Real time Pcr

37
REAL TIME PCR 1

description

Real time Pcr. Limitations of End-Point PCR. Poor Precision Low sensitivity Low resolution Non - Automated Size-based discrimination only Results are not expressed as numbers Ethidium bromide for staining is not very quantitative Post PCR processing. The Evolution of PCR to Real-Time. - PowerPoint PPT Presentation

Transcript of Real time Pcr

Page 1: Real time  Pcr

1

REAL TIME PCR

Page 2: Real time  Pcr

2

LIMITATIONS OF END-POINT PCR• Poor Precision• Low sensitivity• Low resolution• Non - Automated• Size-based discrimination only• Results are not expressed as numbers• Ethidium bromide for staining is not very

quantitative• Post PCR processing

Page 3: Real time  Pcr

3

THE EVOLUTION OF PCR TO REAL-TIME

Traditional PCRs use Agarose gels for detection of amplification at the final phase or end-point of the PCR reaction

Real time PCR has advanced to allow detection before the end-point of the reaction.

Page 4: Real time  Pcr

4

REAL-TIME PCR DETECTIONWhile the reaction is occurring

monitors the fluorescence emitted during the reaction as an indicator of amplicon production at each PCR cycle

Real-Time chemistries allow for the detection of PCR amplification during the early phases of the reaction

Page 5: Real time  Pcr

5

Page 6: Real time  Pcr

6

PCR PHASES• Exponential:

• Exact doubling of product is accumulating at every cycle (assuming 100% reaction efficiency). The reaction is very specific and precise.

• Linear (High Variability):• The reaction components are being consumed, the reaction is slowing,

and products are starting to degrade.• Plateau (End-Point: Gel detection for traditional

methods):• The reaction has stopped, no more products are being made and if left

long enough, the PCR products will begin to degrade.

Page 7: Real time  Pcr

7

Page 8: Real time  Pcr

8

The plateau effect on a fivefold dilution series:The 5-fold dilution series, seems to plateau at the same place

even though the exponential phase clearly shows a difference between the points along the dilution series. This reinforces the fact that if measurements were taken at the plateau phase, the data would

not truly represent the initial amounts of starting target material.

Page 9: Real time  Pcr

9

REAL-TIME PRINCIPLES Three general methods for the quantitative detection:

1. Hydrolysis probes (TaqMan, Beacons).

2. Hybridisation probes (Light Cycler).

3. DNA-binding agents (SYBR Green).

Page 10: Real time  Pcr

10

TAQMAN CHEMISTRY Real-time systems using fluorogenic-labeled

probes that use the 5´ nuclease activity of Taq DNA polymerase. Detecting only specific amplification products. Eliminate post-PCR processing for the analysis of

probe degradation

Page 11: Real time  Pcr

11

A fluorogenic probe enables the detection of a specific PCR product as it accumulates during PCR. The TaqMan® Probe is designed with a high-energy

dye termed a Reporter at the 5 end, and a low-energy molecule termed a Quencher at the 3 end.

While the probe is intact, the proximity of the quencher dye greatly reduces the fluorescence emitted by the reporter dye by fluorescence resonance energy transfer (FRET) through space.

When reporter and quencher are separated the reporter fluoroscence

Page 12: Real time  Pcr

12

HOW TAQMAN SEQUENCE DETECTION CHEMISTRY WORKS

1. If the target sequence is present, the probe anneals downstream from one of the primer sites

2. During extension the probe is cleaved by the 5´ nuclease activity of Taq DNA polymerase.

The reporter dye is separated from the quencher dye, increasing the reporter dye signal.

The probe is removed from the target strand, allowing primer extension to continue to the end of the template strand..

3. Additional reporter dye molecules are cleaved from their respective probes with each cycle.

Page 13: Real time  Pcr

13

• The fluorescence intensity is increased proportional to the amount of amplicon produced.

Page 14: Real time  Pcr

14

MOLECULAR BEACONS

Page 15: Real time  Pcr

15

HYBRIDIZATION PROBES Fluorescence Resonance Energy Transfer

(FRET) Concept Donor fluorophore is excited by appropriate

wavelength Donor energy is transferred non-radiatively to

the acceptor fluorophore Excited acceptor emits at longer wavelength

Increase in fluorescence with each cycle The signal depends only on hybridization

Page 16: Real time  Pcr

16

HOW THE HYBRIDISATION PROBES WORK?

When the two fluorochromes are in close vicinity (1–5 nucleotides apart), the emitted light of the donor fluorochrome will excite the acceptor fluorochrome (FRET).

This results in the emission of fluorescence, which subsequently can be detected during the annealing phase and first part of the extension phase of the PCR reaction.

Page 17: Real time  Pcr

17

SYBR® GREEN I DYE CHEMISTRY Small molecules that bind to

double-stranded DNA can be divided into two classes:Intercalators Minor-groove binders

There are two requirements for a DNA binding dye for real-time detection of PCR: Increased fluorescence when

bound to double-stranded DNA No inhibition of PCR

Page 18: Real time  Pcr

18

HOW THE SYBR GREEN I DYE CHEMISTRY WORKS

When SYBR Green I dye is added to a sample, it immediately binds to all double-stranded DNA present in the sample.

● During the PCR, the DNA Polymerase amplifies the target sequence, and creates the PCR products.● The SYBR Green I dye binds to each new

copy of double-stranded DNA. ● As the PCR progresses, more amplicons are

created. The result is an increase in fluorescence intensity proportionate to the amount of PCR product produced.

Page 19: Real time  Pcr

19

Advantages of SYBR Green I Dye It can be used to monitor the amplification of any

double-stranded DNA sequence. No probe is required, which reduces assay setup

and running costs. Disadvantage of SYBR Green I Dye

SYBR Green I dye chemistry may generate false positive signals SYBR Green I dye binds to nonspecific double-

stranded DNA sequences. Needs fully optimized PCR system -no primer dimers or non-specific

amplicons.When not to choose SYBR Green

Multiplex reactions.

Page 20: Real time  Pcr

20

TERMS USED IN QUANTITATION ANALYSIS

Amplification plot: The plot of fluorescence signal versus cycle number

Page 21: Real time  Pcr

21

Baseline: The initial cycles of PCR, in which there is little change in fluorescence signal

Passive reference: A dye that provides an internal reference to which the reporter dye signal can be normalized during data analysis.

Rn (normalized reporter): The fluorescence emission intensity of the reporter dye divided by the fluorescence emission intensity of the passive reference dye. Normalization is necessary to correct for fluctuations caused by

changes in concentration or volume.

Page 22: Real time  Pcr

22

Rn+: The Rn value of a reaction containing all components, including the template

Rn-:The Rn value of an un-reacted sample. The Rn-value can be obtained from: The early cycles of a real-time PCR run (those cycles prior to a detectable

increase in fluorescence), OR A reaction that does not contain any template

ΔRn (delta Rn): The magnitude of the signal generated by the given set of PCR conditions. [ΔRn = (Rn+) – (Rn-)]

Page 23: Real time  Pcr

23

Threshold: The average standard deviation of Rn for the early PCR cycles, multiplied by an adjustable factor. The threshold should be set in the region associated with an exponential growth of PCR product.

Ct (threshold cycle): The fractional cycle number at which the fluorescence passes the fixed threshold

Page 24: Real time  Pcr

24

Page 25: Real time  Pcr

25

QUANTITATIONBy Real time PCR

Page 26: Real time  Pcr

26

ABSOLUTE QUANTITATION Used to quantitate unknown samples by

interpolating their quantity from a standard curve.

Standard: A sample of known concentration used to construct a standard curve. From a standard curve you can extrapolate the

quantity of an unknown sample. Unknown: A sample containing an unknown

quantity of template. This is the sample whose quantity you want to determine

Page 27: Real time  Pcr

27

Page 28: Real time  Pcr

28

The standard curve resulting from Ct value and standard concentrations should be a linear graph with a high correlation coefficient (> 0.99).

Theoretically a single copy of the target should create a CT value of 40 (if efficiency is 100%), which is the y-intercept in a standard curve experiment

CT value of 40 or more means no amplification and cannot be included in the calculations

Page 29: Real time  Pcr

29

RELATIVE QUANTITATION A relative quantitation assay is used to

analyze changes in gene expression in a given sample relative to another reference sample (such as an untreated control sample). Relative quantitation might be used to measure

gene expression in response to a chemical (drug).

Page 30: Real time  Pcr

30

RELATIVE QUANTITATIONSTANDARD CURVE METHOD

Using a separate standard curve for the target and endogenous control, in separate tubes.

Requires the least amount of optimization and validation.

Page 31: Real time  Pcr

31

RELATIVE QUANTITATIONCOMPARATIVE CT METHOD

The need for a standard curve is eliminated. The target and endogenous control are

amplified in the same tube. Increased throughput:

wells no longer need to be used for the standard curve samples.

the adverse effect of any pipetting and dilution errors made in creating the standard curve samples are eliminated.

A validation experiment must be run to show that the efficiencies of the target and endogenous control amplifications are approximately equal.

Page 32: Real time  Pcr

32

ENDOGENOUS/INTERNAL CONTROL

Usually an abundantly and constantly expressed housekeeping gene

Best to run a validity test for the selected endogenous control

Combination may be used

Page 33: Real time  Pcr

33

UNDERSTANDING CT

CT (threshold cycle) is the intersection between an amplification curve and a threshold line. It is a relative measure of the

concentration of target in the PCR reaction.

The CT values from PCR reactions run under different conditions or with different reagents cannot be compared directly.

Page 34: Real time  Pcr

34

EFFECT OF EFFICIENCY OF A PCR REACTION ON CTThe slope of a standard curve is a

reflection of the amplification efficiencyThe efficiency of the reaction can be

calculated by the following equation:Eff=10(-1/slope) –1

The efficiency of the PCR should be 90-110% (ideal slope = 3.32)

The PCR efficiency is dependent on: the assay and the master mix performancesample quality. length of the ampliconsecondary structureprimer design

Page 35: Real time  Pcr

35

USING THE PCR EQUATIONXn = X0(1 + E)n

Xn = PCR product after cycle n X0 = initial copy number E = amplification efficiency n = cycle number

Xn

X0

cycle number

Page 36: Real time  Pcr

36

EFFECT OF AMPLIFICATION EFFICIENCYXn = X0(1+E)n

Case 1: E = 0.9 Case 2: E = 0.8

Xn = 100 (1+0.9)30 Xn = 100 (1+0.8)30

Xn = 2.3 x 1010 Xn = 4.6 x 109

ResultA difference of 0.1 in amplification

efficiencies created a five-fold difference in thefinal ratio of PCR products after 30 cycles

Page 37: Real time  Pcr

37

REAL-TIME PCR ADVANTAGES

Not influenced by non-specific amplification

Amplification can be monitored in real-time

No post-PCR processing of products (high throughput, low

contamination risk)

Ultra-rapid cycling (30 minutes to 2 hours)

Most specific, sensitive and reproducible

Not much more expensive than conventional PCR (except

equipment cost)