SPECTROSCOPY Downloaded from .
-
Upload
autumn-mitchell -
Category
Documents
-
view
227 -
download
0
Transcript of SPECTROSCOPY Downloaded from .
SPECTROSCOPY
Downloaded from www.pharmacy123.blogfa.com
وکنترول فارمسی کیمیای دیپارتمنت ادویه
2
Definition :
• Spectroscopy - The study of the interaction of electromagnetic radiation with matter .
وکنترول فارمسی کیمیای دیپارتمنت ادویه
3
Electromagnetic radiation :
• An oscillating electric and magnetic field which travels through space
• A discrete series of “particles” that possess a specific energy but have no Mass
BOTH!
وکنترول فارمسی کیمیای دیپارتمنت ادویه
4
Properties of Light• Light can be thought of as a wave or particle.
– The wavelength, , is the distance between crests of a wave (m)
– The frequency, , is the number of oscillations per second (Hz)
m/s 10998.2 8 ccsJ 10626.6 34 hE
وکنترول فارمسی کیمیای دیپارتمنت ادویه
5
Introduction of Spectrometric Analyses
The study how the chemical compound interacts with different wavelengths in a given region of electromagnetic radiation is called spectroscopy or spectrochemical analysis.
The collection of measurements signals (absorbance) of the compound as a function of electromagnetic radiation is called a spectrum.
وکنترول فارمسی کیمیای دیپارتمنت ادویه
6
Energy Absorption
The mechanism of absorption energy is different in the Ultraviolet, Infrared, and Nuclear magnetic resonance regions. However, the fundamental process is the absorption of certain amount of energy.
The energy required for the transition from a state of lower energy to a state of higher energy is directly related to the frequency of electromagnetic radiation that causes the transition.
وکنترول فارمسی کیمیای دیپارتمنت ادویه
7
Regions of the electromagnetic spectrum :
وکنترول فارمسی کیمیای دیپارتمنت ادویه
8
Interaction of e.m.r. with Matter
• Interaction of electromagnetic radiation with matter– The wave-length, , and the wave number, v’, of e.m.r. changes with the
medium it travels through, because of the refractive index of the medium; the frequency, v, however, remains unchanged
– Types of interactions
• Absorption• Reflection• Transmission• Scattering• Refraction
– Each interaction can disclose certain properties of the matter
– When applying e.m.r. of different frequency (thus the energy e.m.r. carried) different type information can be obtained .
refraction
transmission
absorption
reflection scattering
وکنترول فارمسی کیمیای دیپارتمنت ادویه
9
Absorption and Emission of Photons
وکنترول فارمسی کیمیای دیپارتمنت ادویه
10
Wave Number (cycles/cm)
X-Ray UV Visible IR Microwave
200nm 400nm 800nm
Wavelength (nm)
Spectral Distribution of Radiant Energy
وکنترول فارمسی کیمیای دیپارتمنت ادویه
11
V = Wave Number (cm-1)
Wave Length
C = Velocity of Radiation (constant) = 3 x 1010 cm/sec.
= Frequency of Radiation (cycles/sec)
The energy of photon:
h (Planck's constant) = 6.62 x 10-27 (Ergsec)
V =C
E = h = hC
C
= C =
Electromagnetic Radiation
وکنترول فارمسی کیمیای دیپارتمنت ادویه
12
Visible
Ultra violet
Radio
Gamma ray
Hz
cmcm-1Kcal/mol eV
Type
Quantum Transition
Type
spectroscopy
Type
Radiation
Frequency
υ
Wavelength
λ
Wave
Number VEnergy
9.4 x 107 4.9 x 106 3.3 x 1010 3 x 10-11 1021
9.4 x 103 4.9 x 102 3.3 x 106 3 x 10-7 1017
9.4 x 101 4.9 x 100 3.3 x 104 3 x 10-5 1015
9.4 x 10-1 4.9 x 10-2 3.3 x 102 3 x 10-3 1013
9.4 x 10-3 4.9 x 10-4 3.3 x 100 3 x 10-1 1011
9.4 x 10-7 4.9 x 10-8 3.3 x 10-4 3 x 103 107
X-ray
Infrared
Micro-wave
Gamma ray emission
X-ray absorption, emission
UV absorption
IR absorption
Microwave absorption
Nuclear magnetic resonance
Nuclear
Electronic (inner shell)
Molecular vibration
Electronic (outer shell)
Molecular rotation
Magnetically induced spin states
Spectral Properties, Application and Interactions of Electromagnetic Radiation
وکنترول فارمسی کیمیای دیپارتمنت ادویه
13
Atomic Spectra
• Shell structure & energy level of atoms– In an atom there are a number of shells and of subshells
where e-’s can be found – The energy level of each shell & subshell are different
and quantised• The e-’s in the shell closest to the nuclei has the lowest
energy. The higher shell number is, the higher energy it is• The exact energy level of each shell and subshell varies with
substance
• Ground state and excited state of e-’s– Under normal situation an e- stays at the lowest possible
shell - the e- is said to be at its ground state– Upon absorbing energy (excited), an e- can change its
orbital to a higher one - we say the e- is at an excited state.
n = 1
n = 2
n = 3, etc.
energy
E
groundstate
Excitedstate
En
erg
y
n=1
n=2
n=3
n=4
1s2s2p
3s3p
4s3d4p
4d4f
وکنترول فارمسی کیمیای دیپارتمنت ادویه
14
Atomic Spectra
• Electron excitation– The excitation can occur at different
degrees • low E tends to excite the outmost e-’s first• when excited with a high E (photon of high
v) an e- can jump more than one levels• even higher E can tear inner e-’s away from
nuclei
– An e- at its excited state is not stable and tends to return its ground state
– If an e- jumped more than one energy levels because of absorption of a high E, the process of the e- returning to its ground state may take several steps, - i.e. to the nearest low energy level first then down to next …
n = 1
n = 2
n = 3, etc.
energy
E
En
erg
y
n=1
n=2
n=3
n=4
1s2s2p
3s3p
4s3d4p
4d4f
وکنترول فارمسی کیمیای دیپارتمنت ادویه
15
Atomic Spectra
• Atomic spectra– The level and quantities of energy
supplied to excite e-’s can be measured & studied in terms of the frequency and the intensity of an e.m.r. - the absorption spectroscopy
– The level and quantities of energy emitted by excited e-’s, as they return to their ground state, can be measured & studied by means of the emission spectroscopy
– The level & quantities of energy absorbed or emitted (v & intensity of e.m.r.) are specific for a substance
– Atomic spectra are mostly in UV (sometime in visible) regions
n = 1
n = 2
n = 3, etc.
energy
E
En
erg
y
n=1
n=2
n=3
n=4
1s2s2p
3s3p
4s3d4p
4d4f
وکنترول فارمسی کیمیای دیپارتمنت ادویه
16
Absorption Spectroscopy Introduction
A.) Absorption: electromagnetic (light) energy is transferred to atoms, ions, or molecules in the sample. Results in a transition to a higher energy state.
- Transition can be change in electronic levels, vibrations, rotations, translation, etc.
- Concentrate on Molecular Spectrum in UV/Vis (electronic transition)
- Power (P): energy of a beam that reaches a given area per second
- Intensity (I): power per unit solid angle
- P and I related to amplitude2
Eo
E1h Energy required of photon to give this transition:
hE= E1 - Eo
(excited state)
(ground state)
وکنترول فارمسی کیمیای دیپارتمنت ادویه
17
B.) Terms:
1.) Beer’s Law: A = bc
The amount of light absorbed (A) by a sample is dependent on the path length (b), concentration of the sample (c) and a proportionality constant (– molar absorptivity)
Amount of light absorbed is dependent on frequency ()
c
Absorbance is directly proportional to concentration Fe+2
Increasing Fe+2 concentration
وکنترول فارمسی کیمیای دیپارتمنت ادویه
18
B.) Terms:
1.) Beer’s Law: A = bc
Transmittance (T) = P/Po %Transmittance = %T = 100T
Absorbance (A) = log10 Po/P
No light absorbed- % transmittance is 100% absorbance is 0
All light absorbed- % transmittance is 0% absorbance is infinite
وکنترول فارمسی کیمیای دیپارتمنت ادویه
19
Relationship Described in Terms of Beer’s Law
A = Absorbance = bc = -log(%T/100)
= molar absorptivity: constant for a compound at a given frequency () units of L mol-1 cm-1
b = path length: cell distance in cm
c = concentration: sample concentration in moles per liter.
Therefore, by measuring absorbance or percent transmittance at a given frequency can get information related to the amount of sample (c) present with an identified and .
Note: law does not hold at high concentrations, when A > 1
وکنترول فارمسی کیمیای دیپارتمنت ادویه
20
is a measure of the amount of light absorbedper unit concentration at a particular .
Molar absorptivity is a constant for a particular substance, so if the concentration of the solution is halved, so is the
absorbance at sufficiently dilute concentrations.
Molar AbsorptivityA = lc
A
concentration
وکنترول فارمسی کیمیای دیپارتمنت ادویه
21
وکنترول فارمسی کیمیای دیپارتمنت ادویه
22
وکنترول فارمسی کیمیای دیپارتمنت ادویه
23
وکنترول فارمسی کیمیای دیپارتمنت ادویه
24
وکنترول فارمسی کیمیای دیپارتمنت ادویه
25
وکنترول فارمسی کیمیای دیپارتمنت ادویه
26
وکنترول فارمسی کیمیای دیپارتمنت ادویه
27
وکنترول فارمسی کیمیای دیپارتمنت ادویه
28
وکنترول فارمسی کیمیای دیپارتمنت ادویه
29
وکنترول فارمسی کیمیای دیپارتمنت ادویه
30
وکنترول فارمسی کیمیای دیپارتمنت ادویه
31
وکنترول فارمسی کیمیای دیپارتمنت ادویه
32
وکنترول فارمسی کیمیای دیپارتمنت ادویه
33
وکنترول فارمسی کیمیای دیپارتمنت ادویه
34
وکنترول فارمسی کیمیای دیپارتمنت ادویه
35
وکنترول فارمسی کیمیای دیپارتمنت ادویه
36
وکنترول فارمسی کیمیای دیپارتمنت ادویه
37
وکنترول فارمسی کیمیای دیپارتمنت ادویه
38
وکنترول فارمسی کیمیای دیپارتمنت ادویه
39
وکنترول فارمسی کیمیای دیپارتمنت ادویه
40
وکنترول فارمسی کیمیای دیپارتمنت ادویه
41
وکنترول فارمسی کیمیای دیپارتمنت ادویه
42
وکنترول فارمسی کیمیای دیپارتمنت ادویه
43
وکنترول فارمسی کیمیای دیپارتمنت ادویه
44
وکنترول فارمسی کیمیای دیپارتمنت ادویه
45
وکنترول فارمسی کیمیای دیپارتمنت ادویه
46
وکنترول فارمسی کیمیای دیپارتمنت ادویه
47
وکنترول فارمسی کیمیای دیپارتمنت ادویه
48
Cuvettes (sample holder)
• Polystyrene– 340-800 nm
• Methacrylate– 280-800 nm
• Glass– 350-1000 nm
• Suprasil Quartz– 160-2500 nm
وکنترول فارمسی کیمیای دیپارتمنت ادویه
49
وکنترول فارمسی کیمیای دیپارتمنت ادویه
50
وکنترول فارمسی کیمیای دیپارتمنت ادویه
51
وکنترول فارمسی کیمیای دیپارتمنت ادویه
52
وکنترول فارمسی کیمیای دیپارتمنت ادویه
53
وکنترول فارمسی کیمیای دیپارتمنت ادویه
54
UV-Visible Spectrophotometry
بنفش – ماورای سپکتروفوتومتریدید قابل
وکنترول فارمسی کیمیای دیپارتمنت ادویه
56
UV-Visible Spectrophotometry
The absorption of ultraviolet and visible radiation by molecules are dependent upon the electronic structure of the molecule.
So the ultraviolet and visible spectrum are called electronic spectrum.
وکنترول فارمسی کیمیای دیپارتمنت ادویه
57
What does the absorbed light (electromagnetic radiation)
do to the molecule?
high energy UV – ionizes electrons
low energy UV and visible – promotes electrons to higher energy orbitals(absorption of visible light leads to a colored solution)
IR – causes molecules to vibrate (more later)
700 nm 400 nm
IR UV
visibleEnergy increasing
وکنترول فارمسی کیمیای دیپارتمنت ادویه
58
UV/visible light absorption
In organic molecules, electronic transitions to higher energy molecular orbitals – double bonds: *
In transition metals, hydrated ions as Cu++ have splitting of d orbital energies and electronic transitions – weak absorption
In complexed transition metals, charge transfer of electrons from metal to ligand as Cu(NH3)4
++ – strong absorption
Valence electrons
وکنترول فارمسی کیمیای دیپارتمنت ادویه
59
Electronic Excitation
The absorption of light energy by organic compounds in the visible and ultraviolet region involves the promotion of electrons in , , and n-orbitals from the ground state to higher energy states. This is also called energy transition. These higher energy states are molecular orbitals called antibonding.
وکنترول فارمسی کیمیای دیپارتمنت ادویه
60
Ene
rgy
*
*
n
*
*
n
*
n
*
Antibonding
Antibonding
Nonbonding
Bonding
Bonding
وکنترول فارمسی کیمیای دیپارتمنت ادویه
61
Electronic Molecular Energy Levels
The higher energy transitions ( *) occur a shorter wavelength and the low energy transitions (*, n *) occur at longer wavelength.
وکنترول فارمسی کیمیای دیپارتمنت ادویه
62
and * orbitals and * orbitals
وکنترول فارمسی کیمیای دیپارتمنت ادویه
63
Electronic Transitions in Organic Molecules
http://www.cem.msu.edu/~reusch/VirtualText/Spectrpy/UV-Vis/spectrum.htm#uv1
وکنترول فارمسی کیمیای دیپارتمنت ادویه
64
and* orbitals
وکنترول فارمسی کیمیای دیپارتمنت ادویه
65
and * orbitals
وکنترول فارمسی کیمیای دیپارتمنت ادویه
66
Electronic Transitions: *
The * transition involves orbitals that have significant overlap, and the probability is near 1.0 as they are “symmetry allowed”.
وکنترول فارمسی کیمیای دیپارتمنت ادویه
67
* transitions - Triple bonds
Organic compounds with -C≡C- or -C≡N groups, or transition metals complexed by C≡N- or C≡O ligands, usually have “low-lying” * orbitals
وکنترول فارمسی کیمیای دیپارتمنت ادویه
68
Electronic Transitions: n *
The n-orbitals do not overlap at all well with the * orbital, so the probability of this excitation is small. The of the n* transition is about 103 times smaller than for the * transition as it is “symmetry forbidden”.
وکنترول فارمسی کیمیای دیپارتمنت ادویه
69
UV Activity
h
وکنترول فارمسی کیمیای دیپارتمنت ادویه
70
Excited States
وکنترول فارمسی کیمیای دیپارتمنت ادویه
71
Chemical Structure & UV Absorption
What is chromophore ?
•Chromophore is a functional group which absorbs a characteristic ultraviolet or visible region.
• Chromophoric Group ---- The groupings of the molecules which contain the electronic system which is giving rise to absorption in the ultra-violet region.
وکنترول فارمسی کیمیای دیپارتمنت ادویه
72
Chromophore absorptions
Chromophore Example Excitation max, nm Solvent
C=C Ethene
171 15,000 hexane
CC 1-Hexyne
180 10,000 hexane
C=O Ethanal
n
290180
1510,000
hexanehexane
N=O Nitromethane
n
275200
175,000
ethanolethanol
C-X X=Br X=I
Methyl bromide
Methyl Iodide
n
n
205255
200360
hexanehexane
وکنترول فارمسی کیمیای دیپارتمنت ادویه
73
Organic ChromophoresChromophore Transition max(nm) log()
Nitrile (-C≡N) to 160 <1.0
Alkyne (-C≡C-) to 170 3.0
Alkene (-C=C-) to 175 3.0
Alcohol (ROH) to 180 2.5
Ether (ROR) to 180 3.5
Ketone (-C(R)=O) to 180 3.0
to 280 1.5
Aldehyde (–C(H)=O) to 190 2.0
to 290 1.0
Amine (-NR2) to 190 3.5
Acid (-COOH) to 205 1.5
Ester (-COOR) to 205 1.5
Amide (-C(=O)NH2) to 210 1.5
Thiol (-SH) to 210 3.0
Nitro (-NO2) to 271 <1.0
Azo (-N=N-) to 340 <1.0
وکنترول فارمسی کیمیای دیپارتمنت ادویه
74
Single Beam Spectrophotometer
وکنترول فارمسی کیمیای دیپارتمنت ادویه
75
Dual Beam Spectrophotometer
وکنترول فارمسی کیمیای دیپارتمنت ادویه
76
Sample Cells
UV Spectrophotometer
Quartz (crystalline silica)
Visible Spectrophotometer
Glass
وکنترول فارمسی کیمیای دیپارتمنت ادویه
77
Cuvettes (sample holder)
• Polystyrene– 340-800 nm
• Methacrylate– 280-800 nm
• Glass– 350-1000 nm
• Suprasil Quartz– 160-2500 nm
وکنترول فارمسی کیمیای دیپارتمنت ادویه
78
Components of an Instrument for UV/Vis Absorbance Measurements:
1.) Basic Design:
Hitachi Instruments U-3010
Light Source, selector, Sample cell holder, Detector (amplifier, recorder)
وکنترول فارمسی کیمیای دیپارتمنت ادویه
79
a) Desired Properties of Components of UV/Vis:
Light Source Selector Creates Proper Narrow Bandpass:Stable: Selects Desired Constant P Large Light Throughput:
Good Precision Increase PIntense:
Increase PEasier to See Absorbance
Sample Cell Holder DetectorFixed Geometry: Stable
Constant b Sensitive to of InterestTransmits of Interest:
Increase P
وکنترول فارمسی کیمیای دیپارتمنت ادویه
80
b) Light Sources UV/Vis (~ 200 – 800 nm):
1. Deuterium & Hydrogen Lamps (UV range)- continuous source, broad range of frequencies- based on electric excitation of H2 or D2 at Low pressure
40V Electric Arc
Electrode
Filament
D or H Gas2 2
Sealed Quartz Tube
In presence of arc, some of the electrical energy is absorbed by D2 (or H2) which results in the disassociation of the gas and release of light
D2 + Eelect D*2 D’ + D’’ + h(light produced)
Excited state
وکنترول فارمسی کیمیای دیپارتمنت ادویه
81
2. Tungsten Filament Lamp (Vis – Near IR)- continuous source, broad range of frequencies- based on black body radiation:
heat solid filament to glowing, light emitted will be characteristic of temperature more than nature of solid filament
Low pressure (vacuum)
Tungsten Filament
Temperature Dependence of
وکنترول فارمسی کیمیای دیپارتمنت ادویه
82
b) Wavelength Selectors:
1. Monochromator- separates frequencies () from polychromatic light in time or space.- allows only certain ’s to be selected and used.
i.) Dispersing Monochromator:
a) Prism: based on refraction of light and fact that different ’s have different values of refraction index (i) in a medium.
وکنترول فارمسی کیمیای دیپارتمنت ادویه
83
UV vs. IR vs. NMR
• UV has broad peaks relative to IR & NMR
• UV has less information than IR & NMR
• UV spectra are easier to collect
• UV spectra are faster to collect
• UV spectrometers are cheaper
• UV spectra require only nanograms of material or chemicals
وکنترول فارمسی کیمیای دیپارتمنت ادویه
84
Io I
Cell withPathlength, b,
containing solution
lightsource detector
blank where Io = I
concentration 2concentration 1
b
with sample I < Io
The process of light being absorbed by a solution
As concentration increased, less light was transmitted (more light absorbed).
وکنترول فارمسی کیمیای دیپارتمنت ادویه
85
Some terminology
I – intensity where Io is initial intensity
T – transmission or %T = 100 x T(absorption: Abs = 1 – T or %Abs = 100 - %T)
T = I/ Io
A – absorbanceA = - log T = -log I/ Io
وکنترول فارمسی کیمیای دیپارتمنت ادویه
86
Beer’s Law
A = abc
where a – molar absorptivity, b – pathlength, and c – molar concentration
See the Beer’s Law Simulator
وکنترول فارمسی کیمیای دیپارتمنت ادویه
87
Analyze at what wavelength?Scan visible wavelengths from 400 – 650
nm (detector range) to produce an absorption spectrum (A vs. )
Crystal Violet Absorption Spectrum
0
0.2
0.4
0.6
0.8
1
1.2
1.4
200 250 300 350 400 450 500 550 600 650 700 750wavelength, nm
Abso
rban
ce
max
max - wavelength where maximum absorbance occurs
phototube detector range
وکنترول فارمسی کیمیای دیپارتمنت ادویه
88
The BLANK
The blank contains all substances except the analyte.
Is used to set the absorbance to zero:Ablank = 0
This removes any absorption of light due to these substances and the cell.
All measured absorbance is due to analyte.
وکنترول فارمسی کیمیای دیپارتمنت ادویه
89
Light source
Grating
Rotating the gratingchanges the wavelength going through the sample
slits
slits
Sample
filter
Phototube
The components of a Spec-20D
occluder
When blank is the sample Io is determined
otherwise I is measured
Separates white lightinto various colors
detects light &measures intensity
- white light of constant intensity
وکنترول فارمسی کیمیای دیپارتمنت ادویه
90
Uses of visible spectrophotometry
Analysis of unknowns using Beer’s Law calibration curve
Absorbance vs. time graphs for kineticsSingle-point calibration for an
equilibrium constant determinationSpectrophotometric titrations – a way
to follow a reaction if at least one substance is colored – sudden or sharp change in absorbance at equivalence point, a piece-wise function
(Been there, done that!)
وکنترول فارمسی کیمیای دیپارتمنت ادویه
91
Practical Applications
• Medicinal Chemistry– compound ID (steroids, nucleosides)– monitoring isomerization, chirality
• Pharmaceutical Biotechnology– concentration/purity measurements– monitoring conformation of protein drugs
• Pharmacokinetics/Med. Chem.– HPLC monitoring and purification
وکنترول فارمسی کیمیای دیپارتمنت ادویه
92
Quantitative Analysis (Beer’s Law):
1) Widely used for Quantitative Analysis Characterization- wide range of applications (organic & inorganic)- limit of detection 10-4 to 10-5 M (10-6 to 10-7M; current)- moderate to high selectivity- typical accuracy of 1-3% ( can be ~0.1%)- easy to perform, cheap
2) Strategies
a) absorbing species- detect both organic and inorganic compounds
containing any of these species (all the previous examples)
Chromophore Example Excitation max, nm Solvent
C=C Ethene __> * 171 15,000 hexane
CC 1-Hexyne __> * 180 10,000 hexane
C=O Ethanaln __> * __> *
290180
1510,000
hexanehexane
N=O Nitromethanen __> * __> *
275200
175,000
ethanolethanol
C-X X=Br X=I
Methyl bromideMethyl Iodide
n __> *n __> *
205255
200360
hexanehexane
وکنترول فارمسی کیمیای دیپارتمنت ادویه
93
b) non- absorbing species- react with reagent that forms colored product- can also use for absorbing species to lower limit
of detection- items to consider:
, pH, temperature, ionic strength- prepare standard curve (match standards and
samples as much as possible)
Standard Addition Method (spiking the sample)
- used for analytes in a complex matrix where interferences in the UV/Vis for the analyte will occur: i.e. blood, sediment, human serum, etc..
- Method:(1) Prepare several identical aliquots, Vx, of the unknown sample.(2) Add a variable volume, Vs, of a standard solution of known
concentration, cs, to each unknown aliquot.(3) Dilute each solution to an equal volume, Vt.(4) Make instrumental measurements of each sample to get an
instrument response, IR.(5) Calculate unknown concentration, cx, from the following equation.
Note: This method assumes a linear relationship between instrument response and sample concentration.