Specifications Accuracy FS+-1.0% Differential pressure range 0-1kPa Absolute pressure range 80-110kPa(abs) Thermometer range 0-50Response time (63%) 200ms Type of flow Actual volumetric flow rate, standard flow rate, and mass flow rate. Display digit 5 digits Analog output VoltageCycle 0-10V, 15Hz Terminal shape BNC female Serial Communication Speed 9600/19200/38400/57600bps Terminal shape D-sub9 pin Dimensions W260, D350, H140mm, Weight: approx. 8kg Operating ambient environment Temperature: 540, Humidity: 1090%RH no condensationPower consumption 85132VAC, 50/60Hz50VA or below Accessories Diff. press. tube 2 pcs. x 2m long Thermocouple Sheath dia.: 1mm, Sheath length: 250mm, Compensating lead: 2m Power supply cable 3-pin plug, 3m long Option Indicator Separation Type Indicator size W180, D110, H100mm, Weight: approx. 1.5kg Cable 15m length Ethernet output Comm. speed 10BASE-T/100BASE-TX Terminal shape RJ-45 Model ModelLFC-200HF--Body type blank: Built-in Type D: Indicator Separation Type Communication output blank: Serial E: Serial + Ethernet Specifications subject to change without notice in order to improve performance. <2012/09/03> TSUKASA SOKKEN CO., LTD. 1-19-4 Tamazutsumi Setagaya-ku Tokyo 158-0087 JAPAN TEL: +81-3-3703-4391 http://www.sokken.co.jp/ L L a a m m i i n n a a r r F F l l o o w w C C a a l l c c u u l l a a t t o o r r M M O O D D E E L L L L F F C C - - 2 2 0 0 0 0 This product (LFC-200) does not include laminar flow element. Overview This Laminar Flow Calculator Model LFC-200 is a flow calculator designed for the LFE series of Laminar Air Flow meter. This product contains a differential pressure gauge, an absolute pressure gauge and a thermometer, you can calculate easily both volumetric flow rate and mass flow rate after connecting this product to laminar element of LFE series. Adopting with a superior pressure gauge in high accuracy and long-term stability, it provides flow rate accuracy FS+-1.0%. Features z Incorporating a pressure gauge and a thermometer, it easily displays volumetric and mass flow rates. z Usable on all of the laminar elements of LFE series. z Analog and Serial (RS-232C) communication output is standard equipment.

Transcript of Specifications Laminar Flow Calculator LFC-200€¦ · Terminal shape RJ-45 Model...

Page 1: Specifications Laminar Flow Calculator LFC-200€¦ · Terminal shape RJ-45 Model Model:LFC-200HF- - Body type blank: Built-in Type D: Indicator Separation Type Communication output

Specifications

Accuracy FS+-1.0% Differential pressure range 0-1kPa Absolute pressure range 80-110kPa(abs) Thermometer range 0-50℃ Response time (63%) 200ms Type of flow Actual volumetric flow rate, standard flow rate, and mass flow rate.

Display digit 5 digits Analog output Voltage,Cycle 0-10V, 15Hz

Terminal shape BNC female Serial Communication Speed 9600/19200/38400/57600bps

Terminal shape D-sub9 pin Dimensions W260, D350, H140mm, Weight: approx. 8kg Operating ambient environment Temperature: 5~40℃, Humidity: 10~90%RH(no condensation)Power consumption 85~132VAC, 50/60Hz,50VA or below Accessories Diff. press. tube 2 pcs. x 2m long

Thermocouple Sheath dia.: 1mm, Sheath length: 250mm, Compensating lead: 2m

Power supply cable 3-pin plug, 3m long Option

Indicator Separation

Type Indicator size W180, D110, H100mm, Weight: approx. 1.5kg Cable 15m length

Ethernet output Comm. speed 10BASE-T/100BASE-TX Terminal shape RJ-45

Model

Model:LFC-200HF-□-□

Body type blank: Built-in Type D: Indicator Separation Type Communication output blank: Serial E: Serial + Ethernet

Specifications subject to change without notice in order to improve performance. <2012/09/03>

TSUKASA SOKKEN CO., LTD. 1-19-4 Tamazutsumi Setagaya-ku Tokyo 158-0087

JAPAN

TEL: +81-3-3703-4391 http://www.sokken.co.jp/

LLaammiinnaarr FFllooww CCaallccuullaattoorr

MMOODDEELL LLFFCC--220000

This product (LFC-200) does not include laminar flow element.

Overview This Laminar Flow Calculator Model LFC-200 is a flow calculator designed for the LFE series of Laminar Air Flow meter. This product contains a differential pressure gauge, an absolute pressure gauge and a thermometer, you can calculate easily both volumetric flow rate and mass flow rate after connecting this product to laminar element of LFE series. Adopting with a superior pressure gauge in high accuracy and long-term stability, it provides flow rate accuracy FS+-1.0%.

Features

Incorporating a pressure gauge and a thermometer, it easily displays volumetric and mass flow rates.

Usable on all of the laminar elements of LFE series. Analog and Serial (RS-232C) communication output is standard

equipment.

Page 2: Specifications Laminar Flow Calculator LFC-200€¦ · Terminal shape RJ-45 Model Model:LFC-200HF- - Body type blank: Built-in Type D: Indicator Separation Type Communication output

Reliable flow rate accuracy

Guarantee of flow rate accuracy You can easily calculate the flow rate in laminar flow meter. However, if you calculate the flow

rate by pressure gauge and thermometer which you select, it is not easy to confirm the flow rate accuracy. Combination with Sokken LFE + this Sokken Model LFE-200, it guarantees the flow rate accuracy FS+-1.0%.

With an easy connection enables to output a volumetric flow and a mass flow rates

Connecting the two pressure leads between the flow meter and this product, and attaching a thermometer (thermocouple) to flow meter inlet. It is ready for measurement by inputting the inherent flow coefficient on a touch panel screen.

Since this product contains a differential pressure gauge, an absolute pressure gauge and a thermometer, it is able to output the following flow rates; actual volumetric flow rate (volume at the inlet of flow meter under inlet pressure and temperature), standard volume flow rate (volume at 20℃ and 1 atom) and mass flow rate.

Covering wide range of flow rate (from 10mL/min to 500L/s)

Applicable to all the model of LFE series This product is applicable to all of the LFE series of

laminar air flow meter. Therefore the product can use in different flow size (rated flow rate) of flow meters. In details of LFE series, please refer to the LFE series catalogue.

Selected sensors for accuracy and stability

Adapting a superior pressure sensor in accuracy and long-term stability, it fully utilizes the flow range of laminar air flow meter.

Thermo sensor is an easy-to-handle sheath thermocouple with diameter 1mm and it can quickly response to a temperature change of the air.

Display and Output

Selectable flow rate unit on a screen display You can set any flow rate unit in spite of which

flow meter to use thus it is able to display the flow rate at the most suitable for user’s application unit. It adjust automatically the position of decimal point based on an inputted laminar coefficient, you can effectively utilize the number of digits. Analog output with first order lag filter

Analog output flow rate (0-10V) has a function of digital first–order lag filter. You can set any output scaling and filter time constant.

Multifunctional serial communication If you use the standard equipped Serial

(RS-232C) communication, you can operate the acquisition of flow data and setting change through computers. Data acquired through communication includes temperature and pressure in addition to flow rates, you can utilize as a monitor or to record the state of flow.

(The standard of a communication cycle is 200ms. Use the cable length within 15m.)

Option

Indicator separation type

This option includes separated touch panel indicator from the main device, and 15m length cable is connecting the main device and indicator.

Adopting narrow path cable of approximately 7mm outside diameter, it’s easy to handle. You can use it in an engine laboratory at your ease as the coating is oil proof.

In this indicator separation type, an analog output terminal is equipped to the indicator.

Ethernet output communication

Ethernet communication is very useful if it has no serial terminal on a PC. You can set any IP address. If this product is connecting to network, it is able to acquire the data or to operate from long-distance remote. Ethernet communication command is the same as the Serial communication. (Communication period at one to one connection with PC is approx. 200ms. You cannot use an Ethernet communication and a serial

communication simultaneously.)

Sokken LFE: Calibration accuracy +-1% Press. gauge of A: accuracy +-0.1% Thermometer of B: accuracy +-0.5℃ Calculation by a user

Sokken LFE + LFC-200 guarantees,

FS+-1.0% flow rate accuracy

Flow rate accuracy ??? +

Volume metric flow rate ml/s, ml/min, ml/h l/s, l/min, l/h m3/s, m3/min, m3/h

Mass flow rate mg/s, mg/min, mg/h g/s, g/min, g/h kg/s, kg/min, kg/h

Assignable flow rate unit

Flow display screen with large characters