SITRANS F flowmeters SITRANS F US SITRANS FUS380/FUE380/SONOCAL 3000 4/179 Siemens FI 01 · 2007 4 Overview SITRANS FUS380 The 2-track flowmeter SITRANS FUS380 comes as battery or mains powered and is designed to measure water flow in district heating plants, local networks, boiler stations, substations, chiller plants and other general water applications. The flowmeter is approved according to heat meter standards EN 1434 class 2 and OIML R75 class 2 and metrological para- meters are protected against manipulation. The type approved flowmeter is named SITRANS FUE380. Technically the meter types SITRANS FUS380 and SITRANS FUE380 are completely identical, only difference is the calibration limit. Benefits Battery powered up to 6 years Battery back-up option in case of mains power failure Fast measuring frequency 20 Hz/0.5 Hz (230 V AC/Battery) Easy one button straight forward display 2-track measuring principle for optimum accuracy Compact or remote mounting Measures on all district water qualities and water conductivi- ties No pressure drop Long-term stability Galvanic isolated digital output for easy connection to a calcu- lator (potential free) Dynamic range Q min :Q max up to 1:400 MODBUS RTU/RS 232, RS 485 Application The main application for SITRANS FUS380 is measurement of water flow or water flow in heat meter systems in district heating networks or chilled water. Combined with an energy calculator and a pair of temperature sensors, SITRANS FUE380 can be used as part of an energy meter system. For this purpose Siemens offers energy calculator SITRANS FUE950. Design The 2-track design of SITRANS FUS380 ensures maximum ac- curacy under short inlet conditions. The flowmeter consists of a flow sensor pipe, 4 transducers/transducer cables and a trans- mitter SITRANS FUS080. The unit is available in a compact or a remote version with up to 30 meter distance from flowmeter to transmitter. When ordering a compact version the transducer cables are pre-mounted and ready for installation. Compact mounting is only possible up to 120 °C (248 °F). The sensor must be isolated to protect transmitter from heat. The transmitter is available in an IP67/NEMA 4X/6 enclosure. Integration The flowmeter digital output is often used as input for an energy meter or as input for digital systems for remote reading. SITRANS FUS380 has two digital output functions that can be in- dividually selected, and optional MODBUS RTU communication modules. Pulse output rate is defined when ordering. If the flowmeter forms part of an energy meter system for cus- tody transfer, no further approvals are needed, except eventually local approvals on the flowmeter. Overview SONOCAL 3000 SONOCAL 3000 ultrasonic flowmeters are designed with blind display and 115 to 230 V power supply only and approved for custody transfer according to PTB, class C, OIML R75, class 4 and a number of local country approvals. Benefits No pressure drop Reliable and accurate flow measurements Long-term stability with excellent performance Can measure on all district water independent of water quality and conductivity Low cost of ownership Matched pair calibration of complete system Local country approvals Outputs: one analog, pulse and relay Design The SONOCAL 3000 ultrasonic flowmeters consist of a sensor type SONO 3300 CT, transmitter SONO 3000 CT without display incl. remote mounting device and 4 pcs 10 m coaxial cables for connecting the sensor and the transmitter. The sensor is a 2-track direct shot sensor type with flanges and integrated transducers. The transmitter is only available in the shown IP67 (NEMA 4X/6) version. All systems are sealed for custody transfer.

Transcript of SITRANS F flowmeters SITRANS F US - …SITRANS F flowmeters SITRANS F US SITRANS...

Page 1: SITRANS F flowmeters SITRANS F US - …SITRANS F flowmeters SITRANS F US SITRANS FUS380/FUE380/SONOCAL 3000 Siemens FI 01 · 2007 4/179 4 Overview SITRANS FUS380 The 2-track flowmeter

SITRANS F flowmetersSITRANS F US

SITRANS FUS380/FUE380/SONOCAL 3000

4/179Siemens FI 01 · 2007

4

Overview SITRANS FUS380

The 2-track flowmeter SITRANS FUS380 comes as battery or mains powered and is designed to measure water flow in district heating plants, local networks, boiler stations, substations, chiller plants and other general water applications. The flowmeter is approved according to heat meter standards EN 1434 class 2 and OIML R75 class 2 and metrological para-meters are protected against manipulation. The type approved flowmeter is named SITRANS FUE380.Technically the meter types SITRANS FUS380 and SITRANS FUE380 are completely identical, only difference is the calibration limit.

Benefits

• Battery powered up to 6 years• Battery back-up option in case of mains power failure • Fast measuring frequency 20 Hz/0.5 Hz (230 V AC/Battery)• Easy one button straight forward display • 2-track measuring principle for optimum accuracy• Compact or remote mounting• Measures on all district water qualities and water conductivi-

ties• No pressure drop• Long-term stability • Galvanic isolated digital output for easy connection to a calcu-

lator (potential free) • Dynamic range Qmin:Qmax up to 1:400• MODBUS RTU/RS 232, RS 485

Application

The main application for SITRANS FUS380 is measurement of water flow or water flow in heat meter systems in district heating networks or chilled water.Combined with an energy calculator and a pair of temperature sensors, SITRANS FUE380 can be used as part of an energy meter system. For this purpose Siemens offers energy calculator SITRANS FUE950.

Design

The 2-track design of SITRANS FUS380 ensures maximum ac-curacy under short inlet conditions. The flowmeter consists of a

flow sensor pipe, 4 transducers/transducer cables and a trans-mitter SITRANS FUS080.

The unit is available in a compact or a remote version with up to 30 meter distance from flowmeter to transmitter. When ordering a compact version the transducer cables are pre-mounted and ready for installation.

Compact mounting is only possible up to 120 °C (248 °F). The sensor must be isolated to protect transmitter from heat. The transmitter is available in an IP67/NEMA 4X/6 enclosure.

Integration

The flowmeter digital output is often used as input for an energy meter or as input for digital systems for remote reading.SITRANS FUS380 has two digital output functions that can be in-dividually selected, and optional MODBUS RTU communication modules.Pulse output rate is defined when ordering.If the flowmeter forms part of an energy meter system for cus-tody transfer, no further approvals are needed, except eventually local approvals on the flowmeter.

Overview SONOCAL 3000

SONOCAL 3000 ultrasonic flowmeters are designed with blind display and 115 to 230 V power supply only and approved for custody transfer according to PTB, class C, OIML R75, class 4 and a number of local country approvals.

Benefits

• No pressure drop• Reliable and accurate flow measurements• Long-term stability with excellent performance • Can measure on all district water independent of water quality

and conductivity• Low cost of ownership• Matched pair calibration of complete system• Local country approvals• Outputs: one analog, pulse and relay

Design

The SONOCAL 3000 ultrasonic flowmeters consist of a sensor type SONO 3300 CT, transmitter SONO 3000 CT without display incl. remote mounting device and 4 pcs 10 m coaxial cables for connecting the sensor and the transmitter.The sensor is a 2-track direct shot sensor type with flanges and integrated transducers.The transmitter is only available in the shown IP67 (NEMA 4X/6) version.All systems are sealed for custody transfer.

Page 2: SITRANS F flowmeters SITRANS F US - …SITRANS F flowmeters SITRANS F US SITRANS FUS380/FUE380/SONOCAL 3000 Siemens FI 01 · 2007 4/179 4 Overview SITRANS FUS380 The 2-track flowmeter

SITRANS F flowmetersSITRANS F US

SITRANS FUS380/FUE380/SONOCAL 3000

4/180 Siemens FI 01 · 2007

4

Configuration SITRANS FUS380

Selection guide SITRANS FUS380, standard version

Selection guide SITRANS FUE380, type approved version

According to EN 1434, class 2, flowmeter values

According to OIML R75, class 2, flowmeter values

Dynamic range qi:qp: better than 1:100 according to EN 1434 and OIML R75 class 2Low flow cut off: 0.2% of qp (nominal)In order to obtain best pulse output resolution in the range Qmin - Qmax of approx. 100 Hz at qs, two or three values for every dimension can be selected at ordering.1) Other flow ranges - see MLFB ordering table2) In connection with SITRANS FUE950 - other pulse values - see MFLB ordering table

Technical specifications SITRANS FUS380

Flowmeter size nominal to EN 1092-1 DN 100 DN 125 DN 150 DN 200 DN 250 DN 300 DN 350 DN 400 DN 500 DN 600 DN 700 DN 800

Flow range1) Qmax (qs) m3/h 240 400 560 900 1400 2100 2800 3600 5500 8000 10800 14200

Qmin (qi) m3/h 0.6 1.0 1.5 2.5 4.0 5.6 7.0 9.5 14.75 21.50 29.0 38.0

Dyn. range qi:qs 1:400 1:400 1:373 1:360 1:350 1:375 1:400 1:379 1:373 1:372 1:372 1:373

Flowmeter size nominal to EN 1092-1 DN 100 DN 125 DN 150 DN 200 DN 250 DN 300 DN 350 DN 400 DN 500 DN 600 DN 700 DN 800

Flow range1) Qmax (qs) m3/h 120 or 180

200 or 280

300 or 420

500 or 700

800 or 1120

1120 or 1560

1500 or 2100

1900 or 2660

2950 or 4130

4300 or 6020

5800 or 8120

7600 or 10640

Qnom (qp) m3/h 60 100 150 250 400 560 750 950 1475 2150 2900 3800

Qmin (qi) m3/h 0.6 1.0 1.5 2.5 4.0 5.6 7.5 9.5 14.75 21.5 29.0 38.0

Pulse value 2) l/pulse 2.5 2.5 10 10 10 50 50 50 100 100 100 100

Flowmeter size nominal to EN 1092-1 DN 100 DN 125 DN 150 DN 200 DN 250 DN 300 DN 350 DN 400 DN 500 DN 600 DN 700 DN 800

Flow range1) Qmax (qs) m3/h 180 280 420 700 1120 1560 2100 2660 4160 6020 8120 10640

Qnom (qp) m3/h 120 200 300 500 800 1120 1500 1900 2950 4300 5800 7600

Qmin (qi) m3/h 1.2 2.0 3.0 5.0 8.0 11.2 15.0 19.5 29.5 43.0 58.0 76.0

Pulse value 2) l/pulse 2.5 2.5 10 10 10 50 50 50 100 100 100 100

Pipe design 2-track sensor with flanges and integrated transducers wet cali-brated from factory

Nominal size welded version DN 100, 125, 150, 200, 250, 300, 350, 400, 500, 600, 700, 800

Pressure rate PN 16, PN 25, PN 40 EN 1092-1

Pipe material Carbon Steel EN 1.0345 / p235 GH

Transducer design Integrated version welded onto the pipe

Transducer material Stainless steel AISI 316

Sensor operating conditions

Storage -40 … +85 °C (-40 … +185 °F)

Liquid temperature • Remote: 2 … 200 °C (35.6 … 392 °F)

• Compact: 2 … 120 °C (35.6 … 248 °F)

Degree of protection Sensor connection IP67/NEMA 4X/6

Max. flow velocity DN 100 … 800: • FUS380: 8 m/s (26.2 ft/s) • FUE380: 6 m/s (19.7 ft/s)

Transmitter

Display LCD, 8 digits, additional 2 digits and symbols for status informa-tion

Push button One push button for display infor-mation

Communication IrDA – optical communication interface with MODBUS RTU pro-tocol

Add-on modules:• RS 232 serial interface with

MODBUS RTU (Rx/Tx/GND), point to point with max. 15 m ca-ble

• RS 485 serial interface with MODBUS RTU (+/-/GND), multi-drop with up to 32 devices with max. 1000 m cable

MODBUS RTU protocol is an open protocol (further informa-tion available on request)Serial speed 1200, 2400, 4800, 9600, 19200, 38400 Baud

Page 3: SITRANS F flowmeters SITRANS F US - …SITRANS F flowmeters SITRANS F US SITRANS FUS380/FUE380/SONOCAL 3000 Siemens FI 01 · 2007 4/179 4 Overview SITRANS FUS380 The 2-track flowmeter

SITRANS F flowmetersSITRANS F US

SITRANS FUS380/FUE380/SONOCAL 3000

4/181Siemens FI 01 · 2007

4

Type dependent settings

SITRANS FUE380 uncertainty

To ensure continuous accurate measurement, flowmeters must be calibrated. The calibration is conducted at SIEMENS flow fa-cilities accredited according to ISO/IEC 17025 by DANAK or UKAS.

The accreditation bodies DANAK and UKAS have signed the ILAC MRA agreement (International Laboratory Accreditation Corporation - Mutual Recognition Arrangement). Therefore the accreditation ensures international traceability and recognition of the test results in 39 countries world wide, including the US (NIST traceability).

A calibration certificate is shipped with every SITRANS FUS380/ FUE380.

Accuracy SITRANS FUS380:

Standard calibration: Better than 0.5% of rate, 0.5 m/s < v < 8 m/sv < 0.5 m/s, 0.5 + 0.25/v [%]

Typical accuracy SITRANS FUE380:

0.5 + 0.02 qp/q [%]qp according to EN 1434/OIML requirements.

Example: DN 100, qp = 60 m3/h at q = 1.2 m3/h:Accuracy = typical 1.5 %

SITRANS FUE380 fulfils the requirements Ef = ±(2 + 0.02 qp/qi) max. ± 5%, according to EN 1434 and OIML R75, class 2 revised 1. July 2002

Enclosure IP67/NEMA 4X/6 to EN 60529 and DIN 40050

Temperature ambient 0 … 60 °C (32 … 140 °F)

Temperature storage -40 … +85 °C (-40 … +185 °F) (battery included)

Installation Compact on sensor: max. 120 °C (248 °F),Separate: max. 30 m (98.4 ft) from transmitter

Mechanical vibration 2 g, 1 … 800 Hz sinusoidal in all directions to IEC 68-2-6

Design Fibre-glass reinforced polyamide

Power supply • Battery: replaceable 3.6 V LiSOCl (Lithium Thionyl Chlo-ride) battery pack 32 Ah

• Mains: 87 … 265 V AC (50 … 60 Hz)

Measuring rate Battery mode: 0.5 HzMains supply: 20 Hz Back-up mode: 0.5 Hz (at mains supply drop)

Digital output Two passive individual galvani-cally isolated MOS relay outputs, A and B, max. ±35 V AC/DC, 50 mA

Max pulse frequency 100 Hz

Alarm indication Track 1 (F1), track 2 (F2), Low battery indication (F5), qs over-flow (F6), pulse overflow (F7)

Cable length Max. 30 m (98.4 ft) between transmitter and sensor

EMC • Emission EN 61000-6-4 • Immunity EN 61000-6-2

Approvals EN 1434 and OIML R75 Class 2(EN version from 1. july 2002)

FUS380 FUE380

Flow value Predefined settings according to dimension

Predefined according to EN 1434 / OIML R75

Approval No approval Country specific

Flow rate vf 0.02 … 8 m/s (0.065 … 26.2 ft/s)

0.02 … 6 m/s (0.065 … 19.7 ft/s)

Output A Forward / reverse Preset: Forward

Output B Preset: Alarm Preset: Alarm

Output B, function Reverse pulse, alarm, call-up

Preset: Alarm

Pulse value A & B(depending on DN value)

0.5 l/p 1 l/p 2.5 l/p 10 l/p 50 l/p 100 l/p 250 l/p 500 l/p 1 m3/p 2.5 m3/p 5 m3/p 10 m3/p 25 m3/p 50 m3/p 100 m3/p 250 m3/p 500 m3/p 1000 m3/p

Preset: See scheme - previous pagePreset for SITRANS FUE950 or free select-able

Pulse width 5/10/20/50/100/200/ 500 ms

Preset: 5 ms

Flow unit setup Preset: m3/h Preset: m3/h

Volume unit setup Preset: m3 Preset: m3

Page 4: SITRANS F flowmeters SITRANS F US - …SITRANS F flowmeters SITRANS F US SITRANS FUS380/FUE380/SONOCAL 3000 Siemens FI 01 · 2007 4/179 4 Overview SITRANS FUS380 The 2-track flowmeter

SITRANS F flowmetersSITRANS F US

SITRANS FUS380/FUE380/SONOCAL 3000

4/182 Siemens FI 01 · 2007

4

Output configuration SITRANS FUS380/FUE380

Pulse volume: output A/B configured as volume per pulse, cal-culated on forward/reverse or net forward/reverse flow. The volume per pulse is free scaleable (FUS380 only).

Pulse output B can be used as stated above or as alarm or call up function

Call up: the call up output is active until manually reset by use of PDM program. The callup function is activated when an alarm is activated.

Technical specifications SONOCAL 3000

SONO 3300 CT flow sensor

Coaxial cable

Accuracy

Error in measurement at reference conditions; % of measured value

Dynamic range

• 1:20 Pulse output ≤ ± 0.5%

• 1:50 Pulse output ≤ ± 3%

Repeatability ≤ ± 0.25%

Maximum flow velocity 10 m/s (32 ft/s)

Nominal size DN 50, 65, 80, 900, 1000, 1200 (2“, 2,5", 3", 36", 40", 48")

Liquid temperature -10 … +200 °C (14 … 392 °F) depending on approval

Ambient temperature -10 … +160 °C (14 … 320 °F) depending on approval

Enclosure IP67 (NEMA 4X/6)

Process connections

PN designated,EN 1092-1, type 11, B

• PN 16 (DN 50, 65, 80, 900, 1000, 1200 (2”, 2,5", 3", 36", 40", 48"))

• PN 25 (DN 900, DN 1000 (36”, 40”))

• PN 40 (DN 50, 65, 80 (2” ... 3”))

Transducers Integrated version welded into the pipe

Materials

Pipe • DN 50 ... DN 80 (2” ... 3”):Steel GS-16 Mn5, mat. No. 1.1131

• DN 900 ... DN 1200 (36” ... 48”): Steel EN 1.0345

Flange • DN 50, 65, 80, 900, 1000, 1200 (2”, 2,5", 3", 36", 40", 48"):Steel EN 1.0025-S235JRG2

Tranducers Stainless steel AISI 316 or similar

Certificates and approvals

Material certificates The sensor is supplied with a Siemens Certificate of Conformity

Custody transfer PTB class C, OIML R75 class 4, many other local approvals

The sensors are approved according to EU Directive 97/23/EC dated 29 May 1997 regarding fluid in group 1, classified in category III. Design EN 13480.

The first 0.5 m (19.7 inch) of the coaxial cable

Diameter 5.3 mm (0.21”)

Length 0.5 m (19.7”)

Material PTFE

Ambient temperature -200 … +200 °C (-328 ... +392 F)

Coaxial cable > 0.5 m (19.7 inch)

Diameter 5.8 mm (0.23”)

Length 9.5 m (30.4 ft)Total max. length 10 m (32 ft)

Material PE

Ambient temperature -40 … +70 °C (-40 ... +158 °F)

Page 5: SITRANS F flowmeters SITRANS F US - …SITRANS F flowmeters SITRANS F US SITRANS FUS380/FUE380/SONOCAL 3000 Siemens FI 01 · 2007 4/179 4 Overview SITRANS FUS380 The 2-track flowmeter

SITRANS F flowmetersSITRANS F US

SITRANS FUS380/FUE380/SONOCAL 3000

4/183Siemens FI 01 · 2007

4

SONO 3000 CT transmitter without display

Selection guide SITRANS F US SONOCAL series 3000 flow part (DN 50, 65, 80, 900, 1000, 1200) with standard settings

Output

Analog output Individually galvanically isolated, isolation voltage 500 V

• Measurement of Volume flow

• Current 4 ... 20 mA

• Load < 800 Ω

• Time constant 5 s

Pulse output Individually galvanically isolated, isolation voltage 500 V

• Measurement of Volume flow

• Pulse width 5 ms

Pulse values 1-2.5-100 l/pulse depending on size, see selection guide

Output mode

• Passive 5 ... 30 V/max. 200 mA

Relay output

• Time constant/hysteresis 5 s/0.5% F.S.O.

• Change-over relay Error indication

• Load 42 V, 0.5 A

Cut-off: low flow 0.8% F.S.O.

Rated operation conditions

Ambient temperature

• IP67 (NEMA 4X/6) version

- Storage -40 ... +85 °C (-40 ... +185 °F)

Mechanical vibrations 2 g, 1 .. 800 Hz sinusoidal in all directions to IEC 68-2-6

Degree of protection (enclosure) IP67 (NEMA 4X/6) to EN 60529 and DIN 40050

Design

Enclosure material Fibre glass reinforced polyamide

Dimensional drawings and weight See dimensional drawings

Power supply

Supply voltage and power con-sumption

115 … 230 V +10%/-15%, 50 ... 60 Hz, 10 ... 20 VA

Certificates and approvals

EMC Emission EN 61000-6-4Immunity EN 61000-6-2

Low voltage Acoording to EN 61010-1

Approvals

IP67 (NEMA 4X/6) version PTB, class C, OIML R75, class 4

Flowmeter size nominal to EN 1092-1 DN 50 DN 65 DN 80 DN 900 DN 1000 DN 1200

Flow range Qmax (Qs) m3/h 45 72 120 6000 6000 7200

Qn (Qp) m3/h 36 60 100 5000 5000 6000

Qmin m3/h 0.31 0.44 0.7 112 112 162

Flowrate at 20 mA Q20mA m3/h 36 60 100 5000 5000 6000

Pulse output l/pulse 1 1 2.5 100 100 100

Page 6: SITRANS F flowmeters SITRANS F US - …SITRANS F flowmeters SITRANS F US SITRANS FUS380/FUE380/SONOCAL 3000 Siemens FI 01 · 2007 4/179 4 Overview SITRANS FUS380 The 2-track flowmeter

SITRANS F flowmetersSITRANS F US

SITRANS FUS380/FUE380/SONOCAL 3000

4/184 Siemens FI 01 · 2007

4

Selection and Ordering data Order-No. Order code

Flowmeter SITRANS FUS380 (standard) 7 ME 3 4 0 0 -Flowmeter SITRANS FUE380 (type approved) 7 ME 3 4 1 0 -Flowmeter SITRANS F US SONOCAL 3000 7 ME 3 3 2 0 -

77777 - 7777 777

SITRANS FUS380 (standard) 7ME3400

Diameter Qp [m3/h] Qs [m3/h]

DN 100 / 4" 60 60 1 NDN 100 / 4" 60 180 1 QDN 100 / 4" 120 240 1 R

DN 125 / 5" 100 100 1 SDN 125 / 5" 100 280 1 UDN 125 / 5" 200 400 1 V

DN 150 / 6" 150 150 2 ADN 150 / 6" 150 420 2 CDN 150 / 6" 300 560 2 D

DN 200 / 8" 250 250 2 EDN 200 / 8" 250 700 2 GDN 200 / 8" 500 900 2 H

DN 250 / 10" 400 400 2 JDN 250 / 10" 400 1120 2 LDN 250 / 10" 800 1400 2 M

DN 300 / 12" 560 560 2 NDN 300 / 12" 560 1560 2 QDN 300 / 12" 1120 2100 2 R

DN 350 / 14" 750 750 2 SDN 350 / 14" 750 2100 2 UDN 350 / 14" 1500 2800 2 V

DN 400 / 16" 950 950 3 ADN 400 / 16" 950 2660 3 CDN 400 / 16" 1900 3600 3 D

DN 500 / 20" 1475 1475 3 JDN 500 / 20" 1475 4130 3 LDN 500 / 20" 2950 5500 3 M

DN 600 / 24" 2150 2150 3 SDN 600 / 24" 2150 6020 3 UDN 600 / 24" 4300 8000 3 V

DN 700 / 28" 2900 2900 4 EDN 700 / 28" 2900 8120 4 GDN 700 / 28" 5800 10800 4 H

DN 800 / 32" 3800 3800 4 NDN 800 / 32" 3800 10640 4 QDN 800 / 32" 7600 14200 4 R

SITRANS FUE380 (type approved) 7ME3410

Diameter Qp [m3/h] Qs [m3/h]

DN 100 / 4" 601) 120 1 PDN 100 / 4" 601) 180 1 QDN 100 / 4" 1202) 180 1 R

DN 125 / 5" 1001) 200 1 TDN 125 / 5" 1001) 280 1 UDN 125 / 5" 2002) 280 1 V

DN 150 / 6" 1501) 300 2 BDN 150 / 6" 1501) 420 2 CDN 150 / 6" 3002) 420 2 D

DN 200 / 8" 2501) 500 2 FDN 200 / 8" 2501) 700 2 GDN 200 / 8" 5002) 700 2 H

DN 250 / 10" 4001) 800 2 KDN 250 / 10" 4001) 1120 2 LDN 250 / 10" 8002) 1120 2 M

DN 300 / 12" 5601) 1120 2 PDN 300 / 12" 5601) 1560 2 QDN 300 / 12" 11202) 1560 2 R

Diameter Qp [m3/h] Qs [m3/h]

DN 350 / 14" 7501) 1500 2 TDN 350 / 14" 7501) 2100 2 UDN 350 / 14" 15002) 2100 2 V

DN 400 / 16" 9501) 1900 3 BDN 400 / 16" 9501) 2660 3 CDN 400 / 16" 19002) 2660 3 D

DN 500 / 20" 14751) 2950 3 KDN 500 / 20" 14751) 4130 3 LDN 500 / 20" 29502) 4130 3 M

DN 600 / 24" 21501) 4300 3 TDN 600 / 24" 21501) 6020 3 UDN 600 / 24" 43002) 6020 3 V

DN 700 / 28" 29001) 5800 4 FDN 700 / 28" 29001) 8120 4 GDN 700 / 28" 58002) 8120 4 H

DN 800 / 32" 38001) 7600 4 PDN 800 / 32" 38001) 10640 4 QDN 800 / 32" 76002) 10640 4 R

SITRANS F US SONOCAL 3000 7ME3320

Diameter Qp [m3/h] Qs [m3/h]

Cast steel, mat. no 1.1131/GS-16Mn5

DN 50 36 45 1 CDN 65 60 72 1 GDN 80 100 120 1 L

Carbon steel, mat. no 1.0345/P235GH

DN 900 5000 6000 5 CDN 1000 5000 6000 5 LDN 1200 6000 7200 5 U

Flange norm and pressure rating

No pipe (only FUS080) A

EN 1092-1PN 16 (DN 50 ... 1200) CPN 25 (DN 200 ... 1000)PN 40 (DN 50 ... 250)

DE

Compact / remote connection

Compact version, max. 120 °C (248 °F) FUS380/FUE380 only

0

Remote version, SONOCAL 3000 with 10 m (32.8 ft) cable

1

Remote version, max. 200 °C (392 °F)5 m (16.4 ft) 210 m (32.8 ft) 320 m (65.6 ft) 430 m (98.4 ft) 5

Selection and Ordering data Order-No. Order code

Flowmeter SITRANS FUS380 (standard) 7 ME 3 4 0 0 -Flowmeter SITRANS FUE380 (type approved) 7 ME 3 4 1 0 -Flowmeter SITRANS F US SONOCAL 3000 7 ME 3 3 2 0 -

77777 - 7777 777

Page 7: SITRANS F flowmeters SITRANS F US - …SITRANS F flowmeters SITRANS F US SITRANS FUS380/FUE380/SONOCAL 3000 Siemens FI 01 · 2007 4/179 4 Overview SITRANS FUS380 The 2-track flowmeter

SITRANS F flowmetersSITRANS F US

SITRANS FUS380/FUE380/SONOCAL 3000

4/185Siemens FI 01 · 2007

4

Approvals / pulse outputWithout approval (neutral) Selectable pulse output (SONOCAL 3000 standard)

0

Only FUE380 and SONOCAL 3000With approval marksSelectable pulse output (SONOCAL 3000 standard)

1

With approval marks and sealSelectable pulse output (SONOCAL 3000 standard)

2

Without approval (neutral) (FUE380 only)Preset pulse output for FUE950 energy meter

3

With approval marks (FUE380 only)Preset pulse output for FUE950 energy meter

4

With approval marks and seal (FUE380 only)Preset pulse output for FUE950 energy meter

5

Pulse value for SITRANS FUE9501 l/pulse (SONOCAL 3000 DN 50...DN 65) 22.5 l/pulse (SONOCAL 3000 DN 80)(DN 100, DN 125 SITRANS FUE380)

3

10 l/pulse(DN 150 ... DN 250 SITRANS FUE380)

4

50 l/pulse(DN 300 ... DN 400 SITRANS FUE380)

5

100 l/pulse (SONOCAL DN 900...1200)(DN 500 ... DN 800 SITRANS FUE380)

6

Pulse value FUS380/FUE380

250 l/pulse 71 m3/pulse 8

0.5 l/pulse 9 N 0 B5 l/pulse 9 N 0 C25 l/pulse 9 N 0 D

500 l/pulse 9 N 0 E2.5 m3/pulse 9 N 0 F5 m3/pulse 9 N 0 G

10 m3/pulse 9 N 0 H25 m3/pulse 9 N 0 J50 m3/pulse 9 N 0 K

100 m3/pulse 9 N 0 L250 m3/pulse 9 N 0 M500 m3/pulse 9 N 0 N

1000 m3/pulse 9 N 0 P

Transmitter Blind SONO 3000 CT

IP67/NEMA4X/6, 115 ... 230 V AC SONOCAL 3000

A

Transmitter SITRANS FUS080/FUE080

IP67/NEMA 4X/6 115 ... 230 V AC B

IP67/NEMA 4X/6 (3.6 V battery supply) D

IP67/NEMA 4X/6 115 ... 230 V AC, including 3.6 V battery back up

E

IP67/NEMA 4X/6 3.6 V battery version (no battery included)3)

G

Country / approval type 4)

No approval mark A

Only FUE380Denmark, EN 1434/OIML R75 ERussia, EN 1434/OIML R75 M

Selection and Ordering data Order-No. Order code

Flowmeter SITRANS FUS380 (standard) 7 ME 3 4 0 0 -Flowmeter SITRANS FUE380 (type approved) 7 ME 3 4 1 0 -Flowmeter SITRANS F US SONOCAL 3000 7 ME 3 3 2 0 -

77777 - 7777 777

Country/ Approval-type SONOCAL 3000 onlyAustria / OE01 C 050 BCroatia / HR Q-1-1002 C

Czech Republic / TCM 142/97-2741 DDenmark / TS 27.01 076 (OIML R75) EFinland F

Germany / 2216 9802 (PTB) GHungary /Th-8290/4/2004 HLatvia J

Poland / ZT 597/2003 KRomania / RO 176/98 LRussia / 16373/1 (OIML R75) M

Switzerland / ZW 138 NUkraine PYugoslavia / Q-2-17 Q

China Z Q 0 CSlovak Republic Z Q 0 L

Pulse width setup5 ms (always for SONOCAL 3000) 05 ms (standard for SITRANS FUE950) 210 ms 320 ms 4

50 ms 5100 ms 6200 ms 7500 ms 8

1) EN 1434 flow values2) OIML R75 flow values3) Lithium batteries are subject to special transportation regulations accord-

ing to United Nations "Regulation of Dangerous Goods, UN 3090 and UN 3091". Special transport documentation is required to observe these regulations. This may influence both transport time and costs.

4) Other countries in progress

Please also see www.siemens.com/SITRANSForderingfor practical examples of ordering.

Selection and Ordering data Order-No. Order code

Flowmeter SITRANS FUS380 (standard) 7 ME 3 4 0 0 -Flowmeter SITRANS FUE380 (type approved) 7 ME 3 4 1 0 -Flowmeter SITRANS F US SONOCAL 3000 7 ME 3 3 2 0 -

77777 - 7777 777

Page 8: SITRANS F flowmeters SITRANS F US - …SITRANS F flowmeters SITRANS F US SITRANS FUS380/FUE380/SONOCAL 3000 Siemens FI 01 · 2007 4/179 4 Overview SITRANS FUS380 The 2-track flowmeter

SITRANS F flowmetersSITRANS F US

SITRANS FUS380/FUE380/SONOCAL 3000

4/186 Siemens FI 01 · 2007

4

MLFB Ordering example

Customer requires a flowmeter for custody transfer:• DN 250, PN 25, compact version (media temperature max.

120 °C), battery version. • Type approved according to EN 1434, verified and sealed.• Pulse output for energymeter SITRANS FUE950.

Ordering:

FUE380: 7ME3410-2KD05-4DE2

Example of appropriate energy meter:

Energy meter type: FUE950-03110-0R1CB-10300-DK2-00012

Selection and Ordering data Order code

Additional informationPlease add „-Z“ to Order No. and specify Order code(s) and plain text.

Verification FUE380

Verification certificate 2 x 3 points. Max. flow 50 ... 250 m3/h depending on dimension

D10 1)

1) Only selectable in SITRANS FUE380

Verification certificate 2 x 3 points. Max flow 250 ... 1300 m3/h depending on dimension

D11 1)

Verification certificate 2 x 3 points. Max flow 1400 ... 4200 m3/h depending on dimension

D12 1)

Accredited Siemens calibration FUS380/FUE380

Accredited Siemens ISO/IEC 17025 calibration. Max. flow 50 ... 250 m3/h, depending on dimension

D20

Accredited Siemens ISO/IEC 17025 calibration Max. flow 250 ... 1300 m3/h depending on dimension

D21

Accredited Siemens ISO/IEC 17025 calibration Max. flow 1400 ... 4200 m3/h depending on dimension

D22

Verification SONOCAL 3000

Verification certificate: 2 x 3 points, DN 50 ... DN 80(Max. flow 36 ... 180 m3/h depending on dimension)

D10

Verification certificate: 2 x 3 points, DN 900 ... DN 1200(Max. flow 4300 m3/h)

D12

Accredited Siemens calibration SONOCAL 3000

Accredited Siemens ISO/IEC 17025 calibration DN 50 ... DN 80, (Max. flow 36 ... 180 m3/h dep. on dim.)

D20

Accredited Siemens ISO/IEC 17025 calibrationDN 900 ... DN 1200, (Max. flow 4300 m3/h)

D22

Accredited third party calibration SONOCAL 3000

Accredited third party calibration SONOCAL 3000Accredited - Third party ISO/IEC 17025 calibrationDN 50 ... DN 80, (Max. flow 36 ... 180 m3/h dep. on dim.)

D30

Accredited - Third party ISO/IEC 17025 calibrationDN 900 ... DN 1200, (Max. flow 7000 m3/h)

D32

Material certificate

EN 10204-3.1.B F10

Tag name plate

Stainless steel tag with 12 mm characters, max. 15 characters (add plain text)

Y17

Self-adhesive plastic tag with 8 mm characters, max. 15 characters (add plain text)

Y18

Page 9: SITRANS F flowmeters SITRANS F US - …SITRANS F flowmeters SITRANS F US SITRANS FUS380/FUE380/SONOCAL 3000 Siemens FI 01 · 2007 4/179 4 Overview SITRANS FUS380 The 2-track flowmeter

SITRANS F flowmetersSITRANS F US

SITRANS FUS380/FUE380/SONOCAL 3000

4/187Siemens FI 01 · 2007

4

Dimensional drawings

Transmitter IP67/NEMA 4X/6, wall mounting

Pipe Dimensions

Weight electronics 1.5 kg (3.3 lb)

Weight electronics 1.5 kg (3.3 lb) - Means not available

B-5$

-

.$

.-$.

-5.$

B-1$

5-B.$

B-

1$

-

$

1-$

-.B$

-.$

1-$

-$

Size PN 16 PN 25 PN 40

L Weight L Weight L Weight A1 D D1 Lift hug

DN mm kg mm kg mm kg mm mm mm

100 350+0-2 20 - - 350+0-3 16.5 361 220 180 No

125 350+0-2 23 - - 350+0-3 53 374 250 210 No

150 500+0-3 26 - - 500+0-3 32 388 285 240 Yes

200 500+0-3 38 500+0-3 47 500+0-3 55 414 340 295 Yes

250 600+0-3 60 600+0-3 76 600+0-3 91 440 405 355 Yes

300 500+0-3 66 500+0-3 81 - - 466 460 410 Yes

350 550+0-3 94 550+0-3 121 - - - - - -

400 550+0-3 124 550+0-3 153 - - 507 580 525 Yes

500 625+0-3 190 625+0-3 244 - - 558 715 650 Yes

600 750+0-3 303 750+0-3 365 - - 609 840 770 Yes

700 875+0-3 361 875+0-3 552 - - 660 910 840 Yes

800 1000+0-3 494 1000+0-3 770 - - 710 1025 950 Yes

Size PN 16 PN 25 PN 40

L Weight L Weight L Weight A1 D D1 Lift hug

inch inch lb inch lb inch lb inch inch inch

4 13.77+0-0.08 44 - - 13.77+0-0.12 36 14.21 8.66 7.09 No

5 13.77+0-0.08 50 - - 13.77+0-0.12 117 14.72 9.84 8.27 No

6 19.68+0-0.12 57 - - 19.68+0-0.12 71 15.28 11.22 9.45 Yes

8 19.68+0-0.12 84 19.68+0-0.12 104 19.68+0-0.12 121 16.30 13.39 11.61 Yes

10 23.62+0-0.12 132 23.62+0-0.12 168 23.62+0-0.12 201 17.32 15.94 13.98 Yes

12 19.68+0-0.12 146 19.68+0-0.12 179 - - 18.35 18.11 16.14 Yes

14 21.65+0-0.12 207 21.65+0-0.12 267 - - - - - -

16 21.65+0-0.12 273 21.65+0-0.12 337 - - 19.96 22.83 20.67 Yes

20 24.61+0-3 419 24.61+0-3 538 - - 21.97 28.15 25.59 Yes

24 29.53+0-0.12 668 29.53+0-0.12 805 - - 23.98 33.07 30.31 Yes

28 34.45+0-0.12 796 34.45+0-0.12 1217 - - 25.98 35.83 33.07 Yes

32 39.37+0-0.12 1089 39.37+0-0.12 1698 - - 27.95 40.35 37.40 Yes

Page 10: SITRANS F flowmeters SITRANS F US - …SITRANS F flowmeters SITRANS F US SITRANS FUS380/FUE380/SONOCAL 3000 Siemens FI 01 · 2007 4/179 4 Overview SITRANS FUS380 The 2-track flowmeter

SITRANS F flowmetersSITRANS F US

SITRANS FUS380/FUE380/SONOCAL 3000

4/188 Siemens FI 01 · 2007

4

SONO 3300 CT sensor flow part

1) The stated wall thickness for DN 200 ... DN 1200 are minimum values according to the EC Directive on the Pressure Equipment 97/23/EC.

Weight of SONO 3300 CT sensor

SONO 3000 CT transmitter, IP67 (NEMA 4X/6), wall mounting

Add approximately 30 mm (1.18 inch) on each side for cabling, approx. 2 kg (4.4 lb).

Size Build-in length L between the flanges DoOutside diameter

Pipe wall thickness for1)

DN PN 16[mm]

PN 25[mm]

PN 40[mm] [mm]

PN 16[mm]

PN 25[mm]

PN 40[mm]

50 465 ±3 475 ±3 475 ±3 66.6 7.0 7.0 7.065 460 ±3 475 ±3 475 ±3 78.0 7.0 7.0 7.080 380 ±3 400 ±3 400 ±3 92.0 7.0 7.0 7.0900 1230 ±6 1300 ±6 - 914.0 8.8 13.3 -1000 1300 ±6 1370 ±6 - 1016.0 9.7 14.3 -1200 1360 ±6 - - 1220.0 11.3 - -

Size Weight

PN 16 PN 25 PN 40

DN inch kg lbs kg lbs kg lbs

50 2 13 28.6 14 30.8 14 30.8

65 2½ 15 33 16 35.3 16 35.3

80 3 18 39.7 19 42 19 42

900 36 475 1047 835 1841 - -

1000 40 594 1309 1078 2377 - -

1200 48 732 1614 - - - -

B-5$

-

.$

.-$.

-5.$

B-1$

5-B.$

B-

1$

-

$

1-$

-.B$

-.$

1-$

-$

Page 11: SITRANS F flowmeters SITRANS F US - …SITRANS F flowmeters SITRANS F US SITRANS FUS380/FUE380/SONOCAL 3000 Siemens FI 01 · 2007 4/179 4 Overview SITRANS FUS380 The 2-track flowmeter

SITRANS F flowmetersSITRANS F US

SITRANS FUS380/FUE380/SONOCAL 3000

4/189Siemens FI 01 · 2007

4

Schematics SITRANS FUS380

The scheme shows the transducer cable connections between transmitter terminals and respective transducer and the electri-cal connection of the energy calculator SITRANS FUE950.

Schematics SONOCAL 3000

Electrical connections

>

=,

.

..

.

11

&/

11

1B

1.

11

1

'

A

6

+

B 1 5

<; B 1 6<

/

"&!""$

E8 !

F=

>"1. '

"G +

-9*