alterations or additions are made to the structural, electrical, mechanical or pneumatic parts of this system. Such changes may also invalidate the standard terms of warranty.
CopyrightThis manual is provided as an aid to owners of Land Instruments International’s products and contains information proprietary to Land Instruments International. This manual may not, in whole or part, be copied, or reproduced without the expressed written consent of Land Instruments International Ltd.
Copyright © 2016 Land Instruments International.
MARCOM0311, Issue 7, 01 November 2016
Approvals apply in India
Fixed Spot ThermometersSPOT Actuator
Blank
Fixed Spot Thermometers SPOT Actuator
Contents
1 Introduction 1-1
1.1 About the SPOT Actuator 1-1
2 System Components 2-1
3 Installation Drawings 3-1
4 Operating Instructions 4-1
4.1 ConfiguringyourSPOTThermometertoworkwiththeActuator 4-14.2 Powering the Actuator 4-14.3 Controlling the Actuator 4-14.4 Control by PLC using the analogue cable 4-24.4 Control from the Handset 4-24.5 Ethernet Control 4-3
INTRODUCTION
1
1 - 1
SPOT ActuatorFixed Spot Thermometers
User Guide
1 IntroductionThis User Guide gives you instructions on how to install and operate a LAND SPOT Actuator, designed to be used in conjunction with the SPOT AL EQS pyrometer.It is important to read and understand these instructions before installing the equipment.
1.1 About the SPOT ActuatorThe SPOT Actuator provides remotely controlled target alignment of a SPOT pyrometerforindustrialprocessingapplications,specificallyforaluminiumapplications using the SPOT AL EQS (Aluminium Extrusion, Quench and Strip)pyrometer.The SPOT Actuator is a motorised unit which is designed for use at press exit and quench exit positions on aluminium extrusion presses, to align the ALEQSpyrometerwiththeprofile.Itiscapableofoperatinginautomaticoroperator-controlled modes to suit the needs of an individual aluminium plant.Compact,quickandeasilyconfigurable,theSPOTActuatorisalsodesignedfor mounting the SPOT AL EQS pyrometer onto the billet reheat furnace exit, or before the extrusion press entrance face, to scan the SPOT AL EQS pyrometeralongthelengthofthebillet,generatingatemperatureprofile.
Features• Direct Communications - remote controlled or automatic target
alignment by Actuator of SPOT pyrometer;• Accurate Alignment - taking 900 measurement points over 90
degrees,alignsonoptimummeasurementpositiononprofile;• Multiple Locations - reheat furnace exit, press entrance, press exit
and quench exit;• Dynamic Press Speed Control - temperature measurement fed back
to press control system• Rapid Alignment - accurate repositioning following die changes• Dynamic Tracking-smallandwell-definedmeasurementspot
combined with its rapid 15 ms response speed, facilitates precise tracking
• Compact, Quick and Easily Configurable
SYSTEM COMPONENTS
2
2 - 1
Fixed Spot Thermometers
User Guide
SPOT Actuator
2 System Components This section describes the components of the SPOT Actuator system. Before setting up your system, check that you have been supplied with all components. If any components are missing or damage, contact Ametek Land Instruments.
Item Land Part Nº NotesSPOT Actuator 811824 Mount in a suitable
location via the 4 x 8.5 mm mounting holes. Refer to the Installation Drawing, Fig. ?-?.
Unscrew the two M8 x 16 Hex Screws (with spring washers and plain washers) in order to attach the SPOT Light Purge (808015).
SPOT Light Purge 808015 Attach to the Actuator via the two M8 x 16 Hex Screws (and washers).
Refer to the SPOT Light Purge Installation Guide (supplied) forinstructionsonfittingtheSPOT pyrometer into the purge
Actuator Handset 812025 Used to control the Actuator remotely.
Connect to the Actuator via the Handset Cable provided:
2 m Cable Part Nº 8120345 m Cable Part Nº 812035 10 m Cable Part Nº 812036
Fixed Spot Thermometers
User Guide2 - 2
SPOT Actuator
Item Land Part Nº NotesHandset Cable 812034 (2 m)
812035 (5 m)812036 (10 m)
M12 Ethernet M-M X Coded Cable Used to connect the handset to the Actuator.
Actuator Ethernet Cable 812033 M12 Ethernet M-F D Coded Cable 0.5 m
Used to connect the Actuator to the digital input/output of the pyrometer.
Actuator Analogue Cable 812037 0.5 m M-F Analogue Cable
Used to connect the Actuator to the analogue input/output of the pyrometer.
Fig. 2-2 Actuator System Overview
3INSTALLATION DRAWINGS
3 Installation Drawings
OPERATING INSTRUCTIONS
4
Fixed Spot ThermometersSPOT Actuator
Blank
4 - 1
Fixed Spot Thermometers
User Guide
SPOT Actuator
4 Operating Instructions
4.1 Configuring your SPOT Thermometer to work with the ActuatorRefer to the User Guide supplied with your SPOT Thermometer for further instructions on how to access, and adjust settings in the SPOT Thermometer software menu.To work with the actuator, the SPOT Thermometer must be running Firmware version 1018 or later.In the I/O Menu of the SPOT Thermometer software, set the Scan option to On.
4.2 Powering the ActuatorThe actuator can be powered by 18 to 36 V DC or Power over Ethernet (POE+). In both cases, a single cable can be used to power and control the actuator.For DC power, the analogue cable is required, and the Ethernet cable is required for power over Ethernet. The actuator and SPOT typically draw 25 W.If both cables are available, the actuator can be powered one way and controlled the other. Both sources of power can be connected at once.
4.3 Controlling the ActuatorTheactuatorcanbecontrolledinseveraldifferentways:
Analogue Only (1 cable)The PLC triggers the scan with a 24 V digital signal and reads the 4 to 20 mA temperature signal from the SPOT.
Analogue and HandsetAs above, also local control.
Ethernet Only (1 wire)Connect to the Actuator for access to both SPOT and Actuator on separate IP Addresses.Thisallowsyoutoconfigureallsettings,triggerascanandreadscan data using Modbus TCP
Ethernet and HandsetAs above, also local control.
Analogue, Ethernet and HandsetTrigger a scan by any method
Fixed Spot Thermometers
User Guide4 - 2
SPOT Actuator
4.4 Control by PLC using the analogue cableThe analogue cable must be connected from the control system to the actuator if 24 V digital control or 4 to 20 mA analogue output are required, or to power the unit if Power over Ethernet (PoE) is not available.
24V / 0VThe actuator accepts 18 to 36 V. The actuator and SPOT typically draw 25 W.
+mA Out / -mA OutThe current SPOT temperature can be read just as if the cable was connected directly to the instrument.
Relay 1a / Relay 1bThe relay is normally open unless the actuator is in a fault state.
Trigger In+ / Trigger In-The trigger is used to remotely initiate a scan. Connect 24 V to Trigger In+ and 0 V to Trigger In-. The scan is triggered by a rising edge that is sustained foratleast10ms.UnlesstheactuatorhasbeenconfiguredasaBilletScanner, it will automatically calculate and go to the centre of the hottest hot region at the end of the scan.
4.4 Control from the HandsetUse the Left and Right arrow buttons to move the actuator. A short press moves it a single step, holding the button will make it move further.Press the circular button to switch on the LED (short press). To trigger a Scan & Go, press and hold the button for approximately 2 seconds. Unless theactuatorhasbeenconfiguredasaBilletScanner,itwillautomaticallycalculate and go to the centre of the hottest hot region at the end of the scan.The bar of LEDs shows the current position of the actuator within the set scanrange.TheLEDisnormallygreen,flashingslowly.Theflashrateincreases during movement and when the SPOT LED is switched on. The LED goes yellow if the unit is under automatic control (e.g. during a Scan & Go), and red under a fault condition.
Move rightShort Press - Move 1 stepPress & Hold - Move continuously
Short Press - Switch on LEDPress 2 Seconds - Scan & Go
Move leftShort Press - Move 1 step
Press & Hold - Move continuously
4 - 3
Fixed Spot Thermometers
User Guide
SPOT Actuator
4.5 Ethernet ControlThe Ethernet cable must be connected from the control system to the actuator for Ethernet control and/orPoE
4.5.1 Control from the Actuator WebserverThe actuator contains a network switch, so that by connecting to the actuator over Ethernet, it is also possible to connect to the SPOT thermometer either by Modbus TCP or to its Webserver. The SPOT and Actuator are both Modbus slaves/servers and have separate IP addresses. Theactuatorwebserverallowsbasicsettingstobeconfiguredonsiteandallows a scan to be triggered and viewed.
4.5.1.1 Settings
ScanThe start and end points of the scan are set relative to a nominal zero position at the mid-point of the motor’s range of movement. The actuator moves in steps of 0.1 degrees but the scan range is set in integer numbers of degrees.There are two modes of scan. If ‘Hotspot Detect’ is selected, the actuator willanalysescandataandgotothecentreofthehottestHotspotitfinds.AHotspotisdefinedasaregionwherethetemperatureisabovetheThresholdvalue for at least 3 points. If Threshold is set to 0, an automatic algorithm is used whereby the Threshold value is set to the halfway point between the maximum and minimum temperatures recorded during the scan.
Network
TheIPaddress,GatewayandSubnetMaskareallconfigurable.Powercyclethe actuator to reinitialise with the new settings.For connection to the actuator, your device must have a similar but not identicalIPaddress,withintheconfinesoftheSubnetMask.Intheexampleabove,thefirsttwonumbersoftheIPaddressmustbethesamebutatleastoneofthelasttwonumbersmustbedifferentontheconnectingdevice.
Site
If a PLC is connected to read the 4 to 20 mA output from the actuator on startup, the actuator will automatically detect it and the check-box will be ticked automatically. It is only necessary to manually tick the box if an external current loop is connected after power up.The invert handset direction check-box may need to be ticked according to the orientation of the manual controls relative to the actuator.
Fixed Spot Thermometers
User Guide4 - 4
SPOT Actuator
SPOT
The actuator reads the 4 to 20 mA output from the SPOT thermometer, so the 4 mA and 20 mA equivalent temperatures must match those of the SPOT for correct interpretation of temperature. 200 C and 700 C are standard limits fortheALEQSthermometer.HoweveritispossibletoconfigurebothSPOTandactuatortouseadifferentrangeifrequiredforinteractionwithotherconnecting devices.
4.5.1.2 Scan Interface
Press the scan button to trigger a scan. Once the scan has completed, the graph will be updated with the scan data. Click on the ‘<’ or ‘>’ label at the bottom right to invert the graph if necessary to match the real orientation on site.The current position and temperature are updated on page refresh, on update of the settings, and on completion of a scan. If scan mode is ‘Hotspot Detect’,thecurrentpositionandtemperaturewillreflectthecentrepositionof the hottest Hotspot.
4 - 5
Fixed Spot Thermometers
User Guide
SPOT Actuator
4.5.2 Modbus TCPThe actuator contains a network switch, so that by connecting to the actuator over Ethernet, it is also possible to connect to the SPOT thermometer either by Modbus TCP or to its Webserver. The SPOT and Actuator are both Modbus slaves/servers and have separate IP addresses. The actuator would normally be controlled and the scan data read out using Modbus TCP.The actuator is a Modbus TCP slave/server, so the control system must be a Modbus TCP Master. Alternatively an Anybus X-Gateway can be used to interprettoothercommunicationsprotocolssuchasProfibus,EthernetIP.The actuator position is controlled in steps of 0.1 degrees, over at least 90 degrees of movement. For ease of integer-based arithmetic, these are available as integer steps centred at position 600. The position displayed as 0 degrees on the webserver corresponds to Modbus value of 600 for the CurrentPosition and SetPosition registers. AllsettingsandoutputsareconfiguredasModbusHoldingRegisters,asforthe SPOT thermometer. The settings table is shown overleaf.
Fixed Spot Thermometers
User Guide4 - 6
SPOT Actuator
Nº
Nam
eR
ead
/W
rite
Def
ault
V
alu
esD
escr
ipti
on
1Fi
rmw
are
Vers
ion
rVersionnumberofcurrentactuatorfirmware
2Er
rorC
ode
r0x01=flashwriteerror;0x02=flashreaderror;0x04=servo
over
curr
ent
erro
r; 0
x08
rang
e re
strict
ed3
Lim
itSta
rtPo
sitio
nr
Posi
tion
of li
mit
switc
hes
deno
te t
he m
axim
um p
ossi
ble
scan
ran
ge4
Lim
itEnd
Posi
tion
r5
Pyro
Cur
rent
Tem
pr
Cur
rent
tem
pera
ture
mea
sure
d by
the
SPO
T th
erm
omet
er6
Cur
rent
Posi
tion
rCur
rent
pos
ition
of
the
actu
ator
7H
otSpo
tPos
ition
rU
nles
s th
e ‘b
illet
sca
n’ o
ptio
n is
set
, at
the
end
of
a sc
an,
the
data
isanalysedtofindthehotspotposition-thecentreofthehottest
hot
spot
. Th
is r
egis
ter
is u
pdat
ed a
nd t
he a
ctua
tor
mov
ed t
o th
at
posi
tion.
8W
antP
ositi
onr
This
reg
iste
r is
the
des
tinat
ion
of t
he la
test
mov
e.9
Cen
treP
ositi
onr
The
cent
re p
oint
bet
wee
n th
e lim
it sw
itche
s.10
DH
CP
Not
yet
ava
ilabl
e
11,1
2Sub
netM
ask
rw25
5.25
5.0.
0
Networkconfigurationparameters
13,1
4Cur
rent
IP
Add
ress
[2]
r
15,1
6Set
IPAdd
ress
[2]
rw10
.1.1
0.60
17,1
8G
atew
ayAdd
ress
[2]
rw10
.1.1
0.1
19-2
1M
AC A
ddre
ss[3
]r
22Te
mpa
t4m
Arw
200
Thes
e re
gist
ers
mus
t be
set
to
mat
ch t
he s
ettin
gs o
n th
e SPO
T th
erm
omet
er I
/O m
enu,
in e
ither
C o
r F.
Her
e th
ey a
re s
et t
o m
atch
th
e de
faul
ts f
or t
he A
L EQ
S23
Tem
pAt2
0mA
rw70
024
Set
Posi
tion
rw60
0Cha
nge
this
reg
iste
r to
for
ce t
he a
ctua
tor
to m
ove
to t
he n
ew v
alue
.25
Bill
etSca
nrw
0Set
ting
Bill
etSca
n to
1 m
akes
the
act
uato
r m
ove
quic
kly
back
to
the
beginningofthescanreadyfornexttime,ratherthanfindingand
mov
ing
to t
he h
otsp
ot.
26Sca
n&G
o Com
man
drw
0Set
Sca
n&G
o to
1 t
o in
itiat
e a
scan
.
4 - 7
Fixed Spot Thermometers
User Guide
SPOT Actuator
Nº
Nam
eR
ead
/W
rite
Def
ault
V
alu
esD
escr
ipti
on
27Sca
nSta
rtrw
300
The
star
t an
d en
d po
sitio
ns o
f th
e sc
an.
Thes
e ca
n be
set
eith
er w
ay
roun
d: S
canS
tart
gre
ater
tha
n Sca
nEnd
sca
ns in
the
oth
er d
irec
tion.
28Sca
nEnd
rw90
029
Hot
Spo
tThr
esho
ldrw
300
Hotspotsareidentifiedasregionsofthescanbufferthatareabove
this
val
ue.
Deg
rees
C o
r F
can
be u
sed
as lo
ng a
s th
e un
its m
atch
th
ose
of t
he 4
mA a
nd 2
0 m
A t
empe
ratu
res
abov
e. S
et t
o 0
if yo
u w
ish
to a
utom
atic
ally
cal
cula
te a
thr
esho
ld m
idw
ay b
etw
een
the
high
andlowextremesofthescanbuffermeasurements.
30D
irec
tion
rw0
Set
to
1 to
inve
rt t
he le
ft a
nd r
ight
key
s of
the
han
dset
.31
Cur
rent
Loop
back
rw1
This
reg
iste
r is
nor
mal
ly s
et a
utom
atic
ally
, de
pend
ing
on w
heth
er
the
syst
em d
etec
ts t
hat
a PL
C o
r ot
her
curr
ent
loop
rea
ding
dev
ice
has
been
con
nect
ed t
o th
e ou
tput
sid
e of
the
act
uato
r. I
f no
PLC
is
conn
ecte
d, t
he r
egis
ter
mus
t be
set
to
1 to
cre
ate
a cu
rren
t lo
op b
y w
hich
SPO
T te
mpe
ratu
re c
an b
e re
ad in
side
the
act
uato
r.32
Spo
tLED
rw0
Set
to
1 to
sw
itch
on t
he S
POT
LED
33Che
ckLi
mits
rw0
Set
to
1 to
che
ck t
he f
ull r
ange
of
oper
atio
n of
the
act
uato
r34
Fact
oryS
ettin
gsrw
0Set
to
1 to
res
tore
the
act
uato
r’s f
acto
ry d
efau
lt se
ttin
gs35
C o
r F
rwC
Set
to
C o
r F
to c
hang
e th
e un
its u
sed
for
tem
pera
ture
val
ues
disp
laye
d on
the
web
serv
er36
-40
Not
use
d41
-45
Cal
ibra
tion
Pass
wor
d [5
]rw
100-
1299
ScanBuffer
rTe
mpe
ratu
re d
ata
mea
sure
d du
ring
a s
can.
If
the
scan
was
fro
m
posi
tions
300
- 9
00,
then
the
tem
pera
ture
s w
ill b
e st
ored
in
ScanBuffer[300]-ScanBuffer[900].Valuesarestoredas10x
degr
ees
C o
r F.
Fixed Spot Thermometers
User Guide4 - 8
SPOT Actuator
Blank
PRODUCT WARRANTYThank you for purchasing your new product from Land Instruments International. This Land manufacturer’s ‘back-to-base’ warranty covers product malfunctions arising from defects in design or manufacture. The warranty period commences on the instrument despatch date from the Land Instruments International Ltd. factory in Dronfi eld, UK.
36 MONTHS WARRANTYBuilding upon the reputation for reliability and longevity that System 4 and UNO thermometers have earned, Land are delighted to be able to provide our customers with an industry-leading 36 month warranty for the following products:-• SPOT thermometers, accessories and mountings and special instruments based on SPOT.• System 4 thermometers, processors, accessories and mountings and special instruments
based on System 4.• UNO thermometers, accessories and mountings and special instruments based on UNO.• Application-dedicated processors based on LANDMARK® Graphic.• ABTS/S and ABTS/U• FTS• VDT/S and VDT/U• DTT• FLT5/A
This 36 month warranty is provided as standard for all orders for the products listed above received from 1st May 2002.We believe that our customers expect us to set the standard in terms of performance, quality, reliability and value for money. This 36 months warranty, as a part of an on-going program of continuous improvement, is just one way in which Land strive to maintain our position as the temperature measurement partner of choice.
24 MONTHS WARRANTYThe following Land Instruments International products are provided with a 24 months warranty:
• ARC• FTI-E
• NIR
12 MONTHS WARRANTYAll Land Instruments International products not provided with either a 36 month or 24 month warranty (see lists above), are provided with a 12 months warranty.
PRODUCT WARRANTYEXCLUSIONS FROM WARRANTY
It should be noted that costs associated with calibration checks which may be requested during the warranty period are not covered within the warranty.Land reserve the right to charge for service/calibration checks undertaken during the warranty period if the cause is deemed to fall outside the terms of the warranty.This Land manufacturer’s warranty does not cover product malfunction arising from:-• incorrect electrical wiring.• connection to electrical power sources outside the rating of the product.• physical shock (being dropped, etc.) and impact damage.• inappropriate routing, support, physical shock & strain protection, etc. of the lightguide (Fibroptic
thermometers only).• environmental conditions exceeding the IP / NEMA rating of the product.• environmental conditions outside the Ambient Temperature, Humidity and Vibration rating of the product.• environmental contamination (solvent vapours, deposition of airborne contamination, cooling liquids of
non-neutral pH, etc.).• overheating as a result of interruption of water/air fl ow through cooling jackets or of incorrect installation.• inappropriate modifi cation of product (drilling holes in thermometer bodies, etc.).• inappropriate recalibration which results in product calibration being taken outside specifi cation.• improper resealing of thermometer following parameter adjustment (UNO, FLT5/A, etc.).• attempted repair by a non-Land-authorised repair centre.Land Instruments International Ltd • Dronfield S18 1DJ • England • Tel: +44 (0) 1246 417691 • Fax: +44 (0) 1246 410585 Email: [email protected] • www.landinst.comAMETEK Land, Inc . • 150 Freepor t Rd. • P i t t sburgh, PA 15238 • U.S.A. • Te l : +1 (412) 826 4444 Email: [email protected] • www.ametek-land.com
For a complete list of our international offices, please visit www.landinst.com Issue 4: 23 November 2016
MARCOM0290
PRODUCT WARRANTYEXCLUSIONS FROM WARRANTY
It should be noted that costs associated with calibration checks which may be requested during the warranty period are not covered within the warranty.Land reserve the right to charge for service/calibration checks undertaken during the warranty period if the cause is deemed to fall outside the terms of the warranty.This Land manufacturer’s warranty does not cover product malfunction arising from:-• incorrect electrical wiring.• connection to electrical power sources outside the rating of the product.• physical shock (being dropped, etc.) and impact damage.• inappropriate routing, support, physical shock & strain protection, etc. of the lightguide (Fibroptic
thermometers only).• environmental conditions exceeding the IP / NEMA rating of the product.• environmental conditions outside the Ambient Temperature, Humidity and Vibration rating of the product.• environmental contamination (solvent vapours, deposition of airborne contamination, cooling liquids of
non-neutral pH, etc.).• overheating as a result of interruption of water/air fl ow through cooling jackets or of incorrect installation.• inappropriate modifi cation of product (drilling holes in thermometer bodies, etc.).• inappropriate recalibration which results in product calibration being taken outside specifi cation.• improper resealing of thermometer following parameter adjustment (UNO, FLT5/A, etc.).• attempted repair by a non-Land-authorised repair centre.Land Instruments International Ltd • Dronfield S18 1DJ • England • Tel: +44 (0) 1246 417691 • Fax: +44 (0) 1246 410585 Email: [email protected] • www.landinst.comAMETEK Land, Inc . • 150 Freepor t Rd. • P i t t sburgh, PA 15238 • U.S.A. • Te l : +1 (412) 826 4444 Email: [email protected] • www.ametek-land.com
For a complete list of our international offices, please visit www.landinst.com Issue 4: 23 November 2016
MARCOM0290
Blank
For a full list of international offices, please visit our website www.landinst.com Copyright © 2017 LAND Instruments International. Continuous product development may make it necessary to change these details without notice. SPOT Actuator User Guide, Issue 1, 17 August 2017
SPOTA FAMILY OF HIGH PRECISION MULTI-MODE FIXED SPOT NON-CONTACT THERMOMETERS
Applies in the UK Applies in the US Applies in India
Land Instruments InternationalDronfield, S18 1DJ, England
Tel: +44 (0) 1246 417691Email: [email protected] www.landinst.com
AMETEK Land - Americas 150 Freeport Road, Pittsburgh, PA 15238, U.S.A.
Tel: +1 (412) 826 4444Email: [email protected] www.ametek-land.com
AMETEK Land China Service Part A, 1st Floor No. 460 North Fute Road, Waigaoqiao Free Trade Zone 200131 Shanghai, China Tel: +86 21 5868 5111 ext 122Email: [email protected] www.landinst.com
AMETEK Land India Service Divyasree N R Enclave,Block A, 4th Floor, Site No 1, EPIP Industrial Area Whitefield, Bangalore- 560066 Karnataka, India
Tel: +91 - 80 67823240 Email: [email protected] www.landinst.com