Attachment to the decree: BMDW-92.211/0089-I/12/2018 Anton Paar GmbH_17025C Scope of Accreditation of Calibration Laboratory (EN ISO/IEC 17025:2005) Anton Paar GmbH / (ID number: 0630) Valid from 25/09/2018 Fluid and Gas Quantities, Density and Viscosity Density of Liquids No. Measured Quantity // Measurement Range OC 2) Measurement Uncertainty 1) Measurement Conditions Calibrated Device Remarks 1 Density of Liquids // 650 to 1550 kg/m³ 0.02 kg/m³ Calibration by direct comparison with reference liquid Density meters based on the oscillating U-tube principle 2 Density of Liquids // 998 kg/m³ 0.1 kg/m³ Comparison measurement with a calibrated density meter Reference standard The comparison measurement is carried out with a calibrated density meter. The reference standard is intended for internal use only. 1) Best measurement capability according to EA-4/02 for calibrations under laboratory conditions. These are expanded measurement uncertainties with a coverage probability of 95 % and, unless otherwise indicated, with a coverage factor k = 2. Measurement uncertainties without units are relative values in relation to the measurement value unless otherwise mentioned. 2) If marked, on-site calibration is also possible (The measurement uncertainty might be larger as for calibrations done under laboratory conditions). This page is not a binding translation. In case of any ambiguities, the German original text must be used.

Transcript of Scope of Accreditation of Calibration Laboratory (EN ISO ......Scope of Accreditation of Calibration...

Attachment to the decree: BMDW-92.211/0089-I/12/2018 Anton Paar GmbH_17025C

Scope of Accreditation of Calibration Laboratory (EN ISO/IEC 17025:2005)

Anton Paar GmbH / (ID number: 0630)

Valid from 25/09/2018

Fluid and Gas Quantities, Density and Viscosity Density of Liquids

No. Measured Quantity //

Measurement Range OC

2)

Measurement

Uncertainty1)

Measurement Conditions Calibrated Device Remarks

1 Density of Liquids //

650 to 1550 kg/m³

� 0.02 kg/m³ Calibration by direct

comparison with reference

liquid

Density meters based on

the oscillating U-tube

principle

2 Density of Liquids //

998 kg/m³

� 0.1 kg/m³ Comparison measurement with

a calibrated density meter

Reference standard The comparison

measurement is

carried out with a

calibrated density

meter. The reference

standard is intended

for internal use only.

1) Best measurement capability according to EA-4/02 for calibrations under laboratory conditions.

These are expanded measurement uncertainties with a coverage probability of 95 % and, unless otherwise indicated, with a coverage factor k = 2.

Measurement uncertainties without units are relative values in relation to the measurement value unless otherwise mentioned.

2) If marked, on-site calibration is also possible (The measurement uncertainty might be larger as for calibrations done under laboratory conditions).

This page is not a binding translation. In case of any ambiguities, the German original text must be used.

Attachment to the decree: BMDW-92.211/0089-I/12/2018 Anton Paar GmbH_17025C

Scope of Accreditation of Calibration Laboratory (EN ISO/IEC 17025:2005)

Anton Paar GmbH / (ID number: 0630)

Valid from 25/09/2018

Temperature, Humidity and Thermophysical Quantities Temperature

No. Measured Quantity //

Measurement Range OC

2)

Measurement

Uncertainty1)

Measurement Conditions Calibrated Device Remarks

3 Temperature

(fixed point cell) //

0.01 °C

� 1.5 mK Triple point of water (TPW) Standard Platinum

Resistance Thermometer

Fixed point

calibration according

to ITS-90

4 Temperature

(fixed point cell) //

29.7646 °C

� 2 mK Gallium melting point Standard Platinum

Resistance Thermometer

Fixed point

calibration according

to ITS-90

5 Temperature //

0 °C – 200 °C

� 10 mK Calibration bath /

Metrology well

Platinum Resistance

Thermometer

6 Temperature //

0.01 °C – 29.6747 °C

� 2 mK TPW, Gallium Fixed Point Standard Platinum

Resistance Thermometer

Fixed point

calibration with

deviation function

according to ITS-90

7 Temperature //

15 °C – 50 °C

� 15 mK Temperature

measurement chain

Comparison

calibration with

reference

thermometer

This page is not a binding translation. In case of any ambiguities, the German original text must be used.