Resident Physics Series
ACR Mammography Protocols
![Page 2: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/2.jpg)
Mammography Quality Standards Act (MQSA)
“to ensure the safety and reliability of mammography and help detect breast cancer in its earliest, most treatable stages”
Mammography practice under federal control (FDA)
19921992
![Page 3: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/3.jpg)
MQSA Requirements
Yearly physics surveysYearly FDA inspections
Georgia x-ray regulators on contract with FDAFY 2013
No violations: 87%
![Page 4: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/4.jpg)
Facility Accreditation by “Accrediting Body”
American College of Radiology (ACR)State of ArkansasState of IowaState of Texas
![Page 5: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/5.jpg)
![Page 6: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/6.jpg)
FDA Facility & Procedure CountJune, 2006 Late 2013 %
Change
Total certified facilities
8850 8691 -1.8
Total accredited units
13588 12195 -10.3
Certified facilities with FFDM units
1027 8004 +679.4
Accredited FFDM units
1455 12195 +738.1
Annual mammography procedures
33,940,037
38,596,266
+13.7
![Page 7: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/7.jpg)
Number of Accredited FFDM Units
Jun 11Jun 11Nov 03Nov 03
![Page 8: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/8.jpg)
Number of Accredited Units
Jun 11Jun 11Nov 03Nov 03
![Page 9: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/9.jpg)
% FFDM Units
Dec 2003Dec 2003
Jun 2011Jun 2011
![Page 10: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/10.jpg)
FDA-Approved FFDM Units Last 5 YearsDate Manufacture
rModel
6/15/11
Hologic Selenia Encore Direct
4/28/11
Sectra MicroDose Mammography L30 Photon Counting
2/11/11
Hologic Selenia Dimensions Digital Breast Tomosynthesis (DBT)
Direct
2/11/11
Siemens Mammomat Inspiration Direct
11/3/10
Carestream Directview Computed Radiography (CR) CR
2/11/09
Hologic Selenia Dimensions 2D Direct
2/11/09
Hologic Selenia S Direct
2/11/09
Siemens Mammomat Novation S Direct
11/2007
Hologic Selenia with Tungsten target Direct
07/10/06
Fuji Computed Radiography Mammography Suite
CR
04/11/06
General Electric
Senographe Essential Indirect
![Page 11: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/11.jpg)
Mammography QC Manual
RadiologistClinical Image QualityRadiologic TechnologistMedical Physicist
Revised Edition, 1999Revised Edition, 1999
SectionsSections
![Page 12: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/12.jpg)
Quality Assurance (QA)
Every imaging procedure is necessary & appropriate to clinical problem at hand
images contain information critical to solution of that problem
QA should ensureQA should ensure
![Page 13: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/13.jpg)
Quality Assurance (QA)recorded information is
correctly interpretedmade available to patient’s physician in a
timely fashion consistent with image quality objectives,
examination results in lowest possibleradiation exposurecostinconvenience to the patient
![Page 14: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/14.jpg)
Quality Control (QC)
Acceptance Testingdetection of defects in equipment that is
newly installed or has undergone major repair
BaselinesEstablishment of baseline equipment
performance quantitative data when system operating properly
QC is integral part of QAQC is integral part of QA
![Page 15: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/15.jpg)
Quality Control (QC)
DiagnosisDiagnosis of changes in equipment
performance before they become radiologically apparent
VerificationVerification of correction of causes of
deterioration in equipment performance
![Page 16: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/16.jpg)
Facility Responsibility
Designate One Lead Interpreting Physician
![Page 17: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/17.jpg)
Lead Physician’s Responsibilities
Ensure technologists have adequateorientation
based on procedure manualtrainingcontinuing education
Ensure effective QC program
![Page 18: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/18.jpg)
Radiologist’s Responsibilities
Designate one technologist responsible for QC
QC tech can delegate responsibilities
![Page 19: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/19.jpg)
Radiologist’s Responsibilities
Ensure availability of appropriate test equipment
Arrange staffing / scheduling to allow time for QC
![Page 20: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/20.jpg)
Radiologist’s Responsibilities
provide frequent consistent positive & negative feedback to technologists about film quality & QC
Review technologist’s test resultsno less than every 3 monthsmore often if inconsistent
results
![Page 21: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/21.jpg)
Radiologist’s Responsibilities
Select a medical physicistadministers QC programperforms physicist’s tests
Review physicist’s test results
![Page 22: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/22.jpg)
Radiologist’s Responsibilities
Oversee or designate qualified individual to oversee radiation protection program foremployeespatientsindividuals in surrounding area
![Page 23: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/23.jpg)
Radiologist’s Responsibilities
Ensure proper maintenance of records in QC procedures manualemployee qualificationsmammography technique / proceduresquality control / safety / protectioninfection control
![Page 24: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/24.jpg)
Radiologist’s Responsibilities
“The radiologist is ultimately responsible for the quality of films produced under his or her direction and bears ultimate responsibility for both proper QC testing and QA procedures in mammography.”
“The radiologist is ultimately responsible for the quality of films produced under his or her direction and bears ultimate responsibility for both proper QC testing and QA procedures in mammography.”
![Page 25: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/25.jpg)
The Proliferation of Digital MammographyTotal mammography units in US (10/1/2010)
12,445Total digital mammography units (10/1/2010
8748 (70.3%)Total digital mammography units (4/1/2010
or 6 months ago)8195 (65.2%)
Increase in digital units~ 92/month
![Page 26: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/26.jpg)
Physicists & Digital MammoraphyPhysicist must insure equipment meets
manufacturer’s specificationsEquipment specs now defined by
manufacturer, not ACR/FDA
![Page 27: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/27.jpg)
Physicist’s Responsibilities
Note: All physicistNote: All physicist’’s tests are to be done s tests are to be done annuallyannually or after tube replacement or or after tube replacement or major servicemajor service
![Page 28: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/28.jpg)
Physicist’s Responsibilities: Mammography Unit Assembly Evaluation
mechanical stability / identification of sharp edges
receptor stabilitylocks / motions / detentsoperator shieldingthickness scale accuracyindicator lights workingtechnique chart posted (see next
slide)
![Page 29: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/29.jpg)
Technique ChartEven though most mammo is
phototimed, still need technique chartFor given breast size, density, view
specify machine set-upPhototimer mode (auto standard,
contrast, dose, …)Density (-2, -1, 0, …)
![Page 30: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/30.jpg)
Technique Chart
Wrong (my opinion)Display kVp, mAs, target, filter calculated by
machine for given image
Correct (my opinion)Display machine settings selectable by
technologist for this breast and view
Should provide enough guidance to allow a technologist who has not worked in this room or facility to properly set up a study
![Page 31: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/31.jpg)
Physicist’s Responsibilities: Mammography Unit Assembly Evaluation
Cassettes slide smoothly into/out of holderOverride available for auto-decompression
display must indicate when auto-decompression turned off
Manual release of compression if power lost
![Page 32: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/32.jpg)
Collimation Assessment
x-ray light field alignmentbeam does not exceed receptor
by > 2% SIDcompression paddle / receptor
alignment at chest wall within 1% SIDpaddle not visible on image
Image should fill filmMany units by design will not do this
![Page 33: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/33.jpg)
Physicist’s Responsibilities
Focal Spot Performancelimiting resolution pattern
kVp accuracy / reproducibility
Beam quality (HVL)minimum & maximum
minimum: patient dose maximum: image contrast
![Page 34: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/34.jpg)
Automatic Exposure Control (AEC / Phototiming)kVp trackingThickness trackingimage mode tracking (cassette sizes, w
w/o grid)automatic mode tracking
unit selects kVp, target, filterdensity control
even steps of ~ 15-20% Density Step Changes
0
2
4
6
8
10
12
14
16
18
-5 -4 -3 -2 -1 0 1 2 3 4 5
Step
% C
han
ge
Series1
![Page 35: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/35.jpg)
Physicist’s Responsibilities:
Screen Uniformitycompare O.D. of each cassette using
phototimerAEC Reproducibility
![Page 36: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/36.jpg)
Physicist’s Responsibilities
Breast Entrance Exposure, Average Glandular Dose, RMI-156 “accreditation” phantom used for
entrance exposure / average glandular dose
![Page 37: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/37.jpg)
Breast Average Glandular Dose Limits
0.3 rad (300 mrads, 3 mGy) maximum per view for screen-film receptors using a gridSame for film and digital
0.1 rad (100 mrads, 1 mGy) maximum per view for non-grid screen-film receptors
Radiation output rate > 800 mR/sec
![Page 38: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/38.jpg)
Image Quality Evaluationuse RMI-156 “accreditation” phantomrecord
fibersspeck groupsmassesoptical densitiestechnique
![Page 39: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/39.jpg)
A Poor Phantom Image
![Page 40: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/40.jpg)
Artifact Identification
Artifact evaluation / descriptiondetermination of artifact source
processor other equipment
Done with phantomfilm
![Page 41: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/41.jpg)
Physicist’s Responsibilities
Viewing conditionsambient lightviewbox brightness
My experienceAmbient lighting
often ignored
![Page 42: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/42.jpg)
Technologist’s Responsibilities
Dailydarkroom cleanlinessprocessor quality control
sensitometric data
Weeklyscreen cleanlinessviewboxes and viewing
conditionsPhantom images
All QC must not only be performed All QC must not only be performed but must be but must be documenteddocumented!!
Speed
1.11
1.31
1.51
6/2/
97
6/3/
97
6/4/
97
6/5/
97
6/6/
97
6/9/
97
6/10
/97
6/11
/97
6/12
/97
6/13
/97
6/16
/97
6/17
/97
6/18
/97
6/19
/97
6/20
/97
6/23
/97
6/24
/97
6/25
/97
6/26
/97
6/27
/97
6/30
/97
7/1/
97
7/2/
97
7/3/
97
7/7/
97
7/8/
97
7/9/
97
7/10
/97
7/11
/97
7/14
/97
O.D
.
![Page 43: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/43.jpg)
Technologist’s ResponsibilitiesMonthly: Visual Checklist
visual checklistSID indicatorangle indicatorlocksfield lightsmooth motions
![Page 44: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/44.jpg)
Technologist’s ResponsibilitiesMonthly: Visual Checklist
cassette lockIs cassette held firmed in place when tubestand
tilted
Compression device & firm compressionSmooth edgesHolds pressure
hand switch placementvisibilityswitches/ lights/ meterscones/ collimators
![Page 45: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/45.jpg)
Technologist’s ResponsibilitiesQuarterly
Fixer retention analysisFixer affects
archivability of filmsRepeat analysis
breakdown by cause motion positioning technique static etc.
![Page 46: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/46.jpg)
Semi-Annual Technologist’s Responsibilities: Darkroom Fog
must use sensitized film
partially cover previously exposed phantom film in darkroom for 2 minutes
up to .05 O.D. increase acceptable
![Page 47: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/47.jpg)
Semi-annual Technologist’s ResponsibilitiesScreen Film Contact
40 lines/inch Copper meshsubjective resultspoor contact can result if time (15 minutes)
not provided for air to bleed out of cassette after closingNOTE: 15 minutes between cassette loading
& exposure must be provided for all clinical films to insure good contact
![Page 48: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/48.jpg)
Semi-annual Technologist’s ResponsibilitiesScreen Film Contact
![Page 49: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/49.jpg)
Semi-Annual Technologist’s Responsibilities
Compressioncan use bathroom scale covered with
towel25 - 40 pounds for automatic systemsat least 25 pounds for manual
compression
![Page 50: Resident Physics Series ACR Mammography Protocols.](https://reader030.fdocuments.in/reader030/viewer/2022013122/56649ce15503460f949ac309/html5/thumbnails/50.jpg)
The End
Questions?