SINGLE PLANE INNOVA IGS 520 - RENDERING GE Healthcare Project Management...

4
GE Healthcare has made reasonable efforts to ensure this BIM file conforms to requirements of the applicable pre-installation manual. This BIM file was produced to assist with site planning and is not intended for construction or final building modeling. GE Healthcare is not responsible for any damages resulting from use of such file for construction or final building modeling purposes. Compliance with all applicable local, state, and federal laws and regulations is the responsibility of the customer or its designated architects and engineers. Please keep in mind the information in this BIM file is dynamic and is subject to change without notice. It is the responsibility of the customer or its designated architect(s) and engineer(s) to ensure they have the latest equipment or product information. Scale Project number Date Drawn by Checked by GE Healthcare Project Management 12" = 1'-0" A-1 SINGLE PLANE INNOVA IGS 520 - RENDERING 5-110 31.JAN.17 GENPACT -

Transcript of SINGLE PLANE INNOVA IGS 520 - RENDERING GE Healthcare Project Management...

Page 1: SINGLE PLANE INNOVA IGS 520 - RENDERING GE Healthcare Project Management A-1/media/Downloads/us/Support/Site... · 2017. 8. 7. · Project number Date Drawn by Checked by GE Healthcare

GE Healthcare has made reasonable efforts to ensure this BIM file conformsto requirements of the applicable pre-installation manual. This BIM file wasproduced to assist with site planning and is not intended for construction orfinal building modeling. GE Healthcare is not responsible for any damagesresulting from use of such file for construction or final building modelingpurposes. Compliance with all applicable local, state, and federal laws andregulations is the responsibility of the customer or its designated architectsand engineers. Please keep in mind the information in this BIM file isdynamic and is subject to change without notice. It is the responsibility ofthe customer or its designated architect(s) and engineer(s) to ensure theyhave the latest equipment or product information. Scale

Project numberDateDrawn byChecked by

GE HealthcareProject Management

12" = 1'-0"

A-1

SINGLE PLANE INNOVA IGS 520 - RENDERING

5-11031.JAN.17GENPACT

-

Page 2: SINGLE PLANE INNOVA IGS 520 - RENDERING GE Healthcare Project Management A-1/media/Downloads/us/Support/Site... · 2017. 8. 7. · Project number Date Drawn by Checked by GE Healthcare

[6.1

0 m

]

20' -

0"

[2.44 m]

8' - 0"

[9.75 m]

32' - 0"

[2.57 m]

8' - 5"

GE Healthcare has made reasonable efforts to ensure this BIM file conformsto requirements of the applicable pre-installation manual. This BIM file wasproduced to assist with site planning and is not intended for construction orfinal building modeling. GE Healthcare is not responsible for any damagesresulting from use of such file for construction or final building modelingpurposes. Compliance with all applicable local, state, and federal laws andregulations is the responsibility of the customer or its designated architectsand engineers. Please keep in mind the information in this BIM file isdynamic and is subject to change without notice. It is the responsibility ofthe customer or its designated architect(s) and engineer(s) to ensure theyhave the latest equipment or product information. Scale

Project numberDateDrawn byChecked by

GE HealthcareProject Management

1/8" = 1'-0"

A-2

SINGLE PLANE INNOVA IGS 520 - PLAN

5-11031.JAN.17GENPACT

-

Page 3: SINGLE PLANE INNOVA IGS 520 - RENDERING GE Healthcare Project Management A-1/media/Downloads/us/Support/Site... · 2017. 8. 7. · Project number Date Drawn by Checked by GE Healthcare

GE Healthcare has made reasonable efforts to ensure this BIM file conformsto requirements of the applicable pre-installation manual. This BIM file wasproduced to assist with site planning and is not intended for construction orfinal building modeling. GE Healthcare is not responsible for any damagesresulting from use of such file for construction or final building modelingpurposes. Compliance with all applicable local, state, and federal laws andregulations is the responsibility of the customer or its designated architectsand engineers. Please keep in mind the information in this BIM file isdynamic and is subject to change without notice. It is the responsibility ofthe customer or its designated architect(s) and engineer(s) to ensure theyhave the latest equipment or product information. Scale

Project numberDateDrawn byChecked by

GE HealthcareProject Management A-3

SINGLE PLANE INNOVA IGS 520 - SCHEDULE

5-11031.JAN.17GENPACT

Checker

Specialty Equipment Schedule

Description Weight (lbs) BTUExam RoomCABLE DRAPE RAIL 180 0LONGITUDINAL STATIONARY RAIL FORLCD MONITOR

68.00 0

REMOTE CONTROL FOR INJECTOR 4 0OMEGA IV/V TABLE WITH ROTATINGTOP

1300 614

XR BUZZER (LOCATED ABOVECEILING)

2 0

INNOVA POSITIONER 1653 2416INJECTOR HEAD ON TABLE RAIL 15 0COUNTERBALANCED EYEAND|THYROID SHIELD WITH|R96 LAMP

143 0

LARGE DISPLAY MONITOR ONSINGLE|MONITOR SUSPENSION|9 ft. 6in. INBOARD BRIDGE|(MOUNT TWO GEMONITORS ON|BACKSIDE OF LDMONITOR)

784 1706

Specialty Equipment Schedule

Description Weight (lbs) BTUControl RoomBOLUS CHASE HANDSWITCH 2 0REMOTE CONTROL FOR INJECTOR 4 019 in. MONITOR ON WALL SUPPORT 26 20419 in. MONITOR ON WALL SUPPORT 26 204OPERATORS CONSOLE WITH DLKEYPAD

22 546

OPERATORS CONSOLE 22 546BOLUS CHASE HANDSWITCH 2 0AW WORKSTATION 81 1201Equipment RoomINJECTOR ELECTRONICS 37 320LARGE DISPLAY MONITOR CABINET 253 34123 KVA UPS CABINET|(LARGE DISPLAYSUBSYSTEM OPTION)

99 546

CUSTOMER SUPPLIED STORAGECABINET

0 0

150-AMP LOCAL SERVICEDISCONNECT FORLOCK-OUT/|TAG-OUT CAPABILITY.(MAY BE A FUSED|DISCONNECT,CIRCUIT BREAKER OR SAFETYSWITCH.)

0 0

ATLAS CABINET C2 570 1825ATLAS CABINET C1F 1008 3389INNOVA MAIN DISCONNECT,REFERENCE|JUNCTION POINT 'PDB'ON SHEET E1|FOR DETAILEDDESCRIPTION.

326 1532

DETECTOR CHILLER 33 706COOLIX 4100 WATER CHILLER 264 18725COOLIX 4100 AUTOTRANSFORMER 66 15320 KVA UPS CABINET 1170 4061