Specifications & more Elisys Uno
Transcript of Specifications & more Elisys Uno
Elisys Uno
System Overview Cat. No. 17350
Analyzer type 1 plate fully automated ELISA analyzer
Reagent system Open (Human tests predefined)
Throughput (Max.) 1 plate /run
Barcode reader No
Reaction / reading system 96 Well microtiterplate
Operation mode Prescheduled batch with chooseable priority
Languages available English, Portuguese, Russian, Spanish
Operating software Windows 98, 2000,Me, NT4.0, XP, Vista, 7
Tests Measuring modes EIA photometric
Analysis modes Endpoint
Calculation modes Qualitative: cut-off, cut-off standard
Quantitative: absorbance calculation algorithms
Calculation algorithms Point to point, linear regression, cubic spline, dose response,
polynomial 2nd – 4th order, % – absorbance, log-logit, 4 PL
# of programmable tests Unlimited
# of programmable profils Unlimited
# of standards per test Unlimited
Specifications & more
> System Specifications > Consumables > Accessories
Sample/ Control / Calibrator Sample type Blood, plasma, serum, urine
Supported sample vessels Sample cups or primary tubes ø 12–13 mm , max. 100mm height
Samples positions 96 (without predilution)
Sample volume 5 µl -1950 µl
Sample loading Batch
Pre-dilution 90 dilution cuvettes in specific rack, 48 in sample rack
Post-dilution Not automated
Calibrator vessel dimensions ø12-20 mm max. 90 mm height
Sample identification by barcode No
Reagent Reagent positions 27 positions on one rack
Reagent cooling No cooling or heating, reagents need to be at room temperature
Supported reagent vessel dimensions ø 17 mm, max. 100 mm height
ø 25 mm, max. 100 mm height
ø 27 mm , max. 100 mm height
ø 40 mm, max. 100 mm height
Continuous loading reagents No
Reaction Unit / Incubation Capacity 1 controlled position
Mixing capability Yes
Incubation temperature Ambient / 37 °C
Accuracy ± 1.5 °C (96 well plate with 200 µl aqua per cavity)
Uniformity ± 1.5 °C (96 well plate with 200 µl aqua per cavity)
Incubation time Free chooseable (increment 1 minute)
Liquid Pipetting Liquid transportation X-Z pipetting arm
Liquid level detection / Level sense Capacity
Probe crash detection Yes, vertical
Liquid handling Syringe
Clot detection / Bubble detection No
Processing time samples Approx. 18 minutes for 100 µl/well to 96 positions
Processing time reagents Approx. 6 minutes for 100 µl/well to 96 positions
Precision <2.5% CV at <5 µl,<1% CV at >5 µl
Accuracy <10% CV at 20 µl, <5% CV at 100 µl
Features Mixing
Disposable Tips Carbon tips 1 Steel needle (no tips)
Elisys UnoFully Automated ELISA Analyzer
Washing Capacity 2 L Wash buffer bottle
1 L Prime bottle
1 L Rinse bottle
2 L Waste bottle (for waste of 1st washing step)
Drain waste bottle not included
Wash head 8 channels
Plate type Flat and round bottom wells
Wash mode Strip mode
Dispense volume 25-500 µl/well
Precision <5% CV at 200 µl
Residual volume <3 µl U/F Bottom
Fluid alarms Low buffer, prime, rinse, waste full
Features Aspirate, dispense, soak, automatic prime and rinse
Reading Optical system Filterwheel, continuously rotating
Readings Single or double-wavelength simultaneous reading.
Light source Tungsten-Xenon lamp
Spectral range 340-700 nm
Wavelengths filters pre-installed Delivered with 405, 450, 492, 630 nm ±2 nm others on request
Max. # of wavelength filters installed Up to 8 positions
Read time (96 samples) dual wavelength Approx. 180 sec
Read time (96 samples) single wavelength Approx. 180 sec
Dynamic range (measurement range) - 0.2 to 3.0 OD
Accuracy 0 to 1.5 OD ± (1 % of the reading +0.005 OD)
1.5 to 3.0 OD ± (2 % of the reading +0.005 OD)
Detector Photodiode detection
Linearity -0,2 to 3,0 OD ±1%
Data Processing Interfaces Restricted bidirectional RS 232 /USB
LIS/LIMS Yes, free defined text file, manual import export function
ASCII import & export of worklist sample ID Yes from PC triggered by hand
Time scheduling Schedules 1 plate
Multiple assays per plate Up to 8 assays per plate
Quality control module Mean, standard error and Levey-Jennings (basic version)
Report for QC, calibration, controles, blank, patient
Memory capacity Depends on hard disc used
# of programable tests Unlimited
Lot management Included
User interface PC
External PC Yes
Supported printers All windows compatible printers
Elisys UnoFully Automated ELISA Analyzer
Generals PC min. requirements Pentium 133MHz, 128MB Ram SVGA, USB port
Dimensions (W x D x H) Instrument without any components: 102 x 55 x 47 cm
Space required for routine use: 170 x 80 x 80 cm
Packaging: 63 x 105 x 78 cm
Weight: Gross: 70.0 kg, net: 35.0 kg
Electrical requirements Universal AC input: 100–240 V , 50/60 Hz
Power consumption: 160 VA
On-line UPS recommended
Enviromental conditions Operating: temperature 18...35 °C, humidyti< 80 % non condensing
Transport: temperature 5 – 40 °C, humidity < 80 % non-condensing
Consumables Reagent Bottle 15 ml (12 pcs.) Cat.no. 17350/13
Reagent Bottle 30 ml (12 pcs.) Cat.no. 17350/14
Sample Vials 2 ml (500 pcs.) Cat.no. 17350/16
Sample Vial Caps (500 pcs.) Cat.no. 17350/15
Easy-to-Use software
> Graphical interface
> Bi-directional host communication
> Event log documentation
> Report generator; printout format according to user needs
> Password protection
> Assay programming simplified
> Sample and reagent positions can be changed
using the Windows® “drag and drop” feature
> Self-explanatory structure
> Multitasking
Syringe 50 µl
Cat.No. 17355-10
Bottle 15 ml
Cat.No. 17350/13
Sample Vials + Sample Vial Caps
Cat.No. 17350/16 + 17350/15
Standard Accessories (supplied with each instrument)
Quantity Cat. No.
Addendum Sheet for Drain Tube Connection 1 17350/3
Biohazard Risk Sheet 1 17350/18
Reagent Bottle 15 ml (12 pcs.) 1 17350/13
Bottle Set (Wash, Waste, Rinse) 1 17353/9
Drain Tube + 2 Clips + T-Connector 1 17352/48
Maintenance Kit 1 17350/25
Microtiter Plate 1 17360/1
Performance Check Kit 1 17350/11
Plexiglas Cover Red 1 17356/1
Probe Shield Assembly Grey 1 17356/2
Power Cord 1 18999
Prime Bottle with Cap 1 17353/1
Quick Guide 1 17350/90
Quick Start Guide 1 17350/52
Reaction Plate Cover 1 17352/9
Reagent Rack ( Open Rack ) 1 17359/1
Recommended Field Maintenance Procedure 1 17350/72
Sample Rack 96 Positions 1 17359/7
Serial Cable USB 1 17354
Software CD 1 17350/5
Software CD HUMAN Settings 1 17350/6
Syringe 2,5 ml 1 17355-11
Syringe 50 µl 1 17355-10
Unpacking Instructions 1 17350/4
UPS 1 KVA / 230 V 1 18961
UPS 1 KVA / 110 V 1 18961/110
User Manual 1 17350/1
Sample Vials 2 ml ( 50 pcs.) 1 17350/42
Sample Vial Caps (50 pcs.) 1 17350/66
Washer Head + 2 Screws 1 17358
Elisys UnoAccessories
Performance Check Kit
Cat.No. 17350/11
Syringe 2.5 ml
Cat.No. 17355-11
Washer Head + 2 Screws
Cat.No. 17358
UPS 1 KVA / 230 V
Cat.No. 18961
HUMAN Gesellschaft für Biochemica und Diagnostica mbH Max-Planck-Ring 21 · 65205 Wiesbaden · GermanyTel. +49 6122-9988-0 · Fax +49 6122-9988-100 · e-mail: [email protected] · www.human.de
1735
0/20
12-0
8 ©
201
2 H
uman
Gm
bH
Additional Accessories Cat. No.
AID Predilution Rack 17359/1C
Bottle Set (Wash, Waste, Rinse) 17353/9
Blood Bank Rack 17359/10
Combined Dilution / Reagent Rack 17359/1B
Reagent Rack Dengue & H. pylory 17359/3
Reagent Rack HIV & Hepatitis 17359/4
Reagent Rack Hormone 17359/5
Reagent Rack Infectious Diseases 17359/2
Reagent Rack Open Design 17359/1
Rinse Bottle 1 L with Plain Cap 17353/6
Rinse Bottle 1 L with Sensor Cap 17353/2
Sample Rack 96 Round Sample (ø 12 mm) 17359/7
Sample Rack 96 Round Sample (ø 13 mm) 17359/8
Wash Bottle 2 L with Sensor Cap 17353/3
Wash/Waste Bottle 2 L Plain Cap (1 Bottle) 17353/7
Wash/Waste Bottle 2 L Plain Cap (Set of 3 Bottles) 17353/8
Waste Bottle 2 L with Sensor Cap 17353/4
Computer with Windows XP (32 Bit) 18991P
TFT Screen 17901M
Laserprinter 18993L
Servicekits Cat. No.
Starter Spare Part Kit (for up to 5 Elisys Uno) 17350/70
Elisys UnoAccessories
Sample Rack
96 Round Sample
Cat.No. 17359/7
Reagent Rack
HIV & Hepatitis
Cat.No. 17359/4
Reagent Rack
Infectious Diseases
Cat.No. 17359/2
Reagent Rack
Open Design
Cat.No. 17359/1
Reagent Rack
Dengue & H. pylory
Cat.No. 17359/3
AID Predilution Rack
Cat.No. 17359/1C
Blood Bank Rack
Cat.No. 17359/10