Results from the PMT test bench in BCN (10-Oct-03) J.Garra, R.Graciani, E.Graugés, X.Xirgu.

14
Results from the PMT test bench in BCN (10- Oct-03) J.Garra, R.Graciani, E.Graugés, X.Xirgu

Transcript of Results from the PMT test bench in BCN (10-Oct-03) J.Garra, R.Graciani, E.Graugés, X.Xirgu.

Page 1: Results from the PMT test bench in BCN (10-Oct-03) J.Garra, R.Graciani, E.Graugés, X.Xirgu.

Results from the PMT test bench in BCN (10-Oct-03)

J.Garra, R.Graciani, E.Graugés, X.Xirgu

Page 2: Results from the PMT test bench in BCN (10-Oct-03) J.Garra, R.Graciani, E.Graugés, X.Xirgu.

8 stage PMT

CFD base: The output of 1 channel studied to

perform a premilinary analysis on: Signal Shape Gain Number of photo-electrons

Page 3: Results from the PMT test bench in BCN (10-Oct-03) J.Garra, R.Graciani, E.Graugés, X.Xirgu.

CFD active base

Page 4: Results from the PMT test bench in BCN (10-Oct-03) J.Garra, R.Graciani, E.Graugés, X.Xirgu.

Signal Shape

4600 4800 5000 5200 5400 5600 5800 6000 6200 6400

-0.040

-0.035

-0.030

-0.025

-0.020

-0.015

-0.010

-0.005

0.000

0.005

Sig

na

l

Units of time (aprox. 0.1 ns)

Signal taken with a recycled setup

Some oscillations in the signal shape are supposed to be caused by some connection(?) problem.

Page 5: Results from the PMT test bench in BCN (10-Oct-03) J.Garra, R.Graciani, E.Graugés, X.Xirgu.

Gain vs Photo-Statistics

0 20 40 60 80 100 120 140460000

480000

500000

520000

540000

560000

580000

600000

620000

Ga

in

Number of photoelectrons

Dependence gain-number of photoelectrons

• Possible saturation effects might be observed starting around 50 phe and above

Page 6: Results from the PMT test bench in BCN (10-Oct-03) J.Garra, R.Graciani, E.Graugés, X.Xirgu.

Gain

650 700 750 800 850 900

100000

1000000

2 = 500090665.89604

R2 = 0.99727 A = 1.3094E-15 ± 1.4236E-15B = 7.11223 ± 0.16104

Ga

in

-HV (V)

Dependence gain-source voltage

Our desired HV working point should be ~ 600 Volts obtaining an approximate gain of 30k to 50k

Page 7: Results from the PMT test bench in BCN (10-Oct-03) J.Garra, R.Graciani, E.Graugés, X.Xirgu.

8 stage PMT BCN base:

A preliminary study has been performed with a LED source (no fibre)

Gain and Photo-statistics have been looked at from the output of a 1 channel (#64)

Now the setup is ready to get measurements for all the channels at once.

Fatal accident ended up with the 8-stage PMT becoming unusable anymore

Page 8: Results from the PMT test bench in BCN (10-Oct-03) J.Garra, R.Graciani, E.Graugés, X.Xirgu.

BCN active base

Page 9: Results from the PMT test bench in BCN (10-Oct-03) J.Garra, R.Graciani, E.Graugés, X.Xirgu.

Signal Shape

4600 4800 5000 5200 5400 5600 5800 6000 6200 6400

-0.025

-0.020

-0.015

-0.010

-0.005

0.000

Sig

na

l

Units of time (aprox. 0.1 ns)

Signal with R6=R

7=R

8=0

Page 10: Results from the PMT test bench in BCN (10-Oct-03) J.Garra, R.Graciani, E.Graugés, X.Xirgu.

4600 4800 5000 5200 5400 5600 5800 6000 6200 6400-0.05

-0.04

-0.03

-0.02

-0.01

0.00

Sig

nal

Units of time (aprox. 0.1 ns)

Signal with R6=310 K, R

7=R

8=180 K

Signal Shape

Page 11: Results from the PMT test bench in BCN (10-Oct-03) J.Garra, R.Graciani, E.Graugés, X.Xirgu.

The oscillations present in the first attempt (CFD) could be partially removed (with the BCN base) by when the setup has been re-done very carefully and assuring a common grounding for the (ADC/Black Box/Active Base) system.

Still to be tried if that also cured the oscillation observed using the CFD base

Page 12: Results from the PMT test bench in BCN (10-Oct-03) J.Garra, R.Graciani, E.Graugés, X.Xirgu.

12 stage PMT This PMT has already incorporated the

active base by Hamamatsu together with the (PMT/Base) system encapsulation

Scans of 50x50 measurements each have been performed for different angles to ensure the geometrical consistency of the setup (PMT plane parallel to the fiber scaning plane): 0,90,180 and 360 deg. scans

Also scans with different light intensities have been taken to test the uniformity in the linear response of each channel.

Page 13: Results from the PMT test bench in BCN (10-Oct-03) J.Garra, R.Graciani, E.Graugés, X.Xirgu.

2D Scans

Different angles

Different light I

Documento Acrobat

Documento Acrobat

Page 14: Results from the PMT test bench in BCN (10-Oct-03) J.Garra, R.Graciani, E.Graugés, X.Xirgu.

To do list

Let the new team get more experienced

Complete a final version of the fine scans (Motor lost steps, etc…).

Understand any the level of saturation of the signal

Single photo-electron measurement using a 2-stage amplifier board (work in progress)