L 9: X Ray production

Page 3: Raiobiology 9

IAEA 6: X Ray production 3

Introduction

A review is made of:

• The main elements of the of X Rays tube: cathode and anode structure

• The technology constraints of the anode and cathode material

• The rating charts and X Ray tube heat loading capacities

Page 4: Raiobiology 9

IAEA 6: X Ray production 4

Topics

• Basic elements of an X Ray source assembly

• Cathode structure

• Anode structure

• Rating chart

• X Ray generator

• Automatic exposure control

Page 5: Raiobiology 9

IAEA 6: X Ray production 5

Overview

• To become familiar with the technological principles of the X Ray production

Page 6: Raiobiology 9

IAEAInternational Atomic Energy Agency

Part 6: X Ray production

Topic 1: Basic elements of an X Ray source assembly

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

Page 7: Raiobiology 9

IAEA 6: X Ray production 7

Basic elements of the X Ray source assembly

• Generator : power circuit supplying the required potential to the X Ray tube

• X Ray tube and collimator: device producing the X Ray beam

Page 8: Raiobiology 9

IAEA 6: X Ray production 8

X Ray tubes

Page 9: Raiobiology 9

IAEA 6: X Ray production 9

X Ray tube components

• Cathode: heated filament which is the source of the electron beam directed towards the anode• tungsten filament

• Anode (stationary or rotating): impacted by electrons, emits X Rays

• Metal tube housing surrounding glass (or metal) X Ray tube (electrons are traveling in vacuum)

• Shielding material (protection against scattered radiation)

Page 10: Raiobiology 9

IAEA 6: X Ray production 10

X Ray tube components

1: long tungsten filament2 : short tungsten filament3 : real size cathode

1: mark of focal spot

housing cathode

Page 11: Raiobiology 9

IAEAInternational Atomic Energy Agency

Part 6: X Ray production

Topic 2: Cathode structure

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

Page 12: Raiobiology 9

IAEA 6: X Ray production 12

Cathode structure (I)

• Cathode includes filament(s) and associated circuitry

• tungsten material : preferred because of its high melting point (3370°C)

• slow filament evaporation

• no arcing

• minimum deposit of W on glass envelope

• To reduce evaporation the emission temperature of the cathode is reached just before the exposure

• in stand-by, temperature is kept at ± 1500°C so that 2700°C emission temperature can be reached within a second

Page 13: Raiobiology 9

IAEA 6: X Ray production 13

Example of a cathode

Page 14: Raiobiology 9

IAEA 6: X Ray production 14

• Modern tubes have two filaments • a long one : higher current/lower resolution

• a short one : lower current/higher resolution

• Coulomb interaction makes the electron beam divergent on the travel to the anode• lack of electrons producing X Rays

• larger area of target used

• focal spot increased lower image resolution

Focalisation of electrons is crucial !

Cathode structure (I)

Page 15: Raiobiology 9

IAEAInternational Atomic Energy Agency

Part 6: X Ray production

Topic 3: Anode structure

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

Page 16: Raiobiology 9

IAEA 6: X Ray production 16

X Ray tube characteristics

• Anode mechanical constraints• Material : tungsten, rhenium, molybdenum, graphite• Focal spot : surface of anode impacted by electrons• Anode angle• Disk and annular track diameter (rotation frequency

from 3,000 to 10,000 revolutions/minute)• Thickness mass and material (volume) heat

capacity

• Anode thermal constraints• Instantaneous power load (heat unit)• Heat loading time curve• Cooling time curve

Page 17: Raiobiology 9

IAEA 6: X Ray production 17

Anode angle (I)

• The Line-Focus principle• Anode target plate has a shape that is more

rectangular or ellipsoidal than circular • the shape depends on :

• filament size and shape

• focusing cup’s and potential

• distance between cathode and anode

• Image resolution requires a small focal spot

• Heat dissipation requires a large spot

• This conflict is solved by slanting the target face

Page 18: Raiobiology 9

IAEA 6: X Ray production 18

Anode characteristic

1 : anode track2 : anode track

Page 19: Raiobiology 9

IAEA 6: X Ray production 19

THE SMALLER THE ANGLETHE SMALLER THE ANGLETHE BETTER THE RESOLUTIONTHE BETTER THE RESOLUTION

Anode angle (II)

Angle

Incident electron beam width

Apparent focal spot size

Actual focal spot size

Film

Angle

Incident electron beam width

Increased apparent

focal spot size

Actual focal spot size

Film

‘ 

Page 20: Raiobiology 9

IAEA 6: X Ray production 20

Anode heel effect (I)

• Anode angle (from 7° to 20°) induces a variation of the X Ray output in the plane comprising the anode-cathode axis

• Absorption by anode of X photons with low emission angle

• The magnitude of influence of the heel effect on the image depends on factors such as :

• anode angle• size of film• focus to film distance

• Anode aging increases heel effect

Page 21: Raiobiology 9

IAEA 6: X Ray production 21

• The heel effect is not always a negative factor

• It can be used to compensate for different attenuation through parts of the body

• For example:• thoracic spine (thicker part of the patient

towards the cathode side of the tube) • mammography

Anode heel effect (II)

Page 22: Raiobiology 9

IAEA 6: X Ray production 22

Focal spot size and imaging geometry

• Focal spot finite size image unsharpened

• Improving sharpness small focal spot size

• For mammography focal spot size 0.4 mm nominal

• Small focal spot size reduced tube output (longer

exposure time)

• Large focal spot allows high output (shorter exposure time)

• Balance depends on organ movement (fast moving organs

may require larger focus)

Page 23: Raiobiology 9

IAEAInternational Atomic Energy Agency

Part 6: X Ray production

Topic 4: Rating Chart

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

Page 24: Raiobiology 9

IAEA 6: X Ray production 24

Heat loading capacities

• A procedure generates an amount of heat depending on: • kV used, tube current (mA), length of exposure

• type of voltage waveform

• number of exposures taken in rapid sequence

• Heat Unit (HU) [joule] :unit of potential x unit of tube current x unit of time

• The heat generated by various types of X Ray circuits are:

• 1 phase units : HU = kV x mA x s• 3 phase units, 6 pulse : HU = 1.35 kV x mA x s• 3 phase units, 12 pulse: HU = 1.41 kV x mA x s

Page 25: Raiobiology 9

IAEA 6: X Ray production 25

X Ray tube rating chart (I)

• Tube cooling characteristics and focal spot size {mA - time} relationship at constant kV

• intensity decreases with increasing exposure time

• intensity increases with decreasing kV

• Note: higher power reduced exposure time reduced motion unsharpness

Page 26: Raiobiology 9

IAEA 6: X Ray production 26

• Manufacturers combine heat loading characteristics and information about the limits of their X Ray tubes in graphical representations called Tube Rating Charts

• Example:• Tube A: a 300 mA, 0.5 s, 90 kV procedure would damage

the system operated from a 1-phase half wave rectified generator (unacceptable)

• Tube B: a 200 mA, 0.1 s, 120 kV procedure comply with the technical characteristics of the system operated from a 3-phase fully rectified generator (acceptable)

X Ray tube rating chart (II)

Page 27: Raiobiology 9

IAEA 6: X Ray production 27

0.01 0.05 0.1 0.5 1.0 5.0 10.0

700

600

500

400

300

200

100

50 kVp

70 kVp

90 kVp120 kVp

Unacceptable

Exposure time (s)

Tu

be

curr

ent

(mA

)X Ray tube A

half-wave rectified3000 rpm 90 kV

1.0 mm effective focal spot

X Ray tube rating chart (III)

Page 28: Raiobiology 9

IAEA 6: X Ray production 28

0.01 0.05 0.1 0.5 1.0 5.0 10.0

700

600

500

400

300

200

100

50 kVp

70 kVp90 kVp125 kVp

Acceptable

Exposure time (s)

Tu

be

curr

ent

(mA

)

X Ray tube B3 full-wave rectified

10.000 rpm 125 kV1.0 mm effective focal spot

X Ray tube rating chart (IV)

Unacceptable

Page 29: Raiobiology 9

IAEA 6: X Ray production 29

Anode cooling chart (I)

• Heat generated is stored in the anode, and dissipated through the cooling circuit

• A typical cooling chart has:• input curves (heat units stored as a function of time)

• anode cooling curve

• The following graph shows that:• a procedure delivering 500 HU/s can go on

indefinitely

• if it is delivering 1000 HU/s it has to stop after 10 min

• if the anode has stored 120.000 HU, it will take 5 min to cool down completely

Page 30: Raiobiology 9

IAEA 6: X Ray production 30

240

220

200

180

160

140

120

100

80

60

40

20

Elapsed time (min)

He

at

un

its

sto

red

(x

10

00

)

1 2 3 4 5 6 7 8 9 10 11 12 13 14

500 HU/sec1000 HU/sec

350 HU/sec

250 HU/sec

Imput curve

Cooling curve

Maximum Heat Storage Capacity of Anode

Anode cooling chart (II)

Page 31: Raiobiology 9

IAEAInternational Atomic Energy Agency

Part 6: X Ray production

Topic 5: X Ray generator

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

Page 32: Raiobiology 9

IAEA 6: X Ray production 32

X-ray generator (I)

It supplies the X-ray tube with : Current to heat the cathode filament

Potential to accelerate electrons

Automatic control of exposure (power application time)

Energy supply 1000 X-ray beam energy (of which 99.9% is

dissipated as thermal energy)

Page 33: Raiobiology 9

IAEA 6: X Ray production 33

• Generator characteristics have a strong influence on the contrast and sharpness of the radiographic image

• The motion unsharpness can be greatly reduced by a generator allowing an exposure time as short as achievable

• Since the dose at the image plane can be expressed as:

D = k0 . Un . I . T• U: peak voltage (kV)• I: mean current (mA)• T: exposure time (ms)• n: ranging from about 1.5 to 3

X-ray generator (II)

Page 34: Raiobiology 9

IAEA 6: X Ray production 34

• Peak voltage value has an influence on the beam hardness

• It has to be related to medical question• What is the anatomical structure to investigate ?• What is the contrast level needed ?• For a thorax examination : 140 - 150 kV is suitable to

visualize the lung structure • While only 65 kV is necessary to see bone structure

• The ripple “r” of a generator has to be as low as possible

r = [(U - Umin)/U] x 100%

X-ray generator (III)

Page 35: Raiobiology 9

IAEA 6: X Ray production 35

Tube potential wave form (I)

• Conventional generators• single 1-pulse (dental and some mobile systems)

• single 2-pulse (double rectification)

• three 6-pulse

• three 12-pulse

• Constant potential generators (CP)

• HF generators (use of DC choppers to convert 50Hz mains into voltages with frequencies in the kHz range) “Inverter technology”

Page 36: Raiobiology 9

IAEA 6: X Ray production 36

100%

13%

4%

Line voltage

Single phase single pulse

Single phase 2-pulse

Three phase 6-pulse

Three phase 12-pulse

0.02 s

0.01 s

kV ripple (%)

Tube potential wave form (II)

Page 37: Raiobiology 9

IAEA 6: X Ray production 37

The choice of the number of pulses (I)

• Single pulse : low power (<2 kW)

• 2-pulse : low and medium power

• 6-pulse : uses 3-phase mains, medium and high power (manual or automatic compensation for voltage drop)

• 12-pulse : uses two shifted 3-phase system, high power up to 150 kW

Page 38: Raiobiology 9

IAEA 6: X Ray production 38

• CP : eliminates any changes of voltage or tube current • high voltage regulators can control the voltage

AND switch on and off the exposure• voltage can be switched on at any moment

(temporal resolution)• kV ripple <2% thus providing low patient exposure

• HF : combines the advantages of constant potential and conventional generator• reproducibility and consistency of tube voltage • high frame rate possible

The choice of the number of pulses (II)

Page 39: Raiobiology 9

IAEAInternational Atomic Energy Agency

Part 6: X-ray production

Topic 6: Automatic Exposure Control (AEC)

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

Page 40: Raiobiology 9

IAEA 6: X Ray production 40

Automatic exposure control

• Optimal choice of technical parameters in order to avoid repeated exposures (kV, mA)

• Radiation detector behind (or in front of) the film cassette (with due correction)

• Exposure is terminated when the required dose has been integrated

• Compensation for kVp at a given thickness

• Compensation for thickness at a given kVp

Page 41: Raiobiology 9

IAEA 6: X Ray production 41

Automatic exposure control

X Ray tube

Collimator

Beam

Soft tissue

BoneAir

Patient

Table

Grid

Cassette

AEC detectors

Page 42: Raiobiology 9

IAEA 6: X Ray production 42

Automatic exposure control

• Optimal choice of technical parameters in order to avoid repeated exposures (kV, mA)

• Radiation detector behind (or in front of) the film cassette (with due correction)

• Exposure is terminated when the required dose has been integrated

• Compensation for kVp at a given thickness

• Compensation for thickness at a given kVp

Page 43: Raiobiology 9

IAEAInternational Atomic Energy Agency

Part 6: X-ray production

Topic 7: X-ray equipment operation and mode

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

Page 44: Raiobiology 9

IAEA 6: X Ray production 44

X-ray equipment operation mode and application (II)

Radiography and Tomography• Single and 3 generators (inverter technology)

• output : 30 kW at 0.3 focus spot size

• output : 50 - 70 kW at 1.0 focus spot size

• selection of kV and mAs , AEC

Radiography and Fluoroscopy• Under couch equipment, three generator (inverter

technology) - continuous output of 300 - 500 W• output : 50 kW at 1.0 focus size for spot film

• output : 30 kW at 0.6 for fluoroscopy (high resolution)

• priority given to contrast

• automatic settings of kV

Page 45: Raiobiology 9

IAEA 6: X Ray production 45

X-Ray equipment operation mode and application (III)

• Radiography and Fluoroscopy• Over couch equipment, three phase generator (inverter

technology) - continuous output of at least 500 W

• output : 40 kW @ 0.6 focus size for spot film

• output : 70 kW @ 1.0 for fluoroscopy (high resolution)

• priority given to contrast

• automatic settings of kV

• Cardiac angiography • Three phase generator - continuous output 1kW

• output : 30 kW @ 0.4 focus size

• output : 80 kW @ 0.8 focus size

• frame rate : up to 120 fr/s

Page 46: Raiobiology 9

IAEA 6: X Ray production 46

Summary

• The main parts of the system contributing to

the desired X Ray production:

• provide the required source of power

• deliver an appropriate X Ray spectrum

• ensure the optimum adjustment of exposure to

warrant the image quality

Page 47: Raiobiology 9

IAEA 6: X Ray production 47

Where to Get More Information

• Equipment for diagnostic radiology, E. Forster, MTP Press, 1993

• IPSM Report 32, part 1, X-ray tubes and generators

• The Essential Physics of Medical Imaging, Williams and Wilkins. Baltimore:1994

• Manufacturers data sets for different X Ray tubes