    Remote Sensing 29. Remote sensing

    Content

    29.1 Production and use of X-rays

    29.2 Production and use of ultrasound

    29.3 Use of magnetic resonance as an imaging technique

    Learning outcomes

    Candidates should be able to: (a) explain in simple terms the need for remote sensing (non-invasive

    techniques of diagnosis) in medicine

    (b) explain the principles of the production of X-rays by electron

    bombardment of a metal target (c) describe the main features of a modern X-ray tube, including

    control of the intensity and hardness of the X-ray beam

    (d) show an understanding of the use of X-rays in imaging internal

    body structures, including a simple analysis of the causes of

    sharpness and contrast in X-ray imaging 1

  • 7/28/2019 Section VII 29 Remote Sensing

    2/86

    (e) show an understanding of the purpose of computed tomographyor CT scanning

    (f) show an understanding of the principles of CT scanning (g) show an understanding of how the image of an 8-voxel cube can

    be developed using CT scanning

    (h) explain the principles of the generation and detection of ultrasonicwaves using piezo-electric transducers

    (i) explain the main principles behind the use of ultrasound to obtaindiagnostic information about internal structures (j) show an understanding of the meaning of acoustic impedance and

    its importance to the intensity reflection coefficient at a boundary

    (k) recall and solve problems by using the equation I = I0exfor the

    attenuation of X-rays and of ultrasound in matter

    (l) explain the main principles behind the use of magnetic resonanceto obtain diagnostic information about internal structures

    (m) show an understanding of the function of the non-uniformmagnetic field, superimposed on the large constant magnetic field,

    in diagnosis using magnetic resonance.

    2

  • 7/28/2019 Section VII 29 Remote Sensing

    3/86

    Remote Sensing

    Remote sensing is the investigation of an object using equipment that hasno direct contact with the object being investigated

    e.g. an orbiting satellite may be designed so that it can detect small changesin mean sea level. These small changes can then be interpreted to determinethe nature of the rocks under the sea-bed. This investigation enablesinformation to be gathered without actually drilling into the sea-bed

    Medical diagnosis for over 100 years used 2 risky techniques

    Observe the patient externally for fever, pulse-rate, breathing, vomiting,

    skin condition etc. This was part science and part art Carry out investigative invasive surgery which involved a high risk

    causing many patients to die either from trauma of surgery or infection

    Although there is some risk versus benefit with any procedure, nowdiagnostic imaging techniques have been developed that enable externally

    placed equipment to obtain detailed information about internal bodystructures without surgery i.e. non-invasive, from under the skin i.e. a formofremote sensing

    Some of the techniques make use of X-rays

    Ultrasound

    Magnetic resonance imaging(MRI ) 3

  • 7/28/2019 Section VII 29 Remote Sensing

    4/86

    THE DUCK IS DEAD - diagnosis

    A woman brought a very limp duck into a veterinary surgeon. As shelaid her pet on the table, the vet pulled out his stethoscope and listenedto the bird's chest.

    After a moment or two, the vet shook his head sadly and said, "I'msorry, your duck, Cuddles, has passed away."

    The distressed woman wailed, "Are you sure?"

    "Yes, I am sure. The duck is dead," replied the vet.

    "How can you be so sure?" she protested.. "I mean you haven't doneany testing on him or anything. He might just be in a coma orsomething."

    The vet rolled his eyes, turned around and left the room.?????????????

    4

  • 7/28/2019 Section VII 29 Remote Sensing

    5/86

    cont

    He returned a few minutes later with a black Labrador Retriever.As the duck's owner looked on in amazement, the dog stood on hishind legs, put his front paws on the examination table and sniffed theduck from top to bottom. He then looked up at the vet with sad eyesand shook his head.

    The vet patted the dog on the head and took it out of the room.

    !!!!!!!!!!!!!!!!!

    A few minutes later he returned with a cat.

    The cat jumped on the table and also delicately sniffed the bird fromhead to foot. The cat sat back on its haunches, shook its head, meowedsoftly and strolled out of the room.

    5

  • 7/28/2019 Section VII 29 Remote Sensing

    6/86

    Verdict

    The vet looked at the woman and said, "I'm sorry, but as I said, this ismost definitely, 100% certifiably, a dead duck."

    The vet turned to his computer terminal, hit a few keys and produced

    a bill, which he handed to the woman.

    The duck's owner, still in shock, took the bill.

    "$250?" she cried, "$250 just to tell me my duck is dead?"

    6

  • 7/28/2019 Section VII 29 Remote Sensing

    7/86

    Do you know why?

    7

  • 7/28/2019 Section VII 29 Remote Sensing

    8/86

    Verdict justified

    The vet shrugged, "I'm sorry. If you had just taken my word for it, thebill would have been $20, .

    8

  • 7/28/2019 Section VII 29 Remote Sensing

    9/86

    but with the LabReport and the CatScan, it's now $250."

    9

  • 7/28/2019 Section VII 29 Remote Sensing

    10/86

    X-rays

    10

  • 7/28/2019 Section VII 29 Remote Sensing

    11/86

    X-ray machine

    11

  • 7/28/2019 Section VII 29 Remote Sensing

    12/86

    Dental X-ray

    12

  • 7/28/2019 Section VII 29 Remote Sensing

    13/86

    Mobile X-ray machine

    13

  • 7/28/2019 Section VII 29 Remote Sensing

    14/86

    Properties of X-rays

    Produce fluorescence in materials e.g. zinc sulphide Produce latent image (developed to give visible image)

    Penetrate substances opaque to light

    Ionize and excite atoms and molecules

    Have biological effects in living organism

    14

  • 7/28/2019 Section VII 29 Remote Sensing

    15/86

    The production of X-rays

    Whenever a charged particle is accelerated, electromagnetic radiation is emitted

    The frequency of the radiation or emitted photon is proportional to the magnitude

    of the acceleration Whenever high speed electrons are stopped in a metal target, X-ray photons are

    emitted

    X-rays production process: A metal filament(cathode) in an evacuated tube made of material that is opaque to X-rays,

    is heated using a low voltage supply causing electrons to be emitted through the thermioniceffect(the opaque material also reduces background radiation)

    The electrons are then accelerated through a potential difference of between 20 90 kV sothat they have high energy and high speed, but this acceleration is insufficient to cause X-ray radiation to be emitted

    These high energy and high speed electrons are then accelerated towards the target(anode),constructed of metal with a high melting point and high atomic number

    When these electrons are bombarded and strike or are stopped by the target metal, largedecelerations are involved, causing the electrons to lose kinetic energy very rapidly giving

    rise to the emission of X-ray photons known as Bremmstrahlung radiation or brakingradiation(slowing down) This X-ray beam is made to pass out of the tube through a window that is transparent to X-

    rays

    Not all of the energy of the electrons is emitted as X-rays as the majority istransferred as thermal energy in the target rotating metal anode and cooling isnecessary

    15

  • 7/28/2019 Section VII 29 Remote Sensing

    16/86

    Simplified design of an X-ray tube

    16

  • 7/28/2019 Section VII 29 Remote Sensing

    17/86

    X-ray spectrum A typical X-ray spectrum of the variation with wavelength of the intensity

    has 2 distinct components:

    A continuous distribution of wavelengths with a sharp cut-off at theshortest wavelength,0

    Sharp peaks may be observed corresponding to the emission linespectraof the target material and therefore a characteristic of the target

    Continuous X-rays Characteristic X-rays

    0

    17

  • 7/28/2019 Section VII 29 Remote Sensing

    18/86

    Continuous spectrum

    The continuous distribution comes about because the electrons when

    incident on the metal target, will not all have the same decelerations

    but will instead have a wide range of values.

    Since the wavelength of the emitted spectrum is dependent on the

    deceleration, there will be a distribution of wavelengths

    The cut-off wavelength corresponds to an electron that is stopped in

    one collision in the target so that all of its kinetic energy is given upas one X-ray photon

    KE energy is lost in the form of X-ray photons

    Energy of photon depends on how much KE losthence a continuous

    range

    Max energy of photon occurs when all KE of electrons is converted toX-rays.

    18

  • 7/28/2019 Section VII 29 Remote Sensing

    19/86

    Characteristic/discrete X-rays

    Superimposed on the continuous spectrum Produced when an incident electron knocks electrons out of the K-

    shell(lowest shell) of the target atom

    An electron from the L or M shell may

    move into the vacancy in the K-shell, emitting characteristic X-rays

    19

  • 7/28/2019 Section VII 29 Remote Sensing

    20/86

    X-ray spectra

    20

  • 7/28/2019 Section VII 29 Remote Sensing

    21/86

    Discrete X-rays

    M shell

    K shell

    L shell

    K

    K L

    21

  • 7/28/2019 Section VII 29 Remote Sensing

    22/86

    Difference between continuous and characteristic

    spectrum

    Continuous Characteristic

    (i) A continuous range of wavelength A discrete wavelength

    (ii) Produced by loss of KE of incident Produced by electron

    electron transition from higher shellto inner shell

    22

  • 7/28/2019 Section VII 29 Remote Sensing

    23/86

    Quality of beam

    Quality of beamdescribes how penetrating the beam is For monochromatic radiation, the quality is completely described by

    the wavelength

    High quality refers to a very penetrating beam. High quality beam is

    also known as hard X-raywhile low quality is called soft X-ray.

    Quality is specified/measured by HVT (half-value thickness)

    23

  • 7/28/2019 Section VII 29 Remote Sensing

    24/86

    Control of the X-ray beam

    In order that the optimum X-ray image may be obtained, there are 2 factors

    that need to be controlled Hardness is the penetration of the X-ray beam, which determines the

    fraction of the intensity of the incident beam that can penetrate the part of

    the body being X-rayed. In general the shorter the wavelength of the X-rays,

    the greater their penetration

    Intensitythis is the wave power per unit area, and this affects the degreeof blackening of the image

    24

  • 7/28/2019 Section VII 29 Remote Sensing

    25/86

    Hardness/penetration

    The kinetic energy Ek of an electron is equal to the energy gained by the electronwhen it is accelerated from the cathode to the anode i.e.

    Ek = eV where e is the charge of an electron and V the accelerating potentialdifference

    Using E = hc/ at the cut-off wavelength, eV = hc/0hence 0 = hc/(eV)

    i.e. the accelerating potential V thus determines the cut-off wavelength

    The larger the potential difference, the shorter the wavelength

    Therefore the hardness(penetration) of the X-ray beam is controlled byvariation of the accelerating potential difference between the cathode and theanode

    A continuous distribution of wavelengths implies that there will be X-ray

    photons of long wavelengths that would not penetrate the person beinginvestigatedand hence would not contribute towards the X-ray image

    Such X-rays would add to the radiation dose received by the person without

    serving any purpose, so the X-ray beam emerging from the tube frequentlypasses through aluminium filters that absorb these long-wavelength photons

    25

  • 7/28/2019 Section VII 29 Remote Sensing

    26/86

    Intensity

    The intensity of the beamdepends on the number of photons emittedper unit time and hence the number of electrons hitting the metaltarget per unit time

    Since the electrons are produced by thermionic emission, increasingthe heater or filament current in the cathode will increase the rate of

    production of electrons and hence increase the intensity of the X-raybeam

    26

  • 7/28/2019 Section VII 29 Remote Sensing

    27/86

    Example

    The accelerating potential difference between the cathode and the anodeof an X-ray tube is 30 kV. Given that the Planck constant is 6.6 x 10-34 J

    s, the charge on the electron is 1.6 x 10-19 C and the speed of light in free

    space is 3.0 x 108 m s-1, calculate the minimum wavelength of photons in

    the X-ray beam.

    Solution

    For the minimum wavelength,

    Energy gained by electron = energy of photon

    eV = hc/0

    Therefore 0 = 4.1 x 10-11 m

    27

  • 7/28/2019 Section VII 29 Remote Sensing

    28/86

    The X-ray image

    The image is based on the penetration of body parts by X-rays.

    The transmitted X-rays produce a latent image on a photographic film. The

    latent image is developed to give a visible image of the internal organ

    scanned.

    The image is not really an image in the sense of the real image produced by a

    lens

    X-ray is an ionising radiation. It loses energy mainly due to

    (a) absorption and (b) scattering

    The reduction in energy is an exponential decay to the distance traveled.

    When an X-ray beam is incident on the body part of the patient, it can

    penetrate soft tissues(skin, fat, muscleetc) with little loss of intensity

    A photographic film after development will show a dark area corresponding

    to these soft tissues Bone however, causes a greaterattenuation(reduces the intensity by a greater

    extent) than soft tissues and therefore the photographic film will be lighter incolour in areas corresponding to the positions of bones

    The quality of the shadow imageproduced depends on its sharpness andcontrast

    28

  • 7/28/2019 Section VII 29 Remote Sensing

    29/86

    X-ray images

    29

  • 7/28/2019 Section VII 29 Remote Sensing

    30/86

    X-ray images

    30

  • 7/28/2019 Section VII 29 Remote Sensing

    31/86

    X-ray images

    31

  • 7/28/2019 Section VII 29 Remote Sensing

    32/86

    A nebula: Image from Chandrasekhar observatory

    32

  • 7/28/2019 Section VII 29 Remote Sensing

    33/86

    Quality of imaging

    Quality of an X-ray imageis described by sharpnessand contrast Sharpness refers to a clear boundary between different tissues or the

    ease with which the edges of structures can be determined

    A shadow image where the bones and other organs are clearly

    outlined is said to be a sharpimage

    But although an image may be sharp, it may still not be clearly visiblebecause there is little difference in the degree of blackening between

    e.g. bone and surrounding tissue

    An image having a wide range of degrees of blackening is said tohave good contrast

    Contrast refers to different intensity (brightness) in the image ofvarious parts of the internal organ.

    33

  • 7/28/2019 Section VII 29 Remote Sensing

    34/86

    Sharpness

    Image is sharp if the boundary is clearly visible

    A sharp image requires a parallel X-ray beamwhich can be achievedby

    (1) reducing the area of the target anode in the X-ray tube

    (2) limiting the size of the aperture through which the X-ray beampasses

    (3) reducing scattering of the emergent beam

    34

  • 7/28/2019 Section VII 29 Remote Sensing

    35/86

    Sharpness(1) reducing the area of the target anode

    in the X-ray tube

    Secondary/partial shadows/penumbra can cause images to be blur

    The full shadow produces an area that is white on the film

    Where there is no shadow, the image will be black

    In the region of partial shadow or greyness, the image gradually changes from

    white to black

    If the image is to be sharp, this area of greyness must be reduced as muchas possible

    The area of the target anode should be kept to a minimum

    35

  • 7/28/2019 Section VII 29 Remote Sensing

    36/86

    Sharpness(2) limiting the size of the aperture

    through which the X-ray beam passes

    A reduction in the grey area at the edge of the image can also beachieved by limiting the size of the aperture through which the X-ray

    beam passes

    This is achieved by using overlapping metal sheet plates, throughwhich the X-ray beam passes after leaving the tube

    36

  • 7/28/2019 Section VII 29 Remote Sensing

    37/86

    Sharpness - (3) reducing scattering of the emergent

    beam As a result of interactions between photons

    and any substance through which the beampasses(even air), some photons will be

    scattered resulting in loss of sharpness

    These scattered or stray photons may be

    absorbed in a metal lead grid placed in

    front of the photographic film Scattering can also be reduced by reducing

    the distance between patient and film

    37

  • 7/28/2019 Section VII 29 Remote Sensing

    38/86

    Contrast

    Good contrast is achieved when neighbouring

    body organs and tissues absorb the X-rayphotons to very different extents e.g. bone andmuscle

    Not the case e.g. if stomach or blood vessels

    are being investigated

    In such a case to improve contrast, especiallyfor soft tissues, a contrasting mediumis used

    The patient is asked to swallow a solution of

    barium sulphate(barium meal taken orally)

    which is a good absorber of X-ray photons,

    causing the outline of the stomach to show upclearly

    Blood vessels can be made to show up visibly

    by injecting a radio-opaque dye into the

    bloodstream

    38

  • 7/28/2019 Section VII 29 Remote Sensing

    39/86

    Contrast

    Contrast also depends on other factors such as increasing exposure

    time and the use of intensifying cassettes or backing the film with

    fluorescent materials

    39

  • 7/28/2019 Section VII 29 Remote Sensing

    40/86

    The attenuation of X-rays

    When a parallel beam of X-ray photons passes through a medium, absorption

    processes occur that reduce the intensity of the beam

    The intensity is reduced(attenuation) by the same fraction each time thebeam passes through equal thicknesses of the medium no matter what thestarting point is chosen. This thickness of medium is called the half-value-thickness(HVT) and denoted by the symbol x

    The decrease in transmitted intensity is an exponential decrease

    40

  • 7/28/2019 Section VII 29 Remote Sensing

    41/86

    Linear attenuation coefficient

    Consider a parallel beam having an incident intensity I0 and themedium(absorber) having thickness x and the transmitted energy is I

    The transmitted energy is given by the expression

    I = I0 e-x or I = I0exp(-x)

    where is a constant depending on the medium and on the energy of theX-ray photons, known as the linear attenuation coefficient or the linearabsorption coefficientof the medium. The unit is mm-1 or cm-1

    41

  • 7/28/2019 Section VII 29 Remote Sensing

    42/86

    Linear attenuation coefficient and HVT

    For a thickness x (the half-value-thickness) of the medium, the

    intensity I will be equal to I0hence I0 = I0 e

    -x or I0exp(-x)

    which gives x= ln 2 where x is HVT

    In practice, x does not have a precise value as it is constant onlywhen the beam has photons of one energy only.

    Approximate values of linear absorption coefficient for somesubstances are:

    Substance /cm-1

    copper 7

    water 0.3

    fat 0.9

    bone 3

    42

    l

  • 7/28/2019 Section VII 29 Remote Sensing

    43/86

    Example

    The linear absorption coefficient of copper is 0.693 mm-1.

    Calculate:

    (a) the thickness of copper required to reduce the incident intensity by

    50%

    (b) the fraction of the incident intensity of a parallel beam that is

    transmitted through a copper plate of thickness 1.2 cm

    Solution

    (a) using I = I0 e-x or I = I0exp(-x)

    I/I0 = 0.50 = exp(-0.693x)

    ln 0.50 = - 0.693x therefore x= 1.0 mm

    (b) I/I0 = exp(-0.693 x 12)

    I/I0 = 2.4 x 10-4

    43

    C d h (C i )

  • 7/28/2019 Section VII 29 Remote Sensing

    44/86

    Computed tomography(CT scanning)

    The image produced on a photographic plate or film is a shadow or

    flat image, and there is little or no indication of depth i.e. theposition within the body is not apparent. This is a 2-dimensional view

    Soft tissues lying behind structures that are very dense also cannot be

    detected

    Tomography is a technique whereby a 3-dimensional image is

    obtained or constructed by slicingor sectioning the body using aCT scanner through different angles using computer technology andtechniques

    Data from each individual X-ray image and angle of viewing is fedinto a powerful computer enabling a 3-D image of the entire object to

    be reconstructed, which can then be viewed from any angle

    44

  • 7/28/2019 Section VII 29 Remote Sensing

    45/86

    45

    B i i i l f CT CAT

  • 7/28/2019 Section VII 29 Remote Sensing

    46/86

    Basic principles of CT or CAT scan

    Can be illustrated using a simple cube with the aim of producing an image

    of a slice or section through the body from measurements made about its

    axis The section or cube is divided into a series of smaller units called voxels

    which absorbs the X-ray beam to different extents due to its structure

    The intensity transmitted through each voxel alone is given a number

    referred to as a pixel, and these various pixels are built up from

    measurements of the X-ray intensity along different directions through theslice or section

    Operated by using a moving X-ray emitter and detector and a powerful

    computer to store data, reconstruction of these pixels in their correct

    positions is done to display the 3D image of the internal organ being

    scanned

    46

    R i f h li i

  • 7/28/2019 Section VII 29 Remote Sensing

    47/86

    Re-construction of the slice or section

    Suppose a simple cube showing a four-voxel section

    The image of each voxel would have a particular intensity, known asa pixel. The pixels are built up from measurements of X-ray intensitymade along a series of different directions around the section of the

    body

    The number on each voxel is the pixel intensity that is to be

    reproduced Pixel(picture element) is actually a two dimensional unit based on the

    matrix size and the field of view.

    When the CT slice thickness is also factored in, the unit is known as aVoxel, which is a three dimensional unit.

    47

    Detector readings(illustrations - clockwise)

  • 7/28/2019 Section VII 29 Remote Sensing

    48/86

    g ( )

    48

    Fi l tt

  • 7/28/2019 Section VII 29 Remote Sensing

    49/86

    Final pattern

    In order to obtain the original pattern of pixels, two operations must

    be performed.- 1. The background intensity must be removed. The background

    intensity is the total of each set of detector readings. In this case, 14 is

    deducted from each pixel.

    - 2. After deduction of the background, the result must be divided by

    three to allow for the duplication of the views of the section since 4

    sets of readings were taken

    49

  • 7/28/2019 Section VII 29 Remote Sensing

    50/86

    CT cont

    In practice, the image of each section is built up from many small

    pixels, each viewed from many different angles.

    The greater the number of voxels, the better the definition, similar toa digital camera

    In order to build up an image of the whole body, the procedure would

    be repeated for further sections through the body. All the data for all the sections can be stored in the computer memory

    to create a three-dimensional image. Views of the body from different

    angles may constructed

    The collection of the data and its construction into a display on a

    screen requires a powerful computer and complicated softwareprogramming and programs. In fact, the reconstruction of each pixel

    intensity value requires more than one million computations.

    The computer allows for the contrast and brightness of the image to be

    varied so that an optimum image can be obtained50

    Ad t /di d t

  • 7/28/2019 Section VII 29 Remote Sensing

    51/86

    Advantages/disadvantages

    CT scan is expensivebut is indispensable in todays medical practice

    Although historically the images generated were in the axial or

    transverse plane, orthogonal to the long axis of the body

    Modern scanners allow this volume of data to be reformatted in

    various planes or even as volumetric (3D) representations of

    structures. The image of the organ can be viewed in any direction. You can look

    at the organ from the front, top, back, and side. Hence a 3 D image ofthe organ

    Higher resolution compared to Ultrasound

    Excellent for bone imaging

    51

    E i

  • 7/28/2019 Section VII 29 Remote Sensing

    52/86

    Exercise

    1. Compare the image produced during an X-ray investigation and that

    produced in CT scanning

    2. The principles of CT scanning have been understood for some time.

    However scanners could not be developed until large powerful computers

    were made available. By reference to the image produced in a CT scan,

    suggest why such a computer is necessary

    3. A simple object consists of 4 voxels out of 8 voxels (i.e. cube can be thought

    of as 2 slices). The object is scanned from 4 different directions, each at 45to the next. The detector measurements for each individual voxel are summed

    and the result is as shown here. The total of the readings of the detectors in

    any one position is 22. Determine the pattern of the pixels in the voxels

    40

    34

    49

    31

    52

  • 7/28/2019 Section VII 29 Remote Sensing

    53/86

    Ultrasound

    Incident wave

    Reflected wave

    Transmitted wave

    Boundary between media

    53

  • 7/28/2019 Section VII 29 Remote Sensing

    54/86

    Recappiezo-electric effect Piezo-electric devices contain a crystal which can expand and compress

    when external pressure is varied e.g. quartz

    The crystals structure is such that the centre of positive charges coincideswith the centre of negative charges when not stressed

    When expanded, both centres will not coincide.

    When compressed, the centres will be in the opposite direction ascompared to under expansion.

    The separation results in a voltage across the crystal surface and this effectis known as the piezo-electric effect

    54

  • 7/28/2019 Section VII 29 Remote Sensing

    55/86

    Piezo-electric transducer A piezo-electric device is a sensor that detects differences in pressure

    (sound wave)

    Variation in pressure will result in an ac voltage The magnitude of the voltage generated depends on the magnitude of the

    pressure on the crystal and the polarity depends on whether the crystal iscompressed or expanded i.e. whether the pressure is greater than or lessthan the ambient pressure

    A transducer is any device that converts energy from one form to another

    The piezoelectric transducer converts mechanical energy (vibration) intoelectrical energy in the form of ac voltages

    It also can convert electrical voltages back to vibration.

    Hence it acts as a receiver as well as an emitter.

    To detect the voltages, opposite faces of the crystal are coated with a metal

    (silver) and electrical connections are made to these metal films and sincethe voltages are very small they are amplified

    The crystal and its amplifier may be used as a simple microphone forconverting sound signals into electrical signals

    Ultrasound waves may be generated using a piezo-electric crystal such asquartz, as it can convert electrical voltages to vibrations

    55

    Ultrasound

  • 7/28/2019 Section VII 29 Remote Sensing

    56/86

    Ultrasound

    When a potential difference is applied between the electrodes of the crystal,

    an electric field is set up in the crystal which causes forces to act on the ions

    Quartz has a tetrahedral silicate structure with the oxygen ion negativelycharged and the silicon ion positively charged, and as these ions are notheld rigidly in position, they will be displaced slightly when an electric fieldis applied

    The positive ions will be attracted to the negative electrode, and the negative

    ions will be attracted to the positive electrode and depending on the direction

    of the electric field, the crystal will become slightly thinner or thicker

    An ac voltage applied across the electrodes will cause the crystal to vibrate

    with a frequency equal to that of the applied voltage with a small amplitude

    If the frequency of the applied voltage is equal to the natural frequency of

    vibration of the crystal, resonance will occur and the amplitude of vibration

    will be a maximum The dimensions of the crystal can be such that the oscillations are in the

    ultrasound region(> 20 kHz) and this will give rise to ultrasound waves inany medium surrounding the crystal

    In the medical field the ultrasound frequency is in the megahertz region

    Human range of hearing is from 20 Hz to 20,000 Hz56

    The reflection and absorption of ultrasound

  • 7/28/2019 Section VII 29 Remote Sensing

    57/86

    The reflection and absorption of ultrasound

    Ultrasound is typical of many types of waves in that , when it is incident on a

    boundary between 2 media, some of the wave power is reflected and some

    transmitted For a wave of incident intensity I , reflected intensity IR and transmitted

    intensity IT, by conservation of energy,

    I = IR + IT Although for a beam of constant intensity, the sum of the reflected and

    transmitted intensities is constant, their relative magnitudes depends not onlyon the angle of incidenceof the beam on the boundary but also on the mediathemselves

    Hence the relative magnitudes of IRand IT are quantified by reference to thespecific acoustic impedance, Z of each media

    The specific acoustic impedance Z is defined as the product of the density

    of the medium and the speed c of the wave in the medium i.e. Z =c For a wave incident normally on a boundary between 2 media having specific

    acoustic impedances ofZ1 and Z2, the ratio of the reflected intensity IR to theincident intensity I known as the intensity reflection coefficient symbol

    i.e. intensity reflection coefficient = IR/I = (Z2-Z1)2/(Z2+Z1)

    257

  • 7/28/2019 Section VII 29 Remote Sensing

    58/86

    58

    Material Density

    r/kg m3Speed of sound

    c/m s1

    Acoustic impedance

    Z/kg m2

    s1

    air 1.3 330

    blood 1060 1570

    muscle 1075 1590

    bone 1600 4000

    The specific acoustic impedance Z is defined as the product of the density

    of the medium and the speed c of the wave in the medium i.e.

    Z =c

    Typical values of specific acoustic impedance and

  • 7/28/2019 Section VII 29 Remote Sensing

    59/86

    Typical values of specific acoustic impedance and

    speed of ultrasound

    Medium Speed/m s-1 specific acoustic impedance/kg m-2 s-1

    Air 330 430

    Water 1500 1.5 x 106

    Blood 1600 1.6 x 106

    Fat 1500 1.4 x 106

    Muscle 1600 1.7 x 106

    Soft tissue 1600 1.6 x 106

    Bone 4100 5.67.8 x 106

    59

    Example

  • 7/28/2019 Section VII 29 Remote Sensing

    60/86

    Example

    Calculate the intensity reflection coefficient for a parallel beam of ultrasoundincident normally on the boundary between:

    (1) air and soft tissue, specific acoustic impedances of 430 kg m-2 s-1and 1.6 x 106 kg m-2 s-1 respectively

    (2) muscle and bone, specific acoustic impedance of 1.7 x 106 kg m-2 s-1

    and 6.5 x 106 kg m-2 s-1 respectively

    Solution

    Using = IR/I = (Z2-Z1)2/(Z2+Z1)

    2

    (1) = 0.999 almost 1!

    (2) = 0.34

    60

    Linear absorption coefficient

  • 7/28/2019 Section VII 29 Remote Sensing

    61/86

    Linear absorption coefficient

    The intensity reflection coefficient for a boundary between air and soft tissue

    from the last example is almost unity which means that when ultrasound is

    incident on the body, very little ultrasound is transmitted into the body In order to overcome this, it is important that there is no air between the

    transducer and the soft tissue(skin) and this is achieved by using a waterbased jelly whose specific acoustic coefficient is approximately 1.5 x 106kgm-2 s-1

    Once the ultrasound is within the medium, the intensity of the wave is

    reduced by absorption of energy as it passes through the medium

    This causes heating and appropriate frequencies of ultrasound are actually

    used in physiotherapy to assist in sprains and similar injuries

    For a parallel beam, the absorption is approximately exponential and for abeam of ultrasound that is incident normally on a medium of thickness x, the

    transmitted intensity I is related to the incident intensity I0by the expressionI = I0 e

    -kx or I = I0exp(-kx)

    where k is a constant depending on the medium known as the linearabsorption coefficient. The unit is cm-1

    The coefficient k depends not only on the medium but also the frequency of

    the ultrasound61

    Typical values of linear absorption coefficient k of

  • 7/28/2019 Section VII 29 Remote Sensing

    62/86

    Typical values of linear absorption coefficient, k ofultrasound

    Medium Linear absorption coefficient/cm-1

    Water 0.0002

    Bone 0.13

    Muscle 0.23

    Air 1.2

    62

    Example

  • 7/28/2019 Section VII 29 Remote Sensing

    63/86

    Example

    A parallel beam of ultrasound is incident on the surface of a muscle and

    passes through a thickness of 3.5 cm of the muscle. It is then reflected at

    the surface of a bone and returns through the muscle to its surface.Calculate the fraction of the incident intensity that arrives back at the

    surface of the muscle given that the linear absorption coefficient for

    muscle is 0.23 cm-1 and the fraction reflected at bone-muscle interface is

    0.34(from the last example)

    Solution

    The beam passes through a total thickness of 7.0 cm of muscle

    For the attenuation in the muscle,

    using I = I0exp(-kx) = I0exp(-0.23 x 7.0) = 0.20I0Given that the fraction reflected at the bone-muscle interface i.e. is 0.34,

    therefore the fraction received back at surface = 0.34 x 0.20 = 0.068 = 1/15

    63

    Obtaining diagnostic information using ultrasound

  • 7/28/2019 Section VII 29 Remote Sensing

    64/86

    Obtaining diagnostic information using ultrasound

    The ultrasound transducer is placed on the skin with the water based

    jelly excluding any air between the transducer and the skin

    Short pulses of ultrasound are transmitted into the body where theyare partly reflected and partly transmitted at the boundaries between

    media in the body

    The reflected pulses or echoes return to the transducer where they are

    detected and converted into voltage pulses which are amplified and

    processed by electronic circuits to be displayed on a screen oroscilloscope

    The time between the transmitted and reflected pulses givesinformation as to the distance of the boundary from the transducer

    The intensity of the reflected beam gives information as to the

    nature of the boundary

    2 techniques are in common use for the display of an ultrasound scan:

    A-scan

    B-scan

    64

    A-scan

  • 7/28/2019 Section VII 29 Remote Sensing

    65/86

    A-scan

    A short pulse of ultrasound is transmitted into the body through the coupling

    medium

    At each boundary some of the energy of the pulse is transmitted and some isreflected

    The transducer(generator/detector) detects the reflected pulses as it now acts

    as a receiver

    The signal is amplified and displayed on a c.r.o.

    Reflected pulses received at the transducer from deeper in the body tend tohave lower intensity than those reflected from boundaries near the skin

    This is due not only to the energy being absorbed by the various media but

    also on the return of the reflected pulse to the transducer, some of the energy

    of the pulse will again be reflected at intervening boundaries

    To allow for this, echoes received later at the transducer are amplified more

    than those received earlier

    A vertical line is observed on the c.r.o. corresponding to the detection of each

    reflected pulse

    The time-base of the c.r.o. is calibrated so that knowing the speed of the

    ultrasound wave in each medium, the distance between boundaries can be

    determined65

    B-scan

  • 7/28/2019 Section VII 29 Remote Sensing

    66/86

    B-scan

    This consists of a series of A-scans all taken from different angles so

    that a 2-D image can be formed

    The ultrasound probe consisting of a generator/detector, for a B-scandoes not consist of a single crystal, but rather an array of smaller

    crystals each one at a different angle to its neighbours

    The separate signals received from each of the crystals in the probe is

    processed and a pattern of spots is built up to create a 2-D image for

    immediate viewing, photographing or to be stored in computermemory

    66

    Advantages of ultrasound scanning

  • 7/28/2019 Section VII 29 Remote Sensing

    67/86

    Advantages of ultrasound scanning

    Health risk to patient and operator is very much less compared to use

    of X-rays

    The equipment is much more portable and easy to use

    Using higher frequency ultrasound enables greater resolution to be

    obtained i.e. greater details to be seen

    Modern techniques allow for the detection of very low intensity

    reflected pulses, hence boundaries between tissues where there is little

    change in acoustic impedance can be detected

    67

  • 7/28/2019 Section VII 29 Remote Sensing

    68/86

    MRI

    Magnetic resonance imaging

    68

    Proton spin

  • 7/28/2019 Section VII 29 Remote Sensing

    69/86

    Proton spin Many atomic nuclei have a property

    known as spin which causes the

    nuclei to behave as if they were

    small magnets

    Such nuclei have an odd number ofprotons and/or neutrons e.g.hydrogen, carbon, phosphorus

    The proton is not stationary but

    spins about an axis and the spinning

    acts like a tiny current loop which

    generates a magnetic field along the

    spin axis

    The spinning proton behaves like a

    spinning top

    When perfectly balanced, it spins

    about a vertical axis

    69

    Hydrogen atoms

  • 7/28/2019 Section VII 29 Remote Sensing

    70/86

    The magnetic fields of hydrogen atoms are random if there is no external

    magnetic field around. Their individual fields tend to cancel out

    70

    Precession

  • 7/28/2019 Section VII 29 Remote Sensing

    71/86

    ecess o

    When a proton is slightly displaced, it wobbles or precesses about thevertical axis.

    When a external magnetic field Bo

    is applied to these spinning protonnuclei, they tend to line up along the field but this alignment is not perfectand the nuclei rotate about the direction of the magnetic field due to theirspin

    The alignment is either parallel or anti-parallel.

    Parallel alignment: lower energy states, E1

    Anti-parallel alignment: higher energy state, E2 This motion can be modelled as the motion of a top spinning about the

    direction of a gravitational field and this rotation is known as precession

    The spinning about the direction of the magnetic field i.e. the precessabout the external field has a frequency of precession known as theLarmor frequency which depends on the nature of the nucleus and the

    strength of the magnetic field which is the natural frequency of theprecessing proton.

    The frequency is directly proportional to the external magnetic fieldstrength, Bo and is given by f = 42.57 Bo for the hydrogen atom

    For large magnetic fields like 1 T or 2 T, f is of the order of 50 MHz,which is in the radio frequency range. 71

    Precession

  • 7/28/2019 Section VII 29 Remote Sensing

    72/86

    Precession

    72

    Nuclear magnetic resonance(NMR) and magnetic

  • 7/28/2019 Section VII 29 Remote Sensing

    73/86

    g ( ) g

    resonance imaging(MRI)

    If an external stimulus pulse of electromagnetic radiation of radio frequency

    of the same frequency as the Larmor frequency is incident on the precessingnuclei, the nuclei will resonate in phase with each other, absorbing energyand flipping to the higher energy state, E2.

    This frequency is in the radio-frequency(RF) band with a wavelength shorter

    than about 10 cm

    A short time after the incident pulse has ended, the nuclei will return to theirequilibrium state, emitting RF radiation

    This short time between the end of the RF pulse and the re-emitting of the

    radiation is known as the relaxation time

    This entire process of precessing and resonating in phase and emitting RF

    radiation is known as nuclear magnetic resonance

    In practice, there are 2 relaxation processes and it is the time between these 2

    that forms the basis ofmagnetic resonance imaging(MRI)

    Since hydrogen is abundant in body tissues and fluids, hydrogen is the atom

    used in MRI and MRI monitors the concentration of hydrogen nuclei in thebody

    73

    RF stimulus and MRI signal

  • 7/28/2019 Section VII 29 Remote Sensing

    74/86

    RF stimulus and MRI signal A small resultant longitudinal magnetic field MLis developed by the

    protons, parallel to Bo.

    This cannot be measured since it parallel to the large external magneticfield Bo

    Before receiving the RF pulse, there in no resultant transversemagnetic field MTas the protons precess out of phase with each other.(cancel out)

    However, a RF pulse will force them to precess in phase with each

    other. This produces a resultant transverse magnetic field MT which precessesabout the main spinning axis at the same Larmor frequency as the

    protons. This a signal that can be measured.

    According to EM induction law, a changing magnetic field induces ane.m.f in the conductor.

    The resultant MT precessing at 90o

    to the external magnetic field caninduce a small e.m.f. (microvolts) in a detector coil positioned outsidethe patient.

    This forms the MRI signal, the alternating voltage pulse having thesame frequency (the Larmor frequency) as the RF pulse detected.

    Hence, all hydrogen atoms in body act as emitters of RF pulses74

  • 7/28/2019 Section VII 29 Remote Sensing

    75/86

    75

  • 7/28/2019 Section VII 29 Remote Sensing

    76/86

    76

    Principal of the MRI scanner

  • 7/28/2019 Section VII 29 Remote Sensing

    77/86

    The patient is positioned in the scanner between the poles of a large magnet thatproduces a very large uniform magnetic field in excess of 1 Tesla

    This magnetic field causes all the hydrogen nuclei within the person to precess withthe same Larmor frequency

    In order that the hydrogen nuclei in only one small part of the body may bedetected, a non-uniform magnetic field is also applied across the patient

    This non-uniform field is accurately calibrated and results in a different magnitudeof magnetic field strength at each point in the body of the patient

    Since the Larmor frequency is dependent on the strength of the magnetic field,the Larmor frequency will also be different in each part of the patient

    The particular value of the magnetic field strength together with the radio-frequency that is emitted, enables the hydrogen nuclei in the part underinvestigation to be located

    Radio-frequency pulses are produced in coils near the patient which pass into the

    patient The emitted pulses produced as a result of de-excitation of the hydrogen nuclei are

    picked up by the coils

    These signals are processed to construct an image of the number density ofhydrogen atoms in the patient i.e. concentration of hydrogen nuclei

    As the non-uniform magnetic field is changed, atoms in different parts of the

    patients body are detected and displayed as a 3-D image

    77

    Schematic of a MRI scanner

  • 7/28/2019 Section VII 29 Remote Sensing

    78/86

    Schematic of a MRI scanner

    78

    MRI scanner

  • 7/28/2019 Section VII 29 Remote Sensing

    79/86

    MRI scanner

    79

  • 7/28/2019 Section VII 29 Remote Sensing

    80/86

    80

    Slice selection

  • 7/28/2019 Section VII 29 Remote Sensing

    81/86

    Slice selection

    Larmor frequency is proportional to Bo.

    If a slightly different Bo is applied to different sections of the body,

    each section will have its own individual Larmor frequency f.

    Hence, if a pulse of specific RF frequency is sent to the body, only

    those protons having exactly the same precession frequency f will

    resonate. Protons in neighbouring slices will be relatively unaffected

    The variation in the value of external field applied is achieved by using

    a gradient field coil.

    Btotal = Bo + B

    (fixed) (spatially varying)

    81

    MRI Image

  • 7/28/2019 Section VII 29 Remote Sensing

    82/86

    g

    Rf signals from the body are used to build up an image.

    Their origins are located using three perpendicular gradient fields. The signal strength dependson:

    (i) Proton density (hydrogen concentration)

    (ii) tissue type (structure and surroundings)

    (iii) pulse sequence

    (iv) relaxation

    82

    Resolution

  • 7/28/2019 Section VII 29 Remote Sensing

    83/86

    Resolution

    Resolution can be improved using

    (a) large and uniform main magnetic field B

    (b) Large gradient fields

    (c) Small receiver coils close to the body

    83

    Advantages/disadvantages of MRI

  • 7/28/2019 Section VII 29 Remote Sensing

    84/86

    Advantages/disadvantages of MRI

    Advantages

    Safe since no ionising radiations are used

    Any selected plane and orientation can be imaged

    Excellent soft tissue contrast, no contrast agent needed

    Versatile since image varies with many parameters

    Body function and chemistry investigated besides structure

    Disadvantages High capital and running costs

    Image selection and interpretration is complex

    Examination can be claustrophobic, noisy and long

    Hazards with implants, (pacemaker)

    Practical problems associated with large superconducting magnets. Takes time

    For children it may not be possible as the patient has to be very still