Senographe Crystal
Transcript of Senographe Crystal
Senographe CrystalSenographe* Crystal full field digital mammography system provides optimized breast cancer screening and basic diagnostic capabilities, with advanced ergonomic design for the technologist, exceptional patient comfort and seamless workflow connectivity.
Senographe Crystal offers two options on detector configuration, SFOV or LFOV, allowing customers to choose the optimal detector size that adapts to their specific needs.
With enhanced detector performance, the Senographe Crystal offers enhanced image quality to support diagnostic confidence.
Ergonomic design for technologists• Intuitiveuserinterface• Pre-configurableexamprotocols• OnetouchaccesstosetC-armtonextangulationspre-set
in exam protocol, for quick and easy patient positioning•Motorizedmovementsforfastandpreciseoperation
Enhanced patient comfort• Patientfriendlydesign• Easywheelchairaccess• Ergonomicintegratedbucky
Outstanding image quality• EnhancedDetectiveQuantumEfficiency(DQE)• 70μmpixelsizehelpsconfidentlycharacterize micro-calcifications
• AutomaticExposureControl(AEC)transparentlyselects all exposure parameters based on breast radiological properties and breast thickness
GEHealthcare
Image quality
Detector DQE•DQEtypicalvalues:60%at0.5lp/mm• Measurementconditions:Wanodetrack,Rhfilter
Detector•Detectorsize: -LMAM(aka,LFOV):23x30cm -SMAM(aka,SFOV):17x23cm• Pixelsize(pitch):70μm• Acquisitiondynamicrange:14bits• Imagesize(XxY) -LMAM(aka,LFOV):3290x4244pixelsapprox.28MB
per image -SMAM(aka,SFOV):2424x3296pixelsapprox.16MB
per image • PatentedneedlestructureCsIscintillator,singlepiece
construction
Tube technology• X-Raytubetype:I.A.EXM1016T• Anodetargetmaterials–Singletrack:Tungsten(W)• Twofocalspots:0.1and0.3mmoneachtarget perIEC336
• Targetangle:10°/16°degree•Maximalhighvoltage:49kV• Tubecurrent: -80to100mAfrom24to33kVonlargefocalspot -32to40mAfrom26to33kVonsmallfocalspot• Anodesize(tracksdiameter):80mm•Anodeheatstoragecapacity:225kJ(300kHU)• Anodeheatmaximumdissipation:750W•Maxcasingcontinuousdissipation:80W(6.5kHU/s)at40°C• Permanentfiltration:0.5mmBeryllium•Weight:13kg• X-raytubeassembly:self-encasedX-raytube,air-cooled
head• Tubeprotection:softwaremonitoringoftubeload
Grid/breast support• Ergonomicbreastsupportforpatientcomfortandeasy
cleaning• Manualinstallationandremovalofthegridandbreast
support for geometric magnification• Breastsupportmaterial:lowattenuationcarbonfiber
composite•Gridratio:3.5:1• Gridfrequency:41lines/cm• Optimizedgridmotiontohelpensurenogridstructure
is visible in the image• Detectortobreastsupportedge-to-edgedistance≤5mm
Automatic exposureAutomatic Exposure Control (AEC)
Fully automatic mode• AECisafullyautomaticexposuresystemthatselectsall
exposure parameters based on radiological density of the breast and breast thickness for exceptional and consistent imagequality:kV,mAs
Manual mode•Manualselectionofallparameters:kVandmAs
Collimator• Filter:Rhodium:0.05mm• FieldofView(FOV)indetectorplane,incm: -Forstandardcontactviews:23x30@LFOV,
17x23@SFOV -Forgeometricmagnificationviews(optional):SFOVby
default when magnification platform is installed• FieldofView(FOV)selection:manual• FOVsize:selectedmanuallyusingmagneticleadwindow
plate, based on the paddle or geometric magnification platform used
• Lightcenteringdevice:alightautomaticallyswitches on at compression start or it can be turned on with the collimation lamp switch button located on the keypads on both sides of gantry
Compression• Compressionmodes:Motordrivencompressionupto20daN• Dualfoot-pedalsforcolumnheightandcompression
adjustments•Userdefinedmotorizedcompressionforcelimit:5to20daN
Safety features•Gantrylockedwhencompressionforceapplied
Positioner• C-armwithmotorizedrotationandverticalmovement• Sourcetoimagereceptordistance:662mm• Floortoimagereceptordistance:from63.5cmto128.5cm• Rotationangle:±180degrees
User interface• Parametersdisplay -Tubearmsupportrotationangle -Compressedbreastthickness(inmm) -Compressionforce(indaN)• Ergonomiccontrolconsole -Controlsexposure -Providesinformationonsystemstatus -Givesaccesstoadvancedparametersforsystemset-up• Viewnamecanbeeditedatanytime
before the examination is closed
Technical Specifications
Acquisition workstation• Smallfootprint• Dosecalculatedanddisplayedontheimageaftereveryexposure(AverageGlandularDose)
• DualcoreAdvantechworkstation(Inteli3540Processor) -Memory:4GBRAM+0.5MBL2cache+4MBL3cache -Harddisk:1internal1TBdisk,7200RPM -Imagestorage:18,000inSFOVor10,000inLFOV -Ports:oneEthernetport10/100/1000Mbits -CameraLinkvideoboard•Display(standard) -HighResolutionColorLCD2MPmonitor -58.8cm(23”) -1920x1080pixels(landscape) -Luminance:upto250Cd/m2 -Highcontrastratio:1000:1 -Viewingangle:160degrees -Weight:6.8kg(15lb)•ImagePresentation -Postimageprocessingforenhancedlocalcontrast -Automaticwindowing(windowlevelandwindowwidth) -Otherfeatures:zoom,roaming,inversion,flip,rotationof
images, window width and level setting, annotations and measurements
Connectivity•DICOM**3.0platform:- ModalityWorklist- Storage- BasicGrayscalePrint- Verification- DICOM-compliantCD-RDataInterchange
Quality assurance• ComprehensiveQualityControlprogram†(†Qualitycontrolmanualandtoolkit:DetectorCalibrationPMMABlock)
• Datacaneasilybeexportedfordatatracking
Radiation shield• Integratedtothecontrolconsole(height:1850mm)
High voltage generator•Generatortype:highfrequencysingle-phasepowersupply• Power:5kWmax•mAsrange:0.1to320mAs(dependingonkV)• kVrange:20to35kV,in0.5kVsteps• Generatorprotection:softwaremonitoringofgenerator
and tube load
Power supply• Inputfrequency:50/60Hz• Inputvoltage:single-phase220Vto240V
Standard configuration•Motorizedgantry• X-raytubewithrotatingWanode• SFOVorLFOVflatpaneldetector• Acquisitionworkstation -DVD-RW -LCDdisplay -Controlconsole• Pairofdualfoot-pedals•High-frequencygenerator• Faceshield• SFOVorLFOVbuckywithgrid• 17x23(forSFOV)or23x30(forLFOV)contactpaddle•Usermanualandtechnicaldocument•DetectorCalibrationPMMABlock
Options• 1.5and1.8magnificationstand• 12x12magnificationpaddle•Magneticleadwindowplateformagnification• Radiationshield
©2013CopyrightGEHealthcare-HCSDGSBCMC10131DOC1454216–rev.3
GEimaginationatwork
Datasubjecttochange. MarketingCommunicationsGEMedicalSystems.SociétéenCommanditeSimpleaucapitalde85.418.040Euros.283ruedelaMinière–78533BucCedexFranceRCSVersaillesB315013359AGeneralElectriccompany,doingbusinessasGEHealthcareFrance-Paris-Fax:+33(0)130709435Japan-Tokyo-Fax:+81-3-3223-8524Singapore-Fax:+6562917006USA-Milwaukee-Fax:+1-262-521-6123
*TrademarkofGeneralElectricCompany.**TrademarkofNationalElectricalManufacturer’sAssociation.
GE,theGEMonogram,andimaginationatworkaretrademarksoftheGeneralElectricCompany.Weightsanddimensionsshouldbementionedasnon-contractualvaluesbecausetheyvaryslightly with configurations and equipment evolutions.
Physical Specification
Thephysicalspecificationsofeachcomponentareshownasbelow.• Components:Gantryandothers• Packageboxsize:1150(W)x2250(H)x1400(D)mm• Packageboxweight:440kg
Gantry(MainUnit)• Size:650x1108x1999mm
• Weight:250kg
Acquisition Station• Size:705x624x1850mm
• Weight:120kg
C-Arm MovementVerticalTravel
TheentireC-armassembly
can be raised or lowered
on the central stand.
Rotation
TheC-armcanbe
rotated in either
left or right direction.
High Voltage Generator• Size:360xx592xx690mm
• Weight:81kg