smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: [email protected]...

39

Transcript of smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: [email protected]...

Page 1: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –
Page 2: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

web: www.smar.com e-mail: [email protected] Specifications and information are subject to change without notice.

BRAZILSmar Equipamentos Ind. Ltda.Rua Dr. Antonio Furlan Jr., 1028Sertãozinho - SP 14170-480Tel.: +55 16 645-6455Fax: +55 16 645-6450e-mail: [email protected]

ARGENTINASmar ArgentinaSoldado de La Independencia, 1259(1429) Capital Federal – ArgentinaTelefax: 00 (5411) 4776-1300 / 3131e-mail: [email protected]

CHINASmar China Corp.3 Baishiqiao Road, Suite 30233Beijing 100873, P.R.C.Tel.: +86 10 6849-8643Fax: +86 10 6849-9549e-mail: [email protected]

FRANCESmar France S. A. R. L.42, rue du Pavé des GardesF-92370 ChavilleTel.: +33 1 41 15-0220Fax: +33 1 41 15-0219e-mail: [email protected]

GERMANYSmar GmbHRheingaustrasse 955545 Bad KreuznachGermanyTel: + 49 671-794680Fax: + 49 671-7946829e-mail: [email protected]

MEXICOSmar Mexico11, Poniente, No. 1314-1 PBCol. Centro C. P. 72000Ciudad de Puebla, PueblaTel.: +52 22 46-4386Fax: +52 22 46-4386e-mail: [email protected]

SINGAPORESmar Singapore Pte. Ltd.315 Outram Road#06-07, Tan Boon Liat BuildingSingapore 169074Tel.: +65 324-0182Fax: +65 324-0183e-mail: [email protected]

EUASmar International Corporation6001 Stonington Street, Suite 100Houston, TX 77040Tel.: +1 713 849-2021Fax: +1 713 849-2022e-mail: [email protected]

Smar Laboratories Corporation10960 Millridge North, Suite 107Houston, TX 77070Tel.: +1 281 807-1501Fax: +1 281 807-1506e-mail: [email protected]

Smar Research Corporation4250 Veterans Memorial Hwy.Suite 156Holbrook , NY 11741Tel: +1-631-737-3111Fax: +1-631-737-3892e-mail: [email protected]

Page 3: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Introduction III

INTRODUCTION

The DT301 (Touché) Intelligent Concentration/Density Transmitter is an instrument designed for the continuous onlinemeasurement of liquid concentration/density, directly in the industrial process.

The DT301 consists of a capacitive type differential pressure transmitter coupled to a pair of pressure repeaters immersedin the process. The repeaters are connected to the external capacitive sensor through capillary tubes. A temperature sensorlocated between the two pressure repeaters automatically compensates temperature variations in the process

Special techniques in the production and assembly repeaters and temperature sensor, ensure that small variations in theprocess temperature are quickly informed to the transmitter, that calculates the fluid density process accurately through adedicated software.

Depending on the industrial process, density can be expressed in Density, Relative Density, Brix degree, Baumé degree,Plato degree, % of Solids, Concentration, etc.

Designed for process control applications, this 2-wire transmitter generates a 4-20 mA signal proportional to theconcentration/density. Digital communication for remote calibration and monitoring is also provided (Protocol Hart).

The digital technology used in the DT301 allows the choice of several types of transfer functions, an easy interface betweenthe field and the control room and some characteristics that reduce the installation, operation and the maintenance costsconsiderably.

ATTENTION

Get the best results of the DT301 by carefully reading these instructions.

Page 4: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation & Maintenance Instruction Manual - PreliminaryIV

NOTE

This manual is compatible with version 1.XX, where 1 denotes software version and XX softwarerelease. The indication 1.XX means that this manual is compatible with any release of softwareversion 1.

Page 5: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Index V

IndexINTRODUCTION .................................................................................................................................................IIIINDEX...................................................................................................................................................................V

1 - INSTALATION

GENERAL................................................................................................................................................................................................. 1.1RECOMMENDATIONS FOR USE OF THE DT301 ................................................................................................................................ 1.1TRANSMITTERS MODELS OF CONCENTRATION / DENSITY DT301............................................................................................... 1.1ASSEMBLY............................................................................................................................................................................................... 1.1ELETRONIC HOUSING ROTATION....................................................................................................................................................... 1.4ELETRIC WIRING.................................................................................................................................................................................... 1.4MULTIDROP OPERATION ...................................................................................................................................................................... 1.5

2 - OPERATION

FUNCTIONAL DESCRIPTION - SENSOR.............................................................................................................................................. 2.1FUNCTIONAL DESCRIPTION - HARDWARE........................................................................................................................................ 2.1MAIN BOARD ........................................................................................................................................................................................... 2.2FUNCTIONAL DESCRIPTION - SOFTWARE ........................................................................................................................................ 2.2THE DISPLAY........................................................................................................................................................................................... 2.3

3 - PROGRAMMING USING HAND-HELD TERMINAL

THE HAND-HELD TERMINAL ................................................................................................................................................................. 3.1 FITTING THE BATTERY.............................................................................................................................................................. 3.1 TITTING THE DATAPACK AND RAMPACK............................................................................................................................... 3.1 KEYBOARD .................................................................................................................................................................................. 3.1 TUNING THE PROGRAMMER ON AND OFF ............................................................................................................................ 3.2PROGRAMMING TREE USING HT2 ...................................................................................................................................................... 3.2ON-LINE SINGLE UNIT CONFIGURATION............................................................................................................................................ 3.2 INFORMATION - INFO................................................................................................................................................................. 3.4 CONFIGURATION- CONF........................................................................................................................................................... 3.5 CALIBRATION.................................................................................................................................................................. 3.6 DAMPING ......................................................................................................................................................................... 3.6 MEASUREMENT .............................................................................................................................................................. 3.7 CONSTANTS.................................................................................................................................................................... 3.7 DISPLAY ........................................................................................................................................................................... 3.7 FAIL SAFE ........................................................................................................................................................................ 3.8 MONITORING - MONIT ............................................................................................................................................................... 3.8 TRIM FINE AJUST - TRIM ........................................................................................................................................................... 3.8 CURRENT TRIM (4 – 20 mA)........................................................................................................................................... 3.8 PRESSURE TRIM ............................................................................................................................................................ 3.9 LOWER PRESSURE TRIM.............................................................................................................................................. 3.9 UPPER PRESSURE TRIM............................................................................................................................................. 3.10 MAINTENANCE - MAINT ........................................................................................................................................................... 3.10 CODE.............................................................................................................................................................................. 3.10 SERIAL NUMBER........................................................................................................................................................... 3.10 OPERATIONS COUNT – OP_COUNT.......................................................................................................................... 3.10 BACKUP - BACKUP ....................................................................................................................................................... 3.10 PASSWORD - PASSWORD.......................................................................................................................................... 3.10 CONFIGURATION OF THE PASSWORD LEVEL – CONF_LEVEL ............................................................................ 3.11 WRITE PROTECT – WRITE_PROTECTION............................................................................................................... 3.11ON-LINE MULTIDROP CONFIGURATION........................................................................................................................................... 3.11 CONFIGURING THE DT301 FOR MULTIDROP ...................................................................................................................... 3.12

4 - MAINTENANCE

GENERAL................................................................................................................................................................................................. 4.1DIAGNOSTIC WITH HAND-HELD TERMINAL....................................................................................................................................... 4.1ERROR MENSSAGES ............................................................................................................................................................................ 4.1TROUBLESHOOTING ............................................................................................................................................................................. 4.2DISASSEMBLY PROCEDURE ................................................................................................................................................................ 4.3 GROUP OF THE PROBE.............................................................................................................................................................. 4.3 ELECTRONIC CIRCUIT................................................................................................................................................................ 4.3REASSEMBLY PROCEDURE ................................................................................................................................................................. 4.3 GROUP OF THE PROBE.............................................................................................................................................................. 4.3

Page 6: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation & Maintenance Instruction Manual - PreliminaryVI

ELECTRONIC CIRCUIT................................................................................................................................................................ 4.3INTERCHANGEABILITY.......................................................................................................................................................................... 4.4RETURNING MATERIAL ......................................................................................................................................................................... 4.4EXPLODED VIEW DRAW FOR DT301 – SANITARY MODEL .............................................................................................................. 4.5EXPLODED VIEW DRAW FOR DT301 – INDUSTRIAL MODEL........................................................................................................... 4.6SPARE PARTS LIST FOR TRANSMITTER............................................................................................................................................ 4.7SANITARY MODEL PROBE CODE......................................................................................................................................................... 4.8INDUSTRIAL MODEL PROBE CODE ..................................................................................................................................................... 4.8

5 TECHNICAL CHARACTERISTICS

FILLINGS FLUIDS .................................................................................................................................................................................... 5.1FUNCIONAL SPECIFICATIONS.............................................................................................................................................................. 5.1PERFORMANCE SPECIFICATIONS ...................................................................................................................................................... 5.2PHYSICAL SPECIFICATIONS................................................................................................................................................................. 5.2INDUSTRIAL CONCENTRATION / DENSITY TRANSMITTER CODE ................................................................................................. 5.3SANITARY CONCENTRATION / DENSITY TRANSMITTER CODE..................................................................................................... 5.4

Page 7: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Instalation 1.1

1 – INSTALLATION

GENERAL

The overall accuracy of density measurement dependson several variables. Although the transmitter has anoutstanding performance, proper installation is essentialto maximize its performance.

Among all factors, which may affect transmitter accuracy,environmental conditions are the most difficult to control.There are, however, ways of reducing the effects oftemperature, humidity and vibration.

The capacitive sensor of the DT301, that is locatedexternal to the process, is protected of external sourcesof heat by an enclosure with internal thermal insulation.Nevertheless, the transmitter should be installed in way toavoid to the maximum the direct exhibition to the ray’srun.

Humidity is fatal to electronic circuits. In areas subjectedto high relative humidity, the O-rings for the electronichousing covers must be correctly placed and the coversmust be completely closed by tighten them by hand untilyou feel the O-rings being compressed. Do not use toolsto close the covers. Removal of the electronics cover inthe field should be reduced to the minimum necessary,since each time it is removed; the circuits are exposed tothe humidity. The electronic circuit is protected by ahumidity proof coating, but frequent exposures tohumidity may affect the protection provided. It is alsoimportant to keep the covers tightened in place. Everytime they are removed, the threads are exposed tocorrosion, since painting cannot protect these parts.Code-approved sealing methods should be employed onconduit entering the transmitter. The unused outletconnection should be plugged accordingly.

Although the transmitter is virtually insensitive to vibration,installation close to pumps, turbines or other vibratingequipment should be avoided.

Proper winterization (freeze protection) should beemployed to prevent freezing, since this can result in aninoperative transmitter and could even damage the cell.

RECOMMENDATIONS FOR USE OF THE DT301

The process fluid should always cover the two diaphragmrepeaters. (See the Fig. 1.2).

The maximum speed of the process fluid under thediaphragms repeaters must be 0,4 m/s, which in a pipewith diameter of 6 '' it corresponds to a flow of 26 m³/h.These data applies to fluids with close viscosity of thewater. Fluids with very different viscosity should beanalyzed.

This limitation is due to the pressure drop between thediaphragms.The temperature of the process fluid must be within 0°Cand 120°C.

For applications in corrosive fluids, compatible materialsshould be chosen. The materials of the parts that are notin direct contact with the process, but can be subject tothe corrosive atmosphere or drops of the process, shouldalso be observed.

A possible leak of the fill fluid (less than 5 ml), due to ahole in the diaphragm can contaminate the process. Incase that is not allowed, choose a fill fluid compatible withthe process.

Verify if the fill fluid doesn't evaporate in the extremeconditions of temperature and pressure of the process.

TRANSMITTERS MODELS OF CONCENTRATION /DENSITY DT301

DT301I - Industrial model, for general purpose

DT301S - Sanitary model for food, pharmaceuticalindustry and other applications where sanitaryconnections are required.

The industrial model uses connection flanges accordingto ANSI B16.5 or DIN 2526 standards.

The sanitary model uses Tri-Clamp connection to allow aquick and easy connection and disconnection from theprocess. Wetted surface finish is 32Ra as standard.These model meet 3A recommendations so that theprobe surface is free of crevices where food or bacteriacan be collected. 3A is the most widely accepted sanitarystandard in the food, drug and beverage industry.

ASSEMBLY

Both models have two mounting types: top mounting(straight type) and side mounting (curved type).

The Figure 1.1 shows the dimensions of the DT301straight and curved type for industrial and sanitarymodels.

The installation can be done in open or pressurized tanksor through a sampling device, external to the process.

The Figure1.2 shows some mounting examples.

Choose a place for installation that facilitates the accessfor the measuring point and that be free from mechanicalshocks.

Page 8: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation & Maintenance Instruction Manual - Preliminary1.2

Fig. 1.1 – Dimensions Drawing for the DT301

Page 9: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Instalation 1.3

Fig. 1.2 – Mounting types for the DT301

Page 10: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation & Maintenance Instruction Manual - Preliminary1.4

Try to use a valve in the connection to the process beforethe DT301, this simplifies the calibration and maintenanceof the equipment.

ELECTRONIC HOUSING ROTATIONThe electronic housing can be rotated in order to betterposition the digital display. To rotate it, use the HousingRotation Set Screw, see Figure 1.3

WARNING: EXPLOSION PROOF INSTALLATIONS

The electronic housing and the sensor assembly in potentiallyexplosive atmospheres must have a minimum of 6 threadsfully engaged. The provided joint allows 1 turn extra. Try toadjust the display window position by rotating the housingclockwise. If the thread reaches the end before the desiredposition, then rotate the housing counterclockwise, but not bymore than one turn of the thread end. Transmitters have astopper that restricts housing rotation to one turn. See Section4, Figure 4.1.

The digital display can also be rotated. See Section 4,Figure 4.2.

ELECTRIC WIRINGReach the wiring block by removing the ElectricalConnection Cover. The cover locking screw (Figure 1.3)locks this cover. To release the cover, rotate the lockingscrew clockwise.

Fig. 1.3 – Housing Rotation Set Screw

The wiring block has screws on which fork or ring typeterminals can be fastened. See Figure 1.4.

HAZARDOUS AREAS

In hazardous zones with explosion proof requirements the coversmust be tightened with at least 7 turns. In order to avoid moistureor corrosive gases, hand tight the covers until the O-rings arecompressed. Lock the covers closed with the locking screw.

Observed in hazardous zones with intrinsically safe or non-incendive requirements, the circuit entity parameters andapplicable installation procedures.

Obtain cable access to wiring connections by one of the twoconduit outlets. Should seal conduit threads by means of code-approved sealing methods. Plug and seal the unused outletconnection accordingly. Explosion proof, non-incendive andintrinsic safety Factory Mutual certification are standards forDT301.

Should other certifications be necessary, refer to the certificationor specific standard for installation limitations.

Fig. 1.4 – Wiring Block

For convenience, there are two ground terminals: oneinside the cover and one external located close to theconduit entries.

Use of twisted pair (22 AWG or greater than) cables isrecommended.

Avoid routing signal wiring close to power cables orswitching equipment. Plug and seal the unused outletconnection accordingly.

The DT301 has protection against reverse polarity.Connection of the DT301 should be done as inFigure1.6.

ATTENTION:

For proper operation, the Hand-Held Terminal requiresa minimum load of 250 Ohm between it and the powersupply.

Fig. 1.5 – Load Curve

Page 11: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Instalation 1.5

Fig. 1.6 – DT301 Connection Diagram.

MULTIDROP OPERATIONMultidrop connection is formed by several transmitterconnected to a single communication transmission line.Communication between the host and the transmitterstakes place digitally with the transmitters analog outputdeactivated.

The communication with the transmitters and the host(HT2, DCS, Data Acquisition System or PC) can be donewith a Bell 202 Modem using Hart Protocol. Eachtransmitter is identified by a unique address from 1 to 15.Contact Smar Service to specify the requirements formultidrop applications.

Special consideration should be given to the update ratein multidrop configuration communication.

Not however that the updating of the analog output, incontroller mode, is not affected.

The DT301 is factory set to address 0, that means a nonmultidrop operation mode, allowing the transmitter tocommunicative with the Hand-Held Terminal,superimposing the communication on the 4-20 mAsignal. To operate in multidrop mode, the transmitteraddress must be changed to a number from 1 to 15. Thischange deactivates the 4-20 mA analog output sendingit to 4mA.

NOTE:

The current output will be fastened in 4 mA as soon as thetransmitter address it is changed of zero (0) for another onein the multidrop range (1 to 15).

When intrinsic safety is a requirement, special attentionmust be paid to the entity parameters allowed to thatarea:

Where:

Ca, La Barrier Allowable Capacitance and Inductance

Cij, Lij Non protected internalCapacitance/Inductance of transmitterj (j = up to 15)

Cc, Lc Cable capacitance and InductanceVoc Barrier opens circuit voltage

Isc Barrier short circuit current

Vmaxj Maximum allowable voltage to be applied to theinstrument j

Imaxj Maximum allowable current to be applied to theinstrument j

To operate in multidrop mode, it is necessary to seewhich transmitters are connected on the same line. Theconnection of the DT301 in a multidrop net should bemade according to Fig. 1.7.

[ ]jVVoc maxmin≤

[ ]jaxIsc Immin≤

∑=

+≥n

jj CcCiCa

1

LcLiLa j

n

j

+≥ ∑=1

Page 12: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation & Maintenance Instruction Manual - Preliminary1.6

Fig. 1.7 – Wiring Diagram for the DT301 in Multidrop Configuration

Page 13: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Operation 2.1

2 – OPERATION

FUNCTIONAL DESCRIPTION – SENSORThe pressure sensor used by the DT301 SmartConcentration/ Density Transmitter is a capacitive cell, thesame type used by the LD301 Smart PressureTransmitter. This sensor is connected to a probe toaccomplish the measures through of the pressuredifferential reading. The Figure 2.1 schematizes thesensor used by the DT301.

Fig. 2.1 – Capacitive Cell

Where,

P1 and P2 are the pressures applied in cameras H and L.

CH = capacitance between the fixed plate on P1 sideand the sensing diaphragm.

CL = capacitance between the fixed plate on the P2

side and the sensing diaphragm.

d = distance between CH and CL fixed plates.

∆∆d = sensing diaphragm's deflection due to thedifferential pressure ∆P = P1 - P2.

Knowing that the capacitance of a capacitor with flat,parallel plates may be expressed as a function of platearea (A) and distance (d) between the plates:

Where,

εε = dielectric constant of the medium between thecapacitor's plates.

Should CH and CL be considered the capacity of flat andparallel plates with identical areas, then:

However, should the differential pressure (∆P) applied tothe capacitive cell not deflect the sensing diaphragmbeyond d/4, it is possible to assume ∆P as proportionalto ∆d, that is:

∆P α ∆d

By developing the expression (CL - CH)/(CL + CH), itfollows that:

As the distance (d) between the fixed plates CH and CLis constant. It is possible to conclude that the expression(CL - CH)/(CL + CH) is proportional to ∆d and, therefore,to the differential pressure to be measured.

Thus it is possible to conclude that the capacitive cell is apressure sensor formed by two capacitors whosecapacitances vary according to the differential pressureapplied.

FUNCTIONAL DESCRIPTION - HARDWAREThe transmitter blocks Diagram, as it shows the Figure2.2, it describes the circuit used by the DT301functionally.

PROBEPart of the transmitter that is directly in contact with theprocess. Pressure RepeatersIt transfers to the capacitive sensor the differentialpressure detected in the process.

Temperature SensorIt captures the process fluid temperature.

SENSOR BOARDIt implements the transducer that converts the sensorsign for a measure that can be treated by CPU.

Oscillator It generates a proportional frequency to the capacitivegenerated by sensor.

Signal IsolatorIt accomplishes the isolation of signs between the sensorand CPU. The Control signals from the CPU aretransferred through optocouples, and the signal from theoscillator is transferred through transformer

Memory EEPROM It is a non-volatile memory and it contains the specificinformation of the sensor, such as, construction materials,calibration of the sensor, production and customers data.

dA

C∈

=

edd

ACL

∆−∈

=)2/(.

ddA

CH∆+

∈=

)2/(.

Page 14: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation & Maintenance Instruction Manual - Preliminary2.2

Fig. 2.2 – DT301 Block Diagram Software

MAIN BOARD(CPU) Central Processing Unit and PROMThe (CPU) Central Processing Unit is the intelligent partof the transmitter responsible for the management andoperation of the circuits, sign treatment and accomplishthe communication digital with other devices. Fortemporary data storage, CPU uses the memory positionof its internal RAM. The data stored in this RAM are thosethat can be destroyed in the case of energy lack. Thedata that request its retention, CPU stores them in itsmemory it not interns volatile (EEPROM). This EEPROMmemory admits 10.000 recordings in the same memoryposition.The program is stored in a PROM external memory.

D/A ConverterIt converts the digital data from the CPU to an analogsignal with 14-bits resolution.

OutputIt accomplishes the current control in the line of feedingof the transmitter. This current control is made to form togenerate a proportional current to the value of the variablereading. The work range of the transmitter defines thevalues for the currents 4 and 20 MA. The control currentof the DT301 transmitter obeys the specifications of theNAMUR NE-43 norm.

ModemThe function of this circuit is to become possible thechange of information between the HT2 SMARprogrammer and the DT301 transmitter, through protocolHART. The communication sign is symmetrical and itdoesn't affect the level DC in the output of 4-20mA.

Power SupplyThe transmitter gets the energy of the communication linefor its own operation (transmitter to two wires). Theminimum tension for the transmitter operation is of 12Vdc, measure in the block.

Display ControllerIt controls the lit of the liquid crystal Display segments inagreement with the correspondent data for CPU. Theuser has the option of selecting the variable shown in thedisplay, through digital communication. FUNCTIONAL DESCRIPTION - SOFTWARE The Figure 2.3 shows the software functional diagram ofthe DT301 transmitter.

Digital FilterThe digital filter is a lowpass filter with an adjustable time constantDamping. It is used to smooth noisy signals. The Damping valueis the time required for the output reaching 63.2% for a step inputof 100%.

Customer CharacterizationIt calculates the real pressure through layer capacitivereadings and sensor temperature, considering the data offactory calibration stored in EEPROM of the sensor. Thismodule has as output the values of differential pressureand temperature.

Specific Weight CalculationIt calculates the specific weight of the solution, beingtaken in consideration its properties physical chemistries.

Page 15: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Operation 2.3

Fig. 2.3 – Diagram of Blocks of the SoftwareDensity or Concentration CalculationBeing obtained the value of the specific weight, it can bedetermined its density or concentration easily. In thispoint, obtain the value of the PV main variable, so muchin percentage as in engineering. Current CalculationIt makes the correlation of PV with the current values in

THE DISPLAYThe indicator, constituted by the liquid crystal display, itcan show one or two variables in agreement with theuser's selection. When it is shown two variables, theindicator will alternate between both with an interval ofapproximately 3 seconds. Beyond the numeric and alphanumeric fields, the displayshows some alphanumeric icons to indicate thetransmitter states. The Figure 2.4 shows the segmentconfiguration used by DT301 transmitter.

MonitoringThe DT301 transmitter stays continually in the modemonitoring. In this mode, the indication in the displayalternates between the primary and secondary variable,according to the user's configuration. The indicator hasthe capacity to show the value, the engineering unit andthe variable type, simultaneously with most of the stateindications. See in the Figure 2.4 a sample of a DT301indication standard. The display is capable also to show messages andmistakes (See the table 2.1).

Fig. 2.4 – Typical Monitoring Mode Display ShowingPV, in this case 25,0 BRIX

DISPLAY DESCRIPTION

INITThe DT301 is in initializing afterpower on.

FAILFails in the transmitter. See Section 4– Maintenance.

SATPrimary or secondary Variable out ofthe range operation. See Section 4 -Maintenance.

Tabela 2.1 - Display Errors and Messages

Page 16: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Programming Using Hand-Held Terminal 3.1

3 – PROGRAMMING USING HAND-HELD TERMINAL

THE HAND-HELD TERMINAL

\SMAR Hand-Held Terminal (HT2) is the man-machine interfaceused to maximize the advances of digital technology. Theterminal's front face is shown on Figure 3.1.

Fig 3.1 - SMAR Hand-Held Terminal

SMAR Hand-Held Terminal (HT2) Software has the followingfeatures:

§ Transmitter identification and specification data.

§ Remote reranging without using a reference pressure source.

§ Constant current adjustment from 3.9 to 21 mA for loop test.

§ Monitoring of all transmitter variables: PV, PV%, output, error,and process temperature.

§ Diagnosis and determining faults in the processor or thetransmitter.

The operations, which take place between the HT2 and thetransmitter do not interrupt the Pressure measurement, and donot disturb the output signal. The HT2 can be connected on thesame pair of wires as the 4-20 mA signal, up to 2 km away fromthe transmitter. It is necessary to introduce an impedance ( eitherinductive or reactive ) between the power supply and thetransmitters, so that the communication signal may besuperimposed to the power supply level.

The standard model shall not be used in Hazardous Areas.

Fitting the BatteryThe HT2 uses any 9 Volt PP3 size battery, however for longeroperating life, use of alkaline batteries is recommended. Removethe protective case by pulling it firmly downwards. Pull the tab toremove the battery cover and observe polarity. The programmerstandard model cannot be utilized in hazardous areas.

Fitting the Datapack and RampackThe datapack is the module where the DT301 configurationsoftware for the HT2 is stored. Make sure it is marked DT301,and put it into slot B. Another Rampack/Datapack can be usedfor storage of off-line configurations into slot C, although data canbe also stored in the HT2 RAM. After loading the configurationsoftware on the programmer, the software will remain resident onits RAM memory until the programmer is de-energized.

The following programming packs are available for the user:

ü Data Pack – this pack is built based on PROM or EPROMmemories;

ü RAM Pack – this pack is built based on RAM memories;ü Flash Pack – this pack is built based on Flash memories.

ATTENTION:

Should RAM packs be used, do not energize the programmer while thePACK is connected to it!

The RAM Pack could be formatted and the data would be deleted.

KeyboardThe Hand-Held Terminal (HT2) has double function keys. Thefunction labels are indicated on the keys themselves, and abovethem. Simply touching them activates the functions engraved onthe keys, while the functions engraved over the keys are activatedby simultaneously touching the key < SHIFT > and the keycorresponding to the desired function.The following keys are used for DT301 programming.

It is used to turn the programmer on or to return to theprevious decision level on the menu. Should you be

unable to read anything on the display, or if there is somethingdifficult to interpret, simply adjust the contrast. In order to locatethe contrast adjustment, refer to Figure 3.1.

These keys are used to move the cursor between linesand columns on the display.

Used to access the symbols and numbers above thekeys. Must be pressed simultaneously with the desired

key when the HT2 is in the alphanumeric mode. To obtain theselection, just press simultaneously <SHIFT> and the symbol ornumber desired. For data entry that requires only numbers, the<SHIFT> key is not necessary.

Used to delete characters wrongly typed.

Used to leave a space between characters.

Used to confirm an action or complete an input.

Page 17: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation & Maintenance Instruction Manual - Preliminary3.2

TURNING THE PROGRAMMER ON AND OFFWhen the programmer is turned on for the first time, or upon abattery replacement, the following message will be displayed:

00:08a

....................DT301 OFF

Moving the cursor to function OFF and pressing the <EXE> keyor pressing the Key <O> will turn the unit off. If the terminal is ONfor five minutes without any keys being pressed, it will switch itselfOFF.

With the cursor blinking under DT301, press the <EXE> key. Theoperational system is transferred from the plug-in EPROMmemory to the HT2 RAM memory, and the display will show:

HAND HELD TERMINAL MODEL HT2 ....................

Version 1.xx

After a few seconds, the following menu will show the options forconfiguration and the option for return to the previous way.

SMAR - HT2On_Line_Single_UnitOn_Line_MultidropExit

Both <ON/CLEAR> key and the EXIT function allow the user toleave the MENU in use and return to the next higher MENU inhierarchy.

They are quite useful to return to a familiar menu when the useris lost in an unfamiliar operation.

PROGRAMMING TREE USING HT2The Programming tree is a tree-shaped structure with a menu ofall the available software resources, as shown in Figure 3.2.

NOTE:

All transmitters are factory configured with no passwords. To avoidoperation by nonauthorized persons in some critical levels of theProgramming Tree, it is recommended to configure all passwords beforeoperation. See "PASSWORD" option in Maintenance section.

On-Line Single Unit: is used when the HT2 is connected inparallel to a single transmitter and this transmitter has the address0 (Zero).

On-Line Multidrop: is used when the HT2 is connected inparallel to several transmitters (up to 15) and these transmittersare configured with different addresses of 1 to 15(see On-LineMultidrop Operation).

The ON-LINE options can access the complete configurationmenu.

ON-LINE SINGLE UNIT CONFIGURATIONTo configure the transmitter on-line, certify that it is correctlyinstalled, with a suitable power supply and the minimum 250 loadrequired (See Fig. 1.6).

With the cursor blinking under the "ON_LINE_ SINGLE_UNIT"option, press the <EXE> key. The display will show:

Searching for XMTR ...

If the HT2 is connected to a DT301, the display flashes amessage informing that the transmitter is a density/concentrationtransmitter (XMTR) and its software version is 1.XX. Immediatelyafter that, the main menu is displayed. If the HT2 is connected toanother HART device, the HT2 will be limited to a genericdensity/concentration device mode, in which the options arelimited.

Page 18: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Programming Using Hand-Held Terminal 3.3

Fig. 3.2 – Programming Tree Using Hand-Held Terminal

Page 19: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation & Maintenance Instruction Manual - Preliminary3.4

>>>> SMAR DT301 <<<<

Density TransmitterVersion 1.xx

/ TAG XMTR

INFO CONF MONITTRIM MANUT EXIT

The Main Menu includes the following information:

/TAG – this indicates the root node of the main menu;

INFO – here are included the chief information concerning thetransmitter configuration;

CONF – here are included the parameters related with theconfiguration of the transmitter;

MONIT – here the variables monitoring options on the transmitterare shown;

Fig. 3.3 – Terminal Information Items

TRIM – here are the interactive adjustments for the variables inthe transmitter;

MAINT – this includes the maintenance options for the transmitter;

EXIT – this is the way to return to the previous level of the menu.The key < ON > performs the same function.

INFORMATION – INFOUpon the selection of option INFO, the display will show its items.By using the keys < ↓ > and < ↑ >, it is possible to access allinformation items , as shown on Figure 3.3. In order to introducea modification on the selected item just press the key < EXE >.

/INFO TAG XMTR

Tag:XXXXXXXX

/INFO TAG XMTRTag:XXXXXXXXChange it? Y/N

By pressing the key < N >, the item will be displayed again. Should the key < Y > be pressed, it is possible to edit a newcontent and to send it to the transmitter by pressing the key <EXE >.

The option INFO shows the following information concerning thetransmitter:

TAG - 8-character alphanumeric field for identification of thetransmitter. Remember to use the <SHIFT> key for numbers orsymbols.

DESCRIPTOR - 16-character alphanumeric field for additionalidentification of the transmitter. May be used to identify service orlocation.

DATE MODIFIED - The date may be used to identify a relevantdate as the last calibration, the next calibration or the installation.The date is presented in the form of Day, Month, and Year.Month should be entered as a number 1-12. Year as a number1900 - 2155.

MESSAGE - 32-character alphanumeric field for any other IInformation, such as the name of the person who made the lastcalibration, some special care to be taken, or if a ladder is neededfor accessing.

FLANGE TYPES – φ 4” x 150 # ANSI B16.5 RF, φ 4” x 300 #ANSI B16.5 RF, φ 4” x 600 # ANSI B16.5 RF, DN 100 PN25 / 40DIN 2526 – Form D, Special.

FLANGE MATERIALS – Stainless Steel AISI 316, Hastelloy C276, and Special.

O_RINGS MATERIALS- Buna-N, Viton, Teflon or Special.

Page 20: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Programming Using Hand-Held Terminal 3.5

INTEGRAL METER - Installed, None.

DRAIN/VENT MATERIAL – Carbon Steel, 316 SST, Hastelloy,Monel, None, Unknown and Special.

REMOTE SEAL TYPE - Straight type or Curved Type

REMOTE SEAL FLUID FILLING – Silicone DC200 / 350,Silicone DC200 / 20, Glycerin / Water, Sylthern 800 or Special.

SENSOR FLUID ∗∗∗∗ - Silicone, Fluorolube or Special.

REMOTE SEAL DIAPHRAGM - Stainless Steel AISI 316L,Hastelloy C 276 or Special.

SENSOR ISOLATION DIAPHRAGM ∗∗∗∗ - Stainless Steel AISI316, Hastelloy C, Monel, Tantalum or Special.

SENSOR TYPE ∗∗ - This shows the type of sensor.

SENSOR∗∗ RANGE ∗ - This shows the sensor range expressedin the engineering unit chosen by the user refer to UnitConfiguration.

REMARK:

All information items marked with (∗) are not applicable to the DT301and those marked with (∗∗) can not be modified, since they comedirectly from the sensor memory.

CONFIGURATION - CONFThis function affects the 4-20 mA output of the transmitter andthe transmitter display indication. Here the transmitter can bereranged, having the damping adjusted or the outputcharacteristics changed. The Engineering Units displayed on thetransmitter and on the HT2 can also be changed. If the CONFoption is selected in the main menu, the display will show:

Fig. 3.4 –Terminal Configuration Tree

Page 21: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation & Maintenance Instruction Manual - Preliminary3.6

/CONF TAG XMTRCalibrationMeasur = SGU@20oCConstants ...

/CONF TAG XMTRDisp. = [PV] [PV ]Fail_Safe_DownScaleExit

Using the keys < ↓ > and < ↑ > all the configuration items can beaccessed.

If you intend to change one item press the <EXE> key. Threepoints (...) following an items indicates that there is a sub-menuwith more items.

Figure 3.4 shows all options for function CONF in the SMAR HT2 Programmer.

CALIBRATIONIn order to perform calibration, move the cursor to Calibration andpress key < EXE >. The screen will display the followingmessages, depending on the way the keys < ↑ > and < ↓ > areused:

/CONF TAG XMTR 4mA = 10.0 Brix20mA = 96.0 BrixDown_Concent ...

/CONF TAG XMTRUP_CONCENT ...DAMP. = 10.00 SExit

Calibration of the Output CurrentThe output current shall be calibrated so that lower concentrationvalue represents 4.00 mA and the upper concentration valuerepresent 20.00 mA. This may be done by using the Menu:Calibration / Current 4mA and Calibration / Current 20mA. Inorder to calibrate the 4.00 mA current, place the cursor in 4 mAand press key < EXE >. After confirming the intention ofchanging the parameter, enter the new value corresponding tothe lower current, as shown on the tables below:

/CONF TAG XMTR[Down Concentration]4mA = 0 BRIXChange it? Y/N

/CONF TAG XMTR[Down Concentration]Type Lower Range4mA(BRIX) = _

/CONF TAG XMTR[Down Concentration]Type Lower Range4mA(BRIX) = 10.0_

Repeat de same procedure to calibrate 96.00 Brix.

On the example, the chosen working range is 10 to 96 Brix. It isalso possible to carry out the calibration in a reverse manner thatis, opting to enter 96 Brix as 4 mA and 10 Brix as 20 mA.

Measurement CalibrationThe DT301 transmitter is manufactured and calibrated inaccordance with the requirements of the client. During its processinstallation, however, it may be necessary to perform ameasurement adjustment, due to some eventual deviationsoccurred during the installation. Should the required adjustmentinvolve nothing more than the measurement engineering units,refer to item MEASUREMENT described hereafter. Should theadjustment involve altering the measurement values, thereference calibration shall be performed, in accordance with thefollowing steps:

ü Stabilize the process and collect a sample;ü Connect the HT2 Programmer to the DT301 communication

line and select option Down_Concent on the CONF menu;ü Determine in the laboratory the concentration / density value of

the stabilized process;ü After confirming the intention to perform the change confirm

the sample collection so that the value measured in thelaboratory may be entered.

/CONF TAG XMTR4mA = 10.0 BRIX20mA = 96.0 BRIXDown_Concent ...

/CONF TAG XMTR[Down Concentration]

Change it ? Y/N

/CONF TAG XMTR[Down Concentration]Conc 11.3 BRIXCollect sample? Y/N

/CONF TAG XMTR[DownConcentration]Conc = 11.3 BRIXMeasur. (Lab) =

DAMPINGThe damping item in the range menu enables adjustment of theinput damping performed by the software.

The damping is a digital filter which time constant, may beadjusted between 0 and 32 sec. The transmitter has amechanical damping of 0.2 sec.

Placing the cursor on option Damping in the CONF menu, presskey < EXE > in order to enter the new value. The followingsequence of screens shows the interaction necessary to performthe alteration.

Page 22: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Programming Using Hand-Held Terminal 3.7

/CONF TAG XMTR[Down Concentration][Up Concentration ]Damp. = 10.00 s

/CONF TAG XMTR[ Damping Set ]Damp. = 10.00 sChange it ? Y/N

/CONF TAG XMTR[ Damping Set ]Input Damping:Damp. =

MEASUREMENTThis configuration menu function of the SMAR HT 2 Programmermakes it possible to select the type of transference function thetransmitter is expected to perform. There are several functionsrelated with the measurement of density and concentration, andthere is a special function which makes it possible to check the 4to 20 mA current generated by the transmitter. The followingtransference functions have been implemented:

DensityThe transference functions related to the measurement of densitycorrespond to the measurement of absolute densitymeasurement, which takes into consideration the chemicalproperties of the solution and the physical properties of themedium, and to the measurement of specific gravity, which isbased on the density of water. Therefore, it is possible to performmeasurements in the following units: kg/m3, g/cm3, SGU@ 20º C,SGU@ 4ºC.

ConcentrationSuch measurements inform the composition of a solution incomparison with several worldwide accepted units, such as:Baumé Degrees, Plato Degrees, Brix Degrees and INPMDegrees.

Constant OutputThis measurement allows the user to check the consistency ofthe input current generation values between 3.9 and 21 mA. Thisis also an extremely important characteristic while performing theLoop Testing during the Startup of an Industrial plant.

/CONF TAG XMTR[ Constant Output ]Enter_Const._CurrentExit_Const._Current

Example: Upon entering < EXE > while the cursor is on theMeasurement option of the configuration menu, informationrelated with the measurement selection will be displayed.Entering < Y >, the measurement options shown on Figure 3.4may be selected.

CONSTANTSAll measurements performed by the transmitter, with theexception of the fixed current, are calculated in accordance withthe physical and chemical properties of the solution and

the medium, respectively. However, due to intrinsic characteristicsof this transmitter, the only information required for the transmitterto perform correct measurements is the gravity acceleration,expressed in m/s2. Such information is provided in this option ofthe configuration menu.

ATENTION:

Alter this variable only if the local gravity acceleration figure is known.The use of an incorrect figure will cause measurement errors in thespan, which will require a new calibration. In order to correct amistakenly entered information, since return to the normal accelerationfigure, 9.80665, if no adjustments have been made.

Example: How to alter the default value of the gravity accelerationfrom 9.80665 m/s2 to 9.8 m/s2 : by entering < EXE > with thecursor placed on option Constants and then on the variable gl – local gravity, both display types will be presented. Upon entering < Y > to confirm the alteration, the programmer will enter theediting mode, thus allowing a new value to be entered inreplacement for the previous value. By using key < EXE > it ispossible to confirm the alteration that has been performed. It isnot necessary to enter the figures together with key < SHIFT >,since the programmer is prepared to recognize only numerickeys.

/CONF TAG XMTR

gl = 9.8066492Exit

/CONF TAG XMTR

gl = 9.8066492Change it ? Y/N

/CONF TAG XMTR[Down Concentration]Conc = 5.4 BRIXLaboratory

Repeat procedure to calibrate the upper concentration.

DISPLAYThis option makes it possible to configure up to two variables tobe shown on the display of the transmitter. Should the user forusing only one variable, the same variable shall be entered as thesecond variable or, alternatively, none shall be chosen as theoption for the second variable.

Example: With the cursor placed on option Disp., which has beenconfigured to show the same variable twice, enter < EXE >. Thenext screen to be displayed will be the one that makes it possibleto perform the change. By entering < Y >, the configuration ofboth variables will be presented, in accordance with the sequenceof displays shown below. By entering < EXE > with the cursorplaced on each one of the variables, it is possible to select anyvariable among those, which are available for visualization, asshown on Figure 3.4.

/CONF TAG XMTRMensur. = BrixConstants ...Disp. = [PV ][PV ]

Page 23: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation & Maintenance Instruction Manual - Preliminary3.8

/CONF TAG XMTR[ Meter Indication ]Disp. =[PV ][PV ]Change it ? Y/N

/CONF TAG XMTR[ Meter Indication ]First Var. = [PV ]Second Var. = [PV ]

FAIL SAFEThis configuration option includes the possibility of choosing theoutput current action, should a failure occur. The output currentwill remain fixed within the limits of Lower Burnout or UpperBurnout, depending on the failure mode chosen.

The user does not choose the upper and lower burnout currentlimits. These limits are previously determined in accordance withthe transmitter version. The lower current limit is 3.9 mA and thelatest versions comply with the specifications of standard NAMURNE-43, that is , 3.6 mA. In what respects to the upper limit, allversions use the same limit: 21 mA. The selection of DownBurnout and Upper Burnout is done by means of mode switchingdevice, by pushing key < EXE > on option BURNOUT.

/CONF TAG XMTRConstants . . .Disp. =[PV ][PV ]Fail_Safe = UpScale

/CONF TAG XMTR

Fail_Safe = UpScaleChange to Down ? Y/N

MONITORING - MONITThis function allows remote monitoring of 1 of the transmitter'svariables on the display of the HT2. To activate it, select MONITin the main menu. By pushing key < EXE > with the cursor onMONIT, the display will show:

/MONIT TAG XMTR[ CHOOSE VARIABLE ]OUT PV% PVTEMP EXIT

By choosing the PV for monitoring, for example, the programmerdisplay will show the screen and a compassed and intermittentsound signal will indicate each scan update. Should the dataacquisition fail, each failure will be identified by a longer soundsignal.

/MONIT TAG XMTRMonitoring: PV 9.7 BRIXON/CLEAR> to Exit

TRIM FINE ADJUST - TRIMThe TRIM function is used to make the transmitter readingcomply with the user's pressure and current standards.These two types of readouts, Current and Pressure, may beadjusted either in fine tuning, in order to correct minor hardware

deviations or, as in the particular case of Pressure, to determinethe working range selected by the User.

CURRENT TRIM (4-20 mA)When the microprocessor generates a 0% signal, the Digital toAnalog converter and associated electronics are supposed todeliver a 4 mA output. If the signal is 100%, the output should be20 mA.

There might be differences between the factory current standardsand your plant current standard. In this case, you may use thecurrent TRIM adjustment as follows:

§ Connect the transmitter to your plant precision milliammeterwith a 0.03 % precision readout.

§ Select the TRIM function. The display will show:

/TRIM TAG XMTR

Pressure CurrentExit

Move the cursor is blinking under CURRENT, just press the<EXE> key. After a moment, the display will show:

/TRIM TAG XMTR

4mA 20mA StartExit

Select the desired point, e.g., 4 mA and press the <EXE> key.The display will show:

/TRIM TAG XMTR[4mA Current Trim]Enter Line Currenti(mA) = _

/TRIM TAG XMTR[ Trim de i (4mA) ]

Valor de Corrente

If the reading is 4.00 mA, answer Yes <Y>. If the reading is, forexample, 4.012 mA, press <N> (No).

/TRIM TAG XMTR[ Trim de i (4mA) ]Entrar valor de i:i(mA)= _

Type 4.01 and press <EXE>.

The display will repeat the question if the current is correct or not.Answer it properly till the reading is 4.00 mA. Enter the sameprocedure shall be applied for the 20 mA.

PRESSURE TRIMEach sensor has a characteristic curve that establishes a relationbetween the applied pressure and the sensor signal. This curveis determined for each sensor and it is stored in a memory of thesensor circuit. When the sensor is connected to the transmittercircuit, the contents of its memory is made

Page 24: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Programming Using Hand-Held Terminal 3.9

available to the microprocessor, that relate the sensor’s signal tothe measure pressure.

Sometimes the transmitter display and/or the HT2 display readingmay differ from the applied pressure.

The reason may be:

§ The transmitter mounting position.§ The user's pressure standard differs from the factory standard.§ The transmitter had its original characterization shifted by

overpressure, overtemperature or by long term drift.

The pressure TRIM is used to match the reading with the appliedpressure.

There are two types of pressure trim available:

1. LOWER TRIM: Is used to trim the reading at the lower range.The user informs the transmitter the correct reading for theapplied pressure via the HT2.

2. UPPER TRIM: Is used to trim the reading at the upper range.The user informs the transmitter the correct reading for theapplied pressure via the HT2.

By selecting TRIM and PRESSURE , the display will show:Press the following cursor keys: < ↓ > and < ↑ > to select the trimoptions.

Move the cursor to the desired option and press key < EXE >

/TRIM TAG XMTRLower_PressureUpper_PressureExit

ATTENTION:

This is a factory adjustment, so the user must not perform it on the field.Should it be necessary to perform this adjustment on the field, apressure standard with an accuracy equal to or better than 0.015% ofthe pressure to be applied must be used

The upper pressure trim shall be adjusted only after the adjustment ofthe lower pressure trim..

LOWER PRESSURE TRIMThe most common discrepancy is the lower reading. The shift inZERO may be compensated by the LOWER PRESSURE TRIM.To have a digital reading more accurate it is recommended tocalibrate LOWER PRESSURE TRIM with the same valuecalibrated in the LOWER RANGE VALUE.

After consequently selecting TRIM, and LOWER PRESSURE,the display will show:

/TRIM TAG XMTR[Lower Press. Trim ]P(mmH2O20)= 510Is it correct ? Y/N

Should the readout be sufficiently close to the pressure applied,press key < Y > (yes); otherwise, press key < N > (no).Supposing that, in this example, the applied pressure is 50mmH2O, you shall press key < N > in order to make thecorrection.

The display will show:

/TRIM TAG XMTR[Lower Press. Trim ]Enter pressure valueP(mmH2020) = _

Enter the correct value; type 500 followed by <EXE>. The displaywill show:

/TRIM TAG XMTR[Lower Press. Trim ]P(mmH2O20) = 500Is it correct ? Y/N

Since it is now corrected, press <Y> to confirm. The whole curvewas shifted to cancel the offset.

UPPER PRESSURE TRIMUnder some particular operating conditions, the sensormechanical gain may be slightly modified. This reflects on thepressure reading. In order to adjust the reading to the desiredvalue, apply the UPPER RANGE VALUE of the desiredcalibration range and select the UPPER PRESSURE TRIM. Inorder to have a more accurate digital reading it is recommendedto calibrate the UPPER PRESSURE TRIM with the same valueas the calibrated UPPER RANGE VALUE.

The Upper Pressure TRIM corrects any distortions occurred onthe mechanical gain of the sensor, by acting on the transferencecurve span. The Upper Pressure Trim adjustment procedure isidentical to that of the Lower Pressure trim, which has beenexplained on the previous paragraph. All steps shall be followedin the same order, simply changing the reference from lower toupper.

MAINTENANCE - MAINTThis function makes it possible to access all information, whichare relevant for the maintenance of the transmitter. Figure 3.5shows all options offered by this function.

FORMAT - FORMATFormat identifies the " Code of Request " in agreement with theuser's specification.CODEShows the "Ordering Code" according to the Specifications of thecustomer specification.

Example:

DT301I - 1 I S - 1 0 1 - 5 1 - 0

Density / Concentration Transmitter Model DT301I: IndustrialModel; Range: 0.5 to 1.25 g/cm³ (1); Diaphragm made ofStainless Steel 316L (I) and filled with Silicone Oil DC 200 / 20 (S ); equipped with Local Digital Indicator ( 1 ); ElectricalConnection of ½ inches NPT ( 0 ); Straight Assembly ( 1 );Process Connection φ 4 inches ANSI B16.5 ( 4 ); Pressure Class150# ( 1 ); No other special characteristics ( 0 ).

NOTE:

Upon every configuration modification, it is recommendableto reconfigure INFO as well as items MAINT / IDENT. /CODE#.

SERIAL NUMBERThree serial numbers are stored:

Page 25: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation & Maintenance Instruction Manual - Preliminary3.10

Circuit Number - This number is unique to every main circuitboard and can not be changed.

Sensor Number - The serial number of the sensor connected tothe DT301 and can not be altered.

Device ID Number - The number that is written on theidentification plate of the transmitter. This number must be alteredevery time that the main board is changed, in order to avoid digitalcommunication problems.

OP_COUNTEvery time a change is made, there is an increment in therespective change counter.

The monitored items are:

ü Concentration 4/20 mA;

ü Down/Up_Conc (All Down/Upper Concentration);

ü Mensuration (All Mensuration Calibrated);

ü Concentration (All test of loop);

ü Trim_4mA (All trims in 4 mA);

ü Trim_20mA (All trims in 20 mA);

ü Lo_Press_Trim (All Lower Pressure Trims);

ü Up_Press_Trim (All Upper Pressure Trims);

ü Multidrop (any changes in the communication mode as, forexample, multidrop or single transmitter);

ü PSWD / C_LEVEL (any changes in the password or in theconfiguration level);

ü Writing protection (any changes in writing protection);

BACKUPWhen the sensor or main circuit is changed, it is necessary,immediately after the assembly, to transfer the data of the newsensor to the main board. This is done automatically, when thetransmitter is powered on. If necessary, you can force the transferusing the Read_from_sensor option.

PASSWORDThere are three levels of passwords, which may be configured bythe programmer, thus allowing each branch of the programmingtree to have differing accesses as per this hierarchy. In case thepassword has not been configured, the user will have free accessto all branches of the programming tree. In order to configure apassword, access function MANUT.

Display will show:

/MANUT TAG XMTRFormat Op_CountBackup PasswordConf_Level

Select the PASSWORD option, the display will show:

/MANUT TAG XMTR[ Change Password ]Level_1 Level_2Level_3 Exit

Password Level 3 is hierarchically superior to password level 2,which is superior to level 1.

Move the cursor to the level corresponding to the function youwant to password protect and press the <EXE> key. The displaywill show:

Fig. 3.5 - Options of the MANUT Function

Page 26: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Programming Using Hand-Held Terminal 3.11

/MAINT TAG XMTR

Change PasswordLevel_1 ? Y/N

If <N> is pressed, the password is not changed. If <Y> ispressed, the display shows:

/MANUT TAG XMTR

Enter New PasswordPSWL_1 = _

Enter with the password and press the <EXE> key. In case ofactivation of a password or to cancel, the password configured inthis level type six times the space key.

ATTENTION:

In case of loss or forgetfulness of passwords, contact SMARService.

CONFIGURATION OF THE PASSWORD LEVEL -CONF_LEVEL

This function allows the user to modify the default password levelfor each type of operation. The configuration of the passwordlevels (1-3) is stored in the DT301 EEPROM.

Passwords can be configured for all branches of theprogramming tree except MONIT. Should the password of theconfigured level for the specific access to the branch not beactivated, its operation will be free. Should the password level beconfigured, the access to TRIM and MAINT will be completelyprotected; in the case of INFO and CONF, however, it is possibleto access some kinds of information without entering thepassword.

The level configuration of the password presents the followingdefault configuration:

/MAINT TAG XMTR[Level of Passwords]/INFO = 1/CONF = 2

/MAINT TAG XMTR[Level of Passwords]/TRIM = 3/MAINT = 3

WRITE PROTECT - WRITE_PROTECTIONThis feature is used to protect the transmitter configuration fromcorruption via communication or via local adjustment.All data can be protected through by this protection.

The screen relatives of these protections are:

/MAINT TAG XMTR[ Protection Mode ]Not Write ProtectedExit

/MAINT TAG XMTR

Not Write ProtectedEnable ? Y/N

/MANUT TAG XMTR[ Protection Mode ]Write ProtectedExit

ON-LINE MULTIDROP CONFIGURATIONThis operation mode makes it possible to perform the onlineconfiguration of a specific transmitter in a multidrop bus. In orderto accomplish this, the user should know, beforehand, thenetwork address attributed to the target transmitter. So that it isnot necessary to carry out an online investigation in order toidentify the transmitter, that is , calling option IDENT for eachtransmitter on the network until identifying the desired one.

To communicate with a specific transmitter in the way multidrop,select " ON_LINE_MDROP ". After the programmer to detect thetransmitters in the line, select the address of the transmitterwanted starting from the menu. When will appear " PresentAddress = XX. To "Change it ? Y/N", press <N> and the key<EXE> to continue.

IMPORTANT:

By pressing key < N >, after “Present Address = xx; Change if ? Y/N”,the programmer will enter the Online Multidrop Configuration Mode;

By pressing < Y >, the programmer will go to the Address ChangeMode.

The following sequence of screens show an example of how toreach the Online Multidrop Configuration options.

11:23a....................

OFFDt301

SMAR HT2On_Line_Single_UnitOn_Line_MultidropExit

MDROP

Polling Address X

MDROP

AD_1 AD_2 AD_3AD_4 AD_5 AD_6

Place the cursor on the address of the transmitter to beconfigured and press key < EXE >. This will lead to the branch ofthe tree where it is possible to configure or alter the address. Bychoosing “No Address Change” , the configuration options will bedisplayed. From this point on, the configuration procedure will beidentical to the one described for the case of Online Configurationof a Single Transmitter.

Page 27: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation & Maintenance Instruction Manual - Preliminary3.12

MDROP

Present Address = 1Change it ? Y/N

By pressing key < N >, the following screens will be displayed:

>>>> SMAR DT301 <<<<

Density TransmitterVersion 1.xx

/ TAG XMTR

INFO CONF MONITTRIM MAINT EXIT

CONFIGURING THE DT301 FOR MULTIDROP

This operation mode of the programmer makes it possible to builda multidrop bus bar with up to 15 HART instruments. This shallbe done by inserting transmitters with address 0, one by one, untilreaching the total number of transmitters. Initially, connect thetransmitter with address 0 to the bus, then connect it to theprogrammer and reconfigure its new address on the multidropbus. Keeping this transmitter connected to the line, connectanother transmitter with address 0 and repeat the procedure untilthe network is completed. The address change operationcomprises the steps described as follows:

11:23a....................

OFFDt301

A few seconds after selecting "DT301" on the initial menu , thenext menu to be shown is the one, which offers the optionbetween multidrop and single transmitter.

SMAR HT2On_Line_Single_UnitOn_Line_MultidropExit

Placing the cursor on option "ON-LINE MDROP" press key <EXE >. The programmer will then consult the line in order toidentify the transmitters in multidrop connection. The checkingdisplay will be shown as follows:

MDROP

Polling Address X

Where X will scan all the valid addressing space, that is, from 0to 15. As the transmitter is being configured for the first time, theaddresses to be found will be 0 and all the others previouslyconfigured on the network. Therefore, upon placing the cursor onthe located address END_0, press key < EXE >. The sequencefor address alteration will be the confirmation of the alteration andthe selection of a new address among those, which are valid and

available in the bus, as shown on the following screens:

MDROP

AD_1 AD_2 AD_3AD_4 AD_5 AD_6

MDROP

Present Address = 0Change it ? Y/N

MDROPChose new address:AD_3 AD_4 AD_5AD_6 AD_7 AD_8

MDROP

AD_1 AD_2 AD_3

Page 28: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Maintenance 4.1

4 - MAINTENANCE PROCEDURES

GENERAL

SMAR DT301 intelligent pressure transmitters are extensivelytested and inspected before delivery to the end user.Nevertheless, during their design and development, considerationwas given to the possibility of repairs by the end user, ifnecessary.

As main characteristics how much to the maintenance easinessit can be detached its modularity and its reduced number ofelectronic board.

In general, it is recommended that end users do not try to repairprinted circuit boards. Spare circuit boards may be ordered fromSMAR whenever necessary.

The DT301 concentration/density transmitter has been designedto operate for many years without malfunctions. Should theprocess application require periodic cleaning of the repeaterdiaphragms, the flanges may be easily removed and reinstalled.

Should the transmitter eventually require maintenance, it may bechanged in the field. In this case, the possibly damaged sensorshould be returned to SMAR for evaluation and, if necessary,repair. Refer to the item "Returning Materials" at the end of thisSection.

DIAGNOSTIC WITH HAND-HELD TERMINALShould any problem be noticed relating to the transmitter output,investigation may be carried out bythe HHT, as long as power is supplied and communication andthe processing unit are operating normally (see Table 4.1).

The programmer should be connected to the transmitteraccording to the wiring diagram shown on Section 1, Figures, 1.6and 1.7.

ERROR MESSAGES

When communicating using the HHT the user will be informedabout any problem found by the transmitter self-diagnostics.

To follow, some examples of mistake messages shown in theprogrammer's Display.

> OUTPUT SATURATED <

> OUTPUT SATURATED <

> OUTPUT SATURATED <

The mistake messages always are alternate with the informationshown in the first line of programmer SMAR'S Display. The Table4.1 lists the mistake messages. For more details on the correctiveaction it see referred table.

ERROR MESSAGES POTENTIAL SOURCE OF PROBLEM

PARITY ERROR

OVERRUN ERROR

CHECK SUM ERROR

FRAMING ERROR

• The line resistance is not according to load curve.

• Excessive noise or ripple.

• Low level signal.

• Interface damaged.

• Power supply or battery voltage of the HHT lower than 9 V.

NO RESPONSE

• Transmitter line resistance is not according to load curve.

• Transmitter not powered.

• Interface not connected or damaged.

• Transmitter configured in Multidrop mode being accessed by ON LINE SINGLE UNIT.

• Transmitter reversibly powered (polarity is reversed).

• Interface damaged.

• Power supply or battery voltage of the HHT lower than 9 V.

LINE BUSY • Other device using the line.

CMD NOT IMPLEMENTED• Software version not compatible between HHT and transmitter.

• HHT is trying to carry out a DT301 specific command in a transmitter from another manufacturer.

XMTR MALFUNCTION• Sensor disconnected.

• Sensor failure.

COLD START • Start-up or Reset due to power supplies failure.

OUTPUT FIXED• Output in Constant Mode.

• Transmitter in Multidrop mode.

OUTPUT SATURATED • Pressure out of calibrated Span or in fail-safe (Output current in 3.90 or 21.00 mA).

SV OUT OF LIMITS• Temperature out of operating limits.

• Temperature sensor damaged.

PV OUT OF LIMITS • Pressure out of operation limits.

Page 29: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation and Maintenance Instruction Manual – Preliminary4.2

ERROR MESSAGES POTENTIAL SOURCE OF PROBLEM

• Sensor damaged or sensors module not connected.

• Transmitter with false configuration.

LOWER RANGE VALUE TOO HIGH • The 4 mA point was set to a value above a value corresponding to (upper range limit ! minimum span).

LOWER RANGE VALUE TOO LOW • The 4 mA point was set to a value below a value corresponding to (!upper range limit).

UPPER RANGE VALUE TOO HIGH • The 20 mA point was set to a value above the 1.24 H (upper range limit).

UPPER RANGE VALUE TOO LOW • The 20 mA point was set to a value below a value corresponding to (!upper range limit + minimum span).

UPPER & LOWER RANGE VALUESOUT OF LIMITS • C Both the 4 and 20 mA points were outside the sensor's range limit.

SPAN TOO SMALL • The difference, between the 4 and 20 mA points, is less than the 0.75 H (minimum span) allowed by thetransmitter.

APPLIED PROCESS TOO HIGH • The pressure applied on the sensor was above the 1.24 H (upper range limit).

APPLIED PROCESS TOO LOW • The pressure applied on the sensor was below the !1.24 H (upper range limit).

EXCESS CORRECTION • During digital trim, the trim value entered exceeded the factory-characterized value by more than 10% upperrange limit.

PASSED PARAMETER TOO LARGE • Parameter above operating limits.

PASSED PARAMETER TOO SMALL • Parameter below operating limits.

CONTROL LOOP SHOULD BE INMANUAL

• This message appears whenever the possibility exists that the operation will affect the 4-20 mA outputsignal.

CONTROL LOOP MAY BE RETURNEDTO AUTO • After the operation is completed, you are reminded to return the loop to automatic control.

Table 4.1 - Diagnostic Error and Potential Source

TROUBLESHOOTING

Symptom: NO LINE CURRENT

Probable Source of Trouble:

§§ Transmitter Connections

• Check wiring polarity and continuity.• Check for shorts or ground loops.• Check if the power supply connector is connected to main

board.

§§ Power Supply

• Check power supply output. The voltage must be between 12and 45 Vdc at transmitter terminals.

§§ Electronic Circuit Failure

• Check the main board for defect by replacing it with a spareone.

Symptom: NO COMMUNICATION

Probable Source of Trouble:

§§ Terminal Connections

• Check terminal interface connections.• Check if the interface is connected to the wires leading to the

transmitter or to the terminals [COMM] and [-].• Check if the interface is model IF3 (for Hart Protocol).

Transmitter Connections

• Check if connections are according to wiring diagram.

• Check line resistance; it must be equal to or greater than 250Ohm between the transmitter and the power supply.

§§ Power Supply

• Check output of power supply. The voltage at the DT301terminals must be between 12 and 45 V, and ripple less than500 mV.

§ Electronic Circuit Failure

• Locate the failure by alternately replacing the transmitter circuitand the interface with spare parts.

§§ Transmitter Address

• In On Line Multidrop item check if the address is "0."

Symptom: CURRENT OF 21.0 mA or 3.9 mA

Probable Source of Trouble:

§§ Pressure Tap (Piping)

• Verify if blocking valves are fully open.• Check for gas in liquid lines or for liquid in dry lines.• Check the specific gravity of process fluid.• Check process flanges for sediments.• Check the pressure connection.• Check if bypass valves are closed.• Check if pressure applied is not over upper limit of transmitter's

range.

Page 30: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Maintenance 4.3

Sensor to Main Circuit Connection

• Check connection (male and female connectors).

§§ Electronic Circuit Failure

• Check the sensor circuit for damage by replacing it with a spareone.

• Replace sensor.

Symptom: INCORRECT OUTPUT

Probable Source of Trouble:

§§ Transmitter Connections

• Check power supply voltage.• Check for intermittent short circuits, open circuits and

grounding problems.

§§ Noise Measurement Fluid

• Adjust damping

§§ Pressure Tap

• Check the integrity of the circuit by replacing it with a spareone.

§§ Calibration

• Check calibration of the transmitter.

NOTE:

A 21.0 or 3.9 mA current indicates that the transmitter is in safetyoutput or fail-safe output mode. Use the HT2 to investigate the sourceof the problem.

DISASSEMBLY PROCEDURE

WARNING:

Do not disassemble with power on.

Figure 4.3 and 4.4 show transmitter's exploded view and will helpyou to understand the text below.

The numbers between parenthesis are relating to theenumeration of the items of the related drawing.

GROUP OF THE PROBE (16A, 16B, 17A or 17B)

To have access to the probe for cleaning, it is necessary toremove it from the process.

Remove the transmitter loosening the against-flange.

Cleaning should be done carefully in order to avoid damaging ofthe delicate isolating diaphragms. Use of a soft cloth and anonacid solution is recommended.

To remove the sensor from the electronic housing, the electricalconnections (in the field terminal side) and the main boardconnector must be disconnected.

Loosen the hex screw (6) and carefully unscrew the electronichousing from the sensor, observing that the flat cable is notexcessively twisted.

IMPORTANT:

The transmitters have a stopper that can be released to allow thesensor to rotate more than one turn. See Figure 5.1.

CAUTION:

Do not rotate the electronic housing more than 180º withoutdisconnecting the electronic circuit from the sensor and from the powersupply.

Fig. 4.1 – Sensor Rotation Stopper

ELECTRONIC CIRCUITTo remove the circuit board (5), loosen the two screw (3) thatanchor the board.

WARNING:

The board has CMOS components, which may be damaged byelectrostatic discharges. Observe correct procedures for handlingCMOS components. It is also recommended to store the circuit boardsin electrostatic-proof cases.

Pull the main board out of the housing and disconnect the powersupply and the sensor connectors.

REASSEMBLY PROCEDURE

WARNING:

Do not assemble with power on.

GROUP OF THE PROBE (16A, 16B, 17A or 17B)The bolts, nuts, flanges and other parts should be inspected forcorrosion or other eventual damage. Damaged parts should bereplaced.

The fitting of the sensor must be done with the main board out ofthe electronic housing. Mount the sensor to the housing turningclockwise until it stops. Then turn it counterclockwise until it facesthe protective cover (1) parallel to the process flange. Tighten thehex screw (6) to lock the housing to the sensor. Only after thatinstall main board.

ELECTRONIC CIRCUITPlug sensor connector and power supply connector to mainboard.

Attach the display to the main board. Observe the four possiblemounting positions (Figure 4.2). The SMAR mark indicates upposition.

Page 31: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation and Maintenance Instruction Manual – Preliminary4.4

Fig. 4.2 – Four Possible Positions of the Display

Anchor the main board and display with their screw (3).

After tightening the protective cover (1), mounting procedure iscomplete. The transmitter is ready to be energized and tested.

INTERCHANGEABILITY

In order to obtain an, accurate and better temperaturecompensated response, the data of the sensor should betransferred for EEPROM of the main board. That is doneautomatically when the transmitter is energized.

In this operation, then main circuit reads the number of series of

the sensor. If he differs of the number stored in the main board,the circuit will interpreted that there was change of the sensor andit will look for in the new sensor memory its characteristics:- Temperature compensation coefficients.- Sensor's TRIM data, including characterization curve;- Sensor characteristics: type, range, diaphragm material and fill

fluid.

This data must be transferred to the main circuit board.

The other information are stored in the main circuit memory andare not affected by sensor change.

Data transfer from the sensor to the main circuit can also beforced by function MAINT/BACKUP/READ FROM SENSOR.

In the case of change of the main board, the information of thesensor, as described above are up-to-date. Even so, theinformation of the transmitter as upper Value, lower Value,Damping and output unit must be reconfigured.

RETURNING MATERIALS

Should it become necessary to return the transmitter and/orHand-Held Terminal to SMAR, simply contact our office,informing the defective instrument's serial number, and return itto our factory.

In order to speed up analysis and solution of the problem, thedefective item should be returned with a description of the failureobserved, with as much details as possible. Other informationconcerning to the instrument operation, such as service andprocess conditions, is also helpful.

Page 32: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Maintenance 4.5

Fig. 4.3 – Exploded View Draw for DT301 - Sanitary Model

Page 33: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation and Maintenance Instruction Manual – Preliminary4.6

Fig. 4.4 – Exploded View Draw for DT301 - Industrial Model

Page 34: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Maintenance 4.7

ACCESSORIES

ORDERING CODE DESCRIPTION

HT2IF3

DP03DP43BE1

Hand-Held Terminal - 80 characters (PSION Organizer II - Model LZ64)HART Protocol InterfaceGeneral Purpose 32K RAM DatapackDT301 - Version 1.xx, English Language DatapackBattery Eliminator - 9V (110/220 VAC)

SPARE PARTS LIST FOR TRANSMITTER

DESCRIPTION OF PARTS POSITION CODE CATEGORY(NOTE 1)

HOUSING, Aluminum (NOTE 2) . 1/2 - 14 NPT . M20 x 1.5 . PG 13.5 DINHOUSING, 316 Stainless Steel (NOTE 2) . 1/2 - 14 NPT . M20 x 1.5 . PG 13.5 DINCOVER (Includes O-ring) . Aluminum . 316 SSCOVER WITH WINDOW FOR INDICATOR (Includes O-ring) . Aluminum . 316 SSCOVER LOCKING SCREWSENSOR LOCKING SCREWEXTERNAL GROUND SCREWIDENTIFICATION PLATE FIXING SCREW

888

888

1 e 121 e 12

1176139

400-0246400-0247400-0248

400-0249400-0250400-0251

204-0102204-0105

204-0103204-0106204-0120204-0121204-0124204-0116

DIGITAL INDICATOR 4 214-0108TERMINAL HOLDER INSULATOR 10 400-0058MAIN ELECTRONIC CIRCUIT BOARD (NOTE 3) 5 400-0235 AO_RING (NOTE 4) . Cover, Buna-N . Neck, Buna-N . Process Connection, Buna-N

21518

204-0122204-0113400-0236

BBB

TERMINAL HOLDING SCREW . HOUSING, Aluminum . HOUSING, 316 SSMAIN BOARD SCREW FOR HOUSING, Aluminum . Units With indicator . Units Without indicatorMAIN BOARD SCREW FOR HOUSING, 316 SS . Units With indicator . Units Without indicator

1111

33

33

304-0119204-0119

304-0118304-0117

204-0118204-0117

PROCESS CONNECTION - INDUSTRIAL MODEL . Flange 4” – 150# ANSI B-16.5, AISI 316 . Flange 4” – 300# ANSI B-16.5, AISI 316 . Flange 4” – 600# ANSI B-16.5, AISI 316 . Flange DN 100, PN 25 / 40, din 2526 – Form D, AISI 316PROCESS CONNECTION - SANITARY MODEL . Tank Adapter, AISI316 . Tri-Clamp de 4”, AISI304

14141414

1920

400-0237400-0238400-0239400-0240

400-0241400-0242

. Industrial Probe . Sanitary Probe

16A or 16B17 A or 17B

NOTE 5NOTE 5

BB

Note:1) For category A, it is recommended to keep, in stock, 25 parts installed for each set, and for category B, 50.2) Includes Terminal Block, Screws, caps and Identification plate without certification.3) The main board of DT301 and probe are items.4) O_rings are packaged with 12 units5) To specify sensors, use the following tables.

Page 35: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation and Maintenance Instruction Manual – Preliminary4.8

400-0244 Sanitary Model Probe

COD. RANGE

123

0,5 a 1,25 g/cm3

1,0 a 2,5 g/cm3

2,0 a 5,0 g/cm3

COD. Diaphragm MaterialHIZ

Hastelloy C276316L SSTOthers – Specify

COD. Fill FluidASDGTZ

DC 200/350 - Silicone OilDC 200/20 - Silicone OilDC 704 - Silicone OilGlycerin and waterSyltherm 800Others – Specify

COD. Mounting

12

TopSide

400-0244 - 1 H - D 1

400-0243 Industrial Model Probe

COD. RANGE

123

0,5 a 1,25 g/cm3

1,0 a 2,5 g/cm3

2,0 a 5,0 g/cm3

COD. Diaphragm MaterialHIZ

Hastelloy C276316L SSTOthers – Specify

COD. Fill FluidASDGTZ

DC 200/350 - Silicone OilDC 200/20 - Silicone OilDC 704 - Silicone OilGlycerin and waterSyltherm 800Others – Specify

COD. Mounting12

TopSide

400-0243 - 1 H - D 1

Page 36: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Technical Characteristics 5.1

5- TECHNICAL CHARACTERISTICS

Filling FluidsThe filling fluid selection shall take into account its physicalproperties in what concerns to pressure temperature limitsand chemical compatibility with the process fluid. The latter isan important consideration in case the filling fluid happens tocome in contact with the process fluid, should a leakageoccur.

Table 5.1 presents the filling fluids, which are available forthe DT301, together with some physical properties andapplications.

FILLINGFLUID

VISCOSITY(cSt) at 25ºC

DENSITY(g/cm³) at 25ºC

THERMAL EXPANSIONCOEFFICIENT

(1/ºC)APPLICATIONS

Silicone

DC 200 / 2020 0.95 0.00107 General purpose – Standard

Silicone

DC 70439 1.07 0.000799

General purpose ( high temperatureand vacuum )

Fluorolube 50 1.91 0.000874 Oxygen, Chlorine and Nitric Acid

Silicone

DC 200 / 350350 0.97 0.00096 Food and beverage industries

Syltherm 800 10 0.934 0.0009General purpose ( extremetemperatures, positive and negative )

Propylene Glycol

(Neobee M20)9.8 0.90 0.001

Food and beverage, andPharmaceutical areas

Glycerin and Water 12.5 1.13 0.00034 Food industries

Table 5.1 – Properties of filling fluids

Functional Specifications

Output SignalTwo-wire, 4-20 mA with superimposed digital communication.(HART Protocol).

Power Supply12 to 45 Vdc.

Load Limitation

IndicatorOptional 4½ - digit numerical and 5-character alphanumericalLCD indicator.

Hazardous Area CertificationsExplosion proof. Weather proof and intrinsically safe(CENELEC and FM standards) (Pending).

Zero and Span AdjustmentNon-interactive, via digital communication.

Temperature LimitsAmbient: -40 to 85º C (-40 to 185º F)Process: 0 to 120º C (32 to 248º F)Storage: -40 to 100º C (-40 to 212º F)Digital Display: -10 to 60º C (14 to 140º F).

Failure AlarmIn case of sensor or circuit failure, the self-diagnostics drivesthe output to 3.9 or 21.0 mA, according to the user’s choice.

Turn-on TimePerforms within specifications in less than 5.0 seconds afterpower is applied to the transmitter.

Volumetric DisplacementLess than 0.15 cm3 (0.01 in3).

Static Pressure Limit1.7Mpa (17kgf/cm2)

Humidity Limits0 to 100% RH

Damping Adjustment0 to 32 seconds in addition to intrinsic sensor response time(0.2 s) (Via Digital Communication).

ConfigurationTrough digital communication using HART Protocol.

Page 37: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation & Maintenance Instruction Manual - Preliminary5.2

Hand Held Terminal Main Features HT2An interface and a program datapack for DT301 arenecessary for communication.RAM memory: 32 Kbytes.EPROM memory 128 Kbytes, datapack.Display: 80 characters, 4 lines.Power supply: 9 Vdc.Dimensions (LWD): (142 x 78 x 29.3 mm)

Performance Specifications

Reference conditions: temperature 25º C (77ºF), atmosphericpressure, power supply of 24 Vdc, silicone oil fill fluid,isolating diaphragms in 316 L SS and digital trim equal tolower and upper range values

RA

NG

E

ACCURACY(1) ENVIROMENT

STABILITY(For 3

Months)

ZERO (2)STATIC

PRESSURE(per 1 kgf/cm2)

1 ±0.0004 g/cm3

(±0.1 0Bx)0.003 kg/m3 0.021 kg/m3 0.001 kg/m3

2 ±0.0007 g/cm3 0.013 kg/m3 0.083 kg/m3 0.004 kg/m3

3 ±0.0016 g/cm3 0.041 kg/m3 0.521 kg/m3 0.007 kg/m3

(1) Linearity, hysteresis. And repeatability effects are inclued.(2) This is systematic error that can be eliminated by calibrating at the Operating static pressure.

Power Supply Effect±0.005% of calibrated span per volt.

Mounting Position EffectIt can be eliminated after installation. No span effect.

Electro-Magnetic Interference EffectDesigned to comply with IEC 801.

Physical Specifications

Electrical Connection1/2" - 14 NPT, Pg 13.5 or M20 x 1.5 metric.

Process Connection§ Industrial Model: 316 SST Flange φ 4“x150# / 300# /600#

ANSI B16.5, DN100 PN 25/40 DIN 2526 Form D.§ Sanitary Model: 304 SST Tri-clamp 4 “x300#

Wetted Parts§ Isolating Diaphragms

316L SST or Hastelloy C276.§ Probe Material 316L SST

§ Wetted O-Rings (For Sanitary Model) Buna N, VitonTM or PTFE.

Nonwetted Parts§ Electronic Housing Injected aluminum with polyester painting or 316 SST (NEMA 4X, IP67).§ Fill Fluid Silicone (DC200/350, DC200/20, DC704 Syltherm 800) or Glycerin and Water§ Cover Rings Buna N§ Identification Plate 316 SST

MountingSide or top mounting.

Approximate Weights9 kg (20 Ib) – Sanitary Model.13 kg (28 Ib) - Industrial Model.

Page 38: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

Technical Characteristics 5.3

MODELDT301I

INDUSTRIAL CONCENTRATION / DENSITY TRANSMITTER

CODE Range

1

2

3

0,5 a 1,25 g/cm3

1,0 a 2,5 g/cm3

2,0 a 5,0 g/cm3

CODE Diaphragm Material

H

I

Z

Hastelloy C276

316L SST

Others – Specify

CODE Fill Fluid

A

S

D

G

T

Z

DC 200/350

DC 200/20

DC 704

Glycerin and Water

Syltherm 800

Outers – Specify

CODE Indicator Local

0

1

Without Indicator

With Digital Indicator

CODE Electrical Connection

0

A

B

Z

1/2 - 14NPT

M20 X ‘1/2

Pg 13.5 DIN

Others – Specify

CODE Mounting

1

2

Top

Side

CODE Process Connection Size, Rating and Standard

5

5

5

9

Z

1

2

3

C

Z

4” 150# ANSI B – 16.5

4” 300# ANSI B – 16.5

4” 600# ANSI B – 16.5

DN 00 PN25/40 DIN 2526 – FORM D

Others – Specify

CODE Optional Item ( * )

H1

ZZ

316 SST Housing

Special Options – Specify

DT301I - 1 I A - 1 0 1 - 5 1 / *

(*) Leave it Blank for no optional items.

Page 39: smar - Bevingbeving.se/images/files/products/Smar_DT301_Manual.pdf · e-mail: smarinfo@smar.com ARGENTINA Smar Argentina Soldado de La Independencia, 1259 (1429) Capital Federal –

DT301 – Operation & Maintenance Instruction Manual - Preliminary5.4

MODELDT301S SANITARY CONCENTRATION / DENSITY TRANSMITTER

CODE Range

123

0,5 a 1,25 g/cm3

1,0 a 2,5 g/cm3

2,0 a 5,0 g/cm3

CODE Diaphragm MaterialHIZ

Hastelloy C276

316L SST

Others SpecifyCODE Fill Fluid

ASDGTZ

DC 200/350 DC 200/20

DC 704

Water and Glycerin Syltherm 800

Others – SpecifyCODE Local Indicator

01

Without Indicator

With Digital IndicatorCODE Electrical Connection

0ABZ

1/2 - 14NPT

M20 X ‘1/2

Pg 13.5 DIN

Others – SpecifyCODE Mounting

12

Top

SideCODE Process Connection

JZ

Tri-clamp – 4” 300# Others – SpecifyCODE Wetted O-Rings Material

BVTZ

Buna – N Viton

Teflon

Others – SpecifyCODE Tank Adapter

01

Without Tank Adapter (Supplied by Customer) With Tank AdapterCODE Tri – clamp

02

Without Tri-clamp

With Tank AdapterCOD. Optional Item ( * )H1ZZ

316 SST Housing

Special Options – Specify

DT301S - 1 I A - 1 0 1 - J B 0 0 / *

( * ) Leave it Blank for optional items.