SINGLE-PHASE POWER QUALITY ANALYZER 8220 ENGLISH User Manual

Transcript of SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section...

Page 1: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

SINGLE-PHASEPOWER QUALITY ANALYZER 8220

E N G L I S H User Manual

Page 2: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Statement of Compliance

Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments certifies that this instrument has been calibrated using standards and instruments traceable to international standards.

We guarantee that at the time of shipping your instrument has met its published specifications.

An NIST traceable certificate may be requested at the time of purchase, or obtained by returning the instrument to our repair and calibration facility, for a nominal charge.

The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer. For recalibration, please use our calibration services. Refer to our repair and calibration section at www.aemc.com.

Serial #: ____________________________________

Catalog #: __________________________________

Model #: 8220 & 8220 Kits

Please fill in the appropriate date as indicated:

Date Received: _________________________________

Date Calibration Due: _______________________

Chauvin Arnoux®, Inc.d.b.a AEMC® Instrumentswww.aemc.com

Page 3: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 1

Table of Contents

1. INTRODUCTION ................................................................................. 41.1 International Electrical Symbols ...........................................................4

1.2 DefinitionofMeasurementCategories .................................................5

1.3 ReceivingYourShipment .....................................................................5

1.4 OrderingInformation ............................................................................51.4.1 AccessoriesandReplacementParts ......................................61.4.2 ThirdPartyAccessories ..........................................................6

2. PRODUCT FEATURES ......................................................................... 72.1 Description ...........................................................................................7

2.2 ControlFeatures ..................................................................................8

2.3 Display ...............................................................................................11

2.4 PowerSupply .....................................................................................13

2.5 OpticallyCoupledUSBInterface........................................................13

2.6 ButtonFunctions ................................................................................132.6.1 PhotographButton ................................................................132.6.2 ViewButton ...........................................................................142.6.3 DeleteButton ...............................................................152.6.4 PrintButton ..................................................................162.6.5 ArrowButtons .......................................................................162.6.6 WhiteButton .........................................................................172.6.7 YellowButton ........................................................................17

3. SPECIFICATIONS ............................................................................. 183.1 ReferenceConditions.........................................................................18

3.2 ElectricalSpecifications .....................................................................183.2.1 VoltageInputs .......................................................................183.2.2 CurrentInputs .......................................................................193.2.3 Bandwidth .............................................................................193.2.4 PowerSupply ........................................................................203.2.5 AccuracySpecifications(excludingcurrentprobes) ...........21

Page 4: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

2 Power Quality Analyzer Model 8220

3.2.6 AccuracySpecificationsoftheCurrentProbes....................233.2.7 CurrentProbesandSensors ................................................24

3.3 MechanicalSpecifications ..................................................................25

3.4 EnvironmentalSpecifications .............................................................26

3.5 SafetySpecifications ..........................................................................26

4. OPERATION ..................................................................................... 274.1 GeneralOperatingInstructions ..........................................................27

4.2 ConnectingtheCables .......................................................................284.2.1 Single-phaseNetwork ...........................................................284.2.2 BalancedThree-phaseNetwork ...........................................28

4.3 Usingthe5AAdapter/MN1935ACurrentProbe ..............................29

4.4 AutomaticShut-off ..............................................................................29

4.5 InstrumentConfiguration ....................................................................304.6 RotarySwitchModes .........................................................................35

4.6.1 VoltageandCurrent .............................................................354.6.2 PowerandThree-PhaseCalculations ..................................374.6.3 VoltageandCurrentHarmonics ............................................414.6.4 InrushandPhaseRotation ...................................................434.6.5 RotationSpeed(RPM) ..........................................................464.6.6 TemperatureandResistance ................................................46

5. DATAVIEW® SOFTWARE ................................................................... 475.1 InstallingDataView® ...........................................................................47

5.2 PowerPadControlPanel ....................................................................49

6. MAINTENANCE ................................................................................ 506.1 ChangingtheBatteries.......................................................................50

6.2 RechargingtheRechargeableBatteries ............................................51

6.3 Cleaning .............................................................................................51

APPENDIX A: MATHEMATICAL FORMULAS .......................................... 52Half-periodVoltageandCurrentRMSValues .............................................52

MIN/MAXValuesforVoltageandCurrent .................................................52

Page 5: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 3

PeakVoltageandCurrent ...........................................................................53

PeakFactorsforCurrentandVoltage(over1second) ...............................53

1secRMSValuesforVoltageandCurrent .................................................53

HarmonicCalculations ................................................................................54

DifferentPowerLevels1SecinSingle-PhaseConnection .........................54

DifferentPowerLevels1SecinBalancedThree-PhaseConnection .........54

Ratios ..........................................................................................................55

KFactor .......................................................................................................55

VariousTypesofEnergy .............................................................................55

Hysteresis ....................................................................................................56

Diagramof4Quadrants ..............................................................................57

SaturationofInputChannels .......................................................................57

Repair and Calibration ....................................................................... 58

Technical and Sales Assistance ......................................................... 58

Limited Warranty ............................................................................... 59

Warranty Repairs ............................................................................... 59

Page 6: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

4 Power Quality Analyzer Model 8220

CHAPTER 1

INTRODUCTION

Warning • Neveruseoncircuitswithavoltagehigherthan600Vandanovervoltage

categoryhigherthanCATIII.• Use in indoor environments only.• Onlyuseaccessoriesthatarecompliantwiththesafetystandards

(IEC664-1Ed.92)600VCATIIIor300VCATIV.• Onlyusefactoryspecifiedreplacementparts.• Alwaysdisconnectthepowercord,measurementleadsandsensors

beforereplacingthebattery.

1.1 International Electrical Symbols

Signifies that the instrument is protected by double or reinforced insulation.

CAUTION - DANGER! Read the User Manual.

Risk of electric shock. The voltage at the parts marked with this symbol may be dangerous.

Refers to a type A current sensor. This symbol signifies that application around and removal from HAZARDOUS LIVE conductors is permitted.Refers to a type B current sensor. Do not apply around or remove from HAZARD-OUS LIVE conductors without additional protective means (de-energizing the circuit or wearing protective clothing suitable for high voltage work).

Important instructions to read and understand completely.

Important information to acknowledge.

USB socket

The CE marking guarantees conformity with European directives and with regulations covering EMC.

The trash can with a line through it means that in the European Union, the product must undergo selective disposal for the recycling of electric and electronic material, in compliance with Directive WEEE 2002/96/EC.

Page 7: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 5

1.2 Definition of Measurement CategoriesCAT IV: Formeasurementsperformedattheprimaryelectricalsupply(<1000V)

suchasonprimaryovercurrentprotectiondevices,ripplecontrolunits,ormeters.

CAT III: Formeasurementsperformedinthebuildinginstallationatthedistributionlevel such as on hardwired equipment in fixed installation and circuitbreakers.

CAT II: For measurements performed on circuits directly connected to theelectricaldistributionsystem.Examplesaremeasurementsonhouseholdappliancesorportabletools.

1.3 Receiving Your ShipmentMakesure that thecontentsshippedareconsistentwith thepacking list.Notifyyourdistributorofanymissing items. If theequipmentappears tobedamaged,fileaclaimimmediatelywiththecarrierandnotifyyourdistributoratonce,givingadetaileddescriptionof anydamage.Save thedamagedpacking container tosubstantiateyourclaim.Donotuseequipmentwhichisdamagedorappearstobedamaged.

1.4 Ordering Information

Model 8220 ......................................................................................Cat. #2130.90Includes set of two 10 ft (3m) color-coded leads (red/black) with alligator clips, set of safety test probes (red/black), optically isolated USB cable, set of six 1.5V AA batteries, carrying bag, and USB stick supplied with product user manual and DataView® software.

Model 8220 w/MN93-BK .................................................................Cat. #2130.91Includes the Model 8220, one MN93 (240A) current probe (black connector), set of two 10 ft (3m) color-coded leads (red/black) with alligator clips, set of safety test probes (red/black), optically iso-lated USB cable, set of six 1.5V AA batteries, DataView® software, carrying bag, and user manual.

Model 8220 w/SR193-BK ............................................................... .Cat. #2130.92Includes the Model 8220, one SR193 (1200A) current probe (black connector), set of two 10 ft (3m) color-coded leads (red/black) with alligator clips, set of safety test probes (red/black), optically isolated USB cable, set of six 1.5V AA batteries, carrying bag, and USB stick supplied with product user manual and DataView® software.

Model 8220 w/24" AmpFlex® 193-24-BK .......................................Cat. #2130.93Includes the Model 8220, one 24" AmpFlex® 193-24 (6500A) sensor (black connector), set of two 10 ft (3m) color-coded leads (red/black) with alligator clips, set of safety test probes (red/black), optically isolated USB cable, set of six 1.5V AA batteries, carrying bag, and USB stick supplied with product user manual and DataView® software.

Page 8: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

6 Power Quality Analyzer Model 8220

Model 8220 w/MN193-BK ...............................................................Cat. #2130.96Includes the Model 8220, one MN193 (6A/120A) current probe (black connector), set of two 10 ft (3m) color-coded leads (red/black) with alligator clips, set of safety test probes (red/black), optically isolated USB cable, set of six 1.5V AA batteries, carrying bag, and USB stick supplied with product user manual and DataView® software.

Model 8220 w/MA193-10-BK ..........................................................Cat. #2130.97Includes the Model 8220, one MA193 (1000A) current probe (black connector), set of two 10 ft (3m) color-coded leads (red/black) with alligator clips, set of safety test probes (red/black), optically isolated USB cable, set of six 1.5V AA batteries, carrying bag, and USB stick supplied with product user manual and DataView® software.

1.4.1 Accessories and Replacement PartsAdapter110Voutletw/4mmBananaPlug .........................................................Cat. #2118.49SoftCarryingPouch ..........................................................................................Cat. #2119.48SmallClassicCarryingBag ...............................................................................Cat. #2133.72OpticallyIsolatedUSBCable(RS-232Connector) ...........................................Cat. #2135.415AAdapterBox(for1or5Aprobes) ..................................................................Cat. #2140.17AC/DCCurrentProbeModelSL261..................................................................Cat. #1201.51ACCurrentProbeModelMR193-BK.................................................................Cat. #2140.28SetofTwo10ftColor-codedLeadswithAlligatorClips ....................................Cat. #2140.31ACCurrentProbeModelMN93-BK...................................................................Cat. #2140.32ACCurrentProbeModelSR193-BK .................................................................Cat. #2140.33AmpFlex® Sensor 24"193-24-BK ......................................................................Cat. #2140.34AmpFlex® Sensor 36"193-36-BK ......................................................................Cat. #2140.35ACCurrentProbeModelMN193-BK.................................................................Cat. #2140.36MiniFlex® Sensor 10"ModelMA193-10-BK.......................................................Cat. #2140.48110VPowerAdapter ..........................................................................................Cat. #2140.37220VPowerAdapter..........................................................................................Cat. #2140.38BNCAdaptorforusewithSL261 .......................................................................Cat. #2140.40

1.4.2 Third Party Accessories

Converter-EthernettoRS-232,ModelESP901Available Online ........................................................................................ www.bb-elec.com

Order Accessories and Replacement Parts Directly OnlineCheck our Storefront at www.aemc.com/store for availability

Page 9: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 7

CHAPTER 2

PRODUCT FEATURES

2.1 DescriptionTheModel8220isasingle-phaseAC+DCpowermeterwithanelectroluminescentbacklitdigitaldisplayandisratedto600VCATIII.It is a measurement tool for electrical parameters and distribution networkdisturbances, enabling the user to obtain instant measurements of the maincharacteristics of a single-phase network (voltage, current, power, voltage andcurrentharmonics,etc)andtomonitormachineryinoperation(temperature,inrushcurrentanddurationofoperation,resistanceofwindingsandrotationalspeed).Compactandshockresistant,itsergonomicdesignandstraightforwardinterfacemake it user-friendly and intuitive. Within seconds you will be gathering themeasurementsyouwant.TheModel8220’saccuracyisbetterthan1%(excludingcurrentsensors).IthasagreatflexibilityduetoAEMC’srangeofcurrentsensors,measuringfromafewhundredmilliampstoseveralthousandamps.TheModel8220isdesignedfortechnicians,controlandmaintenanceengineers,aswellaselectricalcontractors,inspectorsandeducators.

NOTE: The Model 8220 can be powered by 6 x AA batteries or a 110V/220VAC power adaptor (with batteries removed).

Features:• Measuresupto660VrmsorVDC• Measuresupto6500AAC or 1400ADC(probedependent)• DisplaysMin,MaxandAverageVoltsandAmps,CrestFactor,Peak

valueandK-Factor• CalculatesanddisplaysWatts,VARsandVA,PowerFactorand

DisplacementPowerFactorforsingle-phaseandbalancedthree-phase• Displaystotalharmonicdistortion(THD-FandTHD-R)forvoltageand

current• Displaysindividualharmonicvaluesand%forVoltsandAmpsthrough

the50thharmonic• Capturesanddisplaysinrushcurrent• CalculatesanddisplaysphaserotationandRPM

Page 10: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

8 Power Quality Analyzer Model 8220

• Displaysresistanceupto2000Ω

• Conductscontinuityanddiodetests• Storesuptoninecompletesetsofreadingsforallvolt,amp,powerand

harmonicmeasurements(photographmode)• ConfigurablefromDataView®softwareorfrontpanel• Download stored data to DataView®softwareviaopticalUSBport• IncludesFREEDataView®softwarefordatastorage,real-timewave-

formdisplay,analysisandreportgeneration

2.2 Control Features

Use FactoryPower Supply

Use FactoryCurrent Probe

VoltageInputs

600V CAT III300V CAT IV

COM +

5

6

1

7

8

9

2 3 4

Figure 2-1

Page 11: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 9

ITEM FUNCTION1. External power supply jack

2. Four-pin input for current sensor (MN, SR, AmpFlex, etc.)

3. Negative terminal

4. Positive terminal

5. Electroluminescent backlit digital display

6. Function Buttons:

Photograph ButtonPhotographs measurements, viewable by pressing the View or Print buttons.

Print ButtonPrints all measurements currently displayed on an external printer.

Left Arrow ButtonSelects the previous page, returns to previous step or (if only one page) selects the value.

Right Arrow ButtonSelects the following page, advances to the next step or (if only one page) selects the value.

View ButtonViews saved screen photographs.

Delete ButtonDeletes one or all of the saved photographs or re-initializes the effective maximum and minimum half-cycle values in Inrush mode.

White ButtonAccesses voltage, power, voltage harmonic, motor start current, rotation speed, temperature and information modes and reduces the value in the configuration mode.

Yellow ButtonAccesses the current, balanced three-phase, current harmonic, phase rotation and resistance modes, reduces the value in the configuration mode. Also inhibits automatic extinction.

NOTE: For detailed information on these buttons, see § 2.6

Page 12: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

10 Power Quality Analyzer Model 8220

7. Rotary Functions:

NOTE: To access the second function of a rotary position, turn the switch to the desired position then press the Yellow button.

OFF Turns the Instrument OFF

Voltage and Current modes

Power and Balanced Three-phase modes

Voltage and Current Harmonics modes

Inrush (motor start) and Phase Rotation modes

Rotation speed mode

Temperature and Resistance modes

Configuration mode

NOTE: For detailed information on these modes, see § 4.6.

8. Infrared optically coupled interface for USB cable.

9. Light indicator:ON - indicates the 8220 is running on the optional external power supply.OFF - indicates the 8220 is running on the internal batteries.

Page 13: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 11

2.3 Display

Figure 2-2

PressingtheWhiteandYellowbuttonssimultaneouslytogglesthebacklightONandOFF.

Thedisplayusesthefollowingiconsandabbreviations:

ICON DESCRIPTION

Voltage measurement

Current measurement

Power measurement (active, reactive, apparent)

Calculation associated with connection to a balanced activated three phase network

Voltage or current harmonics measurement

Motor starting (inrush) measurement

Phase rotation

Rotation speed measurement

Temperature measurement

Resistance measurement (up to 2000Ω)

When displayed, indicates the battery is low

Flashes during data transfer of information to the external printer

Page 14: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

12 Power Quality Analyzer Model 8220

Number of active pages compared with total number of pages in the mul-tiple pages modes

Flashes while measurements are being captured

Recalls a list of saved photographs; views a photograph

Deletion of one or all of the photographs. Re-initialization of effective values during half-period of voltage or current.

% Percentage

Ω Resistance in ohms

°C Temperature in degrees Celsius

°F Temperature in degrees Fahrenheit

A Current in amperes

AVG Real RMS value of the signal calculated over one second

CF Crest Factor (current or voltage)

DC Continuous current and voltage component

DPF Displacement Factor (cosine of Φ)

Hz Network frequency in hertz

k Kilo (103)

KF K Factor (for transformers)

MAX Maximum RMS half-cycle voltage or current value

MIN Minimum RMS half-cycle voltage or current value

PEAK Instantaneous maximum (+) or minimum (-) crest value of signal measured

PF Power Factor (ratio of active power on the apparent power)

RMS Real effective value (voltage or current)

RPM Rotation speed in revolutions per minute (rotation per minute)

s Duration in seconds

THD-F Total Harmonic Distortion (or THD)

THD-R Distortion Factor (or DF)

V Voltage in volts

VA Apparent Power (total if 3Φ)

VAR Reactive Power (total if 3Φ)

Active Power (total if 3Φ)

Page 15: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 13

2.4 Power SupplyTheModel8220ispoweredbysix1.5VAA(LR6)alkalinebatteries(standardorrechargeable).See§6forinformationonchangingandrechargingthebatteries.

Batterylifeisdependentontypeofbattery:

Type of power supplyBack-lighting

Without WithAA batteries > 40 hours > 20 hoursNiMH 1800mAh accumulators > 30 hours > 16 hoursNiCd 900mAh accumulators > 15 hours > 8 hours

Low Battery Indication Level:

Level1- The symbolwillflashoncepersecondwhenthebatteryvoltageisweak,buttheinstrumentcanstillbeused.

Level2- The symbolflashesoncepersecond.Inaddition,every10seconds(and repeated 7 times for awholeminute) a beeping sound is emit-tedtogetherwiththemessage“batt”displayedonthescreen.Afteroneminute,theinstrumentturnsitselfoff.

2.5 Optically Coupled USB InterfaceTheopticallycoupledinterface(Figure2-1,item8)providesanisolated,two-wayconnectionbetweentheModel8220andaPCfortransmissionoftheinformationinmemory(photographs,motorstarts,records)andall instantaneousmeasure-mentsandphotographsdisplayedonthescreenoftheModel8220.

ThetransferrateisdeterminedautomaticallybytheModel8220accordingtothesoftwareused;themaximumrateis115.2kbps.

2.6 Button Functions

2.6.1 Photograph Button

This button photographs all themeasurement pages displayed in the followingrotaryswitchpositions: , , .

When the button is pressed, all the pages in the voltage, current, power,voltageharmonicsandcurrentharmonicsmodesarephotographedandthenextscreenisdisplayed.

Page 16: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

14 Power Quality Analyzer Model 8220

Figure 2-3Example of a photograph number.

Thisscreendisplaysthenumberofthephotographinwhichthepageshavebeenstored.Thephotographnumberflashesduring thestorageprocessand the iconisdisplayed.

NOTE: If the memory used for storing the photographs is full, the message

(Memory full) message is displayed.

2.6.2 View Button

Thisbuttonfirstgivesaccesstothelistofphotographs,thenenablestheusertoselectaphotographtobeviewed.

To Display a List of Photographs:

Pressthe button.Anumberofaphotographinthelistwillbedisplayed.Usethe◄►buttonsto:

• Displaythenumberofaspecificphotographtobeviewed(seeillustrationabove).

• Select toselectallphotographs.

ThislistcanbeexitedbypressingeithertheWhiteorYellowbutton( )orthe or button,orbychangingthepositionoftheswitch.

IftheModel8220memorydoesnotcontainanyphotographs, isdisplayedonthescreen.Theinstrumentthenautomaticallyreturnstothepreviewmode.

Page 17: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 15

To Display Pages of a Photograph:

Afterselectingthepagesofaphotographaspreviouslydescribed,pressthe button.

Duringthetimethat thephotographisbeingviewed, the iconflashes inthebottomright-handcornerofthedisplay.Toviewthedifferentpages:

• Selectafunction- , or ontherotaryswitch.• Usethewhiteandyellowbuttons( ) to select a mode.• Usethe◄►buttonstonavigatearoundthedifferentpagesofthe

selected mode.

Toreturntothelistofphotographs,pressthe buttonagain.

2.6.3 Delete Button Thisbuttonisusedtodeleteoneorallofthepreviouslysavedphotographs.

To Delete a Photograph:

Pressthe buttontoviewthelistofphotographs.

Usethe◄►buttonsto:

• Displaythenumberofaspecificphotographtobedeleted(seeillustrationonpreviouspage).

• Select toselectallphotographs.

Pressthe buttontodeletetheselectedphotograph.

WARNING: Selecting will delete all saved photographs.

Afterdeletingaspecificphotograph,oneofthenumbersoftheremainingphoto-graphsisdisplayed.Ifthelistdoesnotcontainanyfurtherphotos,thedisplayindicates (norecording).Thedisplaythenswitchesautomaticallytoameasurementmode.

ThislistcanbeexitedbypressingeithertheWhiteorYellowbutton( )orthe or button,orbychangingthepositionoftheswitch.

If wasselected, isdisplayedonthescreen.Thedisplaythenauto-maticallyswitchestoameasurementmode.

Page 18: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

16 Power Quality Analyzer Model 8220

2.6.4 Print Button

Beforeusingthisbutton,anexternalserialprintermustbeconnectedtotheModel8220usingthespecificserialcableprovidedwiththeprinter.

NOTE: Do not use the optical USB cable provided with the instrument, to connect to the printer. Use the serial cable provided with the printer.

Pressingthisbuttonprintsoutalltheinformationandmeasurementsrelatingtothecurrentlyselectedmode.

NOTE: It is not possible to print while viewing a photograph.

Whendataistransmitting,the iconflashes.

Exampleofaprintoutwhentherotaryswitchisturnedtothe voltagemode:

************************** Mode:voltage Vrms(V)= 225.7 Arms(A)= 112.8 Freq(Hz)= 50.00 Vdc(V)= 3.2 CF= 1.41 Vmax(V)= 229.4 Vmin(V)= 222.6 Vpeak+(V)= 319.6 Vpeak-(V)=-313.3

Figure 2-4

2.6.5 ◄► Buttons

Thesebuttonsenabletheuserto:

• Navigatebetweenpagesforthe , , and modes.

• Activateorde-activatethebalancedthree-phasemodeforthe mode.

• Selectthestepinthe modes.

Page 19: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 17

2.6.6 White Button

Thisbuttonenables:

• Selectionofthefirstfunctionofeachmode,indicatedbythewhitemarkingsaroundtherotaryswitch( , , , , ).

• Decreasingthevaluesinthe mode.

• Exitingthelistofphotographs.

• Displayoftheinformationrelatingtothe8220(serial,softwareandhardwarenumbers).This information is accessedby pressing and holding down thewhitebuttonwhenturningtherotaryswitchfromOFFtoanyotherposition.

2.6.7 Yellow Button

Thisbuttonenables:

• Selectionofthesecondfunctionofeachmode,indicatedbytheyellowmark-ingsaroundtherotaryswitch( , , , , ).

• Increasingthevaluesinthe mode.

• Exitingthelistofphotographs.

• De-activationoftheautomaticshut-off(see§4.4).

Page 20: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

18 Power Quality Analyzer Model 8220

CHAPTER 3

SPECIFICATIONS

3.1 Reference Conditions

Parameter Reference Conditions

Ambient temperature 73° ± 5°F (23° ± 3°C)

Humidity 45 to 75%

Atmospheric pressure 25.4" Hg to 31.3" Hg (860 to 1060 hPa)

Phase-to-earth voltage 50 to 600Vrms without DC (< 0.5%)

Clamp current circuit input voltage 30mVrms to 1Vrms without DC (<0.5%)

AmpFlex® current circuit input voltage 11.8 to 118mVrms without DC (<0.5%)

Network frequency 50 and 60Hz ± 0.1Hz

Phase shift 0° active power / 90° reactive power

Harmonics <0.1%

Balanced three-phase connection Deactivated (OFF)

*All specifications are subject to change without notice.

3.2 Electrical Specifications

3.2.1 Voltage Inputs

Switch Positions: , , , ,

Operating Range:Phase-Phase-0to660VrmsAC/DC*Phase-Neutral-0to600VrmsAC/DC*Provided that the max with 600Vrms in regards to earth is not exceeded.

Input Impedance: 451kΩ

Overload:1.2Vnpermanently;2Vnfor1sec(Vn=nominalvoltage)

Page 21: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 19

Switch Position:

Input Impedance: 450kΩ

Overload: 600Vrmscontinuous

Switch Position:

Voltage in Open Circuit: ≤4.6V

Measurement Current: 500µA

Overload: 600Vrmscontinuous

Threshold for Buzzer: 20Ω(bydefault)

3.2.2 Current Inputs

Operating Range: 0 to 1V

Input Impedance:1MΩforcurrentprobecircuitand12.4kΩforAmpFlex®circuit

Overload: 1.7V

3.2.3 Bandwidth

Measurement Channels:256pointsperperiod,or:

• For50Hz:6.4kHz(256×50÷2)• For60Hz:7.68kHz(256×60÷2)

Analog to -3dB:>to10kHz

Page 22: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

20 Power Quality Analyzer Model 8220

3.2.4 Power Supply

Power Source

Battery: • 6 standard batteries provided (non-rechargeable) in AA format (IEC LR6 – NEDA 15A)

• 6 optional rechargeable batteries (rechargeable elements) NiMH or NiCd in AA format (IEC LR6 – NEDA 15A).

Rechargeable batteries:

Capacity:

(minimum)

NiMh: 1800mAh

NiCd: 900mAhNominal voltage: 1.2V per battery or 7.2V in total.

Battery life Without backlighting With backlightingAA standard

batteries> 40 hours > 20 hours

NiMH 1800mAh rechargeable

batteries> 30 hours > 16 hours

NiCd 900mAh rechargeable

batteries> 15 hours > 8 hours

Temperature:Operating

Storage

32° to 122°F (0°C to 50°C)

Standard batteries: -4° to 158°F (-20°C to 70°C)Rechargeable batteries: -4° to 122°F (-20°C to 50°C)

NOTE: Rechargeable batteries must be recharged on an external charger.

AC Power (externalpowersupply)

Operating Range:120V±10%@60Hz;230V±10%@50Hz(modeldependent)

Max Power: 23.7VA

TheinstrumentmaybepoweredbyanoptionalPowerAdapter110V(Cat.#2140.37)or220V(Cat.#2140.38)

Page 23: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 21

3.2.5 Accuracy Specifications (excluding current probes)

FunctionMeasurement range Display

ResolutionAccuracy

(% of Reading ± ct)Minimum Maximum

Frequency 40Hz 69Hz 0.01Hz ±(1ct)

TRMS voltage 6V 600V (6) 0.1V ±(0.5%+2cts)

DC voltage 6V 600V 0.1V ±(1%+5cts)

TRMS current

ExcludingAmpFlex®

& MiniFlex®*Inom ÷ 1000

[A]1.2 × Inom

[A]

0.1A I < 1000A

±(0.5%+2cts)

1A I ≥ 1000A

±(0.5%+1ct)

AmpFlex®

& MiniFlex®* 10A 6500A

0.1A I < 1000A

±(0.5%+1ct)1A

I ≥ 1000A

DC current 1A 1700A (1)

0.1A I < 1000A

±(1%+1A)1A

I ≥ 1000A

Current Peak

ExcludingAmpFlex®

& MiniFlex®* 0A

1.7 × Inom

[A](2)

0.1A I < 1000A

±(1%+1A)AmpFlex®

& MiniFlex®* 9190A(3) 1A I ≥ 1000A

Half-period TRMS current(4)

ExcludingAmpFlex®

& MiniFlex®*Inom ÷ 100

[A]1.2 × Inom

[A]

0.1A I < 1000A

±(1%+5cts)

1A I ≥ 1000A

±(1%+1ct)

AmpFlex®

& MiniFlex®* 100A 6500A

0.1A I < 1000A

±(1.5%+4A)1A

I ≥ 1000A

Peak voltage 6V 850V (4) 0.1 V ±(1%+5cts)

Half-period TRMS Voltage(5) 6V 600V 0.1V ±(0.8%+5cts)

Peak Factor1 4 0.01 ±(1%+2cts)

4 9.99 0.01 ±(5%+2cts)

Active Power

ExcludingAmpFlex®

& MiniFlex®*0W 9999kW 4 digits

±(1%) Cos φ ≥ 0.8

±(1.5%+10cts) 0.2 ≤ Cos φ < 0.8

AmpFlex®

& MiniFlex®* 0W 9999kW 4 digits

±(1%) Cos φ ≥ 0.8

±(1.5%+10cts) 0.2 ≤ Cos φ < 0.8

Reactive Powers

ExcludingAmpFlex®

& MiniFlex®*0VAR 9999kVAR 4 digits

±(1%) Sin φ ≥ 0.5

±(1.5%+10cts) 0.2 ≤ Sin φ < 0.5

AmpFlex®

& MiniFlex®* 0VAR 9999kVAR 4 digits

±(1.5%) Sin φ ≥ 0.5

±(2.5%+20cts) 0.2 ≤ Sin φ < 0.5

Apparent power 0 9999kVA 4 digits ±(1%)

Power factor -1 1 0.001

±(1.5%) Cos φ ≥ 0.5

±(1.5%+10cts) 0.2 ≤ Cos φ < 0.5

*NOTE: MiniFlex® specified up to 1000A only

Page 24: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

22 Power Quality Analyzer Model 8220

FunctionMeasurement range

Display resolution AccuracyMinimum Maximum

Displacement factor(DPF)

-1 1 0.001±(1°) on φ

±(5cts) on DPFLevel of harmonics

order ∈ [1; 50](VRMS > 50V)

0% 999.9% 0.1% ±(1%+5cts)Without AmpFlex® or MiniFlex®*(IRMS > 3 × Inom ÷ 100)

AmpFlex® or MiniFlex®* (IRMS > Inom ÷ 10)

Global level of harmonics(THD-F) order ≤ 50

0% 999.9% 0.1% ±(1%+5cts)

Distortion factor (THD-R) order ≤ 50

0% 999.9% 0.1% ±(1%+10cts)

K factor 1 99.99 0.01 ±(5%)

Rotation Speed 6 RPM 120 kRPM

0.1 RPMV < 1 kRPM

± (0.5 %)

1 RPM1 kRPM ≤ V < 10 kRPM

10 RPM10 kRPM ≤ V< 100 kRPM

100 RPM V ≥ 100 kRPM

Temperature

-200.0 °C 850.0 °C 0.1 °C ± (1 % + 1 °C) (6)

-328.0 °F 1562 °F

0.1 °FT < 1000°F

± (1.8 % + 2 °F) (6)

1 °FT ³ 1000°F

Resistance 0.0Ω 2000Ω 1 ΩR < 1000Ω ± (1.5 %+2 cts)

(1) 1.2 x 1000 x √2 - 1700A

(2) 1.2 x Inom x √2 = 1.7 x Inom

(3) 6500 x √2 - 9190A

(4) 600 x √2 = 850V

(5) For the phase-to-earth voltage measurement (phase-neutral). For the phase-to-phase voltage measurement (phase-phase), in balanced three-phase mode, it is possible to reach 660VRMS (balanced three-phase network having a phase-neutral voltage of 380VRMS).

(6) An additional influence of 3.5°C must be added in an environment disturbed by radiation fields. In other words, the error in the field of reference in an environment disturbed by radiation fields is ± (1 %+4.5°C).

An additional influence of 6.2°F must be added in an environment disturbed by radiation fields. In other words, the error in the field of reference in an environment disturbed by radiation fields is ± (1 %+8°F).

NOTE: The uncertainties given for power and energy measurements are maximum for Cos φ = 1 or Sin φ = 1 and are typical for the other phase shifts.

Page 25: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 23

3.2.6 Accuracy Specifications of the Current Probes

Thesecharacteristicsarestatedafterlinearization.Theerrorsofthesensorsarecompensatedbyatypicalcorrectioninsidetheinstrument.Thistypicalcorrectionisinphaseandinamplitudeaccordingtothetypeofsensorconnected(detectedautomatically)andthegainofthecurrentacquisitionchainused.The measurement error in RMS current and the phase error are additionalerrors(theymustthereforebeaddedtothoseoftheinstrumentalone)statedasinfluencesonthecalculationsperformedbytheanalyzer(powers,powerfactors,displacementfactors,etc.).

Type of sensor TRMS currentMaximum error on IRMS

(% of Reading ± ct)Maximum error on φ

MR1931000A

[1A; 10A]±(1.5% + 1A)

Not Specified

[10A; 100A] ±(2°)

[100A; 800A] ±(3%)

±(1.5°)[800A; 1200A]±(5%)[1200A; 1400A]

DC only

SR1931000A

[1A; 3A]±(0.8%)

Not Specified

[3A; 10A] ±(1°)

[10A; 100A] ±(0.3%) ±(0.5°)

[100A; 1200A] ±(0.2%) ±(0.3°)

AmpFlex®

3000A[10A; 100A] ±(3%) ±(1°)

[100A; 6500A] ±(2%) ±(0.5°)

MiniFlex®

1000A[10A; 100A] ±(3%) ±(1°)

[100A; 1000A] ±(2%) ±(0.5°)

MN93200A

[0.5A; 2A]±(3% + 1A)

Not Specified

[2A; 10A] ±(6°)

[10A; 100A] ±(2.5% + 1A) ±(3°)

[100A; 240A] ±(1% + 1A) ±(2°)

MN193100A

[100mA; 300mA]±(0.7% + 2mA)

Not Specified

[300mA; 1A] ±(1.5°)

[1A; 120A] ±(0.7%) ±(0.7°)

MN1935A

[5mA; 50mA] ±(1% + 0.1mA) ±(1.7°)

[50mA; 500mA] ±(1%)±(1°)

[500mA; 6A] ±(0.7%)

5AAdapter

[5mA; 50mA] ±(1%) ±(1°)

[50mA; 6A] ±(0.5%) ±(0°)

SL261 10mV/A[0 A ; 40 A] ±(2% + 50mA) ±(0.5°)

[40 A; 100 A] ±(5%) ±(0.5°)

SL261 100mV/A [0 A ; 10 A] ±(1.5% + 50mA) ±(1°)

Page 26: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

24 Power Quality Analyzer Model 8220

3.2.7 Current Probes and Sensors

Wheninstallingprobes,facethearrowontheprobeinthedirectionoftheload.

Model SR193Nominal Range 1000AAC for f ≤1kHz

Measurement Range 1 to 1200AAC max (I >1000A not continuously)

Probe Output Signal 1mVAC/AAC

Maximum Clamping Diameter 2" (52mm)

Safety EN 61010-2-032, Pollution Degree 2, 600V CAT IV, 1000V CAT III

NOTE: Currents <0.5A will be displayed as zero. Neutral current measures down to 0A.

Model MN93Nominal Range 200AAC for f ≤1kHz

Measurement Range 2 to 240AAC max ( I >200A not permanent)

Probe Output Signal 5mVAC/AAC

Maximum Clamping Diameter 0.8" (20mm)

Safety EN 61010-2-032, Pollution Degree 2, 300V CAT IV, 600V CAT III

NOTE: Currents <0.5A will be displayed as zero. Neutral current measures down to 0A.

Model MR193Nominal Range 1000AAC, 1400ADC max

Measurement Range 10 to 1000AAC, 10 to 1300APEAK AC+DC

Probe Output Signal 1mV/A

Maximum Clamping DiameterOne 1.6" (42mm) or two 0.98" (25.4mm) ortwo bus bars 1.96 x 0.19" (50 x 5mm)

Safety EN 61010-2-032, Pollution Degree 2, 300V CAT IV, 600V CAT III

NOTE: Currents <1AAC/DC will be displayed as zero. Neutral current measures down to 0A.

Model MN193Nominal Range 5A and 100AAC

Measurement Range5A: 0.005 to 6AAC max (1 to 1200A with ratio 1000/5 selected)100A: 0.1 to 120AAC max

Probe Output Signal 5A: 200mV/AAC; 100A: 10mV/AAC

Maximum Clamping Diameter 0.8" (20mm)

Safety EN 61010-2-032, Pollution Degree 2, 300V CAT IV, 600V CAT III

The 5A range of the MN193 is designed to work with secondary current transformers. Best accuracy is available when entering the transformer ratio (e.g. 1000/5A). When used to measure 5A direct, the resolution will be limited to 0.1A max.

NOTE: Currents < (Primary x 5) ÷ (Secondary x 1000) or <250mA on the 5A range and <0.2A on the 100A range will be displayed as zero with this probe. Power calculations will also be zeroed when the current is zeroed.

Page 27: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 25

AmpFlex® SensorsNominal Range 3000AAC

Measurement Range 10 to 6500AAC

Probe Output Signal* 140mVAC/3000AAC at 60Hz

SensorLength = 24" (610mm); Ø = 7.64" (190mm)Length = 36" (910mm); Ø = 11.46" (290mm)

Safety EN 61010-2-032, Pollution Degree 2, 600V CAT IV, 1000V CAT III

*Output is proportional to the amplitude and frequency of the measured current.

NOTE: Currents <10A will be displayed as zero. Neutral current measures down to 0A.

MiniFlex® SensorsNominal Range 1000AAC

Measurement Range 10 to 1000AAC

Probe Output Signal* 47µVAC/1000AAC at 60Hz

Sensor Length = 10" (250mm); Ø = 2.75" (70mm)

Safety EN 61010-2-032, Pollution Degree 2, 600V CAT IV, 1000V CAT III

*Output is proportional to the amplitude and frequency of the measured current.

NOTE: Currents <10A will be displayed as zero. Neutral current measures down to 0A.

Model SL 261Nominal Range 50mAAC/DC- 100A Peak+DCAC

Measurement Range 50mA to 10A, 1A to 100A

Probe Output Signal 1A: 100mV/A: 100A to 10mV/A

Maximum Clamping Diameter 0.46" (11.8mm)

Safety EN 61010-2-032, Pollution Degree 2, 600V CAT III 300V CAT IV

3.3 Mechanical SpecificationsDimensions: 8.3x4.3x2.4"(211x108x60mm)

Weight: 1.94lbs(880g)

Shock and Vibration: perEN61010-1

Tightness: IP54perEN60529 (electrical IP2X for the terminals)

Page 28: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

26 Power Quality Analyzer Model 8220

3.4 Environmental SpecificationsReference Temperature: 20to26°C(68to78.8°F)from45to75%RH

Operating Temperature: 0to50°C(32to122°F)from10to85%RH

Storage Temperature:

Withbatteries: -20to50°C(-4to122°F)from10to85%RH

Withoutbatteries:N/A

Recharging Temperature: N/A

Altitude: Operating:0to2000meters(6560ft) Non-Operating:0to10,000meters(32800ft)

3.5 Safety Specifications

Electrical Safety

600VCATIII,PollutionDegree2EN61010-31:2002EN61010-1:2001EN61010-2:1995

Electromagnetic CompatibilityImmunity:EN61236-1A2Emission:EN61236-1A2Electrostaticdischarges:IEC1000-4-2Radiationfieldresistance:IEC1000-4-3Fasttransientsresistance:IEC1000-4-4Electricshockresistance:IEC1000-4-5ConductedRFinterference:IEC1000-4-6InterruptionofVoltage:IEC1000-4-11

Page 29: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 27

CHAPTER 4

OPERATION

4.1 General Operating Instructions

WARNING: Never connect voltages exceeding 600Vrms

TurntherotaryswitchtoanypositionotherthanOFF.Fordetailedinformationonbuttonfunctionsandrotaryswitchmodes,referto§2.6and§4.6.

TheModel8220displaysthreescreensasfollows:

• First Screen:Displayofallsymbolsthatareused,dependingonmode.• Second Screen: Displayofthemodelnumberoftheinstrument.• Third Screen: Displayofthemeasurementpagecorrespondingtotheposi-

tionoftheswitch.

NOTE: If a current sensor is connected, the bottom of the second screen indicates the type of current sensor connected, with the following code:

MR193 1000A Current Probe

MN93 200A Current Probe

MN193 100A or 5A Current Probe

SR193 1000A Current Probe

AmpFlex® 3000A or MiniFlex® 1000A Current Sensor

SL261 10A or 100A Current Probe

Three-phase 5A Adapter

TheModel8220worksfromabatteryonlyifthebatteryissufficientlycharged.TheinstrumentcanbeusedwiththeoptionalACpowerunitconnectedtothepowerjack(seeFigure2-1,item1).

Page 30: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

28 Power Quality Analyzer Model 8220

4.2 Connecting the Cables

ConnectthemeasurementcablestotheModel8220.

• Voltage measurement: COMand(+)terminals.• Current measurement:4-pinconnector.Onthecurrentsensor,donotforget

topositiontheswitch(ifavailable)toasensitivitysuitedtothecurrenttobemeasured.

Themeasurementcablesareconnectedtothecircuittobemonitoredasperthefollowingdiagrams.

4.2.1 Single-phase Network

Allvoltagesmeasuredwillbeline-to-neutral.

The mode(balancedthree-phase)willbede-activated( ).See§4.6.2.

Figure 4-1Single-phase connection

4.2.2 Balanced Three-phase Network

Allvoltagesmeasuredwillbephase-to-phase.

The mode(balancedthree-phase)willbeactivated( ).See§4.6.2.

Figure 4-2Balanced three-phase connection

LNL

L2L1L3

Page 31: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 29

4.3 Using the 5A Adapter / MN193 5A Current ProbeIf the5Aadapter or theMN1935Aclamp is used, anadjustment to the trans-formationratio (primarycurrent (1A to2999A) /secondarycurrent (1Aor5A) isrequired.

Proceedasfollows:

1. Connectthecurrentsensor.

2. SelecttheConfigurationmodebypositioningtherotaryswitchto .

3. Adjustthesensor’sprimarycurrent.• Selectthe parameter(primaryadjustmentpage)withthe◄►but-

tons.• Withthewhiteandyellowbuttons,adjusttheprimarycurrent( )ofthe

transformationratio.See§4.5fordetails.

4. Adjustthesecondarysensorcurrent.• Select the parameter (secondaryadjustmentpage)with the◄►

buttons.• Withthewhiteandyellowbuttons,adjustthesecondarycurrent( )of

thetransformationratioto1or5A.See§4.5fordetails.

4.4 Automatic Shut-off

Bydefault,theautomaticshut-offfeatureisactivatedeachtimetheModel8220isturnedon.

To de-activate:

1. TurntheunitOFFbyturningtherotaryswitchtotheOFFposition.

2. TurntheunitbackONbyturningtherotaryswitchtoanyotherposition.

3. Whenthefirstscreenisdisplayed(displayofallsymbols),pressandholdtheyellowbuttonuntilabeepisheard.

4. Thescreenindicates (Noautomaticshut-off).

Page 32: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

30 Power Quality Analyzer Model 8220

4.5 Instrument Configuration ThisrotaryswitchpositionisdesignedforconfiguringtheparametersusedbytheModel8220.Oncetheconfigurationhasbeensavedinthenon-volatilememoryitcanbeaccessedevenaftertheinstrumentisshutoff,includingduringabatterychange.

Turntherotaryswitchtothe position.Thereare6pagesinthismode.Usethe◄►buttonstomovethroughthem.

Page 1: Parameter

The symbol means start.

ThisparameterisusedtoconfiguretheInrush mode.

Figure 4-3Display of start threshold current.

The parameter sets theeffectivehalf-cycle current value, servingas themotorstartthreshold.When themotor starting current reaches or exceeds this threshold, theModel8220willcountthetimeduringwhichtheeffectivehalf-cyclecurrentvalueisstrictlyinexcessoftheeffectivehalf-cycleendvalue(seeFigure4-14).ThevalueissetusingtheWhiteandYellow( )buttons.Theminimumandmaximumvaluesare0and5999A.

Page 2: ParameterThe symbol means hysteresis.

ThisparameterisusedtoconfiguretheInrush mode.

The parametersetstheeffectivehalf-cyclecurrentvalueservingastheendofmotorstartthreshold.Assoonasthemotorstartingcurrentisoverorequaltotheeffectivehalf-cyclecurrentend(stop)value,theModel8220willstopcountingthestartingtime(seeFigure4-13).

Page 33: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 31

ThevalueissetusingtheWhiteandYellow( )buttons.Thepresetvaluesare0,1,2,5and10%.

NOTE: The effective half-cycle stopping value is calculated using the following formula: [effective half-cycle stopping value] = [effective half-cycle starting value] × (100 – [hysteresis]) ÷ 100.

Page 3: Parameter

The symbol means primary.

Thisparameterconfigurestheprimarycurrentofthetransformationratio( ).

NOTE: This screen is only displayed if the Model 8220 is connected to a 5A adapter or to the MN193 5A clamp. For all of the other sensors, which do not require any adjustment, this screen is not displayed:

Figure 4-4Display of the effective primary current.

The parameterdefinesthevalueoftheprimaryeffectivecurrentfortheMN193clamp(5A)orthe5Aadapter.ThevalueissetusingtheWhiteandYellow( )buttons.Theminimumandmaximumterminalsare0and2999A.

Page 34: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

32 Power Quality Analyzer Model 8220

Page 4: Parameter

The symbol means secondary.

Thisparameterconfigurestheprimarycurrentofthetransformationratio( ).

NOTE: This screen is only displayed if the Model 8220 is connected to a 5A adapter or to the MN193 5A clamp. For all of the other sensors, which do not require any adjustment, this screen is not displayed:

Figure 4-5Display of the effective secondary current.

TheparameterdefinesthevalueofthesecondaryeffectivecurrentfortheMN193clamp(5A)orthe5Aadapter.

ThevalueissetusingtheWhiteandYellow( )buttons.Thepresetvaluesare 1 and 5A.

ThisparameterconfigurestherangeoftheModel8220.

This SL261 screen is displayed only if the Model 8220 is connected to an SL261 probe (AC/DC)

Figure 4-6Display of the parameter concerning the sensitivity used by the probe

Page 35: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 33

Figure 4-7Display of the parameter concerning the sensitivity used on the clamp

TherangedisplayedonthescreenoftheModel8220mustbeconsistentwiththesensitivityusedontheSL261probe10Aforthe10mV/Asensitivity(Figure4-6)and100Aforthe100mV/Asensitivity(Figure4-7).

ThevalueisselectedusingtheWhiteandYellow( )buttons.Thepresetvaluesare10and100(10mV/Aand100mV/A).

Page 5: Parameter

The symbol means event per rotation.

Thisparameterisdesignedtoconfiguretherotationspeedmode( ).The parameterdefinesthenumberofeventsperrotationforthemeasure-mentofrotationspeedforamachineinoperation.

Example: If a tachometer signal provides two pulses per revolutions, this parameter would be set to 2.

ThevalueissetusingtheWhiteandYellow( )buttons.Theminimumandmaximumvaluesare1and99.

Page 6: ParameterThe symbol means threshold.

Thisparameterisdesignedtoconfiguretherotationspeedmode( ).

Page 36: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

34 Power Quality Analyzer Model 8220

Figure 4-8Display of the threshold voltage of the tachometer sensor.

The parameterdefinesthethresholdvoltagevalueusedtodetectanevent(pulseonthetachometersignal).SincethesignalreceivedbytheModel8220canbeunipolarorbipolar,twotypesofthreshold(0.3and1.1V)canbeselected.Therecommendedselectionisasfollows:

• bipolarsignals:0.3Vthreshold• unipolarsignals:1.1Vthreshold

Inbothcases,thehysteresisis0.2V.ThevalueissetusingtheWhiteandYellow( )buttons.Seebelowforthegraphsforthishysteresis.

0.2

0 0.9 1.1

Event

No eventTachometersignal (V)

0.2

0 0.1 0.3

Event

No eventTachometersignal (V)

Figure 4-9

Page 37: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 35

4.6 Rotary Switch ModesForallrotarypositionsthathavetwofunctions,thefollowingapplies:

To access the second function of a rotary position, turn the switch to the desired position then press the Yellow button.

To return to the first function in that rotary position, press the White button.

4.6.1 Voltage and Current

NOTE: The following illustrations are generic examples. Screen values and appearances will vary.

First Function: VOLTAGE MODE

Thereare4pages(screendisplays)inthismode.Usethe◄►buttonstomovethroughthem.

1

2

3

4

Figure 4-10

ITEM 1st Page 2nd Page 3rd Page 4th Page1. Root mean square

value (Vrms)Root mean square value (Vrms)

Effective maximum voltage half-cycle value (Vrms 1/2 max)*

Maximum voltage crest value (VPEAK+)

2. Root mean square value (Arms)

Continuous voltage value (VDC)

Root mean square value (Vrms)

Minimum voltage crest value (VPEAK-)

3. Network frequency (Hz)

Voltage crest factor (VCF)

Effective minimum voltage half-cycle value (Vrms 1/2 max)*

4. # of page displayed / total # of pages

*The effective maximum and minimum half-cycle values may be reinitialized by pressing the Delete button.

Page 38: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

36 Power Quality Analyzer Model 8220

Second Function: CURRENT MODE

Thereare4pages(screendisplays)inthismode.Usethe◄►buttonstomovethroughthem.

1

2

3

4

Figure 4-11

ITEM 1st Page 2nd Page 3rd Page 4th Page1. Root mean square

value (Arms)Root mean square value (Arms)

Effective maximum current half-cycle value (Vrms 1/2max)*

Maximum cur-rentcrest value (VPEAK+)

2. Root mean square value (Vrms)

Continuous current value (ADC) for the MR clamps only.

Root mean square value (Arms)

Minimum cur-rent crest value (VPEAK-)

3. Network frequency (Hz)

Current crest factor (ACF)

Effective minimum minimum half-cycle value (Vrms 1/2max)*

Current K factor (AKF)

4. # of page displayed / total # of pages

*The effective maximum and minimum half-cycle values may be reinitialized by pressing the Delete button.

Page 39: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 37

4.6.2 Power and Three-Phase Calculations

First Function: POWER MODE

Thereare12pagesinthismode.Usethe◄►buttonstomovethroughthem.

1st and 2nd Pages:

1

2

3

4

Figure 4-12

ITEM 1st Page* 2nd Page**1. Active power (W) Continuous power (WDC)2. Reactive power (VAR) Power factor (PF)3. Apparent power (VA) Displacement factor (DPF, also noted as Φ)4. # of page displayed / total # of pages

*The powers displayed are the total powers (sum of 3 phases) if the symbol is displayed.

**The total continuous power (WDC) is not shown if the symbol is displayed.

Page 40: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

38 Power Quality Analyzer Model 8220

4.6.2.1 Energy Measurement Functions (counting)Theenergycountersstartandcountthetotalsforthevarioustypesofenergy(theeight energy counters–4 consumedenergy countersand4generatedenergycounters-arestarted)Energycountingisnotaffectedby:

• Takingphotos• RetrievalofphotosbythePCforviewingwith“PowerAnalyzerTransfer”• Modbusreal-timeseriallinkwith“PowerAnalyzerTransfer”

3rd Page:

Figure 4-13

Counterstatusesare:• On<=>counteroperative• Stop<=>countingstopped(countervalues)• Off<=>counternotoperative(countervaluesreading0)

Iftheinstrumentisnotin“viewaphoto”modewhenpage3/12isinPowerMode, isdisplayed.

• ThebuttongeneratesaswitchfromOFFtoON• Thebuttoneithergeneratesaswitch fromONtoStopor fromStop to

OFF.ThecausesforanautomaticswitchfromONtoStop:

• Currentsensorremoved• Rotaryswitchonapositionotherthan , or • Viewingofaphoto

Page 41: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 39

4th Page:

1

2

3

4

Figure 4-14

ITEM 4th Page1. Number of hours (h)2. Number of minutes (n)3. Number of seconds (s)4. # of page displayed / total # of pages

NOTE: Beyond999h,59m,59s,“---h--m--s”isdisplayed,buttheinternalcount-ingtimecontinuestofunctioncorrectly.

Pages 5/12 to 12/12:

Pages5,6,7and8concernpowerreceivedbythe“Load”side.

Pages9,10,11and12concernpowergeneratedbytheloadsideandthereforereceivedbythe“Supply”side.

1

2

Figure 4-15

Page 42: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

40 Power Quality Analyzer Model 8220

ITEM 5th Page1. Load or supply (SuPP)2. # of page displayed / total # of pages

Powerisdisplayedinthefollowingformats: - [000.1;999.9] - [1.000k;9999k] - [10.0M;999M] - [1.00G;999G]

Beyond999999999999xh(999Gxh)“----”isdisplayedbuttheinternalcounterscontinuetofunctioncorrectly.Theprecisionoftheinternalcountersisgreaterthanthatofthepowerdisplayedontheinstrument(thisisduetodisplaylimitations–numberofdigitsavailable).

Pages6and10outof12concerninductivereactiveenergy“L”.

Pages7and11outof12concerncapacitivereactiveenergy“C”.

Second Function: THREE-PHASE CALCULATIONS

Wheninviewmode,thedisplayindicates or .

: ThecalculationsassociatedwiththeconnectionoftheModel8220toathree-phase balanced network are deactivated.This choice is selectedwhenmeasuringsingle-phasenetworks.

: ThecalculationsassociatedwiththeconnectionoftheModel8220toathree-phasebalancednetworkareactivated.Thischoiceisselectedwhenmeasuringthree-phasebalancednetworks.

Thechoiceismadeusingthe◄ and ►buttons.

Page 43: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 41

4.6.3 Voltage and Current Harmonics

Thismeasuresthetotalharmonicdistortionrateonvoltageandcurrent,aswellastheeffectivevalue,theharmonicdistortionfactor,anycontinuousharmoniccom-ponentandharmonicsuptothe50th.

First Function: VOLTAGE HARMONIC MODE

Thereare52measurementpages in thismode.Use the◄►buttons tomovethroughthem.

1

2

3

4

Figure 4-16

ITEM 1st Page 2nd Page 3rd Page1. Rate of total voltage

harmonic distortion (VTHD-F, also noted as VTHD)

Number of voltage harmonic considered

Number of voltage harmonic considered

2. Root mean square value (Vrms)

The continuous harmonic component value

Effective value of the harmonic component considered

3. Voltage distortion factor (VTHD-R, also noted as VDF)

Percentage of the continu-ous value compared with the effective fundamental value

Percentage of this effective value compared with the effective fundamental value

4. # of page displayed / total # of pages

Page 44: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

42 Power Quality Analyzer Model 8220

Second Function: CURRENT HARMONIC MODE

Thereare52measurementpagesinthismodewhenanMRprobeisconnectedand51measurementpagesforalltheothercurrentprobes.

1

2

3

4

Figure 4-17

ITEM 1st Page 2nd Page 3rd Page1. Rate of total current

harmonic distortion (ATHD-F, also noted as ATHD)

Number of current harmonic considered

Number of current harmonic considered

2. Root mean square value (Arms)

The continuous harmonic component value

Effective value of the harmonic component considered

3. Current distortion factor (ATHD-R, also noted as ADF)

*Percentage of the continuous value compared with the effective fundamental value

**Percentage of this effective value compared with the effective fundamental value. In this example, the fundamental is 100% of itself

4. # of page displayed / total # of pages

*With an MR current probe connected

**Without an MR current probe connected

Page 45: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 43

4.6.4 Inrush and Phase Rotation

First Function: INRUSH MODEThismoderequirestheModel8220tobepre-configured(referto§4.5).Theleads(voltageandcurrent)canbeconnectedeitherinsingle-phaseorbalancedthree-phasemode.Duringsteps2,3and4,itispossibletogobacktothefirststep(step1/4)bypressingthe◄button.

Step 1:

• Turn the rotary switch to the position.Thedisplaywill indicate (ready).

• Pressthe►buttontomovetoStep2.

Step 2:• Themotortobemonitoredshouldnowbeswitchedon.TheModel8220waits

until the effective half-cycle current exceeds the current threshold for theconfiguredmotorstart.

• Theeffectivecurrentcalculatedoveronesecondiscontinuouslydisplayed.TheModel8220automaticallymovesontoStep3.

Step 3:• Assoonasthemotorstartingthresholdisreached,thetimerisstarted(Fig.

4-18,No.2).• Once themotor tobemonitored is running, theModel8220waitsuntil the

effectivehalf-cyclecurrentgoesbelowthecurrentsetthreshold(endofstartthreshold).Theeffectivecurrentcalculatedoveronesecondiscontinuouslydisplayed(Fig.4-18,No.1).TheunitautomaticallymovesontoStep4.

1

2

Figure 4-18The timer (bottom value) is active until the low current threshold has been reached.

Page 46: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

44 Power Quality Analyzer Model 8220

Step 4:• Assoonastheendofmotorstartingthresholdhasbeenreached,theresults

aredisplayed.

1

2

3

Figure 4-19

ITEM MEASUREMENT1. Effective maximum half-cycle value of starting current.2. Absolute instantaneous value of starting current.3. Duration of motor starting (in seconds).

Second Function: PHASE ROTATION MODE

Thismodeenablestheusertodeterminetheorderofthephasesofathree-phasenetworkusingthemethodknownas“2wire”.

NOTE: The activation of the balanced three-phase mode has no effect on deter-mining the order of the phases ( can be or ). The most important point is to respect the voltage connections as directed in 1 and 3.

Step 1:

• Turntherotaryswitchtothe positionthenpresstheYellowbutton.ThedisplaywillindicatethattheModel8220isready.

• OncethetestleadsareconnectedtothesuspectedL1andL2phases,pressthe►button.

• The messageisdisplayedbriefly.

Page 47: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 45

Step 2:• Thedisplaywillshowthefollowingscreen.

Figure 4-18

• TheusershouldthenconnecttheL3phasetotheinput(+)ofthe8220,within10 seconds, afterwhich time the following errormessage (timelimitexceeded) isdisplayed.Theusermuststart thesequenceagainfromthebeginning(Step1).

• Assoonasthetestleadsareconnectedtothecircuit,the8220automaticallymovestoStep3.

NOTE: If the display indicates (Error) an error has been detected during the determination of the order of the phases.

This error may be due to one of the following causes:

- The three-phase network frequency is unstable.- The three-phase network frequency is outside the 40 to 70Hz range.- The voltage signal is too weak (below 10VRMS).- Operations were incorrectly carried out.

To return to the first Step, press the◄button.

Step 3:• Thedisplaywillthenshoweither:

-Correctorderofphases:L1-L2-L3-Incorrectorderofphases:L3-L2-L1

Page 48: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

46 Power Quality Analyzer Model 8220

4.6.5 Rotation Speed (RPM)

ThismoderequirestheModel8220tobepre-configured(referto§4.5).

In this position, theModel 8220measures the rotation speed of a turning ele-ment.

NOTE: The tachometer signal should be applied into the voltage (+) and (COM) terminals of the Model 8220. The unit then measures the interval of time be-tween each signal pulse (event) and subtracts the rotation speed in revolutions per minute.

4.6.6 Temperature and Resistance

In this position, theModel 8220measures the temperature (using an externalprobenotprovided)ortheresistanceofanelectricalcircuit.

First Function: TEMPERATURE MODETheModel8220displaysthetemperaturemeasuredusinga100platinumprobe(notprovided),connected to the(+)and(COM) terminals.Themeasurement isdisplayedsimultaneouslyindegreesCelsius(°C)anddegreesFahrenheit(°F).

Second Function: RESISTANCE MODETwoautomaticrangesareavailable:• 0 to 400Ω:under20Ω(bydefault)abeepingsoundisemitted• 400 to 2000Ω

Page 49: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 47

CHAPTER 5

DATAVIEW® SOFTWARE

5.1 Installing DataView®

DO NOT CONNECT THE INSTRUMENT TO THE PC BEFORE INSTALLING THE SOFTWARE AND DRIVERS.

NOTE:When installing, the user must haveAdministrative access rightsduring the installation.Theusers access rights canbe changedafter theinstallation is complete.DataView®mustbe reinstalled for eachuser inamulti-usersystem.

1. Insert theDataView thumb drive into an availableUSB port on your com-puter. IfAutorun is enabled, anAutoPlaywindowappears on your screen.Click“Openfoldertoviewfiles”todisplaytheDataViewfolder.IfAutorunisnotenabledorallowed,useWindowsExplorertolocateandopentheUSBdrivelabeled“DataView.”

2. WhentheDataViewfolderisopen,findthefileSetup.exelocatedintherootdirectoryoftheUSBdrive,anddouble-clickittoruntheinstallationprogram.

3. TheDataViewsetupscreenappears.Intheupperleftcornerofthescreen,choosethelanguageversionoftheSetupinterface.(AllSetupscreensanddialogswillimmediatelyappearintheselectedlanguage.)Inthelowerleftcorneraretheavailableinstallationoptions.InadditiontotheDataViewsoftware,youcanselect“AdobeReader.”ThislinkstotheAdobewebsitewhereyoucandownloadthelatestversionofReader.ThisprogramisrequiredtoviewDataView.pdfdocuments.TheoptionFirmwareUpgradeslinksto thewebsitewhereyoucancheckfornewfirmwareupdatesfor theinstrument.Finally,UserManualsdisplaysalistof.pdffilescontainedintheUSBdrive thataccompaniesDataView. (DataViewalsocomeswithaHelpsystemthatisinstalledwiththeprogramfiles.)ToinstallDataView,selectDataViewintheOptionslistandclickInstall.

4. SelectthelanguageversionofDataViewyouwanttoinstall(English,French,orSpanish) thenclickNext. (Bydefault, the languageselected instep3 ishighlighted.)

Page 50: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

48 Power Quality Analyzer Model 8220

5. You are now prompted to select the software you want to install. EachAEMCproduct family has its own specially designedControlPanel. If youareperformingaCompleteinstall,bydefaultallavailableControlPanelsareselected (a checkmark next to theControl Panel indicates it is selected).Control Panels take up disk space on the computer; so unless you haveother types of AEMC instruments, we recommend that you select Pow-erPad and deselect the rest. You should also check the option Data-ViewCore,which is a requirement if youplan to createDataView reports.

After you finish selecting and deselectingControl Panels and/orDataViewCore,clickNext.

6. TheSetupprogramnowinformsyouthatitisreadytoinstallDataView.Ifyouwanttoreviewanyofyourpreviousselections,clickthePreviousbuttontoreturntoearlierscreens.Otherwise,clickInstalltobegininstallation.

7. TheInstallShieldprograminstallstheselectedsoftware.Ifanearlierversionofthesoftwareisalreadyinstalledonyourcomputer,foreachselectedpro-gramtheInstallShieldwill:

(a) Askyoutoconfirmtheinstallationoftheprogram.ClickNext.

(b) Displayastatusbarindicatingtheprogressoftheinstallation.

(c) Informyouwhentheprogramisinstalled.ClickFinishtoinstall thenextselectedprogram.

Ifthesoftwareisnotinstalled(oriftheinstalledsoftwareisthesameversionastheselectedsoftware),thesoftwareisinstalledwithoutrequestingconfirmation.

Whenallprogramsare installed,amessageappears informingyouof this.ClickFinishtoreturntotheSetupscreen

8. You cannowselect additionalSetupoptions to install (see step3 above).Whenfinished,clickExit.

9. TheDataViewfoldernowappearsonyourcomputerdesktop,withinwhichisthePowerPad iconandtheicon(s)foranyotherControlPanel(s)youhaveinstalled.

Page 51: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 49

5.2 PowerPad Control PanelClickingtheDataViewiconintheDataViewfolderonyourdesktopopensthecoreDataViewprogram.ClickingthePowerPadControlPaneliconopensthePower-PadControlPanel.

Ingeneral, coreDataView featuresare for creating, viewing,editing, andstoringDataViewreports;whiletheControlPanelisforconnectingto,configuring,viewingmeasurementson,anddownloadingdatafromtheinstrument.YoucanaccessallDataViewfeaturesthrougheithertheDataViewiconortheControlPanelicon.

ForuserswhointeractwithPowerPadseriesinstruments,werecommendprimarilyusingtheControlPanel.However,therearesituationswhereusingthecoreData-Viewiconmaybemoreconvenientforsomeusers,suchaswhenviewingmultiplearchivedreportsfromdifferentAEMCproductfamilies.

ForfurtherinformationaboutusingthePowerPadControlPanel,consulttheHelpsystemthatcomeswiththeproduct.AccessthisHelpbyclickingtheoptionHelpintheControlPanel’smenubaratthetopofthescreen.

Page 52: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

50 Power Quality Analyzer Model 8220

CHAPTER 6

MAINTENANCE

Useonlyfactoryspecifiedreplacementparts.AEMC®willnotbeheldresponsibleforanyaccident,incident,ormalfunctionfollowingarepairdoneotherthanbyitsservicecenterorbyanapprovedrepaircenter.

6.1 Changing the Batteries

WARNING:• When changing the battery, disconnect the instrument from the AC power and

any other inputs and turn the equipment off. There must be a delay of at leastone minute without the battery being connected.

• Do not expose the battery to heat exceeding 212°F (100°C).

• Do not short-circuit the battery terminals.

TheModel8220ispoweredbysixAA(LR6-NEDA15A)batteries(standardorrechargeable).

Toaccessthebatteries,turntheinstrumentoverandturnthelockone-quarterturn(Figure6-1,item2)counter-clockwiseusingacoin(Figure6-1,item3).

Figure 6-1

Page 53: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 51

6.2 Recharging the Rechargeable BatteriesTheModel8220maybepoweredbysixrechargeableNiMHbatteries(availableasanoption)havingacapacityofatleast1800mAh.

Theinstrumentdoesnotrechargeitsownrechargeablebatteries.Thesemustbechargedonanexternalchargerafterremovingthebatteriesfromtheunit.

6.3 Cleaning

Disconnect the instrument from any source of electricity.

• Useasoftcloth,lightlydampenedwithsoapywater

• Wipewithadampclothandthendrywithadrycloth

• Donotsplashwaterdirectlyontheclamp

• Donotusealcohol,solventsorhydrocarbons

Page 54: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

52 Power Quality Analyzer Model 8220

APPENDIX A

Mathematical Formulas for Various Parameters

NOTE the following abbreviations used in this section:

NSHC = numberofsamplesperhalfcycle(betweentwoconsecutivezeros)NSC = numberofsamplespercycleNSS = numberofsamplesinasecond(multipleofNSC)V = voltagephasetoneutralU = voltagephasetophase

Network FrequencyThesamplingislockedtothenetworkfrequencysoastoprovide256samplesperperiod(NECHPER)from40to70Hz.Thislockingisessentialforthecalculationsofthereactivepowers,thecalculationsoflevelsandangles,andthecalculationsgivingtheharmonicmagnitudes.

Half-period Voltage and Current RMS Values

[ ]1Vhalf

:

2∑⋅=(Next Zero)-1

ZeronnV

NSHC

[ ]1Ahalf

:

2∑⋅=(Next Zero)-1

ZeronnA

NSHC

Rms voltage

Rms current

n:sample(0;255)

MIN / MAX Values for Voltage and Current

Vmax=max (Vhalf), Vmin=min (Vhalf)

Amax=max (Ahalf), Amin=min (Ahalf)

Page 55: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 53

Peak Voltage and Current (Updated on each waveform refresh)

Vpp=max (V[n]), Vpm=min (V[n]), n ∈ [0..NSC-1]

App=max (A[n]), Apm=min (V[n]), n ∈ [0..NSC-1]

Peak Factors for Current and Voltage (over 1 second)

[ ]∑−

=

=1

0

212

2

)Vpm,max (VppVcfNSC

nnV

NSC

[ ]∑−

=

=1

0

21)Apm,max (AppAcf

NSC

nnA

NSC

Peak factor voltage

Peak factor current

1 sec RMS Values for Voltage and Current

[ ]∑−

=

⋅= 1

0

21Vrms NSS

nnV

NSS

[ ]∑−

=

⋅= 1

0

21Arms NSS

nnA

NSS

Rms voltage

Rms current

Page 56: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

54 Power Quality Analyzer Model 8220

Harmonic Calculations

Thesearedoneby1024-pointFFT(4periods)withoutwindowing(cf.IEC1000-4-7).Fromtherealandimaginaryparts,thelevelsVharmandAharmarecalculated(withrespecttotheRMSvalueofthefundamental)andtheanglesVphandApharecalculated(phaseshiftwithrespecttothefundamental).

[ ]

[ ]

[ ]

[ ]1Athd , 1

Vthd

50

2

250

2

2

Aharm

nAharm

Vharm

nVharmnn∑∑== ==

The voltageharmonic level (Vharm) ismultiplied by the current harmonic level(Aharm)tocalculatetheapparentpowerharmoniclevel(VAharm).Thepowerhar-monicanglesarecalculatedfromthedifferencebetweenvoltageharmonicanglesandthecurrentharmonicangles.

Different Power Levels 1 Sec in Single-Phase Connection

[ ]W

VA = Vrms Arms

VAR =

or VAR = VA – W if computation method is with harmonics

Active powerNSS-1

0nV

n

.

.

.

∑=

[ ]nA=1

NSS

[ ]Reactive powerNSS-1

0n - NSC / 4VF

n

.∑=

[ ]nAF1NSS

22

Apparent power

Different Power Levels 1 Sec in Balanced Three-Phase Connection

or

Total Active power

Total Reactive power

Total Apparent power

NSS-1

∑=

−×

=

0 n ][].4/[

33- W nANSCnUNSS

RMSRMS AU ⋅⋅=3

3 VA

²² WVAVAR −=

∑=×

=NSS-1

0 n ][].[

33 VAR nAFnUFNSS

if computation methodis without harmonics

U = voltagephasetophasebetweenphase1tophase2A = phase3

Page 57: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 55

Ratios

PF = phase power factorWVA

DPF = cos(φ) phase displacement factor

Tan = tan(φ) phase tangent

cos(φ) =

[ ]Cosine angle between voltagefundamental and phase current

NSS-1

0

2

nVF [ ]nAFn∑=

[ ]NSS-1

0nVF

n∑=

2[ ]NSS-1

0nAF

n∑=

.

K Factor[ ]

[ ]∑

∑=

=

=

=

50

1

2

50

1

22

= Akf n

n

n

n

nAharm

nAharmn

Various Types of Energy

Active energy consumed0for W 3600

WhcTint

≥=∑ W

Active energy generated0for W <

0for W <

Apparent energy consumed

Apparent energy generated

0for W ≥

3600

WhgTint

−= ∑ W

3600

VAhcTint

=∑ VA

3600

VAhgTint

= ∑ VA

≥ Inductive reactive energy consumedfor VAR 0 and W 0

Inductive reactive energy generatedfor VAR < 0 and W < 0

≥ Capacitive reactive energy consumedfor VAR < 0 and W 0

Capacitive reactive energy generatedfor VAR 0 and W < 0

3600VARhLc

Tint= ∑ VAR

3600

VARhCcTint

−= ∑ VAR

3600

VARhLgTint

−= ∑ VAR

3600VARhCg

Tint= ∑ VAR

Page 58: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

56 Power Quality Analyzer Model 8220

HysteresisHysteresis is a filtering principle, often used after the threshold detection hasoccurred.Acorrectsettingofhysteresisvaluewillavoidrepeatedtriggeringwhenthemeasureisvaryingclosetothethreshold.Theeventdetectionisactivatedwhenthemeasureisgoingoverthethresholdbutitcanonlybedeactivatedifthemeasuregoesunderthethresholdminusthevalueofthehysteresis.Thedefaulthysteresisvalueis2%ofthereferencevoltagebutitmaybesetintherangeof[1%,5%]dependingofthevoltagestabilityonthesystem.

Detection of voltage overload

Hysteresis =2% Uref

Swell duration

Threshold of swell =100% Uref

Level to go back =100% - 2% = 98% Uref

Undervoltage or blackout detection

Hysteresis =2% Uref

Level to go back =(100% + 2%)Uref

Threshold = 102% Uref

Duration

Page 59: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 57

Diagram of 4 QuadrantsRepresentationofthe4powerquadrantsusedasapartofthemeasurementofpower .

-W +W

+VAR

-VARProduced Consumed

12

43Reactive Power - from load

Reactive Power - from supply

Saturation of Input ChannelsMonitoringofthesaturationofinputchannelsiscarriedoutwhentheinstrumentisinphotographviewingmodeorinthefollowingmodes:

No monitoring ofthesaturationofinputchannelsiscarriedoutwhentheinstru-mentisininformationdisplaymodeorinthefollowingmodes:

Saturation indicator forvoltage input channel

Saturation indicator forcurrent input channel

NOTE: It is normal for the above screen to be displayed when the current sensor is installed or removed.

Page 60: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

58 Power Quality Analyzer Model 8220

Repair and Calibration

Toensurethatyourinstrumentmeetsfactoryspecifications,werecommendthatitbescheduledbacktoourfactoryServiceCenteratone-yearintervalsforrecalibration,orasrequiredbyotherstandardsorinternalprocedures.

For instrument repair and calibration:Youmust contactourServiceCenter foraCustomerServiceAuthorizationNumber(CSA#).Thiswillensurethatwhenyourinstrumentarrives,itwillbetrackedandprocessedpromptly.PleasewritetheCSA#ontheoutsideoftheshipping container. If the instrument is returned for calibration,weneed toknowifyouwantastandardcalibration,oracalibrationtraceabletoN.I.S.T.(Includescalibrationcertificateplusrecordedcalibrationdata).

Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 15FaradayDrive Dover,NH03820USA Phone:(800)945-2362(Ext.360) (603)749-6434(Ext.360) Fax: (603)742-2346or(603)749-6309 E-mail:[email protected]

(Orcontactyourauthorizeddistributor)Costsforrepair,standardcalibration,andcalibrationtraceabletoN.I.S.T.areavailable.NOTE: You must obtain a CSA# before returning any instrument.

Technical and Sales Assistance

Ifyouareexperiencinganytechnicalproblems,orrequireanyassistancewiththeproperoperationorapplicationofyourinstrument,pleasecall,mail,faxore-mailourtechnicalsupportteam:

ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 200FoxboroughBoulevard Foxborough,MA02035USA Phone:(800)343-1391 (508)698-2115 Fax: (508)698-2118 E-mail:[email protected] www.aemc.com

NOTE: Do not ship Instruments to our Foxborough, MA address.

Page 61: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Power Quality Analyzer Model 8220 59

Limited Warranty

TheModel8220iswarrantedtotheownerforaperiodoftwoyearsfromthedateoforiginalpurchaseagainstdefectsinmanufacture.ThislimitedwarrantyisgivenbyAEMC®Instruments,notbythedistributorfromwhomitwaspur-chased.Thiswarrantyisvoidiftheunithasbeentamperedwith,abusedorifthedefectisrelatedtoservicenotperformedbyAEMC®Instruments.

Full warranty coverage and product registration is available on our website at www.aemc.com/warranty.html.

Please print the online Warranty Coverage Information for your records.

What AEMC® Instruments will do:Ifamalfunctionoccurswithinthewarrantyperiod,youmayreturntheinstrumenttousforrepair,providedwehaveyourwarrantyregistrationinformationonfileoraproofofpurchase.AEMC®Instrumentswill,atitsoption,repairorreplacethefaultymaterial.

REGISTER ONLINE AT:www.aemc.com

Warranty Repairs

What you must do to return an Instrument for Warranty Repair: First, request aCustomerServiceAuthorizationNumber (CSA#) by phoneorbyfaxfromourServiceDepartment(seeaddressbelow),thenreturntheinstrumentalongwith thesignedCSAForm.Pleasewrite theCSA#on theoutsideoftheshippingcontainer.Returntheinstrument,postageorshipmentpre-paidto:

Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments15FaradayDrive•Dover,NH03820USAPhone:(800)945-2362(Ext.360)

(603) 749-6434(Ext.360)Fax: (603)742-2346or(603)749-6309E-mail:[email protected]

Caution:Toprotectyourselfagainstin-transitloss,werecommendyouinsureyourreturnedmaterial.

NOTE: You must obtain a CSA# before returning any instrument.

Page 62: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

Notes:

Page 63: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric
Page 64: SINGLE-PHASE POWER QUALITY ANALYZER 8220 · services. Refer to our repair and calibration section at . Serial #: ... product must undergo selective disposal for the recycling of electric

11/19

99-MAN100297v21

Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments15FaradayDrive•Dover,NH03820USA•Phone:(603)749-6434•Fax:(603)742-2346

www.aemc.com