SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence...

24
Instructions and Parts SaniForce 12:1 Sanitary Pumps For use in sanitary applications to transfer medium to high viscosity fluids. For professional use only. See page 2 for model information, including maximum fluid working pressure. Important Safety Instructions Read all warnings and instructions in this manual. Save these instructions. ti15718a 3A0735N EN

Transcript of SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence...

Page 1: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Instructions and Parts

SaniForce™ 12:1 Sanitary Pumps

For use in sanitary applications to transfer medium to high viscosity fluids. For professional use only.

See page 2 for model information, including maximum fluid working pressure.

Important Safety InstructionsRead all warnings and instructions in this manual. Save these instructions.

ti15718a

3A0735NEN

Page 2: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Models

2 3A0735N

ContentsModels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Pressure Relief Procedure . . . . . . . . . . . . . . . . . 8Flush Before First Use . . . . . . . . . . . . . . . . . . . . 8Adjusting the Pump Speed and Pressure . . . . . . 8Pump Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . 8

Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Flushing Procedure . . . . . . . . . . . . . . . . . . . . . . . 9Cleaning Procedure . . . . . . . . . . . . . . . . . . . . . . 9Tighten Connections . . . . . . . . . . . . . . . . . . . . . . 9

Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . 10

Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Disconnect the Pump . . . . . . . . . . . . . . . . . . . . 11Disassemble the Pump . . . . . . . . . . . . . . . . . . . 11Reassemble the Pump . . . . . . . . . . . . . . . . . . . 13Reconnect the Pump . . . . . . . . . . . . . . . . . . . . . 14

Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Complete Pump Models 24F625

and 24F626 . . . . . . . . . . . . . . . . . . . . . . . . . 16Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Displacement Pump Models 24G761

and 24G762 . . . . . . . . . . . . . . . . . . . . . . . . 18Displacement Pump Models 24G761

and 24G762 . . . . . . . . . . . . . . . . . . . . . . . . 19Kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Performance Chart . . . . . . . . . . . . . . . . . . . . . . . . . 22Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Graco Standard Warranty . . . . . . . . . . . . . . . . . . . 24

ModelsMaximum Air Inlet Pressure: 100 psi (0.7 MPa, 6.9 bar)Maximum Fluid Working Pressure: 1450 psi (10.1 MPa, 100.4 bar)

* For use with viscous, sticky material.

† “Type 3.1” Certified per DS/EN 10204

Pump Model

Displacement Pump Model Pump Type Pump Description Packings

24F625 24G761 Priming Piston Stubby Acetal, PTFE, Nitrile, and UHMWPE24F626 24G762 Priming Piston Stubby, with extra seal* Acetal, PTFE, Nitrile, and UHMWPE24Y567† 24G762 Priming Piston Stubby, with extra seal* Acetal, PTFE, Nitrile, and UHMWPE

All models are compliant with: EC 1935/2004 compliant pumps may be subject to individual national provisions in addition to those specified in the EC regulation. It is the users responsibility to know and follow local laws.

Page 3: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Models

3A0735N 3

Page 4: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Warnings

4 3A0735N

WarningsThe following warnings are for the setup, use, grounding, maintenance, and repair of this equipment. The exclama-tion point symbol alerts you to a general warning and the hazard symbols refer to procedure-specific risks. When these symbols appear in the body of this manual, refer back to these Warnings. Product-specific hazard symbols and warnings not covered in this section may appear throughout the body of this manual where applicable.

WARNINGWARNINGWARNINGWARNING

+

SKIN INJECTION HAZARDHigh-pressure fluid from dispensing device, hose leaks, or ruptured components will pierce skin. This may look like just a cut, but it is a serious injury that can result in amputation. Get immediate surgical treatment.

• Do not point dispensing device at anyone or at any part of the body.

• Do not put your hand over the fluid outlet.

• Do not stop or deflect leaks with your hand, body, glove, or rag.

• Follow the Pressure Relief Procedure when you stop dispensing and before cleaning, checking, or servicing equipment.

• Tighten all fluid connections before operating the equipment.

• Check hoses and couplings daily. Replace worn or damaged parts immediately.

MOVING PARTS HAZARDMoving parts can pinch, cut or amputate fingers and other body parts.

• Keep clear of moving parts.

• Do not operate equipment with protective guards or covers removed.

• Pressurized equipment can start without warning. Before checking, moving, or servicing equipment, follow the Pressure Relief Procedure and disconnect all power sources.

FIRE AND EXPLOSION HAZARDFlammable fumes, such as solvent and paint fumes, in work area can ignite or explode. To help prevent fire and explosion:

• Use equipment only in well ventilated area.

• Eliminate all ignition sources; such as pilot lights, cigarettes, portable electric lamps, and plastic drop cloths (potential static arc).

• Keep work area free of debris, including solvent, rags and gasoline.

• Do not plug or unplug power cords, or turn power or light switches on or off when flammable fumes are present.

• Ground all equipment in the work area. See Grounding instructions.

• Use only grounded hoses.

• Hold gun firmly to side of grounded pail when triggering into pail.

• If there is static sparking or you feel a shock, stop operation immediately. Do not use equipment until you identify and correct the problem.

• Keep a working fire extinguisher in the work area.

Page 5: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Warnings

3A0735N 5

EQUIPMENT MISUSE HAZARDMisuse can cause death or serious injury.

• Do not operate the unit when fatigued or under the influence of drugs or alcohol.

• Do not exceed the maximum working pressure or temperature rating of the lowest rated system component. See Technical Data in all equipment manuals.

• Use fluids and solvents that are compatible with equipment wetted parts. See Technical Data in all equipment manuals. Read fluid and solvent manufacturer’s warnings. For complete information about your material, request MSDS from distributor or retailer.

• Do not leave the work area while equipment is energized or under pressure. Turn off all equipment and follow the Pressure Relief Procedure when equipment is not in use.

• Check equipment daily. Repair or replace worn or damaged parts immediately with genuine manufacturer’s replacement parts only.

• Do not alter or modify equipment.

• Use equipment only for its intended purpose. Call your distributor for information.

• Route hoses and cables away from traffic areas, sharp edges, moving parts, and hot surfaces.

• Do not kink or over bend hoses or use hoses to pull equipment.

• Keep children and animals away from work area.

• Comply with all applicable safety regulations.

SPLATTER HAZARD Hot or toxic fluid can cause serious injury if splashed in the eyes or on skin. During blow off of platen, splatter may occur.

• Use minimum air pressure when removing platen from drum.

TOXIC FLUID OR FUMES HAZARDToxic fluids or fumes can cause serious injury or death if splashed in the eyes or on skin, inhaled, or swallowed.

• Read MSDSs to know the specific hazards of the fluids you are using.

• Store hazardous fluid in approved containers, and dispose of it according to applicable guidelines.

PERSONAL PROTECTIVE EQUIPMENTYou must wear appropriate protective equipment when operating, servicing, or when in the operating area of the equipment to help protect you from serious injury, including eye injury, hearing loss, inhalation of toxic fumes, and burns. This equipment includes but is not limited to:

• Protective eyewear, and hearing protection.

• Respirators, protective clothing, and gloves as recommended by the fluid and solvent manufacturer.

WARNINGWARNINGWARNINGWARNING

Page 6: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Installation

6 3A0735N

Installation

Grounding

Pump: Connect a ground wire (Graco PN 238909) to the ground screw on the bottom cover of the air motor, under the shroud. Connect the other end of the ground wire to a true earth ground.

Air and fluid hoses: use only electrically conductive hoses with a maximum of 500 ft. (150 m) combined hose length to ensure grounding continuity. Check elec-trical resistance of hoses. If total resistance to ground exceeds 25 megohms, replace hose immediately.

Air compressors: follow manufacturer’s recommenda-tions.

Dispense valve: ground through connection to a prop-erly grounded fluid hose and pump.

Material supply container: follow local code.

Container(s) that receive material: follow local code.

Solvent pails used when flushing: follow local code. Use only conductive metal pails, placed on a grounded surface. Do not place the pail on a nonconductive sur-face, such as paper or cardboard, which interrupts grounding continuity.

To maintain grounding continuity when flushing or relieving pressure: hold metal part of the dispense valve firmly to the side of a grounded metal pail, then trigger the valve.

Mounting

Mount the pump on a surface than can support the weight of the pump and accessories, as well as the stress caused during operation. Do not use air or fluid lines to support the pump.

Setup

NOTE: Reference numbers and letters in parentheses in the text refer to the callouts in the figures and the parts drawings.

Accessories are available from Graco. Make certain all accessories are sized and pressure-rated to meet your system requirements.

FIG. 1 is only a guide for selecting and installing system components and accessories. Contact your Graco dis-tributor for assistance in designing a system to suit your particular needs.

Install a bleed-type master air valve (G) close to the pump air inlet (D), to relieve air trapped between it and the air motor.

Install an air filter/regulator (F) in the pump air line, upstream from the bleed valve, to control air inlet pres-sure and to remove harmful dirt and contaminants from your compressed air supply.

The equipment must be grounded. Grounding reduces the risk of static and electric shock by providing an escape wire for the electrical current due to static build up or in the event of a short circuit.

To avoid injury from a falling pump, check the torque on the lift ring (16) and nut (15) before using the lift ring to lift the pump. Torque to 30-36 ft-lb(41-49 N•m).

To avoid contaminating the fluid, pipe the exhaust air to vent outside of the fluid product area, away from people, animals, or food-handling areas.

Page 7: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Installation

3A0735N 7

Install a pump runaway valve (S) in the pump air line to shut off air to the air motor automatically if the pump starts to run too fast.

Install another bleed-type master air valve (G) upstream from all air line accessories and use it to isolate the accessories during cleaning and repair.

On the air drop to the dispense valve (K), install an air regulator (M) to control air pressure to the valve. Install a bleed valve (G) to use as a shutoff when servicing the dispense valve.

Connect air solenoid valves (H) to a timer control (L), and set so the dispense valve (K) will dispense at proper intervals.

Key:

Pump Components (Included)

A Drum Unloader-Mounted Sanitary PumpB Air Exhaust Muffler (may alternately be mounted

remotely, using exhaust hose)C 3/4 npt Exhaust Air OutletD 1/2 npt Air InletE 2 in. Tube Size Flanged Fluid Outlet

System Components/Accessories (sold separately)

F Air Line Filter/RegulatorG Bleed-Type Master Air Valve (required)H Air Solenoid ValveJ Air Line Drain Pipe and ValveK Dispensing NozzleL Pump Ground Wire (required)M Pump Runaway ValveN Storage Vessel

FIG. 1. Typical installation

B

C

G F

J

H

A

E

N

K

Main Air Line

L

M

D

Page 8: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Operation

8 3A0735N

Operation

Pressure Relief Procedure

1. Shut off the air supply to the pump.

2. Close the bleed-type master air valve (required in system).

3. Open the fluid ball valve and/or dispensing valve to relieve fluid pressure.

Flush Before First UseThe sanitary pump was assembled using sanitary lubri-cant on moving parts and was tested in water. Flush the pump thoroughly with an appropriate cleaning solution, and disassemble and sanitize the parts before using the pump. See Flushing Procedure, page 9. Check national, state, and local codes for specific limitations.

Adjusting the Pump Speed and PressureSet pressure regulator to 0 psi. Open the bleed-type master air valve. Adjust the pump air regulator until the pump is running smoothly.

Allow the pump to cycle slowly until all air is pushed out of the lines (the fluid will flow in a steady stream from the fluid outlet) and the pump is primed.

With the air supply turned on, the pump will start when the dispensing valve is opened and stall against pres-sure when the valve is closed. In a circulating system, the pump operates until the air supply is turned off.

If the pump accelerates quickly, or is running too fast, stop the pump immediately and check the fluid supply. If the supply is empty and air has been pumped into the lines, refill the container and prime the pump and lines with fluid. Be sure to eliminate all air from the system.

Pump ShutdownFollow the Pressure Relief Procedure, page 8. Always stop the pump at the bottom of its stroke to prevent fluid from drying on the displacement rod. (The air motor will exhaust at the bottom and top of the stroke.)

NOTICE

Do not expose the air motor to temperatures higher than 120°F (49°C) or the immersed fluid pump to tem-peratures higher than 160°F (71°C). Excessive tem-peratures may damage the pump packings and seals.

Trapped air can cause the pump to cycle unexpectedly, which could result in serious injury from injection, splashing, or moving parts. Relieve pressure when you stop pumping and before cleaning, checking, or servicing equipment. NOTICE

Never allow the pump to run dry of fluid. A dry pump will accelerate to a high speed, possibly damaging itself.

Page 9: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Maintenance

3A0735N 9

Maintenance

Flushing Procedure

NOTE:

• Flush before fluid can dry in the equipment, at the end of the day, before storing, and before repairing equipment.

• Flush at the lowest pressure possible. Check con-nectors for leaks and tighten as necessary.

• Flush with an appropriate cleaning solution.

1. Remove the pump from the fluid container. Operate it at a slow rate to pump out as much fluid as possi-ble.

2. Follow Pressure Relief Procedure, page 8.

3. Place siphon tube in grounded metal pail containing an appropriate cleaning solution.

4. Set pump air regulator to lowest possible fluid pres-sure, and start pump.

5. Run the pump long enough to thoroughly clean the pump and hoses.

6. Follow Pressure Relief Procedure, page 8.

Cleaning ProcedureNOTE: The following instructions are a basic procedure for cleaning a sanitary pump.

• Be sure to follow your national and state sanitary standard codes and local regulations.

• Use appropriate cleaning and disinfecting agents, at intervals appropriate for product processed.

• Follow cleaning product manufacturer’s instructions.

NOTE: The pump must be disassembled to clean it thoroughly.

1. Remove the pump from the fluid container. Operate it at a slow rate to pump out as much fluid as possi-ble.

2. Flush the system thoroughly with an appropriate cleaning solution. See Flushing Procedure,page 9.

3. Follow the Pressure Relief Procedure, page 8.

4. Remove the air and fluid hoses and fittings from the pump.

5. Ram-Mounted Pumps: Loosen the hand screw and lift the upper shroud straight up on the rod.Other Pumps: Remove the upper shroud.

6. Clean thoroughly the surface between the upper and lower shrouds.

7. Disassemble the fluid pump and accessories. See Disassemble the Pump, page 11.

8. Wash all pump parts with an appropriate cleaning solution at the cleaning product manufacturer’s rec-ommended temperature and concentration.

9. Rinse all pump parts again with water and allow them to dry.

10. Inspect all pump parts and reclean if needed.

NOTE: Any damaged rubber parts must be replaced as they could harbor microorganisms that can contaminate the fluid.

11. Immerse all pump parts in an appropriate sanitizer before assembly. Take the pump parts out of the sanitizer one-by-one as needed.

12. Lubricate the moving pump parts and o-rings, pack-ings, and seals with appropriate waterproof sanitary lubricant.

13. Circulate the sanitizing solution through the pump and the system prior to use.

14. Ram-Mounted Pumps: Clean all ram surfaces. Remove and clean the inflatable seal and ram plate. See Manual 3A0591.

Tighten ConnectionsBefore each use, check all hoses for wear or damage. Replace as necessary. Check that all connections are tight and leak-free.

Page 10: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Troubleshooting

10 3A0735N

Troubleshooting1. Follow Pressure Relief Procedure, page 8.

2. Check all possible remedies in the Troubleshooting Chart before disassembling the pump.

Problem Cause Solution

Pump fails to operate. Restricted air line or inadequate air supply.

Clear air line or increase air supply.

Insufficient air pressure; closed or clogged air valves, etc.

Open or clean air valves, etc.

Exhausted fluid supply. Refill fluid supply.

Damaged air motor. Service.

Pump operates, but output low on both strokes.

Restricted air line or inadequate air supply.

Clear air line or increase air supply.

Insufficient air pressure; closed or clogged air valves, etc.

Open or clean air valves, etc.

Exhausted fluid supply. Refill fluid supply.

Obstructed fluid line, valves, dispens-ing valve, etc.

Clear. Relieve pressure and discon-nect fluid line. Turn on air. If pump starts, the fluid line is clogged.

Worn throat packing (103). Replace throat packing.

Damaged cylinder o-ring (123). Replace o-ring.

Pump operates, but output low on down stroke.

Held open or worn fluid inlet valve. Clear or service fluid inlet valve.

Damaged cylinder o-ring (123). Replace o-ring.

Pump operates, but output low on up stroke.

Held open or worn fluid piston or seal (109, 110).

Clear or service fluid piston or seal.

Erratic or accelerated operation. Exhausted fluid supply. Refill fluid supply.

Held open or worn fluid inlet valve. Clear or service fluid inlet valve.

Held open or worn fluid piston or seal (109, 110).

Clear or service fluid piston or seal.

Page 11: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Service

3A0735N 11

Service

Disconnect the Pump

1. Remove the pump from the fluid container. Operate it at a slow rate to pump out as much fluid as possi-ble.

2. Follow the Pressure Relief Procedure, page 8.

3. Remove the air and fluid hoses from the pump. Ram-mounted pumps: leave ram air connected for now.

4. Hold the reducer fitting with a wrench. Use a spanner wrench (T)* to loosen the coupling nut.*A Graco spanner wrench tool (p/n 112887) is avail-able.

5. Lower the coupling nut enough to remove the cou-pling collars.

6. Remove the clamps holding the pump base to the ram or drum. Ram-mounted pumps: Use the ram to lift the air motor. Disconnect air lines.

7. Remove the clamp holding the displacement pump to the air motor. Carry the displacement pump to the bench for service.

Disassemble the PumpNOTE: Pump Repair Kits are available. Purchase the kit separately. See the Parts list on page 19 to select the correct kit for your pump. Kit parts are marked with an *.

1. With the pump on its side, tap the displacement rod (105) with a rubber mallet to drive the priming piston assembly out of the base (102).

2. Hold the valve plate guide (119) with a wrench. Use another wrench to remove the priming piston nut (122).

3. Remove the priming piston (121) and the valve plate (120).

4. Remove the clamp (118) from the base (102). Then remove the base (102). Tap with a rubber mallet to loosen, if needed.

5. Remove the o-ring (123) and the spacer (113). Tap the displacement rod again to drive the displace-ment rod/priming piston rod assembly (105/106) out of the cylinder (101).

6. Hold the rod (106) with a wrench and remove the priming piston guide (119).

Moving parts can pinch, cut or amputate fingers and other body parts. Keep your hands and fingers away from the priming piston during operation and whenever the pump is charged with air.

ti15575a

T*

ti15576a

FIG. 2. Remove piston rod

ti15565a

ti15566a

101123

105/106

113102

119

120122

121

115

118

Page 12: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Service

12 3A0735N

7. Hold the piston rod (106) with a wrench. Use another wrench to loosen the intake valve housing (115), then remove it. If the intake valve housing (115) is stuck to the rod, tap with a plastic rod and mallet.

8. Screw out the packing nut (117) and remove the packing stack (116) from the intake valve housing (115).

9. Remove the valve plate (112). Use wrenches on the dis-placement rod flats (105) and the piston rod flats (106) to dis-connect.

10. Remove the valve plate (108) and the piston (107). Remove the bearings (110) from the piston. (Model 24G761 has two bearings; Model 24G762 has three bearings.)

NOTE: On Displacement Pump Model 24G761, care-fully remove the u-cup seal (109). Use a pick, not a screwdriver, as the packing may deform.

11. Stand cylinder upright. Remove the u-cup seal assembly, which includes the base (103a), insert (103b), o-ring (103d), and u-cup (103c).

12. Refer to the Cleaning Procedure on page 9. Clean the parts and inspect them for wear or damage. Replace them as necessary.FIG. 3. Intake valve housing

FIG. 4. Remove piston and seals

116a116b

116c

116d

117

115 ti15568a

ti15570a

ti15571a

108*109

107

*110

106

FIG. 5. Remove u-cup seal assemblyti15569a

103a

103b*

*103c

*103d

Page 13: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Service

3A0735N 13

Reassemble the PumpNOTE: Any damaged rubber parts must be replaced as they could harbor microorganisms that can contaminate the fluid.

NOTE: Lubricate the o-rings, throat packings, and pis-ton seals with appropriate waterproof sanitary lubricant prior to installation.

1. Install the bearings (110*) and the u-cup seal (109*, 24G761 only) on the piston (107). The lips of the piston u-cup must face up. Install the valve plate (108) and piston assembly on the displacement rod (105). Cutout on piston must face down.

2. Use wrenches on the displacement rod flats and the piston rod flats to reconnect. Torque to 60-70 ft-lb (81-95 N•m).

3. Install the packing stack (116*) into the intake valve housing (115). Maintain the following order: shims (116a), female gland (116b), five v-packings (116c) with the lips facing up, and male gland (116d). Loosely thread packing nut (117) into the valve housing (115).

4. Install the u-cup (103c*) and insert (103b*) into the u-cup seal assembly base (103a). Throat u-cup lips must face down. Then install the o-ring (103d*).

5. Lubricate the displacement rod/piston rod assembly and slide into the cylinder from the top.

6. Install the u-cup assembly (103) in the top of the cyl-inder (101). Reassembly may be difficult. Use steady force until the assembly drops into place.

7. Turn the pump on its side. Use a rubber mallet to tap the displacement rod so the piston rod extends well out of the cylinder.

8. Install the valve plate (112) and the spacer (113) into the bottom of the cylinder. Install the cylinder o-ring (123*).

9. Lubricate the piston rod and the inside of the pack-ing nut (117) and intake valve housing (115). Slide the assembly onto the piston rod, then use two wrenches to tighten the valve housing.

10. Stand the pump upright, placing the cylinder (101) into the base (102). Reattach the clamp to hold the base (102) to the cylinder (101). Replace the pin (125).

11. Turn the pump on its side. Use a rubber mallet to tap the displacement rod so the piston rod extends well out of the cylinder.

12. Hold the displacement rod (105) with a wrench. Screw on the guide (119) and use another wrench to tighten.

13. Install the valve plate (120) and the priming piston (121). Hand tighten the priming piston nut (122).

14. Hold the displacement rod (105) with a wrench. Use another wrench to tighten the priming piston nut (122).

FIG. 6. Intake valve housing

116a116b

116c

116d

117

115 ti15568a

FIG. 7. Valve plate

123*

112113

ti15798a

Page 14: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Service

14 3A0735N

Reconnect the Pump1. Ram-mounted pumps: Set the displacement pump

in place, then attach the clamp. Align the air motor and set it on top of the displacement pump. Attach the clamp. Replace the pin.

2. Other pumps: Slide the assembled displacement pump into position on the air motor pump adapter and reattach the clamp.

3. Hold the motor shaft up with one hand. With your other hand, put the coupling nut (5) on the rod.

4. Put the coupling collars (4) into the coupling nut so large flanges point upward.

5. Gently let the motor shaft drop onto the rod. Use a spanner wrench to tighten the coupling nut securely. A Graco spanner wrench tool (p/n 112887) is avail-able.

6. Connect the remaining air and fluid hoses and the ground wire.

Page 15: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Service

3A0735N 15

Page 16: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Parts

16 3A0735N

Parts

1

2

3

4

5

6

7

8

10

10

9

12

13

1415

16

18

19

20

Apply PTFE tape to threads.1

Hand tighten only.2

Apply sanitary lubricant.3

20

ti16013a

21a

21b

2223

Complete Pump Models 24F625, 24Y567 and 24F626

1

5

3

321b

3

2

21c 3

5

Apply medium-strength (blue) thread locker.

4

4

Apply an appropriate medium-strength thread locker on the lift ring (16) and nut (15) every time it is installed to prevent it from coming loose during operation. Torque to 30-36 ft-lb (41-49 N•m). Do not over tighten.

5

Page 17: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Parts

3A0735N 17

Complete Pump Models 24F625, 24Y567 and 24F626

▲ Replacement Danger and Warning labels, tags, and cards are available at no cost.

Kits

Muffler Kit 16G390

Exhaust Assembly Kit 16G389

Lower Shroud Fastener Kit 16G432

Grommet Kit 16G385

* Order Kit 16G384 for qty. 3 of the piston rod grommet.

Spanner Wrench

Ref Part Description Qty1 24G787 MOTOR, SaniForce; 7.5 in.; see

manual 3A12111

2 DISPLACEMENT PUMP 124G761 Model 24F62524G762 Model 24F626

3 16A939 COUPLER 14 184130 COLLAR, coupling 25 626045 COUPLING 16 16C010 TIE ROD, 13.3 in. (338 mm)

between shoulders3

7 102216 NUT, lock, 5/8-11, sst 38 16G380 SHROUD, upper; includes grom-

mets (Ref. 21)1

9 16G382 SHROUD, lower; includes fasten-ers (Ref. 10) and grommets (Ref. 21)

1

10 118134 SCREW, cap, M8 x 1.25, sst; see Kits

4

12 24G862 FITTING, air inlet, 1/2 npt, Includes Ref. 23

1

13 16C946 FITTING, 3/4 npt 114 165053 O-RING, PTFE 115 16C306 NUT, hand 116 16C009 RING, lift 117▲ 280574 LABEL, warning, not shown 118 512914 MUFFLER; see Kits 119 ----- HOSE, exhaust; see 120 101818 CLAMP, hose; see Kits 221 ----- GROMMET; see Kits 622 16G084 FITTING, air inlet, 1/2 npt 123 166702 O-RING, air inlet, buna-n 1

Ref Part Description Qty

Ref. Part DescriptionQty

.18 512914 MUFFLER, polyethylene 2

Ref. Part DescriptionQty

.18 512914 MUFFLER, polyethylene 119 ----- HOSE, exhaust, 6 ft. 120 101818 CLAMP, hose 2

Ref. Part DescriptionQty

.10 118134 SCREW, cap, M8 x 1.25, sst 4

Ref. Part DescriptionQty

.21a* ----- GROMMET, air motor piston

rod1

21b ----- GROMMET, air fitting 221c ----- GROMMET, tie rod 3

Ref. Part DescriptionQty

.112887 WRENCH, spanner 1

Page 18: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Parts

18 3A0735N

Displacement Pump Models 24G761 and 24G762

101

102

115

116*

117

103b*

103c*

103a

103d*

104

105

106

108

*109

107

*110

112

113

118

118

123*

121

120

119

122

1141

108

107

110* 1

ti15562a

125

125

1

1

5 1

7

1

2

2

1

2

3

6

1

1

1 4

1

Apply appropriate waterproof, sanitary lubricant.1

Model 24G761 has a u-cup (109) and two bearings (110).Model 24G762 has three bearings (110) and no u-cup (109).

2

Piston u-cup lips must face UP.3

V-packing lips must face UP.4

Torque to 60-70 ft-lb (81-95 N•m).5

Cutout on piston must face DOWN.6

Throat u-cup lips must face DOWN.7

6

Page 19: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Parts

3A0735N 19

Displacement Pump Models 24G761 and 24G762

* Parts included in Repair Kit 24G555 (Displacement Pump Model 24G761) or Repair Kit 24G556 (Dis-placement Pump Model 24G762).

Kits

U-Cup Seal Assembly Kit 24C623

U-Cup Seal/O-Ring Kit 24G858

Spanner Wrench

Ref Part Description Qty101 16G430 CYLINDER, pump 1102 16G431 BASE, pump, includes Part 114 1103a ----- BASE, u-cup; see Kits for

U-Cup Seal Assembly1

103b* ----- INSERT; see Kits for U-Cup Seal Assembly

1

103c* ----- U-CUP; see Kits for package of 5 with o-ring (part 103d)

1

103d* ----- O-RING; see Kits for package of 5 with u-cup (part 103c)

1

104 16E465 ADAPTER, tie rod 1105 16G436 ROD, displacement 1106 16G437 ROD, priming piston 1107 181892 PISTON, cylinder 1108 167493 PLATE, piston valve 1109* 108543 U-CUP, UHMWPE, Model

24G761 only1

110* 181897 BEARINGModel 24G761Model 24G762

23

112 16A846 PLATE, intake valve 1

113 16A847 SPACER 1114 167481 SEAT, valve 1115 167482 HOUSING, intake valve seal 1116* ----- STACK, gland/packing 1117 167486 NUT, packing 1118 16G386 CLAMP, assembly 2119 167480 GUIDE, valve plate, pump 1120 167479 PLATE, priming piston valve 1121 172200 PISTON, priming 1122 102533 NUT, plain hex 1123* 16D164 O-RING, 152, nitrile 1125 16F603 PIN, locking, 5/16 in. 2

Ref Part Description Qty

Ref. Part DescriptionQty

.103a ----- BASE 1103b ----- INSERT 1103c ----- U-CUP SEAL 1103d ----- O-RING 1

Ref. Part DescriptionQty

.103c ----- U-CUP SEAL 4103d ----- O-RING 4

Ref. Part DescriptionQty

.112887 WRENCH, spanner 1

Page 20: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Parts

20 3A0735N

Page 21: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Dimensions

3A0735N 21

Dimensions

ti15718a

47.5 in(121 cm)*

21.1 in.(54 cm)*16.1 in.

(41 cm)*

4.1 in.(10.4 cm)

* Add 4.7 in. (12 cm) to allow for full extension of the priming piston rod.

Page 22: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Performance Chart

22 3A0735N

Performance Chart

Cycles per Minute

Fluid Flow gpm (lpm) tested in No. 10 weight oil

Flu

id O

utl

et P

ress

ure

psi

(M

Pa,

bar

)

Air

Flo

w s

cfm

(m

3 /m

in)

KEYA = 100 psi (0.7 MPa, 7 bar)B = 70 psi (0.5 MPa, 5 bar)C = 40 psi (0.3 MPa, 3 bar)

= fluid flow= air consumption

0

A

B

C

A

B

C

1200(8.3, 83)

1000(7.0, 69)

600(4.1, 41)

400(2.8, 28)

2.0(7.6)

6.0(22.7)

4.0(15.1)

120(3.40)

100(2.83)

60(1.70)

20(0.57)

80(2.27)

40(1.13)

0 10 16 30 49

0

800(5.5, 55)

200(1.4, 14)

1400(9.7, 97)

23 36 43

8.0(30.3)

140(3.96)

56

Page 23: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

Technical Data

3A0735N 23

Technical Data

* Sound power at 70 psi (0.48 MPa, 4.8 bar), 20 cpm. Sound power measured per ISO-9614-2.

** Sound pressure was tested 3.28 feet (1 m) from equipment.

Maximum Fluid Working Pressure. . . . . . . . . . . . . . . . . 1450 psi (10.1 MPa, 100.4 bar)

Maximum Air Inlet Pressure. . . . . . . . . . . . . . . . . . . . . . 100 psi (0.7 MPa, 6.9 bar)

Maximum Recommended Pump Speed . . . . . . . . . . . . 60 cycles/min, 8.5 gpm (32 liters/min) delivery

Maximum size pumpable solids 1/4 in. (6.4 mm)

Air Consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . See Performance Chart, page 22

Pump Cycles per Gallon (3.8 Liters) . . . . . . . . . . . . . . . 7.1

Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12:1

Maximum Operating Temperature. . . . . . . . . . . . . . . . . 160°F (71°C)

Maximum Ambient Temperature (air motor) . . . . . . . . . 120°F (49°C)

Air Inlet. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1/2 in. npt(f)

Air Exhaust . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3/4 in. npt(m)

Fluid Inlet Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1 in. (10.4 cm) tube with flange for 6 in. (15.2 cm)

Tri-Clamp®

Fluid Outlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 in. (5.1 cm) Tri-Clamp®

Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122 lb (55.3 kg)

Wetted Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . stainless steel (300 series or other corrosion-resistant grades), Acetal, Buna-N, PTFE, UHMWPE

Sound data

Sound power* . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Sound pressure** . . . . . . . . . . . . . . . . . . . . . . . . . . .

77.2 dBA70.5 dBA

Page 24: SaniForce 12:1 Sanitary Pumps · • Do not operate the unit when fatigued or under the influence of drugs or alcohol. • Do not exceed the maximum working pressure or temperature

All written and visual data contained in this document reflects the latest product information available at the time of publication. Graco reserves the right to make changes at any time without notice.

This manual contains English. MM 3A0735

Graco Headquarters: MinneapolisInternational Offices: Belgium, China, Japan, Korea

GRACO INC. AND SUBSIDIARIES • P.O. BOX 1441 • MINNEAPOLIS MN 55440-1441 • USA

Copyright 2010, Graco Inc. All Graco manufacturing locations are registered to ISO 9001.www.graco.com

Revision N September 2018

Graco Standard WarrantyGraco warrants all equipment referenced in this document which is manufactured by Graco and bearing its name to be free from defects in material and workmanship on the date of sale to the original purchaser for use. With the exception of any special, extended, or limited warranty published by Graco, Graco will, for a period of twelve months from the date of sale, repair or replace any part of the equipment determined by Graco to be defective. This warranty applies only when the equipment is installed, operated and maintained in accordance with Graco’s written recommendations.

This warranty does not cover, and Graco shall not be liable for general wear and tear, or any malfunction, damage or wear caused by faulty installation, misapplication, abrasion, corrosion, inadequate or improper maintenance, negligence, accident, tampering, or substitution of non-Graco component parts. Nor shall Graco be liable for malfunction, damage or wear caused by the incompatibility of Graco equipment with structures, accessories, equipment or materials not supplied by Graco, or the improper design, manufacture, installation, operation or maintenance of structures, accessories, equipment or materials not supplied by Graco.

This warranty is conditioned upon the prepaid return of the equipment claimed to be defective to an authorized Graco distributor for verification of the claimed defect. If the claimed defect is verified, Graco will repair or replace free of charge any defective parts. The equipment will be returned to the original purchaser transportation prepaid. If inspection of the equipment does not disclose any defect in material or workmanship, repairs will be made at a reasonable charge, which charges may include the costs of parts, labor, and transportation.

THIS WARRANTY IS EXCLUSIVE, AND IS IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO WARRANTY OF MERCHANTABILITY OR WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE.

Graco’s sole obligation and buyer’s sole remedy for any breach of warranty shall be as set forth above. The buyer agrees that no other remedy (including, but not limited to, incidental or consequential damages for lost profits, lost sales, injury to person or property, or any other incidental or consequential loss) shall be available. Any action for breach of warranty must be brought within two (2) years of the date of sale.

GRACO MAKES NO WARRANTY, AND DISCLAIMS ALL IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, IN CONNECTION WITH ACCESSORIES, EQUIPMENT, MATERIALS OR COMPONENTS SOLD BUT NOT MANUFACTURED BY GRACO. These items sold, but not manufactured by Graco (such as electric motors, switches, hose, etc.), are subject to the warranty, if any, of their manufacturer. Graco will provide purchaser with reasonable assistance in making any claim for breach of these warranties.

In no event will Graco be liable for indirect, incidental, special or consequential damages resulting from Graco supplying equipment hereunder, or the furnishing, performance, or use of any products or other goods sold hereto, whether due to a breach of contract, breach of warranty, the negligence of Graco, or otherwise.

FOR GRACO CANADA CUSTOMERSThe Parties acknowledge that they have required that the present document, as well as all documents, notices and legal proceedings entered into, given or instituted pursuant hereto or relating directly or indirectly hereto, be drawn up in English. Les parties reconnaissent avoir convenu que la rédaction du présente document sera en Anglais, ainsi que tous documents, avis et procédures judiciaires exécutés, donnés ou intentés, à la suite de ou en rapport, directement ou indirectement, avec les procédures concernées.

Graco InformationFor the latest information about Graco products, visit www.graco.com.

For patent information, see www.graco.com/patents.

TO PLACE AN ORDER, contact your Graco distributor or call to identify the nearest distributor.Phone: 612-623-6921 or Toll Free: 1-800-328-0211 Fax: 612-378-3505