8/12/2019 Siliconic Oil Sat1
http://slidepdf.com/reader/full/siliconic-oil-sat1 1/12
Typology of parts to use for saturation with siliconic oil
The saturation of pore filter is an important step to do before starting a CPTUtest. Pagani Company has a complete kit which is necessary for the saturation
procedure . The following descriptions are made by using a lot of pictures, thisway of explaining make more easy for me to explain and for you to understand I
hope.
1)Mount of mechanical parts tripod of probe The picture1 shows the parts used to mount the tripod of probe(piezocone).
picture1: Mechanical parts - tripod of probeStart mounting by insertion of feet into the tripod main base so fix those firmly
by using a hexagonal screwdriver. The assembling of those part is easy, notice
that tripod support part 2 has a square cross section and it is smaller than part 1so is easy to insert into the part 1.
Now turn the top down of the feet with the mounted base and insert the
“body” of the tr ipod as it shows on the picture2. In the picture 2 is alsoexplained the knobs and their functionality. The tripod of probe is complete
and in the following issue will be explain the probe with the oil kit.
8/12/2019 Siliconic Oil Sat1
http://slidepdf.com/reader/full/siliconic-oil-sat1 2/12
Picture2: Tripod of probe complete
2)Probe and siliconic oil kit
The probe(piezocone) when is sold from Pagani Company has a Not Saturated
pore filter(it is a dummy filter) which is used to keep all the parts of probe
together. Please DO NOT USE this pore filter.
Use the saturated once which are in a plastic bottle into siliconic oil. The
picture4 shows these parts more accurately.
8/12/2019 Siliconic Oil Sat1
http://slidepdf.com/reader/full/siliconic-oil-sat1 3/12
Picture3:Probe and siliconic oil kit
The different parts shown in picture3 will be explained step by step into this quick
reference. Now take the probe and unscrew the head of the probe so you can take off
the non saturated filter. Fix the probe without the 60 degree head into the tripod and
use the knobs to make adjustment of the different parts of the tripod. The picture4show this situation better.
8/12/2019 Siliconic Oil Sat1
http://slidepdf.com/reader/full/siliconic-oil-sat1 4/12
Picture4:Probe fixed to tripod
The second step is to fix the Orange box inside the probe and fill the box with
siliconic oil as it shows in the picture5.
Do the following steps in the following order(do not change order )
8/12/2019 Siliconic Oil Sat1
http://slidepdf.com/reader/full/siliconic-oil-sat1 5/12
a) Insert the orange box and Fill it with siliconic oil
Picture5:fill with siliconic oil most of it
b)Insert the saturated pore filter inside of the orange box quickly
Pay attention, by saying quickly I mean that take of the pore filter from their
bottle and put inside of orange box without losing time otherwise the filter will start to lose saturation.
Picture6:insert the saturated pore filter into the orange box
8/12/2019 Siliconic Oil Sat1
http://slidepdf.com/reader/full/siliconic-oil-sat1 6/12
c)Insert the X-Ring with his support, the 60 degree header and let the
parts for few minutes.
Time necessary to take the bubbles of air away from oil. Use the
syringe to help to take away all the bubbles of air . .
Picture7:insert the X-Ring and 60 degree header into the orange box
e
Picture8:All the parts inside of siliconic oil
8/12/2019 Siliconic Oil Sat1
http://slidepdf.com/reader/full/siliconic-oil-sat1 7/12
d) Screw in all the parts by paying attention that no part goes out of
the siliconic oil.
Picture9screw in all the parts
Consider the right order and orientation of mounting.
e)Mount all the parts under siliconic oil
Picture10:all the parts screw into Piezocone
8/12/2019 Siliconic Oil Sat1
http://slidepdf.com/reader/full/siliconic-oil-sat1 8/12
We have to let all the system in quite for few minutes until there are no more air
bubbles coming out of piezocono.
Thus is necessary to put the condom into the piezocone. This step is made by
moving down carefully the orange box as is shown on the picture 11.
picture 11: Condom into piezocone
8/12/2019 Siliconic Oil Sat1
http://slidepdf.com/reader/full/siliconic-oil-sat1 9/12
The condom is used to prevent the oil leakage during transportation from
saturation place to Test CPTU place. Moreover the siliconic oil is reusable (if
you keep it clean from soil parts or materials which can contaminate it ).
You can spill the oil out of orange box and refill the siliconic oil plastic bottle.
Th e procedure is shown on picture 12 and picture 13.
picture 12: use a funnel device to not lose siliconic oil.
8/12/2019 Siliconic Oil Sat1
http://slidepdf.com/reader/full/siliconic-oil-sat1 10/12
picture 13: Empty the orange box
To take off the orange box from the piezocone, just move the orange box down
till the bottom of piezocone as you see from picture 14 and picture 15.
8/12/2019 Siliconic Oil Sat1
http://slidepdf.com/reader/full/siliconic-oil-sat1 11/12
picture 14 Move down slowly the orange box.
8/12/2019 Siliconic Oil Sat1
http://slidepdf.com/reader/full/siliconic-oil-sat1 12/12
figura 6: Take off the piezocone
The piezocone is ready to start a CPTU test.