UniPixSmartPull OperationmanualRev.1.02 page1/16
Smart Pull Micropipette puller
OperationManualRev.1.02
DesignedinSwitzerland
AssembledinSwitzerland
UniPix,SwitzerlandWeb:www.unipix-lab.com Email:[email protected]
UniPixSmartPull OperationmanualRev.1.02 page2/16
DISCLAIMERThepipettepullerSmartPullisdesignedforthespecificuseofcreatingmicropipettesandnootheruseisrecommended.This instrumentcreates itemsthatshouldonlybeused ina laboratoryenvironment. It isnotintendedforuse,norshouldbeused,inhumanexperimentation,orappliedtohumansinanyway.Thisisnotamedicaldevice.DonotopenorattempttorepairtheinstrumentwithoutexpressedandexplicitinstructionsfromUniPix.Hightemperatureispresentandcouldcauseinjury.Donotallowunauthorizedandoruntrainedoperativestousethisdevice.Anymisusewillbethesoleresponsibilityoftheuser/ownerandUniPixassumesnoimpliedorinferred liability fordirectorconsequentialdamages fromthis instrument if it isoperatedorusedinanywayotherthanforwhichitisdesigned.SAFETYWARNINGSANDPRECAUTIONS• Donotdisassemblethesystem.Referservicingtoqualifiedpersonnel.• Topreventfireorshockhazarddonotexposetheunittorainormoisture.OPERATIONAL• Makesuretooperatetheinstrumentonlywhenthecoverisclosed.• Toavoidburnsdonottouchtheheatingfilament,themetalpartsholdingthefilament,or
theheatedendsofglasspipettesthathavebeenpulled.• OnlyuseUniPixreplacementheatingfilaments.• Failuretocomplywithanyofthefollowingprecautionsmayresultininjurytotheusersof
thisdeviceaswellasthoseworkinginthegeneralareanearthedevice.• The micropipettes created using this instrument are very sharp and relatively fragile.
Contact with the pulled micropipette tips, therefore, should be avoided to preventaccidentallyimpalingyourself.
• Always dispose of micropipettes by placing them into a well-marked and spill-proof“sharps”container.
• Useonlywithcapillaryglass(tubing)recommendedbyUniPix
UniPixSmartPull OperationmanualRev.1.02 page3/16
The micropipette puller you always have dreamed to have Basedonthecutting-edgetechnology, thisaffordableelectrodepullerwillbringyoumultipleadvantagesandsetnewqualitylevelinelectrophysiologymicropipettespreparation.Designedwithatouchscreenwhichhelpyouselectingthepullingsequenceofyourchoiceandmemorizingyourfavoriteparameters,specialattentionwasplacedintheelaborationoftheuserinterfacetosimplifyyourday-to-daylifewhilekeepingthehigheststandard inmicropipettesfabrication. Using a long-standing experience in electrode pulling our engineers have placedspecial attention in the design of the electrode handling, the mechanical drive, force andtemperaturecontroloftheheatingfilament.Programmingofanelectrodepullerhasneverbeensosimpleandasingletouchwillbringyoutothenextlevelofsophistication.Builtaroundfastμ-controllers,theelectrodepullerusesthelatest available technology for the controlof theheating furnacemanagingup to50A in theplatinum-iridiumfilament.Mechanicalandthermalprotectionwillinsurestableandreproducibleconditionsforthepipettefabrication.Simpleandefficientholdingsystemsguarantyaperfectclampoftheglass.
Features
• Motion is controlled by a force-controlled DC motor coupled with a high-resolutionpositionsensing.
• Self-squeeze electrode holding allow efficient manipulation while preventing glassslippingorbreaking.
• Heatingfurnaceusingamechanicallyformedplatinumiridiumfilamentisprotectedinaspecialhousingallowingthermalandmechanicalprotection.
• A4.3inchestouchscreenisimplementedwithspecificuserinterface.• Easyandintuitiveprogrammingontouchscreenmenus• Pre-setpullingprograms,withpossibilityofadjustingparameters• Memoryforover1000customprograms,userpasswordprotected,storedonnon-volatile
memory.
Typical applications
• Patchpipettes• Two-electrodesvoltageclampforXenopusoocytesrecordings• Sharpelectrodes• Intranuclearinjection• XenopusorZebrafishoocytesorembryoinjections• Insectegginjection• Aspirationpipettes• Drugapplicationforpatchclamp
UniPixSmartPull OperationmanualRev.1.02 page4/16
Installation Placetheinstrumentonaflatandstablesurface.
a) Connectthepowersupplytotheinstrumentb) Connectthecabletothepowersupplyandto110V-220Vpowerlinec) Pressthepowerswitchtoturnontheinstrument
Thecoverprotectingthepullingmechanism,electrodeholdersandheatershouldbeclosedwhentheinstrumentisnotinuse.
Programming
The user interface UniPixSmartPullisdesignedtobesimpletouseandeasytoprogram.Everyuserhasthechoicebetween3pullingmodes,eachprimarilydesignedforspecificusages.
Injection pipettes
Thismodeisdesignedtopulltheglassinthreeheating/pullingsteps.Thisisused,forexample,for
patchclamppipettesorshortshanks.
Fine tip
Thismodeisdesignedtopulltheglassintwoheating/pullingsteps.Thismodecanbeusedfor
makingdifferenttipsincludingsharptips.
Patch/HiClamp
Thismodewithasinglepullisdesignedformakinginjectionpipettes.
UniPixSmartPull OperationmanualRev.1.02 page5/16
Anewuserissimplycreatedbyenteringanewname.Hewillberequestedtostorea4letterspassword.Thefactoryparametersforthethreemodesareautomaticallyloaded.Whentheuserisworkingintooneofthepullingmodes,hecanfreelymodifyanyparameterofthemodeinordertooptimizethepullingprocessaccordingtoisrequirements.Thenewparametersareautomaticallystoredforthisuser(unlesstheuserdecidetocancelmodifications).Ifthesameuserwantstohaveseveralvariantsofpullingparametersinthesamemode,hesimplycreatesanewuser.Morethan1000userscanbedeclaredoneachinstrument.
Figure1:OverviewofSmartPullprogramstructure
Commands Thefollowingcommandscanbeentererinsteadoftheusername:emptyspace guestordefaultmode(notethatthismodeisnotpasswordprotected)calibration usefultodeterminetheminimaltemperatureforagivenglassabout displayUPLsoftwareversionandinternalparametersdemo demomode(canbestoppedonlybyturningoffthepuller)
Using Smart Pull
Start up
• Turnontheinstrumentwiththepowerswitch• Aself-testrunsautomaticallyandindicatespossiblemalfunction• Enteryourusername(ifleftblank=guestmode)• Setyourpassword(firsttimethepasswordwillbeset,nopasswordinguestmode)• Repeatpassword
Figure2:Useridentificationandpasswordinputscreens
UniPixSmartPull OperationmanualRev.1.02 page6/16
Micropipette fabrication Selecttheprogram(tipmode):
Figure3:Programselectionscreen
Afterselectingthemode,thedisplaywillswitchtothespecificprogram.Aniconrepresentseachstep.Thepresetvaluesareoptimizedforbestpullingconditions.StartdisplaysoftheSmartPullprogrammodes:
InjectionpipettesFineTipPatch/HiClamp
1step3steps5steps
Figure4:Startscreensofthe3mainprogrammodes
Ifnecessary,youcanadjustallparametersinallsteps.Ifyouwanttochangeaparameterinastep,clickonthecorrespondingicon(seenextparagraphformoredetailsonstepparameters).
Loading the capillary
• PressSTART• Openthecoveroftheinstrument• Verifythatthetwocarriersareinloadconfiguration(closetotheheater)• Opentheholders(1)• Carefullyplacethecapillaryintothecarriers,ascenteredaspossible• Closetheholderbypushingdownbothflaps.Makesuretopressenoughinordertokeep
thecapillarytighttothecarrier
UniPixSmartPull OperationmanualRev.1.02 page7/16
Figure5:Placementoftheglasscapillaryontheholdersandtighteningofthecapillarybeforepulling
Figure6:Pullingprocesslaunchscreen
Pipette pulling
• Closethecoveroftheinstrument• PressPULL(theprogramcanbeabortedbypressingSTOP)• WaituntiltheprogramendsandshowstheLOADscreenagain• Openthecoveroftheinstrument• Carefullyopenthetwoholdersandtakethepulledcapillaries
Figure7:Imageofthefirstheating/pullingstep
Figure8:Imagelastheatingstepandcapillariesafterthelastpullingstep
UniPixSmartPull OperationmanualRev.1.02 page8/16
Pulling next capillary
• Repeattheloadingandpullingoperationsasoftenasnecessary.• Whenyouhavefinished,youcanturnoffthemainswitchoftheinstrument.• Ifyouhavechangedparameters(seesectionbelow)thelastparametersareautomatically
storedfortheuser.
Pulling steps parameters Atypicalpullingstepisdescribedbelow:
• TheheatturnsonatthesetHeat• Waitbeforepulling• Thefilamentheatsup, theglasssoftens,andaweakpulldrawstheglassoutwiththe
programmedForceuntilitreachestheprogrammedPosition.• WhentheprogrammedPositionhasbeenreached,theheatturnsoff
Byclickingontheicon«step1(or3,5)»,youaccessthestepmenuinwhichyoucanaccessthepullingparameters:
• Heat: settheinputcurrentoftheheater• Wait: setthewaitingtimebetweenheatingandpulling• Force: setthepullingforce• Position: setthetargetpositionforpulling
Tochangeaparameter,clickontheparametervalueandusethekeypadtomodifythevalue.PressDonebuttonforgoingbacktomodeprogram
Figure9:Programpullingparametersadjustmentscreenandtypepad
Parametersvalues:• Heat: 0-999 arbitraryunits• Force: 0-999 arbitraryunits• Position: 0-999 arbitraryunits• Wait: 0-999 seconds
UniPixSmartPull OperationmanualRev.1.02 page9/16
Cooling step parameters Aftereachpullingstep,theprogramisexecutingacoolingstep.Byclickingontheicon«step2(or4)»,youaccessthestepmenuinwhichyoucanaccesstheCoolingtimeparameter:
Figure10:Programcoolingtimesettingscreen
Parametervalues:• Coolingtime: 0-999 secondsPresetvaluesofthetipmodesInjectionpipettes:
Step# Heat Wait[s] Position Force
1 850 5 40 900
2 - 5 - -
3 700 5 135 800
4 - 5 - -
5 550 0 800 600
FineTip:
Step# Heat Wait[s] Position Force
1 870 5 140 600
2 - 5 - -
3 480 5 800 800
Patch/HiClamp:
Step# Heat Wait[s] Position Force
1 950 3 800 800
Note:everytimeyoucreateanewuser,youwillhavethethreepreprogrammedtipmodes.
UniPixSmartPull OperationmanualRev.1.02 page10/16
Calibration Theautocalibrationprogramiscalledbyenteringthecommand“calib”insteadoftheusername.Thecalibrationprogramisdesignedtohelptheusertodeterminatethepullingparameterswhenusingnewglasscapillariesorafterreplacingtheheatingfilament.RuntheCalibrationwhen:
• Youchangethefilament• Lotnumbersorcapillarytypeschange• Youcreateormodifyaprogram
Procedure:
1) Onstartingthepullertypecalibrationastheusername2) Mountaglasscapillary3) Press theSTARTkey to run thecalibration programTheheatingpower increases
gradually.Theheatingstopswhentheglassbeginstomove.4) Thevalueofheatingpowerisdisplayed.Thisisthebaselineheatingvalueforagiven
glasstypetested.Itisagoodstartingpointforthefirstcycleofyourprogram.
Figure11:Autocalibrationstartscreen
Connectivity (this feature is not activated in the current product)
SmartPullhasBluetoothandWiFiconnectivity.Allstoredparametersaswellasthelogofinstrumentusagewillbeaccessiblefromacomputeroratablet.Theprogramsandapplications forSmartPullaccesswillbedownloadedonUniPixwebsitewhenthefunctionalitywillbereleased.
UniPixSmartPull OperationmanualRev.1.02 page11/16
Maintenance
Cleaning Routinelycleantheexteriorandthebaseplateoftheunitbywipingthemwithadryclothtoremovedustandfinepiecesofglass.Avoidcontactwiththefilament.If you break glass capillaries into the heater, you can simply remove fragments and dust byblowingairwithanairgunonthefilament.
• Don’tusetoohighairpressure• Don’ttouchthefilamentwithanobject• Weargogglestoprotectyoureyesforglassfragmentswhenyoublowair.• Itisnotrecommendedblowairwithyourmouth.
Replacing the O-ring of the glass capillary holder TheholdershaveanelastomerO-ringtoclampandavoidslidingoftheglasscapillaryduringthepulling.Ifyouobservebadclamping,thismaybeduetodamagedO-ring.Inthiscase,itisrecommendedtoreplacetheO-ringsofthetwoholders.OnlyuseO-ringssoldbyUniPix.O-ringreplacementsteps:
1. TurnOFFthepowersupply2. Removethescrew(1)3. Removetheholderflap,butkeepthesmalltubeinside(2)4. RemovetheoldO-ring(3)andcarefullyplacethenewO-ring5. Screwtheholderflap
Figure11:PictureofcapillaryholderandO-ringlocation
UniPixSmartPull OperationmanualRev.1.02 page12/16
Replacing Heating Filament Ifthefilamentisdamagedofbroken,itmustbereplaced.ItisrecommendedthatyousendtheSmartPULLtoUniPixtodothefilamentreplacement(pleaseaskforthecostofreplacement).Ifyouwanttodothereplacementyourself,pleasecarefullyfollowtheinstructionsbelow:
ONLYUSEthefilamentsprovidedbyUniPixOtherfilamentsmaynotproperlyworkanddamagetheinstrument.
SmartPullreplacementfilamentsareprovidedwithaplasticholderthatholdstheminplaceduringthereplacementprocedure.Pleasefollowcarefullytheinstructions.AvideotutorialwillsoonbeavailableonUniPixwebsite.Filamentreplacementsteps:1. TurnOFFthepowersupply2. Removethegridthatenclosestheheatingfilament(4screws)3. Loosenthe2x2clampscrewsthatholdthefilamentinplace,andthenslideouttheold
filament.4. Slipinanewfilamentandthenretightenthe2x2clampscrews.5. Reinstallthegridthatenclosestheheatingfilamentandretightenthe4screws
Figure11:Pictureofprotectiongridandfilamentafterremovalofthegrid
After replacing the filament, run thecalibration program. In caseof anydysfunction,pleasecontactUniPixserviceatinfo@unipix-lab.comComplementaryinformationonmaintenanceandFAQcanbefoundontheUniPixwebsite:www.unipix-lab.com/upl.php
UniPixSmartPull OperationmanualRev.1.02 page13/16
Troubleshooting
Problem Possiblecause Solution
The puller displayremainsblank
Verify the power supplyconnectionand that theLEDison
Connectthepowersupplyinanotherplug
The puller displayremains blank butthe power switchLEDison
Display system is malfunctioning
Turn OFF the instrument with themain switch, wait 10 seconds, turnONagain.Shouldtheproblempersistcallrepresentative,orcontactUniPix
No display orfrozendisplay
Programbug Turn OFF the instrument with themain switch, wait 10 seconds, turnONagain.Shouldtheproblempersistcall representative, or contact theUniPixwebsite
Glassdoesnotpullwell
Wrongparameters Check the pulling parameters. It isrecommendedtorunthecalibrationprogram to help finding the goodheating/pullingparameters.
Wrongcapillarymaterial UsetherecommendedcapillariesThe glass capillaryslides on thecarrier
Grippernottightenough Tighten more the capillary bypushingtheholderflaps.
Dustorgreaseontheholders. Cleanwithaclothimpregnatedwithalcohol
The filament doesnotheat
Turn off the puller, wait 10secondsbeforeturningon. Ifthefilamentisbrokenortheconnection is loose this willbediagnosedbytheselftest.
Replace the filament only if anapparent damage of the platinum isobserved
UniPixSmartPull OperationmanualRev.1.02 page14/16
Technical specifications
Instrument Dimensions 53cmx40.6cmx30cmWeight 2.75kg(withoutpowersupply)Electrical 115/230volts,50/60HertzpowerlinePower 120W,maxHeatingfilament Platinum/IridiumFilamentPullingForce 0-20N,electroniccontrolledTaperLength 1-10mmCapillaryODrange 0.6–3.0mmCapillaryLength 50mm(min)to150mm(max)Note:SmartPullisdesignedforborosilicateglasscapillaries,notforquartzoraluminosilicate
Glass Capillary & Heating Filament Specifications
Glass Capillary
SmartPullmicropipettepuller isdesignedforusewithaluminosilicate,borosilicateorotherlowermelting-pointglasstubingorrodrangingfrom0.6to3.0mmindiameter.Thisinstrumentdoesnotpullquartzglass.ExamplesofthespecifictypesandsizesofglassthatcanbeusedwithSmartPullarelistedintheUniPixwebsiteatwww.unipix-lab.com.Anyglasswithcomparabletechnicalspecificationscanbeused.
Heating Filament
ThetypeandsizeofglassthatyouchoosemayrequireaHeatingFilamentotherthantheoneinstalledinyourpulleratthefactory.PleaserefertotheHeatingFilamentsectionofthismanualtodeterminetheappropriatestyleandsizeoffilamentnecessaryforpullingthespecificglassyouwouldliketouse.ONLYUSEUNIPIXSUPPLIEDREPLACEMENTHEATINGFILAMENTSINTHISINSTRUMENT
Systematic instructions for replacing the Heating Filament can be found in theMaintenancesectionofthismanual.
Technical Support TechnicalsupportisprovidedbyUniPixbyEmailatinfo@unipix-lab-com.
UniPixSmartPull OperationmanualRev.1.02 page15/16
Limited warranty
• UniPixSwitzerlandlimitsthewarrantyonthisinstrumenttorepairandreplacementofdefectivecomponentsfortwoyearsfromdateofshipment,providedtheinstrumenthasbeenoperatedinaccordancewiththeinstructionsoutlinedinthismanual.
• Abuse,misuse,orunauthorizedrepairswillvoidthiswarranty.• Warrantyworkwillbeperformedonlyatthefactory.• The cost of shipment bothways is paid by UniPix Switzerland during the first three
monthsthiswarrantyisineffect,afterwhichthecostistheresponsibilityofthecustomer.• Thelimitedwarrantyisasstatedaboveandnoimpliedorinferredliabilityfordirector
consequentialdamagesisintended.• Consumables(filaments)areexemptfromthiswarranty.
UniPixSàrl,Ch.Neuf11,CH-1028Préverenges,SwitzerlandEmail:[email protected]
UniPixSmartPull OperationmanualRev.1.02 page16/16
UniPixSàrlCH.Neuf11
CH-1028PréverengesSwitzerland
Email:[email protected]
DECLARATIONOFCONFORMITY
We,UniPixSàrl,Ch.Neuf11,CH-1028Préverenges,Switzerland,declareundersoleresponsibilitythattheproduct:
SmartPull
To which this declaration relates is in conformity with the following standards or othernormativedocuments: Safety: EN61010-1:200And therefore, conformswith the protection requirements of Council Directive 89/336/EECrelating to electromagnetic compatibility andCouncilDirective 73/23/EEC relating to safetyrequirements.
Issuedon: August10,2018_______________________________________ _______________________________PhilippeRenaud DanielBertrandPresident,CEO Vice-president,CTO