Аrea monitor for pulsed radiation - ATOMTEX · The monitor can be used for dosimetric control of...
Transcript of Аrea monitor for pulsed radiation - ATOMTEX · The monitor can be used for dosimetric control of...
RS 485
+12VDC +24VDC
Features
Area monitor for pulse radiation
Operating principle
Flexible and reliable monitor designed for control of pulse radiation during operation of linear electron accelerators or other pulsed action installations
The system consists of АТ1123 X-ray and gamma radiation dosimeters and components of АТ2327 Alarm dosimeter
Purpose
The monitor is structurally divided into measuring channels (1 ndash 32) connected to switchboard unit Each channel comprises a measuring component (АТ1123 Dosimeter) and a connecting component (АТ941 Interface adapter and АТ960 Surge protector) fixed to a wall in close proximity to each other
Any of the channels can be equipped with AT991u light and sound alarm unit with potential-free contacts allowing to connect various actuators such as lock devices etc
Radiation level data from each channel are transmitted over cable into switchboard unit and then to a PC-based operator panel Individual control channel can be mounted inside control room for safety reasons
The monitor operation is controlled by a dedicated SARK2 PXR software allowing reading and analysing measurement data as well as setting and changing configuration Operator panel displays monitored site plan with measurement result values at control points presented in the form of graphs and tables
If one of measuring channels or communication lines fails the rest continue operation in stand-alone mode preserving all alarm measurement and data storage functions
Plastic tissue-equivalent detector
Measurement of pulse radiation from 10 ns
Independent measurement for each channel in wide range of dose rates and energies
Sound and light alarm for each measurement channel in case threshold levels are exceeded
Fault diagnostics
History logging of dose rate levels and cases of threshold crossing
Software for displaying current radiation environment in monitored site on PC screen
Backup power supply unit
The monitor can be used for dosimetric control of continuous gamma radiation
AT1123 dosimeter can be operated in off-line mode as a portable instrument
Computer work station with SARK2 PXR application software
АТ991у
RS 485+24VDC
RS 485+24VDC
RS 485+24VDC Switchboard unit
RS 485 USB adapter
24VDC power adapter
Rechargeable battery
USB
Measuring channel 1
+12VDC
RS 232
RS 485
Potential-free contacts
110-230 VAC50-60 Hz
Measuring channel 2
RS 232
АТ991уPotential-free
contacts
Measuring channel n
+12VDC
RS 232
RS 485
АТ991уPotential-free
contacts
+24VDC
+24VDC
АТ960
АТ941
АТ941
АТ960
АТ941
АТ960
INSTRUMENTS AND TECHNOLOGIES FOR NUCLEAR
MEASUREMENTS AND RADIATION MONITORING
Independent measurement in wide range of dose rates and energies for each channel
Extremely long burn-up lifetime ndash at least 50000 Sv for UDKG-37
Sound and light alarm to indicate exceeded threshold levels for each measuring channel
Fault diagnostics
Logging data on dose rate values and exceeded threshold levels
Software for displaying current radiation environment at controlled site on the PC screen
Backup power supply unit
The monitor can be used for dosimetric control of continuous gamma radiation
The AT1123 dosimeter can be used separately as a portable device
Flexible and reliable system for restricted area beamline radiation control at linear accelerators (LINACs) and other pulse radiation facilities Measurement point may be either in the operators room or directly at the LINAC or facility location
IU-37RS 485
+24VDC
Автоматизированное рабочееместо оператора с прикладным
программным обеспечениемSARK2 PXR
Measuring channel 1 АТ991u
USB
Control unit
RS 485 USB adapter
24VDC power adapter
Rechargeable battery230 В 50 Гц
RS 485+24VDC
Measuring channel n
+12VDC
RS 232
RS 485
АТ991u
+24VDC
АТ941
АТ960
UDKG-37
RS 485+24VDC
IU-37RS 485
+24VDC
Measuring channel 2 АТ991uPotential-free contacts
UDKG-37
RS 485+24VDC
Features
UDKG-37 Detection Device
UDKG-37Detection Device
AТ1123 Dosimeter with AT2327 components
The system is based on UDKG-37 Gamma radiation detection devices AT1123 X-ray and gamma radiation dosimeters and components of the AT2327 Alarm dosimeter The UDKG-37 detection device consists of the BDKG-37 Gamma radiation detection unit and the IU-37 Interface unit Each measuring channel supports connection of an optional electronic information display
Potential-free contacts
Potential-free contacts
Data Display
АТ1123 X-ray and Gamma
Radiation Dosimeter
UDKG-37 Gamma Radiation Detection Devices
Area monitor for pulse radiation
The monitor is structurally divided into measuring channels (1 ndash 32) connected to a switchboard unit Each channel comprises a measuring part (АТ1123 Dosimeter or UDKG-37 Detection device) and a connecting part (АТ941 Interface adapter and АТ960 Surge protector) fixed to a wall in close proximity to each other
Any of the channels can be equipped with a light and sound alarm unit AT991u with potential-free contacts to allow connection to various actuators such as locks etc
Information from each channel about the radiation level is transmitted via cable into the switchboard unit and then to the PC-based operator panel An individual control channel can be mounted inside control room for safety reasons
The ldquoSARK2 PXR software is available to provide the monitor management and to allow reading and analyzing the measurement data as well as configuring and adjusting system settings Operator panel displays graphic representation of monitored site with measurement values at control points in the form of charts and tables
If a measuring channel or communication line fails the rest will continue operation in stand-alone mode maintaining all alarm measurement and data storage functions
OPERATING PRINCIPLE
ldquoSARK2 PXRrdquo SOFTWARE
Functions
Adjust configuration
Trigger alarm in case of exceeded preset for gamma radiation dose and dose rate threshold
System diagnostics
Log measurement results
Purpose The ldquoSARK2 PXRrdquo program collects and processes dosimetric information received from UDKG-37 Gamma radiation detection devices and
AT1123 X-ray and gamma radiation dosimeters and triggers an alarm when the gamma radiation dose rate threshold or accumulated dose rate threshold has been exceeded
Energy dependence in 10 20 MeV range is based on Monte Carlo method and is for reference only
Area monitor for pulse radiation
5 Gikalo st Minsk 220005 Republic of BelarusTelFax +375-17-270-81-42E-mail infoatomtexcom
Alarm 3-stage light alarm and sound alarm
Maximum distance between UDKG-37 AT1123 dosimeterand operator panel
1000 m
Initialisation time le5 min
Power supplya) 110-230 VAC 50-60 Hzb) Rechargeable battery for emergency power supply
Continuous run timeа) ge24 h with AC mains power supply
b) ge6 h with self-contained supply from fully charged battery
Интерфейс
- UDKG-37 Detection Device- AT1123 Dosimeter- АТ991u Alarm Unit- Data Display
RS 485RS 232RS 485RS 485
Operation temperature range
о о-30 С to +50 С-40degС to +60degС (UDKG-37)
-5degС to +40degС (Data Display)
Relative humidity оle95 (with air temperature le35 С without condensation)
Protection class
- Detection Unit BDKG-37- Interface Unit IU-37- AT1123 Dosimeter- АТ941 Interface Adapter- АТ960 Surge Protector- АТ991u Alarm Unit- Data Display- When АТ1123 АТ941 АТ960 are located inside cabinet
IР68IР65IР54IР65IР65IР65IР21IР66
Overall dimensions weight
- Detection Unit BDKG-37- Interface Unit IU-37- AT1123 Dosimeter- АТ941 Interface Adapter- АТ960 Surge Protector- АТ991u Alarm Unit- Data Display
Oslash30х130 mm 025 kg170х80х55 mm 03 kg233x85x67 mm 09 kg206х82х56 mm 04 kg206х82х56 mm 04 kg185х85х95 mm 04 kg644х98х67 mm 40 kg
Measuring channel UDKG-37 АТ1123
DetectorSilicon semiconductor detector
Geiger-Muller counter tubeScintillation tissue-equivalent plastic
Oslash30x15 mm
Measurement range of average ambient dose equivalent rate of pulse radiation
30 microSvs ndash 03 Svs (100 mSvh ndash 1000 Svh)
(pulse repetition rate is not less than 20 cps
duration not less than 1 μs)
30 pSvs ndash 3 mSvs(01 microSvh ndash 10 Svh)
(pulse repetition rateis not less than 10 cps
duration not less than 10 ns)
Measurement range of ambient dose equivalent rate of continuous radiation
1 microSvh ndash 5000 Svh 50 ndash 10 nSvh Svh
Measurement range of ambient dose equivalent of continuous and pulse radiation
ndash 10 nSv ndash 10 Sv
Energy rangepulse radiation continuous radiation
50 keV ndash 10 MeV50 keV ndash 10 MeV
15 ndash 10 keV MeV15 ndash keV 10 MeV
137Typical sensitivity to Cs gamma radiation -1015 cps(microSvh ) for dose rate le01 Svh
-158 mV(Svmiddoth ) for dose rate gt01 Svh70 -1cps(microSvh )
Energy dependence relative 137to 662 keV ( Cs)
plusmn30 (50 keV ndash 10 MeV)plusmn35 (15 ndash 60 )keV keVplusmn25 (60 ndash 10 )keV MeV
-50 max 10 MeV 2 MeV ( ndash 0 )
Limit of intrinsic relative measurement error plusmn25 for dose rate le10 microSvhplusmn15 for dose rate gt10 microSvh
plusmn30 (pulse radiation)plusmn15 (continuous radiation)
Response time for 10-fold dose rate change le10 s for dose rate gt10 microSvh le2 s for dose rate ge01 microSvh
Burn-up life ge50000 Sv ge100 Sv
Design and specifications are subject to change without notice
- Страница 1
- Страница 2
- Страница 3
-
Area monitor for pulse radiation
The monitor is structurally divided into measuring channels (1 ndash 32) connected to a switchboard unit Each channel comprises a measuring part (АТ1123 Dosimeter or UDKG-37 Detection device) and a connecting part (АТ941 Interface adapter and АТ960 Surge protector) fixed to a wall in close proximity to each other
Any of the channels can be equipped with a light and sound alarm unit AT991u with potential-free contacts to allow connection to various actuators such as locks etc
Information from each channel about the radiation level is transmitted via cable into the switchboard unit and then to the PC-based operator panel An individual control channel can be mounted inside control room for safety reasons
The ldquoSARK2 PXR software is available to provide the monitor management and to allow reading and analyzing the measurement data as well as configuring and adjusting system settings Operator panel displays graphic representation of monitored site with measurement values at control points in the form of charts and tables
If a measuring channel or communication line fails the rest will continue operation in stand-alone mode maintaining all alarm measurement and data storage functions
OPERATING PRINCIPLE
ldquoSARK2 PXRrdquo SOFTWARE
Functions
Adjust configuration
Trigger alarm in case of exceeded preset for gamma radiation dose and dose rate threshold
System diagnostics
Log measurement results
Purpose The ldquoSARK2 PXRrdquo program collects and processes dosimetric information received from UDKG-37 Gamma radiation detection devices and
AT1123 X-ray and gamma radiation dosimeters and triggers an alarm when the gamma radiation dose rate threshold or accumulated dose rate threshold has been exceeded
Energy dependence in 10 20 MeV range is based on Monte Carlo method and is for reference only
Area monitor for pulse radiation
5 Gikalo st Minsk 220005 Republic of BelarusTelFax +375-17-270-81-42E-mail infoatomtexcom
Alarm 3-stage light alarm and sound alarm
Maximum distance between UDKG-37 AT1123 dosimeterand operator panel
1000 m
Initialisation time le5 min
Power supplya) 110-230 VAC 50-60 Hzb) Rechargeable battery for emergency power supply
Continuous run timeа) ge24 h with AC mains power supply
b) ge6 h with self-contained supply from fully charged battery
Интерфейс
- UDKG-37 Detection Device- AT1123 Dosimeter- АТ991u Alarm Unit- Data Display
RS 485RS 232RS 485RS 485
Operation temperature range
о о-30 С to +50 С-40degС to +60degС (UDKG-37)
-5degС to +40degС (Data Display)
Relative humidity оle95 (with air temperature le35 С without condensation)
Protection class
- Detection Unit BDKG-37- Interface Unit IU-37- AT1123 Dosimeter- АТ941 Interface Adapter- АТ960 Surge Protector- АТ991u Alarm Unit- Data Display- When АТ1123 АТ941 АТ960 are located inside cabinet
IР68IР65IР54IР65IР65IР65IР21IР66
Overall dimensions weight
- Detection Unit BDKG-37- Interface Unit IU-37- AT1123 Dosimeter- АТ941 Interface Adapter- АТ960 Surge Protector- АТ991u Alarm Unit- Data Display
Oslash30х130 mm 025 kg170х80х55 mm 03 kg233x85x67 mm 09 kg206х82х56 mm 04 kg206х82х56 mm 04 kg185х85х95 mm 04 kg644х98х67 mm 40 kg
Measuring channel UDKG-37 АТ1123
DetectorSilicon semiconductor detector
Geiger-Muller counter tubeScintillation tissue-equivalent plastic
Oslash30x15 mm
Measurement range of average ambient dose equivalent rate of pulse radiation
30 microSvs ndash 03 Svs (100 mSvh ndash 1000 Svh)
(pulse repetition rate is not less than 20 cps
duration not less than 1 μs)
30 pSvs ndash 3 mSvs(01 microSvh ndash 10 Svh)
(pulse repetition rateis not less than 10 cps
duration not less than 10 ns)
Measurement range of ambient dose equivalent rate of continuous radiation
1 microSvh ndash 5000 Svh 50 ndash 10 nSvh Svh
Measurement range of ambient dose equivalent of continuous and pulse radiation
ndash 10 nSv ndash 10 Sv
Energy rangepulse radiation continuous radiation
50 keV ndash 10 MeV50 keV ndash 10 MeV
15 ndash 10 keV MeV15 ndash keV 10 MeV
137Typical sensitivity to Cs gamma radiation -1015 cps(microSvh ) for dose rate le01 Svh
-158 mV(Svmiddoth ) for dose rate gt01 Svh70 -1cps(microSvh )
Energy dependence relative 137to 662 keV ( Cs)
plusmn30 (50 keV ndash 10 MeV)plusmn35 (15 ndash 60 )keV keVplusmn25 (60 ndash 10 )keV MeV
-50 max 10 MeV 2 MeV ( ndash 0 )
Limit of intrinsic relative measurement error plusmn25 for dose rate le10 microSvhplusmn15 for dose rate gt10 microSvh
plusmn30 (pulse radiation)plusmn15 (continuous radiation)
Response time for 10-fold dose rate change le10 s for dose rate gt10 microSvh le2 s for dose rate ge01 microSvh
Burn-up life ge50000 Sv ge100 Sv
Design and specifications are subject to change without notice
- Страница 1
- Страница 2
- Страница 3
-
Energy dependence in 10 20 MeV range is based on Monte Carlo method and is for reference only
Area monitor for pulse radiation
5 Gikalo st Minsk 220005 Republic of BelarusTelFax +375-17-270-81-42E-mail infoatomtexcom
Alarm 3-stage light alarm and sound alarm
Maximum distance between UDKG-37 AT1123 dosimeterand operator panel
1000 m
Initialisation time le5 min
Power supplya) 110-230 VAC 50-60 Hzb) Rechargeable battery for emergency power supply
Continuous run timeа) ge24 h with AC mains power supply
b) ge6 h with self-contained supply from fully charged battery
Интерфейс
- UDKG-37 Detection Device- AT1123 Dosimeter- АТ991u Alarm Unit- Data Display
RS 485RS 232RS 485RS 485
Operation temperature range
о о-30 С to +50 С-40degС to +60degС (UDKG-37)
-5degС to +40degС (Data Display)
Relative humidity оle95 (with air temperature le35 С without condensation)
Protection class
- Detection Unit BDKG-37- Interface Unit IU-37- AT1123 Dosimeter- АТ941 Interface Adapter- АТ960 Surge Protector- АТ991u Alarm Unit- Data Display- When АТ1123 АТ941 АТ960 are located inside cabinet
IР68IР65IР54IР65IР65IР65IР21IР66
Overall dimensions weight
- Detection Unit BDKG-37- Interface Unit IU-37- AT1123 Dosimeter- АТ941 Interface Adapter- АТ960 Surge Protector- АТ991u Alarm Unit- Data Display
Oslash30х130 mm 025 kg170х80х55 mm 03 kg233x85x67 mm 09 kg206х82х56 mm 04 kg206х82х56 mm 04 kg185х85х95 mm 04 kg644х98х67 mm 40 kg
Measuring channel UDKG-37 АТ1123
DetectorSilicon semiconductor detector
Geiger-Muller counter tubeScintillation tissue-equivalent plastic
Oslash30x15 mm
Measurement range of average ambient dose equivalent rate of pulse radiation
30 microSvs ndash 03 Svs (100 mSvh ndash 1000 Svh)
(pulse repetition rate is not less than 20 cps
duration not less than 1 μs)
30 pSvs ndash 3 mSvs(01 microSvh ndash 10 Svh)
(pulse repetition rateis not less than 10 cps
duration not less than 10 ns)
Measurement range of ambient dose equivalent rate of continuous radiation
1 microSvh ndash 5000 Svh 50 ndash 10 nSvh Svh
Measurement range of ambient dose equivalent of continuous and pulse radiation
ndash 10 nSv ndash 10 Sv
Energy rangepulse radiation continuous radiation
50 keV ndash 10 MeV50 keV ndash 10 MeV
15 ndash 10 keV MeV15 ndash keV 10 MeV
137Typical sensitivity to Cs gamma radiation -1015 cps(microSvh ) for dose rate le01 Svh
-158 mV(Svmiddoth ) for dose rate gt01 Svh70 -1cps(microSvh )
Energy dependence relative 137to 662 keV ( Cs)
plusmn30 (50 keV ndash 10 MeV)plusmn35 (15 ndash 60 )keV keVplusmn25 (60 ndash 10 )keV MeV
-50 max 10 MeV 2 MeV ( ndash 0 )
Limit of intrinsic relative measurement error plusmn25 for dose rate le10 microSvhplusmn15 for dose rate gt10 microSvh
plusmn30 (pulse radiation)plusmn15 (continuous radiation)
Response time for 10-fold dose rate change le10 s for dose rate gt10 microSvh le2 s for dose rate ge01 microSvh
Burn-up life ge50000 Sv ge100 Sv
Design and specifications are subject to change without notice