Page [1/29] SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol software. This report describes the checkpoints to be observed with this release so please take the time to check them out. 2. Outline This new version features the following additions and changes to the functions from Ver.3.51. The CR-2 and CR-2 Plus are the products that will use the new version. [Function additions and changes] • Archiving function added • Past study search function added and changed • Operation of transfer button on past study screen improved • Disease name search function added; disease name input method, registration method, search method • Overlay font size GUI created for transfer images • Transfer image trimming function • White mask function during printing added • Thumbnail vertical display function added • 64-bit version of Windows 7 supported • Chinese display files added Note:The study information and image data saved using this control software is saved only temporarily. This means that the information and data must be transferred to a storage server, exported (DICOM, JPEG) or archived without fail. The internal data (saved temporarily), export data and archive data are not backed up by this control software. Use a software application available on the market, the backup function of the operating system, etc. to back up the data separately. Note:If the control software is installed while Windows is being automatically updated, the installation may fail. Therefore, do not install the software while Windows is being automatically updated. Note:The main unit firmware must be CR-2: Ver.2.0.00 or later, and CR-2 Plus: Ver. 1.0.01 or later. If necessary, upgrade the firmware. For details on how to upgrade the firmware, refer to the CR-2/CR-2 Plus Service Manual. Binder: Rics NM2 Voltage: 100V: 220V: Reported by: Y.Matsushita 110V: 230V: Approved by: T. Watanabe 120V: 240V: Date: 2012/03/06 Medical Equipment Technical Service Dept. Report No. 12-004E

Transcript of SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00...

Page 1: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Page [1/29]

Confidential

SERVICE MANUAL REPORT

Concerning Rics NM2 Ver. 4.00

1. Reason for issuing this reportCanon is releasing Ver.4.00 of its Rics NM2 contr2ol software. This report describes the checkpoints to be observed with this release so please take the time to check them out.

2. OutlineThis new version features the following additions and changes to the functions from Ver.3.51. The CR-2 and CR-2 Plus are the products that will use the new version.

[Function additions and changes]• Archiving function added• Past study search function added and changed• Operation of transfer button on past study screen improved• Disease name search function added; disease name input method, registration method, search

method• Overlay font size GUI created for transfer images• Transfer image trimming function• White mask function during printing added• Thumbnail vertical display function added• 64-bit version of Windows 7 supported• Chinese display files added

Note:The study information and image data saved using this control software is saved only temporarily. This means that the information and data must be transferred to a storage server, exported (DICOM, JPEG) or archived without fail. The internal data (saved temporarily), export data and archive data are not backed up by this control software. Use a software application available on the market, the backup function of the operating system, etc. to back up the data separately.

Note:If the control software is installed while Windows is being automatically updated, the installation may fail. Therefore, do not install the software while Windows is being automatically updated.

Note:The main unit firmware must be CR-2: Ver.2.0.00 or later, and CR-2 Plus: Ver. 1.0.01 or later. If necessary, upgrade the firmware. For details on how to upgrade the firmware, refer to the CR-2/CR-2 Plus Service Manual.

Binder: Rics NM2 Voltage: 100V: 220V:

Reported by: Y.Matsushita 110V: 230V:

Approved by: T. Watanabe 120V: 240V:

Date: 2012/03/06

Medical Equipment Technical Service Dept. Report No. 12-004E

Page 2: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [2/29]

3. Compatible hardware

3.1 Compatible hardware• CPU: Core2 Duo 2.4 GHz or faster• RAM: 2GB or more• Hard disk drive: 3.5-inch HDD with a high speed of 7200 rpm or more• Display and screen resolution: WXGA (1280 x 800 pixels or more)• Screen colors: 24 bits or more• Other requirements: Keyboard, Microsoft mouse or compatible mouse

3.2 Compatible operating systems • Microsoft Windows XP Professional, SP3 (x86) (JP, US)• Microsoft Windows 7 Professional, no SP or SP1 (x86) (JP, US)• Microsoft Windows 7 Professional, no SP or SP1 (x64) (JP, US)

Authority to use operating systemsWindows XP: AdministratorWindows 7: User (standard user)

3.3 Product supported• CR-2

Firmware: Ver.2.0.00 or later (Software cannot be used with previous firmware versions)Connected EOS: Dedicated EOS (50D RETINA, 60D RETINA)

• CR-2 PlusFirmware: Ver.1.0.01 or later (Software cannot be used with previous firmware versions)Connected EOS: Dedicated EOS (60D RETINA)

Note:Concerning the connected EOSDo not use the EOS with a CF card or SD card inserted into it.

3.4 Applications• Microsoft .NET Framework Version 4.0 Client Profile• Microsoft .NET Framework Version 3.5 SP1• SQL Server 2005 Express Edition SP4• Canon Retinal Camera VCP Driver (Version6.5) • Canon Easy-PhotoPrint EX JP/US Ver.2.0.0 or later (not included with the product)

Note:When installing the retinal camera driver (Retinal Camera VCP Driver) in a 64-bit operating system, use the installer located in the folder with the security catalog file (slabvcp.cat).

4. StratificationThis will be announced as soon as a decision has been made.

Page 3: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [3/29]

5. Installation

5.1 Procedure for new installation of Rics NM2 Ver. 4.00* This procedure is to be followed when the CR-2/CR-2 Plus control software is installed for the first

time. Note:The procedure for upgrading Rics is to be followed if an existing version of Rics NM2 has

not been uninstalled or if its installation folder has not been deleted so check the next section.

Note:If the PC used meets the following conditions when the control software is to be installed, Windows Update may trigger the automatic updating of the SQL Server, causing the installation to fail. 1) Automatic updating of Windows and the Microsoft product is enabled. 2) The PC is connected to a network. * When an error has occurred, re-install Rics.

(1) Run “setup.exe” among the files which have been downloaded.

(2) Click the [Next] button.

Page 4: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [4/29](3) Choose the destination location for installation, and click the [Next] button.

Default installation destination \Program Files\Canon\RicsNM2* When installing the software in the 64-bit version of Windows 7, the software will be installed

in “Program Files (x86).”

(4) Select the model used (CR-2 or CR-2 Plus), and click the [Next] button.

(5) Click the [Install] button.

Selecting the model.

Page 5: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [5/29](6) When the installation is completed, click the [Finish] button.

5.2 Procedure for upgrading to Rics NM2 Ver. 4.00Note:Bear in mind the following points before starting the installation procedure.

* Be absolutely sure to back up the Rics NM2 version which is being used. * If there is a study in the transfer error status, proceed with the upgrade after completing

the transfer. * If the application crashes while a study is in progress, start up the control software

again, and after finishing or aborting the study in progress, proceed with the upgrade. * Be absolutely sure to proceed with the upgrade after copying the installer from the CD

or USB memory into the local folder (Desktop, for instance.) Note:The settings established in the Rics version used will be carried over by the new version.

(1) Run “setup.exe” among the files which have been downloaded.

(2) Click the [Next] button.

(3) Choose the destination location for installation, and click the [Next] button. Default installation destination \Program Files\Canon\RicsNM2* When installing the software in the 64-bit version of Windows 7, the software will be installed

in “Program Files (x86).” * After this, the buttons for selecting the model to be used will not be displayed, and operation

will transfer straight to the installation screen.

(4) Click the [Install] button.

(5) When the installation is completed, click the [Finish] button. * The display screens are not shown here because they are the same as the ones shown in “5.1

Procedure for new installation of Rics NM2 Ver. 4.00.”

Page 6: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [6/29]6. Additional functions

6.1 Archiving function addedAn archiving function capable of managing the save destinations and automatically saving the data and of simultaneously searching the internal and archive databases by means of past searches has been added.

[Overview of function and changes made]1. Management, searches and reference• Archives managed using databases• Searches of both internal and archive (external) databases on the Study Logs screen• Study forward-feeding enabled2. Automatic processing (* settings required) • Automatic saving of studies when they have been completed• Background saving (same as for transfer)• Folder partitioning setting when data is saved3. Maximum number of studies: 200,000Note:Set the rights to access the drive set as the archive destination as follows:

• Users group: Changed• Administrators group: Full control

Note:The Save (overwrite) function for saving data cannot be used (the data will be saved separately); data cannot be edited.

[How the data was saved previously]

Rics

DB

Incidental information in

XML files

Image files

Study 1

Data saved in internal memory

Data saved in external memory

Transfer

Page 7: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [7/29][After the archiving function has been added][Diagram showing how archives are saved]* The method used differs depending on the selected settings.

Study 5

RicsSave destination drive setting• HDD• Network drive

DB

Archive management

database

Drive: A(HDD)

Drive: B(HDD)

Incidental information in

XML files

Image files

Study 1

Study 2

Incidental information in

XML files

Image files

Study 3

Study 4

Folder name• Data size: 6-digit running number

and up000001 and up

• Date: YYYYMMDD

Data copied onto reliable media by user

Folder name:000001

Folder name:000002

Files saved automatically or manually (management and storage)

Data written onto media by user (management and storage)

000001 000002

When writing data onto a CD, DVD or other medium and managing it, it is recommended that the folder names be marked on the surface of the medium using an oil-based pen, etc.

Data saved in internal memory Transfer

Archives saved

Page 8: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [8/29]6.1.1 Setting method

Setting procedure 1: Using the archiving functionSettings Menu -> “System” under Modality -> “Transfer/Archive” Note:The default setting is “Automatically (when studies are finished)” for Transfer Studies

only. The archive function is set to OFF.

When the Transfer/Archive tab is clicked, the settings related to transfer and archive are displayed. Whether or not to use transfer or archive and the timing at which the data is to be saved can be set.

• When “Manually” has been selected, archiving is possible if menu tab [Study(S)] -> Archive is selected with the Study Logs opened and the cursor positioned on the target data. When the user logs out, a message indicating all non-archived studies will be archived is displayed.

Page 9: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [9/29]Setting procedure 2: Setting the archive destination Settings Menu -> “Study Management” under Modality -> “Archive”Using the Archive tab, the archive destination settings can be checked using the list and the items can be sorted by column. It is also possible to select the setting with which the archive data to be saved will be saved by folder partitioning or select the reference drive which is referenced for the archive data when the data cannot be found in the archive destination.

*1 [Archive Settings]“Japanese: O English: X” is displayed on the selected line for the settings used as the archive

destination. When the [Settings] button is clicked, the archive destination setting screen is displayed. *2 [Folder Partitioning]When archiving and saving the data, the folder is partitioned and the data is saved. Folder partitioning is set here. Partition Size and Saved date are the two folder partitioning options which can be set. *3 [Reference Drive]Set here is the drive to search the study data next if the study data cannot be found at the archive destination when referencing archived studies.

*1

*2

*3

Page 10: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [10/29]*1 [Archive Settings] -> [Create Archive Settings] screen

Archive Name: Archive destination name (specified by the user)

By clicking the [Create] button, the [Create Archive Settings] screen is displayed.

Select the drive name of the archive destination using the pull-down menu.

Note:A drive and folder on a network can be set as the archive destination if the drive has been allocated as a network drive from My Computer (Windows XP: My Computer; Windows 7: Computer).

Note:It is not possible to set a folder, FD, CD, MO or any other removable drive/folder which was mounted automatically in My Network. Set the folder on the HDD of a local machine.

Page 11: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [11/29]*2 [Folder Partitioning] *3 [Reference Drive]

[Folder Partitioning]* After an archive is created, the “RicsNM2_Archive” folder is created immediately under the drive

which was set. Folders are created with the set “Partition Size” at the “Saved date.”

Partition Size:The folders are partitioned in the partition size selected, and the archive data is saved. The folder names are determined using a number that starts with 1. Numbers ranging from 1 to 999999 are 6-digit numbers padded with zeros (example: 000001). For number 1000000 and up, folders are created with the corresponding numbers.

Saved date:The archive data is saved after partitioning the folders by saved date. The format used for the date is “yyyyMMdd.”

[Reference Drive]Set here is the drive to search the study data next if the study data cannot be found at the archive destination when referencing archived studies. When, for instance, referencing from a CD or DVD media, set the drive for that media here. * It may take some time to search depending on the specified location because all the folders of the

reference drive need to be searched.

“Japanese: O English: X” is displayed on the selected line for the settings used as the archive destination.

Page 12: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [12/29]Checking the setting status: Display the archive informationDisplay the screen which displays the archive information set using the archive settings. The screen can be displayed by selecting [Archive Settings Information] from the Study menu on the Rics GUI screen.

[Search and reference illustration]* The method used differs depending on the selected settings. Note:In the case of a past study which has been archived, the study comparison screen

cannot be opened when referencing the drive. Note:Drive referencing assumes that the usage frequency of data is low. (Studies can be

compared by copying them to an HDD and referencing them externally.)Note:Take care not to disconnect the connection with the memory device with the studies until

the referenced studies are closed. For instance, if studies on an external HDD were referenced, then the connection with the HDD was disconnected, and after this operations were performed on the past study screen which was referenced and opened, an error will result, as the images cannot be found.

For a detailed description of the search screen, check the operating instructions and the next section 6.2 “Integrating the past study search screens.”

Study 5

RicsReference drive setting• CD/DVD

DB

Archive management

database

Drive: A(CD)

Drive: B(DVD)

Incidental information in

XML files

Image files

Study 1

Study 2

Incidental information in

XML files

Image files

Study 3

Study 4

Folder nameFolder name is searched from the reference drive if the data is not found during the referencing.

Folder name:000001

Folder name:000002

Automatic or manual search

000001 000002

Page 13: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [13/29]6.2 Integrating the past study search screensA change has been made to the search function so that it is capable of performing combined searches using a function which directly inputs the search conditions. Previously, the search screen was opened using the search button on the past list screen, but this has now been eliminated. Instead, the search conditions can be input on the Study Logs screen to perform combined searches with the new function.

[Study Logs]The Study Logs is used to specify the search conditions from the studies conducted in the past and select the studies to be referenced. The selected studies are displayed on the past study screen. * For details of how to use the list, refer to the operating instructions.

[Types of past studies] (* Details are presented in the previous section “Archiving function added.”)• Temporarily saved data which was saved inside the control software• Data saved as archives outside the control software (archiving function added starting with V4.0)• Past study files which were saved externally in designated folders

Note:Exposure is prohibited while the Study Logs screen is displayed. Note:During log-in, the last search conditions are retained, but the previous display items and

their positions, sort items and screen display positions are not carried over. Note:The maximum number of search results which can be displayed is 10,000. When there

are more than 10,000 hits, an error icon is displayed, and only the first 10,000 results will be displayed.

Note:The past study screen for study comparisons is also changed in the same way.

What happened previously

After the improvements

Improvement [1]

Improvement [2]

Improvement [1]: Study steps reduced• Integration with search screen• Combined searchesImprovement [2]: Facilitated viewing• Function classification• Creation of menus

Page 14: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [14/29]Note:If the search conditions or search destination (database or past log files) have been

changed, click the search start button.

[Types of past studies]• Temporarily saved data which has been saved inside the control software *1• Data saved as archives outside the control software (archiving function added starting with V4.0)• Past study files which have been saved externally in designated folders *2

*1 Files saved temporarily as before “C:\Program Files\Canon\RicsNM2\Database\Images

*2 Files saved in the designated folders prior to V3.51 These files can also be browsed if the studies were conducted using V4.0 or later and the data is stored in the archives.

Note:Under no circumstances must the names of any folders included in the path along which temporarily saved data has been saved be changed using Windows Explorer or some other external tool. If they are changed, it will no longer be possible to display the images in past studies or re-transfer them.

Note:Under no circumstances must the names of any folders and of image data files included in the path along which data temporarily saved in archives has been saved be changed using Windows Explorer or some other external tool. Furthermore, in terms of operation, it is assumed that the data will be written onto CDs or DVDs so after it has been written, it can be moved to another HDD or deleted. However, the XML files and image files in study folders cannot be moved to folders with different folder names.

Note:Studies saved in archives and existing outside Rics cannot be transferred irrespective of whether they have been moved.

[Search method]Click the past study icon on the display screen of Rics GUI to display the Study Logs.

*1

*2 *3

1. Input the search conditions. Input the search conditions (Patient ID, Patient Name, Study Date, Accession No., Study Description and/or Disease).

2. Select the type of past study. In the case of past study files, specify the folder path of the search destination on the [Folder Reference] dialog box, which appears after the [Folder] button has been clicked. * Not only files directly under the

specified folder but files under the subfolders of the folder are also covered by the search.

3. Perform the search. Click the [Find] button (search start button).

Page 15: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [15/29]6.3 Operation of transfer button on past study screen improvedThis improvement ensures that a study continues to be displayed even when the transfer button on the past study screen has been clicked. Previously, when the transfer button on the past study screen was clicked, what happened was that the study was closed and it was not possible to move and transfer to the next study. However, the added function enables the study concerned to remain displayed and “study forward feeding” to be used even when the button is clicked.

The study concerned remains displayed even when its transfer has been executed.

“Study forward feeding” can be used.

What happened previously

The study is closed when its transfer is executed.

After the improvements

Page 16: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [16/29]6.4 Disease name search function addedUsing a function which registers disease names, it has now become possible to search for disease names on the Study Logs screen. Previously, only patient names, IDs, etc. could be searched, and it was not possible to compare images by means of case searches. However, the added function makes it possible to register the names of diseases in the database while studies are in progress and later to search other patients with the same diseases.

6.4.1 Inputting the names of the diseases[Input method]On the Rics GUI screen, input the disease name. Alternatively, the disease name can be selected on the disease name list which has been registered ahead of time, and added.

①Input the name directly into the disease name field. * A multiple number of names can be input, with

each name separated from the next by a comma.

* Up to 64 characters can be input for one study.

②A disease name can be called from the disease name list which has been registered ahead of time, and added.

(1) Click the [+] button.

(2) When a disease name is selected from the list, and the [Add] button is clicked, the name is added.

* A multiple number of disease names can be added, and each name is automatically separated from the next using commas. However, the number of characters which can be input for one study is limited to 64.

Page 17: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [17/29]Note:In the event that the “\” symbol, control characters or characters which do not

correspond to the specified character sets are included in the 64 characters, an error will be displayed and the selected operation cannot be executed. With new studies, this happens when the study is closed, and with past studies, it happens when an operation by which data is saved is performed (when the close or transfer button is clicked, when the study forward or backward feed button is clicked or when the study comparison button is clicked).

6.4.2 Registering the names of diseasesThe names of diseases can be registered on the disease name list which is used when disease names are input. It is possible to register up to 300 disease names on the disease name list. [Registration method 1] Steps taken to register the disease names on the list from the Settings Menu screen

Input the Disease Name in the disease name field. When the [Add] button is clicked, the name is registered at the bottom of the disease name list. • Up to 16 characters per disease name• Disease names containing commas, the “\” symbol, control characters or characters which do not

correspond to the specified character sets cannot be registered. Characters which are not supported by the ANSI character code are not supported.

• Neither is it possible to register disease names with spaces or disease names which are identical to ones which are already registered.

* If there are spaces at the beginning and/or end of a disease name, the name will be registered with these spaces deleted.

The sequence of the names can be changed using the [ ▲ ] and [ ▼ ] buttons.

Page 18: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [18/29]<Reference>The disease name lists are stored in “DiseaseNameList.ini” which is in Settings of the Rics installation folder. Therefore, it is possible to create disease name lists on the Disease name list registration screen, and then use this ini file to distribute them.

Note:If different language settings are used by the distribution source operating system (of the PC which created the disease name list) and the distribution destination operating system, the characters may be garbled or not displayed.

[Registration method 2] Steps taken to register disease names by directly rewriting the DiseaseNameList.ini file.

Open “DiseaseNameList.ini” in Settings of the Rics installation folder using Notepad, for instance. “[Numbers]=” have already been written in DiseaseNameList.ini so disease names can be registered simply by added them to the numbers. * Disease names cannot be displayed unless they have been

entered in compliance with the rules. <Rules>• “[Disease Name]” must be entered on the first line. • “[Number]=” must be entered in front of the disease name. • Only one number will be displayed for one disease name.

When the same “disease name” is entered more than once: Only the disease name at the top is displayed.When the same “number” is entered more than once: Only the disease name at the bottom is displayed.

• The input rules are described below. (Disease names which do not satisfy these rules will not be displayed.)Disease names including commas (,), the \ symbol, control characters or characters which are not supported by specified character setsDisease names allocated beyond the maximum number (300) of names to be registered (such as 301=)Disease names using 17 or more charactersDisease names without “[Number]=” in frontDisease names consisting of only spaces

• The DiseaseNameList.ini file must not be made a read-only file.

Settings folder DiseaseNameList.ini file

Disease name list (ini file) created ahead of time on the Disease name list registration screen of Rics

Exchange

List of files in the Settings folder of the Rics installation folder used

Disease name list for distribution

Page 19: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [19/29]<Precaution>When a change has been made in the character codes or user authority, for instance, it may not be possible to display the disease names or their characters may be garbled. Alternatively, the file can also be used as the disease list file for distribution purposes as mentioned in <Reference> in the previous section. Note:When the file is to be distributed, the operating system and language settings of the

distribution source and distribution destination must be the same. When “DiseaseNameList.ini” is to be saved, it must be saved using the [ANSI] character code.

OSLanguage

settingCharacter

code

Japanese-language operating

system

Japanese ANSI

English-language operating

system

English ANSI

Chinese ANSI

Greek ANSI

Page 20: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [20/29]6.4.3 Searching the disease namesPast studies can be searched by disease name. The following steps are taken to search a particular study from among the past studies which have been saved.

[Search procedure]Search the study by directly inputting the disease name into the “Disease” field which is one of the search conditions for the Study Logs. Alternatively, if the disease name has been registered ahead of time, it can be added to the “Disease”

field using the [+] button, and searched.

Click the past study icon on the display screen of Rics GUI to display the Study Logs.

Search conducted by adding the disease name to the “Disease” field using the [+] button provided that the name of this disease has been registered ahead of time

Search conducted by directly inputting the disease name into the “Disease” field

Page 21: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [21/29]6.5 Overlay font size GUI created for transfer imagesIt is now possible to select the transfer screen overlay font size using GUI. Previously, the overlay character size was fixed, and the user could not set the size. However, this added function makes it possible to adjust the size appropriately using the storage viewer without superimposing the characters onto the images. Note:The overlay font size setting items differ depending on the overlay items. Note:The font size which has been set is reflected only for the overlay font size of the transfer

images. Note:This function works when overlay annotations has been set to ON for annotation in

Storage settings. [Setting procedure]

Select “Storage” on the Settings menu.

1

2

3

Select “Overlay Settings.”

Select “Font Sizes.”

Page 22: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [22/29][Button control]

Example: The sizes are as follows when annotations are overlaid onto the EOS 60D RETINA L size images at the “M” font size setting. Retinal image size: 5184 x 3456 [pixels]Font size: 3456 x 4.00 / 100 = 138.24[pt]

<Reference diagram for font sizes (L, M, S)>* “M” is the recommended size.

Control Status Description

Font size radio buttons

L8.00[%] with the vertical width of the retinal images used

as the reference

M4.00[%] with the vertical width of the retinal images used

as the reference

S2.00[%] with the vertical width of the retinal images used

as the reference

S

M

L

Page 23: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [23/29]6.6 Transfer image trimming functionTrimming settings have been added to the Storage settings. Previously, the transfer images were relatively small in size because of the large mask area and the retinal images could not be viewed in big enough size. However, the mask area has now been reduced, and the retina size has been increased in relative terms in the transfer images, making it possible for larger images to be viewed on the PC that receives and displays them.[Storage registration and editing]The “Create a New Storage” screen is opened when the create button is clicked on the Storage settings screen. The storage information and image compression method can now be set or edited.

* When a check mark has been placed in “Trimming images”, the type of trimming can be selected. • Default

The images are trimmed based on the mask size regardless of the aspect ratio. • 1:1 aspect ratio

After trimming the images based on the mask size, spaces are added so that the aspect ratio of the images is set to 1:1.

• 4:3 aspect ratioAfter trimming the images based on the mask size, spaces are added so that the aspect ratio of the images is set to 4:3.

Note:The trimming function provides trimming only for images with black masks: It does not provide trimming for x2 images or image capture.

Note:In the case of annotations on the images, the font size must be adjusted.

Select “Storage” under DICOM from the Settings Menu.

Page 24: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [24/29]<Reference diagram>

Trimming OFF

Trimming ON, Default

Page 25: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [25/29]

Trimming ON, 4:3 aspect ratio

Trimming ON, 1:1 aspect ratio

Page 26: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [26/29]6.7 White mask function for printing It is now possible to mask retinal images in white for printing. Previously, masking could be applied only in black when printing images, but the addition of white has made it possible to reduce ink consumption. [Setting procedure]

[Reference: Printing comparison]

Select the other “Print” on the Settings Menu.

Select ON or OFF for “Apply white masking to images” under White Mask.[ON] White masking is applied to the retinal images for printing. [OFF] Default. As in the past, black masking is applied to the retinal images for printing.

[OFF]: Black masking (Default) [ON]: White masking

Photos showing printed retinal images

Page 27: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [27/29]6.8 Thumbnail vertical display function This function displays thumbnails of the main images vertically. The thumbnail display screen is positioned at the left of the screen, and the main display screen appears at the right. If there is a large number of thumbnails, hidden thumbnail images can be displayed by operating the scroll bar at the right of the thumbnail display screen. As a result, a larger main image can be displayed when the thumbnails are displayed vertically because the captured image trimming (L size) function described in section 6.1 is supported. [Setting procedure]

Select “Screen” from the Settings Menu.

“Main Screen” tab of “Screen”Select “Vertical” under Thumbnail Display.

Thumbnails displayed horizontally(Default)

Thumbnails displays vertically

Page 28: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [28/29]6.9 64-bit version of Windows 7 supportedThe Professional 64-bit version of Microsoft Windows 7 has been added as an operating system which is supported. * The operating systems which are supported are described in “3.2 Compatible operating systems”

provided above.

6.10 Chinese display files addedThe setting procedure is described below. Note:The only operating system which is supported is Windows 7. Windows XP is not

supported. Note:As a rule, provide support using the English operating system.

(1) Refer to sections 6.3 and 6.4 of SMR Rics_11-001, and carry out all the actions described.

(2) After completing step (1), open the Language\zh folder from the Rics installation CD.

(3) Copy all the files in the zh folder into the Language folder among the folders in which Rics is installed.

Rics installation CD

Rics installation CDAll files inside the zh folder

Language folder in which Rics is installed in the PC

Copy & paste

Page 29: SERVICE MANUAL REPORT - Canon · SERVICE MANUAL REPORT Confidential Concerning Rics NM2 Ver. 4.00 1. Reason for issuing this report Canon is releasing Ver.4.00 of its Rics NM2 contr2ol

Rics NM2 Report No.12-004E Page [29/29]7. Reference information and service checkpoints

7.1 Data size (Unit: byte) The default settings are indicated by the blue background.

[EOS 50D RETINA] *CR-2

[EOS 60D RETINA] *CR-2/CR-2 Plus

7.2 Correlation between capture size and number of pixels[When JPEG is used as the capture format]

[When RAW is used as the capture format]

7.3 Servicing checkpointsThe “Log Gathering Tool (LogManagement)” is provided starting with Ver.3.2 of Rics NM2 so if a service-related problem has arisen, gather the logs, and submit the details of the problem to Canon Inc. * For details on how to use the “Log Gathering Tool (LogManagement),” refer to “5. Log Gathering

Tool (LogManagement)”/300_Repair Guide/Service Manual of CR-2.

Capture modeJPEG format (fine) Lossless format

L M S L M (S1) S (S2)

Color 5M 3M 1.7M 50M 24M 13M

Capture modeJPEG format (fine) Lossless format

L M S (S1) S2 S3 L M S

Color 5M 3M 2M 1.2M 0.2M 50M 34M 16M

FAF (CR-2 Plus)

5M 3M 2M 1.2M 0.2M 17M 11M 5M

EOSL M S S1 S2 S3

W H W H W H W H W H W H

50D 4752 3168 3456 2304 2352 1568 - - - - - -

60D 5184 3456 3456 2304 - - 2592 1728 1920 1280 720 480

EOSL M S

W H W H W H

50D 4752 3168 3267 2178 2376 1584

60D 5184 3456 3888 2592 2592 1728