8
Sodium Analyzer Series 2131 Specifications 70-82-03-60 September 2009 Introduction The 2131 is the most advanced online instrument specifically developed for use in continuous measurement and determination of sodium ion concentration. Common application points in a boiler system for online sodium measurement include: make-up water, condensate, boiler feedwater and saturated steam/main steam. The 2131 sodium analyzer operates automatically and comes complete with true fully automatic calibration. Features Wide Analysis Range (0.01 PPB - 10 PPM) True fully automatic single or 2-point calibration Low maintenance : no change of standards for up to two months (one month for reagent) Grab sample capability No change of diffusion tubing Automatic Range Selection Process calibration-calibrate to sample Two 4-20mA isolated current outputs Protection of electrodes from “Hot sample” Automatic temperature compensation Automatically stores last 10 calibrations and alarm logs No use of highly toxic reagents such as monoethylamine or di-isopropylamine 2131 Sensor Unit The sample flows through the heat exchanger, passes through constant head (used to stabilize minor changes in flow rate) and through the three solenoid valves. A vapor entrainment mechanism adds alkaline buffer vapor to the sample before it enters the sodium measuring flow cell and flows to drain. Calibration and grab sample are controlled automatically by the analyzer. Sodium Analyzer Series 2131 Standard/grab sample liquid is brought into system via the three 3-way solenoid valves. A thermistor is housed in the flowcell and is used to detect the temperature of the sample/standard/ grab sample. Sensor flow diagram

Transcript of Sodium Analyzer Series 2131 Specifications - Eneric Ltdeneric.net/Honeywell/Technical-data/Sodium...

Page 1: Sodium Analyzer Series 2131 Specifications - Eneric Ltdeneric.net/Honeywell/Technical-data/Sodium Analyzer.pdf · application points in a boiler system for online sodium ... Automatically

Sodium Analyzer Series 2131 Specifications 70-82-03-60 September 2009

Introduction

The 2131 is the most advanced online instrument

specifically developed for use in continuous measurement

and determination of sodium ion concentration. Common

application points in a boiler system for online sodium

measurement include: make-up water, condensate, boiler

feedwater and saturated steam/main steam. The 2131

sodium analyzer operates automatically and comes

complete with true fully automatic calibration.

Features

Wide Analysis Range (0.01 PPB - 10 PPM)

True fully automatic single or 2-point calibration

Low maintenance : no change of standards for up to

two months (one month for reagent)

Grab sample capability

No change of diffusion tubing

Automatic Range Selection

Process calibration-calibrate to sample

Two 4-20mA isolated current outputs

Protection of electrodes from “Hot sample”

Automatic temperature compensation

Automatically stores last 10 calibrations and alarm logs

No use of highly toxic reagents such as

monoethylamine or di-isopropylamine

2131 Sensor Unit

The sample flows through the heat exchanger, passes

through constant head (used to stabilize minor changes in

flow rate) and through the three solenoid valves. A vapor

entrainment mechanism adds alkaline buffer vapor to the

sample before it enters the sodium measuring flow cell and

flows to drain. Calibration and grab sample are controlled

automatically by the analyzer.

Sodium Analyzer Series 2131

Standard/grab sample liquid is brought into system via the

three 3-way solenoid valves. A thermistor is housed in the

flowcell and is used to detect the temperature of the

sample/standard/ grab sample.

Sensor flow diagram

HFS Catalog_Without Tab_HighRes.pdf 68 6/8/2011 12:41:07 PM

Page 2: Sodium Analyzer Series 2131 Specifications - Eneric Ltdeneric.net/Honeywell/Technical-data/Sodium Analyzer.pdf · application points in a boiler system for online sodium ... Automatically

Sodium Analyzer Series 2131 2

Sensor Parts

General Specifications

Performance

Range 0.01 ppb - 10 ppm.

Accuracy ± 5% of reading or ± 0.1 ppb (whichever is greater) within ± 5°C of calibration temperature.

Reproducibility ± 5% of reading or ± 0.1 ppb (whichever is greater) at constant temperature.

Response Time 90% of 1-10 ppb step: less than 4 minutes 90% of 100-1 ppb step: less than 6 minutes

Temperature Compensation Automatic.

pH Compensation Alkaline vapor buffer: 0.88 SG (35% W/V) ammonia above 0.1 ppb sodium. 50% diethyl amine solution below 0.1 ppb sodium.

Single or 2-point calibration Manual and fully automatic calibration. Manual initiated by operator, automatic with programmable frequency in days. No manual changes of standards during calibration.

Current Outputs Two isolated 4-20 mA current (analog) outputs.

Digital Outputs Communication interface via RS-232 & RS-485 using MODBUS RTU protocol

Alarms Three voltage-free contacts, alarm points set from transmitter rated at 2A, 250VAC (non-inductive)

Power Wide range of input power supply 90VAC – 250VAC

CE Certification (on request) Meets low voltage and low electromagnetic compatibility directives

HFS Catalog_Without Tab_HighRes.pdf 69 6/8/2011 12:41:07 PM

Page 3: Sodium Analyzer Series 2131 Specifications - Eneric Ltdeneric.net/Honeywell/Technical-data/Sodium Analyzer.pdf · application points in a boiler system for online sodium ... Automatically

Sodium Analyzer Series 2131 3

General Specifications

Operating Conditions

Sample Temperature: 41 – 131 °F (5 - 55°C); Flow 150 - 400 ml/min.

Ambient Temperature 32 - 131 °F (0 - 55°C).

Pressure 5 – 30 psig (0.35 – 2.07 bar)

Composition Sample should be filtered to 60 micron, free of film forming compounds.

Mechanical

Construction

Transmitter – high strength ABS, IP65 (NEMA 4x)

Wet section – high strength ABS and sheet metal, all wetted parts stainless steel or PVC, IP55 (NEMA 4)

Dimensions

Height=31”(78.74 cm) including drain

Width=22”(55.88 cm)

Depth=8.5”(21.59 cm)

Mounting

Four key-style holes:

Height=0.86”(2.18 cm), Top Width=0.281”(0.714 cm), Radius=0.280”(0.711 cm)

Horizontal: 14.7”(37.34 cm), Vertical: 11”(27.94 cm)

Weight 40 lbs. (18.14 kg)

Wiring Eight (8) conduit knockouts

Piping Sample Inlet: 1/4"(0.635 cm) OD Swagelok®

Drain: 3/8"(0.953 cm) OD tube

Finish Corrosion resistant.

Distance Up to 325 feet (100 meters) on request

HFS Catalog_Without Tab_HighRes.pdf 70 6/8/2011 12:41:07 PM

Page 4: Sodium Analyzer Series 2131 Specifications - Eneric Ltdeneric.net/Honeywell/Technical-data/Sodium Analyzer.pdf · application points in a boiler system for online sodium ... Automatically

Sodium Analyzer Series 2131 4

For More Information

Learn more about how Honeywell’s 2131C Sodium

Analyzer can increase performance, reduce downtime

and decrease configuration costs, visit our website

www.honeywell.com/ps/hfs or contact your Honeywell

account manager.

Honeywell Process Solutions

1860 West Rose Garden Lane

Phoenix, Arizona 85027

Tel: 1-800-423-9883 or 1-800-343-0228

www.honeywell.com/ps

Warranty/Remedy

Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Contact your local

sales office for warranty information. If warranted goods are returned to Honeywell during the period of coverage, Honeywell

will repair or replace without charge those items it finds defective. The foregoing is Buyer's sole remedy and is in lieu of all

other warranties, expressed or implied, including those of merchantability and fitness for a particular purpose.

Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this

printing. However, we assume no responsibility for its use.

While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the

customer to determine the suitability of the product in the application.

Specifications are subject to change without notice.

70-82-03-60 September 2009 © 2010 Honeywell International Inc.

HFS Catalog_Without Tab_HighRes.pdf 71 6/8/2011 12:41:07 PM

Page 5: Sodium Analyzer Series 2131 Specifications - Eneric Ltdeneric.net/Honeywell/Technical-data/Sodium Analyzer.pdf · application points in a boiler system for online sodium ... Automatically

Sodium Analyzer Series 2131C Specifications 70-82-03-61 September 2009

Introduction

The 2131C is the most advanced online instrument

specifically developed for use in continuous

measurement and determination of sodium ion

concentration. Common application points in a boiler

system for online sodium measurement include: make-

up water, condensate, boiler feedwater and saturated

steam/main steam.

Features

Wide Analysis Range (0.01 PPB - 10 PPM)

Low maintenance

Very low reagent & standard consumption

No change of diffusion tubing

Automatic Range Selection

Process calibration-calibrate to sample

Two 4-20mA isolated current outputs

Protection of electrodes from “Hot sample”

Automatic temperature compensation

Automatically stores last 10 calibrations and alarm

logs

No use of highly toxic reagents such as

monoethylamine or diisopropylamine

Sodium Analyzer

2131C Sensor Unit

During calibration the 3-way valve is activated, a

standard of known sodium concentration passes

through a heat exchanger to equilibrate the solution’s

temperature to that of the sample. The standard then

enters the constant head unit where flow is stabilized.

Alkaline buffer vapor is added via a vapor

entrainment mechanism before the standard enters

the sodium measuring flow cell and flows to drain.

During normal operating conditions the standard

solution is replaced by the sample.

HFS Catalog_Without Tab_HighRes.pdf 72 6/8/2011 12:41:07 PM

Page 6: Sodium Analyzer Series 2131 Specifications - Eneric Ltdeneric.net/Honeywell/Technical-data/Sodium Analyzer.pdf · application points in a boiler system for online sodium ... Automatically

Sodium Analyzer Series 2131C 2

Sensor Parts

General Specifications

Performance

Range 0.01 ppb - 10 ppm.

Accuracy ± 5% of reading or ± 0.1 ppb (whichever is greater) within ± 5°C of calibration temperature

Reproducibility ± 5% of reading or ± 0.1 ppb (whichever is greater) at constant temperature

Response Time 90% of 1-10 ppb step: less than 4 minutes 90% of 100-1 ppb step: less than 6 minutes

Temperature Compensation Automatic

pH Compensation Alkaline vapor buffer: 0.88 SG (35% W/V) ammonia above 0.1 ppb sodium 50% diethylamine solution below 0.1 ppb sodium

Single or 2-point calibration Manual calibration

Current Outputs Two isolated 4-20 mA current (analog) outputs

Digital Outputs Communication interface via RS-232 & RS-485 using MODBUS RTU protocol

Alarms Three voltage-free contacts, alarm points set from transmitter rated at 2A, 250VAC (non-inductive)

Power Wide range of input power supply 90VAC – 250VAC

CE Certification (on request)

Meets low voltage and low electromagnetic compatibility directives

HFS Catalog_Without Tab_HighRes.pdf 73 6/8/2011 12:41:07 PM

Page 7: Sodium Analyzer Series 2131 Specifications - Eneric Ltdeneric.net/Honeywell/Technical-data/Sodium Analyzer.pdf · application points in a boiler system for online sodium ... Automatically

Sodium Analyzer Series 2131C 3

General Specifications

Operating Conditions

Sample Temperature: 41 – 131 °F (5 - 55°C); Flow 150 - 400 ml/min

Ambient Temperature 32 - 131 °F (0 - 55°C)

Pressure 5 – 30 psig (0.35 – 2.07 bar)

Composition Filter to 60 micron, free of film forming compounds

Mechanical

Construction

Transmitter – high strength ABS, IP65 (NEMA 4x)

Wet section – high strength ABS and sheet metal, all wetted parts stainless steel or PVC, IP55 (NEMA 4)

Dimensions

Transmitter

Height=10.92”(27.74 cm)

Width=7”(17.78 cm)

Depth=3.54”(8.99 cm)

Wet section

Height=18.25”(46.36 cm)

Width=11.25”(28.58 cm)

Depth=7.63”(19.38 cm)

Mounting

Four key-style holes:

Height=0.86”(2.18 cm), Top Width=0.281”(0.714 cm), Radius=0.280”(0.711 cm)

Horizontal: 14.7”(37.34 cm), Vertical: 11”(27.94 cm)

Weight 23 lbs. (10.44 kg)

Wiring Eight (8) conduit knockouts

Piping Sample Inlet: 1/4"(0.635 cm) OD Swagelok®

Drain: 3/8"(0.953 cm) OD tube

Finish Corrosion resistant.

Distance Up to 325 feet (100 meters) on request

HFS Catalog_Without Tab_HighRes.pdf 74 6/8/2011 12:41:07 PM

Page 8: Sodium Analyzer Series 2131 Specifications - Eneric Ltdeneric.net/Honeywell/Technical-data/Sodium Analyzer.pdf · application points in a boiler system for online sodium ... Automatically

Sodium Analyzer Series 2131C 4

For More Information

Learn more about how Honeywell’s 2131C Sodium

Analyzer can increase performance, reduce downtime

and decrease configuration costs, visit our website

www.honeywell.com/ps/hfs or contact your Honeywell

account manager.

Honeywell Process Solutions

1860 West Rose Garden Lane

Phoenix, Arizona 85027

Tel: 1-800-423-9883 or 1-800-343-0228

www.honeywell.com/ps

Warranty/Remedy

Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Contact your local

sales office for warranty information. If warranted goods are returned to Honeywell during the period of coverage, Honeywell

will repair or replace without charge those items it finds defective. The foregoing is Buyer's sole remedy and is in lieu of all

other warranties, expressed or implied, including those of merchantability and fitness for a particular purpose.

Specifications may change without notice. The information we supply is believed to be accurate and reliable as of this

printing. However, we assume no responsibility for its use.

While we provide application assistance personally, through our literature and the Honeywell web site, it is up to the

customer to determine the suitability of the product in the application.

Specifications are subject to change without notice.

Swagelok is a registered product of Swagelok Co.

70-82-03-61 September 2009 © 2010 Honeywell International Inc.

HFS Catalog_Without Tab_HighRes.pdf 75 6/8/2011 12:41:07 PM