Six-Sigma Spo2 Probes
Transcript of Six-Sigma Spo2 Probes
To Select Spo2 probe
Medical Equipment & Systems
Six-Sigma
To select Spo2 probe for use To select Spo2 probe for use in L&T Medical productsin L&T Medical products
LARSEN & TOUBRO 1/23
Green Belt: -Sandip Mohite
Team Members:
-Tejas Bengali
in L&T Medical productsin L&T Medical products
To Select Spo2 probe
Medical Equipment & Systems
Six-Sigma
Brief History
LARSEN & TOUBRO 2/23
To Select Spo2 probe
Medical Equipment & Systems
Six-Sigma
OBJECTIVE:To select Spo2 Probe
SCOPE:Selection from alternatives(No
design changes)
DMADV
LARSEN & TOUBRO 3/23
DELIVERABLES:
Improve product quality.
To Select Spo2 probe
Medical Equipment & Systems
Six-SigmaDMADV
Patient Monitor
Mother Board
LARSEN & TOUBRO 4/23
Key PadNavigation
KeysModules
Spo2-Module
Spo2 Sensor
To Select Spo2 probe
Medical Equipment & Systems
Six-Sigma
Reliability
Technical
Performance
Ease of SourcingAesthetics
Comfort of use
VOB/VOC Kano’s Model
DMADV
LARSEN & TOUBRO 5/23
Biocompatibility
Standard
patient safety standard
Cost
To Select Spo2 probe
Medical Equipment & Systems
Six-Sigma
PUGH MATRIX
DMADV
Criteria Weight Datum A B C DCompliance to Biocompatibility
Standard10 0
Cost 8 0
Reliability 5 0
Technical Performance 8 0
LARSEN & TOUBRO 6/23
*Present Probe – was used as Datum.
Technical Performance 8 0
Strategic Sourcing preference 4 0
Conformance to patient safety
standard10 0
Aesthetics 4 0
Comfort Level 5 0
Total Score 0
To Select Spo2 probe
Medical Equipment & Systems
Six-Sigma
Patient Safety
DMADV
LARSEN & TOUBRO 7/23
To Select Spo2 probe
Medical Equipment & Systems
Six-Sigma
Technical Performance
DMADV
LARSEN & TOUBRO 8/23
To Select Spo2 probe
Medical Equipment & Systems
Six-Sigma
Criteria Weight Datum A B C DCompliance to Biocompatibility
Standard10 0 -1 0 0 0
Cost 8 0 1 1 1 1
Reliability 5 0 0 0 0 1
Technical Performance 8 0 0 0 0 -1
DMADV
PUGH MATRIX
LARSEN & TOUBRO 9/23
Technical Performance 8 0 0 0 0 -1
Strategic Sourcing preference 4 0 1 1 -1 1
Conformance to patient safety
standard10 0 -1 0 0 -1
Aesthetics 4 0 -1 1 1 -1
Comfort Level 5 0 ? ? ? ?
Total Score 0 -12 16 8 -5
To Select Spo2 probe
Medical Equipment & Systems
Six-SigmaDMADV
LARSEN & TOUBRO 10/23
To Select Spo2 probe
Medical Equipment & Systems
Six-Sigma
DOE
Experiment was designed to remove all the biases viz.
1. Previous knowledge
2. Inclination
DMADV
LARSEN & TOUBRO 11/23
3. Effect of day
10 Persons and 6 probes selected for experiment.
1. Variety in terms of gender, age.
Rating from 1-10
-1- Least Comfortable
-10- Most Comfortable
To Select Spo2 probe
Medical Equipment & Systems
Six-SigmaDMADV
Probes of
different makes
labeled
LARSEN & TOUBRO 12/23
Black box
To avoid bias
To Select Spo2 probe
Medical Equipment & Systems
Six-SigmaDMADV
LARSEN & TOUBRO 13/23
To Select Spo2 probe
Medical Equipment & Systems
Six-SigmaDMADV
LARSEN & TOUBRO 14/23
To Select Spo2 probe
Medical Equipment & Systems
Six-Sigma
ANOVA
DMADV
LARSEN & TOUBRO 15/23
• P- Value Indicates that Person, Probe as well as
interaction (PersonXProbe ) are major contributors.
To Select Spo2 probe
Medical Equipment & Systems
Six-Sigma
Residual
Percent
210-1-2
99.9
99
90
50
10
1
0.1
Fitted Value
Residual
10.07.55.02.50.0
2
1
0
-1
-2
Normal Probability Plot of the Residuals Residuals Versus the Fitted Values
Histogram of the Residuals Residuals Versus the Order of the Data
Residual Plots for RATING
DMADV
LARSEN & TOUBRO 16/23
Residual
Frequency
2.251.500.750.00-0.75-1.50-2.25
40
30
20
10
0
Observation Order
Residual
1201101009080706050403020101
2
1
0
-1
-2
Histogram of the Residuals Residuals Versus the Order of the Data
1 and 3. Residuals follow normal distribution.
2. Residuals are symmetrically distributed.
4. Residuals vary randomly with observation order, hence no effect of order.
To Select Spo2 probe
Medical Equipment & Systems
Six-Sigma
ACTUAL RATING
9
8
7
PERSON PROBE
Main Effects Plot (data means) for ACTUAL RATING
DMADV
LARSEN & TOUBRO 17/23
Mean of A
10987654321
6
5
4
654321
Person is more significant than probe
To Select Spo2 probe
Medical Equipment & Systems
Six-Sigma
Mean
10
9
8
7
6
PERSON
3
4
5
6
7
8
9
10
1
2
Interaction Plot (data means) for ACTUAL RATING
DMADV
LARSEN & TOUBRO 18/23
PROBE
654321
5
4
3
2
Interactions suggest people rate same probe differently. People have their
own perception of comfort level.
To Select Spo2 probe
Medical Equipment & Systems
Six-Sigma
SN Ratio AnalysisSN ratios
17
16
Human Probe
Main Effects Plot (data means) for SN ratios
DMADV
LARSEN & TOUBRO 19/23
Mean of S
1
15
14
13
654321
Signal-to-noise: Larger is better
• Probes 2 and 6 show have high SN Ratio, which suggest
that the variance is minimum in their case.
To Select Spo2 probe
Medical Equipment & Systems
Six-Sigma
Mean Ratings
DMADV
Probe Name Probe No.
Comfort
Level
D 4 6.4
C 1 6.55
LARSEN & TOUBRO 20/23
C 1 6.55
E 3 6.7
A 5 6.7
B 6 7.2
Datum 2 7.5
To Select Spo2 probe
Medical Equipment & Systems
Six-SigmaDMADV
LARSEN & TOUBRO 21/23
B & C are same in terms of comfort level,
Since B has higher score in Pugh Matrix
B is selected
To Select Spo2 probe
Medical Equipment & Systems
Six-Sigma
Results
�Cost Avoidance of Rs. 2500 per unit.(Cumulative
till July-2010 is Rs 23L)
�Qualitative and quantitative Feedback from four
sales regions indicate good customer acceptance
DMADV
LARSEN & TOUBRO 22/23
sales regions indicate good customer acceptance
and 20% reduction in initial failures(1-3 months)
To Select Spo2 probe
Medical Equipment & Systems
Six-Sigma
Thank You
LARSEN & TOUBRO 23/23