Siesta i TS 10650E-90 UK Apr08

download Siesta i TS 10650E-90 UK Apr08

of 114

Transcript of Siesta i TS 10650E-90 UK Apr08

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    1/114

    User Manual

    Dameca a/s

    Islevdalvej 211

    DK-2610 Roedovre, Denmark

    Phone: +45 44 50 99 90, Telefax: +45 44 50 99 99E-mail: [email protected], www.dameca.com

    SIESTAiTSAnaesthesia Machine

    Software Version 3.0

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    2/114

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    3/114

    User Manual for SIESTA i TS

    Dameca I

    Conformity according to the

    Council Directive 93/42EEC

    concerning Medical Devices.

    Important !

    We are constantly developing our products, and

    consequently reserve the right to make alterations

    in design and equipment without prior notice.

    DAMECA a/s issues a guarantee against manu-

    facturing and material defects for a period of

    one year from the date of delivery. Excepted

    from this guarantee are rubber articles and

    anaesthesia valves with plastic parts.

    DAMECA a/s will repair or replace all defective

    components during the period of warranty, pro-

    vided (1) that the unit has been used correctly in

    accordance with the instructions for use and (2)

    that the instructions for maintenance have beenobserved carefully.

    Defective products or components must be

    returned to DAMECA a/s with a detailed

    description of the defect in question.

    If repairs are carried out by anyone else than

    DAMECA a/s or by a DAMECA a/s appointed

    agent, the guarantee no longer applies.

    Nobody but DAMECA a/s can issue any otherform for guarantee, either in writing or orally.

    DAMECA a/s has no responsibility for any

    translation of this English version of the manual.

    Preventive maintenance of this unit should

    be carried out at least once a year either by

    DAMECA a/s or by an authorized technician.

    Document No. 10650E-90

    April, 2008

    User Manual for Anaesthesia Machine

    Model SIESTA i TS

    Software Version 3.0

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    4/114

    User Manual for SIESTA i TS

    II Dameca

    Left blank intentionally

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    5/114

    User Manual for SIESTA i TS

    Dameca III

    Contents

    1. Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11.1 Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1

    1.2 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2

    1.3 Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3

    1.4 User Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4

    1.5 Responsibility of the Manufacturer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4

    1.6 Safety Features of the SIESTA i TS Anaesthesia Machine . . . . . . . . . . . . . . . . . . . . . . . 1-4

    2. Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1

    2.1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12.2 Cautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3

    2.3 Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4

    3. Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13.1 Anaesthesia Gas Scavenging System (AGSS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5

    4. Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1

    4.1 Set Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14.2 Function Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7

    4.3 Operation of Electronic Gas Mixer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16

    4.4 Operation of Stopwatch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18

    4.5 Operation of Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19

    5. Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15.1 Respiration Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1

    5.2 Gas Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3

    6. User Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16.1 Setup Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1

    6.2 Trend Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8

    6.3 Alarm Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11

    6.4 Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-15

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    6/114

    User Manual for SIESTA i TS

    IV Dameca

    Contents (continued)

    7. Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17.1 Integrated Breathing System Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1

    7.2 Electronic Gas Mixer Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1

    7.3 Ventilator Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3

    7.4 Multigas Module Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6

    7.5 System Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9

    7.6 Overview of Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-11

    8. Maintenance and Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1

    8.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-18.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1

    8.3 IBS Patient Breathing System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1

    8.4 CO2 Absorber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2

    8.5 Hoses, Y-piece and Manual Breathing Bag . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2

    8.6 Cleaning Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3

    8.7 After Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3

    8.8 Safety Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5

    8.9 Parts for the CO2 Absorber . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5

    8.10 Parts for the IBS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6

    9. Tehcnical Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-19.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1

    9.2 Gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2

    9.3 Electronic Gas Mixer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2

    9.4 Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3

    9.5 Integrated Breathing System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5

    9.6 Multigas Module (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5

    9.7 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9

    9.8 Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9

    Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1EMC Declaration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-1

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    7/114

    User Manual for SIESTA i TS

    Dameca 1-1

    1. Introduction

    This manual is related to EGM software

    version 3.0, ventilator software version2.0 and multigas module (optional) software

    version B7 or higher.

    Damecaanaesthesia machine model SIESTAi

    TS is applicable with most types of inhalation

    anaesthesia. O2 and N2O, or O2 and AIR can

    be administered, and anaesthetic vaporizer(s)

    is connected to a back-bar.

    The gas supply can be obtained from a centralgas supply system or from cylinders. There is

    no difference in performance when supplied

    from cylinders.

    The machine is equipped with:

    An electronic gas mixer (EGM) giving user

    selectable gas mixes (O2/N2O or O2/AIR) in

    a range of 0 20 L/MIN. If N2O is selected

    as a mix to O2, the N2O percentage islimited to max.75% (minimum 25% O2).

    A real time clock and a stopwatch is

    included in the electronic gas mixer.

    A microprocessor controlled anaesthesia

    ventilator with user selectable ventilation

    modes:

    - VCV Volume Controlled Ventilation

    - PCV Pressure Controlled Ventilation

    - SIMV Synchronised Intermittent

    Mandatory Ventilation

    - PSV Pressure Supported Ventilation

    An integrated breathing system (IBS)

    where Bag in Bottle and patient re-breathing

    system (incl. absorber) are integrated into

    one block. Thereby there are only 2 tubes

    (inspiratory and expiratory) from the IBS

    system to the Y-piece.

    An APL (Adjustable Pressure Limiting) valve

    is integrated into the IBS, giving the usera choice between manual ventilation or

    spontaneous breathing.

    A multigas module (optional) which measures

    respiration rate and inspired and expired

    concentrations of O2 , N2O, CO2 and AA

    (anaesthetic agent). The multigas module

    has automatic identification of AA, as well as

    identification of a secondary AA, if present.

    The electronic gas mixer, ventilator and optional

    gas module are all microprocessor controlled,

    with software installed.

    The machine can be equipped with various

    additional equipment such as: Active scavenging system Transfusion stand Arm for patient monitor

    Pin-index reduction valves for 4 L. gascylinders, or stand for 10 L. gas cylinders.

    The design of the machine is very flexible towards

    mounting of extra equipment and additional rails,

    clamps etc. can easily be mounted on the machine.

    1.1 Intended Use

    The SIESTA i TS anaesthesia machine is inten-

    ded for composition of anaesthetic breathing gas

    and manual or automatic ventilation of patients

    undergoing anaesthesia.

    The SIESTA i TS anaesthesia machine is

    intended for adult and paediatric patients.

    Qualified anaesthesia personnel should operate

    the SIESTA i TS anaesthesia machine only.

    Before operating the machine mains electrical

    power and pressurised gas must be connected

    to the machine.

    The machine is not MRI compatible.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    8/114

    User Manual for SIESTA i TS

    1-2 Dameca

    1.2 Description

    The electronic gas mixer (EGM) of the SIESTA i

    TS controls the freshgas flow given to the patientbreathing system. The EGM continuously mea-

    sures this flow to assure that it is in accordance

    with the setting made by the user.

    Besides the EGM controls and displays data

    from electronic gas mixer, ventilator and

    optional multigas module on its 15 display,

    which is touch controlled.

    The ventilator of the SIESTA i TS uses

    compressed AIR or O2 to ventilate the patientthrough the Bag in Bottle system, placed in

    the integrated breathing system (IBS).

    The ventilator continuously measures airway

    pressure and inspired/expired volume (optional).

    If a multigas module is not installed, the ventila-

    tor can also measure inspired O2by means of a

    fuel cell sensor (optional) placed on the inspira-

    tory port of the patient breathing system.

    The integrated breathing system (IBS) of theSIESTA i TS is a semi-closed patient breathing

    system with one-way valves in both the inspira-

    tory and expiratory limb, and an absorber for

    eliminating CO2. The system is suitable for high

    flow (app. the patient minute volume), medium

    flow (app. 1-1.5 L/min) and low flow (app.

    0.5-1 L/min) of freshgas for patient ventilation.

    The IBS is a combination of a Bag in Bottle unit

    and a patient re-breathing circuit. The IBS has

    a standing bellows in order for the user to see

    the controlled ventilation of the patient by the

    movement of the bellows. Besides, in case of

    a leakage, room air will not be drawn into the

    system, and the leakage can be seen immediately

    by the position of the bellows.

    Integrated into the IBS is an APL valve which

    is used for manual ventilation only.

    The IBS uses a CO2 absorber i-SORB which

    contains 900 gr. Sodalime. The absorber isavailable as a re-usable or a disposable version.

    If the absorber is disconnected from the IBS,

    two valves automatically close the connections

    to the absorber, in order not to loose gases.

    The multigas module (optional) of the SIESTA i

    TS uses a gas sample from the Y-piece to mea-

    sure inspired and expired concentrations of O2 ,

    N2O, CO2 and AA (anaesthetic agent). Besides,the multigas module can detect if Ethanol or

    Acetone is expired by the patient, and measure

    the concentration. The multigas module has

    automatic identification of AA.

    If a secondary AA is present, the multigas

    module also identifies this, and measures the

    concentration.

    When gas sample has been measured by themultigas module, it is re-introduced to the pati-

    ent breathing system, on the expiratory side.

    The multigas module does not require re-cali-

    bration, as the calibration is stable for life-time.

    The 3 pcs. integrated power outlets (optional)

    of the Siesta i TS are separated from the mains

    power by a transformer. This transformer

    supplies isolated power to the power outlets.

    See section 9 regarding power rating for the

    power outlets.

    Note:The parts of the SIESTA i TS do notcontain any latex.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    9/114

    User Manual for SIESTA i TS

    Dameca 1-3

    1.3 Performance

    The electronic gas mixer of the SIESTA i TS is

    designed to give a freshgas mixture selected bythe user of:

    0 to 20 L/min O2 / N2O (O2 0-10 L/min,

    N2O 0-10 L/min)

    or 0 to 20 L/min O2 / AIR (O2 0-10 L/min,

    AIR 0-10 L/min)

    This flow is led through the vaporizer rail,

    where anaesthetic agent can be added by meansof a user selectable vaporizer.

    By activating an oxygen by-pass valve, approx.

    45 L/min of O2 will be added to the freshgas

    flow, after the vaporizer rail.

    The SIESTA i TS has an integrated gas

    scavenging system which removes excess

    gases from an open reservoir. This system

    must be connected to an AGSS power unit with

    a flow of 30-40 L/min. If an external AGSS

    power unit is not available, the SIESTA i TS

    can be equipped with an integrated AGSS

    power unit (optional).

    The ventilator of the SIESTA i TS is designed

    to give an inspiratory flow of:

    Continous flow: 2 to 80 L/min Max. Peak flow: 120 L/min.

    In VCV and SIMV mode the Tidal volume,Resp. rate, I:E ratio and Insp. pause user settings

    are continuously controlled by the ventilator in

    order to stay within the 2 to 80 L/min continuous

    flow limits.

    Inspiratory flow = Tidal volume x Resp. rate x

    ((I+E)/I) x (100/(100-Insp. pause))

    The integrated breathing system (IBS) of the

    SIESTA i TS has an expiratory and inspiratoryresistance of:

    60 L/min 30 L/min 5 L/min

    Expiratory 4.5 2.0 0.3

    resistance (x100Pa) (x100Pa) (x100Pa)

    Inspiratory 4.5 2.0 0.3

    resistance (x100Pa) (x100Pa) (x100Pa)

    Compliance = 4 ml/hPa with adult tubings

    (22mm diameter).

    The CO2 absorber i-SORB of the integrated

    breathing system (IBS) is designed to optimize

    the flow through the absorber, so that all of the

    sodalime is being used.

    qusjob8.absorberkurver.tif

    Note:

    The SIESTA i TS contains batteries thatmust be exchanged by a technician atcertain intervals, as prescribed in theservice manual.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    10/114

    User Manual for SIESTA i TS

    1-4 Dameca

    1.4 User Setup

    The SIESTA i TS can be set up to match user

    preferences.

    The following features can be installed when

    the machine is manufactured:

    Integrated patient suction Auxiliary freshgas outlet and integrated

    auxiliary O2 flowmeter Multigas module 3 pcs. integrated power outlets

    Extra O2 , AIR or VAC gas outlet 1 or 2 pcs. pin-index yokes for 4 L. gas cy-

    linders or stand for 2 pcs. 10 L. gas cylinders Power unit for AGSS PSV accessible or not

    The following parameters can be changed

    after the machine has been manufactured:

    Ventilator volume setting based on Tidal

    volume or minute volume. SIMV accessible or not. If multigas module is not installed, oxygen

    monitor based on external fuel cell sensor

    can be activated or de-activated. Volume monitor can be activated or

    de-activated.

    If volume monitor is activated, the volume

    measurement can be configured in two ways:

    Only expired volume

    If the volume monitor is configured to only

    exp. volume, the volume sensor must be placed

    at the expiratory port of the patient breathing

    system. The volume monitor will then measure

    the expired tidal- and minute-volume only.

    Full spirometry

    If the volume monitor is configured to full

    spirometry, the volume sensor must be placed

    at the Y-piece of the patient breathing system.

    The volume monitor will then measure expiredtidal- and minute volume, inspired tidal volume

    (during manual ventilation) and a pressure/

    volume graph (spirometry loop) can be selected

    by the user.

    All these settings can be changed in the service-

    menu, by a super-user or a technician.

    1.5 Responsibility of the

    Manufacturer

    Dameca a/s is responsible for the effects on

    safety, reliability and performance of the

    equipment only if:

    Personnel authorized by Dameca a/s

    carry out assembly, operations, extensions,

    re-adjustments, modifications, periodical

    service or repairs.

    The pneumatic installation of the relevant

    rooms complies with appropriate requirements.

    The equipment is used in accordance with

    this user manual.

    1.6 Safety Features of the SIESTAiTS Anasthesia Machine

    Gas supply

    O2, AIR and N2O supply failure alarm. Built-in

    check valves to prevent reverse flow of gases.

    The ventilator, patient suction (optional) and

    AGSS power unit (optional) is powered by the

    primary driving gas (AIR or O2 , normally AIR).

    The optional patient suction can also be

    powered by VAC (option).

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    11/114

    User Manual for SIESTA i TS

    Dameca 1-5

    If the primary driving gas pressure is lost, a

    driving gas valve automatically switches tothe secondary driving gas source (AIR or O2 ,

    normally O2)

    Electronic gas mixer

    N2O cut-off if O2 supply fails.

    Electronic O2/N2O ratio control (secures a

    minimum of 25% O2 in an O2/N2O mix).

    An MPL (Maximum Pressure Limiting) valveopens if the pressure of the freshgas flow to the

    auxiliary freshgas outlet exceeds 90 (x100Pa).

    Vaporizer rail

    Back-bar for one or two vaporizers. The back-bar for two vaporizers is fitted with interlock

    system to assure that only one vaporizer can

    be activated one at a time.

    Ventilator

    A high-pressure alarm will automatically force

    the ventilator into its expiratory state, avoiding

    dangerous overpressure situations.

    In addition the ventilator has a built-in pressurerelief valve which opens at 90 (x100Pa).

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    12/114

    User Manual for SIESTA i TS

    1-6 Dameca

    Flow Diagram

    qusjob42.SI1LW6~1.wmf

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    13/114

    User Manual for SIESTA i TS

    Dameca 2-1

    2. Safety Precautions

    Read this manual carefully and familiarizeyourself with the machine before you start

    to operate the machine.

    Always check the machine for proper function

    before use.

    A WARNING indicates a situation in which

    the user or the patient may be in danger of

    injury.

    A CAUTION indicates a situation that may

    lead to equipment damage or malfunction.

    A NOTE indicates a point of particular interest

    for more efficient and convenient operation.

    2.1 Warnings

    Do not open service hatches or covers. Nouser-serviceable parts inside the machine.

    A qualified service technician should

    perform service.

    The machine must be connected to correct

    grounded hospital grade electrical outlets

    only.

    Always unplug the electric power cord

    before cleaning or maintenance of the

    machine. Let it dry completely before

    reconnecting it to the electrical outlet.

    Spillage of liquids may cause electrical

    shock.

    Check that all cables are in good working

    condition before connecting the machine

    to mains or patient.

    Do not use this machine in an environmentexceeding the EMC levels in EN60601-1-1-2.

    External patient breathing systems used

    with this machine must conform to EN740

    or similar standard.

    Bacterial filters used with this machine must

    conform to EN 13328-1 or similar standard.

    An anaesthetic vapour delivery deviceconnected to the machine shall comply

    with ISO8835-4 or similar standard.

    Heat and moisture exchangers used with this

    machine must conform to EN ISO 9360 or

    similar standard.

    Humidifiers used with this machine must

    conform to EN ISO 8185 or similar standard.

    Suction units used with this machine

    must conform to EN ISO 10079-1,

    EN ISO 10079-2 and EN ISO 10079-3

    or similar standard.

    Medical gases and anaesthetic agents used

    with this machine must conform to the

    European pharmacopoeia.

    Central gas supply used with this machine

    must comply with EN737-1 or ISO7396 or

    similar standard.

    Note:If the central gas supply fails, theauxiliary gas outlet (optional) ofthis machine will also fail.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    14/114

    User Manual for SIESTA i TS

    2-2 Dameca

    A patient monitor connected to the patient

    breathing system, may cause a leakage,

    even if it is in stand-by mode.

    The AGSS must be turned on whenever the

    machine is in use.

    Do not add, remove or modify any part of

    the machine, unless authorised by Dameca a/s.

    Explosion hazard

    No oil or grease should be used in this

    machine. This also applies to siliconebased lubricants and inflammable oil or

    grease. Only PTFE based oxygen

    compatible lubricants must be used,

    (e.g. DAMECA special grease P/N 36825).

    The machine is not suitable for use in the

    presence of flammable anaesthetic agents

    such as ether and cyclopropane. Only

    anaesthetic agents complying with EN 740

    or similar standards must be used.

    Patient safety

    Do not use the machine for a clinical

    procedure without first ensuring that the

    complete system is operating correctly.

    If the multigas module of the Siesta i TS

    is not installed, the machine should always

    be used with a separate monitor for O2,

    anaesthetic agent and CO2 concentrations.

    This monitor must comply with the relevant

    standards under the EU Medical Device

    Directive.

    If the volume monitor of the Siesta i TS is

    not installed, the machine should always

    be used with a separate monitor for expired

    volume. This monitor must comply with

    the relevant standards or the EU Medical

    Device Directive

    Ensure that all connections are secure and

    check for proper operation of the machine

    before use. In particular the patient breathing

    system and O-rings should be checked toensure that connections are not loose, and

    the patient breathing circuit is not leaking.

    The use of antistatic or electrically condu-

    ctive breathing tubes when utilizing high

    frequency surgery equipment may cause

    burns and is therefore not recommended

    in any application of this machine.

    Constant attention by a qualified individualis required whenever a patient is under

    anaesthesia or connected to a ventilator.

    Some malfunction of equipment requires

    immediate action.

    When using the ventilator, alternative means

    for ventilating the patient must be present.

    When operating the machine, an alternative

    oxygen supply must always be present.

    Use only the anaesthetic agent for which

    the vaporizer is calibrated. Using the wrong

    agent in the vaporizer can be fatal for patients.

    If a re-usable CO2 absorber is used, the

    following must be taken into account during

    change of soda-lime: Soda-lime dust is

    caustic and can produce burns in the respira-

    tory tract if inhaled. Breathing protection

    must be used to ensure that no soda-lime

    dust is inhaled by personnel or patient.

    Do not start any test procedure of the

    electronic gas mixer or the ventilator with

    a patient connected to the machine.

    The test procedures are designed to ensure

    correct and safe function of the machine.

    Do not bypass the test procedures unless

    patient safety requires an immediate starting

    up procedure.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    15/114

    User Manual for SIESTA i TS

    Dameca 2-3

    Note:The auxiliary O2 flowmeter (optional)

    is by-passing the MPL valve. The pres-sure from the flowmeter may exceedthe pressure from the freshgas outlet.

    hetic gases in operating rooms. Observe

    that the ball of the flowmarker is visible.

    2.2 Cautions

    Before connecting the power cord to the

    electrical power supply outlet check that the

    local voltage and frequency are as stated on

    the TYPE label on the rear of the machine.

    Do not attempt to connect other cables or

    connectors to the auxiliary power outlets

    (optional), than the ones intended for thispurpose.

    Maximum load on side rails at table is

    20 kg. Too much heavy equipment attached

    at one side of the machine may cause the

    machine to tilt.

    Maximum total load on table plate

    (including side rails) is 40 kg.

    Maximum load on top plate is 40 kg.

    Maximum torque on siderails is 20 Nm.

    When closing the drawers of the machine,

    observe that nothing gets squeezed or

    damaged, with special attention to fingers.

    Vaporizers used for Fluothane, Halothane or

    other anaesthetic agents containing Tymol or

    other stabilisers must be emptied every other

    week, and the agent that is drained must be

    destroyed.

    Check calibration of vaporizers regularly

    using a proper gas indicator (e.g.refracto-

    meter or similar).

    Authorised service personnel should service

    this machine at least once a year.

    Failure of operation

    If the machine fails to respond as described,

    do not use until qualified service personnel

    have corrected the situation.

    Occupational hazard

    Do not use this machine in insufficiently

    ventilated areas. Use scavenging system to

    avoid increased concentrations of anaest-

    The O2percentage in the patient circuit may

    differ significantly from the O2percentage

    of the freshgas.

    When using soda lime in absorbers, follow

    the manufacturers instructions.

    If the machine has been used with different

    anaesthetic agents, residues of these agents

    (PPM) may still remain in the machine

    (vaporizer rail etc.). By setting a high

    freshgas flow during a long period of

    time these residues can be minimized.

    Means to avoid cross infection between

    patients must be used. See section 8

    regarding maintenance and cleaning.

    When connecting more than one external

    equipment to the auxiliary power outlets

    (optional), a high leakage current from one

    equipment (f.ex. in case of a defective earth

    conductor) will affect the leakage current

    from the other equipment.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    16/114

    User Manual for SIESTA i TS

    2-4 Dameca

    2.3 Symbols

    Type B (EN60601-1) protection

    against electrical shock.

    Attention, consult accompanying

    documents.

    Adjusment for auxiliary O2 or

    patient suction:

    Turn counter-clockwise to increase.

    Turn clockwise to decrease.

    ON/OFF switch for lamp.

    Protective earth.

    Equipotentiality.

    ON/OFF

    WEEE Symbol, meaning that the

    electronics must not be regarded

    as normal waste.

    Note:Tubings used with this product must fulfill the EN12342 standard, andrespiration bags must fulfill the EN1820 standard.

    Vaporizers should be serviced according

    to the manufacturers instructions by an

    authorised service centre.

    Use only original DAMECA hoses and

    accessories; other hoses and accessories

    may cause malfunctions.

    Use only original DAMECA O-rings;

    other O-rings may cause problems if they

    are not of the same size or material.

    Do not leave spare cylinder valves open

    when using the machine on central gas

    outlets.

    Pressure from cylinders may be higher than

    the pressure from central outlets, and the

    cylinder supply could be depleted.

    Disconnect the machine from mains and

    medical gas supply, when not in use, to

    prevent pollution.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    17/114

    User Manual for SIESTA i TS

    DamecaDameca 3-1

    3. Function

    Fig. 3-1 Anaesthesia machine model SIESTA iTS front view

    1. Castors with brake

    2. Drawers

    3. AGSS ON/OFF (option) and flow marker

    4. Auxiliary oxygen flowmeter outlet

    (optional)

    5. Auxiliary freshgas outlet (optional)

    6. CO2 absorber

    7. Lock for absorber

    8. BAG connection

    9. APL valve (for manual ventilation only)

    10. Bellows

    11. O2

    flush

    12. Water trap for multigas module (optional)

    13. Transfusion stand (optional)

    14. Top shelf

    15. Main switch

    16. 15 TFT colour touch screen

    17. Control wheel

    18. ON/OFF switch for light

    19. Emergency flow switch

    20. Back-bar for two interlock vaporizers

    21. Integrated auxiliary oxygen flowmeter

    (optional)

    22. Table with integrated rails

    23. Integrated patient suction (optional)

    24. Auxiliary freshgas outlet switch (optional)

    qusjob42.10653 front.png

    5

    10

    2

    1

    1215

    17

    22

    20

    1

    13

    24

    11

    8

    6

    14

    16

    3

    7

    9

    21

    23

    1819

    IBS

    4

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    18/114

    User Manual for SIESTA i TS

    3-2 Dameca

    Fig. 3-2 Anaesthesia machine model SIESTA i TS rear view

    1. Castors without brake

    2. Lower fan

    3. Yokes for spare cylinders (optional)

    4. Grounding point

    5. Mains power inlet switch

    6. Mains power cord

    7. Connections for patient flow sensor

    (optional)

    8. Auxiliary gas outlet (optional)

    9. Auxiliary AGSS connection

    10. RS232 port

    11. Auxiliary power outlets (optional)

    12. Upper fan

    13. Gas inlets

    14. Insp./Exp. connections for patient

    breathing system

    15. Lock for patient breathing system

    16. Patient suction connection (optional)

    17. Connection for optional O2 fuel cell

    sensor

    qusjob42.10653 bag.png

    11

    2

    13

    10

    1 1

    7

    15

    14

    16

    12

    65

    4

    3

    8

    9

    17

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    19/114

    User Manual for SIESTA i TS

    DamecaDameca 3-3

    Fig. 3-3 Main display if multigas module is installed

    Flow (L/min)O2 N2O

    1.00 2.00

    Rotameter

    Freshgas flow(L/min)

    3.00

    O2set %

    33

    Adult sensor At Y-piece Peak

    17

    Plateau

    17

    PEEP

    3

    Compl. 34 ml/hPa

    hPa

    0

    20

    40

    %

    0

    2

    4

    Stop watch

    00:00:00

    Cancel

    alarms

    Alarm

    menu

    Trend

    menu

    Setup

    menu

    Normal

    screen

    Vent. mode

    VCV

    Status

    VENT

    Exp CO2

    1.8Insp CO2

    0.1

    Resp. meas.

    MVexp(l)5.8

    Exp

    TV(ml) 480

    RR(1/min)12

    8.04.0

    OFF4

    Gas meas. (%)

    Exp Insp

    O2 32 33

    N2O64 63

    SEV0.4 0.6

    MAC 0.4(age: 40)

    PSV settings

    30

    TV set(ml)

    500

    RR set(1/min)

    12

    PEEP(hPa)

    OFF

    I:ERatio

    1:2

    P Insp(hPa)

    (17) More settings

    EGM 5.0 Thursday 6. December 11:13

    Touch screen controlled Selftest bypassed

    qusjob42.Fig. 3-3 with gasmodule.pdf

    1. Window for set freshgas flow

    2. Window for measured freshgas flow

    3. Window for set O2 percentage in

    freshgas flow

    4. Window for set total freshgas flow

    5. Window for alarms and messages

    6. Window for date and real time

    7. Window for measurement of CO2concentrations

    8. Window for measurement of gas

    concentrations

    9. Window for stopwatch

    10. Window for activating normal screen

    11. Window for activating setup menu

    12. Window for activating trend menu

    13. Window for activating alarm menu

    14. Window for activating cancel alarms

    15. Window for measurement of respiration

    data (optional)

    16. Window for machine status VENT/

    MANUAL/STBY)

    17. Window for ventilation mode

    18. Window for more ventilator settings

    19. Window for PSV settings (optional)

    20. Window for set inspiratory pressure

    21. Window for set PEEP

    22. Window for set I:E ratio

    23. Window for set respiration rate

    24. Window for set tidal volume

    25. Window for respiratory graph,

    pressure measurements and

    high alarm setting.

    13

    6

    23

    4

    3

    2

    1

    5 7 98

    10

    11

    12

    14

    15

    22 21 20 19 1824 17

    16

    25

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    20/114

    User Manual for SIESTA i TS

    3-4 Dameca

    Fig. 3-4 Main display if multigas module is not installed

    1. Window for set freshgas flow

    2. Window for measured freshgas flow

    3. Window for set O2 percentage in

    freshgas flow

    4. Window for set total freshgas flow

    5. Window for alarms and messages

    6. Window for date and real time

    7. Window for respiratory graph, pressure

    measurements and high alarm setting

    8. Window for measurement of inspired

    O2 percentage (optional)

    9. Window for stopwatch

    10. Window for activating normal screen

    11. Window for activating setup menu

    12. Window for activating trend menu

    13. Window for activating alarm menu

    14. Window for activating cancel

    alarms

    15. Window for measurement of

    respiration data (optional)

    16. Window for machine status

    (VENT/MANUAL/STBY)

    17. Window for ventilation mode

    18. Window for set inspiratory pause

    19. Window for PSV settings (optional)

    20. Window for set inspiratory pressure

    21. Window for set PEEP

    22. Window for set I:E ratio

    23. Window for set respiration rate

    24. Window for set tidal volume

    Flow (L/min)O2 N2O

    1.00 2.00

    Rotameter

    Freshgas flow(L/min)

    3.00

    O2set %

    33

    hPa

    0

    20

    40

    Stop watch

    00:00:00

    Cancel

    alarms

    Alarm

    menu

    Trend

    menu

    Setup

    menu

    Normal

    screen

    Vent. mode

    VCV

    Status

    VENT

    Resp. meas.

    MVexp(l)5.8

    Exp

    TV(ml) 480

    8.04.0

    PSV settings

    30

    TV set(ml)

    500

    RR set(1/min)

    12

    PEEP(hPa)

    OFF

    I:ERatio

    1:2

    P Insp(hPa)

    (17) More settings

    EGM 5.0 Thursday 6. December 11:13

    Touch screen controlled Selftest bypassed

    O2meas. %

    32

    Adult sensor At Y-piece

    Peak

    17

    Plateau

    17

    PEEP

    3

    Compl. 34 ml/hPa

    qusjob42.fig. 3-4 without gasmodule.PDF

    13

    6

    23

    4

    3

    2

    1

    5 7 98

    10

    11

    12

    14

    15

    22 21 20 19 18 17

    16

    24

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    21/114

    User Manual for SIESTA i TS

    DamecaDameca 3-5

    3.1 Anaesthesia Gas Scavenging System (AGSS)

    The Siesta i TS anaesthesia machine has a

    built-in scavenging system. This systemrequires evacuation flow of 30-40 L/min.,

    in order to remove excess gases from the

    integrated patient breathing system to the

    AGSS outlet on the rear of the machine.

    The evacuation flow can be generated by the

    permanent gas installation in the operating

    room, or it can be generated by an air driven

    ejector in the Siesta i TS machine, activated

    with an ON/OFF switch located at left side ofthe machine, next to the AGSS flow marker.

    A built-in reservoir in the scavenging system is

    open to the ambient air and serves as a relief ofthe ejector vacuum.

    An AGSS connector on the rear of the machine

    is for connection of the machine to the central

    scavenging system.

    Warning:The gases and agents used underanaesthesia can pose a danger ifAGSS is not active, or the operatingroom is not ventilated properly.

    Fig. 3-5 Flow marker

    The air driven ejector is optional.

    An AGSS flowmarker at left side of the

    machine contains a yellow ball which is

    visible when the flow is 25 L/min or more.

    qusjob33.flowmark.pcx

    Note:The AGSS flowmarker is for the IBS(Integrated Breathing System) only.If an external breathing system isused, this system must be equippedwith an indicator for the AGSS flow.

    3.1.1 External Breathing System

    If an external breathing system is used with the

    machine (see section 4.1.3), the excess gases

    from this system can be removed by connecting

    it to the auxiliary AGSS connection on rear sideof the machine (9 on fig. 3-2).

    Remove the plug from the auxiliary AGSS

    connection (30 mm female cone) and connect

    the external breathing system.

    Warning:The external breathing system must

    have an open reservoir integrated intothe AGSS part of the system.Otherwise the auxiliary AGSS flowcan create a sub-atmospheric pressurein the external breathing system,which can be dangerous to the patient.

    When an external breathing system is no longer

    used with the machine, re-mount the plug on the

    auxiliary AGSS connection (30 mm female cone).

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    22/114

    User Manual for SIESTA i TS

    3-6 Dameca

    Left blank intentionally

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    23/114

    User Manual for SIESTA i TS

    DamecaDameca 4-1

    4. Operation

    Note:Before using the machine the first timeit must be electrical safety testedaccording to EN 60601-1, and all

    patient related parts must be cleaned.

    Note:Before connecting the machine to a

    patient, a function test must be carriedout as described in section 4.2.

    Note:Check voltage rating on the type labelat the rear panel of the machine,

    before connecting it to the mains.

    4.1 Set Up

    Connect the machine to the mains.

    Connect the machine to the pressurised

    gases (400 - 600 kPa), vacuum (optional)

    and AGSS. Activate AGSS flow and check that the yellow

    ball in the flow marker (3 on fig. 3-1) appears.

    The AGSS flow can be generated by the

    machine (optional) or by a unit in the AGSS

    installation of the hospital. If the AGSS flow

    must be generated by the machine, it can

    be activated by pressing the red/green pin

    placed above the flow marker (3 on fig. 3-1).

    4.1.1 Integrated breathing system

    Assemble the integrated breathing system

    (see section 8), connect it to the base (1)

    and lock it by using the handle (2). Connect i-SORB CO2 absorber (5) to the

    integrated breathing system, by placing

    it into the bracket underneath the base

    and pulling up, until the lock clicks.

    Caution:If a disposable i-SORB CO2 absorber is

    used with the system, the seal on a newabsorber must be removed, before theabsorber is connected to the integrated

    breathing system.

    qusjob42.10653-00 slange.pdf

    2

    1

    5

    4

    3

    6

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    24/114

    User Manual for SIESTA i TS

    4-2 Dameca

    Connect inspiratory and expiratory hoses

    (3) to the connections marked INSP and

    EXP on the integrated breathing system,

    including Y-piece (4) and bacteria filter(P/N 38477). The bacteria filter can be

    placed at the Y-piece or at the expiratory

    connection of the IBS system. Connect manual respiration bag (6) to the

    connection marked BAG on the integrated

    breathing system. A hose can be placed

    between BAG connection and respiration

    bag to give better flexibility for the user.

    Note:

    If an i-SORB CO2 absorber is notreinserted within 30 sec., a messagewill be shown on the display:Absorber disconnected.

    If the ventilator is in automaticventilation mode, this messagewill be followed by an alarm.

    21-12-200409:47

    qusjob42.10653ny.pdf

    Note:The integrated breathing system mustnot be changed during use.

    If the integrated breathing system isdisconnected, a message will beshown on the display:

    Patientsystem disconnected.

    If the ventilator is in automaticventilation mode, this messagewill be followed by an alarm.

    Note:The respiration bag must be 1-3 L.otherwise the leakage test of theBAG system, included in the selftest,may not work correctly.

    Set the freshgas switch (24 on fig. 3-1) to

    Circle System.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    25/114

    User Manual for SIESTA i TS

    Dameca 4-3

    4.1.2 Changing the CO2 absorber

    4.1.3 External breathing system

    Other breathing systems than the integrated

    breathing system can be used with the

    Siestai TS anaesthesia machine.

    Assemble the breathing system and connect

    it to the auxiliary freshgas outlet (optional)

    placed at left side of the writing table

    (5 on fig. 3-1)

    PUSH

    Remove sealon new absorber.(disposable ver.)

    "CLICK"

    qusjob42.11960uds ab.pdf

    1. Take the handle of the absorber.

    2. Push the knob above the handle of the

    absorber to release it. The two absorber

    valves will close in order to prevent loss

    of gases during change.

    3. Lift the used absorber out of the bracket.

    4. Put the fresh absorber into the bracket.

    5. Lift the handle until the lock gives a click.

    6. The absorber is now attached to the patientsystem.

    Caution:A bacteria filter should always be usedwith the integrated breathing system.Dameca bacteria filter P/N 38477-50(50 pcs.) can be used.

    In order to remove excess gases from the

    breathing system, connect it to the to 30 mm

    female cone marked AUX AGSS on the

    rear of the machine (9 on fig. 3-2).

    See section 3.1.1

    Set the freshgas switch (24 on fig. 3-1)

    to Aux Outlet.

    21-12-200409:47

    qusjob42.10653ny.pdf

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    26/114

    User Manual for SIESTA i TS

    4-4 Dameca

    Caution:If the ventilator is in automaticventilation mode, when the freshgasswitch is set to Aux Outlet, theventilator will stop and an alarmwill start: Ventilator stopped due toexternal freshgasoutlet activated.

    4.1.4 Multigas module water trap

    If an integrated gas module is installed, the water

    trap (12 on fig. 3-1) must be installed prior to use.

    Make sure the water trap is not filled with water,

    and install it on the base on left side of the machine.

    If the water trap is filled with water, it must be

    replaced (P/N 41020-25).

    The capacity of the water trap is 22 ml.

    Attach a sample tube between the water trap and

    the Y-piece, for example via a bacteria filter.

    The sample tube must be 2 to 3 metres long and

    have an internal diameter of at least 1.2 mm.

    Dameca sample tube P/N 42630-100 (100 pcs.,

    2m) or P/N 42630-200 (100 pcs., 3m) can be

    used.

    Adult sensor At Y-piece Peak

    17

    Plateau

    17

    PEEP

    3

    Compl. 34 ml/hPa

    hPa

    0

    20

    40

    30

    PEEP(hPa)

    I:ERatio

    P Insp(hPa)

    Note:

    When the freshgas switch is set toAux Outlet, a message will beshown on the display: Externalfreshgasoutlet activated. In thissituation the patient can only beventilated manually, as the ventilatorcannot be activated.If the user tries to activate the ventilator,a message will be shown on the display:Cannot start ventilator when external

    freshgasoutlet is activated.

    4.1.5 Patient volume sensor

    The machine can be set up with volume monitor

    (15 on fig. 3-3 or 15 on fig. 3-4), which uses a

    flow sensor.

    The machine uses two different flow sensors:

    Paediatric flow sensor is used for patients with

    tidal volume from 20 to 300 ml (insp./exp. flow

    2-35 L/min.).

    Adult flow sensor is used for patients with tidal

    volume from 200 to 1500 ml (insp./exp. flow

    10-120 L/min.).

    If the flow sensor is placed on the expiratory

    side, the expired tidal volume and minute

    volume are measured.

    If the flow sensor is placed at the Y-piece, the

    inspired tidal volume is also measured, and a

    spirometry-graph (pressure, volume) can be

    shown on the main display.

    In the service menu a super-user or a technician

    can define the default value for placement of theflow sensor.

    In the spirometry-menu the user can change the

    placement of the flow sensor. However the ma-

    chine will always go back to default value, after

    selftest or when the machine is powered off.

    If the flow sensor is placed at the Y-piece, be

    sure to place the sensor with the PATIENT

    port facing the patient.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    27/114

    External flowsensor kit P/N 10967-120

    Item Qty Description Part No.

    1 1 Spirometry tube 2 m 42632-091

    2 1 Adult flow sensor, Re-usable 42632-06

    3 1 Paed. flow sensor, Re-usable 42632-07

    4 2 Single-use bacterial filter. 38477

    1

    42

    3

    qusjob41.10967-120.pdf

    Patient port

    Caution:After selftest has been performed, the type of flow sensor (Adult/Paediatric)must be defined in the setup menu.

    User Manual for SIESTA i TS

    Dameca 4-5

    The actual settings for the flow sensor (type and placing) are shown in the window for respira-

    tion graph. If flow sensor type and placing are not shown, this means the machine is installed

    without any flow sensor.

    Note:A bacteria filter must always be placed between the patient and the flowsensor.

    If the external flow sensor is used, tubings between flow sensor and connectors (7 on fig. 3-2)

    on the machine must be connected.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    28/114

    4.1.6 O2 fuel cell sensor

    If an integrated gas module is not installed, the

    machine can be set up with an O2 monitor (8 onfig. 3-4), which uses an external fuel cell sensor.

    Then the main display will look like the

    following:

    Place the O2 sensor on T-piece

    (P/N 42012-07) in the patientbreathing system on the inspiratory

    side. The O2 sensor must always face

    upwards.

    Connect cable between O2 sensor and

    connector (17 on fig. 3-2) on rear panel

    of the machine.

    User Manual for SIESTA i TS

    4-6 Dameca

    Stop watch

    00:00:00

    Setup

    menu

    Normal

    screen

    :13

    ed

    O2meas. %

    32

    O2 Sensor kit P/N 10967-85

    Item Qty Description Part No.1 1 T-piece 42012-07

    2 1 Oxygen sensor 42035-46

    3 1 Cable for oxygen sensor 10967-86

    2

    1

    3

    qusjob41.10967-85.pdf

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    29/114

    User Manual for SIESTA i TS

    Dameca 4-7

    4.2 Function Test

    Turn on the mains power inlet switch on

    rear side of the machine (5 on fig. 3-2).

    If the main switch on front of the machine is set

    to ON during this period, the message will dis-

    appear and the machine will continue to function.

    If the main switch on front of the machine is NOT

    set to ON position during this period, the ma-

    chine will power off when the period has ended.Note:When the mains power switch onrear side of the machine is turnedon, the batteries inside the machinewill be charged at all times, inde-

    pendent of whether the machine isrunning or not. Therefore this switchshould be set to ON at all times.

    Push the main switch on front of the machine

    (15 on fig. 3-1) to set it to ON position.

    Note:If a mains failure alarm is givenwhen the main switch on front ofthe machine is set to ON position,this is most likely caused by themains power inlet switch on rear sideof the machine not being set to ON.

    Note:

    When the main switch on front ofthe machine at a later time is pushedin order to set it to OFF position, amessage will be shown on the screen:Machine closing down. Press ma-ins switch to cancel. Now a powerdelay period of app. 20 seconds isstarted. During this the machinewill continue to function as before.

    Note:The integrated breathing system must

    be assembled and connected, incl. hoses,Y-piece and i-SORB CO

    2

    absorber, be-

    fore a function test can be carried out.

    Warning:The patient must not be connected tothe machine at power-up or duringthe functional test.

    Now the machine will power up and perform an

    initial self-test checking the internal system for

    safe condition and performing test sequence 1.

    This takes approx. 1 minute.

    Note:

    If the machine is equipped with an inte-grated multigas module, this modulewill not be working until approx. 2 min.after activating the main switch on front

    panel. If the machine is equipped withan O2 monitor which uses an external

    fuel cell sensor, this monitor will not beworking until approx. 1 minute afteractivating the main switch on front panel.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    30/114

    User Manual for SIESTA i TS

    4-8 Dameca

    When the initial self-test has been successfully

    completed, the machine is in Stand-by mode

    and the full selftest can be performed.

    4.2.1 FULL self test

    During the FULL selftest the machine checks

    its functions, measures the compliance of the

    system and checks for leakages. The compliance

    and leakage test includes the integrated breathing

    system incl. hoses and Y-piece.

    When the mains switch (15 on fig. 3-1) is set to

    ON by the user, the machine automaticallyperforms test sequence 1 to ensure the correct

    function of its mandatory components. Then test

    sequence 2 is activated.

    Test sequences:

    Start-up

    Display: Do you want to perform FULLTEST: Y/N

    Patient must NOT be connected

    during test

    Select Y or N and press to continue

    Change between Y and Nby turning the

    control wheel, and press enter to select. By selecting Y the FULL selftest is started.

    By selecting N the FULL selftest is bypas-

    sed. This should only be done in case of anemergency.

    Even if the FULL selftest has been started, the

    user can bypass the test by pressing the normal

    screen field (10 on fig. 3-3), and the following

    text will be shown:

    Display: Do you want to bypass selftest:

    Y/N

    Only bypass in case of anemergency

    Select Y or N and press to continue

    Change between Y and Nby turning the

    control wheel, and press enter to select. By selecting Y the FULL selftest is bypas-

    sed. This should only be done in case of

    an emergency. By selecting N the FULL selftest is started

    again from the beginning.

    Warning:If the FULL selftest is bypassed, someof the functions of the machine willnot be checked. The user must thenincrease observation of the patientduring anaesthesia, in order to

    maintain a safe function.

    Note:During the full selftest all alarmsare disabled.

    Note:If a system related alarm (see section 7)is detected after power-up of themachine, the selftest is not started.Instead the alarm is shown on thedisplay. When the alarm is no longeractive, press cancel alarms and theselftest will start.

    When FULL TEST is selected by the user

    (by selecting FULL TEST in the selftest

    menu, included in the setup menu, and pressing

    the control wheel), start-up sequence is

    activated before going to test sequence 1.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    31/114

    User Manual for SIESTA i TS

    Dameca 4-9

    Warning:

    If the FULL selftest is bypassed, thecompliance and leakage of the integra-ted breathing system are not measured.The ventilator therefore uses the datafrom last selftest, and therefore the settidal volume in VCV mode might not

    be correct.

    Note:If the FULL selftest has been bypassed,the text SELFTEST BYPASSEDis shown constantly on the main display.

    Warning:The patient must not be connected tothe machine during the FULL selftest.

    2. Gasmixer test

    Display: Occlude Y-piece (use testadapter).

    Check gas evacuation (flow marker)Set APL valve to 35 hPa

    Open vaporizer

    Press to start

    If you want to leave this test, press

    normal screen

    Place the Y-piece on the test plug at the BAG

    connection of the integrated breathing system.

    Check the yellow ball in the flow marker

    (3 on fig. 3-1) is visible.

    Set the APL valve in patient breathing system

    (9 on fig. 3-1) to 35 hPa, and open the vaporizer.

    Press the control wheel to confirm.

    Display: Electronic gas mixer test is running

    The functionality of the electronic gasmixer

    is checked.

    Warning:

    The FULL selftest must be carriedout at least one time per day, whenthe machine is in use.

    1. System and valves

    Display: System check running

    The internal safety system of the machine

    (watchdog) and the valves for the electronic

    gasmixer and the valves for the ventilator

    are checked during this.

    Note:If the machine is setup with a separate

    patient monitoring device with measure-ment of gas concentrations, and thesample flow is not returned to the

    patient system, the tube for sample flowmust not be connected to the patientsystem during the test. Otherwise theleakage will be too high.

    3. Filling of BAG

    Display: Leakage test of system is running

    The electronic gas mixer generates a flow

    of 10 L/min O2 and 9.75 L/min AIR for

    10 seconds to fill the respiration bag.

    After this, the airway pressure in the patient

    system is checked (should be 25 to 45 hPa).

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    32/114

    User Manual for SIESTA i TS

    4-10 Dameca

    4. Compliance and leakage, VENT

    Display: Leakage test of system is running

    The electronic gas mixer generates a flow of

    10 L/min O2 and 9.75 L/min AIR for 10 seconds

    to fill the bag in the Bag in Bottle system.

    After this the electronic gas mixer generates

    a flow of 5 L/min O2 and the pressure in the

    patient system is checked (should be 1.5 to

    4.5 hPa due to Bag in Bottle).

    Then the compliance of the VENT system ismeasured, based upon the volume required from

    the ventilator to reach a 30 hPa pressure

    in the patient system.

    This 30 hPa pressure must be reached within

    10 sec. at an inspiratory flow of 5 L/min., or an

    error message will be shown on main display.

    The compliance of the system must be from

    2.0 to 10.0 ml/hPa, or an error message willbe shown on main display.

    When the 30 hPa pressure is reached, the

    leakage of the VENT system is measured, based

    upon the pressure drop during time, taking into

    account the compliance of the system.

    If the pressure drops too fast for the leakage to

    be measured, an error message will be shown on

    main display:

    Display: Leakage too high

    Check BiB, connectors and hoses

    Press to repeat test

    When the control wheel is pressed, the compli-

    ance and leakage test will be performed again.

    If the leakage is more than 100 ml/min but

    less than 1000 ml/min, an error message will

    be shown on main display:

    Display: Leakage in VENT system XXX

    ml/min. Accept: Y/N

    Check BiB, connectors and hoses

    Select N to repeat test. Select Y tocontinue

    Change between Y and Nby turning the

    control wheel and press enter to select. By selecting Y the measured leakage is

    accepted, and the next test sequence will

    be performed. By selecting N the measured leakage is not

    accepted, and the compliance and leakage

    test will be performed again.

    If the leakage is more than 1000 ml/min, an

    error message will be shown on main display:

    Display: Leakage in VENT system XXX

    ml/min.

    Check BiB, connectors and hoses

    Press to repeat test

    When the control wheel is pressed, the compli-ance and leakage test will be performed again.

    5. Leakage, BAG

    The BAG/VENT valve in the integrated

    breathing system is set to BAG and the

    pressure in the patient system is measured

    to detect possible leakage from the APL valve

    or the respiration bag. If the pressure has

    dropped to a value below 10 hPa, an error

    message will be shown on main display:

    Display: Leakage in BAG system too high

    (pressure below 10 hPa)

    Check APL valve and resp. bag

    (empty ?)

    Press to repeat test

    When the control wheel is pressed, the compli-

    ance and leakage test will be performed again.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    33/114

    User Manual for SIESTA i TS

    Dameca 4-11

    6. Leakage, BAG/VENT, exp. valve

    and peep valve

    Display: Set APL valve at pat. circuit toSP. Close vaporizer

    Press to continue

    Set the APL valve to SP and close the vaporizer.

    Press the control wheel to confirm.

    Display: Valve check running

    When a 30 hPa pressure is reached on the

    outside of the bellows of the Bag in Bottle,with the BAG/VENT valve in BAG position,

    the leakage from the BAG/VENT valve and

    from the expiration and peep valves inside the

    ventilator is checked based upon the pressure

    drop during time.

    If the pressure drop is too high the ventilator

    gives an error message:

    Display: Leakage in BiB system too highCheck BiB and change-over valve

    Press to repeat test

    When the control wheel is pressed, the leakage

    test for the BAG/VENT valve and the expiration

    and peep valves inside the ventilator will be

    performed again.

    7. Test completedDisplay: Leakage in VENT system: XXX

    ml/min

    Compliance in VENT system: X.X

    ml/hPa

    Press to continue

    When the control wheel is pressed, the next text

    appears on the display:

    Display: Refreshing gases in patient system

    The electronic gasmixer generates a flow of 10L/min O2 and 9.75 L/min. AIR for 10 seconds to

    remove any anaesthetic agent from the patient system.

    Then the nex text appears on the display.

    Display: Check alarm sound present

    Press cancel alarms to confirm

    Check that the alarm sound is present, and press

    cancel alarms. Then the next text appears onthe display:

    Display: Test completed. Check pat.

    suction, vaporizer etc.

    Adjust APL valve

    Remove Y-piece from test adapter

    If vol. monitor is installed, sensor

    must be connected and defined

    Read user manual

    Press to finalize

    The APL valve should be adjusted as

    desired by the user. Y-piece should be removed from the testadapter If the machine is set-up with volume monitor,

    the type of sensor (Adult or Paediatric) must

    be defined in the setup menu by the user. When the control wheel is pressed, the

    machine goes to STBY mode and all user

    settings are set to default values. Then the manual part of the selftest should

    be performed, as described in section 4.2.3 to

    4.2.15

    4.2.2 LC self test

    The LC (Leakage and Compliance) test can be used

    for measuring compliance and checking for leak-

    age if parts of the patient system (f. ex. hoses and

    respiration bag) have been changed between two

    patients.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    34/114

    User Manual for SIESTA i TS

    4-12 Dameca

    During the LC selftest the compliance of the

    system is measured, and the leakage is checked.

    The compliance and leakage test includes the in-

    tegrated breathing system incl. hoses and Y-piece.

    When LC TEST is selected by the user (by

    selecting LC TEST in the selftest menu,

    included in the setup menu, and pressing the

    control wheel), the following test sequences

    are activated.

    Change between Y and N by turning the

    control wheel, and press enter to select. By selecting Y the LC selftest is bypassed.

    By selecting N the LC selftest is startedagain, from the beginning.

    Note:During the LC selftest all alarmsare disabled.

    Test sequences:

    Start-up

    Display: Do you want to perform LC TEST:

    Y/N

    Patient must NOT be connectedduring test

    Select Y or N and press to continue

    Change between Y and Nby turning the

    control wheel, and press enter to select. By selecting Y the LC selftest is started. By selecting N the LC selftest is bypassed.

    Even if the FULL selftest has been started, the

    user can bypass the test by pressing the normal

    screen field (10 on fig. 3-3), and the following

    text will be shown:

    Display: Do you want to bypass selftest:

    Y/N

    Only bypass in case of an

    emergency

    Select Y or N and press to

    continue

    Warning:If parts of the patient system have

    been changed and the LC selftest is notperformed, the compliance and leakageof the integrated breathing system arenot measured. The ventilator thereforeuses the data from last selftest, andtherefore the set tidal volume in VCVmode might not be correct.

    Note:If the LC selftest has been bypassed,the text SELFTEST BYPASSED isshown constantly on the main display.

    Warning:The patient must not be connected tothe machine during the LC selftest.

    Test sequence 2 to 7 from the FULL selftest

    are carried out.

    4.2.3 O2 monitor test

    4.2.3.1 Fuel cell sensor

    If the machine is set up with an O2 monitor

    using an external fuel cell sensor (see section

    4.1.6), the following test must be carried out:

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    35/114

    1. Set status to MANUAL. Y-piece must be

    open to ambient air.

    2. Set freshgas flow to 10 L/min AIR, and waituntil O2 % reading on main display is stable.

    Check that the reading is 20-22 % O2. If not,

    recalibrate the O2-monitor as described in

    section 6.1.6.

    3. Set freshgas flow to 10 L/min O2, and wait

    until O2 % reading on main display is stable.

    Check that the reading is 98-102 % O2. If

    not, recalibrate the O2-monitor as described

    in section 6.1.6.

    4.2.3.2 Multigas module

    If the machine is set up with an integrated

    multigas module, the following test must

    be carried out:

    1. Set status to MANUAL. Y-piece must be

    open to ambient air.

    2. Set freshgas flow to 10 L/min AIR, and waituntil O2 % reading on main display is stable.

    Check that the reading is 19-23 % O2.

    3. Set freshgas flow to 10 L/min O2, and wait

    until O2 % reading on main display is stable.

    Check that the reading is 97-103 % O2.

    4.2.4 Expired minute volume

    If the machine is set up with a volume monitor

    (see section 4.1.5), the following test must be

    carried out:

    1. Attach a test lung to the Y-piece and start

    the ventilator in VCV mode. Wait approx. 1

    minute and read the measured expired minute

    volume on main display. Check that thiscorresponds to the set minute volume

    (TVset x RRset). Check that the insp. and

    exp. valves in the patient system are moving

    during inspiration/expiration.

    4.2.5 Airway pressure high alarm

    1. Attach a test lung to the Y-piece and start the

    ventilator. Read PEAK pressure on main

    display and set airway pressure high alarmlimit 5 hPa below PEAK pressure.

    Check that the airway pressure high alarm

    starts and that the PEAK pressure is now

    reduced by the new airway pressure high

    alarm limit. Set the airway pressure high

    alarm limit as before (above PEAK pressure)

    and check that the alarm stops.

    4.2.6 Mains failure alarm1. Attach a test lung to the Y-piece and start the

    ventilator. Disconnect mains cable to the

    machine from the mains outlet. Check that

    the mains failure alarm starts, and that the

    machine continues to function. Re-connect

    mains cable.

    4.2.7 O2 flush

    1. Set status to MANUAL. Occlude the

    Y-piece. Press O2 flush and check that

    the respiration bag is quickly filled.

    User Manual for SIESTA i TS

    Dameca 4-13

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    36/114

    User Manual for SIESTA i TS

    4-14 Dameca

    4.2.8 O2 alarm

    1. Set freshgas flow 1 L/min O2 and 1 L/min

    N2O. Disconnect O2 gas supply to the ma-chine from the gas outlet and wait until the

    O2 flow column on main display goes to zero.

    Check that the O2 supply pressure is low

    alarm is active, and check that the N2O flow

    column is also zero. Re-connect O2 gas supply.

    4.2.9 Spare cylinder test

    If the machine is equipped with spare cylindersfor O2 or N2O, the following test must be

    performed:

    1. Disconnect the central gas supply tubings

    for O2 and N2O.

    2. Open the valves of the spare O2 and N2O

    cylinders.

    3. Set the Carrier gas in the setup-menu to N2O.

    4. Set Total freshgas flow to 0.0 L/min.

    5. Read the pressure on the spare cylinder

    pressure regulator gauges on rear side

    of the machine.

    6. Check for sufficient cylinder pressure

    (the pressure of the cylinders must never

    be below 10 bar).

    7. Close the spare cylinder valves.

    8. Observe the pressure gauges for at least

    one minute. The pressure must not

    decrease.

    9. Reconnect the central supply tubings to the

    central gas supply.

    4.2.10 Patient suction test

    If the machine is equipped with a patient suction

    unit, the user must check the functionality ofthis unit by the following procedure:

    1. Fully open the regulator on the suction unit.

    2. Remove the plastic tubing from the suction

    container and block the tube opening.

    3. Start suction by activating the ON/OFF switch.

    4. Verify that the vacuum gauge indicates avacuum of at least -0.7 kPa.

    5. Adjust the desired vacuum by means of the

    regulator knob.

    6. Close the ON/OFF switch.

    7. Re-mount the plastic tubing on the suction

    container.

    4.2.11 Vaporizer test

    The user must check the functionality of the

    vaporizer(s) by the following procedure:

    1. Verify that the required vaporizer(s) is

    connected.

    2. Check that the regulator knob on the vapori-

    zer(s) turns freely in the entire regulation

    area and is locked in the 0 position.

    3. Verify that the vaporizer is sufficiently filled.

    Note:Consult the vaporizer user manual.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    37/114

    4.2.12 Aux. O2

    flowmeter test

    If the machine is equipped with an auxiliary O2

    flowmeter unit, the user must check the functio-nality of this unit by the following procedure:

    1. Open the flow meter regulator (turn it

    counter-clockwise).

    2. Check that the flow can reach 10 ltr/min.

    3. Close the regulator (turn it clockwise)

    and check there is no flow.

    4.2.13 Test of internal switches

    1. Remove the CO2 absorber and check the

    following message appears on the display

    (after app. 30 sec.):

    Display: Absorber disconnected

    Re-mount the CO2 absorber and check the

    message disappears.

    2. Remove the integrated breathing system

    (unlock handle) and check the following

    message appears on the display:

    Display: Patientsystem disconnected

    Re-mount the integrated breathing system and

    check the message disappears.

    3. Activate the emergency freshgas switch and

    check the following message appears on the

    display:

    Display: Emergency O2 flow activated

    App. 5 L/min. added to the O2flow setting

    De-activate the switch and check the message

    disappears.

    4. If the machine is equipped with an auxiliary

    freshgas outlet, activate the switch and check

    the following message appears on the display:

    Display: External freshgasoutlet activated

    De-activate the switch and check the messagedisappears.

    4.2.14 Daily check completed

    1. Set status to STBY.

    2. Go to Alarm menu and select Alarm log

    and then Reset.

    3. Go to Alarm menu and select Set todefault.

    4. Press Cancel alarms to remove old

    alarm text from the display.

    User Manual for SIESTA i TS

    Dameca 4-15

    Caution:When the function test has been

    completed, check that the alarm settingsare suitable for the patient, beforeconnecting the patient to the machine.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    38/114

    User Manual for SIESTA i TS

    4-16 Dameca

    Turning the control wheel counter-clockwisewill decrease the flow.

    Turning it clockwise will increase the flow.

    The flow is changed immediately as the user

    changes the set-flow by turning the control-

    wheel.

    2. Setting the O2percentage (3), or the total

    flow (4), by pressing the appropriate field

    on the main display.

    Then the value is highlighted (blue) and

    the gasflow can be changed by turning the

    control-wheel.

    Turning the control wheel counter-clockwise

    will decrease the flow.

    Turning it clockwise will increase the flow.

    The flow is NOT changed, before enter is

    pressed on the control-wheel.

    The freshgas flow values (1), (3) and (4) on

    main display are all set values.

    The freshgas flow bargraphs (2) on main display

    are all measured values.

    If the set values and the measured values do not

    match over a certain period of time, the machine

    will generate an alarm.

    If an oxygen flow less than 300 ml/min is set, a

    message O2 flow critical low is displayed in

    the alarm and message field (5 on fig. 3-3).

    This message can be deleted by pressing

    cancel alarms.

    Flow (L/min)O2 N2O

    1.00 2.00

    Rotameter

    Freshgas flow(L/min)

    3.00

    O2set %

    33

    4

    3

    2

    1

    4.3 Operation of Electronic GasMixer

    When the selftest is completed the electronicgas mixer (EGM) is ready for use.

    4.3.1 Setting of freshgas flow

    The freshgas flow can be set in two ways:

    1. Setting the flow for each gas, by pressing the

    set-flow value (1) on the main display. Then

    the value is highlighted (blue) and the speci-

    fic gasflow can be changed by turning the

    control-wheel.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    39/114

    User Manual for SIESTA i TS

    Dameca 4-17

    4.3.2 Changing carrier gas

    If the machine is set up with both AIR andN2O as carrier gas for O2, this carrier gas

    can be changed in the setup menu.

    Press the setup menu field (11 on fig. 3-3)

    and the first line of the menu will say

    Carrier gas. Press enter on the control wheel to activate

    the Carrier gas setting. Turn the control wheel to the desired carrier

    gas, and press enter on the control wheel to

    confirm the new setting. The gas flow beside the O2-flow will now

    change to the selected carrier gas. Besides

    the carrier gas flow will change to the last

    known flow setting for this specific gas. Leave the menu by pressing the normal

    screen field.

    Note:The O2percentage is a calculated

    value based upon the measured flowdata, and not a value measured by an

    oxygen analyser. In case of a hospitalpipeline cross connection, incorrectvalues will result.

    The O2percentage in the patient

    breathing system may differ signifi-cantly from the O2percentage of

    the freshgas.

    Note:If the O2, AIR or N2O driving gas

    pressure is lower than 2,5 (x100 kPa)the appropriate alarm is activated.

    Note:

    The displayed values are basedon STPD conditions (StandardTemperature Pressure Dry).

    Note:The N2O supply is cut-off in case ofan oxygen failure. When reducing theO2 flow, the N2O flow will be reduced

    proportionally, securing a minimumof 25% O2 in the freshgas. N2O will

    be cut of at an O2 flow less than100 ml/min.

    Note:The freshgas flow can not be activatedif the machine status is STBY. Changeto MANUAL or VENT in order toopen freshgas flow.

    If the total flow setting or the O2 set percentage

    is changed in such a way that the O2, AIR or

    N2O flow becomes less than 100 ml/min, this

    flow will automatically be set to 0 L/min,and the O2 set percentage will be changed

    accordingly.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    40/114

    User Manual for SIESTA i TS

    4-18 Dameca

    4.3.3 Electronic gas mixer default settings

    When the machine is turned OFF or the FULL selftest or the LC selftest has been performed,

    the settings of the electronic gas mixer will change to default values:

    Parameter Factory set value Limits

    Freshgas flow 0.0 L/min 0.0, 0.3-10.0 L/min

    O2 set % 100% Carrier gas AIR: 21-100%

    Carrier gas N2O: 25-100%

    O2-flow 0.0 L/min 0.0, 0.3-10.0 L/min

    AIR-flow 0.0 L/min 0.0, 0.1-10.0 L/min

    N2O-flow 0.0 L/min 0.0, 0.1-10.0 L/minCarrier gas AIR AIR, N2O

    4.4 Operation of stopwatch

    Stop watch

    00:00:00

    Normal

    screenGas meas. (%)

    3

    d

    Start the stopwatch by pressing the

    stopwatch field.

    The colour of the HH:MM:SS will change

    from grey to green and the stopwatch will

    start to count.

    Stop the stopwatch by pressing the

    stopwatch field again, and the HH:MM:SS

    will change from green to grey and the

    stopwatch will stop counting.

    Reset the stopwatch by pressing the

    stopwatch field twice (double-click).

    This can only be done when the stopwatch

    is stopped.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    41/114

    User Manual for SIESTA i TS

    Dameca 4-19

    4.5 Operation of Ventilator

    Flow (L/min)O2 N2O

    1.00 2.00

    Adult sensor At Y-piece Peak

    17

    Plateau

    17

    PEEP

    3

    Compl. 34 ml/hPa

    hPa

    0

    20

    40

    Cancel

    alarms

    Alarm

    menu

    Vent. mode

    VCV

    Status

    VENT

    Resp. meas.

    MVexp(l)5.8

    Exp

    TV(ml) 480

    RR(1/min)12

    8.04.0

    OFF4

    PSV settings

    30

    TV set

    (ml)

    500

    RR set

    (1/min)

    12

    PEEP

    (hPa)

    OFF

    I:E

    Ratio

    1:2

    P Insp

    (hPa)

    (17) More settings

    When the selftest is completed the machine is

    in stand-by mode, ready for automatic ventilation

    after adjustment of the ventilation parameters.

    The following parameters can be set when the

    machine status is STBY, MANUAL orVENT.

    A. Ventilation mode

    During automatic ventilation the following

    ventilation modes are available:

    VCV (Volume Controlled Ventilation)

    PCV (Pressure Controlled Ventilation)

    SIMV (Synchronized Intermittent Mandatory

    Ventilation) (optional)

    PSV (Pressure Supported Ventilation)

    In VCV mode the set tidal volume is delivered

    to the patient during inspiration by a constant

    inspiratory flow.

    This inspiratory flow is adjusted continously

    by the ventilator in order to deliver the set

    tidal volume to the patient, independent of

    system compliance and actual freshgas flow.

    VCV is also known as CMV or IMV.

    In PCV mode the set inspiratory pressure above

    PEEP is applied to the patient during inspira-

    tion, by a decelerating inspiratory flow.

    This inspiratory flow is adjusted continuouslyby the ventilator in order to reach the set

    inspiratory pressure and hold this pressure du-

    ring the inspiration phase, independent of sy-

    stem compliance and actual freshgas flow.

    If the set inspiratory pressure is not reached

    during inspiration, the ventilator generates a

    PRESSURE LIMIT alarm (see section 7.3.8).

    In SIMV mode the patient can trigger a VCV

    inspiration during the expiration phase by

    applying a decrease of the airway pressure in

    the patient system. In the ventilator settings

    menu the user can define at which decrease

    of the airway pressure a VCV inspiration

    should be triggered. The decrease of the airway

    pressure is relative to the set PEEP. If the patient

    has not triggered an inspiration during the

    expiration phase, the ventilator automatically

    gives a VCV inspiration.

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    42/114

    User Manual for SIESTA i TS

    4-20 Dameca

    The time where the patient can trigger an inspi-

    ration is called trigger-window. This trigger-

    window is calculated when the mode is changed

    to SIMV and is equal to the cycle time (inspira-tion + expiration) of the previous mode. When

    the set resp. rate in SIMV is decreased, to force

    the patient to start spontaneous breathing

    during wake-up, the trigger-window is kept

    at the end of the expiration phase, see picturebelow:

    Triggering is never possible during

    insp.time + 200msec.

    Time, pressure graph in SIMV mode.

    For security reasons the patient can never

    trigger a new inspiration, when the ventilator

    is already performing an inspiration, and just

    after an inspiration has taken place.

    Therefore the trigger-window is limited, so

    that it can never enter the inspiration-time

    plus a 200 msec. period thereafter

    The patient can only trigger one inspiration in

    each trigger window.

    In SIMV mode the tidal volume and inspiration

    time should be the same as it was in VCV mode,

    even when the user prolongs the expiration

    time to force the patient to start spontaneous

    breathing during wake-up.

    Therefore the setting of inspiration time and

    volume is different in SIMV mode.

    Inspiration time:

    The inspiration time in SIMV mode is shown

    beneath the I:E ratio setting.

    If the Resp. rate, or Insp. pause is changed, the

    ventilator automatically calculates a new

    I:E ratio, in order to keep the inspiration

    time the same.

    Example:

    Default settings and Resp.rate is set to 20.

    This gives an inspiration time of 1 sec.

    If SIMV mode is selected and the resp.rate is

    then set to 10 (in order to increase expiration

    time in SIMV), the I:E ratio is automatically

    changed from 1:2 to 1:5, in order to keep the

    inspiration time at 1 sec.

    Trigger-window: triggering po ssible

    Time, pressure graph in previous mode.

    Time, pressure graph in SIMV mode.

    Trigger-window: triggering possible

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    43/114

    User Manual for SIESTA i TS

    Dameca 4-21

    In SIMV mode the I:E ratio can go from 1:1

    to 1:9.9, since inverse I:E ratio is not allowed

    in SIMV mode. Therefore Resp. rate and Insp.

    pause cannot be adjusted to a setting whichgives an I:E ratio outside the 1:1 to 1:9.9

    Example:

    Default settings and resp.rate is set to 20.

    This gives an inspiration time of 1 sec.

    If SIMV mode is selected, the resp.rate can

    be set from 6 resp/min (gives I:E ratio of 1:9.0)

    to 30 resp/min (gives I:E ratio of 1:1).

    If the user wants to change the inspiration timein SIMV mode, select I:E ratio and adjust

    with the control wheel. When the I:E ratio is

    adjusted, the new inspiration time is shown on

    the display. However, this new setting will not

    be effective until the control wheel is pressed.

    Set volume:

    If the ventilator is installed with Minute

    volume as set volume, the tidal volume in

    SIMV mode is shown beneath the MV setsetting.

    If the Resp. rate is changed, the ventilator

    automatically calculates a new minute volume

    setting in order to keep the tidal volume the

    same.

    Example:

    Default settings and Resp. rate is set to 20.

    This gives a tidal volume of 300 ml.

    If SIMV mode is selected, and the resp.rate

    is then set to 10 (in order to increase expiration

    time in SIMV), the set minute volume is

    automatically changed from 6.0 to 3.0, in

    order to keep the tidal volume at 300 ml.

    If the user wants to change the tidal volume in

    SIMV mode, select MV set and adjust with

    the control wheel. When the set minute volume

    is adjusted, the new tidal volume is shown onthe display. However this new setting will not

    be effective until the control wheel is pressed.

    Note:In the service-menu it can be definedwhether SIMV mode shall be availableor not. This definition can be changed

    by a super-user or a technician.

    In PSV mode it is only the patient who controls

    the ventilator, as the ventilator does not give any

    mandatory inspirations to the patient. Therefore

    this mode is for patient with a certain degree of

    spontaneous respiration, where the ventilator

    assists the respiration of the patient.

    Inspiration:

    When the patient starts an inspiration, and the

    inspiratory flow at the Y-piece (measured by the

    flow sensor) is higher than the set Insp. trig.,

    the ventilator starts an inspiration, to reach the

    set Support pressure.

    When this support pressure is reached, the

    ventilator maintains the pressure in the rest of

    the inspiration phase.

    Expiration:

    The inspiration stops and the expiration starts if

    one of the following conditions are met:

    1. The pressure reaches the high alarm

    (same as for the other ventilation modes).

  • 8/10/2019 Siesta i TS 10650E-90 UK Apr08

    44/114

    User Manual for SIESTA i TS

    4-22 Dameca

    2. The inspiration time is more than 2.5 sec.

    (if flow sensor is Adult) or 1.5 sec.

    (if flow sensor is Paed.)

    3. The patient inspiratory flow at the Y-piece

    (measured by the flow sensor) is lower than

    the set Exp. trig.

    The Exp. trig. is set as a % of the measured

    peak flow during inspiration.

    Ventilation backup:

    If the patient does not trigger an inspiration

    before the PSV backup period has ended, a

    PSV BACKUP MODE ACTIVATED,SWITCHING TO PCV alarm starts (see section

    7.3.14) and the ventilator automatically switches

    to PCV mode.

    The alarm disappears when the cancel alarms

    button is pressed.

    B. Tidal volume/Minute volume (active in

    VCV or SIMV mode only)

    The tidal volume/minute volume delivered

    by the ventilator in VCV or SIMV mode isautomatically compensated for the compliance

    of the breathing system and for the f