Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

1 of 50

THE BECKMAN COULTER logo, COULTER, DxH, TruColor, and UniCel are trademarks of Beckman Coulter, Inc; COULTER and UniCel are registered in the USPTO.

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

2 of 50

Objectives

FlagsCodes MessagesCBC ParametersCellular Interference

Differential ParametersNRBC ParametersRetic Parameters

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

3 of 50

Flags

* These Flags are also associated with System Messages. See System Messages in HELP for more information.

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

4 of 50

Codes

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

5 of 50

MessagesMessages

– Results related: • Suspect: messages are generated by internal

algorithms to convey that a clinical condition may exist with a specimen based on an abnormal cell population or distribution.

• System: messages indicate an event that may affect the operation of a system, requires operator notification and or entry into a History Log.

– Most accompanied by an R (review) flag– Some are accompanied by N or P flags

• Definitive: defined by the Operator

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

6 of 50

MessagesException or System Status Messages

– Exception: anything with an exception may or may not be released depending on your lab’s release rules criteria. An exception is the default test order.

– System Status: an event that may affect system operation. Accompanied by operator notification and entry into log.

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

7 of 50

Examples

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

8 of 50

CBC Parameters

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

9 of 50

Data Fusion

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

10 of 50

Cellular Interference

• System message Cellular Inter

• WBC & associated parameters, or TNC, flagged with R

• The results are only flagged if we are uncertain of the correction, or can’t correct (not just because a correction occurred).

• Suspect message Cellular Inter

• WBC & associated parameters, or TNC, flagged with R

• For CBC or CBC/Retic panels only, when NRBC are detected but not reported.

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

11 of 50

Red Cell Messages

• Suspect message Dimorphic Reds

• Data pattern consistent with the presence of multiple red cell populations

• Suspect messageRed Cell Agglut

• Data pattern consistent with the presence of agglutinated red cells

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

12 of 50

RBC Fragments/Microcytes

• Suspect message RBC Frag/Micro

• Pattern observed consistent with the presence of RBC fragments and/or microcytic red cells.

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

13 of 50

Giant Platelets

• Suspect message Giant Platelet

• May be accompanied by System message specific to platelet results

• Giant Platelet data patternobserved

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

14 of 50

Platelet Clumps

• System message Platelet Clumps

• PLT & associated parameters flagged with R

• Clumped platelet data pattern observed

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

15 of 50

Platelet Messages

• System message PLT Inter: Debris

• PLT & associated parameters flagged with R

• PLT requires review due to the possible presence of an interference with small platelets

• System message RBC-PLT Overlap

• PLT & associated parameters flagged with R

• PLT histogram shows a possible interference with larger platelets

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

16 of 50

WBC Interference from HGB

• System message WBC Inter: HGB• HGB & associated parameters flagged with R• HGB may not be adequately corrected for

WBC interference

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

17 of 50

Differential Parameters

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

18 of 50

One Cell at a Time . . .

S H E A T H -F L U ID IN

S H E A T H -F L U ID IN

S A M P L E IN

F L O W C E L L

S A M P L E D IL U T IO N

S H E A T H F L U IDS E N S IN G

A R E A

WBCs are hydrodynamically focused

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

19 of 50

Flow Cytometric Digital MorphologyMultiple angles of light scatter, captures 10X more data than previously possible

– AL2 (axial light loss)– LALS (lower angle LS)– LMALS (lower med angle

LS)– UMALS (upper med angle

LS) – MALS (median angle LS)

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

20 of 50

Flow Cytometric Digital Morphology

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

21 of 50

Differential Parameters

MONO

LYMPH

EOSNEUTS

BASO

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

22 of 50

Abnormal Cell Distribution

1

2

3

4

5

6

7

8

1 MO Blasts2 NE Blasts3 Immature Granulocytes4 Band Neutrophils5 LY Blasts6 Variant Lymphocytes7 Small Lymps / Low Opacity8 Non-White Cells

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

23 of 50

LY Blast

• Suspect message LY Blast• Blasts suspected in the LY area of the

dataplot• Data pattern consistent with possible

presence of very immature cells

5PD1 5PD2 Normal 5PD1

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

24 of 50

MO Blast

• Suspect message MO Blast• Blasts suspected in the MO area of the

dataplot• Data pattern consistent with possible

presence of very immature cells

5PD1 5PD2 Normal 5PD1

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

25 of 50

NE Blast

• Suspect message NE Blast• Blasts suspected in the NE area of the

dataplot• Data pattern consistent with possible

presence of very immature cells

5PD1 5PD2 Normal 5PD1

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

26 of 50

Immature Granulocytes

• Suspect message Imm Grans• [Metamyelocytes, myelocytes &/or

promyelocytes] or [myelocytes & promyelocytes]

• Variable sensitivity with default set at high

5PD1 5PD2 Normal 5PD1

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

27 of 50

Left Shift

• Suspect message Left Shift• Metamyelocytes, without more immature

granulocytes• Variable sensitivity with default at high• Can be disabled

5PD1 5PD2 Normal 5PD1

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

28 of 50

Variant LY

• Suspect message Variant LY• Variant (viral), atypical or abnormal LY

suspected• Variable sensitivity with default at high

5PD1 5PD2 Normal 5PD1

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

29 of 50

Abnormal Diff Pattern

• System message Abn Diff Pattern• Results do not match any known template• All Diff results accompanied by R flag

5PD1 5PD2 Normal 5PD1

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

30 of 50

NRBC

• Suspect message NRBC• NRBC suspected in the signature position of

the dataplot• Only in condition where the NRBC module is

disabled and the panel includes a Diff

5PD1 5PD2 Normal 5PD1

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

31 of 50

Aged Sample

• System message Aged Sample• All Diff results accompanied by R flag• The results are only flagged if we are unable

to give a correct differential, not just because aging is present.

5PD1 5PD2 Normal 5PD1

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

32 of 50

Excessive Debris

• System message Excessive Debris: D• All Diff results accompanied by R flag

5PD1

5PD2

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

33 of 50

NE-EO Overlap

• System message NE-EO Overlap• All Diff results flagged with R• Population shift observed (NE and EO not in

correct spatial location)

5PD1 5PD2

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

34 of 50

MO-NE Overlap

• System message MO-NE Overlap• All Diff results flagged with R• Population shift observed (MO and NE not in

correct spatial location)

5PD1

5PD2

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

35 of 50

Low Events: D

• System message Low Events: D• Not enough events available for differential

5PD1

5PD2

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

36 of 50

High Opacity Events

• System message High OP Events• All Diff results accompanied by R flag• Interference in Opacity that has been

observed with bone marrow samples or specimens with platelet clumps.

5PD1

5PD2

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

37 of 50

Low RMALS Events

• System message Low RMALS Events• All Diff results flagged with R• RMALS mode value is too low

5PD1

5PD2

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

38 of 50

NRBC Parameters

PLT/DEBRIS WBC

NRBC

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

39 of 50

Abnormal NRBC Pattern

• System message Abn NRBC Pattern

• NRBC & associated parameters flagged with R

• Fits no known pattern

NRBC1 NRBC2

Normal nRBC

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

40 of 50

NRBC-LY Overlap• System

message NRBC-LY Overlap

• NRBC & associated parameters flagged with R

• Couldn’t separate NRBC from small LY during

• analysis

NRBC1 NRBC2

Normal nRBC

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

41 of 50

NRBC Interference• System

message NRBC Inter

• NRBC & associated parameters flagged with R

• Significant amount of interference in NRBC regions that are not NRBC

NRBC1 NRBC2

Normal nRBC

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

42 of 50

Miscellaneous NRBC Messages

• Detected by Hardware, not algorithm• Flow Cell Clog: N• System Event: N

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

43 of 50

Reticulocyte Parameters

MATURE RBC’S

WBC’S

RETICS

PLT/DEBRIS

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

44 of 50

Abnormal Hemoglobin

• Suspect message Abn Hemoglobin

• ‘Others’ category = green (~ platelets) & orange (not cleared) particles.

• Indicates a possible clearing problem in the presence of abnormal hemoglobin

Normal Retic

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

45 of 50

Sickled Cells

• Suspect message Sickled Cells

• Data pattern consistent with sickled cells

Normal Retic

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

46 of 50

Retic System Messages• System message

RET-RBC Overlap• RET & associated

parameters flagged with R.• Unclear separation of retics

& mature red cells.

• System messageRET Inter: PLT

• RET & associated parameters flagged with R.

• The platelet population may have interfered with the RET.

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

47 of 50

Nucleated Cells• System message

Nucleated Cells• RET & associated

parameters flagged with R

• NRBC or small WBC (especially small LY as in CLL) are visible in the dataplot (blue) and may have interfered with RET measurement

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

48 of 50

Flow Cell Clog• System message

Flow Cell Clog: R• RET and associated

parameters flagged with R

• Hardware detected a flow cell clog during analysis

• It is possible to have a flow cell clog that is severe enough to inhibit data [:::::].

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

49 of 50

System Event During Retic

Examples include (but are not limited to):

• Laser Offset Voltage failure

• Temperature Out of Range

• System message System Event: R

• RET & associated parameters flagged with R

• Hardware parameters were out of limit for an item that could affect RET analysis (e.g. voltage, temperature, pressure).

Rev. A (June 2014) Beckman Coulter Professional Development© 2014 Beckman Coulter, Inc.

50 of 50