Respiration.sensors
-
Upload
ashwani-gaur -
Category
Documents
-
view
216 -
download
0
Transcript of Respiration.sensors
-
8/10/2019 Respiration.sensors
1/2
-
8/10/2019 Respiration.sensors
2/2
viable solution for establishing interoperability between
medical devices. For this reason, CANopen was chosen as the
communication protocol between the measured respiration
pattern and the hospital bed movements. The CANopen
protocol stack was implemented in a Motorola HCS12
microcontroller. The HCS12 also has an analog to digital
converter which is needed for interfacing with the pressure
transducer.
III METHODS
The pressure transducer outputs an analog voltage reading,
which must be converted to a digital signal for storage and
analysis. The HCS12 processor has a
10
bit analog to digital
convertor, and accepts analog voltages in the range
of
0 to
5.12V. The voltage output from the transducer must be
amplified and biased around 2.5V to work in the HCS12, so
an operational amplifier circuit was designed to convert the
pressure transducer voltage to acceptable values.
The pressure transducer is highly sensitive so that it can
detect the slight respiration patterns in humans, but it will also
be susceptible to noise factors. Therefore, a
th
order
Butterworth band pass filter was designed to filter the
respiration waveform from the pressure transducer output. A
typical adults respiration rate is anywhere from
12
to 20
breaths per minute, so a band pass range
of
0.1 to 0.4 Hz was
selected.
The filtered data was continuously stored in the HCS12
microprocessor, and a peak detection algorithm was
implemented to count the peaks in the filtered respiration
waveform. The time between peaks was calculated and used
to represent the instantaneous respiration rate. When a
subjects respiration rate ceased, the peak detection algorithm
detected no peaks, and
if
no peaks were detected for 10
seconds, a bed adjustment was issued.
The bed adjustment is a command issued by the CANopen
protocol. Specifically for this project, a command was
configured to be issued automatically upon a 10 second timer
in which no respiration peaks were detected in the filtered
respiration waveform.
IV. RESULTS AND DISCUSSION
A system was developed that can detect a patient's
respiration rate without the need for any physically attached
equipment. The patient must only be lying in the hospital bed
for the system to accurately detect their respiration rate.
Furthermore, the system is able to automatically respond to
respiration failure by triggering a bed adjustment that alters
the patients sleeping posture.
Due to the nature
of
the monitoring method, the respiration
measurements can be slightly impacted by the weight and
position of the patient. The peak detection algorithm might
need to be customized to the specific patient attributes.
Because
of
limited access to bed functions, the bed
adjustment method employed in this project was a
modification in the overall tilt of the hospital bed. Other
adjustments are possible, like the ability to roll the patient to
one side, but are not implemented in this project. While not
necessary for the project, more in-depth access to the various
bed adjustment functions could provide a more beneficial
treatment to obstructive sleep apnea.
V. CONCLUSIONS
In conclusion, this project demonstrates the possibility to
monitor a patient's respiration rate in an advanced hospital
bed with no need for constraining or invasive medical
equipment. Additionally, with the use
of
a medical device
communication protocol, a further benefit can be realized by
providing an automated first response to a patient suffering
from obstructive sleep apnea.
ACKNOWLEDGMENTS
This study was supported by the University of New
Hampshire. The authors gratefully thank Mr. Frank Hludik
from UNH, Mr. Wayne Smith from UNH, and Mr. William
Seitz, President
of
IXXAT for their help and support.
REFERENCES
[I] Etschberger, Konrad. Controller Area Network Weingarten: IXXAT
Press, 2001.
[2] MC9SI2DP256B: 16-bit Microcontroller.
Motorola
August, 2006
[3] J. Waters, J. Ojala, J. LaCourse (2007). Standardizat ion of Acute
Health Care Digital Communications. Proceedings
of
he 33 d Northeast
Bioengineering Conference.
[4] Cartwright, R. D. (1984). Effect of Sleep Position on Sleep Apnea
Severity. Sleep 7(2): 110-114.
[5] K. Watanabe, T. W., H. Watanabe, H. Ando, T. Ishikawa, K. Kobayashi
(2005). Noninvasive Measurement
of
Heartbeat, Respiration, Snoring
and Body Movements of a Subject in Bed via a Pneumatic Method.
IEEE Transactions on Biomedical Engineering 52(12): 2100-2107.
[6] Morgenthaler, T.
I
(2006). Practice Parameters for the Medical
Therapy
of
Obstructive Sleep Apnea. Sleep 29(8): 1031-1035.
[7] P. Chow, G. N., J. Abisheganaden, Y
TWang
(2000). Respiratory
monitoring using an air-mattress system. Physiol. Meas. 21: 345-35
[8] R. Jokic, A. K., M. Crossley, G. Sridhar, M.F. Fitzpatrick (1999).
Positional Treatment vs Continuous Positive Airway Pressure in
Patients With Positional Obstructive Sleep Apnea Syndrome. Chest
115(3): 771-781.
[9] R.D. Cartwright, S. L., J. Lilie, H. Kravitz (1985). Sleep Position
Training as Treatment for Sleep Apnea Syndrome: A Preliminary
Study. Sleep 8(2): 87-94.
[10] Victor, L. D. (2004). Treatment of Obstructive Sleep Apnea in Primary
Care. American Family Physician 69(3): 561-568.
Authorized licensed use limited to: Rochester Institute of Technology. Downloaded on October 6, 2009 at 11:02 from IEEE Xplore. Restrictions apply.