S.Monteil CaloPiquet 1
August 2010 – A typical online Calo Piquet Analysis -
Outline of the analysis (Fill 1264 – Run #77195)
1. Calorimeter 2DViews, inclusive distributions
2. Reconstruction
3. Hadron and EM triggers performance
1. DAQ/Tell1 summary
2. CF summary.
![Page 2: S.MonteilCaloPiquet1 August 2010 – A typical online Calo Piquet Analysis - Outline of the analysis (Fill 1264 – Run #77195) 1. Calorimeter 2DViews, inclusive.](https://reader036.fdocuments.in/reader036/viewer/2022081603/5697bff81a28abf838cbf11b/html5/thumbnails/2.jpg)
S.Monteil CaloPiquet 2
1. Calorimeters 2D Views
Pick up the distribution in HISTDB/Calorimeters/Calo2DView
Comment: allows to spot hot/dead channels. For instance, 5 dead channels in HCAL (See the CaloPiquet twiki page).
![Page 3: S.MonteilCaloPiquet1 August 2010 – A typical online Calo Piquet Analysis - Outline of the analysis (Fill 1264 – Run #77195) 1. Calorimeter 2DViews, inclusive.](https://reader036.fdocuments.in/reader036/viewer/2022081603/5697bff81a28abf838cbf11b/html5/thumbnails/3.jpg)
S.Monteil CaloPiquet 3
1. Inclusive distributions: XCAL
Pick up the distributions in HISTDB/Shift/XCAL: XCAL Status Overview
Comment: the typical ECAL and HCAL ADC spectra. You can refine the view by looking at the number of saturated channels per event.
![Page 4: S.MonteilCaloPiquet1 August 2010 – A typical online Calo Piquet Analysis - Outline of the analysis (Fill 1264 – Run #77195) 1. Calorimeter 2DViews, inclusive.](https://reader036.fdocuments.in/reader036/viewer/2022081603/5697bff81a28abf838cbf11b/html5/thumbnails/4.jpg)
S.Monteil CaloPiquet 4
1. Inclusive distributions: PRS
Pick up the distributions in HISTDB/Shift/PRS: PRS Status Overview
Comment: PRS measures MIPs at ~15 ADC counts (left plot) and EM component (right plot).
![Page 5: S.MonteilCaloPiquet1 August 2010 – A typical online Calo Piquet Analysis - Outline of the analysis (Fill 1264 – Run #77195) 1. Calorimeter 2DViews, inclusive.](https://reader036.fdocuments.in/reader036/viewer/2022081603/5697bff81a28abf838cbf11b/html5/thumbnails/5.jpg)
S.Monteil CaloPiquet 5
1. Inclusive distributions: SPD multiplicity
Pick up the distributions in HISTDB/L0Calo/SPD/Summary
Comment: those days, we have pretty occupied detectors. 400 hits in average for SPD. Far from the design. Do not report that anyway at the run meeting, even if you fear for your hadronic and/or EM channel …
![Page 6: S.MonteilCaloPiquet1 August 2010 – A typical online Calo Piquet Analysis - Outline of the analysis (Fill 1264 – Run #77195) 1. Calorimeter 2DViews, inclusive.](https://reader036.fdocuments.in/reader036/viewer/2022081603/5697bff81a28abf838cbf11b/html5/thumbnails/6.jpg)
S.Monteil CaloPiquet 6
2. Reconstruction
Pick up the distributions in HISTDB/Calorimeters/Reconstruction
Comment: well, see some online pi0s.
![Page 7: S.MonteilCaloPiquet1 August 2010 – A typical online Calo Piquet Analysis - Outline of the analysis (Fill 1264 – Run #77195) 1. Calorimeter 2DViews, inclusive.](https://reader036.fdocuments.in/reader036/viewer/2022081603/5697bff81a28abf838cbf11b/html5/thumbnails/7.jpg)
S.Monteil CaloPiquet 7
3. Hadron trigger performance
Pick up the distribution in HISTDB/L0Calo/Hadron/Emulator Comparison
Comment: the plots on the left compare the trigger response to the reconstructed response from the DAQ values. Any entry points a hardware failure. In this case (HC24FEB1), all the HW were changed except the cable…
![Page 8: S.MonteilCaloPiquet1 August 2010 – A typical online Calo Piquet Analysis - Outline of the analysis (Fill 1264 – Run #77195) 1. Calorimeter 2DViews, inclusive.](https://reader036.fdocuments.in/reader036/viewer/2022081603/5697bff81a28abf838cbf11b/html5/thumbnails/8.jpg)
S.Monteil CaloPiquet 8
3. Electron trigger performance
Pick up the distribution in HISTDB/L0Calo/Electron/Emulator Comparison
Comment: few errors inner C-side. No obvious trend observed. Make sure the L0Calo guys are aware of it.
![Page 9: S.MonteilCaloPiquet1 August 2010 – A typical online Calo Piquet Analysis - Outline of the analysis (Fill 1264 – Run #77195) 1. Calorimeter 2DViews, inclusive.](https://reader036.fdocuments.in/reader036/viewer/2022081603/5697bff81a28abf838cbf11b/html5/thumbnails/9.jpg)
S.Monteil CaloPiquet 9
3. Photon trigger performance
Pick up the distribution in HISTDB/L0Calo/Photon/Emulator Comparison
Comment: same comment as the previous one is in order. When checking Global pi0 (HISTDB/L0Calo/Global pi0/Emulator Comparison), you’ll find an error-free response (ECAL only). The guilty guy here is hence most likely the PRS.
![Page 10: S.MonteilCaloPiquet1 August 2010 – A typical online Calo Piquet Analysis - Outline of the analysis (Fill 1264 – Run #77195) 1. Calorimeter 2DViews, inclusive.](https://reader036.fdocuments.in/reader036/viewer/2022081603/5697bff81a28abf838cbf11b/html5/thumbnails/10.jpg)
S.Monteil CaloPiquet 10
4. DAQ/Tell1 summary
Pick up the distribution in HISTDB/Calorimeters/Analysis/CaloReadoutStatus
Comment: tells you how the DAQ communications behaved. This run is perfect. An error rate (correlated between all detectors) of few per 10-4 is acceptable [corresponds to a known desync Croc/Tell1 for the first BCID count].
![Page 11: S.MonteilCaloPiquet1 August 2010 – A typical online Calo Piquet Analysis - Outline of the analysis (Fill 1264 – Run #77195) 1. Calorimeter 2DViews, inclusive.](https://reader036.fdocuments.in/reader036/viewer/2022081603/5697bff81a28abf838cbf11b/html5/thumbnails/11.jpg)
S.Monteil CaloPiquet 11
5. Calibration Farm summary
Pick up the distribution in HISTDB/Calorimeters/Analysis/CalibrationData
Comment: overview of the four detectors. They are critical plots since they trigger Shift Crew actions. Comments destinated to the data manager have been updated. In more details for each detector …
![Page 12: S.MonteilCaloPiquet1 August 2010 – A typical online Calo Piquet Analysis - Outline of the analysis (Fill 1264 – Run #77195) 1. Calorimeter 2DViews, inclusive.](https://reader036.fdocuments.in/reader036/viewer/2022081603/5697bff81a28abf838cbf11b/html5/thumbnails/12.jpg)
S.Monteil CaloPiquet 12
5. Calibration Farm summary: ECAL.
Pick up the distribution in HISTDB/Calorimeters/Analysis/CalibrationData
Comment: give statements on the detector, provided that the LED monitoring system behaves as expected.
If the number of channels in error/fatal notice are large (>300), most likely linked [for the time being] to a LEDTSB configuration problem. Action: reconfigure LEDTSB.
The shift crew is notified to do so. Actions in nominal operation (hopefully soon) are also given in Shift page.
GainVariationPmt2PinNoiseNoGainMonitorLargeLEDNoiseLEDNoiseOutRangeLEDLowLEDSignalUnexpectedSignal
PedestalChi2LargePedestalNoisePedestalShiftOverNoisePedestalNoiseAveragePedestalNoiseLargePedestalShiftPedestalShift
![Page 13: S.MonteilCaloPiquet1 August 2010 – A typical online Calo Piquet Analysis - Outline of the analysis (Fill 1264 – Run #77195) 1. Calorimeter 2DViews, inclusive.](https://reader036.fdocuments.in/reader036/viewer/2022081603/5697bff81a28abf838cbf11b/html5/thumbnails/13.jpg)
S.Monteil CaloPiquet 13
5. Calibration Farm summary: HCAL.
Pick up the distribution in HISTDB/Calorimeters/Analysis/CalibrationData
Comment: stable so far.
The nominal indications to the shift crew are standing.
GainVariationPmt2PinNoiseNoGainMonitorLargeLEDNoiseLEDNoiseOutRangeLEDLowLEDSignalUnexpectedSignal
PedestalChi2LargePedestalNoisePedestalShiftOverNoisePedestalNoiseAveragePedestalNoiseLargePedestalShiftPedestalShift
![Page 14: S.MonteilCaloPiquet1 August 2010 – A typical online Calo Piquet Analysis - Outline of the analysis (Fill 1264 – Run #77195) 1. Calorimeter 2DViews, inclusive.](https://reader036.fdocuments.in/reader036/viewer/2022081603/5697bff81a28abf838cbf11b/html5/thumbnails/14.jpg)
S.Monteil CaloPiquet 14
5. Calibration Farm summary: PRS.
Pick up the distribution in HISTDB/Calorimeters/Analysis/CalibrationData
Comments: under development.
LEDs are not enabled. The corresponding 4 monitored quantities should be found in UNKNOWN state at each data SaveSet.
Pedestal mean and sigma are measured for the two VFE alternances. A goodness of fit is given in each case (~300 warnings is currently the expectation).
No action must be taken so far. Yet, warn L. Li Gioi or S. Monteil for any deviating behaviour.
Chi^2 of the fit odd
Pedestal sigma odd
Pedestal mean odd
Chi^2 of the fit even
Pedestal sigma even
Pedestal mean even
![Page 15: S.MonteilCaloPiquet1 August 2010 – A typical online Calo Piquet Analysis - Outline of the analysis (Fill 1264 – Run #77195) 1. Calorimeter 2DViews, inclusive.](https://reader036.fdocuments.in/reader036/viewer/2022081603/5697bff81a28abf838cbf11b/html5/thumbnails/15.jpg)
S.Monteil CaloPiquet 15
5. Calibration Farm summary: SPD.
Pick up the distribution in HISTDB/Calorimeters/Analysis/CalibrationData
Comments: under development.
Again LEDs are not enabled. The corresponding 2 monitored quantities should be found in UNKNOWN state at each data SaveSet.
Count the number of hits for each alternance. No signal expected apart the electronic noise.
Actions: diagnose noisy channels. In case of Error/Fatal, you’ll have most likely to mask the channel in the trigger path..
Electronic noise odd
Electronic noise even
![Page 16: S.MonteilCaloPiquet1 August 2010 – A typical online Calo Piquet Analysis - Outline of the analysis (Fill 1264 – Run #77195) 1. Calorimeter 2DViews, inclusive.](https://reader036.fdocuments.in/reader036/viewer/2022081603/5697bff81a28abf838cbf11b/html5/thumbnails/16.jpg)
S.Monteil CaloPiquet 16
6. Conclusion.
It was a nice run …