s800NFSA Noise Lna

download s800NFSA Noise Lna

of 24

Transcript of s800NFSA Noise Lna

  • 8/6/2019 s800NFSA Noise Lna

    1/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    1

    Agilent ESA Series Spectrum Analyzers

    Option 219 -Noise Figure Measurements Personality and Hardware

    Overview:

    A key measurement in the development of devices and systems is its noise figure; considering

    that this figure of merit is one of the key limiting factors in the overall systems performance. The

    ability to make accurate noise figure measurements have significant financial benefits since aproduct with a guaranteed low noise figure commands a premium price. A manufacturer can

    however only claim this reward, if every unit manufactured can be shown to meet its

    specification.

    Noise Figure Measurements are required in production to essentially demonstrate thatthe product meets its specification needs.

    For the upstream customer interested in the low noise figure component; perhaps the system

    integrator who is putting together a receiver, noise figure measurements is also of interest. Theywant to confirm that they are getting the performance that they have paid a premium for, and itwill indeed meet their design needs.

    Noise Figure Measurements are required downstream for customers to confirm thatthey are getting the performance that they have paid for.

    The engineer also has a choice to ignore this important parameter and instead over design the

    other aspects of their communications link.

    a) They can raise the transmitted signal power. This usually translates to a more costlydesign in terms of engineering time spent, or higher rated components. For the case of a

    satellite, where the transmitter needs to generate large wattages at the frequenciesrequired, amplifiers fabricated via the regular chip process are no longer adequate. The

    adoption of high power amplifiers like klystrons or magnetrons are required which are

    orders of magnitude larger than the regular microchip PAs (BJTs, FETs). This translates

    to a huge investment in paying the Boeings, Ariannes, Lockheed Martins, Lorals or

    Northop Grummans of this world to send the larger payload into space.

    b) Another way is to increase the amount of power, that the receive antennas intercepts.This translates to a larger receive antenna aperture size which after a while becomes

    impractical because of environmental issues and other government regulations.

    Measuring and then minimizing the noise generated in the components of your communications

    system, is the alternative and recommended approach. Particularly, considering that approacheslike amplification and improving directivity has the same effect on the signal as on the noise.

    Agilents noise figure measurement systems are an easy to use set of tools that automate

    these measurements, providing high accuracy to meet many of your customers needs at avastly reduced cost.

  • 8/6/2019 s800NFSA Noise Lna

    2/24

  • 8/6/2019 s800NFSA Noise Lna

    3/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    3

    Fundamentals of noise figure measurements: There is not enough time to go through the intricacies of a

    noise figure measurement so only a summary of equations relevant to the understanding of how a

    measurement is made will be shown.

    Fig 1: Theory refresher 1

    DUTN

    1=kGB(T

    c+T

    e)

    N2=kGB(Th+Te)

    GZs@TeZs@Th

    Zs@TcY FACTOR

    Y=1

    2

    N

    N=

    NaKTcBG

    NaKThBG

    +

    +=

    KTeBGKTcBG

    KTeBGKThBG

    +

    +=

    )(

    )(

    TeTcKBG

    TeThKBG

    +

    +=

    )(

    )(

    TeTc

    TeTh

    +

    +

    Y=mX+C

    YTc+YTe=Th+Te Te(Y-1)=Th Ytc

    Te= 1

    Y

    YTcTh

    (1)

    Na =Added Noise generatedin D.U.T.

    N1 = Noise Power levelwith noise source on

    N2 = Noise Power levelwith noise source off

    Output Noise Power

    N1

    Na

    N2

    TsThTcTe

    Temperature of

    Source Impedance

    F (Noise Figure) =

    GKTsB

    NaGKTsB

    GNi

    NaGNi

    )GNiNa(GSi

    NiSi

    NoSo

    NiSi

    +=

    +=

    +

    =

    F (Noise Figure) =Ts

    )TeTs(

    GKTsB

    )TeTs(GKB

    GKTsB

    GKTeBGKTsB +=

    +=

    +

    F (Noise Figure) =Ts

    )TeTs( + FTs=Ts+Te FTs-Ts=Te

    Te=Ts(F-1) (2) from (1) Ts(F-1) =1

    Y

    YTcTh

    Assume Ts=Tc TcF= Tc+1

    Y

    YTcTh

    F = 1 +1Y

    TcTcY

    TcTh

    =1Y

    TcTcY

    TcTh1Y

    +

    Rearranging F = 1 +1Y

    TcTcY

    TcTh

    =

    1Y

    )1Tc

    Tc(Y)1Tc

    Th(

    F =1Y

    )Tc

    TcTc(Y)

    Tc

    TcTh(

    F =1Y

    )Tc

    TcTh(

    ENR=)

    Tc

    TcTh(

    F = 1Y

    ENR

    NF(dB) = 10Log (F) = 10Log ENR 10 Log (Y-1) (3)

    So If we measure the Value of Y and we Tell the analyzer what the ENR is usually provided with

    the noise source, then we should be able to measure the noise figure of the device under test.

  • 8/6/2019 s800NFSA Noise Lna

    4/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    4

    Lab 0. Setting up ESA

    Switch to the noise figure measurement personality (Opt. 219).

    Spectrum analyzers can make many different types of measurements. The noise figure personality is only

    one of many modes that the ESA-E series can be operated in. This makes it a cost-effective way to expandthe capability of an essential engineering tool. If the ESA is not already in thenoise figure measurement

    personality mode..

    Instructions: ESA-E series SpectrumAnalyzer

    Keystrokes: ESA-E series SpectrumAnalyzer

    Switch to the Noise figure measurement

    personality.

    [Preset][Mode]({More 1 of 2} if necessary){Noise

    Figure}

    Lab 1. Entering the Excess Noise Ratio (ENR) Values

    All demonstrations use the Agilent N4002A SNS Series smart noise source, mini-circuits mixer, amplifier

    and 70 MHz band pass filter provided in the Demo accessory kit housed in the front panel of the ESA.

    Note (1): In the following keystrokes, {} = soft key and [ ] = hard key.

    Note (2): Optional settings are in smaller Italic font for your information.

    Lets proceed with providing one of the sources of data required for the Noise Figure Calculation shown in

    equation 1, The ENR Data.

    Entering the ENR table for a noise source manually or automatically:

    The noise source used for this demonstration is the N4002A smart noise source. This noise source has a

    calibrated range of 10 MHz to 26.5 GHz. The SNS series broadband noise sources work with the ESA-E

    series to simplify measurement set-up and improve accuracy. Only available with Agilent instruments, they

    provide the following advantages.

    Automatic download of Excess Noise Ratio (ENR) data to the ESA, speeding overall setuptime.

    Electronic storage of ENR calibration data which all but eliminates the opportunity for user

    error Automatic sensing of the ambient temperature of the measurement environment allowing the

    ESA to compensate for these changes during the measurement cycle increasing the accuracy and

    reliability of noise figure measurements.

    Agilent is always sensitive to conservation of investment and so also provides an interface [+28V (pulsed)

    noise source drive output] for the large installed base of 346 series noise sources. These noise sources do

    not operate automatically as do the SNS Series, however, they are available with disks containing the noise

    source ENR data which facilitates the process.

    Figure 2: N4002A SNS Series

    Noise Source available with

    Agilent instruments only

  • 8/6/2019 s800NFSA Noise Lna

    5/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    5

    The ESA allows you to set the preference for which noise source drive output you want to use. Once

    calibration data is entered into the measurement personality, system calibration and DUT measurements

    can be made. In most cases a common ENR table can be used for calibration and measurements, however,

    in the case of mixers, for example, the frequency range of the source for measurements may be outside the

    range for calibration, and therefore two sources are required. This is an instance where the calibration ENR

    table will be different from the measurement ENR table, so the common table function is turned off.

    Step 1: Preferred step with Smart noise source.

    Instructions: ESA-E series Spectrum

    Analyzer

    Keystrokes: ESA-E series Spectrum

    Analyzer

    Connect the SNS to the ESA using the 11730A

    cable.

    No key presses are required for this step.

    It is important to note, though that the SNS

    should not be disconnected while the upload

    process is ongoing

    Automatic upload of ENR data from SNS Series

    noise source

    [Meas Setup]{ENR}{SNS Setup}{Preferences

    Norm SNS} Toggles to SNS if on Norm.

    {Auto Load ENR} On Off Toggles to On if Off.

    Verify that the data has correctly transferred over [Return]{Meas & Cal Table}

    Simply connect the N4002A smart noise source to the ESA noise source output. using a 11730A cable to

    automatically transfer the ENR data to the NFA. This simplifies the process and reduces the possibility

    of user error due to incorrect entry.

    Figure 3: Automatic upload of ENR data from SNS EEPROM

  • 8/6/2019 s800NFSA Noise Lna

    6/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    6

    Figure 4: Common ENR Table with ENR data for N4002A smart noise source

    If the noise source preference was a normal 346 Series noise source, then this is the same interface that

    you could have used to either enter the ENR data. The 346 Series come with a disk containing the noise

    source ENR data. Load it as follows.

    Instructions: ESA-E series Spectrum Analyzer Keystrokes: ESA-E series Spectrum Analyzer

    Load the ENR numbers from disk

    Change directory to A:\ if not already selected

    [File]{Load}{Type}{More 1 of 3}{ENR

    Meas/Common Table}

    {Dir Select}, use [] or [] arrows to select driveA for the floppy then press {Dir Select}.

    Highlight the ENR file name using [] or []andthen press {Load Now}

    If you misplace the disks, you can enter the ENR

    values manually from annotation on the Noise

    source as follows. Add Excess Noise Ratio (ENR)

    serial number and model number

    {Meas Setup}, {ENR}, {Meas & Cal Table}

    [Enter]{Serial #}.

    Use the numeric pad and alpha editor to enter the

    serial number

    {model ID} and enter the model number using the

    alpha editor and numeric key pad

    Adding ENR values versus frequency

    The table auto sorts by frequency

    Press {Index} 1, {Frequency} 10 {MHz}, {ENR

    Value}, 13.14 {dB}. Repeat the process for index 2

    and so on.

    Instructions: ESA-E series Spectrum

    Analyzer

    Keystrokes: ESA-E series Spectrum

    AnalyzerSaving the calibration data to a floppy or the

    internal memory of the ESAPress [File], {Save}, {Dir Select}, use [] or [] toselect the drive A for the floppy, then press {Dir

    Select}.

    Press {Type}{More 1 of 3}{ENR Meas/Common

    Table}

    Press {Name} and use the Alpha Editor to name the

    file. When finished entering the name, press

    [Return] and {Save Now}.

  • 8/6/2019 s800NFSA Noise Lna

    7/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    7

    Noise figure measurement process summary

    Measuring the noise figure of a device also requires knowledge of the measurement

    system. Once the noise figure of the measurement instrument is known and the gain of

    the device under test (DUT) is known, then the noise figure of DUT can be calculated,

    after the overall noise figure is measured.1. Enter the excess noise ratio ENR values in dB of the noise source. (Done)2. Calibrate the measurement personality.3. Make the Noise figure measurement using the Y-Factor method described above

    in equation 3.

    Model the measurement system as another stage with its own noise figure and gain

    So during calibration we make a measurement on just Stage 2, which is the measurement system connected

    to the D.U.T. This will allow us to measure F2 as in equation 3. After that, we also measure F12 , which is

    the noise figure of the cascade. Finally, we need to determine the gain of the D.U.T. which will allow us to

    report the noise figure of the D.U.T., by itself. This is done as described infigure 8: Theory refresh 3.First

    lets proceed with the calibration .

    F12 (Noise Figure)kToBGG

    GkToBGGNaNakToBGG

    No

    NoSiGG

    kToB

    Si

    NoSo

    Ni

    Si

    21

    21212

    2121

    ++===

    Na1 = kTe1G1B =kTo(F1-1)G1B and Na2 = kTe2G2B =kTo(F2-1)G2 B from equation 2

    F12 (Noise Figure) =kToBGG

    GkToBGGB1)G-kTo(FB1)G-kTo(F

    21

    2121122 ++ =kToBGG

    11F)G

    1-F(GkToBG

    21

    11

    221

    ++

    F12 (Noise Figure) = 11

    2 F)G

    1-F( +

    1

    21

    G

    1-FF + (4)

    Hence F1 (Noise Figure) =

    1

    212

    G

    1-FF (5) Fig 5: Theory refresher 2

    kToBkToBG1

    Na1

    kToBG1G2

    B G1 Na1

    Input

    Noise

    kToB

    B, G2,Na2

    Si SiG1 SiG1G2F1 F2

    Na1G2

    Na2Total

    Noise

    Added

    Total

    Noise

    Power

    Out ut

    F12

    Sta e 1

    Stage 2

  • 8/6/2019 s800NFSA Noise Lna

    8/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    8

    Lab 2: Calibration of the noise figure measurement personality and HW (Opt. 219)

    For accurate and correct measurements of noise figure for a DUT, the system must first be calibrated.

    Calibration is required because the measurement system has its inherent noise figure that must be known

    and corrected for before a DUT can be measured as shown above.

    Following is the calibration process:

    1: Select the frequency range appropriate for the D.U.T.

    2: Set the number of points and set the number of averages.Any jitter in the calibration step will add to the

    measurement uncertainty of all subsequent measurements. Therefore a long averaging time should be

    used for calibration in order to reduce this source of uncertainty to a negligible level.3: If the device under test does not have gain or if the gain is low then it is recommended to have the built-

    in preamplifier on before calibration.

    Now Perform a system calibration

    Instructions: ESA-E series Spectrum Analyzer Keystrokes: ESA-E series SpectrumAnalyzer

    Access the DUT Setup diagram to obtain guidance onhow to setup connections for calibration of an amplifier

    as a DUT.Press the tab keys to navigate your way around the form.

    When the form Highlights the diagram field"Blue" you

    should use the softkey to change the parameter.

    [Mode Setup]{DUT Setup.}

    {Calibration}

    Connect the SNS to the ESA input connector as described

    by the diagram for an amplifier

    No key presses are required for this step.

    Instructions: ESA-E series Spectrum

    Analyzer

    Keystrokes: ESA-E series Spectrum

    Analyzer

    Set the start frequency [Frequency] {Start Freq}, 10 MHz

    Set the stop frequency [Frequency] {Stop Freq}, 3 GHz

    Set the number of points at which to measure [Frequency] {Points}, 30 {enter}

    Figure 6: Device under test

    setup form. The DUT setupform allows the user to

    prepare the DLP for

    measuring specific devices

    and setups, and provides

    information on how to setup

    the instrumentation for either

    calibration as show above or

    measurement.

  • 8/6/2019 s800NFSA Noise Lna

    9/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    9

    Set the averaging function to 15 averages [Meas Setup] {Avg Number On} 15 {Enter}

    Calibrate the measurement personality [Meas Setup] {Calibrate} {Calibrate}

    While the analyzer is calibrating, a quick refresher on how gain is measured

    G1(D.U.T)

    =)CAL(Gradient

    )MEAS(Gradient=

    CcalHcal

    1cal2cal

    CtotHtot

    1tot2tot

    T-T

    N-N

    T-T

    N-N

    (6)

    Figure 7: Calibration process

    on ESA. The ESA is set to

    sweep a number of times(defined by averaging) across

    the user defined frequency

    range, cycling through all the

    defined attenuator settings to

    ensure that a corrected noise

    figure measurement can be

    made.

    Since we are switching the same noise source between the same Thot and Tcold

    during calibration and measurement, make THtot =THcal and TCtot=TCcal

    G1(D.U.T) =1cal2cal

    1tot2tot

    N-N

    N-N(7) Figure 8: Theory refresh 3.

    N1cal

    Nacal

    N2cal

    TsTH calTc cal

    Calibration(Measurement System)

    Output Noise Power

    N1tot

    Natot

    N2tot

    TsTH totTc tot

    DUT Measurement(DUT & System)

    Gradient =CcalHcal

    1cal2cal

    T-TN-N

    =

    CcalHcal

    2ccal2hcal

    T-T

    BGkT-BGkT

    =

    CcalHcal

    ccalhcal2

    T-T

    )T-(TkBG

    Gradient(CAL) = kBG2

    Gradient =CtotHtot

    1tot2tot

    T-TN-N

    =

    CtotHtot

    21ctot21htot

    T-T

    GBGT-GBGkT

    =

    CtotHtot

    ctothtot21

    T-T

    )T-(TGkBG

    Gradient(MEAS = kBG1G2

  • 8/6/2019 s800NFSA Noise Lna

    10/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    10

    Equation 7 is the ratio of difference in output noise power measured with noise source on/off for

    calibration and measurement setups.

    Lab 3: Noise figure and gain measurements

    Now that the measurement personality is calibrated with the noise source connected directly to the input, it

    is a simple matter to makecorrectednoise figure and gain measurements on a device. The calibration

    indicator on the top of your screen has changed from a red uncorr to a green corr. With no D.U.Tconnected, both the Gain and Noise Figure are at ~ 0db as expected. This is because the analyzer is

    displaying corrected results with the noise contribution of the measurement system removed. Since the

    input is noise, there is invariably some variation, however considering that most D.U.Ts have gain, from

    equation 5 (pg. 7), this value should become negligible

    To see the second stage noise figure (F2) that has been calibrated out, proceed as follows.

    Instructions: ESA-E series Spectrum

    Analyzer

    Keystrokes: ESA-E series Spectrum

    Analyzer

    You can reduce averaging or turn it off for a

    faster measurement

    Setup an uncorrected display to display the

    approximate noise figure of the measurement

    system

    [Meas Setup]{Avg On Off}

    [Input/Output] {Noise figure Corrections}{Noise

    figure Corrections On Off}

    Figure 9: Corrected NF Gain

    measurement after calibration

  • 8/6/2019 s800NFSA Noise Lna

    11/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    11

    Instructions: ESA-E series Spectrum

    Analyzer

    Keystrokes: ESA-E series Spectrum

    AnalyzerToggle back to making a corrected measurement to

    proceed with rest of lab.

    [Input/Output] {Noise figure Corrections}{Noise

    figure Corrections On Off}

    Disconnect the noise source from the input and connect the device under test to the input and connect the

    noise source to the DUT as shown in figure below.

    Figure 11: Device under test setup form for an amplifier measurement.

    As soon as the DUT is connected to the ESA, a measurement will begin on the noise figure and gain of the

    amplifier because the system is already sweeping. The user has the flexibility to select single or continuous

    sweep based on what the intention is. Additionally, the user can specify a lower number of averages to

    permit a faster measurement.

    Access this menu by pressing

    [Mode Setup]{DUT Setup.}Press the tab keys to navigate

    your way around the form.

    When the form Highlights the

    diagram field"Blue" you

    should use the softkey to

    change the parameter to

    Measurement

    Figure 10: Uncorrected

    measurement, showing the

    approximate NF of the

    measurement system.

  • 8/6/2019 s800NFSA Noise Lna

    12/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    12

    Using the display featuresThe noise figure measurement personality has many features to help you interpret and analyze noise figure

    measurements.

    Perform display scaling

    Instructions: ESA-E series SpectrumAnalyzer

    Keystrokes: ESA-E series SpectrumAnalyzer

    Restart the measurement if necessary

    Expand the trace to fit the graph for a better view of

    the measurement using the Auto Scale function as

    shown in figure below.

    Press [ESC] or [Return] to get out of DUT Setup

    screen.

    {Restart} for a faster response after changing DUT.

    Press [Amplitude] use [Next Window] to highlight

    the graph to be expanded then press {Auto Scale}

    Figure 12: Display of Noise figure and gain after auto scaling

    Manually configure your display via the AMPLITUDE key as follows..

    Instructions: ESA-E series Spectrum

    Analyzer

    Keystrokes: ESA-E series Spectrum Analyzer

    Set the scale of the graphical view. Press [Next Window] to highlight the graph to be

    changed. In this case the Gain window.

    Press {Scale/Div} and enter the new value 2 {dB}

    Set the reference value

    Set the position of the reference.

    Press [Amplitude] {Ref Value} and enter the value 20

    and press dB

    Move the position of the Reference Value by toggling

    through {Ref Position Ctr}. Notice what happens

    More Display Features.

    Select and Zoom Active Window:

    This feature allows you to highlight a window and then enlarge it for closer analysis.

    Scale and reference levelvaluesThe scale in dB per division and the

    reference values can be adjusted to

    give an optimized view of the

    measured results. The scale per

    division can be adjusted from 0 to 20

    dB. The Reference level can be placed

    at the top of the graph, in the center or

    at the bottom. The reference level is

    also adjustable from 100 dB to +100

    dB.

    Use the Auto Scale feature to give thebroadest view of the measured trace.

    The lowest point will be placed at the

    bottom of the graph and the highest

  • 8/6/2019 s800NFSA Noise Lna

    13/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    13

    Instructions: ESA-E series SpectrumAnalyzer

    Keystrokes: ESA-E series SpectrumAnalyzer

    Highlight the window of interest Press [Next Window] until the window you want is

    highlighted

    Enlarge the window for closer analysis Press [Zoom]

    Switch to another window (figure 13) Press [Next Window]

    Figure 13: Full screen display of Noise figure and device gain

    General, Markers and Source tabs

    There are three tabs available at the bottom of the screen. These tabs are accessed using the [ ] and []tab arrow keys on the front panel of the ESA. The General tab shows information about BW, number of

    points, Tcold value, loss, attenuator setting and internal preamplifier setting.

    The Marker tab gives the frequency, noise figure and gain at each of the marker readings. The Source tab

    has information about the noise source including serial number and model identification.

    Instructions: ESA-E series Spectrum

    Analyzer

    Keystrokes: ESA-E series Spectrum

    Analyzer

    View the General table at the bottom of display. Use the Right and Left Tab keys at the bottom of

    the front panel to scroll through the tabsView the Source tab at the bottom of the display. Use the Right and Left Tab keys.

    Figure 14: General information display

    Figure 15: Noise source information: Cal information is blank because user selected common ENR table

    Markers

  • 8/6/2019 s800NFSA Noise Lna

    14/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    14

    A total of four normal markers can be placed on the graphical display. The placement of the markers is

    limited to the number of equally spaced calibration points. For example, if there are 11 calibration points

    then the markers can be placed on each of the vertical graticule lines. Each of the normal markers can be

    changed to delta markers. For example marker 2 will change to marker 2 and 2R where 2R is the reference

    and 2 would be the delta.

    Instructions: ESA-E series SpectrumAnalyzer

    Keystrokes: ESA-E series SpectrumAnalyzer

    The marker function operates the same as the

    standard ESA-E Series.

    To turn marker on, press [Marker].

    Turn on marker 2 Press {Select Marker 2} and press {Normal}

    Active delta marker 2.

    The marker table under the graphical display

    reflects the delta marker information.

    First place the marker to a reference point using

    knob or up/down arrows. Press {Delta}. Move the

    marker relative to the reference marker.

    Switch between displaying the absolute frequency

    of the delta marker and the reference marker

    frequency.

    Press {Delta Pair}. Note the change in frequency

    above the graphical display.

    Note that it is not necessary to toggle to the marker tab when selecting marker functions. This is

    automatically displayed. Additionally, the analyzer toggles to the general tab once markers are turned off.

    Change format of the active window

    The default view of the window is the graphical mode with noise figure in the top and gain in the bottom.The two graphs can be combined to display both traces on one graph. There are two other views available;

    the table mode, which some users prefer, and meter mode which provides quick and easy to read

    measurement information while facilitating testing.

    Instructions: ESA-E series Spectrum

    Analyzer

    Keystrokes: ESA-E series Spectrum

    Analyzer

    To combine both traces on one graph, see figure

    below.

    Press [View/ Trace]{Combined on}.

    Figure 16: Display of markers

    and delta markers on the ESA

    allow noise figure and gain to

    be read along the entire sweep

  • 8/6/2019 s800NFSA Noise Lna

    15/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    15

    Activate the table mode, see figure below Press [View/ Trace]{Table}

    Activate the meter mode Press [View/ Trace]{Meter}

    Change the measurement result you are displayingThe default display of the ESA is noise figure and gain., however there is a separate facility for displaying

    6 different types of measurement results. This can be done independently for the two windows. Depending

    on which one is chosen

    Instructions: ESA-E series SpectrumAnalyzer

    Keystrokes: ESA-E series SpectrumAnalyzer

    Activate the table mode, see figure below Press [View/ Trace]{Result A}{Teffective}

    Scale the view appropriately Press [Amplitude]{Auto Scale}

    Figure 17: Different measurement result type on ESA.

    Creating and testing to limit lines

    In the manufacturing environment, it may be necessary to increase manufacturing through-put. This could

    be implemented by inserting pass/fail limit lines. By using this function, the operator can quickly and

    Figure 17: Combined display mode on the ESA and Table display mode on ESA.

    Teffective from equation 2 is

    proportional to noise figure and so the

    graph should follow a similar trend

  • 8/6/2019 s800NFSA Noise Lna

    16/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    16

    efficiently quantify noise figure and gain dramatically reducing the time spent testing each DUT. Up to four

    limit lines can be setup two for the upper graph and two for the lower graph. The two limit lines for the

    upper graph are designated with up arrows (soft keys), and the limit lines for the lower graph are designated

    with down arrows (soft keys). The limit lines can be designated as upper limit or lower limit and each can

    have a test pass/fail indicator.

    Instructions: ESA-E series SpectrumAnalyzer

    Keystrokes: ESA-E series SpectrumAnalyzer

    Change view back to default noise figure Press [View/Trace]{Result A}{Noise

    Figure}{Noise Figure(dB)}

    Open the limit line editor, select upper limit for the

    upper graph and turn on the limit test.[Display]{limits}{limit 1}{Edit}, use [][] tabkeys under display to highlight Limit. Press {On},

    move to Type, press {Upper}, move to Display

    {On}, move to Test {On}.

    Insert limit values for 10 MHz, 100MHz, 1, 2 and 3

    GHz

    Limit lines values take on the value of the set of

    results it is being applied to.

    Use [][] tab keys to highlight point 1. Press{Frequency 10 MHz}, {limit Value}[3]{ x 1},

    {Connected Yes}, {Point 2} {Frequency 100

    MHz}, {Limit Value}[ 4]{ x 1}, {Connected Yes},

    {Point 3},{Frequency 1 GHz}, {Limit Value{[ 5]{

    x 1}, {Connected Yes}, {Point 4},{Frequency 2GHz}, {Limit Value}[{ 5]{ x 1}, {Connected

    Yes},{Point 5},{Frequency 3 GHz}, {Limit

    Value}[ 8]{ x 1}, {Connected Yes}.

    Figure 18:Limit line editor on the ESA.

  • 8/6/2019 s800NFSA Noise Lna

    17/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    17

    Figure 19: Measurement with limit test on, showing a passing noise figure limit test

    Lab 4: Noise Figure Uncertainty Calculator

    Option 219, noise figure personality has a built-in uncertainty calculator. To calculate the overall

    measurement uncertainty, simply choose the default noise source (N4002A for example), enter the input

    and output match of the device under test and the gain/noise figure of the DUT from the measurement

    display and the value of the uncertainty will be calculated. There are some default values for the instrument(ESA) already entered.

    Using the built-in uncertainty calculator to measure the uncertainty of the measurement for:

    Instructions: ESA-E series Spectrum

    Analyzer

    Keystrokes: ESA-E series Spectrum Analyzer

    Select uncertainty calculator Press [Mode Setup], {Uncertainty Calculator}

    Choose N4002A as the noise source Use [] [] tab keys to highlight Noise SourceModel box. Press {N4002A}

    Setup the instrument NF and Gain uncertainty for

    ENR and frequency range of measurement.

    Instrument noise figure ~ what you saw in lab 3

    during uncorrected measurement after calibration

    Use [] [] tab keys to highlightInstrument Noise Figure and enter 8dB.

    Highlight Noise figure uncertainty and enter 0.41 dB

    Highlight Gain uncertainty and enter 0-.83 dB

    Enter the Noise Figure and Gain values from the Using tab keys to highlight DUT Noise Figure and

    Example:Frequency 525 MHz

    D.U.T. with 24.85 dB gain

    D.U.T. NF: 3.45 dB

    N4002A ENR (12 16 dB) F

  • 8/6/2019 s800NFSA Noise Lna

    18/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    18

    measurements graph or marker table enter 3.45 dB. (To view the marker table, press

    [Return] and to return to the calculator press

    {Uncertainty Calculator}.

    Then highlight DUT Gain and enter 24.45 dB.

    The input and output match of the DUT is

    determined from the specifications sheet or

    measured using a network analyzer.

    Highlight DUT Input Match and enter 1.5, Highlight

    DUT Output Match and enter 1.5.

    The measurement Uncertainty is then calculated and

    the results is display at the bottom of the form.

    SNS allows easy correction for physical temperature

    While on uncertainty.

    Remember from figure 1, page 3:

    ENR= )Tc

    TcTh(

    : We assumed that Ts=Tc=To

    The default value for Tcold used in the ESA is 296.5K. This is assumed to be the value of the noise source

    physical temperature when in the off condition. The ENR vs frequency tables, characterizing each noise

    source that was entered in Lab 1 are referenced to this. This may very well not be the case and if Tcold is

    not 296.5K, the ENR will not be correct leading to an error in the noise figure measurement shown below.

    Figure 20: Uncertainty

    calculator display showing

    result for the example above.

    If you have the time, you can

    observe the trends of

    uncertainty changes, i.e.

    increase in uncertainty as the

    gain decreases etc

  • 8/6/2019 s800NFSA Noise Lna

    19/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    19

    Figure 21: Magnitude of error in noise figure measurement if true Tcold is not what the instrument expects

    For a D.U.T. NF of 3.5 dB, and a temp error of 8K, there is almost an additional 0.1 dB that needs to be

    added to the overall uncertainty budget!

    This is why the SNS on the ESA is so valuable. They allow automatic sensing of the ambient temperature

    of the measurement environment allowing the ESA to compensate for these changes during the

    measurement cycle increasing the accuracy and reliability of noise figure measurements.

    Instructions: ESA-E series SpectrumAnalyzer

    Keystrokes: ESA-E series Spectrum Analyzer

    Automatic temperature sensing of Smart Noise

    Source

    [Meas Setup] {ENR}{Tcold}{Preferences Norm SNS}

    Note that the SNS Tcold is ON. Because of this, the

    Tcold reported in the general tab is not the Default of

    296.5K. It is the automatically sensed ambient

    temperature of 307.15K

    Use the default Tcold value {SNS Tcold On Off}{User Tcold Default User}

    You can also define your own Tcold, or sense

    this from the SNS instantaneously

    {User Tcold Default User}{User Tcold from SNS}

  • 8/6/2019 s800NFSA Noise Lna

    20/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    20

    Lab 5: Noise figure measurements using a mixer as the DUT

    When a down conversion is included in the noise figure measurement, for example measuring the noisefigure of a mixer, there are some additional setups to consider. For this example let us use a mixer as a

    downconverter with an IF at 70MHz, LO at 3GHZ and both RF sidebands are used, 2930 and 3070MHz i.e.

    a DSB measurement.

    The measurement as well as calibration is made at the IF frequency.

    When an IF frequency is chosen, it is a good idea to keep the frequency as low as possible in orderto avoid large differences in ENR values between the upper and lower sidebands when using DSB

    mode. This is because it is the ENR value at the LO that is used in the measurement (compromise

    since it is centered between the 2 sidebands)

    Since this device has some loss, it is recommended that the internal preamp be used.

    Compensate for two sidebands by selecting double side band.

    Any broadband noise in the LO will directly affect results. This can be solved by either a high passor low pass filter at the LO port that removes the noise at the IF frequency. Place an IF filter at the

    input of the spectrum analyzer to remove LO feed through. Usually mixers have around 20dB ofisolation between the LO-IF port so the high powered LO will seriously affect results.

    Disconnect the amplifier from the ESA

    Instructions: ESA-E series SpectrumAnalyzer

    Keystrokes: ESA-E series SpectrumAnalyzer

    Setup the ESA for down conversion measurements

    as shown in the diagrams below.

    It is recommended that the internal preamp be used

    when measuring devices that have low gain.

    Press {Meas Setup} {More 1 of 2} {Restore Meas

    Defaults}

    Press {Meas Setup} {Int Preamp On Off}

    Press [Mode Setup] {DUT Setup}{Down Conv}

    Setup the LO frequency (figure 23) on the ESA Move to Ext LO frequency using tab keys the

    enter 3GHz

    Tab to Sideband and choose DSBSetup the fixed IF frequency [Frequency] {Freq mode}{Fixed}{Fixed

    frequency}{70 MHz}

    Calibration: connect the noise source to the input

    of the ESA.

    Press [Meas setup]{Calibrate}{Calibrate}

    Figure 22: Automatic

    sensing of ambient

    temperature or user

    definable Tcold..

  • 8/6/2019 s800NFSA Noise Lna

    21/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    21

    Instructions: ESA-E series SpectrumAnalyzer

    Keystrokes: ESA-E series SpectrumAnalyzer

    Measure the DUT: Connect the mixer IF (I) port to

    the ESA, the LO (L) port to the signal source and

    the RF (R) to the noise source.

    To add more averaging, press [Meas Setup] then

    {Avg Number On)

    Instructions: ESGC E4438C Vector Signalgenerator

    Keystrokes: ESG E4438C Vector Signalgenerator

    Setup the source for + 7 dBm at 3 GHz On E4438C press [Frequency][ 3] GHz

    [Amplitude] [7]{dBm}[RF On]

    Instructions: ESA-E series Spectrum

    Analyzer

    Keystrokes: ESA-E series Spectrum

    Analyzer

    Change the ESA display to meter mode, more

    appropriate for a single frequency measurment

    [View/Trace]{Meter}

    Figure 23: Setup for measuring

    a down-convertor (mixer) at a

    fixed LO and fixed IF

    Figure 24: The meter view

    showing noise figure and

    Gain (conversion loss) of a

  • 8/6/2019 s800NFSA Noise Lna

    22/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    22

    The measurement is valid for a DSB device under test, provided that there isnt too much variation of the

    frequency response of the D.U.T. and the ESA. This measurement could be used to estimate the SSB noise

    figure without having to worry much about image reject filtering. The next page outlines the differences

    and how one can correct for this. The details of mixer measurements are out of the scope of this lab.

    For a S.S.B application, the analyzer should see the image frequency from just one sideband. So as a result,if the analyzer is using the D.S.B. measurement to approximate the S.S.B case, it will see the gain as twice

    what it should be. And since noise figure =(Noise power out)/(Gain * Noise power in}, the noise figure will

    be of what it should be in the S.S.B case. So a loss compensation of 3dB (gain of 3 dB) needs to be

    applied to make the S.S.B equivalent measurement.

    Loss CompensationCompensate for losses before and after the DUT. These losses can be fixed or varied versus frequency. The

    figures below shows the before and after setup menu. The before menu as well as the after can be

    setup to have a table of losses versus frequency or a fixed value. For fixed values, input a value for loss in

    dB and temperature in K. In our example, lets use a fixed gain (negative loss) after the DUT since we are

    making a fixed frequency measurement.

    Instructions: ESA-E series Spectrum

    Analyzer

    Keystrokes: ESA-E series Spectrum Analyzer

    Enter a fixed loss value to use as compensation

    after the DUT

    Press [Input/Output], {Loss Comp}and {Setup}. Use

    the tab keys under the display to highlight the box

    labeled {Loss Compensation after DUT} and press

    {fixed}. Tab to {Fixed Value} and enter 3 dB.

    The temperature of the loss must also be entered.

    In this case room temperature.

    Tab to {Temperature} in the before box and enter

    290 K

    The loss compensation could also have been setup vs frequency in a tabular format

    F LO

    Fif

    Noisepower

    Freq

    LSB

    USB

    Measurementfrequency (IF)

    For D.S.B. measurement1) During Calibration, the analyzer

    uses ENR values at I.F. frequency

    2) During measurement, the analyzer

    uses the average of the U.S.B. and

    L.S.B. ENR valuesFif

    0

  • 8/6/2019 s800NFSA Noise Lna

    23/24

    SA 308: SA EnhancementsNoise Figure Lab

    Noise Figure Lab S800 November 2003

    ESA-E Series, Option 219

    23

    Figure 25: Loss Compensation selection table Noise figure after loss compensation

    Narrowband noise figure measurementsThe noise figure measurement personality has the capability to reduce the resolution bandwidth allowing

    narrow bandwidth measurements. As the bandwidth narrows, the measurement jitter increases. It is

    recommended that the number of averages increase to reduce the jitter. In this measurement, the deviceunder test will be an amplifier, band limited, with a band pass filter which has a center frequency of 70

    MHz. The start and stop frequencies are set and the number of points to be measured are set. In this

    example 30 points will be used. The figure below is an example of a narrowband measurement.

    Instructions: ESA-E series SpectrumAnalyzer

    Keystrokes: ESA-E series SpectrumAnalyzer

    Setup the ESA for narrow band measurements. Press [Mode Setup] {DUT Setup}{Amplifier}

    Connect the noise source directly to the spectrum

    analyzer

    ESA-E SeriesSpectrum

    Top of unit front panel

    Smart NoiseSource

    N-TYPE TO SMAAMP

    SMA TO SMAAmp. Power

    Mini CircuitsFilter

    Mini Circuits Amplifier

    11730 SNS cable

    SNS output

    connector

    Figure 26: Connection

    diagram for narrow band

    measurements

  • 8/6/2019 s800NFSA Noise Lna

    24/24

    SA 308: SA EnhancementsNoise Figure Lab

    N i Fi L b S800 N b 2003 24

    Set start and stop frequency [Frequency]{Freq Mode Swept}{Start Freq, 50

    MHz}{Stop Freq, 90 MHz}

    Set the number of measurement points [Frequency]{Points} 30 enter

    Set the resolution bandwidth [BW/Avg], 300 kHz

    To reduce measurement jitter add averaging then

    calibrate

    [Meas Setup]{Avg Number}15{Enter},

    {Calibrate}{Calibrate}

    To measure the DUT connect the filter to the noisesource and the other end to the input of the amplifier

    and the output of the amplifier to SA input.

    To achieve the display shown below [Trace/View], {Combined On}

    [Amplitude]{Auto Scale}

    Figure 25: Narrow band measurements display of filter in cascade with amplifier.